06.07.2018

Comparison of image noise and image quality between full-dose abdominal computed tomography scans reconstructed with weighted filtered back projection and half-dose scans reconstructed with improved sinogram-affirmed iterative reconstruction (SAFIRE*)

verfasst von: Stephen Choy, Dennis Parhar, Kevin Lian, Heiko Schmiedeskamp, Luck Louis, Timothy O’Connell, Patrick McLaughlin, Savvas Nicolaou

Erschienen in: Abdominal Radiology | Ausgabe 1/2019

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To retrospectively compare the image noise, signal-to-noise ratio (SNR), and subjective image quality between CT images acquired with a dual-source, split-dose imaging protocol reconstructed at full and half doses with weighted filtered back projection (wFBP) and an improved sinogram-affirmed iterative reconstruction algorithm (SAFIRE*).

Methods

Fifty-three consecutive patients underwent contrast-enhanced CT of the abdomen using a standardized dual-source, single energy CT protocol. Half-dose images were retrospectively generated using data from one detector only. Full-dose datasets were reconstructed with wFBP, while half-dose datasets were reconstructed with wFBP and SAFIRE* strengths 1–5. Region of interest analysis was performed to assess SNR and noise. Diagnostic acceptability, subjective noise, and spatial resolution were graded on a 10-point scale by two readers. Statistical analysis was carried out with repeated measures analysis of variance, Wilcoxon signed rank test, and Cohen’s κ test.

Results

With the increasing strengths of SAFIRE*, a progressive reduction in noise and increase in SNR (p < 0.01) was observed. There was a statistically significant decrease in objective noise and increase in SNR in half-dose SAFIRE* strength 4 and 5 reconstructions compared to full-dose reconstructions using wFBP (p < 0.01). Qualitative analysis revealed a progressive increase in diagnostic acceptability, decrease in subjective noise and increase in spatial resolution for half-dose images reconstructed with the increasing strengths of SAFIRE* (p < 0.01).

