CT radiation dose reduction in patients with total hip arthroplasties using model-based iterative reconstruction and orthopaedic metal artefact reduction

verfasst von: Ruud H. H. Wellenberg, Jochen A. C. van Osch, Henk J. Boelhouwers, Mireille A. Edens, Geert J. Streekstra, Harmen B. Ettema, Martijn F. Boomsma

Erschienen in: Skeletal Radiology | Ausgabe 11/2019

Abstract

Objective

To evaluate the impact of radiation dose reduction on image quality in patients with metal-on-metal total hip arthroplasties (THAs) using model-based iterative reconstruction (MBIR) combined with orthopaedic metal artefact reduction (O-MAR).

Materials and methods

Patients with metal-on-metal THAs received a pelvic CT with a full (FD) and a reduced radiation dose (RD) with −20%, −40%, −57%, or −80% CT radiation dose respectively, when assigned to group 1, 2, 3, or 4 respectively. FD acquisitions were reconstructed with iterative reconstruction, iDose4. RD acquisitions were additionally reconstructed with iterative model-based reconstruction (IMR) levels 1–3 with different levels of noise suppression. CT numbers, noise and contrast-to-noise ratios were measured in muscle, fat and bladder. Subjective image quality was evaluated on seven aspects including artefacts, osseous structures, prosthetic components and soft tissues.

Results

Seventy-six patients were randomly assigned to one of the four groups. While reducing radiation dose by 20%, 40%, 57%, or 80% in combination with IMR, CT numbers remained constant. Compared with iDose4, the noise decreased (p < 0.001) and contrast-to-noise ratios increased (p < 0.001) with IMR. O-MAR improved CT number accuracy in the bladder and reduced noise in the bladder, muscle and fat (p < 0.01). Subjective image quality was rated lower on RD IMR images than FD iDose4 images on all seven aspects (p < 0.05) and was not related to the applied radiation dose reduction.

Conclusion

In RD IMR with O-MAR images, CT numbers remained constant, noise decreased and contrast-to-noise ratios between muscle and fat increased compared with FD iDose4 with O-MAR images in patients with metal-on-metal THAs. Subjective image quality reduced, regardless of the degree of radiation dose reduction.
Hinweise

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

In computed tomography (CT), the use of iterative reconstruction (IR) and model-based iterative reconstruction (MBIR) techniques enables a reduction in the CT radiation dose while improving image quality compared with standard filtered back-projection (FBP) [1].
Iterative reconstruction, or so-called hybrid reconstruction techniques, are blending techniques that combine IR images with FBP [2]. Some commercially available IR techniques are iDose4 (Philips), Safire (Siemens), ASIR (GE) and AIDR 3D (Toshiba). Philips’ MBIR algorithm iterative model-based reconstruction (IMR) is a full iterative reconstruction technique, which improves the reconstruction process by incorporating system models and photon statistics [3]. Other vendors have similar MBIR techniques named ADMIRE (Siemens), Veo (GE) and FIRST (Toshiba). With MBIR, images can be reconstructed at RDs and low noise levels. MBIR, furthermore, reduces the size of metal artefacts and allows an equal or better visibility of the bone–metal interfaces and improves the assessment of soft tissue surrounding implants compared with FBP [4, 5]. The downside of MBIR is the fact that images may appear smooth [6, 7]. More importantly, the possible loss of small details and structures could impede its use in musculoskeletal CT imaging.
In addition, metal hardware impairs the diagnostic value of CT in musculoskeletal imaging. Severe metal artefacts in the case of large-head metal-on-metal total hip arthroplasties (THAs) are mainly caused by extensive photon starvation and affect a reliable diagnosis of prosthesis-related soft-tissue and bone abnormalities. Philips’ orthopaedic metal artefact reduction algorithm, O-MAR, reduces severe metal artefacts in large implants [812]. Similar MAR software techniques by other vendors are iMAR (Siemens), SmartMAR (GE) and SEMAR (Toshiba). O-MAR post-processes the projection data and provides more regular attenuation profiles before image reconstruction. These more regular attenuation profiles can improve the general performance of iDose4 and IMR. A previous THA phantom study showed that the combined use of IMR and O-MAR enabled a radiation dose reduction of 83% where CT number accuracy, signal-to-noise ratios and contrast-to-noise ratios (CNRs) increased and noise decreased compared with the iterative reconstruction technique iDose4 and O-MAR [13]. However, a clinical validation study is essential to determine if the CT radiation dose can be reduced in THA patients by using IMR and O-MAR as well.
Therefore, the aim of this study was to quantitatively and qualitatively assess image quality using IMR with O-MAR while reducing CT radiation dose up to 80% in patients with large-head metal-on-metal total hip prostheses compared with full-dose (FD) iDose4 with O-MAR.

