Comparison of Inline R2* MRI versus FerriScan for liver iron quantification in patients on chelation therapy for iron overload: preliminary results

verfasst von: Gerard M. Healy, Stephan A. R. Kannengiesser, Osvaldo Espin-Garcia, Richard Ward, Kevin H. M. Kuo, Kartik S. Jhaveri

Erschienen in: European Radiology | Ausgabe 12/2021

Einloggen, um Zugang zu erhalten

Abstract

Objectives

MRI quantification of liver iron concentration (LIC) using R2 or R2* relaxometry requires offline post-processing causing reporting delays, administrative overhead, and added costs. A prototype 3D multi-gradient-echo pulse sequence, with inline post-processing, allows immediate calculation of LIC from an R2* map (inline R2*-LIC) without offline processing. We compared inline R2*-LIC to FerriScan and offline R2* calibration methods.

Methods

Forty patients (25 women, 15 men; age 18–82 years), prospectively underwent FerriScan and the prototype sequence, which produces two R2* maps, with and without fat modeling, as well as an inline R2*-LIC map derived from the R2* map with fat modeling, with informed consent. For each map, the following contours were drawn: ROIs, whole-axial-liver contour, and an exact copy of contour utilized by FerriScan. LIC values from the FerriScan report and those calculated using an alternative R2 calibration were the reference standards. Results were compared using Pearson and interclass correlation coefficients (PCC, ICC), linear regression, Bland-Altman analysis, and estimation of area under the receiver operator curve (ROC-AUC).

Results

Inline R2*-LIC demonstrated good agreement with the reference standards. Compared to FerriScan, inline R2*-LIC with whole-axial-liver contour, ROIs, and FerriScan contour demonstrated PCC of 94.8%, 94.8%, and 92%; ICC 93%, 92.7%, and 90.2%; regression slopes 1.004, 0.974, and 1.031; mean bias 5.54%, 10.91%, and 0.36%; and ROC-AUC estimates 0.903, 0.906, and 0.890 respectively. Agreement was maintained when adjusted for sex, age, diagnosis, liver fat content, and fat-water swap.

Conclusion

Inline R2*-LIC provides robust and comparable quantification of LIC compared to FerriScan, without the need for offline post-processing.

