Erschienen in: Breast Cancer Research and Treatment 1/2015

Open Access 01.01.2015 | Brief Report

Confocal fluorescence microscopy for rapid evaluation of invasive tumor cellularity of inflammatory breast carcinoma core needle biopsies

verfasst von: Jessica Dobbs, Savitri Krishnamurthy, Matthew Kyrish, Ana Paula Benveniste, Wei Yang, Rebecca Richards-Kortum

Erschienen in: Breast Cancer Research and Treatment | Ausgabe 1/2015

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Abstract

Tissue sampling is a problematic issue for inflammatory breast carcinoma, and immediate evaluation following core needle biopsy is needed to evaluate specimen adequacy. We sought to determine if confocal fluorescence microscopy provides sufficient resolution to evaluate specimen adequacy by comparing invasive tumor cellularity estimated from standard histologic images to invasive tumor cellularity estimated from confocal images of breast core needle biopsy specimens. Grayscale confocal fluorescence images of breast core needle biopsy specimens were acquired following proflavine application. A breast-dedicated pathologist evaluated invasive tumor cellularity in histologic images with hematoxylin and eosin staining and in grayscale and false-colored confocal images of cores. Agreement between cellularity estimates was quantified using a kappa coefficient. 23 cores from 23 patients with suspected inflammatory breast carcinoma were imaged. Confocal images were acquired in an average of less than 2 min per core. Invasive tumor cellularity estimated from histologic and grayscale confocal images showed moderate agreement by kappa coefficient: κ = 0.48 ± 0.09 (p < 0.001). Grayscale confocal images require less than 2 min for acquisition and allow for evaluation of invasive tumor cellularity in breast core needle biopsy specimens with moderate agreement to histologic images. We show that confocal fluorescence microscopy can be performed immediately following specimen acquisition and could indicate the need for additional biopsies at the initial visit.

Introduction

Inflammatory breast carcinoma (IBC) is a rare, aggressive type of breast cancer that is characterized clinically by tenderness, erythema, and edema involving at least one-third of the entire breast [14]. This variant of breast cancer can be clinically mistaken for mastitis and treated with antibiotics resulting in a delay in the recognition of malignancy. The redness and edema of the breast is attributed to the frequent involvement of the lymphovascular channels in the breast. Pathologic diagnosis can be made using skin punch biopsy and/or core needle biopsy (CNB) of the affected breast. The presence of lymphovascular tumor emboli in the skin biopsy and invasive mammary carcinoma in the core biopsy can establish the diagnosis and initiate the clinical management of the patients. Tissue sampling can be a particularly problematic issue in IBC, since most cases present with no discrete mass and therefore accurate targeting of the area of breast for image-guided CNBs can be difficult [5, 6]. Shah and colleagues showed that sampling error can lead to inadequate CNB specimens, which have an insufficient amount of tissue to make a diagnosis, in as many as 9.1 % of cases [7]. In cases with sampling error, several more days are required to perform repeated biopsies and histologic assessment for additional tissue specimens [7, 8]. Immediate evaluation of CNBs for assessment of tumor cellularity can be extremely valuable to ensure procurement of adequate tissue not just for diagnosis of the invasive carcinoma but also for ancillary studies in one setting. Touch preparation and frozen section are the currently available techniques for rapid evaluation of breast CNB specimen adequacy. While touch preparation of breast CNBs can be useful to ascertain the presence or absence of tumor cells in the CNBs, it cannot indicate the extent of tumor cellularity in the CNB and whether the malignancy is in situ, invasive, or both. In addition, the process of touching and rolling the CNBs on the glass slides may cause distortion, which makes them less than optimal for accurate histopathological examination. Frozen section is generally not preferred for immediate assessment of CNBs because of loss of the precious tissue during the cutting process. A technique is needed to evaluate the adequacy of IBC CNB specimens as soon as they are procured in a clinical setting, without tissue preparation that may compromise the quality of the core. The technique should be able to immediately indicate to the radiologist if the area of interest has been accurately targeted and if additional biopsies are required [7].
Several studies have established the utility of confocal microscopy for breast tissue assessment [4, 912]. Schiffhauer and colleagues demonstrated the feasibility of confocal reflectance microscopy for identifying the presence of neoplasia in unfixed breast CNB specimens in real time [10]. We recently showed that confocal fluorescence microscopy can be performed in a clinical setting to acquire images of fresh breast tissue with sufficient resolution to identify neoplastic and non-neoplastic breast cytomorphological and architectural features [4].
The objective of this work was to determine the feasibility of using confocal fluorescence microscopy in a clinical setting to evaluate the adequacy of IBC CNB specimens by estimating the amount of invasive tumor cellularity in each specimen. To meet this objective, we performed a proof-of-concept study to estimate invasive tumor cellularity from grayscale confocal images and confocal images that were false-colored to mimic histologic staining; results were compared to the assessment from the gold standard of histologic images obtained from conventional light microscopic examination of formalin-fixed and paraffin-embedded tissue sections of the core biopsies stained by hematoxylin and eosin.

