Comparison of intraocular pressure as measured by three different non-contact tonometers and goldmann applanation tonometer for non-glaucomatous subjects

verfasst von: Seung Pil Bang, Chong Eun Lee, Yu Cheol Kim

Erschienen in: BMC Ophthalmology | Ausgabe 1/2017

Abstract

Background

To compare the measurement of intraocular pressure (IOP) among the three different non-contact tonometers (NCT) and the Goldmann applanation tonometer (GAT) for non-glaucomatous subjects.

Methods

In 52 eyes of 52 non-glaucomatous subjects, IOP was measured sequentially with the Canon TX-20P, the Nidek NT-530P, the Topcon CT-1P, and the GAT at the same time. We evaluated the IOP-measurement agreement among the tonometers as well as the factors affecting the measurements.

Results

A significant positive correlation was shown between the IOP values obtained with GAT and each NCT. The Canon TX-20P showed statistically the most significant agreement with the GAT (ICC 0.906, 95% CI 0.837–0.946). In an analysis of the Bland-Altman plots, the Canon TX-20P also showed the largest mean bias (1.38 mmHg) but the narrowest limits of agreement (LoA) (95% LoA; ± 3.43 mmHg). The Topcon CT-1P showed the smallest mean bias (0.48 mmHg) but the widest LoA (95% LoA; ± 4.16 mmHg). The Topcon CT-1P and Nidek NT-530P both showed a significantly positive correlation between increasing central corneal thickness (CCT) and increasing IOP.

Conclusion

There was a statistically significant correlation between each of the three different NCT and the GAT measurements. IOP measured with the Canon TX-20P and Topcon CT-1P tended to be higher, and with the Nidek NT-530P lower, than with the GAT. Practitioners need to know the properties of their own NCTs and their respective measurement tendencies.
Abkürzungen
C-IOP
Intraocular pressure values obtained with the Canon TX-20P
CCT
Central corneal thickness
D
Diopters
GAT
Goldmann applanation tonometer
G-IOP
Intraocular pressure values obtained with the GAT
ICC
Intraclass correlation coefficient
IOP
Intraocular pressure
IRB
Institutional Review Board
LoA
Limits of agreement
NCT
Non-contact tonometers
N-IOP
Intraocular pressure values obtained with the Nidek NT-530P
SD
Standard deviation
T-IOP
Intraocular pressure values obtained with the Topcon CT-1P

Background

Currently, the Goldmann applanation tonometer (GAT) is considered to be the gold standard for IOP measurement in clinical practice [16]. However, since their introduction, non-contact tonometers (NCTs) have become well established in clinical practice. NCT is a rapid, simple and objective method of intraocular pressure (IOP) measurement that can be performed by ancillary staff without the use of corneal anaesthesia [7].
Not all NCTs perform similarly. Although earlier models showed, relative to the GAT, a wider bias and a poor sensitivity for detection of elevated IOP [1, 3, 7, 8], the accuracies of subsequent models are much improved [918]. The Canon TX-20P, Nidek NT-530P, and Topcon CT-1P are the new automatic tonometers currently marketed. Perhaps surprisingly, there are in the literature, to our knowledge, few reports available on the performance of the Canon TX-20P or Nidek NT-530P [1015], and none on the Topcon CT-1P. Also, various studies have compared just one or two of the NCTs with the GAT, and there have been few studies comparing the accuracy of various tonometers for non-glaucomatous healthy adult subjects [10, 14, 1921].
The purposes of this study were to evaluate the reliability of the three new NCTs and to compare their IOP measurements with those obtained by the GAT for non-glaucomatous subjects. For each tonometer, the relationships between the measured ocular parameters and the IOP measurements also were examined.

