CXCL13 in laboratory diagnosis of Lyme neuroborreliosis—the performance of the recomBead and ReaScan CXCL13 assays in human cerebrospinal fluid samples

verfasst von: Sofie Haglund, Malin Lager, Paula Gyllemark, Gärda Andersson, Oskar Ekelund, Martin Sundqvist, Anna J. Henningsson

Erschienen in: European Journal of Clinical Microbiology & Infectious Diseases | Ausgabe 1/2022

Abstract

The chemokine CXCL13 is used as complement to serology in the diagnostics of Lyme neuroborreliosis (LNB). We evaluated and compared the semi-quantitative, cassette-based ReaScan CXCL13 assay with the quantitative recomBead CXCL13 assay using a collection of 209 cerebrospinal fluid samples. The categorical agreement between results interpreted as negative, grey zone, and positive by the two methods was 87%. The diagnostic sensitivity was higher using the recomBead assay, whereas specificity was higher using ReaScan. Few manual steps, and a short turn-around time with no batching of samples makes the ReaScan CXCL13 assay an attractive complement to serology in the diagnostics of LNB.
Hinweise

Supplementary Information

The online version contains supplementary material available at https://​doi.​org/​10.​1007/​s10096-021-04350-y.

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

Lyme neuroborreliosis (LNB), the most common form of disseminated infection by bacteria in the Borrelia burgdorferi sensu lato complex in Europe [1], typically presents as cranial nerve palsy, sub-acute meningitis or other neurological impairments [2]. Definite LNB is defined as (i) neurological symptoms attributable to LNB, (ii) pleocytosis in cerebrospinal fluid (CSF, mononuclear cells > 5/µL CSF), and (iii) intrathecally produced anti-Borrelia antibodies (elevated antibody index, here referred to as positive Bb AI) [3].
The B cell-attracting chemokine CXCL13 has in previous studies been shown to be a useful and reliable complement to serology as it is elevated in the CSF of patients with very early LNB where the sensitivity of antibody tests may be low. Furthermore, CXCL13 has been proven useful in the monitoring of response to antibiotic treatment [47], and for the discriminating of acute LNB from other central nervous system (CNS) disorders [810]. However, the appropriate cut-off level varies between studies and has been debated [614].
The main aim of this work was to evaluate the rapid and semi-quantitative ReaScan CXCL13 assays (Reagena Ltd, Toivala, Finland) performance and compare it to the quantitative recomBead CXCL13 assay (Mikrogen Diagnostik, GmbH, Neuried, Germany), which was used in clinical routine in the laboratory in Jönköping as a complement to serology in selected cases and run once a week following batching of samples [1517]. Secondly, we aimed to investigate the impact of the pre-analytical storage conditions on the test results.

Materials and methods

Study population

CSF samples (n = 209) from definite LNB (n = 53), probable early LNB (n = 16) and non-LNB (n = 140) patients were retrospectively included from the Clinical Microbiology Laboratories in Region Jönköping County, Region Kronoberg, and Örebro University Hospital. Patients with symptoms strongly suggestive of LNB, such as subacute meningitis or peripheral facial palsy, and CSF pleocytosis, but, due to short symptom duration, with negative Bb AI, were in our study classified as probable early LNB [3]. They were all considered as, and treated for, LNB by their physician and responded well to the treatment. If the criteria for definite or probable early LNB were not fulfilled, the patient was classified as non-LNB. The non-LNB group included patients both with and without CSF pleocytosis, whereof 22 had other confirmed CNS infections. Demographic and clinical data were retrieved from the patients’ medical records or the medical referral (Table 1). Samples were collected in tubes without additives and either analysed as fresh samples, or stored at − 20 °C before the CXCL13 analysis.
Table 1
Patient characteristics, n = 209
 
