19.08.2017 | Sleep Breathing Physiology and Disorders • Original Article

Comparison of two home sleep testing devices with different strategies for diagnosis of OSA

verfasst von: Tyler Gumb, Akosua Twumasi, Shahnaz Alimokhtari, Alan Perez, Kathleen Black, David M. Rapoport, Jag Sunderram, Indu Ayappa

Erschienen in: Sleep and Breathing | Ausgabe 1/2018

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Home sleep testing devices are being widely used in diagnosis/screening for obstructive sleep apnea (OSA). We examined differences in OSA metrics obtained from two devices with divergent home monitoring strategies, the Apnea Risk Evaluation System (ARES™, multiple signals plus forehead reflectance oximetry) and the Nonin WristOx2™ (single channel finger transmission pulse oximeter), compared to differences from night-night variability of OSA.

Methods

One hundred fifty-two male/26 female subjects (BMI = 30.3 ± 5.6 kg/m2, age = 52.5 ± 8.9 years) were recruited without regard to OSA symptoms and simultaneously wore both ARES™ and Nonin WristOx2™ for two nights (n = 351 nights). Automated analysis of the WristOx2 yielded oxygen desaturation index (ODIOx2, ≥4% O2 dips/h), and automated analysis with manual editing of ARES™ yielded AHI4ARES (apneas + hypopneas with ≥4% O2 dips/h) and RDIARES (apneas + hypopneas with ≥4% O2 dips/h or arousal surrogates). Baseline awake oxygen saturation, percent time < 90% O2 saturation (%time < 90%O2Sat), and O2 signal loss were compared between the two methods.

Results

Correlation between AHI4ARES and ODIOx2 was high (ICC = 0.9, 95% CI = 0.87–0.92, p < 0.001, bias ± SD = 0.7 ± 6.1 events/h). Agreement values for OSA diagnosis (77–85%) between devices were similar to those seen from night-to-night variability of OSA using a single device. Awake baseline O2 saturation was significantly higher in the ARES™ (96.2 ± 1.6%) than WristOx2™ (92.2 ± 2.1%, p < 0.01). There was a significantly lower %time < 90%O2Sat reported by the ARES™ compared to WristOx2 (median (IQR) 0.5 (0.0, 2.6) vs. 2.1 (0.3, 9.7), p < 0.001), and the correlation was low (ICC = 0.2).

