01.09.2012 | Cataract

Comparison of intraocular lens power prediction using immersion ultrasound and optical biometry with and without formula optimization

verfasst von: Gabor Nemeth, Attila Nagy, Andras Berta, Laszlo Modis Jr

Erschienen in: Graefe's Archive for Clinical and Experimental Ophthalmology | Ausgabe 9/2012

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Comparison of postoperative refraction results using ultrasound biometry with closed immersion shell and optical biometry.

Patients and method

Three hundred and sixty-four eyes of 306 patients (age: 70.6 ± 12.8 years) underwent cataract surgery where intraocular lenses calculated by SRK/T formula were implanted. In 159 cases immersion ultrasonic biometry, in 205 eyes optical biometry was used. Differences between predicted and actual postoperative refractions were calculated both prior to and after optimization with the SRK/T formula, after which we analysed the similar data in the case of Holladay, Haigis, and Hoffer-Q formulas. Mean absolute error (MAE) and the percentage rate of patients within ±0.5 and ±1.0 D difference in the predicted error were calculated with these four formulas.

Results

MAE was 0.5–0.7 D in cases of both methods with SRK/T, Holladay, and Hoffer-Q formula, but higher with Haigis formula. With no optimization, 60–65 % of the patients were under 0.5 D error in the immersion group (except for Haigis formula). Using the optical method, this value was slightly higher (62-67 %), however, in this case, Haigis formula also did not perform so well (45 %). Refraction results significantly improved with Holladay, Hoffer-Q, and Haigis formulas in both groups. The rate of patients under 0.5 D error increased to 65 % by the immersion technique, and up to 80 % by the optical one.

