Assessment of hepatic arterial hemodynamics with 4D flow MRI: in vitro analysis of motion and spatial resolution related error and in vivo feasibility study in 20 volunteers

verfasst von: Ivan P. Dimov, Cyril Tous, Ning Li, Maxime Barat, Tim Bomberna, Charlotte Debbaut, Ning Jin, Gerald Moran, An Tang, Gilles Soulez

Erschienen in: European Radiology | Ausgabe 12/2022

Einloggen, um Zugang zu erhalten

Abstract

Objectives

To assess the ability of four-dimensional (4D) flow MRI to measure hepatic arterial hemodynamics by determining the effects of spatial resolution and respiratory motion suppression in vitro and its applicability in vivo with comparison to two-dimensional (2D) phase-contrast MRI.

Methods

A dynamic hepatic artery phantom and 20 consecutive volunteers were scanned. The accuracies of Cartesian 4D flow sequences with k-space reordering and navigator gating at four spatial resolutions (0.5- to 1-mm isotropic) and navigator acceptance windows (± 8 to ± 2 mm) and one 2D phase-contrast sequence (0.5-mm in -plane) were assessed in vitro at 3 T. Two sequences centered on gastroduodenal and hepatic artery branches were assessed in vivo for intra - and interobserver agreement and compared to 2D phase-contrast.

Results

In vitro, higher spatial resolution led to a greater decrease in error than narrower navigator window (30.5 to −4.67% vs −6.64 to −4.67% for flow). In vivo, hepatic and gastroduodenal arteries were more often visualized with the higher resolution sequence (90 vs 71%). Despite similar interobserver agreement (κ = 0.660 and 0.704), the higher resolution sequence had lower variability for area (CV = 20.04 vs 30.67%), flow (CV = 34.92 vs 51.99%), and average velocity (CV = 26.47 vs 44.76%). 4D flow had lower differences between inflow and outflow at the hepatic artery bifurcation (11.03 ± 5.05% and 15.69 ± 6.14%) than 2D phase-contrast (28.77 ± 21.01%).

Conclusion

High-resolution 4D flow can assess hepatic artery anatomy and hemodynamics with improved accuracy, greater vessel visibility, better interobserver reliability, and internal consistency.

