Open Access 01.12.2022 | COVID-19 | Research

Diagnostic power of one-step and two-step RT-qPCR methods to SARS‑CoV‑2 detection

verfasst von: Asra Malekshahi, Sayyad Khanizadeh, Shirzad Fallahi, Gholamreza Talei, Mehdi Birjandi, Faezeh Hajizadeh

Erschienen in: BMC Infectious Diseases | Ausgabe 1/2022

Abstract

Background

Coronavirus-2019 (COVID-2019) is a novel coronavirus known as Acute Respiratory Syndrome (SARS-CoV-2). The premier standard test for SARS-CoV-2 diagnosis is a one-step RT-qPCR method, which requires specific probes and reagents. Therefore, detection on a large scale is expensive and cannot be very accurate.

Methods

A cost-effective technique based on SYBR green was evaluated in the current study. The specific primers for S and N genes were designed, then performed the cross-reactivity test with other coronavirus and respiratory viruses positive samples. Moreover, the analytical sensitivity test was carried out with 8 dilutions (1:10). Lastly, the SARS-CoV-2 clinical samples (n = 210) were tested by these two methods, and receiver operating characteristic (ROC) analysis was performed to investigate the incremental diagnostic value of each gene in the study methods.

Results

The two-step method detected up to 6th dilutions of the SARS-CoV-2 samples and did not show any amplification of the positive samples of other respiratory viruses. ROC analysis revealed a diagnostic ability of the two-step method for SARS-CoV-2 with an area under the ROC curve of ≥ 0.7 (P ˂ 0.05) and relatively high sensitivity and specificity. The combination of N and S genes increased the sensitivity up to 88%, specificity up to 86%, and area under the ROC curve up to 0.85 (95% confidence interval (95% CI) 0.72 to 0.93, P = 0.0461).

Conclusion

Our findings indicated that the two-step method has comparable sensitivity and specificity to the one-step method. Therefore, this method can be considered a potential diagnostic method for diagnosing and monitoring COVID-19 patients. It suggests that when the one-step RT-qPCR method is not available, the two-step RT-qPCR can be used to identify SARS-CoV-2.
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
COVID-19
Coronavirus disease 19
SARS-CoV-2
Severe acute respiratory syndrome coronavirus 2
RT-qPCR
Quantitative reverse transcription PCR
ddPCR
Droplet digital PCR
LAMP
Loop-mediated isothermal amplification

Introduction

Coronavirus disease-2019 (COVID-2019) is known as Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2). The COVID-19 outbreak by the new coronavirus strain was recognized as a pandemic by the World Health Organization (WHO) on 11 March 2020 [1]. COVID-19 has high infections and mortality worldwide; moreover, there have been other consequences, including damage to education, social interactions, and the world economy [2, 3]. Isolation, social distancing, and early detection have been the main tools in the fight to interrupt the chain of viral transmission. Some individuals may be asymptomatic, making it difficult to diagnose the virus [4]. The SARS-CoV-2 testing capacity is an important issue worldwide. Development of the testing capacity is critical for quick detection and screening of cases during outbreaks, particularly in developing countries where the availability of supplies or infrastructure to carry out the real-time PCR test is limited [5]. Consequently, to develop alternative methods for SARS-CoV-2 detection that might be faster or cheaper to perform, different research has been conducted, such as sample pooling of RNA extracts [6], two steps endpoint RT‑PCR with agarose gel electrophoresis [7], loop-mediated isothermal amplification (LAMP) [8, 9], multiplex PCR [10], droplet digital PCR (ddPCR) [11] or even protocols based on CRISPR–Cas12 [12]. Reverse transcription-quantitative real-time PCR (RT-qPCR) is the foremost test for diagnosing SARS-CoV-2 in a clinic [13]. The RT-qPCR is working based on the amplification of the RNA target followed by incremental fluorescence dye. It is performed at two levels, i.e., TaqMan probe-based and SYBR Green-based RT-qPCR. However, each has some advantages and disadvantages. Although both methods utilize expensive reagents and equipment [14, 15], the TaqMan probe-based method is more costly because of using a specific probe.
On the other hand, the SYBR Green-based method has a lower accuracy as primer tends to form non-specific products and primer-dimer, although trained technicians can avoid it. In addition, this method has been proposed and used for testing different types of viruses [13, 16, 17]. Altogether, the one-step method is preferred to the two-step manner [15, 18]. The present study compared the one-step TaqMan probe-based method and the two-step SYBR Green-based method.