CSF mononuclear cells > 5/µL CSF
CSF Bb AI
Age (years, mean, range)
Male (%)
Symptom duration (days, mean, range)
Definite LNB (53)
 + 
 + 
42 (2–87)
46
15 (1–280)1
Probable LNB (16)
 + 
 −
21 (3–76)
47
3 (1–28)2
Non-LNB (140)
 − (73)/ + (67)
 −
47 (4–83)
49
30 (1–3650)3
CSF cerebrospinal fluid, Bb AI Borrelia specific antibody index, LNB Lyme neuroborreliosis. 1Data missing in three cases, 2data missing in three cases, 3data missing in 52 cases

CXCL13 methods

The ReaScan CXCL13 is a rapid cassette-based immuno chromatographic system. The reader values are translated to semi-quantitative CXCL13 concentrations interpreted as < 250 pg/mL (negative), CXCL13 250–500 pg/mL (grey zone), and CXCL13 > 500 pg/mL (positive/suspected LNB), based on the instructions by the manufacturer.
In the quantitative recomBead CXCL13 assay, based on the Luminex xMAP® technology, samples are analysed in batches. The range of measurement is 9–1000 pg/mL. The CXCL13 concentration is interpreted as CXCL13 < 190 pg/mL (negative), CXCL13 190–300 pg/mL (grey zone), and CXCL13 > 300 pg/mL (positive/suspected LNB), based on the instructions by the manufacturer.

Statistical analyses

Both the ReaScan and recomBead CXCL13 assays are CE/IVD validated by their manufacturers. Their recommended cut-off values were applied in the study, since these are most likely the cut-offs used by laboratories using these assays. Grey zone results were interpreted as positives in the evaluation of sensitivity and specificity, based on the tradition in our laboratory. In the calculations, patients with probable early LNB were considered as LNB, as they were diagnosed with and treated for LNB by their physician. The correlation between variables was investigated by Spearman rank correlation analysis. A P value < 0.05 was considered statistically significant. The impact of repeated freeze thawing (− 20 °C) and cold storage at 4–8 °C for up to one week was compared with results based on fresh samples with ReaScan CXCL13.

Results

Overall, the categorical agreement between ReaScan CXCL13 and recomBead CXCL13 over the three CXCL13 concentration intervals was 87%. The reader values following analysis with ReaScan correlated with the CXCL13 concentration received after analysis with the recomBead assay; rs = 0.76, P < 0.001.
The categorical agreement between the two methods is presented in Table 2. When all grey zone results were handled as positives, the overall agreement between the two methods was 90%; in negative samples it was 98%, and in positive samples 74%.
Table 2
Categorical agreement between the ReaScan and recomBead CXCL13 assays, with recomBead as reference (n = 209)
recomBead1
 
ReaScan2
  
Agreement (%)
  
Negative
Grey zone
Positive
 
Negative
140
137
2
1
98
Grey zone
10
10
0
0
0
Positive
59
8
7
44
75
1recomBead CXCL13 negative; CXCL13 < 190 pg/mL, grey zone; CXCL13 191–300 pg/mL, positive; CXCL13 > 300 pg/mL. 2ReaScan CXCL13 negative; CXCL13 < 250 pg/mL, grey zone; CXCL13 250–500 pg/mL, positive; CXCL13 > 500 pg/mL
The diagnostic sensitivity, including the entire sample collection (n = 209), was lower for the ReaScan CXCL13 assay (78%) compared with the recomBead assay (86%). However, the specificity of ReaScan CXCL13 was higher (95% vs. 91%), (Table 3). In patients with probable early LNB, ReaScan showed a sensitivity of 43% and recomBead 50%.
Table 3
Results from all cerebrospinal fluid samples combined, analysed with the ReaScan CXCL13 assay and the recomBead CXCL13 assay, n = 209
 
LNB (n = 69)1
Non-LNB (n = 140)2
Sensitivity
Specificity
ReaScan CXCL13
  Positive
54
7
  
  Negative
15
133
78%
95%
recomBead CXCL13
  Positive
59
13
  
  Negative
10
127
86%
91%
LNB Lyme neuroborreliosis. 1The LNB group comprised both definite LNB and patients with short symptom duration; probable early LNB. 2The non-LNB group included patients both with and without cerebrospinal fluid pleocytosis
In samples from patients with CNS infections other than LNB (n = 22), including 10 patients with tick-borne encephalitis (TBE), the categorical interpretation of the results differed between the methods in six cases (Supplementary Table 1).
The study of the impact of pre-analytical storage conditions on the ReaScan CXCL13 assays performance indicated that cold storage for up to three days or freeze thawing up to two cycles have minor effects on the analytical performance, at least at higher concentrations of CXCL13 (Supplementary Table 2).

