Serum Sirolimus Levels After Implantation of Third Generation Drug Eluting Cobalt Chromium Coronary Stent in Ductus Arteriosus in Neonates with Duct-Dependent Pulmonary Circulation

verfasst von: Kothandam Sivakumar, Sreeja Pavithran, Bhushan Sonawane, Monica Rajendran, Rajeshkumar Ramasamy

Erschienen in: Pediatric Cardiology | Ausgabe 7/2020

Einloggen, um Zugang zu erhalten

Abstract

Ductal stenting (DS) palliates duct-dependent lesions using coronary stents. Sirolimus-eluting stents have replaced bare-metal stents in coronary interventions. Concerns exist about sirolimus levels in neonates. Therapeutic immunosuppressive sirolimus level is 5–15 ng/ml. After neonatal DS, drug levels were assessed at 24 h, 7 days and monthly thereafter till they were undetectable. Clinical course, ductal patency till their final corrective surgery was analyzed. The exact quantity of sirolimus in each stent was known. Twelve neonates with median age of 5.5 days received sirolimus-eluting stents, one stent in nine and two in the rest. The lesions were pulmonary atresia intact ventricular septum(PAIVS) in four, univentricular lesions with pulmonary atresia in four, biventricular lesions with pulmonary atresia in three and right ventricular rhabdomyoma in one neonate. If single stents up to 22 mm length, 24-h drug levels were less than 5 ng/ml. Even though 24-h levels were above 5 ng/ml in patients with single longer stent or two stents, it reduced to very low levels by seventh day. Two hospital deaths included rhabdomyoma with complete heart block and post-valvotomy cardiac failure for PAIVS. Stent patency after valvotomy for PAIVS exceeded three years. Patency was retained for 8–27 months till their elective corrective surgery in others. Sirolimus levels were acceptable at 24 h in all neonates receiving single stent under 22 mm length. In patients needing two stents, drug levels were in immunosuppressive range at 24 h but reduced rapidly within 7 days. The palliation provided by sirolimus-eluting DS was sufficiently long to provide clinical benefit.
Literatur
1.
Zurück zum Zitat Peters B, Ewert P, Berger F (2009) The role of stents in the treatment of congenital heart disease: current status and future perspectives. Ann Pediatr Cardiol 2:3–23CrossRefPubMed Peters B, Ewert P, Berger F (2009) The role of stents in the treatment of congenital heart disease: current status and future perspectives. Ann Pediatr Cardiol 2:3–23CrossRefPubMed
2.
Zurück zum Zitat Wells WJ, Yu RJ, Batra AS, Monforte H, Sintek C, Starnes VA (2005) Obstruction in modified Blalock shunts: a quantitative analysis with clinical correlation. Ann Thorac Surg 79:2072–2076CrossRef Wells WJ, Yu RJ, Batra AS, Monforte H, Sintek C, Starnes VA (2005) Obstruction in modified Blalock shunts: a quantitative analysis with clinical correlation. Ann Thorac Surg 79:2072–2076CrossRef
3.
Zurück zum Zitat Coe JY, Olley PM (1991) A novel method to maintain ductus arteriosus patency. J Am Coll Cardiol 18:837–841CrossRef Coe JY, Olley PM (1991) A novel method to maintain ductus arteriosus patency. J Am Coll Cardiol 18:837–841CrossRef
4.
Zurück zum Zitat Gibbs JL, Uzun O, Blackburn ME, Wren C, Hamilton JR, Watterson KG (1999) Fate of the stented arterial duct. Circulation 99:2621–2625CrossRef Gibbs JL, Uzun O, Blackburn ME, Wren C, Hamilton JR, Watterson KG (1999) Fate of the stented arterial duct. Circulation 99:2621–2625CrossRef
5.