Conclusions

Half-dose CT images reconstructed with SAFIRE* at strength 4 and 5 have superior image quality compared to full-dose images reconstructed with wFBP. SAFIRE* potentially allows dose reductions in the order of 50% over wFBP.
Literatur
1.
Zurück zum Zitat Mettler FA, Thomadsen BR, Bhargavan M, et al. (2008) Medical radiation exposure in the U.S. in 2006: preliminary results. Health Phys 95(5):502–507CrossRefPubMed Mettler FA, Thomadsen BR, Bhargavan M, et al. (2008) Medical radiation exposure in the U.S. in 2006: preliminary results. Health Phys 95(5):502–507CrossRefPubMed
2.
Zurück zum Zitat Gordic S, Desbiolles L, Stolzmann P, et al. (2014) Advanced modelled iterative reconstruction for abdominal CT: qualitative and quantitative evaluation. Clin Radiol 69(12):e497–e504CrossRefPubMed Gordic S, Desbiolles L, Stolzmann P, et al. (2014) Advanced modelled iterative reconstruction for abdominal CT: qualitative and quantitative evaluation. Clin Radiol 69(12):e497–e504CrossRefPubMed
3.
Zurück zum Zitat Marin D, Nelson RC, Schindera ST, et al. (2010) Low-tube-voltage, high-tube-current multidetector abdominal CT: improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm–initial clinical experience. Radiology 254(1):145–153CrossRefPubMed Marin D, Nelson RC, Schindera ST, et al. (2010) Low-tube-voltage, high-tube-current multidetector abdominal CT: improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm–initial clinical experience. Radiology 254(1):145–153CrossRefPubMed
4.
Zurück zum Zitat Kordolaimi SD, Argentos S, Pantos I, Kelekis NL, Efstathopoulos EP (2013) A new era in computed tomographic dose optimization: the impact of iterative reconstruction on image quality and radiation dose. J Comput Assist Tomogr 37(6):924–931CrossRefPubMed Kordolaimi SD, Argentos S, Pantos I, Kelekis NL, Efstathopoulos EP (2013) A new era in computed tomographic dose optimization: the impact of iterative reconstruction on image quality and radiation dose. J Comput Assist Tomogr 37(6):924–931CrossRefPubMed
5.
Zurück zum Zitat Martinsen ACT, Sæther HK, Hol PK, Olsen DR, Skaane P (2012) Iterative reconstruction reduces abdominal CT dose. Eur J Radiol 81(7):1483–1487CrossRefPubMed Martinsen ACT, Sæther HK, Hol PK, Olsen DR, Skaane P (2012) Iterative reconstruction reduces abdominal CT dose. Eur J Radiol 81(7):1483–1487CrossRefPubMed
6.
Zurück zum Zitat Schabel C, Fenchel M, Schmidt B, et al. (2013) Clinical evaluation and potential radiation dose reduction of the novel sinogram-affirmed iterative reconstruction technique (SAFIRE) in abdominal computed tomography angiography. Acad Radiol 20(2):165–172CrossRefPubMed Schabel C, Fenchel M, Schmidt B, et al. (2013) Clinical evaluation and potential radiation dose reduction of the novel sinogram-affirmed iterative reconstruction technique (SAFIRE) in abdominal computed tomography angiography. Acad Radiol 20(2):165–172CrossRefPubMed
7.
Zurück zum Zitat Leipsic J, Labounty TM, Heilbron B, et al. (2010) Adaptive statistical iterative reconstruction: assessment of image noise and image quality in coronary CT angiography. AJR Am J Roentgenol 195(3):649–654CrossRefPubMed Leipsic J, Labounty TM, Heilbron B, et al. (2010) Adaptive statistical iterative reconstruction: assessment of image noise and image quality in coronary CT angiography. AJR Am J Roentgenol 195(3):649–654CrossRefPubMed
8.
Zurück zum Zitat Vardhanabhuti V, Ilyas S, Gutteridge C, Freeman SJ, Roobottom CA (2013) Comparison of image quality between filtered back-projection and the adaptive statistical and novel model-based iterative reconstruction techniques in abdominal CT for renal calculi. Insights Imaging 4(5):661–669CrossRefPubMedPubMedCentral Vardhanabhuti V, Ilyas S, Gutteridge C, Freeman SJ, Roobottom CA (2013) Comparison of image quality between filtered back-projection and the adaptive statistical and novel model-based iterative reconstruction techniques in abdominal CT for renal calculi. Insights Imaging 4(5):661–669CrossRefPubMedPubMedCentral
9.
Zurück zum Zitat Singh S, Kalra MK, Hsieh J, et al. (2010) Abdominal CT: comparison of adaptive statistical iterative and filtered back projection reconstruction techniques. Radiology 257(2):373–383CrossRefPubMed Singh S, Kalra MK, Hsieh J, et al. (2010) Abdominal CT: comparison of adaptive statistical iterative and filtered back projection reconstruction techniques. Radiology 257(2):373–383CrossRefPubMed
10.