Discussion

In this study, which included CSF samples from patients classified based on clinical and laboratory parameters, the recomBead CXCL13 assay showed a somewhat higher diagnostic sensitivity, whereas the diagnostic specificity was slightly higher using ReaScan CXCL13. These observations are in agreement with previous comparisons between the ReaScan CXCL13 assay and enzyme-linked immunosorbent assays including non-LNB populations comprising a mix of patients with and without CFS pleocytosis [18, 19]. The specificity noticed in our comparison (95%) corresponded with the level suggested as sufficient for diagnostic accuracy by others [14].
The categorical agreement between the assays studied here was 87% overall, 98% in samples interpreted as negative using recomBead CXCL13, and 75% in positive samples. The observed discrepancies may be due to differences in antigen recognition by the detection antibodies used, a feature common to many unstandardised immunological assays aiming to detect the same antigen [2022].
The correlation between the reader values of ReaScan CXCL13 and the concentration of CXCL13 measured using recomBead was strong (rs = 0.76). However, samples over the measurement range of the recomBead CXCL13 assay were not further diluted in our comparison, and therefore the correlation may actually have been underestimated.
CXCL13 may increase in CSF in other inflammatory conditions of the CNS than LNB, however usually to lower concentrations [810]. When the performance of the assays was studied using samples from 22 patients with other CNS infections, including TBE, which is an important differential diagnosis to LNB, six discrepant results were noticed. Three samples, one from a patient with enterovirus meningitis and two with TBE, generated low values indicating negative results with ReaScan CXCL13, but grey zone or high concentrations with recomBead CXCL13. Another two samples, one from a patient with herpes simplex 2 meningitis and one with meningitis due to Streptococcus pneumoniae, showed grey zone results with the ReaScan CXCL13, but high concentrations with recomBead CXCL13. However, one sample from a patient with TBE generated a high value indicating a positive result with ReaScan, but a low concentration with recomBead. Taken together, these findings may reflect the higher specificity for LNB noticed for ReaScan CXCL13 in our study, but also that CXCL13 in TBE merits further investigation and that clinicians should be aware of the fact that CXCL13 may be elevated in the CSF from TBE patients as well.
The stability of the CXCL13 chemokine may be affected by the pre-analytical storage conditions, which may lead to erroneous results. Even though cold-storage for up to three days had minor impact on the results, our data indicate that samples should be stored frozen (− 20 °C) if it is not possible to analyse the fresh CSF samples immediately. This observation is supported by the results presented by Hytönen et al. [10]. They demonstrated that up to five freeze–thaw cycles did not affect the CXCL13 concentration significantly. However, those data were based on one sample only. Our samples did not undergo more than two freeze–thaw cycles, and therefore we can assume that they accurately reflect the CXCL13 concentration in fresh samples.
The time to an available test result using ReaScan CXCL13 was approximately 30 min, whereas the analytical turnaround time (TAT) for recomBead CXCL13 is approximately 4 h. Apart from the gain in TAT for the ReaScan CXCL13 assay, the possibility of immediate single-unit analysis of the samples after arrival to the laboratory, without batching of samples, which in our laboratory may be up to one week, is a major advantage for the ReaScan CXCL13 assay.
In conclusion, our results demonstrate that the recomBead and the ReaScan CXCL13 assays show comparable performance in the identification of LNB. The advantages with fewer manual steps and shorter TAT with no batching of samples makes the ReaScan CXCL13 assay an attractive alternative suitable even for smaller laboratories, and the rapid test results may support clinicians in their treatment decisions.