Materials and methods

Patient selection

Patients with unilateral or bilateral metal-on-metal total hip prosthesis who were scheduled for a routine 5- or 10-year follow-up CT scan for the investigation of progression or regression of pseudo tumour formation were prospectively included between October 2016 and September 2017 (Table 1). Institutional Review Board (IRB) approval was received for acquiring an additional RD CT scan in patients older than 60 years, with an explicit written informed consent for participation in this study (NL58001.075.16). Exclusion criteria were previous participation in the study, pregnancy or concomitant participation in a study in which the patient is exposed to X-rays. In total, 76 of the 120 patients who qualified to receive a 5- to 10-year follow-up CT were willing to participate in this study.
Table 1
Patient characteristics including age, sex, unilateral and bilateral prosthesis, location of the metal-on-metal total hip arthroplasty (THA), reduced dose (RD) and full-dose (FD) dose–length product (DLP) and CT dose index (CTDI)
 
Group 1
Group 2
Group 3
Group 4
Total
Number of patients
20
19
18
19
76
Age (years)
72.0 ± 5.1 (61–83)
72.1 ± 5.8 (63–81)
73.6 ± 4.3 (68–85)
74.6 ± 3.6 (68–82)
73.0 ± 4.8 (61–85)
Sex (male/female)
9/11
10/9
5/13
11/8
35/41
Unilateral/bilateral
10/10
10/9
5/13
5/14
30/46
Metal-on-metal THA (left/right/both)
9/9/2
4/13/2
11/7/0
9/7/3
33/36/7
DLP FD (mGy–cm)
561.3 (446.3–839.5)
544.9 (436.4–686.2)
550.8 (454.9–699.3)
706.8 (467.8–1,007.0)
560.0 (446.3–839.5)
DLP RD (mGy–cm)
459.2 (356.8–697.8)
321.0 (257.2–408.8)
228.4 (190.3–297.7)
145.7 (101.5–267.5)
 
CTDI FD (mGy)
27.2 (23.0–41.5)
24.9 (20.2–32.2)
25.1 (21.8–34.5)
29.4 (20.7–50.9)
26.5 (21.0–41.5)
CTDI RD (mGy)
22.1 (18.3–33.1)
14.2 (11.5–18.7)
10.2 (8.8–14.0)
6.2 (4.4–12.4)
 

Image acquisition and reconstruction

Patients were scanned on an iCT 256-slice CT scanner (Philips Healthcare, Best, The Netherlands) and received a consecutive FD pelvic CT and an RD pelvic CT with 20%, 40%, 57% or 80% reduced CT radiation dose for patients in group 1, 2, 3 or 4 respectively. Patients were randomly assigned to one of the four groups (Table 1). No intravenous contrast medium was used. Scanning parameters were 140 kVp, 1.0-mm slice thickness, 0.5-mm increment, 500-mm field-of-view, 0.398 pitch and a 768 × 768 image matrix. Dose modulation and dose right index (DRI) were used to reduce the radiation dose of the RD scans based on the FD scans. Images were reconstructed with iterative reconstruction, iDose4 and IMR, which can be used at 7 and 3 different levels of noise suppression respectively (Philips Healthcare, Best, The Netherlands). All FD scans were reconstructed using iDose4 level 4, which is the middle level in terms of noise suppression. Level 4 was chosen as this is used in current clinical practice. All RD scans were reconstructed using iDose4 level 4 and IMR levels 1, 2 and 3 (Fig. 1). Filter type was matched for iDose4 and IMR reconstruction using filter D and filter SharpPlus respectively. All images were reconstructed with and without the use of O-MAR (Philips Healthcare, Best, The Netherlands).

Quantitative analysis

A thin axial slice, containing the largest diameter of the head of the metal-on-metal prosthesis with most severe artefacts was used for the CT measurements. CT numbers, noise or standard deviation (SD) and CNRs between muscle and fat were measured in the gluteus maximus muscle, fat and bladder. CNRs were determined by dividing the absolute CT number difference between muscle and fat by the average SD of muscle and fat. Circular regions of interest (ROIs) were placed (by RW) in homogeneous areas where ROIs in muscle and fat were placed at the side with the least artefacts where the ROI in the bladder was placed within the artefact area. Size of the different ROIs was optimised for each patient as differences in anatomy preclude the use of uniform ROIs. Patient movement was limited as the FD and RD scans were acquired sequentially. However, as some patient movement could occur, FD and RD ROI templates were made for each patient scan (Fig. 2).

Qualitative analysis

Images reconstructed with O-MAR were used for further qualitative analysis. Thin axial FD images reconstructed with iDose4 level 4 and RD images reconstructed with IMR level 1 were analysed (Fig. 1). IMR level 1 was chosen because quantitative results were already better than FD results in terms of noise and CNR. IMR levels 2 and 3 result in even lower noise values and higher CNR, but also increases the plastic appearance. Images were blinded and were randomly and independently evaluated on seven aspects using four-point Likert scales by two musculoskeletal radiologists with 9 years’ (MB, observer 1) and 23 years' (HB, observer 2) experience. We furthermore precluded the consecutive presentation of FD and RD images of the same patient.
Artefacts (1) were classified as 0: severe artefacts with a poor diagnostic quality, 1: moderate artefacts with an impaired diagnostic quality, 2: minor artefacts with a good diagnostic quality, or 3: absence of artefacts with excellent diagnostic quality. The delineation of bony structures (2), evaluation of bone density, destruction and sclerosis (3) and evaluation of prosthetic components (4) were classified as 0: poor evaluation with very low confidence, 1: impaired evaluation with low confidence, 2: good evaluation with medium confidence, or 3: excellent evaluation with high confidence. Evaluation and presence of pseudo tumours (5), evaluation of atrophy/hypertrophy and fatty infiltration of relevant hip muscles, including the obturator internus, gluteus medius, quadratus femoris/gluteus maximus and iliopsoas/tensor fasciae latae (6) and evaluation of the recto-uterine pouch, vesico-uterine pouch and bladder wall (7) were classified as 0: poor anatomic recognition and delineation with very low confidence, 1: impaired anatomic recognition and delineation with low confidence, 2: good anatomic recognition and delineation with medium confidence, or 3: excellent anatomical recognition and delineation and high confidence. Observers were able to switch between both bone window (1,600/400) and soft-tissue window (350/40). The training session consisted of a joint meeting with both radiologists where several random cases were evaluated on all seven aspects. The training cases included iDose4 and IMR images with different grades of pathology, artefact severity and radiation dose reduction of patients with unilateral and bilateral prostheses to provide a fair representation of the population and to elaborate on differences in scores.

