01.02.2012 | Computed Tomography

CT image quality improvement using adaptive iterative dose reduction with wide-volume acquisition on 320-detector CT

verfasst von: Alban Gervaise, Benoît Osemont, Sophie Lecocq, Alain Noel, Emilien Micard, Jacques Felblinger, Alain Blum

Erschienen in: European Radiology | Ausgabe 2/2012

Einloggen, um Zugang zu erhalten

Abstract

Objectives

To evaluate the impact of Adaptive Iterative Dose Reduction (AIDR) on image quality and radiation dose in phantom and patient studies.

Methods

A phantom was examined in volumetric mode on a 320-detector CT at different tube currents from 25 to 550 mAs. CT images were reconstructed with AIDR and with Filtered Back Projection (FBP) reconstruction algorithm. Image noise, Contrast-to-Noise Ratio (CNR), Signal-to-Noise Ratio (SNR) and spatial resolution were compared between FBP and AIDR images. AIDR was then tested on 15 CT examinations of the lumbar spine in a prospective study. Again, FBP and AIDR images were compared. Image noise and SNR were analysed using a Wilcoxon signed-rank test.

Results

In the phantom, spatial resolution assessment showed no significant difference between FBP and AIDR reconstructions. Image noise was lower with AIDR than with FBP images with a mean reduction of 40%. CNR and SNR were also improved with AIDR. In patients, quantitative and subjective evaluation showed that image noise was significantly lower with AIDR than with FBP. SNR was also greater with AIDR than with FBP.

Conclusion

Compared to traditional FBP reconstruction techniques, AIDR significantly improves image quality and has the potential to decrease radiation dose.