Methods

Fifty-two eyes of 52 non-glaucomatous healthy subjects aged 21 to 85 years were included in the study. The study was approved by the Institutional Review Board (IRB) of Keimyung University Dongsan Medical Center (IRB no. 2015–12-052) and was performed in accordance with the tenets of the Declaration of Helsinki. None of the subjects had any history of ocular pathology affecting IOP, refractive surgery or trauma, and none had worn contact lenses within the two-week period prior to IOP measurement. Subjects were excluded if they had regular astigmatism greater than 3.50 diopters (D) or any irregular astigmatism. To avoid the double-organ bias, one eye per individual was randomly included in the analysis using a table of random numbers for randomization [22]. In a fixed sequence, all of the subjects were examined with the three NCTs, the GAT and ultrasound pachymeter (850, Humphrey Instruments, Inc., San Leandro, CA, USA) to obtain IOP and central corneal thickness (CCT) measurements, respectively. According to the GAT IOP values, the eyes were categorized into low-teen (<14 mmHg), mid-teen (≥14 and ≤17 mmHg) and high-teen (>17 mmHg) groups.
IOP measurements were made by the same experienced ancillary staff using the three NCTs. The order in which the instruments were used was the Canon TX-20P, followed by the Nidek NT-530P, the Topcon CT-1P and the GAT. Each of the tonometers was calibrated according to the manufacturer’s guidelines prior to its use in this study. In manual measurements using the Nidek NT-530P, on the other hand, the operator aligns the cornea by superimposing a reflection of the target from the subject’s cornea on a stationary ring and depresses the trigger when the cornea is aligned. In the present study, the mean of three measurements was used so as to avoid the effect of fluctuations caused by the cardiac pulse cycle.
IOP measurements were taken with the GAT (AT900; Haag-Streit, Köniz, Switzerland) according to the standard procedures. Before acquisition, one drop of 0.5% proparacaine hydrochloride eye drops (Alcaine®, Alcon Laboratories Inc., Fort Worth, TX, USA) was instilled and a fluorescein strip was applied to the inferior conjunctival fornix. To avoid error introduced by topical anesthesia, the GAT was applied five minutes after eyedrop instillation [23]. The last IOP measurement was obtained using GAT to avoid a corneal-compression-induced aqueous outflow increase that would have affected subsequent IOP readings [24, 25]. Also, the mean of three measurements of GAT was used and each IOP readings were masked to the one clinician (SPB) performing the measurements. Between each instrumentation application, the subjects were allowed a five-minute rest period to recover from the aqueous outflow. All of the measurements were taken between 11:30 am and 1:30 pm in order to minimize the effects of diurnal IOP variation [26].
Pearson correlation analysis, the intraclass correlation coefficient (ICC), the paired t-test were used to assess the correlation, consistency and agreement among the IOP measurements provided by each instrument. We also constructed Bland-Altman plots using Medcalc version 15.2 (Ostend, Belgium) to compare the bias in the IOP measurements of each NCT relative to the GAT. Simple linear regression analysis was used to assess the correlations between CCT and the IOP measurements of each tonometer. The paired t-test was used to analyze the difference between the IOP measurements of each NCT and the GAT in each subgroup divided by the GAT IOP values. Data were analyzed using SPSS version 22 (IBM, Armonk, NY, USA).

Results

Of the 52 participants, 22 were male and 30 were female, with a mean age of 50.56 ± 17.25 years (range: 21–85 years). The mean IOP across all subjects was 17.23 ± 2.94 mmHg (range: 11–21 mmHg) with the Canon TX-20P, 14.87 ± 3.25 mmHg (range: 8–21 mmHg) with the Nidek NT-530P, 16.33 ± 3.01 mmHg (range: 9–21 mmHg) with the Topcon CT-1P, and 15.85 ± 3.05 mmHg (range: 8–21 mmHg) with the GAT. The mean CCT as measured with the ultrasound pachymeter was 541.44 ± 28.49 μm (range: 469–605 μm).