Material and method

Samples recruitment

Positive (n = 110) and negative (n = 100) specimens were recruited from the COVID-19 Reference Laboratory, Lorestan University of Medical Sciences Khorramabad, Iran; based on Taq Man RT-qPCR results recommended by WHO [19]. All positive subjects have a cycle threshold (Ct) value < 40, and negative samples have a Ct value > 40. Swab samples were collected in the safety level 2 laboratory. Our study was performed between June 2019 and November 2021, and the Committee approved the investigation of Clinical Research Ethicals in Lorestan University of Medical Sciences, Khorramabad, Iran (Ethical code: IR.LUMS.REC.1400.115). Demographic characteristics were recorded at the time of sampling.

Sample preparation

The swab was inserted into the nasopharyngeal and oropharyngeal ducts of individuals who had COVID-19 symptoms and put the swabs in 2 ml of fresh viral transport media (VTM). The samples were transferred into the refrigerator for RT-qPCR assessment and stored at 4 °C.

Pooled sample and serial dilution

We used the pooling sample process to enhance collected samples’ analytical capacity and efficiency of collected samples. The 20 positive specimens were collected into a container and diluted in the ratio of 1:10 (8-times). Afterward, the copy number of each concentration was calculated at Pishtaztab Lab using by SARS-CoV-2 standard sample.

Viral genome extraction and RT-qPCR

According to the manufacturer’s protocol, total RNA was extracted by RNJia Virus Kit solutions (Rojeh Technology Company, Tehran, Iran). The total amount of extracted RNA was evaluated using a nanodrop 1000 spectrophotometer (Wilmington, USA) and qualified by agarose gel electrophoresis. Reverse transcription of extracted RNA into cDNA in a two-step manner was applied using a cDNA synthesis kit (Takara, Japan). Quantitative reverse transcription PCR (RT-qPCR) was implemented using SYBR green (Prime Script RT Master Mix, Takara) and TaqMan probe (Pishtaztab, Tehran, Iran), and MIC PCR (BioMolecular System, Australia). For SYBR Green-based RT-qPCR method, Master Mix Green SYBR Green and specific primers designed for S and N genes were used. For this method, each 20 μl reaction contained 10 μl of Master Mix, 0.6 μl of primer (0.25 μM final concentration each), 7.8 μl of nuclease-free water, and 1 μl of cDNA with the following conditions: pre-denaturation at 95 °C for 35 s, followed by 40 cycles of denaturation at 95 °C for 5 s, and extension at 62 °C for 1 min. The reverse and forward primer sequences for the two-step method (SYBR Green-based RT-qPCR method) (Table 1) were then used.
Table 1
Sequence of primers designed for two step RT-qPCR method
Gene
Primer
Primer length (bp)
Product length
S
F 5ʹ-TCCATCAAAACCAAGCAAGA-3ʹ
20
245 bp
 
R 5ʹ-CACCAAAGGTCCAACCAGAA-3ʹ
20
NC_045512v2:23,986 + 24,230
N
F 5ʹ-CTGGACTTCCCTATGGTGCT-3ʹ
20
284 bp
 
R 5ʹ-ATTGCCAGCCATTCTAGCAG-3ʹ
20
NC_045512v2:28,629 + 28,912
The TaqMan probe-based method was performed using a commercial one-step RTqPCR Kit (Pishtaztab, Tehran, Iran). This kit’s probe and primer mixtures were designed using the dual-target gene method, which simultaneously identifies protected genomic sequences of the RdRp region and the N-protein nucleocapsid. The one-step qPCR was performed according to the manufacturer’s instructions. Briefly, each 20 μl reaction contained 9 μl of Master Mix (Enzyme mix + 5× buffer real-time PCR), 1 μl of primer, probe (0.3 μM final concentration each), 5 μl of nuclease-free water, and 5 μl of RNA. Thermal cycling was: 50 °C for 20 min for reverse transcription, inactivation of reverse transcriptase at 95 °C for 3 min, and then 40 cycles of 94 °C for 10 s and 55 °C for the 40 s.
Non-template control (nuclease-free water) was included in every qPCR run as a negative control.

Determination of analytical specificity

To confirm the analytical specificity of the SYBR Green real-time RT-PCR, the RNA was extracted from OC43, 229E, NL63, Influenza A virus (H1N1), human rhinovirus, and human respiratory syncytial virus samples. After the cDNA synthesis, the two-step manner PCR used the SARS-CoV-2 primers was carried out on the samples.

Static analysis

The crud data was analyzed statically by SPSS (19.03) and MedCalc (14.8.1) software. The ROC curve analysis was applied to find the sensitivity and specificity of each two methods. Quantitative and qualitative variables were presented as mean ± SD and frequency (percent), respectively. A P-value < 0.05 was considered statistically significant.