Materials and methods

Core needle biopsy specimens were obtained from patients with clinical signs and symptoms of IBC. The protocol for tissue acquisition was reviewed and approved by the Institutional Review Boards at The University of Texas MD Anderson Cancer Center and Rice University. Ultrasound-guided CNBs were performed per standard clinical protocol. Single cores were prepared for confocal imaging within 5 min of the core needle biopsy procedure; each core was typically 1 × 4 × 12 mm3 in size. Cores were kept moist in phosphate buffered saline (PBS) (pH 7.4) prior to imaging. A solution of 0.01 % proflavine in sterile PBS was applied topically to the core for approximately 1 min; proflavine is a fluorescent dye that preferentially stains nuclei in confocal fluorescence images [1316].
The procedure for imaging cores with confocal fluorescence microscopy has been described previously [4]. Prior to imaging, a core was positioned on the microscope stage. Each core was imaged using a scanning confocal microscope (Vivascope 2500®; Caliber Imaging and Diagnostics Inc., Rochester, New York) at 488 nm excitation with a 550 ± 44 nm bandpass filter with a 30× water immersion lens (numerical aperture = 0.8). At the center of the 750 × 750 µm2 region of interest (ROI), lateral resolution of the confocal microscope is 1.0 µm and axial resolution is 5.0 µm. Initially, illumination power was set to 2.1 mW (±0.4 mW) and adjusted to maximize signal and avoid saturation. To create a composite image of the core at a single axial depth, the microscope stage was scanned in a grid pattern (maximum area: 12.2 × 12.2 mm2); composite images of cores were typically 4 × 12.2 mm2. After imaging, the cores were stored in PBS and submitted to the IBC tissue registry for processing, including standard histologic preparation, fixation, and staining with hematoxylin and eosin (H&E).
Grayscale fluorescence images of cores were acquired with the confocal microscope. The grayscale confocal fluorescence images were false-colored to resemble histologic staining by combining the experimentally determined color values for hematoxylin and eosin reported by Gareau [17] into a single set of color values: H = [1.3 0.75 1.88], where [red (k = 1), green (k = 2), and blue (k = 3)]. False color was applied to each pixel of confocal images (C) as follows:
$$C_{x,y,k} = 1 - F_{x,y} (2 - H_{k} ),$$
where F is the grayscale confocal fluorescence image with each pixel normalized to the maximum [17]. The individual grayscale confocal images were automatically stitched together immediately following acquisition by the confocal microscope to form a composite grayscale image of each core. False-colored confocal images were manually stitched together, using the grayscale confocal image as a guide, to form a composite false-colored image of each core.
A board-certified, dedicated breast pathologist (author S. Krishnamurthy) assessed H&E-stained slides with a 10× objective lens and evaluated confocal images at comparable magnification. Invasive tumor cellularity was estimated in three images for each core: one histologic image prepared with standard fixation in neutral buffered formalin, routinely processed and embedded in paraffin followed by H&E staining, one composite grayscale confocal image, and one false-colored confocal image— using standard histologic criteria [18]. Tumor cellularity was defined as the percentage of the core comprised of invasive breast carcinoma [19].
A kappa coefficient was calculated for each combination of confocal and histologic image types (histologic vs. grayscale confocal images and histologic vs. false-colored confocal images), in order to compare the agreement between invasive tumor cellularity estimated from histologic, grayscale confocal, and false-colored confocal images. In this calculation of inter-rater agreement, each image type was considered to be one “rater.”

