Comparison of short axis and long axis acquisitions of T1 and extracellular volume mapping using MOLLI and SASHA in patients with myocardial infarction and healthy volunteers

verfasst von: Christos G. Xanthis, David Nordlund, Robert Jablonowski, Håkan Arheden

Erschienen in: BMC Medical Imaging | Ausgabe 1/2019

Abstract

Background

Although previous studies have examined the impact of slice position in volumetric measurements in Cardiovascular Magnetic Resonance (CMR) imaging, very limited data are available today comparing T1 and Extra-Cellular Volume (ECV) measurements from short and long axis acquisitions. The purpose of this study was to investigate the impact of slice position and orientation on T1 and ECV measurements using the MOdified Look-Locker Inversion recovery (MOLLI) and Saturation recovery single-shot acquisition (SASHA) sequence in patients with myocardial infarction and in healthy volunteers.

Methods

Eight (8) healthy volunteers with no medical history and eight (8) patients with myocardial infarction were included in this study. MOLLI and SASHA were utilized and short-axis and long-axis images were acquired. T1 and ECV measurements were performed by drawing same size regions of interest on the myocardium as well in the blood pool at the intersections of the short axis and long axis images.

Results

In healthy volunteers, there were no statistically significant differences in native T1 and ECV values between short axis and long axis acquisitions using MOLLI (two-chamber, three-chamber and four-chamber) and SASHA (three-chamber). In patients, there were no statistically significant differences in native T1 and ECV values between short axis and 3-chamber long axis acquisitions in both remote and affected myocardium using MOLLI and SASHA.

Conclusions

Long axis measurements of myocardial T1 and ECV using MOLLI and SASHA exhibit good agreement with the corresponding short axis measurements allowing for fast and reliable myocardial tissue characterization in cases where shortening of the overall imaging acquisition is required.
Abkürzungen
bSSFP
Balanced Steady-State Free Precession
CMR
Cardiovascular Magnetic Resonance
ECV
Extracellular volume
LAX
Long axis
MOLLI
MOdified Look-Locker Inversion recovery
ROI
Region of Interest
SASHA
Saturation recovery single-shot acquisition
SAX
Short axis

Background

In the field of Cardiovascular Magnetic Resonance (CMR) imaging, quantitative measures of myocardial and blood T1 before and after contrast injection enabled the calculation of myocardial extracellular volume fraction (ECV), an important myocardial biomarker [1, 2]. Several studies have demonstrated the potential of ECV measurement in the assessment of a variety of myocardial pathologies [3, 4] and in the guidance of therapy [5].
Myocardial and blood T1 values are most often determined on short-axis images of the left ventricle of the heart, usually including one apical, one midventricular and one basal slice. Recent advancements in the field of cardiac T1 mapping have allowed for fast generation of a single-slice T1 map within a single breath-hold acquisition. Today, the most commonly used T1 mapping technique in CMR is the MOdified Look-Locker Inversion recovery (MOLLI) [6] whereas the Saturation recovery single-shot acquisition (SASHA) [7] T1 mapping technique has been proposed as a means of mitigating the T1-underestimation in MOLLI [7, 8].
It has previously been shown that several factors affect the performance of MOLLI and SASHA on T1 and ECV mapping in terms of accuracy and precision [9]. Different T1 mapping methods and parameter sets result in different ranges of T1 and ECV values for normal myocardium and blood [8, 10, 11]. Although previous studies have examined the impact of slice position in measuring other cardiac parameters in CMR imaging (such as left ventricular mass, volume and ejection fraction) [1214], very limited data are available today comparing T1 [15] and ECV measurements from short and long axis acquisitions [16, 17]. Despite this, current recommendations on T1 and ECV mapping in patients with global/diffuse cardiac disease [18] suggest adding a long axis map to aid analysis. Therefore, the specific aim of this study was to investigate the impact of slice position and orientation on T1 and ECV measurements. MOLLI and SASHA sequences were utilized in healthy volunteers and in patients with myocardial infarction to investigate the impact of slice position and orientation on T1 and ECV measurements.