Results

Subjects and specimens

A total of 210 swap samples into VTM, including 110 positives and 100 negatives, were recruited in the current investigation. We matched the study groups regarding age, gender, location, lung disease, and clinical features, so there were no statistically remarkable differences (Table 2). Afterward, the positive samples were pooled together and diluted for 8-times, and finally, the genome COVID-19 copy numbers and Ct values for each concentration were calculated (Table 3). Furthermore, the standard curves for S and N genes in two-step and RdRp and N genes in one-step RT-qPCR have been shown in Fig. 1.
Table 2
Demographic characteristics of the samples
Data
Finding
Total n = 210
Positive sample
n = 110
Negative sample
n = 100
P value
n (%)
n (%)
n (%)
 
Gender
Male
52.00%
53.50%
50.40%
0.083
Female
48.00%
46.50%
49.60%
 
Age (years)
11–40
60.50%
59.80%
61.30%
 
41–70
35.50%
36.20%
34.70%
0.076
70<
3.90%
3.90%
3.90%
 
Location
City
53.90%
70.10%
37.80%
Village
46.10%
29.90%
62.20%
 
Lung disease
Yes
1.60%
3.10%
0.00%
No
98.40%
96.90%
100.00%
 
Fever
Yes
9.10%
9.40%
8.70%
0.158
No
90.90%
90.60%
91.30%
 
Sore throat
Yes
25.60%
31.50%
19.70%
0.063
No
74.40%
68.50%
80.30%
 
Headache
Yes
27.20%
28.30%
26%
0.118
No
72.80%
71.70%
74%
 
Cough
Yes
20.50%
24.40%
16.50%
0.068
No
79.50%
75.60%
83.50%
 
Table 3
Ct value and melting temperatures (Tm) of the amplified serial dilution of the one step RT-qPCR and two step RT-qPCR
Standard dilution
Copies/ml
One step RT-qPCR RdRp gene
One-step RT-qPCR N gene
Two step RT-qPCR S gene
Two step RT-qPCR N gene
Ct (Threshold 2)
Ct (Threshold 2)
\(\mathrm{Ct }(\mathrm{Threshold }0.2 )\)
Tm (°C)
\(\mathrm{Ct }(\mathrm{Threshold }0.2 )\)
Tm (°C)
0.1
109
19.97
16.44
18.56
83
19.93
85
0.01
108
21.49
19.14
24.13
83
25.49
85
0.001
107
24.03
21.56
27.52
83
28.14
85
0.0001
106
27.5
23.82
32.06
83
32.31
85
0.00001
105
31.56
27.31
36.05
83
36.91
85
0.000001
104
34.66
30.77
39.41
83
39.18
85.14
0.0000001
103
36.74
33.23
41.08
83.19
N.D
0.00000001
102
40.01
37.83
N.D
N.D

Analytical specificity of two-step RT-qPCR techniques

The analytical specificity results of the SYBR Green RT-qPCR showed no cross-attachment of the SARS-CoV-2 primers with OC43, 229E, NL63, H1N1, human rhinovirus, and human respiratory syncytial virus genes, indicating that the assay had well specificity.

RT-qPCR techniques validation in clinical samples

Each RT-qPCR manner was validated in terms of accuracy. We carried out the two-step and one-step RT-qPCR on positive and negative samples. The two-step method could detect 92% of the S gene and 88% of the N gene in positive samples. As well, it could recognize 84% of negative samples, correctly. Amplification and melt curves of clinical samples are shown in Fig. 2A–D. On the other hand, correctly, the one-step RT-qPCR method detected 92% of RdRp gene and 96% of N gene in positive and 86% of negative samples. The detection power of the one-step manner was 2.5% higher than that two-step method, but it was not significant (P = 0.086). The accuracy of the two-step method was 88% for the S gene and 86% for the N gene. The one-step was 86% for the RdRp gene and 94% for the N gene. In addition, the positive predictive value (PPV) and negative predictive value (NPV) in the two-step method for the S gene were 85% and 91%; also, they were 84% and 87% for the N gene, respectively. Furthermore, the one-step method had the PPV, and NPV values were 82% and 90% for the RdRp gene and 92% and 95% for the N gene, respectively (Table 4). There were no significant differences between PPV and NPV compared to the two methods (P > 0.05).
Table 4
Clinical performance of the compared one-step and two-step RT-qPCR for diagnosis SARS-CoV-2
Method
Target
Accuracy (%)
PPV (%)
NPV (%)
Two step RT-qPCR (SYBR Green-based)
S gene
88
85
91
N gene
86
84
87
One step RT-qPCR (Prob-based)
RdRp gene
86
82
90
N gene
94
92
95

Visualization of RT-qPCR amplicon

To investigate the specificity of amplification products, the two-step RT-qPCR amplicon of S and N genes were separated by gel electrophoresis. Positive samples of the S and N genes generated specific bands on the agarose gel (Fig. 3A and B). Then, the bands were visualized using a gel doc system. In addition, we did not observe any bands from negative samples (Fig. 3C and D).