Key Points

  • This study showed that Adaptive Iterative Dose Reduction (AIDR) reduces image noise.
  • In a phantom image noise was reduced without altering spatial resolution.
  • In patients AIDR reduced the image noise in lumbar spine CT.
  • AIDR can potentially reduce the dose for lumbar spine CT by 52%.
Literatur
1.
Zurück zum Zitat 2007 CT market summary report (2007) Des Plaines, IL: IMV Medical Information Division 2007 CT market summary report (2007) Des Plaines, IL: IMV Medical Information Division
2.
Zurück zum Zitat Brenner DJ, Hall EJ (2007) Computed tomography an increasing source of radiation exposure. N Engl J Med 357:2277–2284PubMedCrossRef Brenner DJ, Hall EJ (2007) Computed tomography an increasing source of radiation exposure. N Engl J Med 357:2277–2284PubMedCrossRef
3.
Zurück zum Zitat Sodickson A, Baeyens PF, Andriole KP et al (2009) Recurrent CT, cumulative radiation exposure and associated radiation-induced cancer risks from CT of adults. Radiology 252:175–184CrossRef Sodickson A, Baeyens PF, Andriole KP et al (2009) Recurrent CT, cumulative radiation exposure and associated radiation-induced cancer risks from CT of adults. Radiology 252:175–184CrossRef
4.
Zurück zum Zitat Smith-Bindman R, Lipson J, Marcus R et al (2009) Radiation dose associated with common computed tomography examinations and the associated lifetime attributable risk of cancer. Arch Intern Med 169:2078–2086PubMedCrossRef Smith-Bindman R, Lipson J, Marcus R et al (2009) Radiation dose associated with common computed tomography examinations and the associated lifetime attributable risk of cancer. Arch Intern Med 169:2078–2086PubMedCrossRef
5.
Zurück zum Zitat Berrington de González A, Mahesh M, Kim KP et al (2009) Projected cancer risks from computed tomography scans performed in the United States in 2007. Arch Intern Med 169:2071–2077PubMedCrossRef Berrington de González A, Mahesh M, Kim KP et al (2009) Projected cancer risks from computed tomography scans performed in the United States in 2007. Arch Intern Med 169:2071–2077PubMedCrossRef
6.
Zurück zum Zitat Mulkens TH, Bellinck P, Baeyaert M et al (2005) Use of an automatic exposure control mechanism for dose optimization in multi-detector row CT examinations: clinical evaluation. Radiology 237:213–223PubMedCrossRef Mulkens TH, Bellinck P, Baeyaert M et al (2005) Use of an automatic exposure control mechanism for dose optimization in multi-detector row CT examinations: clinical evaluation. Radiology 237:213–223PubMedCrossRef
7.
Zurück zum Zitat Kalender WA, Buchenau S, Deak P et al (2008) Technical approaches to the optimisation of CT. Phys Med 24:71–79PubMedCrossRef Kalender WA, Buchenau S, Deak P et al (2008) Technical approaches to the optimisation of CT. Phys Med 24:71–79PubMedCrossRef
8.
Zurück zum Zitat Alkadhi H, Leschka S (2011) Radiation dose of cardiac computed tomography—what has been achieved and what needs to be done. Eur Radiol 21:505–509PubMedCrossRef Alkadhi H, Leschka S (2011) Radiation dose of cardiac computed tomography—what has been achieved and what needs to be done. Eur Radiol 21:505–509PubMedCrossRef
9.
Zurück zum Zitat Christner JA, Zavaletta VA, Eusemann CD, Walz-Flannigan AI, McCollough CH (2010) Dose reduction in helical CT: dynamically adjustable z-axis X-ray beam collimation. AJR Am J Roentgenol 194:W49–W55PubMedCrossRef Christner JA, Zavaletta VA, Eusemann CD, Walz-Flannigan AI, McCollough CH (2010) Dose reduction in helical CT: dynamically adjustable z-axis X-ray beam collimation. AJR Am J Roentgenol 194:W49–W55PubMedCrossRef
10.
Zurück zum Zitat Gervaise A, Louis M, Batch T et al (2010) Dose reduction at CT of the lumbar spine using a 320-detector scanner: initial results. J Radiol 91:779–785PubMedCrossRef Gervaise A, Louis M, Batch T et al (2010) Dose reduction at CT of the lumbar spine using a 320-detector scanner: initial results. J Radiol 91:779–785PubMedCrossRef
11.
Zurück zum Zitat Xu J, Mahesh M, Tsui BM (2009) Is iterative reconstruction ready for MDCT? J Am Coll Radiol 6:274–276PubMedCrossRef Xu J, Mahesh M, Tsui BM (2009) Is iterative reconstruction ready for MDCT? J Am Coll Radiol 6:274–276PubMedCrossRef
12.
Zurück zum Zitat Hara AK, Paden RG, Silva AC, Kujak JL, Lawder HJ, Pavlicek W (2009) Iterative reconstruction technique for reducing body radiation dose at CT: feasibility study. AJR Am J Roentgenol 193:764–771PubMedCrossRef Hara AK, Paden RG, Silva AC, Kujak JL, Lawder HJ, Pavlicek W (2009) Iterative reconstruction technique for reducing body radiation dose at CT: feasibility study. AJR Am J Roentgenol 193:764–771PubMedCrossRef
13.
Zurück zum Zitat Thibault JB, Sauer KD, Bouman CA, Hsieh J (2007) A three-dimensional statistical approach to improved image quality for multislice helical CT. Med Phys 34:4526–4544PubMedCrossRef Thibault JB, Sauer KD, Bouman CA, Hsieh J (2007) A three-dimensional statistical approach to improved image quality for multislice helical CT. Med Phys 34:4526–4544PubMedCrossRef
14.
Zurück zum Zitat Gupta AK, Nelson RC, Johnson GA, Paulson EK, Delong DM, Yoshizumi TT (2003) Optimization of eight-element multi-detector row helical CT technology for evaluation of the abdomen. Radiology 227:739–745PubMedCrossRef Gupta AK, Nelson RC, Johnson GA, Paulson EK, Delong DM, Yoshizumi TT (2003) Optimization of eight-element multi-detector row helical CT technology for evaluation of the abdomen. Radiology 227:739–745PubMedCrossRef