Key Points

Motion-suppressed Cartesian four-dimensional (4D) flow MRI with higher spatial resolution provides more accurate measurements even when accepted respiratory motion exceeds voxel size.
4D flow MRI with higher spatial resolution provides substantial interobserver agreement for visualization of hepatic artery branches.
Lower peak and average velocities and a trend toward better internal consistency were observed with 4D flow MRI as compared to 2D phase-contrast.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Borhani AA, Elsayes KM, Catania R et al (2021) Imaging evaluation of living liver donor candidates: techniques, protocols, and anatomy. Radiographics 6:1572–1591 Borhani AA, Elsayes KM, Catania R et al (2021) Imaging evaluation of living liver donor candidates: techniques, protocols, and anatomy. Radiographics 6:1572–1591
2.
Zurück zum Zitat Ohta K, Shimohira M, Hashizume T et al (2019) Identification of the feeding arteries of hepatocellular carcinomas by performing dual arterial phase CT during pre-transarterial chemoembolization angiography. Abdom Radiol (NY) 6:2276–2282 Ohta K, Shimohira M, Hashizume T et al (2019) Identification of the feeding arteries of hepatocellular carcinomas by performing dual arterial phase CT during pre-transarterial chemoembolization angiography. Abdom Radiol (NY) 6:2276–2282
3.
Zurück zum Zitat Bogren HG, Buonocore MH (1999) 4D magnetic resonance velocity mapping of blood flow patterns in the aorta in young vs. elderly normal subjects. J Magn Reson Imaging 5:861–869 CrossRef Bogren HG, Buonocore MH (1999) 4D magnetic resonance velocity mapping of blood flow patterns in the aorta in young vs. elderly normal subjects. J Magn Reson Imaging 5:861–869 CrossRef
4.
Zurück zum Zitat Carlisle KM, Halliwell M, Read AE, Wells PN (1992) Estimation of total hepatic blood flow by duplex ultrasound. Gut 1:92–97 CrossRef Carlisle KM, Halliwell M, Read AE, Wells PN (1992) Estimation of total hepatic blood flow by duplex ultrasound. Gut 1:92–97 CrossRef
5.
Zurück zum Zitat Numata K, Tanaka K, Kiba T et al (1999) Hepatic arterial resistance after mixed-meal ingestion in healthy subjects and patients with chronic liver disease. J Clin Ultrasound 5:239–248 CrossRef Numata K, Tanaka K, Kiba T et al (1999) Hepatic arterial resistance after mixed-meal ingestion in healthy subjects and patients with chronic liver disease. J Clin Ultrasound 5:239–248 CrossRef
6.
Zurück zum Zitat Yzet T, Bouzerar R, Baledent O et al (2008) Dynamic measurements of total hepatic blood flow with phase contrast MRI. Eur J Rad 1:119–124 Yzet T, Bouzerar R, Baledent O et al (2008) Dynamic measurements of total hepatic blood flow with phase contrast MRI. Eur J Rad 1:119–124
7.
Zurück zum Zitat Jin N, Lewandowski RJ, Omary RA, Larson AC (2009) Respiratory self-gating for free-breathing abdominal phase-contrast blood flow measurements. J Magn Reson Imaging 4:860–868 CrossRef Jin N, Lewandowski RJ, Omary RA, Larson AC (2009) Respiratory self-gating for free-breathing abdominal phase-contrast blood flow measurements. J Magn Reson Imaging 4:860–868 CrossRef
8.
Zurück zum Zitat Bane O, Peti S, Wagner M et al (2019) Hemodynamic measurements with an abdominal 4D flow MRI sequence with spiral sampling and compressed sensing in patients with chronic liver disease. J Magn Reson Imaging 4:994–1005 CrossRef Bane O, Peti S, Wagner M et al (2019) Hemodynamic measurements with an abdominal 4D flow MRI sequence with spiral sampling and compressed sensing in patients with chronic liver disease. J Magn Reson Imaging 4:994–1005 CrossRef
9.
Zurück zum Zitat Brunsing RL, Brown D, Almahoud H et al (2021) Quantification of the hemodynamic changes of cirrhosis with free-breathing self-navigated MRI. J Magn Reson Imaging 5:1410–1421 CrossRef Brunsing RL, Brown D, Almahoud H et al (2021) Quantification of the hemodynamic changes of cirrhosis with free-breathing self-navigated MRI. J Magn Reson Imaging 5:1410–1421 CrossRef
10.
Zurück zum Zitat Parekh K, Markl M, Rose M et al (2017) 4D flow MR imaging of the portal venous system: a feasibility study in children. Eur Radiol 2:832–840 CrossRef Parekh K, Markl M, Rose M et al (2017) 4D flow MR imaging of the portal venous system: a feasibility study in children. Eur Radiol 2:832–840 CrossRef
11.
Zurück zum Zitat Stankovic Z, Jung B, Collins J et al (2014) Reproducibility study of four-dimensional flow MRI of arterial and portal venous liver hemodynamics: influence of spatio-temporal resolution. Magn Reson Med 2:477–484 CrossRef Stankovic Z, Jung B, Collins J et al (2014) Reproducibility study of four-dimensional flow MRI of arterial and portal venous liver hemodynamics: influence of spatio-temporal resolution. Magn Reson Med 2:477–484 CrossRef
12.