ROC curve analysis

To investigate and compare the potential diagnostic value of one-step and two-step RT-qPCR techniques to identify SARS-CoV-2, the ROC curve analysis was performed using MedCalc—version 14.8.1 (Fig. 4). The detection power of two-step method for N gene was revealed an area under the curve (AUC) of 0.79 (95% confidence interval (95% CI) 0.66 to 0.89; P = 0.048) with a sensitivity of 88% and specificity of 84% (Fig. 4A). Furthermore, the AUC of 0.90 (95% CI 0.79 to 0.97, P = 0.047) with a sensitivity 88% and specificity 88% were obtained for the S gene (Fig. 4B). The one-step manner showed a sensitivity of 92%, specificity of 80%, and the AUC of 0.95 (95% confidence interval (95% CI) 0.86 to 0.99, P = 0.023) for RdRp gene (Fig. 4C). Moreover, the AUC of 0.95 (95% confidence interval (95% CI) 0.86 to 0.99, P = 0.035) was calculated for N gene detection with a sensitivity of 92% and specificity 96% (Fig. 4D). The combination of the S gene and the N gene in two-step got the AUC 0.85 (95% CI 0.72 to 0.93) with 88% (68.8 to 97.5) sensitivity and 86% specificity (66.35 to 96.5) (P = 0.046) (Fig. 4E). Also, the combination of the RdRp gene and the N gene in one-step got the AUC 0.95 (95% CI 0.86 to 0.99) with 92% (74.45 to 99.8) sensitivity and 88% specificity (69.04 to 96.55) (P = 0.044) (Fig. 4F and Table 5).
Table 5
ROC curve analysis of two step RT-qPCR a and one-step RT-qPCR Ct
Target
Method
AUC (95% CI)
Sensitivity (%) (95% CI)
Specificity (%) (95% CI)
Cut-off value
P-value
S gene
Two step RT-qPCR
0.90 (0.79–0.97)
88 (68.8–97.5)
88 (68.8–97.5)
≤ 40
0.0471
N gene
Two step RT-qPCR
0.79 (0.66–0.89)
88 (68.8–97.5)
84 (63.9–95.5)
≤ 40
0.0489
RdRp gene
One step RT-qPCR
0.95 (0.86–0.99)
92 (74.0–99)
80 (59.3–93.2)
≤ 40
0.0234
N gene
One step RT-qPCR
0.95 (0.86–0.99)
92 (74.0–99)
96 (79.6–99.9)
≤ 40
0.0356
S and N gene
Two step RT-qPCR
0.85 (0.72–0.93)
88 (68.8–97.5)
86 (66.3–96.5)
≤ 40
0.0461
RdRp and N gene
One step RT-qPCR
0.95 (0.86–0.99)
92 (74.45–99.8)
88 (69.4–96.55)
≤ 40
0.0441