IOP comparisons

A significant positive correlation was shown between the IOP values obtained with the GAT (G-IOP) and the Canon TX-20P (C-IOP: R = 0.829, p < 0.001), the Nidek NT-530P (N-IOP: R = 0.799, p < 0.001) and the Topcon CT-1P (T-IOP: R = 0.755, p < 0.001), respectively (Fig. 1). Also, a significantly high consistency by ICC was observed between the G-IOP and the C-IOP (ICC = 0.906, 95% CI: 0.837–0.946, p < 0.001), the N-IOP (ICC = 0.887, 95% CI: 0.803–0.935, p < 0.001) and the T-IOP (ICC = 0.861, 95% CI: 0.757–0.920, p < 0.001), respectively.
Figure 2 shows a Bland-Altman scatterplot comparing the three NCT and GAT readings. The mean of the differences between the C-IOP and G-IOP was 1.38 ± 1.75 mmHg (95% LoA, −2.05 to +4.82 mmHg), between the N-IOP and G-IOP, −0.98 ± 2.00 mmHg (95% LoA, −4.91 to +2.95 mmHg), and between the T-IOP and G-IOP, 0.48 ± 2.12 mmHg (95% LoA, −3.67 to +4.63 mmHg). The narrowest 95% LoA, as indicative of the highest consistency, was that between the C-IOP and G-IOP. There was no significant linear relationship between the difference and the mean of each NCT/GAT pair, which fact indicated good equality between the 3 NCTs and the GAT.
There was a significant mean difference between, respectively, the Canon TX-20P and Nidek NT-530P and GAT IOP measurements but not between the Topcon CT-1P and GAT measurements. The mean difference between the C-IOP and G-IOP was 1.38 mmHg (p < 0.001), and between the N-IOP and G-IOP, 0.98 mmHg (p = 0.001), as noted above (Table 1). There was no significant mean difference between the IOP measurements of each NCT and those of the GAT in six of the nine subgroups divided by G-IOP; significant differences were found in the low- and mid-teen groups of the C-IOP and in the high-teen group of N-IOP. The mean differences between the C-IOP and G-IOP were 1.71 mmHg (p = 0.013) in the low-teen group and 1.77 mmHg (p < 0.001) in the mid-teen group. The mean difference between the N-IOP and G-IOP was 1.75 mmHg (p = 0.001) in the high-teen group (Table 2).
Table 1
Pairwise comparative analyses between each NCT and GAT
NCT device
Mean (±SD) (mmHg)
Mean difference (±SD) with G-IOP
95% LoA
p value
C-IOP
17.23 ± 2.94
1.38 ± 1.75
−2.05 to 4.82
<0.001
N-IOP
14.87 ± 3.25
−0.98 ± 2.00
−4.91 to 2.95
0.001
T-IOP
16.33 ± 3.01
0.48 ± 2.12
−3.67 to 4.63
0.108
C-IOP IOP value measured by Canon NCT, N-IOP IOP value measured by Nidek NCT, T-IOP IOP value measured by Topcon NCT, G-IOP IOP value measured by GAT, LoA limits of agreement
P value by paired t-test
Table 2
Pairwise comparative analyses between each NCT and GAT in subgroups divided by G-IOP
Subgroupsa
n
C-IOP – G-IOP
N-IOP – G-IOP
T-IOP – G-IOP
Low-teen
14
1.71 ± 2.23 ( p = 0.013)
−1.14 ± 2.57 (p = 0.120)
1.14 ± 1.96 (p = 0.048)
Mid-teen
22
1.77 ± 1.34 ( p < 0.001)
−0.32 ± 1.67 (p = 0.382)
1.00 ± 2.07 (p = 0.034)
High-teen
16
0.56 ± 1.59 (p = 0.178)
−1.75 ± 1.65 ( p = 0.001)
−0.81 ± 1.83 (p = 0.097)
C-IOP IOP value measured by Canon NCT, N-IOP Nidek NCT, T-IOP Topcon NCT, G-IOP GAT
Values are presented as the mean ± standard deviation. P value by paired t-test
aCategorized into low-teen (<14 mmHg), mid-teen (≥14 and ≤17 mmHg) and high-teen (>17 mmHg) groups

Influence of CCT

There was a significant positive correlation between CCT and T-IOP (R = 0.390, p = 0.004), the IOP increasing with increasing corneal thickness; linear regression analysis showed a mean change of 0.4 mmHg in T-IOP per 10 μm variation in CCT. Meanwhile, there was a weak positive correlation between CCT and N-IOP (R = 0.325, p = 0.019); linear regression analysis also showed a mean change of 0.4 mmHg in N-IOP per 10 μm variation in CCT. Contrastingly, no correlation was detected between CCT and C-IOP (R = 0.237, p = 0.091) or G-IOP (R = 0.149, p = 0.293), indicating no significant effect of CCT on C-IOP or G-IOP.