Results

A total of 23 cores were collected from 23 patients with untreated, clinically suspected IBC [1] (Table 1). Twenty of the cores contained neoplastic tissue, identified by standard histologic criteria in histologic images. The average time to acquire a composite confocal image was 1 min, 54 s (Table 1). False coloring produced confocal images that visually resembled tissue sections stained with hematoxylin and eosin.
Table 1
Percent invasive tumor cellularity in IBC cores estimated in histologic, grayscale confocal, and false-colored confocal images. Imaging time indicates the time to acquire a composite grayscale confocal image. The average imaging time was 01:54 mm:ss. Histologic IBC Type and Tumor Grade indicate the histologic type of inflammatory breast cancer and tumor grade, respectively, based on histologic assessment of all tissue specimens (including the core reviewed in this study) for each corresponding patient case
Adequacy for tumor diagnosis
Core #
Estimated invasive tumor cellularity in images of core needle biopsy specimens
Imaging time (MM:SS)
Histologic IBC type
Tumor grade
Histology (H&E)
Confocal (grayscale)
Confocal (false coloring)
Adequate
1
90 %
75–80 %
80 %
00:53
Invasive ductal carcinoma
2
2
5 %
5 %
5 %
00:53
Invasive ductal carcinoma with mucinous carcinoma
3
3
10 %
25 %
20 %
02:41
Invasive ductal carcinoma
3
4
20 %
5 %
20 %
01:47
Invasive ductal carcinoma
2
5
75 %
75 %
20 %
02:51
Invasive ductal carcinoma
3
6
25 %
20 %
20 %
02:07
Invasive ductal carcinoma
3
7
20 %
20 %
20 %
02:54
Invasive lobular carcinoma
3
8
100 %
100 %
30 %
02:51
Invasive ductal carcinoma
3
9
5 %
5–10 %
5 %
02:51
Invasive ductal carcinoma
3
10
25–30 %
25 %
40 %
01:46
Invasive lobular carcinoma
2
11
95 %
80–95 %
90 %
02:33
Invasive ductal carcinoma
3
12
30 %
30 %
30 %
02:09
Invasive ductal carcinoma
3
13
30 %
20 %
50 %
01:01
Invasive ductal carcinoma
3
14
95 %
70 %
90 %
01:04
Invasive ductal carcinoma
3
15
80 %
75 %
90 %
01:04
Invasive ductal carcinoma with micropapillary features
2
16
70 %
90–95 %
95 %
01:04
Invasive ductal carcinoma
3
17
20 %
10 %
10 %
01:04
Invasive ductal and lobular carcinoma
1
18
50 %
10–15 %
15 %
02:51
Invasive lobular carcinoma
3
19
25 %
30 %
25 %
01:47
Invasive ductal carcinoma
2
20
30–40 %
40 %
20 %
01:25
Invasive ductal carcinoma
3
Inadequate
21
0 %
0 %
5 %
02:51
Invasive ductal carcinoma
3
22
0 %
0 %
0 %
01:04
Invasive ductal carcinoma
3
23
0 %
0 %
0 %
01:47
Invasive ductal carcinoma
2
H&E hematoxylin and eosin stain
Representative histologic, grayscale confocal, and false-colored confocal images of cores that contain IBC tissue are shown in Figs. 1 and 2. Figure 1 shows a core that contains both neoplastic and non-neoplastic tissue, with an estimated 25 % cellularity in the histologic image; cellularity was estimated as 30 and 25 % in grayscale confocal and false-colored confocal images, respectively. Figure 2 shows a core with 80 % cellularity estimated in the histologic image, 75 % cellularity estimated in the grayscale confocal image, and 90 % cellularity estimated in the false-colored confocal image.
Three cores acquired from patients diagnosed with IBC were considered to be inadequate as a result of sampling error; the histologic image contained no neoplastic tissue despite a clinical diagnosis of IBC. Representative images of an inadequate core consisting of adipose tissue and no neoplastic tissue, based on assessment of the histologic image, are shown in Fig. 3. All three images of the core estimated cellularity as 0 %.
Estimates of tumor cellularity based on histologic, grayscale confocal, and false-colored confocal images for all specimens are summarized in Table 1. Figure 4 illustrates the agreement between tumor cellularity estimated in histologic and grayscale confocal images (Fig. 4a) and in histologic and false-colored confocal images (Fig. 4b). Each data point represents a core needle biopsy specimen from which tumor cellularity was estimated. Dashed lines indicate the linear fit for each set of data points and Pearson coefficient (R 2) values quantify the linear fit of the data. Tumor cellularity estimates were more consistent between histologic and grayscale confocal images (R 2 = 0.88) than between histologic and false-colored confocal images (R 2 = 0.61).
A kappa coefficient was calculated [20, 21] to quantify the agreement between histologic images and either grayscale or false-colored images. Standard error and p values for kappa were also quantified (Table 2). The kappa value calculated for agreement between histologic and grayscale confocal images indicates moderate agreement based on defined categories for kappa values [21, 22]; κ = 0.48 ± 0.09 (p < 0.001). Agreement between histologic and false-colored confocal images was fair and not statistically significant; κ = 0.28 ± 0.26 (p = 0.14).
Table 2
Inter-rater agreement of estimated invasive tumor cellularity between the types of images, quantified by the kappa coefficient and standard error [20]. Level of agreement is based on categories described by Viera and Garrett [21]. P value was calculated using a normal distribution
Inter-rater agreement—two techniques
Image types
Kappa coefficient
Standard error
Level of agreement
P value
Histologic versus grayscale confocal
0.477
0.088
Moderate
3.43E−08
Histologic versus false-colored confocal
0.280
0.258
Fair
0.139

