01.02.2020 | Original Article

Comparison of image quality between synthetic and patients' electrocardiogram-gated 320-row pediatric cardiac computed tomography

Erschienen in: Pediatric Radiology | Ausgabe 2/2020

Einloggen, um Zugang zu erhalten

Abstract

Background

Pediatric cardiac computed tomography (CT) can be acquired without electrode placement by using synthetic electrocardiogram (ECG).

Objective

To determine whether the depiction of gross cardiac structures and coronary arteries in 320-row pediatric CT is not inferior when CT is gated with synthetic ECG at 150 beats per minute (bpm), compared to the patients’ own ECG.

Materials and methods

Sixty 320-row CT examinations performed in children younger than 3 years old with congenital cardiac anomaly were enrolled in this retrospective study. Thirty examinations were scanned using the children’s own ECG for gating and 30 examinations were scanned using synthetic ECG at 150 bpm. The image quality was compared between the two gating modes using a 3-point scale to delineate the following anatomical structures: atrial septum, ventricular septum, right atrium, right ventricle, left atrium, left ventricle, main pulmonary artery, ascending aorta, aortic arch including the patent ductus arteriosus, descending aorta, right coronary artery and left main trunk. Beam-hardening artifacts from contrast enhancement material were evaluated using a 3-point scale, and the overall image quality was evaluated using a 5-point scale.

Results

Synthetic ECG was not inferior to the patients’ ECG in depicting each structure, beam-hardening artifact and overall image quality. Average indices were clinically acceptable imaging quality, except for subjective image quality of mid and distal coronary arteries.