Discussion

This study compared the IOP-measurement performances of the three new NCTs with the GAT. With the same subjects at the same moment, the IOPs from each NCT were obtained and demonstrated the following tendencies in terms of range: the IOP measurements with the Canon TX-20P and the Topcon CT-1P tended to be higher, whereas those with the Nidek NT-530P tended to be lower than with the GAT. Considering the order of the tonometer measurements, the greatest IOP values of C-IOP could be due to the ‘massage effect’, though the NCTs were less affected than was the GAT [24, 25], and regardless, the five-minute interval between each measurement, enough time to recover from enhanced aqueous outflow due to corneal compression, would minimize such effect.
In terms of study samples, the mean IOPs recorded using the four tonometers were close to those obtained by other authors for healthy subjects (15–16 mmHg on average, SD 2.5–3.0 mmHg) [3, 27]. Similarly, the mean CCT of 541.4 μm determined using the ultrasound pachymeter was close to the population mean of 530.9 μm and within the range found in normal Korean subjects [28].
The relationships observed between the CCT and IOP measurements obtained by the Nidek NT-530P and Topcon CT-1P tonometers are in agreement with the general assumption that CCT affects NCT readings, showing a positive correlation. According to previous reports, CCT affects both GAT and NCT readings [29, 30], the latter being more affected than the former [3032]. The moderate correlations observed in this study (R = 0.325 and 0.390) could be attributed to the limited range of CCT in our population (469–605 μm). Other authors argue that for corneas with a CCT less than 575 μm, IOP measurements are unaffected [5, 33]. In our study population, only 6 of the 52 eyes showed a CCT greater than 575 μm.
Through Pearson correlation analysis (Fig. 1), we observed a highly significant relationship between the pressure readings provided by each NCT and those offered by the GAT, confirming the good relationships between NCT- and GAT-determined IOPs reported by others [33]. These results indicate that the three NCTs provide statistically significant predictions of G-IOP measurements.
Many variables affect IOP measurements, including astigmatism, corneal thickness, corneal biomechanics/hysteresis, accommodation, respiration, heart rate and rhythm [34], and diurnal variation [26]. Considering these sources of error, some authors advocate that ±3.0 mmHg should be the maximum clinically acceptable error [19]. In our study, according to the Bland-Altman plots, the 95% LoA between C-IOP and G-IOP were close to this clinically acceptable error (±3.43 mmHg), 92% of the differences falling within this range. This means that the N-IOP showed 8% clinical error in our study subjects, all of whose IOP was within the normal range. In case of N-IOP, the LoA were wider (±3.92 mmHg), and 89% of the differences were within the clinically acceptable limits. The LoA of T-IOP was the widest (±4.16 mmHg), and 87% of the differences were within the clinically acceptable limits. These differences of LoA among the three NCTs might be accounted for in the differences in the alignment methods: the Canon TX-20P and Topcon CT-1P make parallel automatically, whereas the Nidek NT-530P aligns manually. Manual alignments might allow the operator to more accurately adjust subjects’ cornea in the three, axial, vertical and lateral dimensions. The reason for the narrower LoA of the Canon TX-20P than that of the Nidek NT-530P should be clarified by further evaluation with a larger subject cohort.
While there was a statistically significant difference between C-IOP and G-IOP as well as between N-IOP and G-IOP, none such was detected between T-IOP and G-IOP, suggesting that any such differences as might exist are independent of the mean IOP values. Thus, we could argue that the Topcon CT-1P offers relatively closer agreement with the GAT than does the Canon TX-20P or Nidek NT-530P. Furthermore, in the subgroup comparison study by G-IOP, the low- and mid-teen subgroups between C-IOP and G-IOP and the high-teen subgroup between N-IOP and G-IOP showed significant mean differences, demonstrating the tendency of discrepancy between the Canon TX-20P and GAT as well as between the Nidek NT-530P and GAT in these G-IOP ranges.
There are some limitations to this study. The fact that we did not test the instruments across the entire useful range is the main one. We also excluded subjects with irregular astigmatism or astigmatism of more than 3.5 diopters, and thus, our findings may hold true only for subjects with similar refraction characteristics. Furthermore, the sample size of this study was relatively small for the subgroup analysis; future studies will need to include larger populations of ocular hypertensive and glaucoma subjects, with or without irregular/severe astigmatism, for whom any bias between instruments could have clinical implications. Additionally, we were unable to randomize the order in which the NCTs were used due to restraints on access to the equipment. Therefore, we were unable to avoid potential systematic errors and are unable to ensure that the lack of randomization did not influence the outcome. We did implement a five-minute interval between each measurement to minimize the ‘massage effect’. One other limitation was the fact that we did not carry out CCT correction separately with each tonopachymeter; future studies should also investigate the results of CCT correction and comparison of CCT values from the ultrasound pachymeter with those from each tonopachymeter.
In conclusion, this comparative study shows that the only Canon TX-20P provides reliable and repeatable IOP measurements not influenced by CCT within a relatively normal/restricted range of corneal thickness. However, without CCT correction, the Canon TX-20P, Nidek NT-530P and Topcon CT-1P all offer similar accuracy to the GAT. Despite the relatively small sample size of this study and the absence of high IOPs, the instruments demonstrated good performance in measuring IOP in ocular normotensive non-glaucomatous subjects. These results suggest the instruments will prove useful for rapid IOP testing in clinical practice. Indeed, practitioners should keep in mind the differences in the tendencies of the IOP values as measured by the respective NCTs.

Conclusion

Without CCT correction, the Canon TX-20P, Nidek NT-530P and Topcon CT-1P all offer similar accuracy to the GAT.

Acknowledgments

A summary of this paper was presented as a narrative at the 115th Spring Meeting of the Korean Ophthalmology Society, Busan, Korea, April 9-10, 2016.

Funding

This work was supported the National Research Foundation of Korea (NRF) grant funded by the Korean government (MSIP; no. 2014R1A5A2010008). The sponsor had no role in the design or conduct of this research.