Methods

Eight (8) healthy volunteers with no medical history (5 men, 3 women, age 25 ± 5 years) and eight (8) patients (7 men, 1 woman, age 66 ± 10 years) with myocardial infarction and without any renal impairment were included in this study. The study was approved by the regional ethics committee and all subjects provided written informed consent (The regional ethics committee, Lund, Sweden. Ethics applications numbers: 541/2004 and 815/2016). Blood for hematocrit analysis was sampled from a peripheral vein from the subjects approximately 30 min after lying down, just before Gd-contrast administration.

MR protocol

All subjects underwent CMR on a MAGNETOM Aera 1.5 T scanner (Siemens Healthcare, Erlangen, Germany) using a 30-channel coil (body array and spine array). In healthy volunteers, a prototype MOLLI sequence was used to acquire a single midventricular short-axis image and three long-axis images (two-chamber, three-chamber and four-chamber) whereas a prototype SASHA sequence was used to acquire a midventricular short-axis image and a single long-axis (three-chamber) image. Pre-contrast MOLLI T1 mapping was performed using an acquisition scheme of 5s(3s)3s whereas post-contrast MOLLI T1 mapping was performed using an acquisition scheme of 4s(1s)3s(1s)2s. The SASHA scheme remained the same before and after contrast administration. Previous studies [7, 8] have shown that these pulse sequences are heart-rate independent. Post-contrast mapping was performed approximately 30 min after injection of 0.2 mmol/kg Gd-DOTA (Dotarem, Guerbet, Roissy, France). In patients, the same MR protocol was utilized for the acquisition of a single midventricular short-axis image and a single long-axis image (two-chamber or three-chamber). MOLLI and SASHA T1 maps were acquired at the same slice locations.

Image analysis

The relaxation time parameters were estimated through a ROI-based curve fitting on the in-line, motion-corrected image series derived from both pre- and post-contrast MOLLI and SASHA acquisitions. All images were analyzed using the software Segment, version 2.0R5453 (http://​segment.​heiberg.​se) [19]. T1 measurements were performed by drawing same size regions of interest (ROIs) on the myocardium (ROI area: 0.1 cm2) as well in the blood pool (ROI area: 0.8 cm2) at the intersections of the short axis and long axis images (Fig. 1), therefore the same tissue area was evaluated twice (once in short axis and once in long axis view). In healthy volunteers, the ROIs were placed at the center of the myocardial wall and special care was taken so as to avoid signal contamination from adjacent blood. In patients, ROIs of infarction and myocardium-at-risk (MaR) were considered as affected myocardium. Contrast enhanced SSFP (CE-SSFP) and late gadolinium enhancement (LGE) images were used to detect regional myocardial edema and fibrosis. Special care was taken so as to place the ROIs within a single tissue type area (remote, edema or fibrosis) and avoid signal contamination from adjacent tissue types. The placement of the ROIs was performed by an experienced reviewer (DN: 5 years of CMR experience). T1-values were measured both before and after a gadolinium (Gd) based contrast injection and myocardial ECV was calculated according to the following Eq. [1]:
$$ Myocardial\ ECV=\left(1- Hct\right)\frac{1/ Myocardial\ T{1}_{post\ contrast}-1/ Myocardial\ T{1}_{pre\ contrast}}{1/ Blood\ T{1}_{post\ contrast}-1/ Blood\ T{1}_{pre\ contrast}} $$
(1)

Statistical analysis

Graphpad Prism version 7.03 (Graphpad Software Inc. La Jolla, USA) was used to perform statistical analysis. Values are presented as mean ± SD. Student’s two tailed t-test for paired data was utilized for comparison of different acquisitions. Differences with a p-value < 0.05 were considered to be statistically significant. Bland-Altman plots were also used to compare the short-axis native T1 and ECV values against the T1 values and ECV values extracted from the corresponding long-axis image.