Conclusions

OSA severity metrics predominantly dependent on change in oxygen saturation and metrics used in diagnosis of OSA (AHI4 and ODI) correlated well across devices tested. However, differences in cumulative oxygen desaturation measures (i.e., %time < 90%O2Sat) between the devices suggest that caution is needed when interpreting this metric particularly in populations likely to have significant hypoxia.
Literatur
1.
Zurück zum Zitat Hang LW, Wang HL, Chen JH, Hsu JC, Lin HH, Chung WS et al (2015) Validation of overnight oximetry to diagnose patients with moderate to severe obstructive sleep apnea. BMC Pulm Med 15:24CrossRefPubMedPubMedCentral Hang LW, Wang HL, Chen JH, Hsu JC, Lin HH, Chung WS et al (2015) Validation of overnight oximetry to diagnose patients with moderate to severe obstructive sleep apnea. BMC Pulm Med 15:24CrossRefPubMedPubMedCentral
2.
Zurück zum Zitat Chai-Coetzer CL, Antic NA, Hamilton GS, McArdle N, Wong K, Yee BJ et al (2017) Physician decision making and clinical outcomes with laboratory polysomnography or limited-channel sleep studies for obstructive sleep apnea: a randomized trial. Ann Intern Med 166(5):332–340CrossRefPubMed Chai-Coetzer CL, Antic NA, Hamilton GS, McArdle N, Wong K, Yee BJ et al (2017) Physician decision making and clinical outcomes with laboratory polysomnography or limited-channel sleep studies for obstructive sleep apnea: a randomized trial. Ann Intern Med 166(5):332–340CrossRefPubMed
3.
Zurück zum Zitat Collop NA, Anderson WM, Boehlecke B, Claman D, Goldberg R, Gottlieb DJ et al (2007) Clinical guidelines for the use of unattended portable monitors in the diagnosis of obstructive sleep apnea in adult patients. Portable Monitoring Task Force of the American Academy of Sleep Medicine. J Clin Sleep Med 3(7):737–747PubMed Collop NA, Anderson WM, Boehlecke B, Claman D, Goldberg R, Gottlieb DJ et al (2007) Clinical guidelines for the use of unattended portable monitors in the diagnosis of obstructive sleep apnea in adult patients. Portable Monitoring Task Force of the American Academy of Sleep Medicine. J Clin Sleep Med 3(7):737–747PubMed
4.
Zurück zum Zitat Ayappa I, Norman RG, Seelall V, Rapoport DM (2008) Validation of a self-applied unattended monitor for sleep disordered breathing. J Clin Sleep Med 4(1):26–37PubMedPubMedCentral Ayappa I, Norman RG, Seelall V, Rapoport DM (2008) Validation of a self-applied unattended monitor for sleep disordered breathing. J Clin Sleep Med 4(1):26–37PubMedPubMedCentral
5.
Zurück zum Zitat Vazquez JC, Tsai WH, Flemons WW, Masuda A, Brant R, Hajduk E et al (2000) Automated analysis of digital oximetry in the diagnosis of obstructive sleep apnoea. Thorax 55(4):302–307CrossRefPubMedPubMedCentral Vazquez JC, Tsai WH, Flemons WW, Masuda A, Brant R, Hajduk E et al (2000) Automated analysis of digital oximetry in the diagnosis of obstructive sleep apnoea. Thorax 55(4):302–307CrossRefPubMedPubMedCentral
6.
Zurück zum Zitat Zou D, Grote L, Peker Y, Lindblad U, Hedner J (2006) Validation a portable monitoring device for sleep apnea diagnosis in a population based cohort using synchronized home polysomnography. Sleep 29(3):367–374CrossRefPubMed Zou D, Grote L, Peker Y, Lindblad U, Hedner J (2006) Validation a portable monitoring device for sleep apnea diagnosis in a population based cohort using synchronized home polysomnography. Sleep 29(3):367–374CrossRefPubMed
7.