Discussion

SARS-CoV-2 led to a global public health emergency state with the rapid spread and elevated morbidity and mortality rates [5]. Given the widespread epidemic of the SARS-CoV-2 virus and the possibility of a recurrence in the future, it is essential to provide an accurate, optimal, and low-cost diagnostic test for SARS-CoV-2 to prevent and control recurrence. The qPCR is a sensitive and specific method that does not need post-PCR steps. This assay is usually employed in medical diagnostic laboratories [20]. TaqMan probe-based RT-qPCR method is used to detect SARS-CoV-2 as a standard WHO protocol [3]. The Charity and Centers for Disease Control and Prevention (CDC) protocols use four different probes to determine a SARS-CoV-2 infection. These protocols contain reagents and probes that are infrequent and expensive [3]. The TaqMan probe-based RT-qPCR method is hard to access in the laboratories of developing countries [5]. In addition, frequent false results enhance the cost of the test as the sample measurement must be repeated 3-times [21].
The SYBR Green technique has a simple design and is a low-cost method but that is not specific. This method is mainly employed for the diagnosis and amplification of genomes. Generally, using specific primers leads to obtaining specific results [22]. Therefore, applying efficient primers in an SYBR Green-based method can turn into a low-cost alternative with high specificity and sensitivity [5]. The current study has evaluated the sensitivity and specificity of one- and two-step RT-qPCR for SARS-CoV-2 detection. PCR signals can be considerably impaired by the non-specific binding of primer [3]. Thus, we first assessed the specificity of the primers by the BLAST option on the NCBI website. In addition, we tested the positive clinical samples, including OC43, 229E, NL63, and Influenza a Virus (H1N1); human rhinovirus, a human respiratory syncytial virus with a two-step method in which it did not show any cross-reactivity.
In contrast to previous studies, our observations indicate that the SYBR green-based method is not precise for detecting SARS-CoV-2. Also, the sensitivity of the one-step method was higher than that two-step assay [22]. The results obtained from the diluted pool samples by the two-step RT-qPCR method detected amplification of the S and N genes up to 10−6. On the other hand, the one-step method detected all diluted pool samples’ concentrations. These results showed 100-fold more sensitivity for the one-step manner. It’s can be because of using the probes in this method [20]. More RNA or cDNA loadings can be added to the reaction mix to enhance the test’s sensitivity and improve the lower viral loads detection [5]. The mutations in target genes can also reduce the sensitivity of diagnostic tests and increase the false-negative results. Thus, monitoring the SARS-CoV-2 sequence for mutations that may affect the PCR assay is recommended [23]. The obtained results on positive clinical samples by the tow-step method showed a specific peak in the melting curve, revealing no non-specific products. Based on previous research, non-specific products during SYBR Green-based qPCR are considered false positives; accordingly, the melting curve should be analyzed [15]. In line with the previous investigation, our results showed that among assessed SARS-CoV-2 genes, the RdRp gene had the lowest specificity between the two-step and one-step techniques. This may be the conservation of the RdRp gene region among beta coronaviruses [24]. An occasional two-step analysis of negative samples may show unspecific sigmoidal fluorescence curves [14]. However, the stabile primers to targeted genes can significantly reduce such errors. Therefore, we propose that when RT-qPCR signal analysis is in linear mode, the negative result obtained from both primers can confirm the absence of the virus.

Conclusion

The SARS-CoV-2 samples were recruited to evaluate the advantages and limitations of one-step and two-step qPCR methods. Our results proposed that the two-step method is practical for diagnosing SARS-CoV-2. We have set up two-step protocols that can be done using reverse transcriptase (MMLV-RT) and a variety of qPCR mixes. Combining this and easy RNA preparation methods would significantly improve the SARS-CoV-2 detection test capacities. Generally, the sensitivity and reliability of the two-step qPCR method are almost similar to the one-step qPCR manner. Besides decreasing expenses in medical diagnostic laboratories, this method will ease viral detection.
The obtained data from ROC curve analysis for every two manners of RT-qPCR showed specificity lower than 90%; however, we did not observe a cross-attachment of SARS-CoV-2 primers with other tested viruses. It’s maybe resulting from the reaction situation or the small sample size. More studies are needed to compare the one-step TaqMan probe-based method and the two-step SYBR Green-based method.

Acknowledgements

The authors gratefully acknowledge Department of Medical Virology, Faculty of Medicine, Lorestan University of Medical Sciences, Khorramabad, Iran; and Hepatitis Research Center, Lorestan University of Medical Sciences, Khorramabad, Iran.

Declarations

Our study was performed between June 2019 and November 2021, and the investigation was approved by the Committee of Clinical Research Ethical in Lorestan University of Medical Sciences, Khorramabad, Iran (Ethical code: IR.LUMS.REC.1400.115). Written informed consent was obtained from all participants in the study. All methods in this study were carried out in accordance with relevant guidelines and regulations the institutional research committee and with the 1964 Helsinki Declaration.
Not applicable.

