Cryoneedle artifacts are frequently observed in MRI-guided cryoablations, and may obscure visualization of critical anatomy and compromise needle placement accuracy. This work experimentally investigated the contributing factors of these artifacts to identify effective mitigation strategies.
Materials and Methods
Ex vivo porcine tissue with inserted cryoneedles was imaged on a 1.5-Tesla MRI. Fast spin echo (FSE) and spoiled gradient echo (GRE) sequences with echo times from 1.04 to 60 ms and specific absorption rates (SARs) from 0.01 to 2.1 W/kg were used. During MRI, cryoneedle temperatures were monitored using fiber-optic sensors. Configurations with one to three cryoneedles oriented at 0-degree or 45-degree angles to the patient table were investigated. The body coil was used for transmit/receive, both with and without an additional receive-only surface loop coil. Artifact width and intensity were measured for analysis.
Results
Cryoneedle artifact widths were unrelated to echo time for both FSE (p = 0.6) and GRE (p = 0.3) and were smaller in GRE than in FSE images (p << 0.05). Artifact widths correlated with cryoneedle temperature elevations (r2 = 0.969, p << 0.05) but were not correlated with SAR (GRE: p = 0.3; FSE: p = 0.5). The artifact intensity with the cryoneedle oriented at 0 degrees increased with a greater number of cryoneedles in the tissue (p = 0.006), and when the surface loop coil was used (p = 0.008).
Conclusion
Clinically observed cryoneedle artifacts compromising treatment efficacy can be indicative of tissue radiofrequency heating risk, and effectively mitigated by either using GRE-based sequences or adjusting coil/cryoneedle configurations.
Graphical Abstract
×
Anzeige
Bitte loggen Sie sich ein, um Zugang zu diesem Inhalt zu erhalten
Zwei US-Onkologen argumentieren im renommierten New England Journal of Medicine, dass bestimmte Formen der Nachsorge Krebsüberlebenden mehr schaden als nützen. Aber wie kommen sie darauf?
Betreuer müssen auch die Patientenverfügung psychisch erkrankter Menschen berücksichtigen, wenn sie sich darin gegen eine Zwangsbehandlung aussprechen, betont der Bundesgerichtshof.
Wird der Kontrastmittelverbrauch im Zuge einer perkutanen Koronarintervention gesenkt, reduziert sich das Risiko der Patienten für einen akuten Nierenschaden. Ob sich das in der Praxis durchsetzt, haben US-Mediziner erforscht.
Die Computertomografie im Niedrigdosisbereich gilt als probates Instrument im Screening auf Lungenkrebs. Kandidaten dafür sind in erster Linie (Ex-)Raucher. Fündig wird die Low-Dose-CT aber auch in anderen Fällen, wie eine Studie ergeben hat.