Statistical analysis

Shapiro–Wilk test and visual inspections were performed to determine whether data were normally distributed. Wilcoxon signed rank test was used to compare CT numbers, noise and CNR of the FD and RD images and O-MAR versus no O-MAR results due to relatively small sample sizes. Wilcoxon signed rank test was also used to compare subjective image quality scores by radiologists of FD and RD images. Differences in Likert scores in FD and RD images between groups were analysed using Fisher’s exact test. A significance level of 5% was used for all tests and all tests were two-tailed. SPSS software, version 24 was used.

Results

In total 20, 19, 18 and 19 patients were included in groups 1, 2, 3 and 4 respectively (Table 1). Mean dose–length products (DLPs) decreased from 650 to 535 mGy–cm (−18%), 600 to 359 mGy–cm (−40%), 621 to 272 mGy–cm (−56%) and 725 to 173 mGy–cm (−76%) in groups 1, 2, 3 and 4 respectively.

Quantitative analysis

Orthopaedic metal artefact reduction (O-MAR) reduced noise in the bladder by 37%, 48%, 42% and 32% and increased mean CT values in the bladder (Fig. 3) with 66, 22, 14 and 66 HU in groups 1, 2, 3 and 4 respectively (p < 0.01). In muscle and fat, CT numbers were not statistically different in O-MAR and no O-MAR images. O-MAR furthermore decreased noise in ROIs placed in muscle and fat in RD and FD reconstructions (p < 0.001). O-MAR increased the CNRs between muscle and fat in FD and in RD reconstructions (p < 0.001).
Computed tomography numbers of muscle were slightly lower in IMR 1, 2 and 3 images compared with iDose4 level 4 images (p < 0.001; Fig. 4a). Similar results were seen in fat (p < 0.001). When reducing the radiation dose by 20%, 40%, 57% or 80%, CNRs decreased when comparing FD iDose4 level 4 + O-MAR images with RD iDose4 level 4 + O-MAR images with 8%, 19%, 25% and 35% respectively (Fig. 5). With IMR, noise decreased (p < 0.001) and CNRs (p < 0.001) increased in RD IMR 1, 2 and 3 images compared with the FD iDose4 level 4 images (Figs. 4b and 5). In RD IMR level 1 with O-MAR images, CNRs between muscle and fat were on average 78%, 62%, 52% and 40% higher compared with FD iDose4 level 4 with O-MAR images in groups 1, 2, 3 and 4 respectively (Fig. 5).

Subjective analysis within observers

For observer 1, mean subjective image quality scores were lower in RD IMR with O-MAR compared with FD iDose4 with O-MAR regarding the following aspects: artefacts and diagnostic quality (p < 0.001), delineation of bony structures (p < 0.001), evaluation of bone density, destruction and sclerosis (p < 0.001) and evaluation of the recto-uterine pouch, the vesico-uterine pouch and the bladder wall (p < 0.01). For observer 2, average subjective image quality scores in RD IMR with O-MAR were lower compared with FD iDose4 with O-MAR on all seven aspects: 1 (p < 0.01), 2 (p < 0.01), 3 (p < 0.01), 4 (p < 0.05), 5 (p < 0.05), 6 (p < 0.001) and 7 (p < 0.001; Fig. 6).
Averaging the results of both observers resulted in overall lower image quality scores on all seven aspects: 1 (p < 0.001), 2 (p < 0.001), 3 (p < 0.001), 4 (p < 0.05), 5 (p < 0.05), 6 (p < 0.001) and 7 (p < 0.001). There was no statistically significant difference in subjective image quality scores among the four groups. In RD IMR images with radiation dose reduced by 20%, 40%, 57% and 80%, mean Likert scores were on average 0.17, 0.25, 0.27 and 0.24 lower respectively compared with FD iDose4 images.

Subjective analysis between observers

Subjective image quality scores were statistically different between the two observers with respect to aspects 3 (p < 0.05), 5 (p < 0.01) and 7 (p < 0.001) in FD iDose4 images and aspects 5 (p < 0.001) and 6 (p < 0.01) in FD IMR level 1 images (p < 0.05). Cohen’s Kappa values of 0.16 and lower were found, indicating no or poor reliability between observers for FD images and RD images.