Discussion

The objective of this work was to determine the feasibility of using confocal fluorescence microscopy for immediate assessment of invasive tumor cellularity of CNBs in patients with IBC. Composite grayscale confocal images of CNB specimens were acquired in an average time of 2 min. Statistical analysis shows that there is a moderate agreement between tumor cellularity estimated from histologic images and grayscale confocal images. False-coloring the confocal images using a single analog for histologic staining does not improve agreement. These findings demonstrate that confocal fluorescence microscopy could be used to assess adequacy of IBC CNB specimens. This technique may be useful to ensure procurement of high-quality CNBs not only for routine histopathological diagnosis but also for ancillary studies including immunohistochemistry and molecular testing.
Of the three CNB specimens that were determined to be inadequate by histologic assessment, one had benign lobules and fibrous tissue (core #21) and two consisted completely of adipose tissue (cores #22, 23). These cores were identified as inadequate based on tumor cellularity estimated as 0–5 % in histologic, grayscale confocal, and false-colored confocal images (Table 1). Of the 20 CNB specimens that were determined to be adequate by histologic assessment, 20 were diagnosed as invasive carcinoma; morphological features of invasive carcinoma were also recognized in grayscale confocal and false-colored confocal images. One CNB specimen was diagnosed as mucinous carcinoma (core #2) based on standard histologic criteria [18] and was also recognized as such on the confocal images.
IBC is an aggressive type of breast cancer, which is known to metastasize through the lymphatic system and typically presents as a non-palpable lesion [13]. To obtain a rapid and accurate diagnosis of IBC, there must be a reliable method to sample tissue. Although inter-rater agreement is typically high for histologic diagnosis [2326], the rate of inadequate samples is a limitation of CNB as a sampling technique for breast cancer diagnosis [7, 8, 23, 24, 27]. Shah and colleagues showed that the rate of sampling error leading to inadequate CNB specimens can be as high as 9.1 % [7]. Therefore, there is a need for a technique that can evaluate specimen adequacy [7] with a relatively high sensitivity and specificity.
Previous studies have illustrated the feasibility of evaluating breast tissue specimens with confocal microscopy [4, 912]. Schiffhauer and colleagues evaluated confocal reflectance microscopy for screening unfixed breast CNB specimens with a range of neoplastic and non-neoplastic histologic features. Their study showed that there were adequate morphologic and cellular features in confocal reflectance images to allow correlation to histologic images [10]. In this study, we chose to focus on inflammatory breast cancer, because we wanted to determine if assessment of confocal fluorescence microscopy images of fresh cores would show which cores had insufficient tumor tissue for diagnosis and therefore identify cases for which a repeat biopsy was needed. Our previous work showed that confocal fluorescence microscopy images provide sufficient detail to distinguish between neoplastic breast features, such as ductal carcinoma in situ and invasive ductal and lobular carcinoma, and non-neoplastic breast features including benign ducts, lobules, and hyperplasia. [4] We characterized inter-rater agreement between pathologists who assessed confocal fluorescence with a kappa coefficient for multiple ratings per subject. The kappa coefficient described in our previous work (0.51, p < 0.001) between pathologists assessing confocal images indicated moderate agreement based on defined categories of kappa coefficients [21, 22]. This level of agreement is similar to the moderate agreement between a single rater’s assessment of grayscale confocal and histologic images shown in Table 2 (0.477, p < 0.001). The findings of these two studies suggest that confocal fluorescence images could potentially be used to assess adequacy of breast core needle biopsies with many histologic features.
False coloring to mimic histologic staining in confocal images of breast tissue has been demonstrated in two studies [17, 28]. Gareau showed that the false-coloring technique using confocal reflectance and fluorescence microscopy to mimic hematoxylin and eosin staining, respectively, produced multimodal confocal images mimicking the appearance of corresponding histologic images [17]. Bini and colleagues used false coloring for tissue obtained from Mohs skin excisions and demonstrated correlation between confocal mosaics and histologic images [28]. Previous studies showed that false coloring improved correlation between confocal and histologic images [21, 22]. However, in contrast to the previous studies which used both reflectance and fluorescence images to mimic hematoxylin and eosin staining, we used only fluorescence images to mimic histologic staining of the tissue section. It was not possible to use multimodal confocal images because reflectance images were not collected for every CNB specimen and because some confocal reflectance images were not aligned laterally or axially with the confocal fluorescence images of the corresponding core needle biopsy specimen.
In conclusion, we present the results of a proof-of-concept study comparing estimated invasive tumor cellularity between grayscale confocal, false-colored confocal, and histologic images of 23 IBC CNB specimens. Grayscale confocal images require an average of only 2 min for acquisition and allow immediate evaluation of invasive tumor cellularity in IBC CNB specimens; agreement with histologic images is moderate. We show that confocal fluorescence microscopy could be used for immediate assessment of CNB specimen adequacy, so that additional biopsies for diagnosis, bio-banking, or genetic sequencing studies can be performed on site without need for a repeat visit. Additional larger prospective studies are warranted to validate our findings for the potential utilization of confocal fluorescence microscopy in routine pathology practice as a robust technique for immediate assessment of specimen adequacy of IBC CNBs.