Zurück zum Zitat Dyvorne H, Knight-Greenfield A, Jajamovich G et al (2015) Abdominal 4D flow MR imaging in a breath hold: combination of spiral sampling and dynamic compressed sensing for highly accelerated acquisition. Radiology 1:245–254 CrossRef Dyvorne H, Knight-Greenfield A, Jajamovich G et al (2015) Abdominal 4D flow MR imaging in a breath hold: combination of spiral sampling and dynamic compressed sensing for highly accelerated acquisition. Radiology 1:245–254 CrossRef
13.
Zurück zum Zitat Dyverfeldt P, Ebbers T (2017) Comparison of respiratory motion suppression techniques for 4D flow MRI. Magn Reson Med 5:1877–1882 CrossRef Dyverfeldt P, Ebbers T (2017) Comparison of respiratory motion suppression techniques for 4D flow MRI. Magn Reson Med 5:1877–1882 CrossRef
14.
Zurück zum Zitat Bomberna T, Koudehi GA, Claerebout C et al (2021) Transarterial drug delivery for liver cancer: numerical simulations and experimental validation of particle distribution in patient-specific livers. Expert Opin Drug Deliv 3:409–422 CrossRef Bomberna T, Koudehi GA, Claerebout C et al (2021) Transarterial drug delivery for liver cancer: numerical simulations and experimental validation of particle distribution in patient-specific livers. Expert Opin Drug Deliv 3:409–422 CrossRef
15.
Zurück zum Zitat Song R, Tipirneni A, Johnson P et al (2011) Evaluation of respiratory liver and kidney movements for MRI navigator gating. J Magn Reson Imaging 1:143–148 CrossRef Song R, Tipirneni A, Johnson P et al (2011) Evaluation of respiratory liver and kidney movements for MRI navigator gating. J Magn Reson Imaging 1:143–148 CrossRef
16.
Zurück zum Zitat Korin HW, Ehman RL, Riederer SJ et al (1992) Respiratory kinematics of the upper abdominal organs: a quantitative study. Magn Reson Med 1:172–178 CrossRef Korin HW, Ehman RL, Riederer SJ et al (1992) Respiratory kinematics of the upper abdominal organs: a quantitative study. Magn Reson Med 1:172–178 CrossRef
17.
Zurück zum Zitat Zoli M (1999) Total and functional hepatic blood flow decrease in parallel with ageing. Age Ageing 1:29–33 CrossRef Zoli M (1999) Total and functional hepatic blood flow decrease in parallel with ageing. Age Ageing 1:29–33 CrossRef
18.
Zurück zum Zitat Yzet T, Bouzerar R, Allart J-D et al (2010) Hepatic vascular flow measurements by phase contrast MRI and doppler echography: a comparative and reproducibility study. J Magn Reson Imaging 3:579–588 CrossRef Yzet T, Bouzerar R, Allart J-D et al (2010) Hepatic vascular flow measurements by phase contrast MRI and doppler echography: a comparative and reproducibility study. J Magn Reson Imaging 3:579–588 CrossRef
19.
Zurück zum Zitat McNaughton DA, Abu-Yousef MM (2011) Doppler US of the liver made simple. Radiographics 1:161–188 McNaughton DA, Abu-Yousef MM (2011) Doppler US of the liver made simple. Radiographics 1:161–188
20.
Zurück zum Zitat Ma LE, Markl M, Chow K et al (2019) Aortic 4D flow MRI in 2 minutes using compressed sensing, respiratory controlled adaptive k-space reordering, and inline reconstruction. Magn Reson Med 6:3675–3690 CrossRef Ma LE, Markl M, Chow K et al (2019) Aortic 4D flow MRI in 2 minutes using compressed sensing, respiratory controlled adaptive k-space reordering, and inline reconstruction. Magn Reson Med 6:3675–3690 CrossRef
21.
Zurück zum Zitat Markl M, Harloff A, Bley TA et al (2007) Time-resolved 3D MR velocity mapping at 3T: improved navigator-gated assessment of vascular anatomy and blood flow. J Magn Reson Imaging 4:824–831 CrossRef Markl M, Harloff A, Bley TA et al (2007) Time-resolved 3D MR velocity mapping at 3T: improved navigator-gated assessment of vascular anatomy and blood flow. J Magn Reson Imaging 4:824–831 CrossRef
22.
Zurück zum Zitat Hess AT, Bissell MM, Ntusi NAB et al (2015) Aortic 4D flow: quantification of signal-to-noise ratio as a function of field strength and contrast enhancement for 1.5T, 3T, and 7T. Magn Reson Med 5:1864–1871 CrossRef Hess AT, Bissell MM, Ntusi NAB et al (2015) Aortic 4D flow: quantification of signal-to-noise ratio as a function of field strength and contrast enhancement for 1.5T, 3T, and 7T. Magn Reson Med 5:1864–1871 CrossRef
23.
Zurück zum Zitat Heiberg E, Sjögren J, Ugander M et al (2010) Design and validation of Segment - freely available software for cardiovascular image analysis. BMC Med Imaging 1:1 CrossRef Heiberg E, Sjögren J, Ugander M et al (2010) Design and validation of Segment - freely available software for cardiovascular image analysis. BMC Med Imaging 1:1 CrossRef
24.
Zurück zum Zitat Martin Bland J, Altman D (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 8476:307–310 CrossRef Martin Bland J, Altman D (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 8476:307–310 CrossRef
25.