Zurück zum Zitat Hardie AD, Tipnis SV, Rieter WJ, Rissing MS, De Cecco CN (2013) Physician preference between low-dose computed tomography with a sinogram-affirmed iterative reconstruction algorithm and routine-dose computed tomography with filtered back projection in abdominopelvic imaging. J Comput Assist Tomogr 37(6):932–936CrossRefPubMed Hardie AD, Tipnis SV, Rieter WJ, Rissing MS, De Cecco CN (2013) Physician preference between low-dose computed tomography with a sinogram-affirmed iterative reconstruction algorithm and routine-dose computed tomography with filtered back projection in abdominopelvic imaging. J Comput Assist Tomogr 37(6):932–936CrossRefPubMed
11.
Zurück zum Zitat Kalra MK, Woisetschläger M, Dahlström N, et al. (2013) Sinogram-affirmed iterative reconstruction of low-dose chest CT: effect on image quality and radiation dose. AJR Am J Roentgenol 201(2):W235–W244CrossRefPubMed Kalra MK, Woisetschläger M, Dahlström N, et al. (2013) Sinogram-affirmed iterative reconstruction of low-dose chest CT: effect on image quality and radiation dose. AJR Am J Roentgenol 201(2):W235–W244CrossRefPubMed
12.
Zurück zum Zitat Silva AC, Lawder HJ, Hara A, Kujak J, Pavlicek W (2010) Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm. AJR Am J Roentgenol 194(1):191–199CrossRefPubMed Silva AC, Lawder HJ, Hara A, Kujak J, Pavlicek W (2010) Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm. AJR Am J Roentgenol 194(1):191–199CrossRefPubMed
13.
Zurück zum Zitat Koc G, Courtier JL, Phelps A, Marcovici PA, MacKenzie JD (2014) Computed tomography depiction of small pediatric vessels with model-based iterative reconstruction. Pediatr Radiol 44(7):787–794CrossRefPubMed Koc G, Courtier JL, Phelps A, Marcovici PA, MacKenzie JD (2014) Computed tomography depiction of small pediatric vessels with model-based iterative reconstruction. Pediatr Radiol 44(7):787–794CrossRefPubMed
14.
Zurück zum Zitat Chen C-M, Lin Y-Y, Hsu M-Y, et al. (2016) Performance of adaptive iterative dose reduction 3D integrated with automatic tube current modulation in radiation dose and image noise reduction compared with filtered-back projection for 80-kVp abdominal CT: anthropomorphic phantom and patient study. Eur J Radiol 85(9):1666–1672CrossRefPubMed Chen C-M, Lin Y-Y, Hsu M-Y, et al. (2016) Performance of adaptive iterative dose reduction 3D integrated with automatic tube current modulation in radiation dose and image noise reduction compared with filtered-back projection for 80-kVp abdominal CT: anthropomorphic phantom and patient study. Eur J Radiol 85(9):1666–1672CrossRefPubMed
15.
Zurück zum Zitat Schaller F, Sedlmair M, Raupach R, Uder M, Lell M (2016) Noise reduction in abdominal computed tomography applying iterative reconstruction (ADMIRE). Acad Radiol 23(10):1230–1238CrossRefPubMed Schaller F, Sedlmair M, Raupach R, Uder M, Lell M (2016) Noise reduction in abdominal computed tomography applying iterative reconstruction (ADMIRE). Acad Radiol 23(10):1230–1238CrossRefPubMed
16.
Zurück zum Zitat Yasaka K, Katsura M, Akahane M, et al. (2013) Model-based iterative reconstruction for reduction of radiation dose in abdominopelvic CT: comparison to adaptive statistical iterative reconstruction. SpringerPlus 2(1):209CrossRefPubMedPubMedCentral Yasaka K, Katsura M, Akahane M, et al. (2013) Model-based iterative reconstruction for reduction of radiation dose in abdominopelvic CT: comparison to adaptive statistical iterative reconstruction. SpringerPlus 2(1):209CrossRefPubMedPubMedCentral
17.
Zurück zum Zitat Park SB, Kim YS, Lee JB, Park HJ (2015) Knowledge-based iterative model reconstruction (IMR) algorithm in ultralow-dose CT for evaluation of urolithiasis: evaluation of radiation dose reduction, image quality, and diagnostic performance. Abdom Imaging 40(8):3137–3146CrossRefPubMed Park SB, Kim YS, Lee JB, Park HJ (2015) Knowledge-based iterative model reconstruction (IMR) algorithm in ultralow-dose CT for evaluation of urolithiasis: evaluation of radiation dose reduction, image quality, and diagnostic performance. Abdom Imaging 40(8):3137–3146CrossRefPubMed
18.
Zurück zum Zitat Menzel H, Schibilla H, Teunen D (2000) European guidelines on quality criteria for computed tomography. Luxembourg: European Commission Menzel H, Schibilla H, Teunen D (2000) European guidelines on quality criteria for computed tomography. Luxembourg: European Commission
19.
Zurück zum Zitat Brady SL, Yee BS, Kaufman RA (2012) Characterization of adaptive statistical iterative reconstruction algorithm for dose reduction in CT: a pediatric oncology perspective. Med Phys 39(9):5520–5531CrossRefPubMed Brady SL, Yee BS, Kaufman RA (2012) Characterization of adaptive statistical iterative reconstruction algorithm for dose reduction in CT: a pediatric oncology perspective. Med Phys 39(9):5520–5531CrossRefPubMed
20.