Results

SAX vs LAX T1 and ECV values in healthy volunteers using MOLLI and SASHA

In healthy volunteers, there were no statistically significant differences in native T1 and ECV values between short axis and long axis acquisitions (p > 0.05) using MOLLI acquisitions. Figure 2 presents the mean native T1 values derived from the ROIs placed in three LAX slices (2 chamber view, 3 chamber view and 4 chamber view) against the corresponding ROIs in the midventricular SAX plane. The mean native T1 values were 999 ± 58 ms vs. 982 ± 61 ms (SAX vs LAX 2ch, n = 16, p = 0.46), 963 ± 35 ms vs. 950 ± 51 ms (SAX vs LAX 3ch, n = 16, p = 0.35) and 973 ± 41 ms vs. 981 ± 46 ms (SAX vs LAX 4ch, n = 16, p = 0.53). The corresponding Bland-Altman plots for native T1 measurement in SAX view against the 3-chamber and 4 chamber LAX views are shown in Fig. 3. Figure 4 presents the mean ECV values derived from the ROIs placed in the same three LAX slices against the corresponding ROIs in the midventricular SAX plane. The mean ECV values were 27.5 ± 6.1% vs. 28.2 ± 5.1% (SAX vs LAX 2ch, n = 16, p = 0.59), 24.9 ± 4% vs. 24.2 ± 3.4% (SAX vs LAX 3ch, n = 16, p = 0.36) and 26.2 ± 3.2% vs. 26.3 ± 3.4% (SAX vs LAX 4ch, n = 16, p = 0.83). The corresponding Bland-Altman plots for ECV measurement in SAX view against the 3-chamber and 4 chamber LAX views are shown in Fig. 5.
Native T1 and ECV measurements in SAX and 3-chamber LAX views using MOLLI and SASHA acquisitions are presented in Figs. 6 and 7 respectively. The mean native T1 values were 963 ± 35 ms vs. 950 ± 51 ms (SAX vs LAX 3ch, n = 16, p = 0.35) for MOLLI whereas the corresponding mean T1 values for SASHA were 1181 ± 38 ms vs. 1215 ± 83 ms (SAX vs LAX 3ch, n = 14, p = 0.13). The mean ECV values were 24.9 ± 4% vs. 24.2 ± 3.4% (SAX vs LAX 3ch, n = 16, p = 0.36) for MOLLI whereas the corresponding mean ECV values for SASHA were 22.8 ± 2.5% vs. 21.2 ± 2.6% (SAX vs LAX 3ch, n = 14, p = 0.07). The corresponding Bland-Altman plots for ECV measurement in SAX view against the 3-chamber LAX view using SASHA is shown in Fig. 8. In one volunteer data acquired using SASHA was excluded due to motion artifacts. No registration distortion [20] was observed in the motion corrected, T1-weighted image series derived from both MOLLI and SASHA acquisitions in healthy volunteers.