Conclusions

According to our results, optical biometry offers only slightly better outcomes compared to those of immersion shell with no optimized formulas. However, in case of new generation formulas with both methods, the optimization of IOL-constants give significantly better results.
Literatur
1.
Zurück zum Zitat Norrby S (2008) Sources of error in intraocular lens power calculation. J Cataract Refract Surg 34:368–376 PubMedCrossRef Norrby S (2008) Sources of error in intraocular lens power calculation. J Cataract Refract Surg 34:368–376 PubMedCrossRef
2.
Zurück zum Zitat Olsen T (1992) Sources of error in intraocular lens power calculation. J Cataract Refract Surg 18:125–129 PubMed Olsen T (1992) Sources of error in intraocular lens power calculation. J Cataract Refract Surg 18:125–129 PubMed
3.
Zurück zum Zitat Gale RP, Saldana M, Johnston RL, Zuberbuhler B, McKibbin M (2009) Benchmark standards for refractive outcomes after NHS cataract surgery. Eye 23:149–152 PubMedCrossRef Gale RP, Saldana M, Johnston RL, Zuberbuhler B, McKibbin M (2009) Benchmark standards for refractive outcomes after NHS cataract surgery. Eye 23:149–152 PubMedCrossRef
4.
Zurück zum Zitat Kiss B, Findl O, Menapace R, Wirtitsch M, Petternel V, Drexler W, Rainer G, Georgopoulos M, Hitzenberger CK, Fercher AF (2002) Refractive outcome of cataract surgery using partial coherence interferometry and ultrasound biometry: clinical feasibility study of a commercial prototype II. J Cataract Refract Surg 28:230–234 PubMedCrossRef Kiss B, Findl O, Menapace R, Wirtitsch M, Petternel V, Drexler W, Rainer G, Georgopoulos M, Hitzenberger CK, Fercher AF (2002) Refractive outcome of cataract surgery using partial coherence interferometry and ultrasound biometry: clinical feasibility study of a commercial prototype II. J Cataract Refract Surg 28:230–234 PubMedCrossRef
5.
Zurück zum Zitat Olsen T (2007) Improved accuracy of intraocular lens power calculation with the Zeiss IOLMaster. Acta Ophthalmol Scand 85:84–87 PubMedCrossRef Olsen T (2007) Improved accuracy of intraocular lens power calculation with the Zeiss IOLMaster. Acta Ophthalmol Scand 85:84–87 PubMedCrossRef
6.
Zurück zum Zitat Charalampidou S, Cassidy L, Ng E, Loughman J, Nolan J, Stack J, Beatty S (2010) Effect on refractive outcomes after cataract surgery of intraocular lens constant personalization using the Haigis formula. J Cataract Refract Surg 36:1081–1089 PubMedCrossRef Charalampidou S, Cassidy L, Ng E, Loughman J, Nolan J, Stack J, Beatty S (2010) Effect on refractive outcomes after cataract surgery of intraocular lens constant personalization using the Haigis formula. J Cataract Refract Surg 36:1081–1089 PubMedCrossRef
7.
Zurück zum Zitat Raymond S, Favilla I, Santamaria L (2009) Comparing ultrasound biometry with partial coherence interferometry for intraocular lens power calculations: a randomized study. Invest Ophthalmol Vis Sci 50:2547–2552 PubMedCrossRef Raymond S, Favilla I, Santamaria L (2009) Comparing ultrasound biometry with partial coherence interferometry for intraocular lens power calculations: a randomized study. Invest Ophthalmol Vis Sci 50:2547–2552 PubMedCrossRef
8.
Zurück zum Zitat Aristodemou P, Knox Cartwright NE, Sparrow JM, Johnston RL (2011) Formula choice: Hoffer Q, Holladay 1, or SRK/T and refractive outcomes in 8108 eyes after cataract surgery with biometry by partial coherence interferometry. J Cataract Refract Surg 37:63–71 PubMedCrossRef Aristodemou P, Knox Cartwright NE, Sparrow JM, Johnston RL (2011) Formula choice: Hoffer Q, Holladay 1, or SRK/T and refractive outcomes in 8108 eyes after cataract surgery with biometry by partial coherence interferometry. J Cataract Refract Surg 37:63–71 PubMedCrossRef
9.
Zurück zum Zitat Holladay JT, Prager TC, Ruiz RS, Lewis JW, Rosenthal H (1986) Improving the predictability of intraocular lens power calculations. Arch Ophthalmol 104:539–541 PubMedCrossRef Holladay JT, Prager TC, Ruiz RS, Lewis JW, Rosenthal H (1986) Improving the predictability of intraocular lens power calculations. Arch Ophthalmol 104:539–541 PubMedCrossRef