Key Points

• In patients being treated for iron overload with chelation therapy, liver iron concentration (LIC) is regularly assessed in order to monitor and adjust therapy.
• Magnetic resonance imaging (MRI) is commonly used to quantify LIC. Several R 2 and R 2 * methods are available, all of which require offline post-processing.
• A novel R 2 * MRI method allows for immediate calculation of LIC and provides comparable quantification of LIC to the FerriScan and recently published alternative R 2 * methods.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Marsella M, Borgna-Pignatti C (2014) Transfusional iron overload and iron chelation therapy in thalassemia major and sickle cell disease. Hematol Oncol Clin North Am 28:703–727 viCrossRef Marsella M, Borgna-Pignatti C (2014) Transfusional iron overload and iron chelation therapy in thalassemia major and sickle cell disease. Hematol Oncol Clin North Am 28:703–727 viCrossRef
2.
Zurück zum Zitat Taher AT, Cappellini MD, Aydinok Y et al (2016) Optimising iron chelation therapy with deferasirox for non-transfusion-dependent thalassaemia patients: 1-year results from the THETIS study. Blood Cells Mol Dis 57:23–29CrossRef Taher AT, Cappellini MD, Aydinok Y et al (2016) Optimising iron chelation therapy with deferasirox for non-transfusion-dependent thalassaemia patients: 1-year results from the THETIS study. Blood Cells Mol Dis 57:23–29CrossRef
3.
Zurück zum Zitat Hernando D, Levin YS, Sirlin CB, Reeder SB (2014) Quantification of liver iron with MRI: state of the art and remaining challenges. J Magn Reson Imaging 40:1003–1021CrossRef Hernando D, Levin YS, Sirlin CB, Reeder SB (2014) Quantification of liver iron with MRI: state of the art and remaining challenges. J Magn Reson Imaging 40:1003–1021CrossRef
4.
Zurück zum Zitat Sirlin CB, Reeder SB (2010) Magnetic resonance imaging quantification of liver iron. Magn Reson Imaging Clin N Am 18:359–381 ixCrossRef Sirlin CB, Reeder SB (2010) Magnetic resonance imaging quantification of liver iron. Magn Reson Imaging Clin N Am 18:359–381 ixCrossRef
5.
Zurück zum Zitat St Pierre TG, Clark PR, Chua-anusorn W et al (2005) Noninvasive measurement and imaging of liver iron concentrations using proton magnetic resonance. Blood 105:855–861CrossRef St Pierre TG, Clark PR, Chua-anusorn W et al (2005) Noninvasive measurement and imaging of liver iron concentrations using proton magnetic resonance. Blood 105:855–861CrossRef
6.
Zurück zum Zitat St Pierre TG, El-Beshlawy A, Elalfy M et al (2014) Multicenter validation of spin-density projection-assisted R2-MRI for the noninvasive measurement of liver iron concentration. Magn Reson Med 71:2215–2223CrossRef St Pierre TG, El-Beshlawy A, Elalfy M et al (2014) Multicenter validation of spin-density projection-assisted R2-MRI for the noninvasive measurement of liver iron concentration. Magn Reson Med 71:2215–2223CrossRef
7.
Zurück zum Zitat Doyle E, Ghugre N, Coates TD, Wood JC (2020) Fixing the MRI R2-iron calibration in liver. Am J Hematol 95:E120–e122CrossRef Doyle E, Ghugre N, Coates TD, Wood JC (2020) Fixing the MRI R2-iron calibration in liver. Am J Hematol 95:E120–e122CrossRef
8.
Zurück zum Zitat Wood JC, Enriquez C, Ghugre N et al (2005) MRI R2 and R2* mapping accurately estimates hepatic iron concentration in transfusion-dependent thalassemia and sickle cell disease patients. Blood 106:1460–1465CrossRef Wood JC, Enriquez C, Ghugre N et al (2005) MRI R2 and R2* mapping accurately estimates hepatic iron concentration in transfusion-dependent thalassemia and sickle cell disease patients. Blood 106:1460–1465CrossRef
9.
Zurück zum Zitat Anderson LJ, Holden S, Davis B et al (2001) Cardiovascular T2-star (T2*) magnetic resonance for the early diagnosis of myocardial iron overload. Eur Heart J 22:2171–2179CrossRef Anderson LJ, Holden S, Davis B et al (2001) Cardiovascular T2-star (T2*) magnetic resonance for the early diagnosis of myocardial iron overload. Eur Heart J 22:2171–2179CrossRef
10.
Zurück zum Zitat Gandon Y, Olivie D, Guyader D et al (2004) Non-invasive assessment of hepatic iron stores by MRI. Lancet 363:357–362CrossRef Gandon Y, Olivie D, Guyader D et al (2004) Non-invasive assessment of hepatic iron stores by MRI. Lancet 363:357–362CrossRef
12.
Zurück zum Zitat Garbowski MW, Carpenter JP, Smith G et al (2014) Biopsy-based calibration of T2* magnetic resonance for estimation of liver iron concentration and comparison with R2 Ferriscan. J Cardiovasc Magn Reson 16:40CrossRef Garbowski MW, Carpenter JP, Smith G et al (2014) Biopsy-based calibration of T2* magnetic resonance for estimation of liver iron concentration and comparison with R2 Ferriscan. J Cardiovasc Magn Reson 16:40CrossRef
13.
Zurück zum Zitat Jhaveri KS, Kannengiesser SAR, Ward R, Kuo K, Sussman MS (2019) Prospective evaluation of an R2* method for assessing liver iron concentration (LIC) against FerriScan: derivation of the calibration curve and characterization of the nature and source of uncertainty in the relationship. J Magn Reson Imaging 49:1467–1474CrossRef Jhaveri KS, Kannengiesser SAR, Ward R, Kuo K, Sussman MS (2019) Prospective evaluation of an R2* method for assessing liver iron concentration (LIC) against FerriScan: derivation of the calibration curve and characterization of the nature and source of uncertainty in the relationship. J Magn Reson Imaging 49:1467–1474CrossRef
14.
Zurück zum Zitat Henninger B, Zoller H, Rauch S et al (2015) R2* relaxometry for the quantification of hepatic iron overload: biopsy-based calibration and comparison with the literature. Rofo 187:472–479CrossRef Henninger B, Zoller H, Rauch S et al (2015) R2* relaxometry for the quantification of hepatic iron overload: biopsy-based calibration and comparison with the literature. Rofo 187:472–479CrossRef
15.
16.
Zurück zum Zitat Feng Y, He T, Gatehouse PD et al (2013) Improved MRI R2 * relaxometry of iron-loaded liver with noise correction. Magn Reson Med 70:1765–1774CrossRef Feng Y, He T, Gatehouse PD et al (2013) Improved MRI R2 * relaxometry of iron-loaded liver with noise correction. Magn Reson Med 70:1765–1774CrossRef
17.
Zurück zum Zitat Tsai YL, Wu CJ, Shaw S, Yu PC, Nien HH, Lui LT (2018) Quantitative analysis of respiration-induced motion of each liver segment with helical computed tomography and 4-dimensional computed tomography. Radiat Oncol 13:59CrossRef Tsai YL, Wu CJ, Shaw S, Yu PC, Nien HH, Lui LT (2018) Quantitative analysis of respiration-induced motion of each liver segment with helical computed tomography and 4-dimensional computed tomography. Radiat Oncol 13:59CrossRef