Acknowledgements

The authors would like to thank Anna-Karin Åberg and Anna-Maria Delis, Department of Laboratory Medicine, Örebro University Hospital, Örebro, for excellent technical assistance.

Declarations

Ethics approval

This study was performed in line with the principles of the Declaration of Helsinki and was approved by the Regional Ethical Review Board in Linköping, Sweden, Dnr 2012/246–31, and 2018/525–31.
All participants gave informed consent to have their samples stored in a biobank for future research, development of diagnostic methods and quality control.
Not applicable.

Conflict of interest

Anna J Henningsson has a collaborative research agreement with Reagena Ltd within the Eurostars programme, co-funded by the European Union and the EUREKA network. Apart from this, the authors have no conflicts of interest to declare that are relevant to the content of this article.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

© Springer Medizin

Bis 11. April 2024 bestellen und im ersten Jahr 50 % sparen!

e.Med Innere Medizin

Kombi-Abonnement

Mit e.Med Innere Medizin erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Innere Medizin, den Premium-Inhalten der internistischen Fachzeitschriften, inklusive einer gedruckten internistischen Zeitschrift Ihrer Wahl.

© Springer Medizin

Bis 11. April 2024 bestellen und im ersten Jahr 50 % sparen!

Anhänge

Supplementary Information

Below is the link to the electronic supplementary material.
Literatur
1.
Zurück zum Zitat Cimmino MA (1998) Relative frequency of Lyme borreliosis and of its clinical manifestations in Europe. European Community Concerted Action on Risk Assessment in Lyme Borreliosis. Infection 26(5):298–300CrossRef Cimmino MA (1998) Relative frequency of Lyme borreliosis and of its clinical manifestations in Europe. European Community Concerted Action on Risk Assessment in Lyme Borreliosis. Infection 26(5):298–300CrossRef
2.
Zurück zum Zitat Stanek G, Strle F (2009) Lyme borreliosis: a European perspective on diagnosis and clinical management. Curr Opin Infect Dis 22(5):450–454CrossRef Stanek G, Strle F (2009) Lyme borreliosis: a European perspective on diagnosis and clinical management. Curr Opin Infect Dis 22(5):450–454CrossRef
3.
Zurück zum Zitat Mygland A, Ljostad U, Fingerle V, Rupprecht T, Schmutzhard E, Steiner I, European Federation of Neurological S (2010) EFNS guidelines on the diagnosis and management of European Lyme neuroborreliosis. Eur J Neurol 17(1):8–16 (e11-14)CrossRef Mygland A, Ljostad U, Fingerle V, Rupprecht T, Schmutzhard E, Steiner I, European Federation of Neurological S (2010) EFNS guidelines on the diagnosis and management of European Lyme neuroborreliosis. Eur J Neurol 17(1):8–16 (e11-14)CrossRef
4.
Zurück zum Zitat Ljostad U, Skarpaas T, Mygland A (2007) Clinical usefulness of intrathecal antibody testing in acute Lyme neuroborreliosis. Eur J Neurol 14(8):873–876CrossRef Ljostad U, Skarpaas T, Mygland A (2007) Clinical usefulness of intrathecal antibody testing in acute Lyme neuroborreliosis. Eur J Neurol 14(8):873–876CrossRef
5.