Zurück zum Zitat Sivakumar K, Bhagyavathy A, Coelho R, Satish R, Krishnan P (2012) Longevity of neonatal ductal stenting for congenital heart diseases with duct dependent pulmonary circulation. Congenit Heart Dis 7:526–533CrossRef Sivakumar K, Bhagyavathy A, Coelho R, Satish R, Krishnan P (2012) Longevity of neonatal ductal stenting for congenital heart diseases with duct dependent pulmonary circulation. Congenit Heart Dis 7:526–533CrossRef
6.
Zurück zum Zitat Costa MA, Simon DI (2005) Molecular basis of restenosis and drug-eluting stents. Circulation 111:2257–2273CrossRef Costa MA, Simon DI (2005) Molecular basis of restenosis and drug-eluting stents. Circulation 111:2257–2273CrossRef
7.
Zurück zum Zitat Babapulle MN, Eisenberg MJ (2002) Coated stents for the prevention of restenosis: part 1. Circulation 106:2734–2740CrossRef Babapulle MN, Eisenberg MJ (2002) Coated stents for the prevention of restenosis: part 1. Circulation 106:2734–2740CrossRef
8.
Zurück zum Zitat Lee KJ, Hinek A, Chaturvedi RR, Almeida CL, Honjo O, Koren G, Benson LN (2009) Rapamycin-eluting stents in the arterial duct: experimental observations in the pig model. Circulation 119:2078–2085CrossRef Lee KJ, Hinek A, Chaturvedi RR, Almeida CL, Honjo O, Koren G, Benson LN (2009) Rapamycin-eluting stents in the arterial duct: experimental observations in the pig model. Circulation 119:2078–2085CrossRef
9.
Zurück zum Zitat Mueller GC, Dodge-Khatami A, Weil J (2010) First experience with a new drug eluting balloon for the treatment of congenital pulmonary vein stenosis in a neonate. Cardiol Young 20:455–458CrossRef Mueller GC, Dodge-Khatami A, Weil J (2010) First experience with a new drug eluting balloon for the treatment of congenital pulmonary vein stenosis in a neonate. Cardiol Young 20:455–458CrossRef
10.
Zurück zum Zitat Lobach NE, Pollock-Barziv SM, West LJ, Dipchand AI (2005) Sirolimus immunosuppression in pediatric heart transplant recipients: a single-center experience. J Heart Lung Transplant 24:184–189CrossRef Lobach NE, Pollock-Barziv SM, West LJ, Dipchand AI (2005) Sirolimus immunosuppression in pediatric heart transplant recipients: a single-center experience. J Heart Lung Transplant 24:184–189CrossRef
11.
Zurück zum Zitat Riede FT, Schneider P, Dahnert I (2007) Transient sirolimus levels after implantation of a sirolimus eluting stent in an infant. Clin Res Cardiol 96:508–510CrossRef Riede FT, Schneider P, Dahnert I (2007) Transient sirolimus levels after implantation of a sirolimus eluting stent in an infant. Clin Res Cardiol 96:508–510CrossRef
12.
Zurück zum Zitat Li SS, Cheng BC, Lee SH (2005) Giant coronary aneurysm formation after sirolimus-eluting stent implantation in Kawasaki disease. Circulation 112:e105–e107CrossRef Li SS, Cheng BC, Lee SH (2005) Giant coronary aneurysm formation after sirolimus-eluting stent implantation in Kawasaki disease. Circulation 112:e105–e107CrossRef
13.
Zurück zum Zitat Rashid A, Ringewald J, Saucedo J (2005) Placement of a sirolimus coated stent in the distal left internal mammary artery of an 8-year-old boy. J Invasive Cardiol 17:329–330 Rashid A, Ringewald J, Saucedo J (2005) Placement of a sirolimus coated stent in the distal left internal mammary artery of an 8-year-old boy. J Invasive Cardiol 17:329–330
14.
Zurück zum Zitat Lee KJ, Seto W, Benson L, Chaturvedi RR (2015) Pharmacokinetics of sirolimus eluting stents implanted in neonatal arterial duct. Circ Cardiovasc Interv 8:e002233 Lee KJ, Seto W, Benson L, Chaturvedi RR (2015) Pharmacokinetics of sirolimus eluting stents implanted in neonatal arterial duct. Circ Cardiovasc Interv 8:e002233