Conclusion

Pediatric cardiac CT in patients younger than 3 years old can be acquired using synthetic ECG gating, with image quality not inferior to the patients’ ECG.
Literatur
1.
Zurück zum Zitat Huang B, Law MW, Mak HK et al (2009) Pediatric 64-MDCT coronary angiography with ECG-modulated tube current: radiation dose and cancer risk. AJR Am J Roentgenol 193:539–544CrossRef Huang B, Law MW, Mak HK et al (2009) Pediatric 64-MDCT coronary angiography with ECG-modulated tube current: radiation dose and cancer risk. AJR Am J Roentgenol 193:539–544CrossRef
2.
Zurück zum Zitat Brenner D, Elliston C, Hall E, Berdon W (2001) Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR Am J Roentgenol 176:289–296CrossRef Brenner D, Elliston C, Hall E, Berdon W (2001) Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR Am J Roentgenol 176:289–296CrossRef
3.
Zurück zum Zitat Meinel FG, Henzler T, Schoepf UJ et al (2015) ECG-synchronized CT angiography in 324 consecutive pediatric patients: spectrum of indications and trends in radiation dose. Pediatr Cardiol 36:569–578CrossRef Meinel FG, Henzler T, Schoepf UJ et al (2015) ECG-synchronized CT angiography in 324 consecutive pediatric patients: spectrum of indications and trends in radiation dose. Pediatr Cardiol 36:569–578CrossRef
4.
Zurück zum Zitat Herzog C, Mulvihill DM, Nguyen SA et al (2008) Pediatric cardiovascular CT angiography: radiation dose reduction using automatic anatomic tube current modulation. AJR Am J Roentgenol 190:1232–1240CrossRef Herzog C, Mulvihill DM, Nguyen SA et al (2008) Pediatric cardiovascular CT angiography: radiation dose reduction using automatic anatomic tube current modulation. AJR Am J Roentgenol 190:1232–1240CrossRef
5.
Zurück zum Zitat Einstein AJ, Elliston CD, Arai AE et al (2010) Radiation dose from single-heartbeat coronary CT angiography performed with a 320-detector row volume scanner. Radiology 254:698–706CrossRef Einstein AJ, Elliston CD, Arai AE et al (2010) Radiation dose from single-heartbeat coronary CT angiography performed with a 320-detector row volume scanner. Radiology 254:698–706CrossRef
6.
Zurück zum Zitat Zhang T, Wang W, Luo Z et al (2012) Initial experience on the application of 320-row CT angiography with low-dose prospective ECG-triggered in children with congenital heart disease. Int J Cardiovasc Imaging 28:1787–1797CrossRef Zhang T, Wang W, Luo Z et al (2012) Initial experience on the application of 320-row CT angiography with low-dose prospective ECG-triggered in children with congenital heart disease. Int J Cardiovasc Imaging 28:1787–1797CrossRef
7.
Zurück zum Zitat Al-Mousily F, Shifrin RY, Fricker FJ et al (2011) Use of 320-detector computed tomographic angiography for infants and young children with congenital heart disease. Pediatr Cardiol 32:426–432CrossRef Al-Mousily F, Shifrin RY, Fricker FJ et al (2011) Use of 320-detector computed tomographic angiography for infants and young children with congenital heart disease. Pediatr Cardiol 32:426–432CrossRef
8.
Zurück zum Zitat Yamasaki Y, Kawanami S, Kamitani T et al (2018) Free-breathing 320-row computed tomographic angiography with low-tube voltage and hybrid iterative reconstruction in infants with complex congenital heart disease. Clin Imaging 50:147–156CrossRef Yamasaki Y, Kawanami S, Kamitani T et al (2018) Free-breathing 320-row computed tomographic angiography with low-tube voltage and hybrid iterative reconstruction in infants with complex congenital heart disease. Clin Imaging 50:147–156CrossRef
9.
Zurück zum Zitat Shirota G, Maeda E, Namiki Y et al (2017) Pediatric 320-row cardiac computed tomography using electrocardiogram-gated model-based full iterative reconstruction. Pediatr Radiol 47:1463–1470CrossRef Shirota G, Maeda E, Namiki Y et al (2017) Pediatric 320-row cardiac computed tomography using electrocardiogram-gated model-based full iterative reconstruction. Pediatr Radiol 47:1463–1470CrossRef
10.
Zurück zum Zitat Maeda E, Yamamoto K, Kanno S et al (2016) Diagnostic phase of calcium scoring scan applied as the center of acquisition window of coronary computed tomography angiography improves image quality in minimal acquisition window scan (Target CTA Mode) using the second generation 320-row CT. ScientificWorldJournal 2016:1017851 Maeda E, Yamamoto K, Kanno S et al (2016) Diagnostic phase of calcium scoring scan applied as the center of acquisition window of coronary computed tomography angiography improves image quality in minimal acquisition window scan (Target CTA Mode) using the second generation 320-row CT. ScientificWorldJournal 2016:1017851
11.
Zurück zum Zitat Winkler MA, von Herrmann PF, Brooks MA et al (2013) Synthetic ECG-gated cardiac computed tomography for in vivo imaging of the temporary total artificial heart. Circulation 127:e4–e5CrossRef Winkler MA, von Herrmann PF, Brooks MA et al (2013) Synthetic ECG-gated cardiac computed tomography for in vivo imaging of the temporary total artificial heart. Circulation 127:e4–e5CrossRef
12.
Zurück zum Zitat Kimiya T, Sekiguchi S, Yagihashi T et al (2017) Sedation protocol with fasting and shorter sleep leads to magnetic resonance imaging success. Pediatr Int 59:1087–1090CrossRef Kimiya T, Sekiguchi S, Yagihashi T et al (2017) Sedation protocol with fasting and shorter sleep leads to magnetic resonance imaging success. Pediatr Int 59:1087–1090CrossRef
13.
Zurück zum Zitat Deak PD, Smal Y, Kalender WA (2010) Multisection CT protocols: sex- and age-specific conversion factors used to determine effective dose from dose-length product. Radiology 257:158–166CrossRef Deak PD, Smal Y, Kalender WA (2010) Multisection CT protocols: sex- and age-specific conversion factors used to determine effective dose from dose-length product. Radiology 257:158–166CrossRef
14.
Zurück zum Zitat Boone J, Strauss K, Cody D et al (2011) Size-specific dose estimates (SSDE) in pediatric and adult body CT examinations. Report of AAPM Task Group 204. College Park, Md: American Association of Physicists in Medicine Boone J, Strauss K, Cody D et al (2011) Size-specific dose estimates (SSDE) in pediatric and adult body CT examinations. Report of AAPM Task Group 204. College Park, Md: American Association of Physicists in Medicine
15.
Zurück zum Zitat Saake M, Lell MM, Rompel O et al (2014) Contrast medium application in pediatric high-pitch cardiovascular CT angiography: manual or power injection? J Cardiovasc Comput Tomogr 8:315–322CrossRef Saake M, Lell MM, Rompel O et al (2014) Contrast medium application in pediatric high-pitch cardiovascular CT angiography: manual or power injection? J Cardiovasc Comput Tomogr 8:315–322CrossRef
16.
Zurück zum Zitat Kundel HL, Polansky M (2003) Measurement of observer agreement. Radiology 228:303–308CrossRef Kundel HL, Polansky M (2003) Measurement of observer agreement. Radiology 228:303–308CrossRef
17.
Zurück zum Zitat Fleming S, Thompson M, Stevens R et al (2011) Normal ranges of heart rate and respiratory rate in children from the birth to 18 years of age: a systemic review of observational studies. Lancet 377:1011–1018CrossRef Fleming S, Thompson M, Stevens R et al (2011) Normal ranges of heart rate and respiratory rate in children from the birth to 18 years of age: a systemic review of observational studies. Lancet 377:1011–1018CrossRef
Metadaten
Titel
Comparison of image quality between synthetic and patients' electrocardiogram-gated 320-row pediatric cardiac computed tomography
Publikationsdatum
01.02.2020
Erschienen in
Pediatric Radiology / Ausgabe 2/2020
Print ISSN: 0301-0449
Elektronische ISSN: 1432-1998
DOI
https://doi.org/10.1007/s00247-019-04541-y

Weitere Artikel der Ausgabe 2/2020

Pediatric Radiology 2/2020 Zur Ausgabe

Hermes

Hermes

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.