Comparison of treatment position with mask immobilization and standard diagnostic setup in intracranial MRI radiotherapy simulation

verfasst von: Vesna Mekiš, Valerija Žager Marciuš, Dominika Rogina, Laura Dolenc, Nejc Mekiš

Erschienen in: Strahlentherapie und Onkologie | Ausgabe 7/2021

Einloggen, um Zugang zu erhalten

Abstract

Purpose

This study aims to compare the quality of images resulting from magnetic resonance imaging of patients who underwent intracranial MRI simulation using two different setups (treatment position with mask immobilization and standard diagnostic setup). Due to a larger number of channels and lack of mask immobilization in the standard diagnostic setup, we would like to evaluate whether this is an appropriate technique for MRI treatment planning.

Methods

In total, 70 patients who underwent MR imaging of the brain at 1.5T were included in the study (48 for 6‑channel flex coil, 22 for 24-channel HNU face bill coil). Contrast-enhanced 3D T1w and T2 FLAIR images were acquired. Images were subjectively compared for artifact appearance and general image quality by three radiographers. Objective comparison of contrast rate, contrast-to-noise ratio, and signal-to-noise ratio was also performed.

Results

FLAIR and contrast-enhanced 3D T1w images showed various artifacts, such as susceptibility and movement artifacts. There were no statistically significant differences regarding the evaluation of movement artifacts between two coils and two different immobilization methods. There were also no statistically significant differences (p > 0.05) between the 6‑channel flex coil and 24-channel HNU face bill coil regarding qualitative general image quality and objective measures.

Conclusion

There were no statistically significant differences between the occurrence of movement artifacts, overall image quality, and objective image quality in treatment position with mask immobilization and standard diagnostic setup. Based on this result, we can conclude that a standard diagnostic setup is also applicable in intracranial MRI treatment planning with no loss to image quality. Registration of the imaging plans was not performed in this study; therefore, it might still be necessary to perform measurements of tumor delineation matching and geometrical accuracy acceptance in our institution.
Literatur
17.
Zurück zum Zitat Somasundaram K, Kalavathi P (2012) Analysis of imaging artifacts in MR brain images. Orient J Comput Sci Technol 5:135–141 Somasundaram K, Kalavathi P (2012) Analysis of imaging artifacts in MR brain images. Orient J Comput Sci Technol 5:135–141
Metadaten
Titel
Comparison of treatment position with mask immobilization and standard diagnostic setup in intracranial MRI radiotherapy simulation
verfasst von
Vesna Mekiš
Valerija Žager Marciuš
Dominika Rogina
Laura Dolenc
Nejc Mekiš
Publikationsdatum
21.04.2021
Verlag
Springer Berlin Heidelberg
Erschienen in
Strahlentherapie und Onkologie / Ausgabe 7/2021
Print ISSN: 0179-7158
Elektronische ISSN: 1439-099X
DOI
https://doi.org/10.1007/s00066-021-01776-3

Weitere Artikel der Ausgabe 7/2021

Strahlentherapie und Onkologie 7/2021 Zur Ausgabe

Update Onkologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.