Purpose

To evaluate the complication rate in adult subjects with open cavities that were implanted with the Vibrant Soundbridge implant, using the round window (RW) vibroplasty procedure.

Methods

From 2009 to 2014, 21 adult subjects with mixed hearing loss, all with sequel from open tympanoplasty surgery, underwent RW vibroplasty (RW-VPL). Surgical complications were recorded and a standard minimal approach was used as a basis for all the cases that needed revision.

Results

The mean follow-up was 42 months (range 12–76). Complications occurred in nearly half of the cases and included: cable extrusion (23.8%), hardware failure (14.3%), profound hearing loss (9.5%), and inadequate RW coupling (9.5%). A minimal endaural approach (MEA) was used in the majority of the cases (86.7%), while the extended endaural approach was adopted for those patients requiring explantation with or without replacement (14.3%).

Conclusions

RW-VPL can be considered a possible option for the rehabilitation of auditory impairment derived from an open tympanoplasty procedure due to cholesteatoma. The procedure may lead to minor/major complications that may require a surgical revision. By adopting an MEA, it has been possible to manage all the situations in which functionality of the device is worth being preserved.
Literatur
1.
Zurück zum Zitat Beltrame AM, Martini A, Prosser S, Giarbini N, Streitberger C (2009) Coupling the Vibrant Soundbridge to cochlea round window: auditory results in patients with mixed hearing loss. Otol Neurotol 30:194–201CrossRefPubMed Beltrame AM, Martini A, Prosser S, Giarbini N, Streitberger C (2009) Coupling the Vibrant Soundbridge to cochlea round window: auditory results in patients with mixed hearing loss. Otol Neurotol 30:194–201CrossRefPubMed
2.
Zurück zum Zitat Colletti V, Soli SD, Carner M, Colletti L (2006) Treatment of mixed hearing losses via implantation of a vibratory transducer on the round window. Int J Audiol 45(10):600–608CrossRefPubMed Colletti V, Soli SD, Carner M, Colletti L (2006) Treatment of mixed hearing losses via implantation of a vibratory transducer on the round window. Int J Audiol 45(10):600–608CrossRefPubMed
3.
Zurück zum Zitat Lassaletta L, Calvino M, Sánchez-Cuadrado I, Pérez-Mora RM, Muñoz E, Gavilán J (2015) Pros and cons of round window vibroplasty in open cavities: audiological, surgical, and quality of life outcomes. Otol Neurotol 36:944–952CrossRefPubMed Lassaletta L, Calvino M, Sánchez-Cuadrado I, Pérez-Mora RM, Muñoz E, Gavilán J (2015) Pros and cons of round window vibroplasty in open cavities: audiological, surgical, and quality of life outcomes. Otol Neurotol 36:944–952CrossRefPubMed
4.
Zurück zum Zitat Mojallal H, Schwab B, Hinze AL, Giere T, Lenarz T (2015) Retrospective audiological analysis of bone conduction versus round window vibratory stimulation in patients with mixed hearing loss. Int J Audiol 54:391–400CrossRefPubMed Mojallal H, Schwab B, Hinze AL, Giere T, Lenarz T (2015) Retrospective audiological analysis of bone conduction versus round window vibratory stimulation in patients with mixed hearing loss. Int J Audiol 54:391–400CrossRefPubMed
5.
Zurück zum Zitat Barbara M, Volpini L, Ciotti M, Filippi C, Covelli E, Monini S, D’Ambrosio F (2015) Cone beam computed tomography after round window vibroplasty: do the radiological findings match the auditory outcome? Acta Otolaryngol 135:369–375CrossRefPubMed Barbara M, Volpini L, Ciotti M, Filippi C, Covelli E, Monini S, D’Ambrosio F (2015) Cone beam computed tomography after round window vibroplasty: do the radiological findings match the auditory outcome? Acta Otolaryngol 135:369–375CrossRefPubMed
6.
Zurück zum Zitat Perez R, Adelman C, Chordekar S, de Jong MA, Sohmer H (2014) The mechanism of direct stimulation of the cochlea by vibrating the round window. J Basic Clin Physiol Pharmacol 25:273–276CrossRefPubMed Perez R, Adelman C, Chordekar S, de Jong MA, Sohmer H (2014) The mechanism of direct stimulation of the cochlea by vibrating the round window. J Basic Clin Physiol Pharmacol 25:273–276CrossRefPubMed
7.
Zurück zum Zitat Zwartenkot JW, Mulder JJS, Snik AFM, Mylanus EAM (2016) Active middle ear implantation: longterm medical and technical follow-up, implant survival and complication. Otol Neurotol 37:513–519CrossRefPubMed Zwartenkot JW, Mulder JJS, Snik AFM, Mylanus EAM (2016) Active middle ear implantation: longterm medical and technical follow-up, implant survival and complication. Otol Neurotol 37:513–519CrossRefPubMed
8.
Zurück zum Zitat Böheim K, Mlynski R, Lenarz T, Schlögel M, Hagen R (2012) Round window vibroplasty: long-term results. Acta Otolaryngol 132:1042–1048CrossRefPubMed Böheim K, Mlynski R, Lenarz T, Schlögel M, Hagen R (2012) Round window vibroplasty: long-term results. Acta Otolaryngol 132:1042–1048CrossRefPubMed
9.
Zurück zum Zitat Colletti V, Mandalà M, Colletti L (2012) Electrocochleography in round window Vibrant Soundbridge implantation. Otolaryngol Head Neck Surg 146:633–640CrossRefPubMed Colletti V, Mandalà M, Colletti L (2012) Electrocochleography in round window Vibrant Soundbridge implantation. Otolaryngol Head Neck Surg 146:633–640CrossRefPubMed
10.
Zurück zum Zitat Verhaert N, Mojallal H, Schwab B (2013) Indications and outcome of subtotal petrosectomy for active middle ear implants. Eur Arch Otorhinolaryngol 270:1243–1248CrossRefPubMed Verhaert N, Mojallal H, Schwab B (2013) Indications and outcome of subtotal petrosectomy for active middle ear implants. Eur Arch Otorhinolaryngol 270:1243–1248CrossRefPubMed
11.
Zurück zum Zitat Linder T, Schlegel C, DeMin N, van der Westhuizen S (2009) Active middle ear implants in patients undergoing subtotal petrosectomy: new application for the Vibrant Soundbridge device and its implication for lateral cranium base surgery. Otol Neurotol 30:41–47CrossRefPubMed Linder T, Schlegel C, DeMin N, van der Westhuizen S (2009) Active middle ear implants in patients undergoing subtotal petrosectomy: new application for the Vibrant Soundbridge device and its implication for lateral cranium base surgery. Otol Neurotol 30:41–47CrossRefPubMed
Metadaten
Titel
Complications after round window vibroplasty
verfasst von
Maurizio Barbara
Luigi Volpini
Edoardo Covelli
Martina Romeo
Chiara Filippi
Simonetta Monini
Publikationsdatum
27.03.2019
Verlag
Springer Berlin Heidelberg
Erschienen in
European Archives of Oto-Rhino-Laryngology / Ausgabe 6/2019
Print ISSN: 0937-4477
Elektronische ISSN: 1434-4726
DOI
https://doi.org/10.1007/s00405-019-05402-y

Weitere Artikel der Ausgabe 6/2019

European Archives of Oto-Rhino-Laryngology 6/2019 Zur Ausgabe

Update HNO

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert – ganz bequem per eMail.