Discussion

Our study reveals that CT numbers remained constant, noise decreased and CNRs between muscle and fat increased in RD IMR reconstructions compared with FD iDose4 reconstructions for all levels of IMR while reducing CT radiation dose by 20%, 40%, 57% or 80%. By reducing metal artefacts, O-MAR reduced noise, improved CT number accuracy by bringing these values closer to expected CT numbers in the bladder and improved CNRs between muscle and fat. However, subjective image quality scores were lower in RD IMR level 1 with O-MAR images compared with FD iDose4 level 4 with O-MAR images, regardless of the degree of radiation dose reduction.
Large head metal-on-metal THAs composed of a cobalt chromium molybdenum alloy caused severe metal artefacts, mainly because of extensive photon starvation. These artefacts were reduced, which resulted in an improvement of CT number accuracy in the bladder and a reduction of noise. O-MAR reduced artefacts in the bladder by increasing CT numbers and reducing noise, as illustrated in Fig. 3 (p < 0.001). O-MAR also had a positive effect on noise measured in muscle and fat, which resulted in higher CNRs on O-MAR images compared with no O-MAR images, as CT numbers remained constant. Quantitative results with respect to metal artefact reduction are in concordance with a previous phantom study [13] and other patient and phantom studies, also focusing on the CT imaging of THAs [9, 10, 1417]. Despite the fact that O-MAR improves image quality by reducing metal artefacts, we recommend additionally evaluating conventional CT images without the use of MAR software, because MAR algorithms may introduce secondary artefacts and could degrade the depiction of bone trabeculae and bone cortex [11, 12, 1822].
Filtered back-projection is a fast and robust reconstruction technique, but it cannot deal with artefacts and noise when reducing radiation dose (Fig. 7). We addressed noise as the SD of pixel intensities within an ROI, and are aware that both noise and artefact affect the SD. However, in a previous study we showed that noise reduction by O-MAR is mainly caused by a reduction of metal artefacts resulting in a lower SD, as O-MAR has no influence on images without any metal artefacts [9]. As CT numbers remained similar and noise decreased, CNR improved in RD IMR reconstructions compared with FD iDose4 reconstructions with and without the use of O-MAR (Fig. 5). As described in a previous phantom study, lower noise values and higher CNR values are found on RD images compared with FD images, even on −80% RD images [13]. Boudabbous et al. and Kuya et al. found that MBIR furthermore reduces the size of artefacts, and that MBIR allows equal or better visibility of the bone–metal interface and improved assessment of the soft tissue surrounding implants compared with FBP [4, 5]. Yasaka et al. also found that streak artefacts were reduced using MBIR [6]. In contrast to these findings, no distinct metal artefact reduction was observed when using IMR (Figs. 7, 8).
Mean Likert scores (rated from 0 to 3), with respect to artefact and diagnostic quality of 1.59 and 1.56 were found for observers 1 and 2 respectively. In general, image quality was inferior in RD IMR results compared with FD iDose4 results. Low Kappa values of 0.16 and lower indicated poor or no agreement between the two observers in RD and FD images. Differences in subjective image quality scoring are not uncommon as there will be a variation in the perception of image quality and viewing strategy among observers, which was also found by Kataria et al. [23]. Despite the fact that there were differences in subjective image quality scores between observers, it is more important to focus on differences between RD and FD scores within observers. For observer 1, the image quality of RD IMR images was rated as inferior to FD iDose4 images with respect to: artefacts and diagnostic quality (p < 0.001), delineation of bony structures (p < 0.001), evaluation of bone density, destruction and sclerosis (p < 0.001) and the recto-uterine pouch, the vesico-uterine pouch and the bladder wall (p < 0.01; Fig. 6). Similar image quality scores were seen in FD and RD results regarding the evaluation of prosthetic components, pseudo-tumours and the evaluation of relevant hip musculature. Regarding observer 2, the quality of RD IMR images was rated as inferior compared with FD iDose4 images on all seven aspects (Fig. 6). Image quality scores were statistically different between the two observers regarding aspects 3, 5 and 7 on FD iDose4 images and aspects 5 and 6 on RD IMR images. In most of these aspects, image quality scores of observer 2 were lower compared with observer 1. This can be explained by the fact that observer 2 was less experienced in evaluating IMR images. Second, observer 1 is more experienced in evaluating metal-on-metal THAs and pseudo-tumours.
When focusing on the differences in image quality scores between RD IMR results and FD iDose4 results, agreement among observers improved. In most cases, image quality of FD and RD images was rated equal for observer 1 (64.7%) and observer 2 (61.5%). In 25.9% and 30.5% of the cases the image quality of the RD scan was rated 1 point lower compared with the FD scan for observers 1 and 2 respectively. In only 1.7% and 0.9% of the cases was the RD image quality rated 2 points lower compared with the FD scan for observers 1 and 2 respectively. Higher image quality scores were seen in 7.7% and 7.1% of the cases for observers 1 and 2 respectively.
Image quality of RD images reconstructed with IMR is rated lower than images reconstructed with iDose4, regardless of the amount of radiation dose reduction. Nevertheless, it can easily be observed that image quality is insufficient in RD IMR results at 80% reduced radiation dose, especially in bilateral metal-on-metal THAs (Fig. 8). The combination of IMR and O-MAR results in a blurring or smoothening effect, especially at the level of the large head of the THA. When there is no or little metal involved, for example in the stem of the prosthesis, this effect was not observed (Figs. 10, 11). When dealing with bilateral metal-on-metal THAs, the use of IMR combined with aggressive radiation dose reduction is therefore discouraged.