Zurück zum Zitat Rupprecht TA, Pfister HW, Angele B, Kastenbauer S, Wilske B, Koedel U (2005) The chemokine CXCL13 (BLC): a putative diagnostic marker for neuroborreliosis. Neurology 65(3):448–450CrossRef Rupprecht TA, Pfister HW, Angele B, Kastenbauer S, Wilske B, Koedel U (2005) The chemokine CXCL13 (BLC): a putative diagnostic marker for neuroborreliosis. Neurology 65(3):448–450CrossRef
6.
Zurück zum Zitat Sillanpaa H, Skogman BH, Sarvas H, Seppala IJ, Lahdenne P (2013) Cerebrospinal fluid chemokine CXCL13 in the diagnosis of neuroborreliosis in children. Scand J Infect Dis 45(7):526–530CrossRef Sillanpaa H, Skogman BH, Sarvas H, Seppala IJ, Lahdenne P (2013) Cerebrospinal fluid chemokine CXCL13 in the diagnosis of neuroborreliosis in children. Scand J Infect Dis 45(7):526–530CrossRef
7.
Zurück zum Zitat Bremell D, Mattsson N, Edsbagge M, Blennow K, Andreasson U, Wikkelso C, Zetterberg H, Hagberg L (2013) Cerebrospinal fluid CXCL13 in Lyme neuroborreliosis and asymptomatic HIV infection. BMC Neurol 13:2CrossRef Bremell D, Mattsson N, Edsbagge M, Blennow K, Andreasson U, Wikkelso C, Zetterberg H, Hagberg L (2013) Cerebrospinal fluid CXCL13 in Lyme neuroborreliosis and asymptomatic HIV infection. BMC Neurol 13:2CrossRef
8.
Zurück zum Zitat Schmidt C, Plate A, Angele B, Pfister HW, Wick M, Koedel U, Rupprecht TA (2011) A prospective study on the role of CXCL13 in Lyme neuroborreliosis. Neurology 76(12):1051–1058CrossRef Schmidt C, Plate A, Angele B, Pfister HW, Wick M, Koedel U, Rupprecht TA (2011) A prospective study on the role of CXCL13 in Lyme neuroborreliosis. Neurology 76(12):1051–1058CrossRef
9.
Zurück zum Zitat van Burgel ND, Bakels F, Kroes AC, van Dam AP (2011) Discriminating Lyme neuroborreliosis from other neuroinflammatory diseases by levels of CXCL13 in cerebrospinal fluid. J Clin Microbiol 49(5):2027–2030CrossRef van Burgel ND, Bakels F, Kroes AC, van Dam AP (2011) Discriminating Lyme neuroborreliosis from other neuroinflammatory diseases by levels of CXCL13 in cerebrospinal fluid. J Clin Microbiol 49(5):2027–2030CrossRef
10.
Zurück zum Zitat Hytonen J, Kortela E, Waris M, Puustinen J, Salo J, Oksi J (2014) CXCL13 and neopterin concentrations in cerebrospinal fluid of patients with Lyme neuroborreliosis and other diseases that cause neuroinflammation. J Neuroinflammation 11:103CrossRef Hytonen J, Kortela E, Waris M, Puustinen J, Salo J, Oksi J (2014) CXCL13 and neopterin concentrations in cerebrospinal fluid of patients with Lyme neuroborreliosis and other diseases that cause neuroinflammation. J Neuroinflammation 11:103CrossRef
11.
Zurück zum Zitat Senel M, Rupprecht TA, Tumani H, Pfister HW, Ludolph AC, Brettschneider J (2010) The chemokine CXCL13 in acute neuroborreliosis. J Neurol Neurosurg Psychiatry 81(8):929–933CrossRef Senel M, Rupprecht TA, Tumani H, Pfister HW, Ludolph AC, Brettschneider J (2010) The chemokine CXCL13 in acute neuroborreliosis. J Neurol Neurosurg Psychiatry 81(8):929–933CrossRef
12.