Competing interests

The authors declare no conflict of interests.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Chang F-Y, Chen H-C, Chen P-J, Ho M-S, Hsieh S-L, Lin J-C, et al. Immunologic aspects of characteristics, diagnosis, and treatment of coronavirus disease 2019 (COVID-19). J Biomed Sci. 2020;27(1):72.CrossRef Chang F-Y, Chen H-C, Chen P-J, Ho M-S, Hsieh S-L, Lin J-C, et al. Immunologic aspects of characteristics, diagnosis, and treatment of coronavirus disease 2019 (COVID-19). J Biomed Sci. 2020;27(1):72.CrossRef
2.
Zurück zum Zitat Liang Y, Wang M-L, Chien C-S, Yarmishyn AA, Yang Y-P, Lai W-Y, et al. Highlight of immune pathogenic response and hematopathologic effect in SARS-CoV, MERS-CoV, and SARS-Cov-2 infection. Front Immunol. 2020;11:1022.CrossRef Liang Y, Wang M-L, Chien C-S, Yarmishyn AA, Yang Y-P, Lai W-Y, et al. Highlight of immune pathogenic response and hematopathologic effect in SARS-CoV, MERS-CoV, and SARS-Cov-2 infection. Front Immunol. 2020;11:1022.CrossRef
3.
Zurück zum Zitat Marinowic DR, Zanirati G, Rodrigues FVF, Grahl MVC, Alcará AM, Machado DC, et al. A new SYBR Green real-time PCR to detect SARS-CoV-2. Sci Rep. 2021;11(1):2224.CrossRef Marinowic DR, Zanirati G, Rodrigues FVF, Grahl MVC, Alcará AM, Machado DC, et al. A new SYBR Green real-time PCR to detect SARS-CoV-2. Sci Rep. 2021;11(1):2224.CrossRef
4.
Zurück zum Zitat Covid-19 National Emergency Response Center E, Case Management Team KCfDC, Prevention. Early epidemiological and clinical characteristics of 28 cases of coronavirus disease in South Korea. Osong Public Health Res Perspect. 2020;11(1):8–14.CrossRef Covid-19 National Emergency Response Center E, Case Management Team KCfDC, Prevention. Early epidemiological and clinical characteristics of 28 cases of coronavirus disease in South Korea. Osong Public Health Res Perspect. 2020;11(1):8–14.CrossRef
5.
Zurück zum Zitat Dorlass EG, Monteiro CO, Viana AO, Soares CP, Machado RRG, Thomazelli LM, et al. Lower cost alternatives for molecular diagnosis of COVID-19: conventional RT-PCR and SYBR Green-based RT-qPCR. Braz J Microbiol. 2020;51(3):1117–23.CrossRef Dorlass EG, Monteiro CO, Viana AO, Soares CP, Machado RRG, Thomazelli LM, et al. Lower cost alternatives for molecular diagnosis of COVID-19: conventional RT-PCR and SYBR Green-based RT-qPCR. Braz J Microbiol. 2020;51(3):1117–23.CrossRef
6.
Zurück zum Zitat Freire-Paspuel B, Vega-Mariño P, Velez A, Cruz M, Garcia-Bereguiain MA. Sample pooling of RNA extracts to speed up SARS-CoV-2 diagnosis using CDC FDA EUA RT-qPCR kit. Virus Res. 2020;290:198173.CrossRef Freire-Paspuel B, Vega-Mariño P, Velez A, Cruz M, Garcia-Bereguiain MA. Sample pooling of RNA extracts to speed up SARS-CoV-2 diagnosis using CDC FDA EUA RT-qPCR kit. Virus Res. 2020;290:198173.CrossRef
7.
Zurück zum Zitat Figueroa S, Freire-Paspuel B, Vega-Mariño P, Velez A, Cruz M, Cardenas WB, et al. High sensitivity-low cost detection of SARS-CoV-2 by two steps end point RT-PCR with agarose gel electrophoresis visualization. Sci Rep. 2021;11(1):21658.CrossRef Figueroa S, Freire-Paspuel B, Vega-Mariño P, Velez A, Cruz M, Cardenas WB, et al. High sensitivity-low cost detection of SARS-CoV-2 by two steps end point RT-PCR with agarose gel electrophoresis visualization. Sci Rep. 2021;11(1):21658.CrossRef
8.
Zurück zum Zitat Jiang M, Pan W, Arasthfer A, Fang W, Ling L, Fang H, et al. Development and validation of a rapid, single-step reverse transcriptase loop-mediated isothermal amplification (RT-LAMP) system potentially to be used for reliable and high-throughput screening of COVID-19. Front Cell Infect Microbiol. 2020;10:331.CrossRef Jiang M, Pan W, Arasthfer A, Fang W, Ling L, Fang H, et al. Development and validation of a rapid, single-step reverse transcriptase loop-mediated isothermal amplification (RT-LAMP) system potentially to be used for reliable and high-throughput screening of COVID-19. Front Cell Infect Microbiol. 2020;10:331.CrossRef
