Erschienen in:
28.05.2020 | COVID-19 | Brief Report
Zur Zeit gratis
The Allplex 2019-nCoV (Seegene) assay: which performances are for SARS-CoV-2 infection diagnosis?
verfasst von:
Eric Farfour, Philippe Lesprit, Benoit Visseaux, Tiffany Pascreau, Emilie Jolly, Nadira Houhou, Laurence Mazaux, Marianne Asso-Bonnet, Marc Vasse, on behalf of the SARS-CoV-2 Foch Hospital study group
Erschienen in:
European Journal of Clinical Microbiology & Infectious Diseases
|
Ausgabe 10/2020
Einloggen, um Zugang zu erhalten
Abstract
Several commercial assays for SARS-CoV-2 RT-PCR are available but few of them were assessed. We evaluate the Allplex 2019-nCoV (Seegene) assay using 41 nasopharyngeal samples. The rates of agreement were 92.7% and 100% with the GeneFinder COVID-19 plus (Elitech) and the diagnosis of the infectious disease specialist respectively. Four samples display a Ct < 22.0 for the E and RdRp genes while the N gene was not detected, suggesting a variability of the viral sequence. There was no cross-reactivity with other respiratory viruses. The Allplex 2019-nCoV appears as a reliable method, but additional evaluations using more samples are needed. RT-PCR assays should probably include at least 2 viral targets.