15.
Zurück zum Zitat Akin I, Schneider H, Ince H, Kische S, Rehders TC, Chatterjee T, Nienabar CA (2011) Second and third generation drug eluting coronary stents: progress and safety. Herz 36:190–196CrossRef Akin I, Schneider H, Ince H, Kische S, Rehders TC, Chatterjee T, Nienabar CA (2011) Second and third generation drug eluting coronary stents: progress and safety. Herz 36:190–196CrossRef
16.
Zurück zum Zitat Dani S, Costa RA, Joshi H, Shah J, Pandya R, Virmani R et al (2013) First in human evaluation of the novel Biomime sirolimus eluting coronary stent with bioabsorbable polymer for the treatment of single denovo lesions located in native coronary vessels—results from the Merit-1 trial. Eurointervention 9:493–500CrossRef Dani S, Costa RA, Joshi H, Shah J, Pandya R, Virmani R et al (2013) First in human evaluation of the novel Biomime sirolimus eluting coronary stent with bioabsorbable polymer for the treatment of single denovo lesions located in native coronary vessels—results from the Merit-1 trial. Eurointervention 9:493–500CrossRef
17.
Zurück zum Zitat Alwi M, Choo KK, Latiff HA, Kandavello G, Samion H, Mulyadi MD (2004) Initial results and medium-term follow-up of stent implantation of patent ductus arteriosus in duct-dependent pulmonary circulation. J Am Coll Cardiol 44:438–445CrossRef Alwi M, Choo KK, Latiff HA, Kandavello G, Samion H, Mulyadi MD (2004) Initial results and medium-term follow-up of stent implantation of patent ductus arteriosus in duct-dependent pulmonary circulation. J Am Coll Cardiol 44:438–445CrossRef
18.
Zurück zum Zitat Boshoff DE, Michel-Behnke I, Schranz D, Gewillig M (2007) Stenting the neonatal arterial duct. Expert Rev Cardiovasc Ther 5:893–901CrossRef Boshoff DE, Michel-Behnke I, Schranz D, Gewillig M (2007) Stenting the neonatal arterial duct. Expert Rev Cardiovasc Ther 5:893–901CrossRef
19.
Zurück zum Zitat Schneider M, Zartner P, Sidiropoulos A, Konertz W, Hausdorf G (1998) Stent implantation of the arterial duct in newborns with duct-dependent circulation. Eur Heart J 19:1401–1409CrossRef Schneider M, Zartner P, Sidiropoulos A, Konertz W, Hausdorf G (1998) Stent implantation of the arterial duct in newborns with duct-dependent circulation. Eur Heart J 19:1401–1409CrossRef
20.
Zurück zum Zitat Benson L, Arsdell GV (2018) Comparisons between ductal stenting and Blalock-Taussig shunts for infants with ductal-dependent pulmonary circulation. Circulation 137:602–604CrossRef Benson L, Arsdell GV (2018) Comparisons between ductal stenting and Blalock-Taussig shunts for infants with ductal-dependent pulmonary circulation. Circulation 137:602–604CrossRef
21.
Zurück zum Zitat Bentham JR, Zava NK, Harrison WJ, Shauq A, Kalantre A, Derrick G et al (2018) Duct stenting versus modified Blalock-Taussig shunt in neonates with duct dependent pulmonary blood flow. Circulation 137:581–588CrossRef Bentham JR, Zava NK, Harrison WJ, Shauq A, Kalantre A, Derrick G et al (2018) Duct stenting versus modified Blalock-Taussig shunt in neonates with duct dependent pulmonary blood flow. Circulation 137:581–588CrossRef
22.
Zurück zum Zitat Glatz AC, Petit CJ, Goldstein BH, Kelleman MS, McCracken CE, McDonnell A, Buckey T et al (2018) Comparison between patent ductus arteriosus stent and modified Blalock-Taussig shunt as palliation for infants with ductal-dependent pulmonary blood flow. Circulation 137:589–601CrossRef Glatz AC, Petit CJ, Goldstein BH, Kelleman MS, McCracken CE, McDonnell A, Buckey T et al (2018) Comparison between patent ductus arteriosus stent and modified Blalock-Taussig shunt as palliation for infants with ductal-dependent pulmonary blood flow. Circulation 137:589–601CrossRef