9.
Zurück zum Zitat Zhu N, Zhang D, Wang W, Li X, Yang B, Song J, et al. A novel coronavirus from patients with pneumonia in China, 2019. N Engl J Med. 2020;382(8):727–33.CrossRef Zhu N, Zhang D, Wang W, Li X, Yang B, Song J, et al. A novel coronavirus from patients with pneumonia in China, 2019. N Engl J Med. 2020;382(8):727–33.CrossRef
10.
Zurück zum Zitat Kudo E, Israelow B, Vogels CBF, Lu P, Wyllie AL, Tokuyama M, et al. Detection of SARS-CoV-2 RNA by multiplex RT-qPCR. PLoS Biol. 2020;18(10):e3000867.CrossRef Kudo E, Israelow B, Vogels CBF, Lu P, Wyllie AL, Tokuyama M, et al. Detection of SARS-CoV-2 RNA by multiplex RT-qPCR. PLoS Biol. 2020;18(10):e3000867.CrossRef
11.
Zurück zum Zitat Suo T, Liu X, Feng J, Guo M, Hu W, Guo D, et al. ddPCR: a more accurate tool for SARS-CoV-2 detection in low viral load specimens. Emerg Microbes Infect. 2020;9(1):1259–68.CrossRef Suo T, Liu X, Feng J, Guo M, Hu W, Guo D, et al. ddPCR: a more accurate tool for SARS-CoV-2 detection in low viral load specimens. Emerg Microbes Infect. 2020;9(1):1259–68.CrossRef
12.
Zurück zum Zitat Lucia C, Federico P-B, Alejandra GC. An ultrasensitive, rapid, and portable coronavirus SARS-CoV-2 sequence detection method based on CRISPR-Cas12. bioRxiv. 2020:2020.02.29.971127. Lucia C, Federico P-B, Alejandra GC. An ultrasensitive, rapid, and portable coronavirus SARS-CoV-2 sequence detection method based on CRISPR-Cas12. bioRxiv. 2020:2020.02.29.971127.
13.
Zurück zum Zitat Espy MJ, Uhl JR, Sloan LM, Buckwalter SP, Jones MF, Vetter EA, et al. Real-time PCR in clinical microbiology: applications for routine laboratory testing. Clin Microbiol Rev. 2006;19(1):165–256.CrossRef Espy MJ, Uhl JR, Sloan LM, Buckwalter SP, Jones MF, Vetter EA, et al. Real-time PCR in clinical microbiology: applications for routine laboratory testing. Clin Microbiol Rev. 2006;19(1):165–256.CrossRef
14.
Zurück zum Zitat Fuchs Wightman F, Godoy Herz MA, Muñoz JC, Stigliano JN, Bragado L, Moreno NN, et al. A DNA intercalating dye-based RT-qPCR alternative to diagnose SARS-CoV-2. RNA Biol. 2021;18(12):2218–25.CrossRef Fuchs Wightman F, Godoy Herz MA, Muñoz JC, Stigliano JN, Bragado L, Moreno NN, et al. A DNA intercalating dye-based RT-qPCR alternative to diagnose SARS-CoV-2. RNA Biol. 2021;18(12):2218–25.CrossRef
15.
Zurück zum Zitat Pereira-Gómez M, Fajardo A, Echeverría N, López-Tort F, Perbolianachis P, Costábile A, et al. Evaluation of SYBR Green real time PCR for detecting SARS-CoV-2 from clinical samples. J Virol Methods. 2020;289:114035.CrossRef Pereira-Gómez M, Fajardo A, Echeverría N, López-Tort F, Perbolianachis P, Costábile A, et al. Evaluation of SYBR Green real time PCR for detecting SARS-CoV-2 from clinical samples. J Virol Methods. 2020;289:114035.CrossRef
16.
Zurück zum Zitat Gomes-Ruiz AC, Nascimento RT, de Paula SO, da Fonseca BAL. SYBR green and TaqMan real-time PCR assays are equivalent for the diagnosis of dengue virus type 3 infections. J Med Virol. 2006;78(6):760–3.CrossRef Gomes-Ruiz AC, Nascimento RT, de Paula SO, da Fonseca BAL. SYBR green and TaqMan real-time PCR assays are equivalent for the diagnosis of dengue virus type 3 infections. J Med Virol. 2006;78(6):760–3.CrossRef
17.
Zurück zum Zitat Kumar B, Kumar P, Rajput R, Daga MK, Singh V, Khanna M. Comparative reproducibility of SYBR Green I and TaqManreal-time PCR chemistries for the analysis of matrix andhemagglutinin genes of Influenza A viruses. Int J Collab Res Intern Med Public Health. 2012;4:1346–52. Kumar B, Kumar P, Rajput R, Daga MK, Singh V, Khanna M. Comparative reproducibility of SYBR Green I and TaqManreal-time PCR chemistries for the analysis of matrix andhemagglutinin genes of Influenza A viruses. Int J Collab Res Intern Med Public Health. 2012;4:1346–52.
18.