Zurück zum Zitat Solomon JB, Christianson O, Samei E (2012) Quantitative comparison of noise texture across CT scanners from different manufacturers. Med Phys 39(10):6048–6055CrossRefPubMed Solomon JB, Christianson O, Samei E (2012) Quantitative comparison of noise texture across CT scanners from different manufacturers. Med Phys 39(10):6048–6055CrossRefPubMed
21.
Zurück zum Zitat Burgess AE, Li X, Abbey CK (1997) Visual signal detectability with two noise components: anomalous masking effects. J Opt Soc Am A Opt Image Sci Vis 14(9):2420–2442CrossRefPubMed Burgess AE, Li X, Abbey CK (1997) Visual signal detectability with two noise components: anomalous masking effects. J Opt Soc Am A Opt Image Sci Vis 14(9):2420–2442CrossRefPubMed
22.
Zurück zum Zitat Notohamiprodjo S, Deak Z, Meurer F, et al. (2015) Image quality of iterative reconstruction in cranial CT imaging: comparison of model-based iterative reconstruction (MBIR) and adaptive statistical iterative reconstruction (ASiR). Eur Radiol 25(1):140–146CrossRefPubMed Notohamiprodjo S, Deak Z, Meurer F, et al. (2015) Image quality of iterative reconstruction in cranial CT imaging: comparison of model-based iterative reconstruction (MBIR) and adaptive statistical iterative reconstruction (ASiR). Eur Radiol 25(1):140–146CrossRefPubMed
23.
Zurück zum Zitat Shuman WP, Chan KT, Busey JM, et al. (2014) Standard and reduced radiation dose liver CT images: adaptive statistical iterative reconstruction versus model-based iterative reconstruction-comparison of findings and image quality. Radiology 273(3):793–800CrossRefPubMed Shuman WP, Chan KT, Busey JM, et al. (2014) Standard and reduced radiation dose liver CT images: adaptive statistical iterative reconstruction versus model-based iterative reconstruction-comparison of findings and image quality. Radiology 273(3):793–800CrossRefPubMed
24.
Zurück zum Zitat Deák Z, Grimm JM, Treitl M, et al. (2013) Filtered back projection, adaptive statistical iterative reconstruction, and a model-based iterative reconstruction in abdominal CT: an experimental clinical study. Radiology 266(1):197–206CrossRefPubMed Deák Z, Grimm JM, Treitl M, et al. (2013) Filtered back projection, adaptive statistical iterative reconstruction, and a model-based iterative reconstruction in abdominal CT: an experimental clinical study. Radiology 266(1):197–206CrossRefPubMed
25.
Zurück zum Zitat Volders D, Bols A, Haspeslagh M, Coenegrachts K (2013) Model-based iterative reconstruction and adaptive statistical iterative reconstruction techniques in abdominal CT: comparison of image quality in the detection of colorectal liver metastases. Radiology 269(2):469–474CrossRefPubMed Volders D, Bols A, Haspeslagh M, Coenegrachts K (2013) Model-based iterative reconstruction and adaptive statistical iterative reconstruction techniques in abdominal CT: comparison of image quality in the detection of colorectal liver metastases. Radiology 269(2):469–474CrossRefPubMed
26.
Zurück zum Zitat Boedeker KL, McNitt-Gray MF (2007) Application of the noise power spectrum in modern diagnostic MDCT: part II. Noise power spectra and signal to noise. Phys Med Biol 52(14):4047–4061CrossRefPubMed Boedeker KL, McNitt-Gray MF (2007) Application of the noise power spectrum in modern diagnostic MDCT: part II. Noise power spectra and signal to noise. Phys Med Biol 52(14):4047–4061CrossRefPubMed
27.
Zurück zum Zitat Solomon J, Mileto A, Ramirez-Giraldo JC, Samei E (2015) Diagnostic performance of an advanced modeled iterative reconstruction algorithm for low-contrast detectability with a third-generation dual-source multidetector CT scanner: potential for radiation dose reduction in a multireader study. Radiology 275(3):735–745CrossRefPubMed Solomon J, Mileto A, Ramirez-Giraldo JC, Samei E (2015) Diagnostic performance of an advanced modeled iterative reconstruction algorithm for low-contrast detectability with a third-generation dual-source multidetector CT scanner: potential for radiation dose reduction in a multireader study. Radiology 275(3):735–745CrossRefPubMed
Metadaten
Titel
Comparison of image noise and image quality between full-dose abdominal computed tomography scans reconstructed with weighted filtered back projection and half-dose scans reconstructed with improved sinogram-affirmed iterative reconstruction (SAFIRE*)
verfasst von
Stephen Choy
Dennis Parhar
Kevin Lian
Heiko Schmiedeskamp
Luck Louis
Timothy O’Connell
Patrick McLaughlin
Savvas Nicolaou
Publikationsdatum
06.07.2018
Verlag
Springer US
Erschienen in
Abdominal Radiology / Ausgabe 1/2019
Print ISSN: 2366-004X
Elektronische ISSN: 2366-0058
DOI
https://doi.org/10.1007/s00261-018-1687-9

Weitere Artikel der Ausgabe 1/2019

Abdominal Radiology 1/2019 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.