Availability of data and materials

All data are available upon request.
This was a retrospective study; therefore, informed written informed consent could not be obtained from the patients. Nevertheless, all patient data were anonymized. Moreover, verbal informed consent was obtained from all patients before examination. This study was approved by the Institutional Review Board (IRB) of Keimyung University Dongsan Medical Center (IRB no. 2015–12-052) and was performed in accordance with the tenets of the Declaration of Helsinki.
Not applicable.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Hansen MK. Clinical comparison of the XPERT non-contact tonometer and the conventional Goldmann applanation tonometer. Acta Ophthalmol Scand. 1995;73(2):176–80.CrossRefPubMed Hansen MK. Clinical comparison of the XPERT non-contact tonometer and the conventional Goldmann applanation tonometer. Acta Ophthalmol Scand. 1995;73(2):176–80.CrossRefPubMed
2.
Zurück zum Zitat Beatty S, Nischal KK, Jones H, Eagling EM. Effect of applanation tonometry on mean corneal curvature. J Cataract Refract Surg. 1996;22(7):970–1.CrossRefPubMed Beatty S, Nischal KK, Jones H, Eagling EM. Effect of applanation tonometry on mean corneal curvature. J Cataract Refract Surg. 1996;22(7):970–1.CrossRefPubMed
3.
Zurück zum Zitat Cho P, Lui T. Comparison of the performance of the Nidek NT-2000 noncontact tonometer with the Keeler Pulsair 2000 and the Goldmann applanation tonometer. Optom Vis Sci. 1997;74(1):51–8.CrossRefPubMed Cho P, Lui T. Comparison of the performance of the Nidek NT-2000 noncontact tonometer with the Keeler Pulsair 2000 and the Goldmann applanation tonometer. Optom Vis Sci. 1997;74(1):51–8.CrossRefPubMed
4.
Zurück zum Zitat Jorge J, Diaz-Rey JA, Gonzalez-Meijome JM, Almeida JB, Parafita MA. Clinical performance of the Reichert AT550: a new non-contact tonometer. Ophthalmic Physiol Opt. 2002;22(6):560–4.CrossRefPubMed Jorge J, Diaz-Rey JA, Gonzalez-Meijome JM, Almeida JB, Parafita MA. Clinical performance of the Reichert AT550: a new non-contact tonometer. Ophthalmic Physiol Opt. 2002;22(6):560–4.CrossRefPubMed
5.
Zurück zum Zitat Garcia Resua C, Giraldez Fernandez MJ, Cervino Exposito A, Gonzalez Perez J, Yebra-Pimentel E. Clinical evaluation of the new TGDc-01 "PRA" palpebral tonometer: comparison with contact and non-contact tonometry. Optom Vis Sci. 2005;82(2):143–50.CrossRefPubMed Garcia Resua C, Giraldez Fernandez MJ, Cervino Exposito A, Gonzalez Perez J, Yebra-Pimentel E. Clinical evaluation of the new TGDc-01 "PRA" palpebral tonometer: comparison with contact and non-contact tonometry. Optom Vis Sci. 2005;82(2):143–50.CrossRefPubMed
6.
Zurück zum Zitat van der Jagt LH, Jansonius NM. Three portable tonometers, the TGDc-01, the ICARE and the Tonopen XL, compared with each other and with Goldmann applanation tonometry. Ophthalmic Physiol Opt. 2005;25(5):429–35.CrossRefPubMed van der Jagt LH, Jansonius NM. Three portable tonometers, the TGDc-01, the ICARE and the Tonopen XL, compared with each other and with Goldmann applanation tonometry. Ophthalmic Physiol Opt. 2005;25(5):429–35.CrossRefPubMed
7.
Zurück zum Zitat Shields MB. The non-contact tonometer. Its value and limitations. Surv Ophthalmol. 1980;24(4):211–9.CrossRefPubMed Shields MB. The non-contact tonometer. Its value and limitations. Surv Ophthalmol. 1980;24(4):211–9.CrossRefPubMed
8.
Zurück zum Zitat Moreno-Montanes J, Gomez-Demmel E, Lajara-Blesa J, Aliseda-Perez de Madrid D. Comparative study of three non-contact tonometers and the Goldmann tonometer. Ophthalmologica. 1994;208(3):115–8.CrossRefPubMed Moreno-Montanes J, Gomez-Demmel E, Lajara-Blesa J, Aliseda-Perez de Madrid D. Comparative study of three non-contact tonometers and the Goldmann tonometer. Ophthalmologica. 1994;208(3):115–8.CrossRefPubMed
9.
