27.02.2019 | Hepatobiliary

Comparison of navigator-gated and breath-held image acquisition techniques for multi-echo quantitative dixon imaging of the liver in children and young adults

Zeitschrift:
Abdominal Radiology
Autoren:
Leah A. Gilligan, Jonathan R. Dillman, Jean A. Tkach, Andrew T. Trout
Wichtige Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Abstract

Purpose

Acquired over a breath hold, multi-echo Dixon (mDixon) magnetic resonance imaging (MRI) of the liver can be used to quantify proton density fat fraction (PDFF) and iron-related signal decay. However, young, obese, and co-morbid patients may have limited breath holding capacity and could benefit from a motion-robust mDixon acquisition. The purpose of this study was to compare hepatic PDFF and R2* values between navigator-gated and breath-held mDixon MRI acquisition techniques in children and young adults with suspected liver disease.

Materials and methods

This retrospective study was institutional review board-approved with a waiver of informed consent. Patients who underwent liver MRI with breath-held and navigator-gated mDixon sequences between January 2017 and July 2018 were included. One reviewer, blinded to sequence, measured PDFF and R2* on four images from each sequence. Another blinded reviewer graded respiratory motion (5-point Likert scale). Pearson correlation (r), Lin’s concordance coefficients (rc), and Bland–Altman analyses were used to assess agreement between techniques. Frequency of clinically limiting motion (score ≥ 3) was compared with Fisher’s exact test.

Results

Forty-two patients were included (15 female, 27 male; mean age: 15.7 ± 4.6 years). Mean PDFF and R2* were 16.6 ± 13.1% and 29.3 ± 4.7 s−1 (breath-held) versus 17.0 ± 13.2% and 29.6 ± 5.2 s−1 (navigator-gated). PDFF agreed almost perfectly between sequences (rc = 0.997, 95% CI 0.994–0.998; mean bias: 0.3%; 95% limits of agreement: − 2.4 to +1.7%), while R2* values correlated very strongly but with poor agreement (r = 0.837, rc = 0.832, 95% CI 0.716–0.910). Navigator-gated images exhibited significantly higher frequency of clinically limiting respiratory motion (88% vs. 48%, p = 0.0001).

Conclusion

Despite greater respiratory motion artifact, a free-breathing navigator-gated mDixon sequence produces PDFF values with almost perfect agreement to a breath-held sequence.

Bitte loggen Sie sich ein, um Zugang zu diesem Inhalt zu erhalten

★ PREMIUM-INHALT
e.Med Interdisziplinär

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de. Zusätzlich können Sie eine Zeitschrift Ihrer Wahl in gedruckter Form beziehen – ohne Aufpreis.

Weitere Produktempfehlungen anzeigen
Literatur
Über diesen Artikel
  1. Sie können e.Med Radiologie 14 Tage kostenlos testen (keine Print-Zeitschrift enthalten). Der Test läuft automatisch und formlos aus. Es kann nur einmal getestet werden.

Neu im Fachgebiet Radiologie

Meistgelesene Bücher aus der Radiologie

2016 | Buch

Medizinische Fremdkörper in der Bildgebung

Thorax, Abdomen, Gefäße und Kinder

Dieses einzigartige Buch enthält ca. 1.600 hochwertige radiologische Abbildungen und Fotos iatrogen eingebrachter Fremdmaterialien im Röntgenbild und CT.

Herausgeber:
Dr. med. Daniela Kildal

2011 | Buch

Atlas Klinische Neuroradiologie des Gehirns

Radiologie lebt von Bildern! Der vorliegende Atlas trägt dieser Tatsache Rechnung. Sie finden zu jedem Krankheitsbild des Gehirns Referenzbilder zum Abgleichen mit eigenen Befunden.

Autoren:
Priv.-Doz. Dr. med. Jennifer Linn, Prof. Dr. med. Martin Wiesmann, Prof. Dr. med. Hartmut Brückmann

Mail Icon II Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Radiologie und bleiben Sie gut informiert – ganz bequem per eMail.

Bildnachweise