SAX vs LAX T1 and ECV values in patients using MOLLI and SASHA

In patients, there were no statistically significant differences in native T1 and ECV values between short axis and 3-chamber long axis acquisitions (p > 0.05). The mean native T1 values extracted from ROIs placed in remote myocardium were 982 ± 79 ms vs. 981 ± 99 ms (SAX vs LAX 3ch, n = 7, p = 0.95) for MOLLI whereas the corresponding mean T1 values for SASHA were 1083 ± 147 ms vs. 1154 ± 145 ms (SAX vs LAX 3ch, n = 6, p = 0.14). The mean ECV values from the same ROIs placed in remote myocardium were 24.4 ± 3.9% vs. 24.6 ± 3.8% (SAX vs LAX 3ch, n = 7, p = 0.88) for MOLLI whereas the corresponding mean ECV values for SASHA were 20.2 ± 3.1% vs. 20.1 ± 5.3% (SAX vs LAX 3ch, n = 6, p = 0.94). In affected myocardium, the mean native T1 values were 1212 ± 106 ms vs. 1199 ± 113 ms (SAX vs LAX 3ch, n = 9, p = 0.48) for MOLLI whereas the corresponding mean T1 values for SASHA were 1407 ± 87 ms vs. 1416 ± 139 ms (SAX vs LAX 3ch, n = 9, p = 0.82). The mean ECV values from the same ROIs placed in affected myocardium were 46.2 ± 8.9% vs. 50.6 ± 14.7% (SAX vs LAX 3ch, n = 9, p = 0.28) for MOLLI whereas the corresponding mean ECV values for SASHA were 44.3 ± 11.7% vs. 41.3 ± 10.2% (SAX vs LAX 3ch, n = 9, p = 0.43). Figures 9 presents the mean T1 values extracted from remote and affected myocardium in SAX and 3-chamber LAX views using MOLLI and SASHA. The corresponding ECV values are shown in Fig. 10. In one patient data acquired using SASHA was excluded due to motion artifacts. No registration distortion [20] was observed in the motion corrected, T1-weighted image series derived from both MOLLI and SASHA acquisitions in patients.
Lastly, as Fig. 9 shows, there was a statistically significant difference (p < 0.05) in T1 measurements between remote and diseased myocardium in both MOLLI and SASHA for both short axis and 3-chamber long axis acquisitions (MOLLI: SAX p = 0.0014, LAX p = 0.0026 – SASHA: SAX p = 0.0025, LAX p = 0.0019). However, there was no statistically significant difference (p > 0.05) between the MOLLI-based T1 measurements of diseased myocardium and the SASHA-based T1 measurements of remote myocardium for both short axis (p = 0.06) and 3-chamber long axis (p = 0.26) acquisitions.

Discussion

This study investigates T1 and ECV quantified from long-axis acquisitions compared to short-axis acquisitions using MOLLI and SASHA in healthy volunteers and patients. Long-axis acquisitions showed no statistically significant differences in native T1 and ECV values compared to short-axis acquisitions using both MOLLI and SASHA methods. The two-chamber long axis acquisition presented the lowest agreement with the short-axis acquisition compared to the other two long axis acquisitions. Both MOLLI and SASHA showed similarly tight limits of agreement when comparing the ECV measurements taken with short axis and three-chamber long axis acquisitions although SASHA showed larger bias than MOLLI. Lastly, the statistical hypothesis testing presented in Fig. 9 indicates that MOLLI and SASHA should not be used interchangeably in T1 mapping for characterizing different tissue types in myocardium.
In the last several years, CMR parametric mapping has been used as a non-invasive tool for quantifying myocardial tissue alternations in myocardial disease. Recently, the Society for Cardiovascular Magnetic Resonance (SCMR) endorsed by the European Association for Cardiovascular Imaging (EACVI) published a consensus statement with recommendations on T1 and ECV mapping [18]. The guidelines suggested that an optional single long axis map should be acquired in cases of global/diffuse diseases whereas acquisition of at least one long axis map is considered mandatory in cases of patchy diseases. Nowadays, MOLLI is mainly used as the preferred technique in cardiac T1 and ECV mapping [21] whereas SASHA has been proposed in the literature as a means of mitigating the T1-underestimation in MOLLI [7, 8]. However, there is limited work today in the literature that has evaluated the performance of T1 and ECV mapping in myocardial tissue characterization using long-axis against short-axis acquisitions. Moreover, although SASHA is actively being studied today, its performance in T1 and ECV measurements under different slice orientations has not been investigated before.
This study adds on the research that other groups have already performed in order to investigate the impact of slice position in CMR imaging. In particular, comparison of short and long axis methods had been the focus of other previous studies in CMR for volumetric measurements. Harizolan et al. [13] showed that there was no significant difference in both patient with myocardial infarction and control groups between end-diastolic volume determined from short axis slice and end-diastolic volume determined by 2-chamber, 3-chamber and 4-chamber long axis slices. In a more recent study, Huttin et al. [14] showed that measurements of left-ventricle volumes and ejection fraction utilizing a biplane long axis MRI study allowed for accurate, fast and reliable assessment of left-ventricle function and exhibited good correlation with the short axis measurements. Despite that the long axis acquisition presented a systematic underestimation of left-ventricles volumes compared to the short axis acquisition, the authors suggested that a biplane long axis MRI study should be considered to shorten the overall imaging acquisition in an acute clinical setting.
In the field of T1 and ECV mapping, results of the current work were similar to results previously reported in the literature. Nacif et al. [15] showed that there was no significant difference in global myocardial T1 values between four-chamber long-axis and mid-ventricular short-axis measurements using the MOLLI T1 mapping technique on a group of healthy volunteers. In another study, Bohnen et al. [16] compared global myocardial T1 and ECV values between short-axis and three different long-axis slices (two-chamber, three-chamber and four-chamber views) on patients with suspected myocarditis. Although the authors showed that there were no significant differences on ECV measurements between short and long axis slices, significantly lower median native myocardial T1 values on long axis slices were reported compared to short-axis slices. The latter was attributed to issues related to slice orientation (such as through-plane motion and partial volume effects) but also to the heterogeneity of myocardial injury in myocarditis. In the current study, although no significant differences on T1 and ECV measurements were shown between short and long axis slices using MOLLI, Figs. 2 and 4 presented a larger intra-slice variability of the MOLLI-based T1 and ECV estimates within the midventricular short axis image compared to the inter-slice variability (short axis vs long axis). In a similar manner, previous studies [11, 22, 23] have demonstrated significant regional variations of native T1 values in SAX slices of normal subjects. These differences were not considered representative of a true difference in tissue composition but were attributed to other factors, such as inadequate B0-shimming around the heart (off-resonance issues) [9], receiver coil sensitivity and distance of the receiver coil elements from the region of interest [24].
Compared to the previous two studies [15, 16], the current study did not measure the mean T1 and ECV values of the entire myocardium within the slice to evaluate the performance of long-axis acquisitions against the short-axis acquisitions. In this study, T1 measurements and ECV calculations were performed using an ROI-based analysis by drawing same size ROIs on the myocardium as well in the blood pool at the intersections of the short axis and long axis images. This approach was considered more representative for evaluating the differences between different slice orientations since it eliminates any T1 variability caused by biological focal abnormalities in the myocardium and enhances the investigation of any T1 variability caused by the technical design of the quantitative approach. In a similar approach, Caballeros et al. [17] showed that there was no significant difference in myocardial T1 values and ECV measurement between short-axis and four-chamber long axis analysis on groups of patients with various diseases.