Acknowledgments

The authors would like to thank Dongsuk Shin for his assistance with the code for false-coloring confocal fluorescence images, and acknowledge support from the Department of Defense Era of Hope Award W81XWH-09-1-0410, and the Susan G. Komen for the Cure grant KG091020. This work was also supported in part by The State of Texas Grant for Rare and Aggressive Cancers through the Morgan Welch Inflammatory Breast Cancer Research Program.

Conflicts of interest

The authors declare that they have no conflicts of interest.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Gynäkologie

Kombi-Abonnement

Mit e.Med Gynäkologie erhalten Sie Zugang zu CME-Fortbildungen der beiden Fachgebiete, den Premium-Inhalten der Fachzeitschriften, inklusive einer gedruckten gynäkologischen oder urologischen Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Yamauchi H, Woodward WA, Valero V, Alvarez RH, Lucci A, Buchholz TA et al (2012) Inflammatory breast cancer: what we know and what we need to learn. Oncologist 17:891–899PubMedCentralPubMedCrossRef Yamauchi H, Woodward WA, Valero V, Alvarez RH, Lucci A, Buchholz TA et al (2012) Inflammatory breast cancer: what we know and what we need to learn. Oncologist 17:891–899PubMedCentralPubMedCrossRef
2.
Zurück zum Zitat Hance KW, Anderson WF, Devesa SS, Young HA, Levine PH (2005) Trends in inflammatory breast carcinoma incidence and survival: the surveillance, epidemiology, and end results program at the National Cancer Institute. J Natl Cancer Inst 97:966–975PubMedCentralPubMedCrossRef Hance KW, Anderson WF, Devesa SS, Young HA, Levine PH (2005) Trends in inflammatory breast carcinoma incidence and survival: the surveillance, epidemiology, and end results program at the National Cancer Institute. J Natl Cancer Inst 97:966–975PubMedCentralPubMedCrossRef
3.
Zurück zum Zitat Anderson WF, Chu KC, Chang S (2003) Inflammatory breast carcinoma and noninflammatory locally advanced breast carcinoma: distinct clinicopathologic entities? J Clin Oncol 21:2254–2259PubMedCrossRef Anderson WF, Chu KC, Chang S (2003) Inflammatory breast carcinoma and noninflammatory locally advanced breast carcinoma: distinct clinicopathologic entities? J Clin Oncol 21:2254–2259PubMedCrossRef
4.
Zurück zum Zitat Dobbs JL, Ding H, Benveniste AP, Kuerer HM, Krishnamurthy S, Yang W, Richards-Kortum R (2013) Feasibility of confocal fluorescence microscopy for real-time evaluation of neoplasia in fresh human breast tissue. J Biomed Opt. doi:10.1117/1.JBO.18.10.106016 PubMedCentral Dobbs JL, Ding H, Benveniste AP, Kuerer HM, Krishnamurthy S, Yang W, Richards-Kortum R (2013) Feasibility of confocal fluorescence microscopy for real-time evaluation of neoplasia in fresh human breast tissue. J Biomed Opt. doi:10.​1117/​1.​JBO.​18.​10.​106016 PubMedCentral
5.
Zurück zum Zitat Kumar N, Sayed S, Moloo Z, Wasike R (2011) Fine-needle aspiration in suspected inflammatory breast cancer: case series with emphasis on approach to specimen adequacy. Acta Cytol 55:239–244PubMedCrossRef Kumar N, Sayed S, Moloo Z, Wasike R (2011) Fine-needle aspiration in suspected inflammatory breast cancer: case series with emphasis on approach to specimen adequacy. Acta Cytol 55:239–244PubMedCrossRef
6.