Zurück zum Zitat Watzinger F, Ebner K, Lion T. Detection and monitoring of virus infections by real-time PCR. Mol Aspects Med. 2006;27(2–3):254–98.CrossRef Watzinger F, Ebner K, Lion T. Detection and monitoring of virus infections by real-time PCR. Mol Aspects Med. 2006;27(2–3):254–98.CrossRef
19.
Zurück zum Zitat Candel FJ, Barreiro P, San Román J, Abanades JC, Barba R, Barberán J, et al. Recommendations for use of antigenic tests in the diagnosis of acute SARS-CoV-2 infection in the second pandemic wave: attitude in different clinical settings. Rev Esp Quimioter. 2020;33(6):466–84.CrossRef Candel FJ, Barreiro P, San Román J, Abanades JC, Barba R, Barberán J, et al. Recommendations for use of antigenic tests in the diagnosis of acute SARS-CoV-2 infection in the second pandemic wave: attitude in different clinical settings. Rev Esp Quimioter. 2020;33(6):466–84.CrossRef
20.
Zurück zum Zitat Pereira-Gómez M, Fajardo Á, Echeverría N, López-Tort F, Perbolianachis P, Costábile A, et al. Evaluation of SYBR Green real time PCR for detecting SARS-CoV-2 from clinical samples. J Virol Methods. 2021;289:114035.CrossRef Pereira-Gómez M, Fajardo Á, Echeverría N, López-Tort F, Perbolianachis P, Costábile A, et al. Evaluation of SYBR Green real time PCR for detecting SARS-CoV-2 from clinical samples. J Virol Methods. 2021;289:114035.CrossRef
21.
Zurück zum Zitat Zhou X, Zhang T, Song D, Huang T, Peng Q, Chen Y, et al. Comparison and evaluation of conventional RT-PCR, SYBR green I and TaqMan real-time RT-PCR assays for the detection of porcine epidemic diarrhea virus. Mol Cell Probes. 2017;33:36–41.CrossRef Zhou X, Zhang T, Song D, Huang T, Peng Q, Chen Y, et al. Comparison and evaluation of conventional RT-PCR, SYBR green I and TaqMan real-time RT-PCR assays for the detection of porcine epidemic diarrhea virus. Mol Cell Probes. 2017;33:36–41.CrossRef
22.
Zurück zum Zitat Tajadini M, Panjehpour M, Javanmard SH. Comparison of SYBR Green and TaqMan methods in quantitative real-time polymerase chain reaction analysis of four adenosine receptor subtypes. Adv Biomed Res. 2014;3:85.CrossRef Tajadini M, Panjehpour M, Javanmard SH. Comparison of SYBR Green and TaqMan methods in quantitative real-time polymerase chain reaction analysis of four adenosine receptor subtypes. Adv Biomed Res. 2014;3:85.CrossRef
23.
Zurück zum Zitat Kenyeres B, Ánosi N, Bányai K, Mátyus M, Orosz L, Kiss A, et al. Comparison of four PCR and two point of care assays used in the laboratory detection of SARS-CoV-2. J Virol Methods. 2021;293:114165.CrossRef Kenyeres B, Ánosi N, Bányai K, Mátyus M, Orosz L, Kiss A, et al. Comparison of four PCR and two point of care assays used in the laboratory detection of SARS-CoV-2. J Virol Methods. 2021;293:114165.CrossRef
24.
Zurück zum Zitat Chan Jasper F-W, Yip Cyril C-Y, To Kelvin K-W, Tang Tommy H-C, Wong Sally C-Y, Leung K-H, et al. Improved molecular diagnosis of COVID-19 by the novel, highly sensitive and specific COVID-19-RdRp/Hel real-time reverse transcription-PCR assay validated in vitro and with clinical specimens. J Clin Microbiol. 2020;58(5):e00310-20.CrossRef Chan Jasper F-W, Yip Cyril C-Y, To Kelvin K-W, Tang Tommy H-C, Wong Sally C-Y, Leung K-H, et al. Improved molecular diagnosis of COVID-19 by the novel, highly sensitive and specific COVID-19-RdRp/Hel real-time reverse transcription-PCR assay validated in vitro and with clinical specimens. J Clin Microbiol. 2020;58(5):e00310-20.CrossRef
Metadaten
Titel
Diagnostic power of one-step and two-step RT-qPCR methods to SARS‑CoV‑2 detection
verfasst von
Asra Malekshahi
Sayyad Khanizadeh
Shirzad Fallahi
Gholamreza Talei
Mehdi Birjandi
Faezeh Hajizadeh
Publikationsdatum
01.12.2022
Verlag
BioMed Central
Schlagwort
COVID-19
Erschienen in
BMC Infectious Diseases / Ausgabe 1/2022
Elektronische ISSN: 1471-2334
DOI
https://doi.org/10.1186/s12879-022-07478-0

Weitere Artikel der Ausgabe 1/2022

BMC Infectious Diseases 1/2022 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.