18.
Zurück zum Zitat Pavitt HL, Aydinok Y, El-Beshlawy A et al (2011) The effect of reducing repetition time TR on the measurement of liver R2 for the purpose of measuring liver iron concentration. Magn Reson Med 65:1346–1351CrossRef Pavitt HL, Aydinok Y, El-Beshlawy A et al (2011) The effect of reducing repetition time TR on the measurement of liver R2 for the purpose of measuring liver iron concentration. Magn Reson Med 65:1346–1351CrossRef
19.
Zurück zum Zitat Obuchowski NA (2006) An ROC-type measure of diagnostic accuracy when the gold standard is continuous-scale. Stat Med 25:481–493CrossRef Obuchowski NA (2006) An ROC-type measure of diagnostic accuracy when the gold standard is continuous-scale. Stat Med 25:481–493CrossRef
20.
Zurück zum Zitat Obuchowski NA, Reeves AP, Huang EP et al (2015) Quantitative imaging biomarkers: a review of statistical methods for computer algorithm comparisons. Stat Methods Med Res 24:68–106CrossRef Obuchowski NA, Reeves AP, Huang EP et al (2015) Quantitative imaging biomarkers: a review of statistical methods for computer algorithm comparisons. Stat Methods Med Res 24:68–106CrossRef
21.
Zurück zum Zitat Henninger B, Alustiza J, Garbowski M, Gandon Y (2020) Practical guide to quantification of hepatic iron with MRI. Eur Radiol 30:383–393CrossRef Henninger B, Alustiza J, Garbowski M, Gandon Y (2020) Practical guide to quantification of hepatic iron with MRI. Eur Radiol 30:383–393CrossRef
22.
Zurück zum Zitat Mandal S, Sodhi KS, Bansal D et al (2017) MRI for quantification of liver and cardiac iron in thalassemia major patients: pilot study in Indian population. Indian J Pediatr 84:276–282CrossRef Mandal S, Sodhi KS, Bansal D et al (2017) MRI for quantification of liver and cardiac iron in thalassemia major patients: pilot study in Indian population. Indian J Pediatr 84:276–282CrossRef
23.
Zurück zum Zitat Henninger B, Zoller H, Kannengiesser S, Zhong X, Jaschke W, Kremser C (2017) 3D multiecho dixon for the evaluation of hepatic iron and fat in a clinical setting. J Magn Reson Imaging 46:793–800CrossRef Henninger B, Zoller H, Kannengiesser S, Zhong X, Jaschke W, Kremser C (2017) 3D multiecho dixon for the evaluation of hepatic iron and fat in a clinical setting. J Magn Reson Imaging 46:793–800CrossRef
24.
Zurück zum Zitat Serai SD, Smith EA, Trout AT, Dillman JR (2018) Agreement between manual relaxometry and semi-automated scanner-based multi-echo Dixon technique for measuring liver T2* in a pediatric and young adult population. Pediatr Radiol 48:94–100CrossRef Serai SD, Smith EA, Trout AT, Dillman JR (2018) Agreement between manual relaxometry and semi-automated scanner-based multi-echo Dixon technique for measuring liver T2* in a pediatric and young adult population. Pediatr Radiol 48:94–100CrossRef
25.
Zurück zum Zitat Zhong X, Nickel MD, Kannengiesser SA, Dale BM, Kiefer B, Bashir MR (2014) Liver fat quantification using a multi-step adaptive fitting approach with multi-echo GRE imaging. Magn Reson Med 72:1353–1365CrossRef Zhong X, Nickel MD, Kannengiesser SA, Dale BM, Kiefer B, Bashir MR (2014) Liver fat quantification using a multi-step adaptive fitting approach with multi-echo GRE imaging. Magn Reson Med 72:1353–1365CrossRef
26.
Zurück zum Zitat Reeder SB, Hu HH, Sirlin CB (2012) Proton density fat-fraction: a standardized MR-based biomarker of tissue fat concentration. J Magn Reson Imaging 36:1011–1014CrossRef Reeder SB, Hu HH, Sirlin CB (2012) Proton density fat-fraction: a standardized MR-based biomarker of tissue fat concentration. J Magn Reson Imaging 36:1011–1014CrossRef
27.
Zurück zum Zitat Yokoo T, Serai SD, Pirasteh A et al (2018) Linearity, bias, and precision of hepatic proton density fat fraction measurements by using MR imaging: a meta-analysis. Radiology 286:486–498CrossRef Yokoo T, Serai SD, Pirasteh A et al (2018) Linearity, bias, and precision of hepatic proton density fat fraction measurements by using MR imaging: a meta-analysis. Radiology 286:486–498CrossRef
28.
Zurück zum Zitat Karlsson M, Ekstedt M, Dahlström N et al (2019) Liver R2* is affected by both iron and fat: a dual biopsy-validated study of chronic liver disease. J Magn Reson Imaging 50:325–333CrossRef Karlsson M, Ekstedt M, Dahlström N et al (2019) Liver R2* is affected by both iron and fat: a dual biopsy-validated study of chronic liver disease. J Magn Reson Imaging 50:325–333CrossRef
29.
Zurück zum Zitat Rose C, Vandevenne P, Bourgeois E, Cambier N, Ernst O (2006) Liver iron content assessment by routine and simple magnetic resonance imaging procedure in highly transfused patients. Eur J Haematol 77:145–149CrossRef Rose C, Vandevenne P, Bourgeois E, Cambier N, Ernst O (2006) Liver iron content assessment by routine and simple magnetic resonance imaging procedure in highly transfused patients. Eur J Haematol 77:145–149CrossRef
30.
Zurück zum Zitat McCarville MB, Hillenbrand CM, Loeffler RB et al (2010) Comparison of whole liver and small region-of-interest measurements of MRI liver R2* in children with iron overload. Pediatr Radiol 40:1360–1367CrossRef McCarville MB, Hillenbrand CM, Loeffler RB et al (2010) Comparison of whole liver and small region-of-interest measurements of MRI liver R2* in children with iron overload. Pediatr Radiol 40:1360–1367CrossRef
31.
Zurück zum Zitat Stocker D, Bashir MR, Kannengiesser SAR, Reiner CS (2018) Accuracy of automated liver contouring, fat fraction, and R2* measurement on gradient multiecho magnetic resonance images. J Comput Assist Tomogr 42:697–706CrossRef Stocker D, Bashir MR, Kannengiesser SAR, Reiner CS (2018) Accuracy of automated liver contouring, fat fraction, and R2* measurement on gradient multiecho magnetic resonance images. J Comput Assist Tomogr 42:697–706CrossRef
Metadaten
Titel
Comparison of Inline R2* MRI versus FerriScan for liver iron quantification in patients on chelation therapy for iron overload: preliminary results
verfasst von
Gerard M. Healy
Stephan A. R. Kannengiesser
Osvaldo Espin-Garcia
Richard Ward
Kevin H. M. Kuo
Kartik S. Jhaveri
Publikationsdatum
26.05.2021
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 12/2021
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-021-08019-0