Zurück zum Zitat Tjernberg I, Henningsson AJ, Eliasson I, Forsberg P, Ernerudh J (2011) Diagnostic performance of cerebrospinal fluid chemokine CXCL13 and antibodies to the C6-peptide in Lyme neuroborreliosis. J Infect 62(2):149–158CrossRef Tjernberg I, Henningsson AJ, Eliasson I, Forsberg P, Ernerudh J (2011) Diagnostic performance of cerebrospinal fluid chemokine CXCL13 and antibodies to the C6-peptide in Lyme neuroborreliosis. J Infect 62(2):149–158CrossRef
13.
Zurück zum Zitat Cerar T, Ogrinc K, Lotric-Furlan S, Kobal J, Levicnik-Stezinar S, Strle F, Ruzic-Sabljic E (2013) Diagnostic value of cytokines and chemokines in Lyme neuroborreliosis. Clin Vaccine Immunol 20(10):1578–1584CrossRef Cerar T, Ogrinc K, Lotric-Furlan S, Kobal J, Levicnik-Stezinar S, Strle F, Ruzic-Sabljic E (2013) Diagnostic value of cytokines and chemokines in Lyme neuroborreliosis. Clin Vaccine Immunol 20(10):1578–1584CrossRef
14.
Zurück zum Zitat Rupprecht TA, Manz KM, Fingerle V, Lechner C, Klein M, Pfirrmann M, Koedel U (2018) Diagnostic value of cerebrospinal fluid CXCL13 for acute Lyme neuroborreliosis. A systematic review and meta-analysis. Clin Microbiol Infect 24(12):1234–1240CrossRef Rupprecht TA, Manz KM, Fingerle V, Lechner C, Klein M, Pfirrmann M, Koedel U (2018) Diagnostic value of cerebrospinal fluid CXCL13 for acute Lyme neuroborreliosis. A systematic review and meta-analysis. Clin Microbiol Infect 24(12):1234–1240CrossRef
15.
Zurück zum Zitat Skogman BH, Lager M, Henningsson AJ, Tjernberg I (2017) The recomBead Borrelia antibody index, CXCL13 and total IgM index for laboratory diagnosis of Lyme neuroborreliosis in children. Eur J Clin Microbiol Infect Dis 36(11):2221–2229CrossRef Skogman BH, Lager M, Henningsson AJ, Tjernberg I (2017) The recomBead Borrelia antibody index, CXCL13 and total IgM index for laboratory diagnosis of Lyme neuroborreliosis in children. Eur J Clin Microbiol Infect Dis 36(11):2221–2229CrossRef
16.
Zurück zum Zitat Henningsson AJ, Lager M, Brannstrom R, Tjernberg I, Skogman BH (2018) The chemokine CXCL13 in cerebrospinal fluid in children with Lyme neuroborreliosis. Eur J Clin Microbiol Infect Dis 37(10):1983–1991CrossRef Henningsson AJ, Lager M, Brannstrom R, Tjernberg I, Skogman BH (2018) The chemokine CXCL13 in cerebrospinal fluid in children with Lyme neuroborreliosis. Eur J Clin Microbiol Infect Dis 37(10):1983–1991CrossRef
17.
Zurück zum Zitat Henningsson AJ, Gyllemark P, Lager M, Skogman BH, Tjernberg I (2016) Evaluation of two assays for CXCL13 analysis in cerebrospinal fluid for laboratory diagnosis of Lyme neuroborreliosis. APMIS 124(11):985–990CrossRef Henningsson AJ, Gyllemark P, Lager M, Skogman BH, Tjernberg I (2016) Evaluation of two assays for CXCL13 analysis in cerebrospinal fluid for laboratory diagnosis of Lyme neuroborreliosis. APMIS 124(11):985–990CrossRef
18.
Zurück zum Zitat Pietikainen A, Oksi J, Hytonen J (2018) Point-of-care testing for CXCL13 in Lyme neuroborreliosis. Diagn Microbiol Infect Dis 91(3):226–228CrossRef Pietikainen A, Oksi J, Hytonen J (2018) Point-of-care testing for CXCL13 in Lyme neuroborreliosis. Diagn Microbiol Infect Dis 91(3):226–228CrossRef
19.