10.
Zurück zum Zitat Schelenz J, Kammann J (1989) Comparison of contact and immersion technique for axial length measurement and implant power calculation. J Cataract Refract Surg 15:425–428 PubMed Schelenz J, Kammann J (1989) Comparison of contact and immersion technique for axial length measurement and implant power calculation. J Cataract Refract Surg 15:425–428 PubMed
11.
Zurück zum Zitat Shammas HJF (1984) A comparison of immersion and contact techniques for axial length measurements. J Am Intraocul Implant Soc 10:444–447 PubMed Shammas HJF (1984) A comparison of immersion and contact techniques for axial length measurements. J Am Intraocul Implant Soc 10:444–447 PubMed
12.
Zurück zum Zitat Olsen T, Nielsen PJ (1989) Immersion versus contact technique in the measurement of axial length by ultrasound. Acta Ophthalmol 67:101–102 Olsen T, Nielsen PJ (1989) Immersion versus contact technique in the measurement of axial length by ultrasound. Acta Ophthalmol 67:101–102
13.
Zurück zum Zitat Haigis W, Lege B, Miller N, Schneider B (2000) Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis. Graefes Arch Clin Exp Ophthalmol 238:765–773 PubMedCrossRef Haigis W, Lege B, Miller N, Schneider B (2000) Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis. Graefes Arch Clin Exp Ophthalmol 238:765–773 PubMedCrossRef
14.
Zurück zum Zitat Madge SN, Khong CH, Lamont M, Bansal A, Antcliff RJ (2005) Optimization of biometry for intraocular lens implantation using the Zeiss IOLMaster. Acta Ophthalmol Scand 83:436–438 Madge SN, Khong CH, Lamont M, Bansal A, Antcliff RJ (2005) Optimization of biometry for intraocular lens implantation using the Zeiss IOLMaster. Acta Ophthalmol Scand 83:436–438
15.
Zurück zum Zitat Findl O, Drexler W, Menapace R, Heinzl H, Hitzenberger CK, Fercher AF (2001) Improved prediction of intraocular lens power using partial coherence interferometry. J Cataract Refract Surg 27:861–867 PubMedCrossRef Findl O, Drexler W, Menapace R, Heinzl H, Hitzenberger CK, Fercher AF (2001) Improved prediction of intraocular lens power using partial coherence interferometry. J Cataract Refract Surg 27:861–867 PubMedCrossRef
16.
Zurück zum Zitat Hoffmann PC, Hütz WW, Eckhardt HB, Heuring AH (1998) Intraocular lens calculation and ultrasound biometry: immersion and contact procedures. Klin Monatsbl Augenheilkd 213:161–165 PubMedCrossRef Hoffmann PC, Hütz WW, Eckhardt HB, Heuring AH (1998) Intraocular lens calculation and ultrasound biometry: immersion and contact procedures. Klin Monatsbl Augenheilkd 213:161–165 PubMedCrossRef
17.
Zurück zum Zitat Hennessy MP, Franzco CDG (2003) Contact versus immersion biometry of axial length before cataract surgery. J Cataract Refract Surg 29:2195–2198 PubMedCrossRef Hennessy MP, Franzco CDG (2003) Contact versus immersion biometry of axial length before cataract surgery. J Cataract Refract Surg 29:2195–2198 PubMedCrossRef
18.
Zurück zum Zitat Hitzenberger CK, Drexler W, Dolezal C, Skorpik F, Juchem M, Fercher AF, Gnad HD (1993) Measurement of the axial length of cataract eyes by laser Doppler interferometry. Invest Ophthalmol Vis Sci 34:1886–1893 PubMed Hitzenberger CK, Drexler W, Dolezal C, Skorpik F, Juchem M, Fercher AF, Gnad HD (1993) Measurement of the axial length of cataract eyes by laser Doppler interferometry. Invest Ophthalmol Vis Sci 34:1886–1893 PubMed
19.
Zurück zum Zitat Nemeth J, Fekete O, Pesztenlehrer N (2003) Optical and ultrasound measurement of axial length and anterior chamber depth for intraocular lens power calculation. J Cataract Refract Surg 29:85–88 PubMedCrossRef Nemeth J, Fekete O, Pesztenlehrer N (2003) Optical and ultrasound measurement of axial length and anterior chamber depth for intraocular lens power calculation. J Cataract Refract Surg 29:85–88 PubMedCrossRef
20.