We acknowledge that commonly used metrics of image quality, noise and CNR have limited utility in fully assessing image quality for IR and MBIR algorithms [2426]. Samei et al. suggested that MBIR can potentially make better use of projection data to reduce CT dose by approximately a factor of 2 and furthermore found that MBIR shows improved spatial resolution for high-contrast tasks, but reduced performance for low-contrast tasks at an RD, which may influence low-contrast object detectability [25]. They showed that MBIR has a dose reduction potential of 46–84% compared with ASIR 50%, which is comparable with iDose4. MBIR techniques improve spatial resolution while reducing noise at the same time [13, 24, 2729]. We quantitatively assessed CT image quality by measuring CT numbers, noise and CNR, but did not assess spatial resolution. Millon et al. assessed spatial resolution, low-contrast detectability, noise and CNR in a Catphan phantom and chest cadaver study using the same 256-slice iCT scanner and MBIR technique [28]. By determining edge-spread functions, the modulation transfer function, which is a measure of spatial resolution, was calculated. They found that MBIR improved the low-contrast detectability and enabled a CT radiation dose reduction. At an RD, spatial resolution became dose- and contrast-dependent. With an 80% radiation dose reduction, an increasing loss of structures on IMR images was observed and small low-contrast structures started to disappear, which was the case for both FBP and IMR [28]. Therefore, when applying large dose reductions, diagnostic performance could be compromised.
In our opinion, radiation dose can be reduced in THA imaging using MBIR combined with O-MAR, depending on the diagnostic purpose. Especially when evaluating the correct placement and status of prosthetic components in situ, RD protocols are sufficient. We have to keep in mind that image quality is already impaired when imaging a large-head MoM prosthesis; thus, it is unrealistic to expect high image quality scores anyway, especially at the height of the prosthetic head. Despite the fact that image quality was lower in RD IMR results, most soft-tissue abnormalities can be assessed, albeit with reduced diagnostic confidence. Mixed results were found regarding the use of MBIR techniques in soft tissues by other studies. Vardhanabhuti et al., Pooler et al. and Padole et al. reported that small lesions could be missed on RD images using IR and MBIR. Diagnostic confidence and reader confidence for detecting low-contrast lesions were reduced in RD MBIR results compared with the standard dose [7, 30, 31]. In high-contrast regions, a greater reduction of CT radiation dose may be applicable [30].
In a phantom study, Subhas et al. investigated lesion detectability near hardware using MBIR combined with MAR software versus FBP at reduced radiation dose levels and found that a 50% CT radiation dose reduction did not compromise the accuracy compared with FBP. They found that MBIR with MAR was also significantly more sensitive than FBP in detecting smaller lesions and lesions near large amounts of metal artefact [32]. Kataria et al. found that a radiation dose reduction can be applied; however, the diagnostic confidence in evaluating low-contrast objects and details in the liver was impaired using MBIR [23]. It is difficult to compare the outcomes of these studies with our own findings as the number of patients with abnormalities was limited in this population. With respect to evaluating the delineation of bony structures and assessment of bone density, bone destruction and sclerosis, the use of FD iDose4 reconstructions is advised. High-contrast details disappear, which could result in the loss of small details and structures such as fractures.
This study has limitations. Observers were blinded to iDose4 and IMR results, but one could easily identify IMR images owing to their plastic appearance, which was also encountered in other studies [7, 23]. An extensive training session for both observers could have enhanced the agreement between the observers. Radiologists are used to evaluating images containing noise and adaptation time is required to get used to MBIR images. Additionally, one of the observers was more familiar with evaluating MBIR images, with evaluating metal-on-metal THAs and with the assessment of pseudo-tumour formation. As this study involves follow-up CT scans of patients with metal-on-metal THAs, a relatively low incidence of soft tissue and bone abnormalities was observed. It is therefore hard to determine if clinically relevant findings would have been missed. Figures 911 illustrate that findings are also assessed in RD IMR results. It furthermore needs to be noted that the results of this study are generated with the specific IMR reconstruction algorithm using a 256-slice iCT scanner. Results cannot be compared 1-on-1 with MBIR techniques or scanner hardware of other vendors.
In conclusion, CT numbers remained constant, noise decreased and CNRs between muscle and fat increased in RD IMR reconstructions compared with FD iDose4 reconstructions for all levels of IMR while reducing CT radiation dose by 20%, 40%, 57% and 80%. Subjective image quality of RD IMR with O-MAR images decreased compared with FD iDose4 with O-MAR images in patients with large metal-on-metal THAs, regardless of the degree of radiation dose reduction. Care should be taken when using IMR in combination with reduced CT radiation dose in patients with THAs, especially in the evaluation of osseous structures and when focussing on small details and structures.