Zurück zum Zitat Ziegler K, Rath A, Schoerner C, Meyer R, Bertsch T, Erbguth F, Bogdan C, Steinmann J, Held J (2020) Comparative analysis of the Euroimmun CXCL13 enzyme-linked immunosorbent assay and the ReaScan lateral flow immunoassay for diagnosis of Lyme neuroborreliosis. J Clin Microbiol 58(9):e00207-20CrossRef Ziegler K, Rath A, Schoerner C, Meyer R, Bertsch T, Erbguth F, Bogdan C, Steinmann J, Held J (2020) Comparative analysis of the Euroimmun CXCL13 enzyme-linked immunosorbent assay and the ReaScan lateral flow immunoassay for diagnosis of Lyme neuroborreliosis. J Clin Microbiol 58(9):e00207-20CrossRef
20.
Zurück zum Zitat Labaere D, Smismans A, Van Olmen A, Christiaens P, D’Haens G, Moons V, Cuyle PJ, Frans J, Bossuyt P (2014) Comparison of six different calprotectin assays for the assessment of inflammatory bowel disease. United European Gastroenterol J 2(1):30–37CrossRef Labaere D, Smismans A, Van Olmen A, Christiaens P, D’Haens G, Moons V, Cuyle PJ, Frans J, Bossuyt P (2014) Comparison of six different calprotectin assays for the assessment of inflammatory bowel disease. United European Gastroenterol J 2(1):30–37CrossRef
21.
Zurück zum Zitat Lager M, Dessau RB, Wilhelmsson P, Nyman D, Jensen GF, Matussek A, Lindgren PE, Henningsson AJ, ScandTick Biobank Study G, Baqir H, Serrander L, Johansson M, Tjernberg I, Skarstein I, Ulvestad E, Grude N, Pedersen AB, Bredberg A, Veflingstad R, Wass L, Aleke J, Nordberg M, Nyberg C, Perander L, Bojesson C, Sjoberg E, Lorentzen AR, Eikeland R, Noraas S, Henriksson GA (1933) Petranyi G 2019 Serological diagnostics of Lyme borreliosis: comparison of assays in twelve clinical laboratories in Northern Europe. Eur J Clin Microbiol Infect Dis 38(10):1945 Lager M, Dessau RB, Wilhelmsson P, Nyman D, Jensen GF, Matussek A, Lindgren PE, Henningsson AJ, ScandTick Biobank Study G, Baqir H, Serrander L, Johansson M, Tjernberg I, Skarstein I, Ulvestad E, Grude N, Pedersen AB, Bredberg A, Veflingstad R, Wass L, Aleke J, Nordberg M, Nyberg C, Perander L, Bojesson C, Sjoberg E, Lorentzen AR, Eikeland R, Noraas S, Henriksson GA (1933) Petranyi G 2019 Serological diagnostics of Lyme borreliosis: comparison of assays in twelve clinical laboratories in Northern Europe. Eur J Clin Microbiol Infect Dis 38(10):1945
22.
Zurück zum Zitat Khoury DS, Wheatley AK, Ramuta MD, Reynaldi A, Cromer D, Subbarao K, O’Connor DH, Kent SJ, Davenport MP (2020) Measuring immunity to SARS-CoV-2 infection: comparing assays and animal models. Nat Rev Immunol 20(12):727–738CrossRef Khoury DS, Wheatley AK, Ramuta MD, Reynaldi A, Cromer D, Subbarao K, O’Connor DH, Kent SJ, Davenport MP (2020) Measuring immunity to SARS-CoV-2 infection: comparing assays and animal models. Nat Rev Immunol 20(12):727–738CrossRef
Metadaten
Titel
CXCL13 in laboratory diagnosis of Lyme neuroborreliosis—the performance of the recomBead and ReaScan CXCL13 assays in human cerebrospinal fluid samples
verfasst von
Sofie Haglund
Malin Lager
Paula Gyllemark
Gärda Andersson
Oskar Ekelund
Martin Sundqvist
Anna J. Henningsson
Publikationsdatum
09.10.2021
Verlag
Springer Berlin Heidelberg
Erschienen in
European Journal of Clinical Microbiology & Infectious Diseases / Ausgabe 1/2022
Print ISSN: 0934-9723
Elektronische ISSN: 1435-4373
DOI
https://doi.org/10.1007/s10096-021-04350-y

Weitere Artikel der Ausgabe 1/2022