Kinematic 4D CT case-control study of wrist in dart throwing motion “in vivo”: comparison with other maneuvers

verfasst von: Sinan Orkut, Romain Gillet, Gabriela Hossu, Charles Lombard, Alain Blum, Lionel Athlani, Pedro Augusto Gondim Teixeira

Erschienen in: European Radiology | Ausgabe 11/2022

Einloggen, um Zugang zu erhalten

Abstract

Objectives

To compare the diagnostic performance of scapholunate gap (SLG) measurements acquired with dart throwing (DT), radio-ulnar deviation (RUD), and clenching fist (CF) maneuvers on 4D CT for the identification of scapholunate instability.

Methods

In this prospective study, 47 patients with suspected scapholunate interosseous ligament (SLIL) tears were evaluated from March 2015 to March 2020 with semiautomatic quantitative analysis on 4D CT. Five parameters (median, maximal value, range, and coefficient of variation) for SLG, lunocapitate angle (LCA), and radioscaphoid angle (RSA) obtained during DT maneuver were evaluated in patients with and without SLIL tears. CT arthrography was used as the gold standard for the SLIL status. The SLG values obtained were also compared with those obtained during CF and RUD maneuvers.

Results

Significant differences in all SLG- and LCA-derived parameters are found between patients with and without SLIL tears with DT (p < 0.003). The best diagnostic performance for the diagnosis of SLIL tears was obtained with median and maximal SLG values (sensitivity and specificity of 86–89% and 95%) and with maximal and range LCA values (sensitivity and specificity of 86% and 74%). No significant differences were observed for RSA values (p > 0.275). The SLG range obtained with DT maneuver was the only dynamic parameter statistically different between patients with partial and complete torn SLIL (p = 0.037).

Conclusion

4D CT of the wrist during DT showed a similar performance than RUD and a better performance than CF for the differentiation between patients with and without SLIL tears.