Acknowledgements

The authors would like to thank radiologist and professor in musculoskeletal radiology Mario Maas, MD, PhD, from the Amsterdam University Medical Centre (Amsterdam UMC), Amsterdam, the Netherlands, for his expert clinical input during this and many other studies. Also, we thank clinical scientist CT Julien Milles, PhD, from Philips Healthcare, Eindhoven, the Netherlands, for enabling the use of IMR at our department and for further research support.

Compliance with ethical standards

Conflicts of interest

None of the authors has any conflicts of interest to disclose.
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Orthopädie & Unfallchirurgie

Kombi-Abonnement

Mit e.Med Orthopädie & Unfallchirurgie erhalten Sie Zugang zu CME-Fortbildungen der Fachgebiete, den Premium-Inhalten der dazugehörigen Fachzeitschriften, inklusive einer gedruckten Zeitschrift Ihrer Wahl.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
2.
Zurück zum Zitat Khawaja RDA, Singh S, Otrakji A, et al. Dose reduction in pediatric abdominal CT: use of iterative reconstruction techniques across different CT platforms. Pediatr Radiol. 2015;45(7):1046–55.CrossRef Khawaja RDA, Singh S, Otrakji A, et al. Dose reduction in pediatric abdominal CT: use of iterative reconstruction techniques across different CT platforms. Pediatr Radiol. 2015;45(7):1046–55.CrossRef
3.
Zurück zum Zitat Mehta D, Thompson R, Morton T, Dhanantwari A, Shefer E. Iterative model reconstruction: simultaneously lowered computed tomography radiation dose and improved image quality. Med Phys Int J 2013;1:147-155. Mehta D, Thompson R, Morton T, Dhanantwari A, Shefer E. Iterative model reconstruction: simultaneously lowered computed tomography radiation dose and improved image quality. Med Phys Int J 2013;1:147-155.
4.
Zurück zum Zitat Boudabbous S, Arditi D, Paulin E, Syrogiannopoulou A, Becker C, Montet X. Model-based iterative reconstruction (MBIR) for the reduction of metal artifacts on CT. Am J Roentgenol. 2015;205(2):380–5.CrossRef Boudabbous S, Arditi D, Paulin E, Syrogiannopoulou A, Becker C, Montet X. Model-based iterative reconstruction (MBIR) for the reduction of metal artifacts on CT. Am J Roentgenol. 2015;205(2):380–5.CrossRef
5.
Zurück zum Zitat Kuya K, Shinohara Y, Kato A, Sakamoto M, Kurosaki M, Ogawa T. Reduction of metal artifacts due to dental hardware in computed tomography angiography: assessment of the utility of model-based iterative reconstruction. Neuroradiology. 2017;59(3):231–5.CrossRef Kuya K, Shinohara Y, Kato A, Sakamoto M, Kurosaki M, Ogawa T. Reduction of metal artifacts due to dental hardware in computed tomography angiography: assessment of the utility of model-based iterative reconstruction. Neuroradiology. 2017;59(3):231–5.CrossRef
6.
Zurück zum Zitat Yasaka K, Katsura M, Akahane M, Sato J, Matsuda I, Ohtomo K. Model-based iterative reconstruction for reduction of radiation dose in abdominopelvic CT: comparison to adaptive statistical iterative reconstruction. Springerplus. 2013;2(1):209.CrossRef Yasaka K, Katsura M, Akahane M, Sato J, Matsuda I, Ohtomo K. Model-based iterative reconstruction for reduction of radiation dose in abdominopelvic CT: comparison to adaptive statistical iterative reconstruction. Springerplus. 2013;2(1):209.CrossRef
7.
Zurück zum Zitat Padole A, Singh S, Lira D, et al. Assessment of filtered back projection, adaptive statistical, and model-based iterative reconstruction for reduced dose abdominal computed tomography. J Comput Assist Tomogr. 2015;39(4):462–7.CrossRef Padole A, Singh S, Lira D, et al. Assessment of filtered back projection, adaptive statistical, and model-based iterative reconstruction for reduced dose abdominal computed tomography. J Comput Assist Tomogr. 2015;39(4):462–7.CrossRef
8.
Zurück zum Zitat Li H, Noel C, Chen H, et al. Clinical evaluation of a commercial orthopedic metal artifact reduction tool for CT simulations in radiation therapy. Med Phys. 2012;39:7507–17.CrossRef Li H, Noel C, Chen H, et al. Clinical evaluation of a commercial orthopedic metal artifact reduction tool for CT simulations in radiation therapy. Med Phys. 2012;39:7507–17.CrossRef
9.
Zurück zum Zitat Wellenberg RHH, Boomsma MF, van Osch JAC, et al. Computed tomography imaging of a hip prosthesis using iterative model-based reconstruction and Orthopaedic metal artefact reduction: a quantitative analysis. J Comput Assist Tomogr. 2016;40(6):971–8.CrossRef Wellenberg RHH, Boomsma MF, van Osch JAC, et al. Computed tomography imaging of a hip prosthesis using iterative model-based reconstruction and Orthopaedic metal artefact reduction: a quantitative analysis. J Comput Assist Tomogr. 2016;40(6):971–8.CrossRef
10.
Zurück zum Zitat Boomsma MF, Warringa N, Edens MA, et al. Quantitative analysis of orthopedic metal artefact reduction in 64-slice computed tomography scans in large head metal-on-metal total hip replacement, a phantom study. Springerplus. 2016;5:405-15.CrossRef Boomsma MF, Warringa N, Edens MA, et al. Quantitative analysis of orthopedic metal artefact reduction in 64-slice computed tomography scans in large head metal-on-metal total hip replacement, a phantom study. Springerplus. 2016;5:405-15.CrossRef
11.
Zurück zum Zitat Shim E, Kang Y, Ahn JM, et al. Metal artifact reduction for orthopedic implants (O-MAR): usefulness in CT evaluation of reverse total shoulder arthroplasty. Am J Roentgenol. 2017;209(4):860–6.CrossRef Shim E, Kang Y, Ahn JM, et al. Metal artifact reduction for orthopedic implants (O-MAR): usefulness in CT evaluation of reverse total shoulder arthroplasty. Am J Roentgenol. 2017;209(4):860–6.CrossRef