Zurück zum Zitat Westbrook PR, Levendowski DJ, Cvetinovic M, Zavora T, Velimirovic V, Henninger D et al (2005) Description and validation of the apnea risk evaluation system: a novel method to diagnose sleep apnea-hypopnea in the home. Chest 128(4):2166–2175CrossRefPubMed Westbrook PR, Levendowski DJ, Cvetinovic M, Zavora T, Velimirovic V, Henninger D et al (2005) Description and validation of the apnea risk evaluation system: a novel method to diagnose sleep apnea-hypopnea in the home. Chest 128(4):2166–2175CrossRefPubMed
8.
Zurück zum Zitat Kapur VK, Auckley DH, Chowdhuri S, Kuhlmann DC, Mehra R, Ramar K et al (2017) Clinical practice guideline for diagnostic testing for adult obstructive sleep apnea: an American Academy of Sleep Medicine Clinical Practice Guideline. J Clin Sleep Med 13(3):479–504CrossRefPubMedPubMedCentral Kapur VK, Auckley DH, Chowdhuri S, Kuhlmann DC, Mehra R, Ramar K et al (2017) Clinical practice guideline for diagnostic testing for adult obstructive sleep apnea: an American Academy of Sleep Medicine Clinical Practice Guideline. J Clin Sleep Med 13(3):479–504CrossRefPubMedPubMedCentral
9.
Zurück zum Zitat Kisch-Wedel H, Bernreuter P, Kemming G, Albert M, Zwissler B (2009) Does the estimation of light attenuation in tissue increase the accuracy of reflectance pulse oximetry at low oxygen saturations in vivo? IEEE Trans Biomed Eng 56(9):2271–2279CrossRefPubMed Kisch-Wedel H, Bernreuter P, Kemming G, Albert M, Zwissler B (2009) Does the estimation of light attenuation in tissue increase the accuracy of reflectance pulse oximetry at low oxygen saturations in vivo? IEEE Trans Biomed Eng 56(9):2271–2279CrossRefPubMed
10.
Zurück zum Zitat Zafar S, Ayappa I, Norman RG, Krieger AC, Walsleben JA, Rapoport DM (2005) Choice of oximeter affects apnea-hypopnea index. Chest 127(1):80–88CrossRefPubMed Zafar S, Ayappa I, Norman RG, Krieger AC, Walsleben JA, Rapoport DM (2005) Choice of oximeter affects apnea-hypopnea index. Chest 127(1):80–88CrossRefPubMed
11.
Zurück zum Zitat Davila DG, Richards KC, Marshall BL, O’Sullivan PS, Osbahr LA, Huddleston RB et al (2003) Oximeter’s acquisition parameter influences the profile of respiratory disturbances. Sleep 26(1):91–95PubMed Davila DG, Richards KC, Marshall BL, O’Sullivan PS, Osbahr LA, Huddleston RB et al (2003) Oximeter’s acquisition parameter influences the profile of respiratory disturbances. Sleep 26(1):91–95PubMed
12.
Zurück zum Zitat Cross TJ, Keller-Ross M, Issa A, Wentz R, Taylor B, Johnson B (2015) The impact of averaging window length on the “desaturation indexes during overnight pulse oximetry at high-altitude”. Sleep 38(8):1331–1334CrossRefPubMedPubMedCentral Cross TJ, Keller-Ross M, Issa A, Wentz R, Taylor B, Johnson B (2015) The impact of averaging window length on the “desaturation indexes during overnight pulse oximetry at high-altitude”. Sleep 38(8):1331–1334CrossRefPubMedPubMedCentral
13.
Zurück zum Zitat Stepnowsky C, Levendowski D, Popovic D, Ayappa I, Rapoport DM (2013) Scoring accuracy of automated sleep staging from a bipolar electroocular recording compared to manual scoring by multiple raters. Sleep Med 14(11):1199–1207CrossRefPubMed Stepnowsky C, Levendowski D, Popovic D, Ayappa I, Rapoport DM (2013) Scoring accuracy of automated sleep staging from a bipolar electroocular recording compared to manual scoring by multiple raters. Sleep Med 14(11):1199–1207CrossRefPubMed
14.
Zurück zum Zitat Flemons WW, Littner MR (2003) Measuring agreement between diagnostic devices. Chest 124(4):1535–1542CrossRefPubMed Flemons WW, Littner MR (2003) Measuring agreement between diagnostic devices. Chest 124(4):1535–1542CrossRefPubMed
15.
Zurück zum Zitat Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310CrossRefPubMed Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310CrossRefPubMed
16.