Zurück zum Zitat Iwase T, Takahashi K, Gomi N, Horii R, Akiyama F (2006) Present state of and problems with core needle biopsy for non-palpable breast lesions. Breast Cancer 13:32–37 Tokyo, JapanPubMedCrossRef Iwase T, Takahashi K, Gomi N, Horii R, Akiyama F (2006) Present state of and problems with core needle biopsy for non-palpable breast lesions. Breast Cancer 13:32–37 Tokyo, JapanPubMedCrossRef
7.
Zurück zum Zitat Shah VI, Raju U, Chitale D, Deshpande V, Gregory N, Strand V (2003) False-negative core needle biopsies of the breast—an analysis of clinical, radiologic, and pathologic findings in 27 consecutive cases of missed breast cancer. Cancer 97:1824–1831PubMedCrossRef Shah VI, Raju U, Chitale D, Deshpande V, Gregory N, Strand V (2003) False-negative core needle biopsies of the breast—an analysis of clinical, radiologic, and pathologic findings in 27 consecutive cases of missed breast cancer. Cancer 97:1824–1831PubMedCrossRef
8.
Zurück zum Zitat Dillon MF, Hill ADK, Quinn CM, O’Doherty A, McDermott EW, O’Higgins N (2005) The accuracy of ultrasound, stereotactic, and clinical core biopsies in the diagnosis of breast cancer, with an analysis of false-negative cases. Ann Surg 242:701–707PubMedCentralPubMedCrossRef Dillon MF, Hill ADK, Quinn CM, O’Doherty A, McDermott EW, O’Higgins N (2005) The accuracy of ultrasound, stereotactic, and clinical core biopsies in the diagnosis of breast cancer, with an analysis of false-negative cases. Ann Surg 242:701–707PubMedCentralPubMedCrossRef
9.
Zurück zum Zitat Parrish A, Halama E, Tilli MT, Freedman M, Furth PA (2005) Reflectance confocal microscopy for characterization of mammary ductal structures and development of neoplasia in genetically engineered mouse models of breast cancer. J Biomed Opt. doi: 10.1117/1.2065827 Parrish A, Halama E, Tilli MT, Freedman M, Furth PA (2005) Reflectance confocal microscopy for characterization of mammary ductal structures and development of neoplasia in genetically engineered mouse models of breast cancer. J Biomed Opt. doi: 10.​1117/​1.​2065827
10.
Zurück zum Zitat Schiffhauer LM, Boger JN, Bonfiglio TA, Zavislan JM, Zuley M, Fox CA (2009) Confocal Microscopy of Unfixed Breast Needle Core Biopsies: A Comparison to Fixed and Stained Sections. BMC Cancer. doi: 10.1186/1471-2407-9-265 Schiffhauer LM, Boger JN, Bonfiglio TA, Zavislan JM, Zuley M, Fox CA (2009) Confocal Microscopy of Unfixed Breast Needle Core Biopsies: A Comparison to Fixed and Stained Sections. BMC Cancer. doi: 10.​1186/​1471-2407-9-265
11.
Zurück zum Zitat Tilli MT, Cabrera MC, Parrish AR, Torre KM, Sidawy MK, Gallagher AL, et al (2007) Real-time imaging and characterization of human breast tissue by reflectance confocal microscopy. J Biomed Opt. doi: 10.1117/1.2799187 Tilli MT, Cabrera MC, Parrish AR, Torre KM, Sidawy MK, Gallagher AL, et al (2007) Real-time imaging and characterization of human breast tissue by reflectance confocal microscopy. J Biomed Opt. doi: 10.​1117/​1.​2799187
12.
Zurück zum Zitat Abeytunge S, Li Y, Larson B, Peterson G, Seltzer E, Toledo-Crow R, et al (2013) Confocal microscopy with strip mosaicing for rapid imaging over large areas of excised tissue. J Biomed Opt. doi: 10.1117/1.JBO.18.6.061227 Abeytunge S, Li Y, Larson B, Peterson G, Seltzer E, Toledo-Crow R, et al (2013) Confocal microscopy with strip mosaicing for rapid imaging over large areas of excised tissue. J Biomed Opt. doi: 10.​1117/​1.​JBO.​18.​6.​061227
13.