12.
Zurück zum Zitat Andersson KM, Nowik P, Persliden J, Thunberg P, Norrman E. Metal artefact reduction in CT imaging of hip prostheses—an evaluation of commercial techniques provided by four vendors. Br J Radiol. 2015;88(1052):20140473.CrossRef Andersson KM, Nowik P, Persliden J, Thunberg P, Norrman E. Metal artefact reduction in CT imaging of hip prostheses—an evaluation of commercial techniques provided by four vendors. Br J Radiol. 2015;88(1052):20140473.CrossRef
13.
Zurück zum Zitat Wellenberg RHH, Boomsma MF, van Osch JAC, et al. Low-dose CT imaging of a total hip arthroplasty phantom using model-based iterative reconstruction and orthopedic metal artifact reduction. Skeletal Radiol. 2017;46(5):623–32.CrossRef Wellenberg RHH, Boomsma MF, van Osch JAC, et al. Low-dose CT imaging of a total hip arthroplasty phantom using model-based iterative reconstruction and orthopedic metal artifact reduction. Skeletal Radiol. 2017;46(5):623–32.CrossRef
14.
Zurück zum Zitat Gondim Teixeira P, Meyer J, Baumann C, et al. Total hip prosthesis CT with single-energy projection-based metallic artifact reduction: impact on the visualization of specific periprosthetic soft tissue structures. Skeletal Radiol. 2014;43(9):1237–46.CrossRef Gondim Teixeira P, Meyer J, Baumann C, et al. Total hip prosthesis CT with single-energy projection-based metallic artifact reduction: impact on the visualization of specific periprosthetic soft tissue structures. Skeletal Radiol. 2014;43(9):1237–46.CrossRef
15.
Zurück zum Zitat Yasaka K, Maeda E, Hanaoka S, Katsura M, Sato J, Ohtomo K. Single-energy metal artifact reduction for helical computed tomography of the pelvis in patients with metal hip prostheses. Jpn J Radiol. 2016;34(9):625–32.CrossRef Yasaka K, Maeda E, Hanaoka S, Katsura M, Sato J, Ohtomo K. Single-energy metal artifact reduction for helical computed tomography of the pelvis in patients with metal hip prostheses. Jpn J Radiol. 2016;34(9):625–32.CrossRef
16.
Zurück zum Zitat Sonoda A, Nitta N, Ushio N, et al. Evaluation of the quality of CT images acquired with the single energy metal artifact reduction (SEMAR) algorithm in patients with hip and dental prostheses and aneurysm embolization coils. Jpn J Radiol. 2015;33(11):710–6.CrossRef Sonoda A, Nitta N, Ushio N, et al. Evaluation of the quality of CT images acquired with the single energy metal artifact reduction (SEMAR) algorithm in patients with hip and dental prostheses and aneurysm embolization coils. Jpn J Radiol. 2015;33(11):710–6.CrossRef
17.
Zurück zum Zitat Subhas N, Primak AN, Obuchowski NA, et al. Iterative metal artifact reduction: evaluation and optimization of technique. Skeletal Radiol. 2014;43(12):1729–35.CrossRef Subhas N, Primak AN, Obuchowski NA, et al. Iterative metal artifact reduction: evaluation and optimization of technique. Skeletal Radiol. 2014;43(12):1729–35.CrossRef
18.
Zurück zum Zitat Han SC, Chung YE, Lee YH, Park KK, Kim MJ, Kim KW. Metal artifact reduction software used with abdominopelvic dual-energy CT of patients with metal hip prostheses: assessment of image quality and clinical feasibility. AJR Am J Roentgenol. 2014;203(4):788–95.CrossRef Han SC, Chung YE, Lee YH, Park KK, Kim MJ, Kim KW. Metal artifact reduction software used with abdominopelvic dual-energy CT of patients with metal hip prostheses: assessment of image quality and clinical feasibility. AJR Am J Roentgenol. 2014;203(4):788–95.CrossRef
19.
Zurück zum Zitat Reynoso E, Capunay C, Rasumoff A, et al. Periprosthetic artifact reduction using virtual monochromatic imaging derived from gemstone dual-energy computed tomography and dedicated software. J Comput Assist Tomogr. 2016;40(4):649–57.CrossRef Reynoso E, Capunay C, Rasumoff A, et al. Periprosthetic artifact reduction using virtual monochromatic imaging derived from gemstone dual-energy computed tomography and dedicated software. J Comput Assist Tomogr. 2016;40(4):649–57.CrossRef
20.
Zurück zum Zitat Huang J, Kerns J, Nute J, et al. An evaluation of three commercially available metal artifact reduction methods for CT imaging. Phys Med Biol. 2015;60(3):1047–67.CrossRef Huang J, Kerns J, Nute J, et al. An evaluation of three commercially available metal artifact reduction methods for CT imaging. Phys Med Biol. 2015;60(3):1047–67.CrossRef
21.
Zurück zum Zitat Pessis E, Sverzut JM, Campagna R, Guerini H, Feydy A, Drapé J-L. Reduction of metal artifact with dual-energy CT: virtual monospectral imaging with fast kilovoltage switching and metal artifact reduction software. Semin Musculoskelet Radiol. 2015;19(5):446–55.CrossRef Pessis E, Sverzut JM, Campagna R, Guerini H, Feydy A, Drapé J-L. Reduction of metal artifact with dual-energy CT: virtual monospectral imaging with fast kilovoltage switching and metal artifact reduction software. Semin Musculoskelet Radiol. 2015;19(5):446–55.CrossRef
22.
Zurück zum Zitat Bolstad K, Flatabø S, Aadnevik D, Dalehaug I, Vetti N. Metal artifact reduction in CT, a phantom study: subjective and objective evaluation of four commercial metal artifact reduction algorithms when used on three different orthopedic metal implants. Acta Radiol. 2018;59(9):111–1118.CrossRef Bolstad K, Flatabø S, Aadnevik D, Dalehaug I, Vetti N. Metal artifact reduction in CT, a phantom study: subjective and objective evaluation of four commercial metal artifact reduction algorithms when used on three different orthopedic metal implants. Acta Radiol. 2018;59(9):111–1118.CrossRef