Zurück zum Zitat Tonnu PA, Ho T, Sharma K, White E, Bunce C, Garway-Heath D. A comparison of four methods of tonometry: method agreement and interobserver variability. Br J Ophthalmol. 2005;89(7):847–50.CrossRefPubMedPubMedCentral Tonnu PA, Ho T, Sharma K, White E, Bunce C, Garway-Heath D. A comparison of four methods of tonometry: method agreement and interobserver variability. Br J Ophthalmol. 2005;89(7):847–50.CrossRefPubMedPubMedCentral
10.
Zurück zum Zitat Jansson U, Brautaset R, Cerviño A, Nilsson M. A comparison of the canon TX-20P™ non-contact tonometer and pachymeter in healthy eyes. Int J Ophthalmic Pract. 2012;3(3):96–102.CrossRef Jansson U, Brautaset R, Cerviño A, Nilsson M. A comparison of the canon TX-20P™ non-contact tonometer and pachymeter in healthy eyes. Int J Ophthalmic Pract. 2012;3(3):96–102.CrossRef
11.
Zurück zum Zitat Tranchina L, Lombardo M, Oddone F, Serrao S, Schiano Lomoriello D, Ducoli P. Influence of corneal biomechanical properties on intraocular pressure differences between an air-puff tonometer and the Goldmann applanation tonometer. J Glaucoma. 2013;22(5):416–21.CrossRefPubMed Tranchina L, Lombardo M, Oddone F, Serrao S, Schiano Lomoriello D, Ducoli P. Influence of corneal biomechanical properties on intraocular pressure differences between an air-puff tonometer and the Goldmann applanation tonometer. J Glaucoma. 2013;22(5):416–21.CrossRefPubMed
12.
Zurück zum Zitat Anton A, Neuburger M, Bohringer D, Jordan JF. Comparative measurement of intraocular pressure by Icare tonometry and Airpuff tonometry in healthy subjects and patients wearing therapeutic soft contact lenses. Graefes Arch Clin Exp Ophthalmol. 2013;251(7):1791–5.CrossRefPubMed Anton A, Neuburger M, Bohringer D, Jordan JF. Comparative measurement of intraocular pressure by Icare tonometry and Airpuff tonometry in healthy subjects and patients wearing therapeutic soft contact lenses. Graefes Arch Clin Exp Ophthalmol. 2013;251(7):1791–5.CrossRefPubMed
13.
Zurück zum Zitat Wei W, Fan Z, Wang L, Li Z, Jiao W, Li Y. Correlation analysis between central corneal thickness and intraocular pressure in juveniles in northern China: the Jinan city eye study. PLoS One. 2014;9(8):e104842.CrossRefPubMedPubMedCentral Wei W, Fan Z, Wang L, Li Z, Jiao W, Li Y. Correlation analysis between central corneal thickness and intraocular pressure in juveniles in northern China: the Jinan city eye study. PLoS One. 2014;9(8):e104842.CrossRefPubMedPubMedCentral
14.
Zurück zum Zitat Garcia-Resua C, Pena-Verdeal H, Minones M, Giraldez MJ, Yebra-Pimentel E. Reliability of the non-contact tono-pachymeter Tonopachy NT-530P in healthy eyes. Clin Exp Optom. 2013;96(3):286–94.CrossRefPubMed Garcia-Resua C, Pena-Verdeal H, Minones M, Giraldez MJ, Yebra-Pimentel E. Reliability of the non-contact tono-pachymeter Tonopachy NT-530P in healthy eyes. Clin Exp Optom. 2013;96(3):286–94.CrossRefPubMed
15.
Zurück zum Zitat Fujimura F, Kamiya K, Fujiwara K, Shoji N, Shimizu K. Repeatability and reproducibility of measurements using a NT-530P noncontact tono/pachymeter and correlation of central corneal thickness with intraocular pressure. Biomed Res Int. 2013;2013:370592.CrossRefPubMedPubMedCentral Fujimura F, Kamiya K, Fujiwara K, Shoji N, Shimizu K. Repeatability and reproducibility of measurements using a NT-530P noncontact tono/pachymeter and correlation of central corneal thickness with intraocular pressure. Biomed Res Int. 2013;2013:370592.CrossRefPubMedPubMedCentral
16.
Zurück zum Zitat Ogbuehi KC. Assessment of the accuracy and reliability of the Topcon CT80 non-contact tonometer. Clin Exp Optom. 2006;89(5):310–4.CrossRefPubMed Ogbuehi KC. Assessment of the accuracy and reliability of the Topcon CT80 non-contact tonometer. Clin Exp Optom. 2006;89(5):310–4.CrossRefPubMed
17.