15.
Zurück zum Zitat Pontana F, Duhamel A, Pagniez J et al (2011) Chest computed tomography using iterative reconstruction vs. filtered back projection (part 2): image quality of low-dose CT examinations in 80 patients. Eur Radiol 21:636–643PubMedCrossRef Pontana F, Duhamel A, Pagniez J et al (2011) Chest computed tomography using iterative reconstruction vs. filtered back projection (part 2): image quality of low-dose CT examinations in 80 patients. Eur Radiol 21:636–643PubMedCrossRef
16.
Zurück zum Zitat Boedeker KL, McNitt-Gray MF (2007) Application of the noise power spectrum in modern diagnostic MDCT: part II. Noise power spectra and signal to noise. Phys Med Biol 52:4047–4061PubMedCrossRef Boedeker KL, McNitt-Gray MF (2007) Application of the noise power spectrum in modern diagnostic MDCT: part II. Noise power spectra and signal to noise. Phys Med Biol 52:4047–4061PubMedCrossRef
17.
Zurück zum Zitat von Falck C, Galanski M, Shin HO (2010) Informatics in radiology: sliding-thin-slab averaging for improved depiction of low-contrast lesions with radiation dose savings at thin-section CT. Radiographics 30:317–326CrossRef von Falck C, Galanski M, Shin HO (2010) Informatics in radiology: sliding-thin-slab averaging for improved depiction of low-contrast lesions with radiation dose savings at thin-section CT. Radiographics 30:317–326CrossRef
18.
Zurück zum Zitat Prakash P, Kalra MK, Ackman JB et al (2010) Diffuse lung disease: CT of the chest with adaptive statistical iterative reconstruction technique. Radiology 256:261–269PubMedCrossRef Prakash P, Kalra MK, Ackman JB et al (2010) Diffuse lung disease: CT of the chest with adaptive statistical iterative reconstruction technique. Radiology 256:261–269PubMedCrossRef
19.
Zurück zum Zitat Ghetti C, Ortenzia O, Serreli G (2011) CT iterative reconstruction in image space: a phantom study. Phys Med [Epub ahead of print] Ghetti C, Ortenzia O, Serreli G (2011) CT iterative reconstruction in image space: a phantom study. Phys Med [Epub ahead of print]
20.
Zurück zum Zitat Spangfort EV (1972) The lumbar disc herniation: a computer-aided analysis of 2,504 operations. Acta Orthop Scand Suppl 142:1–95PubMed Spangfort EV (1972) The lumbar disc herniation: a computer-aided analysis of 2,504 operations. Acta Orthop Scand Suppl 142:1–95PubMed
21.
Zurück zum Zitat Silva A, Lawder H, Hara A, Kujak J, Pavlicek W (2010) Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm. AJR Am J Roentgenol 194:191–199PubMedCrossRef Silva A, Lawder H, Hara A, Kujak J, Pavlicek W (2010) Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm. AJR Am J Roentgenol 194:191–199PubMedCrossRef
22.
Zurück zum Zitat Pontana F, Pagniez J, Flohr T et al (2011) Chest computed tomography using iterative reconstruction vs. filtered back projection (part 1): evaluation of image noise reduction in 32 patients. Eur Radiol 21:627–635PubMedCrossRef Pontana F, Pagniez J, Flohr T et al (2011) Chest computed tomography using iterative reconstruction vs. filtered back projection (part 1): evaluation of image noise reduction in 32 patients. Eur Radiol 21:627–635PubMedCrossRef
23.
Zurück zum Zitat Mitsumori LM, Shuman WP, Busey JM, Kolokythas O, Koprowicz KM (2011) Adaptive statistical iterative reconstruction versus filtered back projection in the same patient: 64 channel liver CT image quality and patient radiation dose. Eur Radiol [Epub ahead of print] Mitsumori LM, Shuman WP, Busey JM, Kolokythas O, Koprowicz KM (2011) Adaptive statistical iterative reconstruction versus filtered back projection in the same patient: 64 channel liver CT image quality and patient radiation dose. Eur Radiol [Epub ahead of print]
24.
Zurück zum Zitat Fazel R, Krumholz HM, Wang Y et al (2009) Exposure to low-dose ionizing radiation from medical imaging procedures. N Engl J Med 361:849–857PubMedCrossRef Fazel R, Krumholz HM, Wang Y et al (2009) Exposure to low-dose ionizing radiation from medical imaging procedures. N Engl J Med 361:849–857PubMedCrossRef
25.
Zurück zum Zitat Tins B (2010) Technical aspects of CT imaging of the spine. Insights Imaging 1:349–359CrossRef Tins B (2010) Technical aspects of CT imaging of the spine. Insights Imaging 1:349–359CrossRef
26.
Zurück zum Zitat Bohy P, de Maertelaer V, Roquigny A, Keyzer C, Tack D, Genevois PA (2007) Multidetector CT in patients suspected of having lumbar disk herniation: comparison of standard-dose and simulated low-dose techniques. Radiology 244:231–524CrossRef Bohy P, de Maertelaer V, Roquigny A, Keyzer C, Tack D, Genevois PA (2007) Multidetector CT in patients suspected of having lumbar disk herniation: comparison of standard-dose and simulated low-dose techniques. Radiology 244:231–524CrossRef
Metadaten
Titel
CT image quality improvement using adaptive iterative dose reduction with wide-volume acquisition on 320-detector CT
verfasst von
Alban Gervaise
Benoît Osemont
Sophie Lecocq
Alain Noel
Emilien Micard
Jacques Felblinger
Alain Blum
Publikationsdatum
01.02.2012
Verlag
Springer-Verlag
Erschienen in
European Radiology / Ausgabe 2/2012
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-011-2271-7

Weitere Artikel der Ausgabe 2/2012

European Radiology 2/2012 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.