Limitations

In this study, some limitations apply. The number of subjects is small in both groups. Larger scale studies are required to detect potential subtle differences between different acquisition techniques in T1 and ECV measurements. Moreover, the design of the current study does not allow for a direct comparison between MOLLI and SASHA on the inter-slice variability of the T1 and ECV estimates using the two-chamber and four-chamber LAX images since these images were not acquired neither in healthy volunteers nor in patients. Lastly, the post contrast myocardial T1 values were not presented in this study. Post contrast T1 mapping is considered more variable and depend on several factors such as time elapsed between contrast agent administration and renal clearance [25].

Conclusions

In conclusion, long axis measurements of myocardial T1 and ECV using MOLLI and SASHA exhibit good agreement with the corresponding short axis measurements allowing for fast and reliable myocardial tissue characterization. This may be of high importance in clinical cases where shortening of the overall imaging acquisition is required. Moreover, the ROI-based design of the current study may be utilized in other studies that are focused on myocardial tissue characterization in order to evaluate the differences between different slice orientations, especially in cases with focal native T1 abnormalities.

Acknowledgements

The authors would like to acknowledge the Master Research Agreement between Skåne University Hospital and Siemens Healthcare as well as the technical support provided in this context by Drs. Andreas Greiser and Kelvin Chow for providing the prototype sequence.

Funding

The Medical Faculty of Lund University, The Swedish Heart and Lung Foundation, Region of Scania, Sweden and Skåne University Hospital, Sweden. These funding sources had no role in the design of this study and in the collection, analysis, and interpretation of data and in writing the manuscript.