23.
Zurück zum Zitat Kataria B, Althén JN, Smedby Ö, Persson A, Sökjer H, Sandborg M. Assessment of image quality in abdominal CT: potential dose reduction with model-based iterative reconstruction. Eur Radiol 2018;28(6):2464–73.CrossRef Kataria B, Althén JN, Smedby Ö, Persson A, Sökjer H, Sandborg M. Assessment of image quality in abdominal CT: potential dose reduction with model-based iterative reconstruction. Eur Radiol 2018;28(6):2464–73.CrossRef
24.
Zurück zum Zitat Richard S, Husarik DB, Yadava G, Murphy SN, Samei E. Towards task-based assessment of CT performance: system and object MTF across different reconstruction algorithms. Med Phys. 2012;39(7):4115–22.CrossRef Richard S, Husarik DB, Yadava G, Murphy SN, Samei E. Towards task-based assessment of CT performance: system and object MTF across different reconstruction algorithms. Med Phys. 2012;39(7):4115–22.CrossRef
25.
Zurück zum Zitat Samei E, Richard S. Assessment of the dose reduction potential of a model-based iterative reconstruction algorithm using a task-based performance metrology. Med Phys 2015;42:314–23.CrossRef Samei E, Richard S. Assessment of the dose reduction potential of a model-based iterative reconstruction algorithm using a task-based performance metrology. Med Phys 2015;42:314–23.CrossRef
26.
Zurück zum Zitat Solomon J, Mileto A, Ramirez Giraldo JC, Samei E. Diagnostic performance of an advanced modeled iterative reconstruction algorithm for multidetector CT scanner: potential for radiation dose reduction in a multireader study 1. Radiology. 2015;275(3):735–45.CrossRef Solomon J, Mileto A, Ramirez Giraldo JC, Samei E. Diagnostic performance of an advanced modeled iterative reconstruction algorithm for multidetector CT scanner: potential for radiation dose reduction in a multireader study 1. Radiology. 2015;275(3):735–45.CrossRef
27.
Zurück zum Zitat Löve A, Olsson M-L, Siemund R, Stålhammar F, Björkman-Burtscher IM, Söderberg M. Six iterative reconstruction algorithms in brain CT: a phantom study on image quality at different radiation dose levels. Br J Radiol. 2013;86(1031):1–11.CrossRef Löve A, Olsson M-L, Siemund R, Stålhammar F, Björkman-Burtscher IM, Söderberg M. Six iterative reconstruction algorithms in brain CT: a phantom study on image quality at different radiation dose levels. Br J Radiol. 2013;86(1031):1–11.CrossRef
28.
Zurück zum Zitat Millon D, Vlassenbroek A, Van Maanen AG, Cambier SE, Coche EE. Low contrast detectability and spatial resolution with model-based iterative reconstructions of MDCT images: a phantom and cadaveric study. Eur Radiol. 2017;27(3):927–37.CrossRef Millon D, Vlassenbroek A, Van Maanen AG, Cambier SE, Coche EE. Low contrast detectability and spatial resolution with model-based iterative reconstructions of MDCT images: a phantom and cadaveric study. Eur Radiol. 2017;27(3):927–37.CrossRef
29.
Zurück zum Zitat Neroladaki A, Botsikas D, Boudabbous S, Becker CD, Montet X. Computed tomography of the chest with model-based iterative reconstruction using a radiation exposure similar to chest X-ray examination: preliminary observations. Eur Radiol. 2013;23(2):360–6.CrossRef Neroladaki A, Botsikas D, Boudabbous S, Becker CD, Montet X. Computed tomography of the chest with model-based iterative reconstruction using a radiation exposure similar to chest X-ray examination: preliminary observations. Eur Radiol. 2013;23(2):360–6.CrossRef
30.
Zurück zum Zitat Pooler BD, Lubner MG, Kim DH, et al. Prospective evaluation of reduced dose computed tomography for the detection of low-contrast liver lesions: direct comparison with concurrent standard dose imaging. Eur Radiol. 2017;27(5):2055–66.CrossRef Pooler BD, Lubner MG, Kim DH, et al. Prospective evaluation of reduced dose computed tomography for the detection of low-contrast liver lesions: direct comparison with concurrent standard dose imaging. Eur Radiol. 2017;27(5):2055–66.CrossRef
31.
Zurück zum Zitat Vardhanabhuti V, Riordan RD, Mitchell GR, Hyde C, Roobottom CA. Image comparative assessment using iterative reconstructions. Invest Radiol. 2014;49(4):209–16.CrossRef Vardhanabhuti V, Riordan RD, Mitchell GR, Hyde C, Roobottom CA. Image comparative assessment using iterative reconstructions. Invest Radiol. 2014;49(4):209–16.CrossRef
32.
Zurück zum Zitat Subhas N, Pursyko CP, Polster JM, et al. Dose reduction with dedicated CT metal artifact reduction algorithm: CT phantom study. Am J Roentgenol. 2018;210(3):593–600.CrossRef Subhas N, Pursyko CP, Polster JM, et al. Dose reduction with dedicated CT metal artifact reduction algorithm: CT phantom study. Am J Roentgenol. 2018;210(3):593–600.CrossRef
Metadaten
Titel
CT radiation dose reduction in patients with total hip arthroplasties using model-based iterative reconstruction and orthopaedic metal artefact reduction
verfasst von
Ruud H. H. Wellenberg
Jochen A. C. van Osch
Henk J. Boelhouwers
Mireille A. Edens
Geert J. Streekstra
Harmen B. Ettema
Martijn F. Boomsma
Publikationsdatum
24.04.2019
Verlag
Springer Berlin Heidelberg
Erschienen in
Skeletal Radiology / Ausgabe 11/2019
Print ISSN: 0364-2348
Elektronische ISSN: 1432-2161
DOI
https://doi.org/10.1007/s00256-019-03206-z

Weitere Artikel der Ausgabe 11/2019

Skeletal Radiology 11/2019 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.