Zurück zum Zitat Muldoon TJ, Anandasabapathy S, Maru D, Richards-Kortum R (2008) High-resolution imaging in Barrett’s esophagus: a novel, low-cost endoscopic microscope. Gastrointest Endosc 68:737–744PubMedCentralPubMedCrossRef Muldoon TJ, Anandasabapathy S, Maru D, Richards-Kortum R (2008) High-resolution imaging in Barrett’s esophagus: a novel, low-cost endoscopic microscope. Gastrointest Endosc 68:737–744PubMedCentralPubMedCrossRef
14.
Zurück zum Zitat Muldoon TJ, Pierce MC, Nida DL, Williams MD, Gillenwater A, Richards-Kortum R (2007) Subcellular-resolution molecular imaging within living tissue by fiber microendoscopy. Opt Express 15:16413–16423PubMedCentralPubMedCrossRef Muldoon TJ, Pierce MC, Nida DL, Williams MD, Gillenwater A, Richards-Kortum R (2007) Subcellular-resolution molecular imaging within living tissue by fiber microendoscopy. Opt Express 15:16413–16423PubMedCentralPubMedCrossRef
15.
Zurück zum Zitat Rosbach KJ, Shin D, Muldoon TJ, Quraishi MA, Middleton LP, Hunt KK et al (2010) High-resolution fiber optic microscopy with fluorescent contrast enhancement for the identification of axillary lymph node metastases in breast cancer: a pilot study. Biomed Opt Express 1:911–922PubMedCentralPubMedCrossRef Rosbach KJ, Shin D, Muldoon TJ, Quraishi MA, Middleton LP, Hunt KK et al (2010) High-resolution fiber optic microscopy with fluorescent contrast enhancement for the identification of axillary lymph node metastases in breast cancer: a pilot study. Biomed Opt Express 1:911–922PubMedCentralPubMedCrossRef
16.
Zurück zum Zitat Pierce MC, Javier DJ, Richards-Kortum R (2008) Optical contrast agents and imaging systems for detection and diagnosis of cancer. Int J Cancer 123:1979–1990PubMedCentralPubMedCrossRef Pierce MC, Javier DJ, Richards-Kortum R (2008) Optical contrast agents and imaging systems for detection and diagnosis of cancer. Int J Cancer 123:1979–1990PubMedCentralPubMedCrossRef
17.
Zurück zum Zitat Gareau DS (2009) Feasibility of digitally stained multimodal confocal mosaics to simulate histopathology. J Biomed Opt. doi: 10.1117/1.3149853 Gareau DS (2009) Feasibility of digitally stained multimodal confocal mosaics to simulate histopathology. J Biomed Opt. doi: 10.​1117/​1.​3149853
18.
Zurück zum Zitat Rosen PP (2008) Rosen’s breast pathology, 3rd edn. Lippincott Williams & Wilkins, Philadelphia Rosen PP (2008) Rosen’s breast pathology, 3rd edn. Lippincott Williams & Wilkins, Philadelphia
19.
Zurück zum Zitat Rajan R, Poniecka A, Smith TL, Yang Y, Frye D, Pusztai L et al (2004) Change in tumor cellularity of breast carcinoma after neoadjuvant chemotherapy as a variable in the pathologic assessment of response. Cancer 100:1365–1373PubMedCrossRef Rajan R, Poniecka A, Smith TL, Yang Y, Frye D, Pusztai L et al (2004) Change in tumor cellularity of breast carcinoma after neoadjuvant chemotherapy as a variable in the pathologic assessment of response. Cancer 100:1365–1373PubMedCrossRef
20.
Zurück zum Zitat Fleiss JL, Levin B, Paik MC (2003) The measurement of interrater agreement. Statistical methods for rates and proportions, 3rd edn. Wiley, HobokenCrossRef Fleiss JL, Levin B, Paik MC (2003) The measurement of interrater agreement. Statistical methods for rates and proportions, 3rd edn. Wiley, HobokenCrossRef
21.
Zurück zum Zitat Viera AJ, Garrett JM (2005) Understanding interobserver agreement: the kappa statistic. Fam Med 37:360–363PubMed Viera AJ, Garrett JM (2005) Understanding interobserver agreement: the kappa statistic. Fam Med 37:360–363PubMed