Zurück zum Zitat Levendowski D, Steward D, Woodson BT, Olmstead R, Popovic D, Westbrook P (2009) The impact of obstructive sleep apnea variability measured in-lab versus in-home on sample size calculations. Int Arch Med 2(1):2CrossRefPubMedPubMedCentral Levendowski D, Steward D, Woodson BT, Olmstead R, Popovic D, Westbrook P (2009) The impact of obstructive sleep apnea variability measured in-lab versus in-home on sample size calculations. Int Arch Med 2(1):2CrossRefPubMedPubMedCentral
17.
Zurück zum Zitat Newell J, Mairesse O, Verbanck P, Neu D (2012) Is a one-night stay in the lab really enough to conclude? First-night effect and night-to-night variability in polysomnographic recordings among different clinical population samples. Psychiatry Res 200(2–3):795–801CrossRefPubMed Newell J, Mairesse O, Verbanck P, Neu D (2012) Is a one-night stay in the lab really enough to conclude? First-night effect and night-to-night variability in polysomnographic recordings among different clinical population samples. Psychiatry Res 200(2–3):795–801CrossRefPubMed
18.
Zurück zum Zitat Le Bon O, Hoffmann G, Tecco J, Staner L, Noseda A, Pelc I et al (2000) Mild to moderate sleep respiratory events: one negative night may not be enough. Chest 118(2):353–359CrossRefPubMed Le Bon O, Hoffmann G, Tecco J, Staner L, Noseda A, Pelc I et al (2000) Mild to moderate sleep respiratory events: one negative night may not be enough. Chest 118(2):353–359CrossRefPubMed
19.
Zurück zum Zitat Prasad B, Usmani S, Steffen AD, Van Dongen HP, Pack FM, Strakovsky I et al (2016) Short-term variability in apnea-hypopnea index during extended home portable monitoring. J Clin Sleep Med 12(6):855–863CrossRefPubMedPubMedCentral Prasad B, Usmani S, Steffen AD, Van Dongen HP, Pack FM, Strakovsky I et al (2016) Short-term variability in apnea-hypopnea index during extended home portable monitoring. J Clin Sleep Med 12(6):855–863CrossRefPubMedPubMedCentral
20.
Zurück zum Zitat Nigro CA, Aimaretti S, Gonzalez S, Rhodius E (2009) Validation of the WristOx 3100 oximeter for the diagnosis of sleep apnea/hypopnea syndrome. Sleep Breath 13(2):127–136CrossRefPubMed Nigro CA, Aimaretti S, Gonzalez S, Rhodius E (2009) Validation of the WristOx 3100 oximeter for the diagnosis of sleep apnea/hypopnea syndrome. Sleep Breath 13(2):127–136CrossRefPubMed
21.
Zurück zum Zitat Romem A, Romem A, Koldobskiy D, Scharf SM (2014) Diagnosis of obstructive sleep apnea using pulse oximeter derived photoplethysmographic signals. J Clin Sleep Med 10(3):285–290PubMedPubMedCentral Romem A, Romem A, Koldobskiy D, Scharf SM (2014) Diagnosis of obstructive sleep apnea using pulse oximeter derived photoplethysmographic signals. J Clin Sleep Med 10(3):285–290PubMedPubMedCentral
22.
Zurück zum Zitat Chiner E, Signes-Costa J, Arriero JM, Marco J, Fuentes I, Sergado A (1999) Nocturnal oximetry for the diagnosis of the sleep apnoea hypopnoea syndrome: a method to reduce the number of polysomnographies? Thorax 54(11):968–971CrossRefPubMedPubMedCentral Chiner E, Signes-Costa J, Arriero JM, Marco J, Fuentes I, Sergado A (1999) Nocturnal oximetry for the diagnosis of the sleep apnoea hypopnoea syndrome: a method to reduce the number of polysomnographies? Thorax 54(11):968–971CrossRefPubMedPubMedCentral
23.
Zurück zum Zitat Collop NA, Tracy SL, Kapur V, Mehra R, Kuhlmann D, Fleishman SA et al (2011) Obstructive sleep apnea devices for out-of-center (OOC) testing: technology evaluation. J Clin Sleep Med 7(5):531–548PubMedPubMedCentral Collop NA, Tracy SL, Kapur V, Mehra R, Kuhlmann D, Fleishman SA et al (2011) Obstructive sleep apnea devices for out-of-center (OOC) testing: technology evaluation. J Clin Sleep Med 7(5):531–548PubMedPubMedCentral