23.
Zurück zum Zitat Feltes TF, Bacha E, Beekman RH III, Cheatham JP, Feinstein JA, Gomes AS et al (2011) Indications for cardiac catheterization and intervention in pediatric cardiac disease. Circulation 123:2607–2652CrossRef Feltes TF, Bacha E, Beekman RH III, Cheatham JP, Feinstein JA, Gomes AS et al (2011) Indications for cardiac catheterization and intervention in pediatric cardiac disease. Circulation 123:2607–2652CrossRef
24.
Zurück zum Zitat Aggarwal V, Dhillon GS, Penny DJ, Gowda ST, Qureshi AM (2019) Drug-eluting stents compared to bare metal stents for stenting the ductus arteriosus in infants with ductal-dependent pulmonary blood flow. Am J Cardiol 124:952–959CrossRef Aggarwal V, Dhillon GS, Penny DJ, Gowda ST, Qureshi AM (2019) Drug-eluting stents compared to bare metal stents for stenting the ductus arteriosus in infants with ductal-dependent pulmonary blood flow. Am J Cardiol 124:952–959CrossRef
25.
Zurück zum Zitat Abizaid A, Costa JR Jr (2010) New drug-eluting stents. An overview on biodegradable and polymer-free next-generation stent systems. Circ Cardiovasc Interv 3:384–393CrossRef Abizaid A, Costa JR Jr (2010) New drug-eluting stents. An overview on biodegradable and polymer-free next-generation stent systems. Circ Cardiovasc Interv 3:384–393CrossRef
26.
Zurück zum Zitat Meier-Kriesche HU, Kaplan B (2000) Toxicity and efficacy of sirolimus: relationship to whole blood concentration. Clin Ther 22(Suppl B):B93–B100CrossRef Meier-Kriesche HU, Kaplan B (2000) Toxicity and efficacy of sirolimus: relationship to whole blood concentration. Clin Ther 22(Suppl B):B93–B100CrossRef
27.
Zurück zum Zitat Qureshi AM, Goldstein BH, Glatz AC, Agrawal H, Aggarwal V, Ligon RA et al (2019) Classification scheme for ductal morphology in cyanotic patients with ductal dependent pulmonary blood flow and association with outcomes of patent ductus arteriosus stenting. Catheter Cardiovasc Interv 93:933–943CrossRef Qureshi AM, Goldstein BH, Glatz AC, Agrawal H, Aggarwal V, Ligon RA et al (2019) Classification scheme for ductal morphology in cyanotic patients with ductal dependent pulmonary blood flow and association with outcomes of patent ductus arteriosus stenting. Catheter Cardiovasc Interv 93:933–943CrossRef
28.
Zurück zum Zitat Fenton KN, Siewers RD, Rebovich B, Pigula FA (2003) Interim mortality in infants with systemic-to-pulmonary artery shunts. Ann Thorac Surg 76:152–156CrossRef Fenton KN, Siewers RD, Rebovich B, Pigula FA (2003) Interim mortality in infants with systemic-to-pulmonary artery shunts. Ann Thorac Surg 76:152–156CrossRef
Metadaten
Titel
Serum Sirolimus Levels After Implantation of Third Generation Drug Eluting Cobalt Chromium Coronary Stent in Ductus Arteriosus in Neonates with Duct-Dependent Pulmonary Circulation
verfasst von
Kothandam Sivakumar
Sreeja Pavithran
Bhushan Sonawane
Monica Rajendran
Rajeshkumar Ramasamy
Publikationsdatum
30.05.2020
Verlag
Springer US
Erschienen in
Pediatric Cardiology / Ausgabe 7/2020
Print ISSN: 0172-0643
Elektronische ISSN: 1432-1971
DOI
https://doi.org/10.1007/s00246-020-02381-4

Weitere Artikel der Ausgabe 7/2020

Pediatric Cardiology 7/2020 Zur Ausgabe

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.