Condensation of humidified air in the inflation line of a polyurethane cuff precludes correct continuous pressure monitoring during mechanical ventilation

verfasst von: Herbert Spapen, Walter Moeyersons, Wim Stiers, Geert Desmet, Emiel Suys

Erschienen in: Journal of Anesthesia | Ausgabe 6/2014

Einloggen, um Zugang zu erhalten

Abstract

Within continuously controlled limits of cuff pressure, an endotracheal tube cuff made of polyurethane (PU) prevents secretion inflow better than a conventional polyvinylchloride cuff. However, the temperature difference between ventilator gas and the air inside a PU cuff produces condensation droplets that accumulate in the cuff inflation canal. We investigated whether condensation influenced continuous cuff pressure monitoring. A PU-cuffed tube was placed into an artificial trachea and connected to a ventilator and test lung. An additional line was inserted at the distal cuff end to directly measure pressure inside the cuff. Methylene blue instillation via the inflation line mimicked condensation. Inspiratory (Pinsp) and expiratory (Pexp) pressures were continuously recorded. Six consecutive experiments were performed comparing pressures at baseline (T0) and at 24 h (T24). Shortly after dye instillation, pressures recorded at the inflation canal became permanently fixed at 25 cmH2O. In contrast, pressures measured directly in the cuff progressively decreased (mean Pinsp 30 ± 3 vs. 20 ± 2 cmH2O and mean Pexp 25 ± 0 vs. 12 ± 2 cmH2O, T0 vs. T24, both P < 0.05). Thus, condensation in the inflation line of a PU-cuffed tube renders continuous cuff pressure monitoring unreliable, thereby increasing the risk for microaspiration.
Literatur
2.
Zurück zum Zitat Seegobin RD, van Hasselt GL. Endotracheal cuff pressure and tracheal mucosal blood flow: endoscopic study of effects of four large volume cuffs. Br Med J (Clin Res Ed). 1984;288:965–8.CrossRef Seegobin RD, van Hasselt GL. Endotracheal cuff pressure and tracheal mucosal blood flow: endoscopic study of effects of four large volume cuffs. Br Med J (Clin Res Ed). 1984;288:965–8.CrossRef
3.
Zurück zum Zitat Nseir S, Brisson H, Marquette CH, Chaud P, Di Pompeo C, Diarra M, Durocher A. Variations in endotracheal cuff pressure in intubated critically ill patients: prevalence and risk factors. Eur J Anaesthesiol. 2009;26:229–34.PubMedCrossRef Nseir S, Brisson H, Marquette CH, Chaud P, Di Pompeo C, Diarra M, Durocher A. Variations in endotracheal cuff pressure in intubated critically ill patients: prevalence and risk factors. Eur J Anaesthesiol. 2009;26:229–34.PubMedCrossRef
4.
Zurück zum Zitat Dullenkopf A, Gerber A, Weiss M. Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube. Intensive Care Med. 2003;29:1849–53.PubMedCrossRef Dullenkopf A, Gerber A, Weiss M. Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube. Intensive Care Med. 2003;29:1849–53.PubMedCrossRef
5.
Zurück zum Zitat Ouanes I, Lyazidi A, Danin PE, Rana N, Di Bari A, Abroug F, Louis B, Brochard L. Mechanical influences on fluid leakage past the tracheal tube cuff in a benchtop model. Intensive Care Med. 2011;37:695–700.PubMedCrossRef Ouanes I, Lyazidi A, Danin PE, Rana N, Di Bari A, Abroug F, Louis B, Brochard L. Mechanical influences on fluid leakage past the tracheal tube cuff in a benchtop model. Intensive Care Med. 2011;37:695–700.PubMedCrossRef
6.
Zurück zum Zitat Zanella A, Scaravilli V, Isgrò S, Milan M, Cressoni M, Patroniti N, Fumagalli R, Pesenti A. Fluid leakage across tracheal tube cuff, effect of different cuff material, shape, and positive expiratory pressure: a bench-top study. Intensive Care Med. 2011;37:343–7.PubMedCrossRef Zanella A, Scaravilli V, Isgrò S, Milan M, Cressoni M, Patroniti N, Fumagalli R, Pesenti A. Fluid leakage across tracheal tube cuff, effect of different cuff material, shape, and positive expiratory pressure: a bench-top study. Intensive Care Med. 2011;37:343–7.PubMedCrossRef
7.
Zurück zum Zitat Li Bassi G, Ranzani OT, Marti JD, Giunta V, Luque N, Isetta V, Ferrer M, Farre R, Pimentel GL, Torres A. An in vitro study to assess determinant features associated with fluid sealing in the design of endotracheal tube cuffs and exerted tracheal pressures. Crit Care Med. 2013;41:518–26.PubMedCrossRef Li Bassi G, Ranzani OT, Marti JD, Giunta V, Luque N, Isetta V, Ferrer M, Farre R, Pimentel GL, Torres A. An in vitro study to assess determinant features associated with fluid sealing in the design of endotracheal tube cuffs and exerted tracheal pressures. Crit Care Med. 2013;41:518–26.PubMedCrossRef
8.
Zurück zum Zitat Rouzé A, Nseir S. Continuous control of tracheal cuff pressure for the prevention of ventilator-associated pneumonia in critically ill patients: where is the evidence? Curr Opin Crit Care. 2013;19:440–7.PubMedCrossRef Rouzé A, Nseir S. Continuous control of tracheal cuff pressure for the prevention of ventilator-associated pneumonia in critically ill patients: where is the evidence? Curr Opin Crit Care. 2013;19:440–7.PubMedCrossRef
9.
Zurück zum Zitat Jaillette E, Zerimech F, De Jonckheere J, Makris D, Balduyck M, Durocher A, Duhamel A, Nseir S. Efficiency of a pneumatic device in controlling cuff pressure of polyurethane-cuffed tracheal tubes: a randomized controlled study. BMC Anesthesiol. 2013;13:50.PubMedCentralPubMedCrossRef Jaillette E, Zerimech F, De Jonckheere J, Makris D, Balduyck M, Durocher A, Duhamel A, Nseir S. Efficiency of a pneumatic device in controlling cuff pressure of polyurethane-cuffed tracheal tubes: a randomized controlled study. BMC Anesthesiol. 2013;13:50.PubMedCentralPubMedCrossRef
Metadaten
Titel
Condensation of humidified air in the inflation line of a polyurethane cuff precludes correct continuous pressure monitoring during mechanical ventilation
verfasst von
Herbert Spapen
Walter Moeyersons
Wim Stiers
Geert Desmet
Emiel Suys
Publikationsdatum
01.12.2014
Verlag
Springer Japan
Erschienen in
Journal of Anesthesia / Ausgabe 6/2014
Print ISSN: 0913-8668
Elektronische ISSN: 1438-8359
DOI
https://doi.org/10.1007/s00540-014-1849-z

Weitere Artikel der Ausgabe 6/2014

Journal of Anesthesia 6/2014 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.