Zurück zum Zitat Farhood QK. Comparative evaluation of intraocular pressure with an air-puff tonometer versus a Goldmann applanation tonometer. Clin Ophthalmol. 2013;7:23–7.PubMed Farhood QK. Comparative evaluation of intraocular pressure with an air-puff tonometer versus a Goldmann applanation tonometer. Clin Ophthalmol. 2013;7:23–7.PubMed
18.
Zurück zum Zitat Tamcelik N, Atalay E, Cicik E, Ozkok A. Comparability of Icare pro rebound Tonometer with Goldmann Applanation and noncontact Tonometer in a wide range of intraocular pressure and central corneal thickness. Ophthalmic Res. 2015;54(1):18–25.CrossRefPubMed Tamcelik N, Atalay E, Cicik E, Ozkok A. Comparability of Icare pro rebound Tonometer with Goldmann Applanation and noncontact Tonometer in a wide range of intraocular pressure and central corneal thickness. Ophthalmic Res. 2015;54(1):18–25.CrossRefPubMed
19.
Zurück zum Zitat Garcia-Resua C, Giraldez Fernandez MJ, Yebra-Pimentel E, Garcia-Montero S. Clinical evaluation of the canon TX-10 noncontact tonometer in healthy eyes. Eur J Ophthalmol. 2010;20(3):523–30.PubMed Garcia-Resua C, Giraldez Fernandez MJ, Yebra-Pimentel E, Garcia-Montero S. Clinical evaluation of the canon TX-10 noncontact tonometer in healthy eyes. Eur J Ophthalmol. 2010;20(3):523–30.PubMed
20.
Zurück zum Zitat Pelit A, Altan-Yaycioglu R, Pelit A, Akova YA. Effect of corneal thickness on intraocular pressure measurements with the Pascal dynamic contour, canon TX-10 non-contact and Goldmann applanation tonometers in healthy subjects. Clin Exp Optom. 2009;92(1):14–8.CrossRefPubMed Pelit A, Altan-Yaycioglu R, Pelit A, Akova YA. Effect of corneal thickness on intraocular pressure measurements with the Pascal dynamic contour, canon TX-10 non-contact and Goldmann applanation tonometers in healthy subjects. Clin Exp Optom. 2009;92(1):14–8.CrossRefPubMed
21.
Zurück zum Zitat Yilmaz I, Altan C, Aygit ED, Alagoz C, Baz O, Ahmet S, Urvasizoglu S, Yasa D, Demirok A. Comparison of three methods of tonometry in normal subjects: Goldmann applanation tonometer, non-contact airpuff tonometer, and Tono-pen XL. Clin Ophthalmol. 2014;8:1069–74.CrossRefPubMedPubMedCentral Yilmaz I, Altan C, Aygit ED, Alagoz C, Baz O, Ahmet S, Urvasizoglu S, Yasa D, Demirok A. Comparison of three methods of tonometry in normal subjects: Goldmann applanation tonometer, non-contact airpuff tonometer, and Tono-pen XL. Clin Ophthalmol. 2014;8:1069–74.CrossRefPubMedPubMedCentral
22.
Zurück zum Zitat Esen F, Kostek M, Emekli AS, Eraslan M. Double-organ bias in published randomized controlled trials of glaucoma. J Glaucoma. 2016;25(6):520–2.CrossRefPubMed Esen F, Kostek M, Emekli AS, Eraslan M. Double-organ bias in published randomized controlled trials of glaucoma. J Glaucoma. 2016;25(6):520–2.CrossRefPubMed
23.
Zurück zum Zitat Almubrad TM, Ogbuehi KC. Clinical investigation of the effect of topical anesthesia on intraocular pressure. Clin Ophthalmol. 2007;1(3):305–9.PubMedPubMedCentral Almubrad TM, Ogbuehi KC. Clinical investigation of the effect of topical anesthesia on intraocular pressure. Clin Ophthalmol. 2007;1(3):305–9.PubMedPubMedCentral
24.
Zurück zum Zitat AlMubrad TM, Ogbuehi KC. The effect of repeated applanation on subsequent IOP measurements. Clin Exp Optom. 2008;91(6):524–9.CrossRefPubMed AlMubrad TM, Ogbuehi KC. The effect of repeated applanation on subsequent IOP measurements. Clin Exp Optom. 2008;91(6):524–9.CrossRefPubMed
25.
Zurück zum Zitat Almubrad TM, Ogbuehi KC. On repeated corneal applanation with the Goldmann and two non-contact tonometers. Clin Exp Optom. 2010;93(2):77–82.CrossRefPubMed Almubrad TM, Ogbuehi KC. On repeated corneal applanation with the Goldmann and two non-contact tonometers. Clin Exp Optom. 2010;93(2):77–82.CrossRefPubMed
26.