Availability of data and materials

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.
The study was approved by the Regional ethics committee and all subjects provided written informed consent (The regional ethics committee, Lund, Sweden. Ethics applications numbers: 541/2004 and 815/2016).
All subjects provided written consent for publication of this study and accompanying images.

Competing interests

The authors declare that they have no competing interests in relation to the content of this study.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Arheden H, et al. Measurement of the distribution volume of gadopentetate dimeglumine at echo-planar MR imaging to quantify myocardial infarction: comparison with 99mTc-DTPA autoradiography in rats. Radiology. 1999;211(3):698–708.PubMedCrossRef Arheden H, et al. Measurement of the distribution volume of gadopentetate dimeglumine at echo-planar MR imaging to quantify myocardial infarction: comparison with 99mTc-DTPA autoradiography in rats. Radiology. 1999;211(3):698–708.PubMedCrossRef
2.
Zurück zum Zitat Arheden H, et al. Reperfused rat myocardium subjected to various durations of ischemia: estimation of the distribution volume of contrast material with echo-planar MR imaging. Radiology. 2000;215(2):520–8.PubMedCrossRef Arheden H, et al. Reperfused rat myocardium subjected to various durations of ischemia: estimation of the distribution volume of contrast material with echo-planar MR imaging. Radiology. 2000;215(2):520–8.PubMedCrossRef
3.
Zurück zum Zitat Kellman P, et al. Extracellular volume fraction mapping in the myocardium, part 2: initial clinical experience. J Cardiovasc Magn Reson. 2012;14:64.PubMedPubMedCentralCrossRef Kellman P, et al. Extracellular volume fraction mapping in the myocardium, part 2: initial clinical experience. J Cardiovasc Magn Reson. 2012;14:64.PubMedPubMedCentralCrossRef
4.
Zurück zum Zitat Kellman P, et al. Extracellular volume fraction mapping in the myocardium, part 1: evaluation of an automated method. J Cardiovasc Magn Reson. 2012;14:63.PubMedPubMedCentralCrossRef Kellman P, et al. Extracellular volume fraction mapping in the myocardium, part 1: evaluation of an automated method. J Cardiovasc Magn Reson. 2012;14:63.PubMedPubMedCentralCrossRef
5.
Zurück zum Zitat Moon JC, et al. Myocardial T1 mapping and extracellular volume quantification: a Society for Cardiovascular Magnetic Resonance (SCMR) and CMR working Group of the European Society of cardiology consensus statement. J Cardiovasc Magn Reson. 2013;15:92.PubMedPubMedCentralCrossRef Moon JC, et al. Myocardial T1 mapping and extracellular volume quantification: a Society for Cardiovascular Magnetic Resonance (SCMR) and CMR working Group of the European Society of cardiology consensus statement. J Cardiovasc Magn Reson. 2013;15:92.PubMedPubMedCentralCrossRef
6.
Zurück zum Zitat Messroghli DR, et al. Modified look-locker inversion recovery (MOLLI) for high-resolution T1 mapping of the heart. Magn Reson Med. 2004;52(1):141–6.PubMedCrossRef Messroghli DR, et al. Modified look-locker inversion recovery (MOLLI) for high-resolution T1 mapping of the heart. Magn Reson Med. 2004;52(1):141–6.PubMedCrossRef
7.
Zurück zum Zitat Chow K, et al. Saturation recovery single-shot acquisition (SASHA) for myocardial T (1) mapping. Magn Reson Med. 2014;71(6):2082–95.PubMedCrossRef Chow K, et al. Saturation recovery single-shot acquisition (SASHA) for myocardial T (1) mapping. Magn Reson Med. 2014;71(6):2082–95.PubMedCrossRef
8.