Zurück zum Zitat Lenhart PD, Hutchinson AK, Lynn MJ, Lambert SR (2010) Partial coherence interferometry versus immersion ultrasonography for axial length measurement in children. J Cataract Refract Surg 36:2100–2104 PubMedCrossRef Lenhart PD, Hutchinson AK, Lynn MJ, Lambert SR (2010) Partial coherence interferometry versus immersion ultrasonography for axial length measurement in children. J Cataract Refract Surg 36:2100–2104 PubMedCrossRef
21.
Zurück zum Zitat Hrebcová J, Skorkovská S, Vasků A (2009) Comparison of contact and immersion techniques of ultrasound biometry in terms of target postoperative refraction. Cesk Slov Oftalmol 65:143–146 PubMed Hrebcová J, Skorkovská S, Vasků A (2009) Comparison of contact and immersion techniques of ultrasound biometry in terms of target postoperative refraction. Cesk Slov Oftalmol 65:143–146 PubMed
22.
Zurück zum Zitat Narvaez J, Cherwek DH, Stulting RD, Waldron R, Zimmerman GJ, Wessels IE, Waring GO III (2008) Comparing immersion ultrasound with partial coherence interferometry for intraocular lens power calculation. Ophthalmic Surg Lasers Imaging 39:30–34 PubMedCrossRef Narvaez J, Cherwek DH, Stulting RD, Waldron R, Zimmerman GJ, Wessels IE, Waring GO III (2008) Comparing immersion ultrasound with partial coherence interferometry for intraocular lens power calculation. Ophthalmic Surg Lasers Imaging 39:30–34 PubMedCrossRef
23.
Zurück zum Zitat Montés-Micó R, Carones F, Buttacchio A, Ferrer-Blasco T, Madrid-Costa D (2011) Comparison of immersion ultrasound, partial coherence interferometry, and low coherence reflectometry for ocular biometry in cataract patients. J Refract Surg 27:665–671 PubMedCrossRef Montés-Micó R, Carones F, Buttacchio A, Ferrer-Blasco T, Madrid-Costa D (2011) Comparison of immersion ultrasound, partial coherence interferometry, and low coherence reflectometry for ocular biometry in cataract patients. J Refract Surg 27:665–671 PubMedCrossRef
24.
Zurück zum Zitat Packer M, Fine IH, Hoffman RS, Coffman PG, Brown LK (2002) Immersion A-scan compared with partial coherence interferometry; outcomes analysis. J Cataract Refract Surg 28:239–242 PubMedCrossRef Packer M, Fine IH, Hoffman RS, Coffman PG, Brown LK (2002) Immersion A-scan compared with partial coherence interferometry; outcomes analysis. J Cataract Refract Surg 28:239–242 PubMedCrossRef
25.
Zurück zum Zitat Eleftheriadis H (2003) IOLMaster biometry: refractive results of 100 consecutive cases. Br J Ophthalmol 87:960–963 PubMedCrossRef Eleftheriadis H (2003) IOLMaster biometry: refractive results of 100 consecutive cases. Br J Ophthalmol 87:960–963 PubMedCrossRef
26.
Zurück zum Zitat Percival SPB, Vyas AV, Setty SS, Manvikar S (2002) The influence of implant design on accuracy of postoperative refraction. Eye 16:309–315 PubMedCrossRef Percival SPB, Vyas AV, Setty SS, Manvikar S (2002) The influence of implant design on accuracy of postoperative refraction. Eye 16:309–315 PubMedCrossRef
27.
Zurück zum Zitat Retzlaff JA, Sanders DR, Kraff MC (1990) Development of the SRK/T intraocular lens implant power calculation formula. J Cataract Refract Surg 16:333–340, correction, 528 PubMed Retzlaff JA, Sanders DR, Kraff MC (1990) Development of the SRK/T intraocular lens implant power calculation formula. J Cataract Refract Surg 16:333–340, correction, 528 PubMed
28.
Zurück zum Zitat Royal College of Ophthalmologists (2007) Cataract surgery guidelines (2004, updated 2007). Royal College of Ophthalmologists, London, p 25 Royal College of Ophthalmologists (2007) Cataract surgery guidelines (2004, updated 2007). Royal College of Ophthalmologists, London, p 25
29.
Zurück zum Zitat Sheard RM, Smith GT, Cooke DL (2010) Improving the prediction accuracy of the SRK/T formula: the T2 formula. J Cataract Refract Surg 36:1829–1834 PubMedCrossRef Sheard RM, Smith GT, Cooke DL (2010) Improving the prediction accuracy of the SRK/T formula: the T2 formula. J Cataract Refract Surg 36:1829–1834 PubMedCrossRef
30.
Zurück zum Zitat Sarver EJ (2011) Comments on: Improving the prediction accuracy of the SRK/T formula: The T2 formula. J Cataract Refract Surg 37:795, letter PubMedCrossRef Sarver EJ (2011) Comments on: Improving the prediction accuracy of the SRK/T formula: The T2 formula. J Cataract Refract Surg 37:795, letter PubMedCrossRef