Zurück zum Zitat Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 1:159 CrossRef Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 1:159 CrossRef
26.
Zurück zum Zitat Tang C, Blatter DD, Parker DL (1993) Accuracy of phase-contrast flow measurements in the presence of partial-volume effects. J Magn Reson Imaging 2:377–385 CrossRef Tang C, Blatter DD, Parker DL (1993) Accuracy of phase-contrast flow measurements in the presence of partial-volume effects. J Magn Reson Imaging 2:377–385 CrossRef
27.
Zurück zum Zitat Bailes DR, Gilderdale DJ, Bydder GM et al (1985) Respiratory ordered phase encoding (ROPE): a method for reducing respiratory motion artefacts in MR imaging. J Comput Assist Tomogr 4:835–838 CrossRef Bailes DR, Gilderdale DJ, Bydder GM et al (1985) Respiratory ordered phase encoding (ROPE): a method for reducing respiratory motion artefacts in MR imaging. J Comput Assist Tomogr 4:835–838 CrossRef
28.
Zurück zum Zitat Mano Y, Takehara Y, Sakaguchi T et al (2013) Hemodynamic assessment of celiaco-mesenteric anastomosis in patients with pancreaticoduodenal artery aneurysm concomitant with celiac artery occlusion using flow-sensitive four-dimensional magnetic resonance imaging. Eur J Vasc Endovasc Surg 3:321–328 CrossRef Mano Y, Takehara Y, Sakaguchi T et al (2013) Hemodynamic assessment of celiaco-mesenteric anastomosis in patients with pancreaticoduodenal artery aneurysm concomitant with celiac artery occlusion using flow-sensitive four-dimensional magnetic resonance imaging. Eur J Vasc Endovasc Surg 3:321–328 CrossRef
29.
Zurück zum Zitat Haquin A, Sigovan M, Si-Mohamed S et al (2017) Phase-contrast MRI evaluation of haemodynamic changes induces by a coeliac axis stenosis in the gastroduodenal artery. Br J Radiol 1072:20160802 CrossRef Haquin A, Sigovan M, Si-Mohamed S et al (2017) Phase-contrast MRI evaluation of haemodynamic changes induces by a coeliac axis stenosis in the gastroduodenal artery. Br J Radiol 1072:20160802 CrossRef
30.
Zurück zum Zitat Bannas P, Roldán-Alzate A, Johnson KM et al (2016) Longitudinal monitoring of hepatic blood flow before and after TIPS by using 4D-flow MR imaging. Radiology 2:574–582 CrossRef Bannas P, Roldán-Alzate A, Johnson KM et al (2016) Longitudinal monitoring of hepatic blood flow before and after TIPS by using 4D-flow MR imaging. Radiology 2:574–582 CrossRef
31.
Zurück zum Zitat Stankovic Z, Rössle M, Euringer W et al (2015) Effect of TIPS placement on portal and splanchnic arterial blood flow in 4-dimensional flow MRI. Eur Radiol 9:2634–2640 CrossRef Stankovic Z, Rössle M, Euringer W et al (2015) Effect of TIPS placement on portal and splanchnic arterial blood flow in 4-dimensional flow MRI. Eur Radiol 9:2634–2640 CrossRef
32.
Zurück zum Zitat Roldán-Alzate A, Frydrychowicz A, Said A et al (2015) Impaired regulation of portal venous flow in response to a meal challenge as quantified by 4D flow MRI. J Magn Reson Imaging 4:1009–1017 CrossRef Roldán-Alzate A, Frydrychowicz A, Said A et al (2015) Impaired regulation of portal venous flow in response to a meal challenge as quantified by 4D flow MRI. J Magn Reson Imaging 4:1009–1017 CrossRef
33.
Zurück zum Zitat Lotz J, Meier C, Leppert A, Galanski M (2002) Cardiovascular flow measurement with phase-contrast MR imaging: basic facts and implementation. Radiographics 3:651–671 Lotz J, Meier C, Leppert A, Galanski M (2002) Cardiovascular flow measurement with phase-contrast MR imaging: basic facts and implementation. Radiographics 3:651–671
34.
Zurück zum Zitat David A, Le Touze D, Warin-Fresse K et al (2019) In-vitro validation of 4D flow MRI measurements with an experimental pulsatile flow model. Diagn Interv Imaging 1:17–23 CrossRef David A, Le Touze D, Warin-Fresse K et al (2019) In-vitro validation of 4D flow MRI measurements with an experimental pulsatile flow model. Diagn Interv Imaging 1:17–23 CrossRef
35.
Zurück zum Zitat Schnell S, Ansari SA, Wu C et al (2017) Accelerated dual-venc 4D flow MRI for neurovascular applications. J Magn Reson Imaging 1:102–114 CrossRef Schnell S, Ansari SA, Wu C et al (2017) Accelerated dual-venc 4D flow MRI for neurovascular applications. J Magn Reson Imaging 1:102–114 CrossRef
Metadaten
Titel
Assessment of hepatic arterial hemodynamics with 4D flow MRI: in vitro analysis of motion and spatial resolution related error and in vivo feasibility study in 20 volunteers
verfasst von
Ivan P. Dimov
Cyril Tous
Ning Li
Maxime Barat
Tim Bomberna
Charlotte Debbaut
Ning Jin
Gerald Moran
An Tang
Gilles Soulez
Publikationsdatum
22.06.2022
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 12/2022
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-022-08890-5

Weitere Artikel der Ausgabe 12/2022

European Radiology 12/2022 Zur Ausgabe

Neu im Fachgebiet Radiologie

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert – ganz bequem per eMail.