Zurück zum Zitat Kellman P, et al. Optimized saturation recovery protocols for T1-mapping in the heart: influence of sampling strategies on precision. J Cardiovasc Magn Reson. 2014;16(1):55.PubMedPubMedCentralCrossRef Kellman P, et al. Optimized saturation recovery protocols for T1-mapping in the heart: influence of sampling strategies on precision. J Cardiovasc Magn Reson. 2014;16(1):55.PubMedPubMedCentralCrossRef
10.
Zurück zum Zitat Higgins D, Moon J. Review of T1 mapping methods: comparative effectiveness including reproducibility issues. Curr Cardiovasc Imaging Rep. 2014;7(3):1–10.CrossRef Higgins D, Moon J. Review of T1 mapping methods: comparative effectiveness including reproducibility issues. Curr Cardiovasc Imaging Rep. 2014;7(3):1–10.CrossRef
11.
Zurück zum Zitat Xanthis CG, et al. Parallel simulations for QUAntifying RElaxation magnetic resonance constants (SQUAREMR): an example towards accurate MOLLI T1 measurements. J Cardiovasc Magn Reson. 2015;17:104.PubMedPubMedCentralCrossRef Xanthis CG, et al. Parallel simulations for QUAntifying RElaxation magnetic resonance constants (SQUAREMR): an example towards accurate MOLLI T1 measurements. J Cardiovasc Magn Reson. 2015;17:104.PubMedPubMedCentralCrossRef
12.
Zurück zum Zitat Childs H, et al. Comparison of long and short axis quantification of left ventricular volume parameters by cardiovascular magnetic resonance, with ex-vivo validation. J Cardiovasc Magn Reson. 2011;13:40.PubMedPubMedCentralCrossRef Childs H, et al. Comparison of long and short axis quantification of left ventricular volume parameters by cardiovascular magnetic resonance, with ex-vivo validation. J Cardiovasc Magn Reson. 2011;13:40.PubMedPubMedCentralCrossRef
13.
Zurück zum Zitat Hazirolan T, et al. Comparison of short and long axis methods in cardiac MR imaging and echocardiography for left ventricular function. Diagn Interv Radiol. 2007;13(1):33–8.PubMed Hazirolan T, et al. Comparison of short and long axis methods in cardiac MR imaging and echocardiography for left ventricular function. Diagn Interv Radiol. 2007;13(1):33–8.PubMed
14.
Zurück zum Zitat Huttin O, et al. Assessment of left ventricular ejection fraction calculation on long-axis views from cardiac magnetic resonance imaging in patients with acute myocardial infarction. Medicine (Baltimore). 2015;94(43):e1856.CrossRef Huttin O, et al. Assessment of left ventricular ejection fraction calculation on long-axis views from cardiac magnetic resonance imaging in patients with acute myocardial infarction. Medicine (Baltimore). 2015;94(43):e1856.CrossRef
15.
16.
Zurück zum Zitat Bohnen S, et al. T1 mapping cardiovascular magnetic resonance imaging to detect myocarditis-impact of slice orientation on the diagnostic performance. Eur J Radiol. 2017;86:6–12.PubMedCrossRef Bohnen S, et al. T1 mapping cardiovascular magnetic resonance imaging to detect myocarditis-impact of slice orientation on the diagnostic performance. Eur J Radiol. 2017;86:6–12.PubMedCrossRef
17.
Zurück zum Zitat Caballeros FM, et al. Comparison of cardiac imaging planes for quantification of T1 maps and myocardial extra-cellular volume (ECV), in ECR 2016. European Society of Cardiology. Vienna; 2016. Caballeros FM, et al. Comparison of cardiac imaging planes for quantification of T1 maps and myocardial extra-cellular volume (ECV), in ECR 2016. European Society of Cardiology. Vienna; 2016.
18.