22.
Zurück zum Zitat Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174PubMedCrossRef Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174PubMedCrossRef
23.
Zurück zum Zitat Perrot N, Jalaguier-Coudray A, Frey I, Thomassin-Naggara I, Chopier J (2013) US-guided core needle biopsy: false-negatives. How to reduce them? Eur J Radiol 82:424–426PubMedCrossRef Perrot N, Jalaguier-Coudray A, Frey I, Thomassin-Naggara I, Chopier J (2013) US-guided core needle biopsy: false-negatives. How to reduce them? Eur J Radiol 82:424–426PubMedCrossRef
24.
Zurück zum Zitat Youk JH, Kim E-K, Kim MJ, Kwak JY, Son EJ (2010) Analysis of false-negative results after US-guided 14-gauge core needle breast biopsy. Eur Radiol 20:782–789PubMedCrossRef Youk JH, Kim E-K, Kim MJ, Kwak JY, Son EJ (2010) Analysis of false-negative results after US-guided 14-gauge core needle breast biopsy. Eur Radiol 20:782–789PubMedCrossRef
25.
Zurück zum Zitat Garg S, Mohan H, Bal A, Attri AK, Kochhar S (2007) A comparative analysis of core needle biopsy and fine-needle aspiration cytology in the evaluation of palpable mammographically suspicious breast detected lesions. Diagn Cytopathol 35:681–689PubMedCrossRef Garg S, Mohan H, Bal A, Attri AK, Kochhar S (2007) A comparative analysis of core needle biopsy and fine-needle aspiration cytology in the evaluation of palpable mammographically suspicious breast detected lesions. Diagn Cytopathol 35:681–689PubMedCrossRef
26.
Zurück zum Zitat Collins LC, Connolly JL, Page DL, Goulart RA, Pisano ED, Fajardo LL et al (2004) Diagnostic agreement in the evaluation of image-guided breast core needle biopsies - Results from a randomized clinical. Am J Surg Pathol 28:126–131PubMedCrossRef Collins LC, Connolly JL, Page DL, Goulart RA, Pisano ED, Fajardo LL et al (2004) Diagnostic agreement in the evaluation of image-guided breast core needle biopsies - Results from a randomized clinical. Am J Surg Pathol 28:126–131PubMedCrossRef
27.
Zurück zum Zitat Peter D, Gruenhagen J, Wenke R, Schaefer FKW, Schreer I (2008) False-negative results after stereotactically guided vacuum biopsy. Eur Radiol 18:177–182PubMedCrossRef Peter D, Gruenhagen J, Wenke R, Schaefer FKW, Schreer I (2008) False-negative results after stereotactically guided vacuum biopsy. Eur Radiol 18:177–182PubMedCrossRef
28.
Zurück zum Zitat Bini J, Spain J, Nehal K, Hazelwood V, DiMarzio C, Rajadhyaksha M (2011) Confocal mosaicing microscopy of human skin ex vivo: spectral analysis for digital staining to simulate histology-like appearance. J Biomed Opt. doi: 10.1117/1.3596742 Bini J, Spain J, Nehal K, Hazelwood V, DiMarzio C, Rajadhyaksha M (2011) Confocal mosaicing microscopy of human skin ex vivo: spectral analysis for digital staining to simulate histology-like appearance. J Biomed Opt. doi: 10.​1117/​1.​3596742
Metadaten
Titel
Confocal fluorescence microscopy for rapid evaluation of invasive tumor cellularity of inflammatory breast carcinoma core needle biopsies
verfasst von
Jessica Dobbs
Savitri Krishnamurthy
Matthew Kyrish
Ana Paula Benveniste
Wei Yang
Rebecca Richards-Kortum
Publikationsdatum
01.01.2015
Verlag
Springer US
Erschienen in
Breast Cancer Research and Treatment / Ausgabe 1/2015
Print ISSN: 0167-6806
Elektronische ISSN: 1573-7217
DOI
https://doi.org/10.1007/s10549-014-3182-5