Key Points

• Four-dimensional computed tomography can dynamically assess scapholunate instability.
• The best results for differentiating between patients with and without SLIL tears were obtained with SLG median and maximal values.
• The dart throwing and radio-ulnar deviation maneuvers yielded the best results for the dynamic evaluation of scapholunate instability.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Kitay A, Wolfe SW (2012) Scapholunate instability: current concepts in diagnosis and management. J Hand Surg Am 37:21752175uCrossRef Kitay A, Wolfe SW (2012) Scapholunate instability: current concepts in diagnosis and management. J Hand Surg Am 37:21752175uCrossRef
2.
Zurück zum Zitat White NJ, Rollick NC (2015) Injuries of the scapholunate interosseous ligament: an update. J Am Acad Orthop Surg 23:691SurgCrossRef White NJ, Rollick NC (2015) Injuries of the scapholunate interosseous ligament: an update. J Am Acad Orthop Surg 23:691SurgCrossRef
3.
Zurück zum Zitat Halpenny D, Courtney K, Torreggiani WC (2012) Dynamic four-dimensional 320 section CT and carpal bone injury one injury pal bone injury injury ry urg1755175scapholunate instability. Clin Radiol 67:185CrossRef Halpenny D, Courtney K, Torreggiani WC (2012) Dynamic four-dimensional 320 section CT and carpal bone injury one injury pal bone injury injury ry urg1755175scapholunate instability. Clin Radiol 67:185CrossRef
4.
Zurück zum Zitat Pliefke J, Stengel D, Rademacher G, Mutze S, Ekkernkamp A, Eisenschenk A (2008) Diagnostic accuracy of plain radiographs and cineradiography in diagnosing traumatic scapholunate dissociation. Skeletal Radiol 37:139CrossRef Pliefke J, Stengel D, Rademacher G, Mutze S, Ekkernkamp A, Eisenschenk A (2008) Diagnostic accuracy of plain radiographs and cineradiography in diagnosing traumatic scapholunate dissociation. Skeletal Radiol 37:139CrossRef
5.
Zurück zum Zitat Cheriex KCAL, Sulkers GSI, Terra MP et al (2017) Scapholunate dissociation; diagnostics made easy. Eur J Radiol 92:45CrossRef Cheriex KCAL, Sulkers GSI, Terra MP et al (2017) Scapholunate dissociation; diagnostics made easy. Eur J Radiol 92:45CrossRef
6.
Zurück zum Zitat Kwon BC, Baek GH (2008) Fluoroscopic diagnosis of scapholunate interosseous ligament injuries in distal radius fractures. Clin Orthop Relat Res 466:969CrossRef Kwon BC, Baek GH (2008) Fluoroscopic diagnosis of scapholunate interosseous ligament injuries in distal radius fractures. Clin Orthop Relat Res 466:969CrossRef
7.
Zurück zum Zitat Lee YH, Choi YR, Kim S et al (2013) Intrinsic ligament and triangular fibrocartilage complex (TFCC) tears of the wrist: comparison of isovolumetric 3D-THRIVE sequence MR arthrography and conventional MR image at 3T. Magn Reson Imaging 31:221CrossRef Lee YH, Choi YR, Kim S et al (2013) Intrinsic ligament and triangular fibrocartilage complex (TFCC) tears of the wrist: comparison of isovolumetric 3D-THRIVE sequence MR arthrography and conventional MR image at 3T. Magn Reson Imaging 31:221CrossRef
8.
Zurück zum Zitat Bille B, Harley B, Cohen H (2007) A comparison of CT arthrography of the wrist to findings during wrist arthroscopy. J Hand Surg Am 32:834CrossRef Bille B, Harley B, Cohen H (2007) A comparison of CT arthrography of the wrist to findings during wrist arthroscopy. J Hand Surg Am 32:834CrossRef
9.
Zurück zum Zitat Linscheid RL, Dobyns JH, Beabout JW, Bryan RS (2002) Traumatic instability of the wrist: diagnosis, classification, and pathomechanics. J Bone Joint Surg Am 84-A:142CrossRef Linscheid RL, Dobyns JH, Beabout JW, Bryan RS (2002) Traumatic instability of the wrist: diagnosis, classification, and pathomechanics. J Bone Joint Surg Am 84-A:142CrossRef
10.
Zurück zum Zitat Lindau TR (2016) The role of arthroscopy in carpal instability. J Hand Surg Eur 41:35CrossRef Lindau TR (2016) The role of arthroscopy in carpal instability. J Hand Surg Eur 41:35CrossRef
11.
Zurück zum Zitat Gondim Teixeira PA, Formery A-S, Hossu G et al (2017) Evidence-based recommendations for musculoskeletal kinematic 4D-CT studies using wide area-detector scanners: a phantom study with cadaveric correlation. Eur Radiol 27:437CrossRef Gondim Teixeira PA, Formery A-S, Hossu G et al (2017) Evidence-based recommendations for musculoskeletal kinematic 4D-CT studies using wide area-detector scanners: a phantom study with cadaveric correlation. Eur Radiol 27:437CrossRef
12.
Zurück zum Zitat Garcia-Elias M, Alomar Serrallach X, Monill Serra J (2014) Dart-throwing motion in patients with scapholunate instability: a dynamic four-dimensional computed tomography study. J Hand Surg Eur 39:346CrossRef Garcia-Elias M, Alomar Serrallach X, Monill Serra J (2014) Dart-throwing motion in patients with scapholunate instability: a dynamic four-dimensional computed tomography study. J Hand Surg Eur 39:346CrossRef
13.
Zurück zum Zitat Leng S, Zhao K, Qu M, An K-N, Berger R, McCollough CH (2011) Dynamic CT technique for assessment of wrist joint instabilities. Med Phys 38:S50hysCrossRef Leng S, Zhao K, Qu M, An K-N, Berger R, McCollough CH (2011) Dynamic CT technique for assessment of wrist joint instabilities. Med Phys 38:S50hysCrossRef
14.
Zurück zum Zitat Demehri S, Hafezi-Nejad N, Morelli JN et al (2016) Scapholunate kinematics of asymptomatic wrists in comparison with symptomatic contralateral wrists using four-dimensional CT examinations: initial clinical experience. Skeletal Radiol 45:437CrossRef Demehri S, Hafezi-Nejad N, Morelli JN et al (2016) Scapholunate kinematics of asymptomatic wrists in comparison with symptomatic contralateral wrists using four-dimensional CT examinations: initial clinical experience. Skeletal Radiol 45:437CrossRef
15.