Zurück zum Zitat Liu JH, Bouligny RP, Kripke DF, Weinreb RN. Nocturnal elevation of intraocular pressure is detectable in the sitting position. Invest Ophthalmol Vis Sci. 2003;44(10):4439–42.CrossRefPubMed Liu JH, Bouligny RP, Kripke DF, Weinreb RN. Nocturnal elevation of intraocular pressure is detectable in the sitting position. Invest Ophthalmol Vis Sci. 2003;44(10):4439–42.CrossRefPubMed
27.
Zurück zum Zitat Doughty MJ, Zaman ML. Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol. 2000;44(5):367–408.CrossRefPubMed Doughty MJ, Zaman ML. Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol. 2000;44(5):367–408.CrossRefPubMed
28.
Zurück zum Zitat Hwang YH, Kim HK, Sohn YH, Namil Study Group KGS. Central corneal thickness in a Korean population: the Namil study. Invest Ophthalmol Vis Sci. 2012;53(11):6851–5.CrossRefPubMed Hwang YH, Kim HK, Sohn YH, Namil Study Group KGS. Central corneal thickness in a Korean population: the Namil study. Invest Ophthalmol Vis Sci. 2012;53(11):6851–5.CrossRefPubMed
29.
Zurück zum Zitat Regine F, Scuderi GL, Cesareo M, Ricci F, Cedrone C, Nucci C. Validity and limitations of the Nidek NT-4000 non-contact tonometer: a clinical study. Ophthalmic Physiol Opt. 2006;26(1):33–9.CrossRefPubMed Regine F, Scuderi GL, Cesareo M, Ricci F, Cedrone C, Nucci C. Validity and limitations of the Nidek NT-4000 non-contact tonometer: a clinical study. Ophthalmic Physiol Opt. 2006;26(1):33–9.CrossRefPubMed
30.
Zurück zum Zitat Recep OF, Hasiripi H, Cagil N, Sarikatipoglu H. Relation between corneal thickness and intraocular pressure measurement by noncontact and applanation tonometry. J Cataract Refract Surg. 2001;27(11):1787–91.CrossRefPubMed Recep OF, Hasiripi H, Cagil N, Sarikatipoglu H. Relation between corneal thickness and intraocular pressure measurement by noncontact and applanation tonometry. J Cataract Refract Surg. 2001;27(11):1787–91.CrossRefPubMed
31.
Zurück zum Zitat Stodtmeister R. Applanation tonometry and correction according to corneal thickness. Acta Ophthalmol Scand. 1998;76(3):319–24.CrossRefPubMed Stodtmeister R. Applanation tonometry and correction according to corneal thickness. Acta Ophthalmol Scand. 1998;76(3):319–24.CrossRefPubMed
32.
Zurück zum Zitat Babalola OE, Kehinde AV, Iloegbunam AC, Akinbinu T, Moghalu C, Onuoha I. A comparison of the Goldmann applanation and non-contact (Keeler Pulsair EasyEye) tonometers and the effect of central corneal thickness in indigenous African eyes. Ophthalmic Physiol Opt. 2009;29(2):182–8.CrossRefPubMed Babalola OE, Kehinde AV, Iloegbunam AC, Akinbinu T, Moghalu C, Onuoha I. A comparison of the Goldmann applanation and non-contact (Keeler Pulsair EasyEye) tonometers and the effect of central corneal thickness in indigenous African eyes. Ophthalmic Physiol Opt. 2009;29(2):182–8.CrossRefPubMed
33.
Zurück zum Zitat Lee YG, Kim JH, Kim NR, Kim CY, Lee ES. Comparison between Tonopachy and other tonometric and pachymetric devices. Optom Vis Sci. 2011;88(7):843–9.CrossRefPubMed Lee YG, Kim JH, Kim NR, Kim CY, Lee ES. Comparison between Tonopachy and other tonometric and pachymetric devices. Optom Vis Sci. 2011;88(7):843–9.CrossRefPubMed
34.
Zurück zum Zitat Whitacre MM, Stein R. Sources of error with use of Goldmann-type tonometers. Surv Ophthalmol. 1993;38(1):1–30.CrossRefPubMed Whitacre MM, Stein R. Sources of error with use of Goldmann-type tonometers. Surv Ophthalmol. 1993;38(1):1–30.CrossRefPubMed
Metadaten
Titel
Comparison of intraocular pressure as measured by three different non-contact tonometers and goldmann applanation tonometer for non-glaucomatous subjects
verfasst von
Seung Pil Bang
Chong Eun Lee
Yu Cheol Kim
Publikationsdatum
01.12.2017
Verlag
BioMed Central
Erschienen in
BMC Ophthalmology / Ausgabe 1/2017
Elektronische ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-017-0593-1

Weitere Artikel der Ausgabe 1/2017

BMC Ophthalmology 1/2017 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.