24.
Zurück zum Zitat Dawson A, Loving RT, Gordon RM, Abel SL, Loewy D, Kripke DF et al (2015) Type III home sleep testing versus pulse oximetry: is the respiratory disturbance index better than the oxygen desaturation index to predict the apnoea-hypopnoea index measured during laboratory polysomnography? BMJ Open 5(6):e007956CrossRefPubMedPubMedCentral Dawson A, Loving RT, Gordon RM, Abel SL, Loewy D, Kripke DF et al (2015) Type III home sleep testing versus pulse oximetry: is the respiratory disturbance index better than the oxygen desaturation index to predict the apnoea-hypopnoea index measured during laboratory polysomnography? BMJ Open 5(6):e007956CrossRefPubMedPubMedCentral
25.
Zurück zum Zitat Yamaura K, Nanishi N, Higashi M, Hoka S (2014) Effects of thermoregulatory vasoconstriction on pulse hemoglobin measurements using a co-oximeter in patients undergoing surgery. J Clin Anesth 26(8):643–647CrossRefPubMed Yamaura K, Nanishi N, Higashi M, Hoka S (2014) Effects of thermoregulatory vasoconstriction on pulse hemoglobin measurements using a co-oximeter in patients undergoing surgery. J Clin Anesth 26(8):643–647CrossRefPubMed
26.
Zurück zum Zitat The Report of an American Academy of Sleep Medicine Task Force (1999) Sleep-related breathing disorders in adults: recommendations for syndrome definition and measurement techniques in clinical research. Sleep 22(5):667–689 The Report of an American Academy of Sleep Medicine Task Force (1999) Sleep-related breathing disorders in adults: recommendations for syndrome definition and measurement techniques in clinical research. Sleep 22(5):667–689
27.
Zurück zum Zitat Hosselet J, Ayappa I, Norman RG, Krieger AC, Rapoport DM (2001) Classification of sleep-disordered breathing. Am J Respir Crit Care Med 163(2):398–405CrossRefPubMed Hosselet J, Ayappa I, Norman RG, Krieger AC, Rapoport DM (2001) Classification of sleep-disordered breathing. Am J Respir Crit Care Med 163(2):398–405CrossRefPubMed
28.
Zurück zum Zitat Masdeu MJ, Ayappa I, Hwang D, Mooney AM, Rapoport DM (2010) Impact of clinical assessment on use of data from unattended limited monitoring as opposed to full-in lab PSG in sleep disordered breathing. J Clin Sleep Med 6(1):51–58PubMedPubMedCentral Masdeu MJ, Ayappa I, Hwang D, Mooney AM, Rapoport DM (2010) Impact of clinical assessment on use of data from unattended limited monitoring as opposed to full-in lab PSG in sleep disordered breathing. J Clin Sleep Med 6(1):51–58PubMedPubMedCentral
29.
Zurück zum Zitat Ayappa I, Rapaport BS, Norman RG, Rapoport DM (2005) Immediate consequences of respiratory events in sleep disordered breathing. Sleep Med 6(2):123–130CrossRefPubMed Ayappa I, Rapaport BS, Norman RG, Rapoport DM (2005) Immediate consequences of respiratory events in sleep disordered breathing. Sleep Med 6(2):123–130CrossRefPubMed
Metadaten
Titel
Comparison of two home sleep testing devices with different strategies for diagnosis of OSA
verfasst von
Tyler Gumb
Akosua Twumasi
Shahnaz Alimokhtari
Alan Perez
Kathleen Black
David M. Rapoport
Jag Sunderram
Indu Ayappa
Publikationsdatum
19.08.2017
Verlag
Springer International Publishing
Erschienen in
Sleep and Breathing / Ausgabe 1/2018
Print ISSN: 1520-9512
Elektronische ISSN: 1522-1709
DOI
https://doi.org/10.1007/s11325-017-1547-9

Weitere Artikel der Ausgabe 1/2018

Sleep and Breathing 1/2018 Zur Ausgabe

Sleep Breathing Physiology and Disorders • Original Article

Comorbid insomnia and sleep apnea in Veterans with post-traumatic stress disorder

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.