Zurück zum Zitat Abou Arab W, Rauch A, Chawki MB et al (2018) Scapholunate instability: improved detection with semi-automated kinematic CT analysis during stress maneuvers. Eur Radiol 28:4397CrossRef Abou Arab W, Rauch A, Chawki MB et al (2018) Scapholunate instability: improved detection with semi-automated kinematic CT analysis during stress maneuvers. Eur Radiol 28:4397CrossRef
16.
Zurück zum Zitat Rauch A, Arab WA, Dap F, Dautel G, Blum A, Gondim Teixeira PA (2018) Four-dimensional CT analysis of wrist kinematics during radioulnar deviation. Radiology 289:750–758CrossRef Rauch A, Arab WA, Dap F, Dautel G, Blum A, Gondim Teixeira PA (2018) Four-dimensional CT analysis of wrist kinematics during radioulnar deviation. Radiology 289:750–758CrossRef
17.
Zurück zum Zitat Carr R, MacLean S, Slavotinek J, Bain GI (2019) Four-dimensional computed tomography scanning for dynamic wrist disorders: prospective analysis and recommendations for clinical utility. J Wrist Surg 8:161aphyCrossRef Carr R, MacLean S, Slavotinek J, Bain GI (2019) Four-dimensional computed tomography scanning for dynamic wrist disorders: prospective analysis and recommendations for clinical utility. J Wrist Surg 8:161aphyCrossRef
18.
Zurück zum Zitat Kakar S, Breighner RE, Leng S et al (2016) The role of dynamic (4D) CT in the detection of scapholunate ligament injury. J Wrist Surg 5:306CrossRef Kakar S, Breighner RE, Leng S et al (2016) The role of dynamic (4D) CT in the detection of scapholunate ligament injury. J Wrist Surg 5:306CrossRef
19.
Zurück zum Zitat Athlani L, Rouizi K, Granero J et al (2020) Assessment of scapholunate instability with dynamic computed tomography. J Hand Surg Eur 45:375CrossRef Athlani L, Rouizi K, Granero J et al (2020) Assessment of scapholunate instability with dynamic computed tomography. J Hand Surg Eur 45:375CrossRef
20.
Zurück zum Zitat Schriever T, Olivecrona H, Wilcke M (2021) There is motion between the scaphoid and the lunate during the dart-throwing motion. J Plast Surg Hand Surg 55:294CrossRef Schriever T, Olivecrona H, Wilcke M (2021) There is motion between the scaphoid and the lunate during the dart-throwing motion. J Plast Surg Hand Surg 55:294CrossRef
21.
Zurück zum Zitat Kaufman-Cohen Y, Friedman J, Levanon Y, Jacobi G, Doron N, Portnoy S (2018) Wrist plane of motion and range during daily activities. Am J Occup Ther 72:7206205080p1–7206205080p10CrossRef Kaufman-Cohen Y, Friedman J, Levanon Y, Jacobi G, Doron N, Portnoy S (2018) Wrist plane of motion and range during daily activities. Am J Occup Ther 72:7206205080p1–7206205080p10CrossRef
22.
Zurück zum Zitat Gervaise A, Osemont B, Lecocq S et al (2012) CT image quality improvement using Adaptive Iterative Dose Reduction with wide-volume acquisition on 320-detector CT. Eur Radiol 22:295CrossRef Gervaise A, Osemont B, Lecocq S et al (2012) CT image quality improvement using Adaptive Iterative Dose Reduction with wide-volume acquisition on 320-detector CT. Eur Radiol 22:295CrossRef
23.
Zurück zum Zitat Rajan PV, Day CS (2015) Scapholunate interosseous ligament anatomy and biomechanics. J Hand Surg Am 40:1692CrossRef Rajan PV, Day CS (2015) Scapholunate interosseous ligament anatomy and biomechanics. J Hand Surg Am 40:1692CrossRef
24.
Zurück zum Zitat Dietrich TJ, Toms AP, Cerezal L et al (2021) Interdisciplinary consensus statements on imaging of scapholunate joint instability. Eur Radiol 31:94460-detCrossRef Dietrich TJ, Toms AP, Cerezal L et al (2021) Interdisciplinary consensus statements on imaging of scapholunate joint instability. Eur Radiol 31:94460-detCrossRef
25.
Zurück zum Zitat Ramamurthy NK, Chojnowski AJ, Toms AP (2016) Imaging in carpal instability. J Hand Surg Eur 41:22CrossRef Ramamurthy NK, Chojnowski AJ, Toms AP (2016) Imaging in carpal instability. J Hand Surg Eur 41:22CrossRef
26.
Zurück zum Zitat Brigstocke GHO, Hearnden A, Holt C, Whatling G (2014) In-vivo confirmation of the use of the dart thrower’s motion during activities of daily living. J Hand Surg Eur 39(4):373–378 Brigstocke GHO, Hearnden A, Holt C, Whatling G (2014) In-vivo confirmation of the use of the dart thrower’s motion during activities of daily living. J Hand Surg Eur 39(4):373–378
27.
Zurück zum Zitat Vardakastani V, Bell H, Mee S, Brigstocke G, Kedgley AE (2018) Clinical measurement of the dart throwing motion of the wrist: variability, accuracy and correction. J Hand Surg Eur 43:723CrossRef Vardakastani V, Bell H, Mee S, Brigstocke G, Kedgley AE (2018) Clinical measurement of the dart throwing motion of the wrist: variability, accuracy and correction. J Hand Surg Eur 43:723CrossRef
28.
Zurück zum Zitat Gondim Teixeira PA, Formery A-S, Balazuc G et al (2019) Comparison between subtalar joint quantitative kinematic 4-D CT parameters in healthy volunteers and patients with joint stiffness or chronic ankle instability: a preliminary study. Eur J Radiol 114:76CrossRef Gondim Teixeira PA, Formery A-S, Balazuc G et al (2019) Comparison between subtalar joint quantitative kinematic 4-D CT parameters in healthy volunteers and patients with joint stiffness or chronic ankle instability: a preliminary study. Eur J Radiol 114:76CrossRef
Metadaten
Titel
Kinematic 4D CT case-control study of wrist in dart throwing motion “in vivo”: comparison with other maneuvers
verfasst von
Sinan Orkut
Romain Gillet
Gabriela Hossu
Charles Lombard
Alain Blum
Lionel Athlani
Pedro Augusto Gondim Teixeira
Publikationsdatum
21.04.2022
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 11/2022
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-022-08746-y

Weitere Artikel der Ausgabe 11/2022

European Radiology 11/2022 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.