Zurück zum Zitat Messroghli DR, et al. Clinical recommendations for cardiovascular magnetic resonance mapping of T1, T2, T2* and extracellular volume: a consensus statement by the Society for Cardiovascular Magnetic Resonance (SCMR) endorsed by the European Association for Cardiovascular Imaging (EACVI). J Cardiovasc Magn Reson. 2017;19(1):75.PubMedPubMedCentralCrossRef Messroghli DR, et al. Clinical recommendations for cardiovascular magnetic resonance mapping of T1, T2, T2* and extracellular volume: a consensus statement by the Society for Cardiovascular Magnetic Resonance (SCMR) endorsed by the European Association for Cardiovascular Imaging (EACVI). J Cardiovasc Magn Reson. 2017;19(1):75.PubMedPubMedCentralCrossRef
19.
20.
Zurück zum Zitat Weingartner S, et al. Myocardial T1-mapping at 3T using saturation-recovery: reference values, precision and comparison with MOLLI. J Cardiovasc Magn Reson. 2016;18(1):84.PubMedPubMedCentralCrossRef Weingartner S, et al. Myocardial T1-mapping at 3T using saturation-recovery: reference values, precision and comparison with MOLLI. J Cardiovasc Magn Reson. 2016;18(1):84.PubMedPubMedCentralCrossRef
21.
Zurück zum Zitat Child N, et al. Comparison of MOLLI, shMOLLLI, and SASHA in discrimination between health and disease and relationship with histologically derived collagen volume fraction. Eur Heart J Cardiovasc Imaging. 2017;19(7):768–76.CrossRef Child N, et al. Comparison of MOLLI, shMOLLLI, and SASHA in discrimination between health and disease and relationship with histologically derived collagen volume fraction. Eur Heart J Cardiovasc Imaging. 2017;19(7):768–76.CrossRef
22.
Zurück zum Zitat Dabir D, et al. Reference values for healthy human myocardium using a T1 mapping methodology: results from the international T1 multicenter cardiovascular magnetic resonance study. J Cardiovasc Magn Reson. 2014;16:69.PubMedPubMedCentralCrossRef Dabir D, et al. Reference values for healthy human myocardium using a T1 mapping methodology: results from the international T1 multicenter cardiovascular magnetic resonance study. J Cardiovasc Magn Reson. 2014;16:69.PubMedPubMedCentralCrossRef
23.
Zurück zum Zitat Piechnik SK, et al. Shortened Modified look-locker inversion recovery (ShMOLLI) for clinical myocardial T1-mapping at 1.5 and 3 T within a 9 heartbeat breathhold. J Cardiovasc Magn Reson. 2010;12:69.PubMedPubMedCentralCrossRef Piechnik SK, et al. Shortened Modified look-locker inversion recovery (ShMOLLI) for clinical myocardial T1-mapping at 1.5 and 3 T within a 9 heartbeat breathhold. J Cardiovasc Magn Reson. 2010;12:69.PubMedPubMedCentralCrossRef
24.
Zurück zum Zitat Rogers T, Puntmann VO. T1 mapping - beware regional variations. Eur Heart J Cardiovasc Imaging. 2014;15(11):1302.PubMedCrossRef Rogers T, Puntmann VO. T1 mapping - beware regional variations. Eur Heart J Cardiovasc Imaging. 2014;15(11):1302.PubMedCrossRef
25.
Zurück zum Zitat Haaf P, et al. Cardiac T1 mapping and extracellular volume (ECV) in clinical practice: a comprehensive review. J Cardiovasc Magn Reson. 2016;18(1):89.PubMedPubMedCentralCrossRef Haaf P, et al. Cardiac T1 mapping and extracellular volume (ECV) in clinical practice: a comprehensive review. J Cardiovasc Magn Reson. 2016;18(1):89.PubMedPubMedCentralCrossRef
Metadaten
Titel
Comparison of short axis and long axis acquisitions of T1 and extracellular volume mapping using MOLLI and SASHA in patients with myocardial infarction and healthy volunteers
verfasst von
Christos G. Xanthis
David Nordlund
Robert Jablonowski
Håkan Arheden
Publikationsdatum
01.12.2019
Verlag
BioMed Central
Erschienen in
BMC Medical Imaging / Ausgabe 1/2019
Elektronische ISSN: 1471-2342
DOI
https://doi.org/10.1186/s12880-019-0320-x

Weitere Artikel der Ausgabe 1/2019

BMC Medical Imaging 1/2019 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.