Conductive heating mattress leads to ECG changes that mimic pacemaker spikes

verfasst von: A. Bräuer, R. Franke, A. von Hammerstein-Equord

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 3/2021

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Abstract

Hypothermia is a common perioperative complication. To prevent perioperative hypothermia amongst other things electrical heating mattresses are used. We have made an observation with the use of an electrical heating mattress that may confuse users. In this case the ECG monitoring suddenly showed spikes that looked like spikes from an implanted pacemaker. When turning off the heating mattress the spikes disappeared and returned after turning on the heating mattress again.
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Hypothermia is a common perioperative complication that increases morbidity of the patients [1]. Therefore, guidelines all over the world recommend active warming therapy. Forced-air warming is the most important method for perioperative warming, but in many guidelines conductive warming methods are recommended as another form of active warming therapy [2]. Although the efficacy of active warming with conductive warming methods is still debated [35] it is used quite often.
We have made an observation with the use of an electrical heating mattress that may confuse users. In this case a patient undergoing Video Assisted Thoracic Surgery (VATS) in the lateral decubitus position was actively warmed from underneath using a UniqueTemp° Jelly blanket (Geratherm Medical AG, Geschwenda, Germany) and both legs were warmed by two leg blankets (UniqueTemp° Geratherm Medical AG, Geschwenda, Germany) after positioning on the operating table.
During surgery the ECG monitoring suddenly showed spikes that looked like spikes from an implanted pacemaker.
These spikes had no correlation to the QRS complex and the pattern looked like an ECG of a patient with pacemaker dysfunction, although the patient did not have an implanted pacemaker. When turning off the heating mattress the spikes disappeared (Fig. 1) and returned after turning on the heating mattress again.
It seems that these spikes were caused by interference of the electrically driven conductive heating blankets with the ECG. The warming system uses carbon-fibers that are heated with a maximum of 35.5 V direct current by a control unit. These current-carrying carbon fibers or the control unit can generate an electromagnetic field that may have caused these artefacts. The artefacts may be falsely interpreted as pacemaker dysfunction in a patient that has a pacemaker implanted.
The device was controlled after use and no technical problems could be detected. The manufacturer claims that the device meets the requirements of Council Directive 93/42 EEC of 14.06.93 on Medical Devices and of the standard DIN EN 60601-1 (Medical electrical equipment, Part 1: General requirements for safety), DIN EN 60601-1-2 (Medical electrical equipment Part 1–2: General requirements for safety—Collateral standard: Electromagnetic compatibility—Requirements and Tests), DIN EN 12470-4 and DIN EN 60601-2-35 (Medical electrical equipment, Part 2: Particular requirements for the safety of blankets, pads and mattresses intended for the warming of patients in medical use) and that a DIN EN ISO 13485 certified quality management system ensures that these requirements are met and entitles the manufacturer to use the CE 0118 label.
In the instruction manual the manufacturer states “Do not place the control unit directly on top of ECG or EEG equipment. Otherwise, faults may arise, e.g. the devices may produce false readings.” However, we placed the control unit on an IV pole far away from the monitoring equipment. The monitoring system (Philips IntelliVue MX500, Philips Medizin Systeme Boeblingen GmbH, Boeblingen Germany) is mounted on the ceiling and had no direct or indirect contact to the control unit and or the heating mattress.
In our view users of the system should be aware of the possibility that an electrical heating mattresses can cause artifacts that look like pacemaker spikes even if all safety precautions given by the manufacturer are take into consideration.

Acknowledgements

Open Access funding provided by Projekt DEAL.

Compliance with ethical standards

Conflict of interest

Prof. Dr. A. Bräuer is a member of the advisory board of 3M Europe and has received payments from 3M Germany, 3M Europe, 3M Asia Pacific Pte Ltd. for consultancy work. R. Franke and A. von Hammerstein-Equord had loaned a UniqueTemp° Jelly blanket and the two UniqueTemp° leg blankets from Geratherm Medical AG, Geschwenda, Germany for a study.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Literatur
1.
Zurück zum Zitat Sessler DI. Perioperative thermoregulation and heat balance. Lancet. 2016;387:2655–64.CrossRef Sessler DI. Perioperative thermoregulation and heat balance. Lancet. 2016;387:2655–64.CrossRef
3.
Zurück zum Zitat Emmert A, Franke R, Brandes IF, et al. Comparison of conductive and convective warming in patients undergoing video-assisted thoracic surgery: a prospective randomized clinical trial. Thorac Cardiovasc Surg. 2017;65:362–6.CrossRef Emmert A, Franke R, Brandes IF, et al. Comparison of conductive and convective warming in patients undergoing video-assisted thoracic surgery: a prospective randomized clinical trial. Thorac Cardiovasc Surg. 2017;65:362–6.CrossRef
4.
Zurück zum Zitat Negishi C, Hasegawa K, Mukai S, et al. Resistive-heating and forced-air warming are comparably effective. Anesth Analg. 2003;96:1683–7.CrossRef Negishi C, Hasegawa K, Mukai S, et al. Resistive-heating and forced-air warming are comparably effective. Anesth Analg. 2003;96:1683–7.CrossRef
5.
Zurück zum Zitat Röder G, Sessler DI, Roth G, et al. Intra-operative rewarming with Hot Dog® resistive heating and forced-air heating: a trial of lower-body warming. Anaesthesia. 2011;66:667–74.CrossRef Röder G, Sessler DI, Roth G, et al. Intra-operative rewarming with Hot Dog® resistive heating and forced-air heating: a trial of lower-body warming. Anaesthesia. 2011;66:667–74.CrossRef
Metadaten
Titel
Conductive heating mattress leads to ECG changes that mimic pacemaker spikes
verfasst von
A. Bräuer
R. Franke
A. von Hammerstein-Equord
Publikationsdatum
15.07.2020
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 3/2021
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-020-00520-z

Weitere Artikel der Ausgabe 3/2021

Journal of Clinical Monitoring and Computing 3/2021 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.