Neu im Fachgebiet Radiologie

Abdominale CT bei Kindern: 40% mit Zufallsbefunden

Wird bei Kindern mit stumpfem Trauma eine CT des Bauchraums veranlasst, sind in rund 40% der Fälle Auffälligkeiten zu sehen, die nichts mit dem Trauma zu tun haben. Die allerwenigsten davon sind klinisch relevant.

Genügt die biparametrische MRT für die Prostatadiagnostik?

Die multiparametrische Magnetresonanztomografie hat einen festen Platz im Screening auf klinisch signifikante Prostatakarzinome. Ob auch ein biparametrisches Vorgehen ausreicht, ist in einer Metaanalyse untersucht worden.

Höhere Trefferquoten bei Brustkrebsscreening dank KI?

Künstliche Intelligenz unterstützt bei der Auswertung von Mammografie-Screenings und senkt somit den Arbeitsaufwand für Radiologen. Wie wirken sich diese Technologien auf die Trefferquote und die Falsch-positiv-Rate aus? Das hat jetzt eine Studie aus Schweden untersucht.

KI-gestütztes Mammografiescreening überzeugt im Praxistest

Mit dem Einsatz künstlicher Intelligenz lässt sich die Detektionsrate im Mammografiescreening offenbar deutlich steigern. Mehr unnötige Zusatzuntersuchungen sind laut der Studie aus Deutschland nicht zu befürchten.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.