31.
Zurück zum Zitat Sheard R, Smith G, Cooke D (2011) Reply: comments on study of SKR/T versus T2 formula, comments on: improving the prediction accuracy of the SRK/T formula: the T2 formula and what constitutes "clinically significant" improvement in implant power calculation accuracy? J Cataract Refract Surg 37:796–798, letter CrossRef Sheard R, Smith G, Cooke D (2011) Reply: comments on study of SKR/T versus T2 formula, comments on: improving the prediction accuracy of the SRK/T formula: the T2 formula and what constitutes "clinically significant" improvement in implant power calculation accuracy? J Cataract Refract Surg 37:796–798, letter CrossRef
32.
Zurück zum Zitat Lee AC, Qazi MA, Pepose JS (2008) Biometry and intraocular lens power calculation. Curr Opin Ophthalmol 19:13–17 PubMedCrossRef Lee AC, Qazi MA, Pepose JS (2008) Biometry and intraocular lens power calculation. Curr Opin Ophthalmol 19:13–17 PubMedCrossRef
33.
Zurück zum Zitat Olsen T, Corydon L, Gimbel H (1995) Intraocular lens power calculation with an improved anterior chamber depth prediction algorithm. J Cataract Refract Surg 21:313–319 PubMed Olsen T, Corydon L, Gimbel H (1995) Intraocular lens power calculation with an improved anterior chamber depth prediction algorithm. J Cataract Refract Surg 21:313–319 PubMed
34.
Zurück zum Zitat Holladay JT, Prager TC, Chandler TY, Musgrove KH, Lewis JW, Ruiz RS (1988) A three-part system for refining intraocular lens power calculations. J Cataract Refract Surg 14:17–24 PubMed Holladay JT, Prager TC, Chandler TY, Musgrove KH, Lewis JW, Ruiz RS (1988) A three-part system for refining intraocular lens power calculations. J Cataract Refract Surg 14:17–24 PubMed
35.
Zurück zum Zitat Hoffer KJ (1993) The Hoffer Q formula: a comparison of theoretic and regression formulas. J Cataract Refract Surg 19:700–712, errata 1994; 20:677 PubMed Hoffer KJ (1993) The Hoffer Q formula: a comparison of theoretic and regression formulas. J Cataract Refract Surg 19:700–712, errata 1994; 20:677 PubMed
36.
Zurück zum Zitat Petermeier K, Gekeler F, Messias A, Spitzer MS, Haigis W, Szurman P (2009) Intraocular lens power calculation and optimized constants for highly myopic eyes. J Cataract Refract Surg 35:1575–1581 PubMedCrossRef Petermeier K, Gekeler F, Messias A, Spitzer MS, Haigis W, Szurman P (2009) Intraocular lens power calculation and optimized constants for highly myopic eyes. J Cataract Refract Surg 35:1575–1581 PubMedCrossRef
37.
Zurück zum Zitat Holladay JT (1997) Standardizing constants for ultrasonic biometry, keratometry, and intraocular lens power calculations. J Cataract Refract Surg 23:1356–1370 PubMed Holladay JT (1997) Standardizing constants for ultrasonic biometry, keratometry, and intraocular lens power calculations. J Cataract Refract Surg 23:1356–1370 PubMed
38.
Zurück zum Zitat Aristodemou P, Knox Cartwright NE, Sparrow JM, Johnston RL (2011) Intraocular lens formula constant optimization and partial coherence interferometry biometry: Refractive outcomes in 8108 eyes after cataract surgery. J Cataract Refract Surg 37:50–62 PubMedCrossRef Aristodemou P, Knox Cartwright NE, Sparrow JM, Johnston RL (2011) Intraocular lens formula constant optimization and partial coherence interferometry biometry: Refractive outcomes in 8108 eyes after cataract surgery. J Cataract Refract Surg 37:50–62 PubMedCrossRef
Metadaten
Titel
Comparison of intraocular lens power prediction using immersion ultrasound and optical biometry with and without formula optimization
verfasst von
Gabor Nemeth
Attila Nagy
Andras Berta
Laszlo Modis Jr
Publikationsdatum
01.09.2012
Verlag
Springer-Verlag
Erschienen in
Graefe's Archive for Clinical and Experimental Ophthalmology / Ausgabe 9/2012
Print ISSN: 0721-832X
Elektronische ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-012-2013-9

Weitere Artikel der Ausgabe 9/2012

Graefe's Archive for Clinical and Experimental Ophthalmology 9/2012 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde