Skip to main content
Erschienen in: Journal of Medical Case Reports 1/2018

Open Access 01.12.2018 | Case report

Sudden death after valve-in-valve procedure due to delayed coronary obstruction: a case report

verfasst von: Angelo Buscaglia, Giacomo Tini, Gian Paolo Bezante, Claudio Brunelli, Manrico Balbi

Erschienen in: Journal of Medical Case Reports | Ausgabe 1/2018

Abstract

Background

Valve-in-valve transcatheter aortic valve implantation for degenerated aortic bioprostheses is an effective option for patients at high risk for redo surgery, even if it may be burdened by complications more common in specific settings, such as, coronary artery obstruction.

Case presentation

We present the case of a Caucasic 84-year-old woman with degeneration of a previously implanted aortic Mitroflow bioprosthesis. She underwent a valve-in-valve transcatheter aortic valve implantation with a CoreValve® bioprosthesis. End-procedure coronary angiography demonstrated maintained perfusion of both coronary arteries. However, few hours later, she experienced sudden cardiac death. An autopsy showed that Mitroflow prosthesis leaflets were higher than the left main coronary ostium, and no other possible cause for the sudden death. Fatality was thus ascribed to left main coronary ostium obstruction due to apposition of the Mitroflow leaflet pushed upward by the late expansion of CoreValve®.

Conclusions

Coronary artery obstruction is a frequently fatal complication which usually presents just after valve implantation, but, as reported in our case, it may also have a delayed presentation. Accurate patient’s selection and intraoperative preventive measures can reduce this eventuality.
Hinweise

Electronic supplementary material

The online version of this article (https://​doi.​org/​10.​1186/​s13256-018-1785-x) contains supplementary material, which is available to authorized users.
Angelo Buscaglia and Giacomo Tini contributed equally to this work.
Abkürzungen
CT
Computed tomography
LMCO
Left main coronary ostium
PCI
Percutaneous coronary intervention
TAVI
Transcatheter aortic valve implantation
TEE
Transesophageal echocardiogram
TTE
Transthoracic echocardiography
VIV
Valve-in-valve

Background

In recent years, valve-in-valve (VIV) transcatheter aortic valve implantation (TAVI) procedures for correction of degenerated aortic bioprostheses appeared to be a feasible and satisfactory option for patients deemed at high risk for redo surgery [1]. However, the efficacy of this method has been reported to be lower in specific settings. In particular, there are concerns about higher residual transvalvular gradients and rate of coronary artery obstruction [2].
We present a case of sudden death which occurred few hours after a successful VIV intervention. The only possible explanation for the event, as demonstrated by autopsy, appeared to be left main coronary ostium (LMCO) obstruction caused by the previously surgically implanted aortic valve’s leaflets, pushed upward by the expansion of the CoreValve® used for the VIV procedure.

Case presentation

We report the case of a retired Caucasic 84-year-old woman who required a VIV procedure due to the degeneration of a previously implanted aortic bioprosthesis. Her cardiologic history started in 2006 when she experienced syncope and was then diagnosed as having severe aortic stenosis (mean transvalvular gradient 44 mmHg) and severe mitral regurgitation. She reported no previous clinical events. She underwent aortic valve replacement with a Mitroflow number 21 bioprosthesis and a Carpentier-Edwards Physio mitral annuloplasty ring implantation. After surgery, she suffered from brady-tachy syndrome and needed a pacemaker implantation.
The initial signs of prosthesis degeneration were found at a routine transthoracic echocardiography (TTE) in 2011, with a transvalvular mean gradient of 26 mmHg. However, she was asymptomatic and meanwhile she was diagnosed as having an indolent myeloma, thus a conservative approach was chosen.
In January 2014, she started complaining of epigastric discomfort and dyspnea for minimal exertion. A TTE was repeated and showed a further increase of the mean transvalvular gradient (35 mmHg) and occurrence of moderate paraprosthetic regurgitation due to detachment of the anterior edge of the aortic prosthesis ring.
In February 2014 she was admitted to our department for an episode of pulmonary edema with angina. A physical examination revealed bilateral crackles, 3/6 systolic ejection murmur, and leg swelling. Neurological evaluation was normal. Her electrocardiogram showed transient diffuse ST segment depression and troponin values were slightly elevated (peak, 0.08 ng/ml; reference values, < 0.015 ng/ml). In addition, laboratory tests showed a mild anemia (hemoglobin values, 11 g/dl; reference values, 13–17 g/dl) and a stage 3 chronic kidney disease (serum creatinine, 1.1 mg/dl; reference values, 0.5–1 mg/dl; and glomerular filtration rate, 47 mg/dl). Her hepatic function was normal, as well as white blood cells and platelets count (white blood cells, 6500/ml; reference values, 4500–9800; and platelets count, 300 000/ml; reference values, 150,000–450,000).
A transesophageal echocardiogram (TEE) showed hypomobility of the non-coronary cusp, moderate paraprosthetic regurgitation, and severe intraprosthetic regurgitation. Due to the frailty of our patient, in consideration of age, previous cardiac surgery, and concurrent hematologic disease, after a Heart Team discussion, a VIV TAVI was proposed. Pre-procedural investigations included a coronary angiogram, showing absence of coronary artery disease, and a computed tomography (CT) angiography to calculate the diameter of the Mitroflow prosthesis (17 mm) and the LMCO height from the valvular annulus (11.6 mm).
VIV implantation was performed via right femoral artery using a CoreValve® prosthesis number 23 (Additional file 1). A 6 Fr guide catheter, via right omeral artery, was used for the left main coronary artery cannulation as protection (Figs. 1 and 2).
Post-procedural angiography confirmed maintained coronary perfusion (Fig. 3), even after the removal of the guide wire. It also showed correct position of the prosthesis and its normal functioning with a transvalvular gradient of 12 mmHg and a mild intraprosthetic regurgitation (Fig. 4).
Our patient was hemodynamically stable and was transferred to the cardiothoracic intensive care unit. Two hours after the end of the procedure, she experienced a sudden cardiac arrest with asystole and electromechanical dissociation. Resuscitation maneuvers were ineffective.
An autopsy was performed to investigate the cause of death. The CoreValve® prosthesis was removed from the aortic root with no signs of damage or thrombus formation (Fig. 5). The underlying Mitroflow valve appeared free of calcium or thrombi, but its leaflets appeared higher than the LMCO (Fig. 6). No other possible cause for the sudden cardiac death could be found. Our hypothesis was: a delayed occlusion of the LMCO by the Mitroflow leaflets, pushed upward by the late expansion of the CoreValve® prosthesis.

Discussion

Here we described a case of VIV TAVI complicated by coronary obstruction and sudden cardiac death. Even if the procedure is performed with coronary protection and perfusion is maintained at the end of the intervention, our case shows that this severe adverse event may also occur hours after the end of the TAVI, due to late expansion of the prosthesis.
Use of bioprostheses rather than mechanical valves in surgical aortic valve replacement has increased in recent years [3]. Bioprostheses are characterized by progressive leaflet degeneration, which may ultimately result in valve stenosis, regurgitation, or both, possibly leading to symptomatic heart failure [4, 5]. In the case of bioprosthesis degeneration, even if reoperation is considered the standard treatment, patients are frequently at high risk for a second conventional surgical intervention [6, 7]. VIV TAVI inside the previously surgically implanted failed aortic bioprosthesis has become a reliable alternative to repeat surgery, because it is less invasive and has a high rate of procedural success [1]. Moreover, VIV procedures have lower rates of intervalvular leaks and lower rates of atrioventricular blocks, due to the protection given by the sewing ring of the previously implanted prosthesis that renders a compression of the conduction system less probable [1].
However, the efficacy of VIV procedures has been reported to be lower in specific settings [8, 9]. Worse outcomes are in fact reported for patients with small surgical valves and in those with a stenotic presentation [8]. The poor distensibility of the bioprosthetic ring accounts for elevated residual transaortic mean gradients after VIV TAVI [2]. SAPIEN valves are characterized by annular leaflets and a relation is observed between postoperative gradients after VIV procedure and the size of surgical valve [1]. The same relation is not seen with CoreValve®, due to its supra-annular leaflets that provide a larger orifice [1].
VIV interventions may also be complicated by the life-threatening occurrence of coronary obstruction. The frequency of this severe event is reported to be higher in VIV than in native valve procedures, up to 3.5% versus 0.7% of cases [2]. In a recent series by Jabbour et al., coronary obstructions in TAVI were found to be significantly more common after VIV procedures [9]. Notably, in the study, delayed coronary obstructions were investigated, distinguishing them as early (occurring < 24 hours to 7 days after the procedure) and late delayed (> 7 days). VIV was a risk factor for both.
Coronary obstructions are usually secondary to the displacement of the leaflet of the underlying valve toward a coronary ostium, most commonly the left one [912], as in the case presented.
In addition to the anatomic features known to possibly predispose to coronary obstruction in the case of native valve procedures, namely shallow sinus of Valsalva or low-lying coronary ostium [9, 10], in VIV the risk is also dependent on the characteristics of the previously implanted surgical bioprosthesis and on the relationship of its leaflets and posts with the coronary ostia [2].
Coronary obstructions are reported to be more common with stenotic than with regurgitant bioprostheses, supra-annular or internally stented bioprostheses, bulky bioprosthetic leaflets and reimplanted coronary arteries [12]. Internally stented bioprostheses, such as Mitroflow, have relatively long leaflets that may extend outward in a tubular fashion after a VIV implantation [12]. Stentless bioprostheses are usually implanted in a supra-annular position, thus the valve leaflets are closer to the coronary ostia [13].
If there is suspicion of coronary obstruction an immediate angiography must be carried out, and if diagnosis is confirmed, percutaneous coronary intervention (PCI) should be considered or in the case of PCI failure, emergent coronary artery bypass should be considered [12]. In a patient with high risk for coronary obstruction additional preventive measures are advised, such as coronary protection with a guidewire, leaving an undeployed stent in the distal portion of the coronary ready to be pulled back and implanted in the ostium if needed [10, 11]. In selected cases, a preventive stent implantation with chimney technique, even in a case of maintained perfusion of the coronary, may be reasonable. In our center, after the case described here, this technique was successfully performed in a subsequent case of a VIV TAVI on a previous Mitroflow number 21 prosthesis dysfunction. In addition, the use of a retrievable valve may permit rapid prosthesis removal in the case of coronary obstruction evidence [12]. The selection of a smaller sized valve provides less displacement of bioprosthesis leaflets than the low implantation does [1].
Coronary obstructions after TAVI may be acute or delayed, as in the case presented (early delayed, according to Jabbour et al. [9]). No other cause for the sudden death experienced by our patient except for LMCO obstruction was demonstrated by an autoptic examination. It must, in fact, be considered that an implanted bioprosthesis may reach full expansion only in the first hours after the completion of the VIV procedure; thus, displacement of the underlying valve’s leaflets may not be immediate [9, 13]. This mechanism is particularly true for CoreValve® prostheses [13]. Notably, the bioprosthesis our patient received in 2006 was a Mitroflow number 21. Other authors have previously underlined the need to cautiously execute or even to avoid VIV procedures on Mitroflow bioprostheses [14].
A delayed occurrence of coronary obstruction renders this complication, already burdened by an extremely high mortality, even more difficult to manage. Therefore, an adequate and careful assessment of patients who receive indication for a VIV procedure is of paramount importance. Even though rare, in fact, some complications are sudden and fatal. Since VIV procedures have shown satisfactory results, the best way to limit unsuccessfulness is prevention of complications by accurate pre-selection of patients.

Conclusions

VIV TAVI for previously implanted degenerated aortic bioprostheses is a satisfactory alternative method to the repeating of surgery, it is less invasive and has a high rate of procedural success. However, coronary obstruction, a life-threatening complication, seems to occur more frequently in VIV than in native valve TAVI.
This complication usually occurs at the time of implantation, but, as reported in our case, it may also have a delayed presentation. This highlights the importance of adequate preoperative selection of patients as a means of prevention.

Availability of data and materials

Data sharing is not applicable to this article as no datasets were generated or analyzed during the current study.
Approval of the institutional ethics committee was not required, because this is a case report without any experimental trial.
Written informed consent was obtained from the patient’s next-of-kin for publication of this case report and any accompanying images. A copy of the written consent is available for review by the Editor-in-Chief of this journal.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Allgemeinmedizin

Kombi-Abonnement

Mit e.Med Allgemeinmedizin erhalten Sie Zugang zu allen CME-Fortbildungen und Premium-Inhalten der allgemeinmedizinischen Zeitschriften, inklusive einer gedruckten Allgemeinmedizin-Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Webb JG, Dvir D. Transcatheter aortic valve replacement for bioprosthetic aortic valve failure: the valve-in-valve procedure. Circulation. 2013;127(25):2542–50.CrossRefPubMed Webb JG, Dvir D. Transcatheter aortic valve replacement for bioprosthetic aortic valve failure: the valve-in-valve procedure. Circulation. 2013;127(25):2542–50.CrossRefPubMed
2.
Zurück zum Zitat Dvir D, Webb J, Brecker S, Bleiziffer S, Hildick-Smith D, Colombo A, Descoutures F, Hengstenberg C, Moat NE, Bekeredjian R, Napodano M, Testa L, Lefevre T, Guetta V, Nissen H, Hernández JM, Roy D, Teles RC, Segev A, Dumonteil N, Fiorina C, Gotzmann M, Tchetche D, Abdel-Wahab M, De Marco F, Baumbach A, Laborde JC, Kornowski R. Transcatheter aortic valve replacement for degenerative bioprosthetic surgical valves: results from the global valve-in-valve registry. Circulation. 2012;126:2335–44.CrossRefPubMed Dvir D, Webb J, Brecker S, Bleiziffer S, Hildick-Smith D, Colombo A, Descoutures F, Hengstenberg C, Moat NE, Bekeredjian R, Napodano M, Testa L, Lefevre T, Guetta V, Nissen H, Hernández JM, Roy D, Teles RC, Segev A, Dumonteil N, Fiorina C, Gotzmann M, Tchetche D, Abdel-Wahab M, De Marco F, Baumbach A, Laborde JC, Kornowski R. Transcatheter aortic valve replacement for degenerative bioprosthetic surgical valves: results from the global valve-in-valve registry. Circulation. 2012;126:2335–44.CrossRefPubMed
3.
Zurück zum Zitat Brown JM, O’Brien SM, Wu C, Sikora JA, Griffith BP, Gammie JS. Isolated aortic valve replacement in North America comprising 108,687 patients in 10 years: changes in risks, valve types, and outcomes in the Society of Thoracic Surgeons National Database. J Thorac Cardiovasc Surg. 2009;137(1):82–90.CrossRefPubMed Brown JM, O’Brien SM, Wu C, Sikora JA, Griffith BP, Gammie JS. Isolated aortic valve replacement in North America comprising 108,687 patients in 10 years: changes in risks, valve types, and outcomes in the Society of Thoracic Surgeons National Database. J Thorac Cardiovasc Surg. 2009;137(1):82–90.CrossRefPubMed
4.
Zurück zum Zitat Brennan JM, Edwards FH, Zhao Y, O’Brien S, Booth ME, Dokholyan RS, Douglas PS, Peterson ED, DEcIDE AVR (Developing Evidence to Inform Decisions about Effectiveness–Aortic Valve Replacement) Research Team. Long-term safety and effectiveness of mechanical vs biologic aortic valve prostheses in older patients: results from the Society of Thoracic Surgeons Adult Cardiac Surgery National Database. Circulation. 2013;127(16):1647–55.CrossRefPubMed Brennan JM, Edwards FH, Zhao Y, O’Brien S, Booth ME, Dokholyan RS, Douglas PS, Peterson ED, DEcIDE AVR (Developing Evidence to Inform Decisions about Effectiveness–Aortic Valve Replacement) Research Team. Long-term safety and effectiveness of mechanical vs biologic aortic valve prostheses in older patients: results from the Society of Thoracic Surgeons Adult Cardiac Surgery National Database. Circulation. 2013;127(16):1647–55.CrossRefPubMed
5.
Zurück zum Zitat David TE, Ivanov J, Armstrong S, Feindel CM, Cohen G. Late results of heart valve replacement with the Hancock II bioprosthesis. J Thorac Cardiovasc Surg. 2001;121(2):268–77.CrossRefPubMed David TE, Ivanov J, Armstrong S, Feindel CM, Cohen G. Late results of heart valve replacement with the Hancock II bioprosthesis. J Thorac Cardiovasc Surg. 2001;121(2):268–77.CrossRefPubMed
6.
Zurück zum Zitat Maganti M, Rao V, Armstrong S, Feindel CM, Scully HE, David TE. Redo valvular surgery in elderly patients. Ann Thorac Surg. 2009;87(2):521–5.CrossRefPubMed Maganti M, Rao V, Armstrong S, Feindel CM, Scully HE, David TE. Redo valvular surgery in elderly patients. Ann Thorac Surg. 2009;87(2):521–5.CrossRefPubMed
7.
Zurück zum Zitat Kirsch M, Nakashima K, Kubota S, Houël R, Hillion ML, Loisance D. The risk of reoperative heart valve procedures in octogenarian patients. J Heart Valve Dis. 2004;13(6):991–6.PubMed Kirsch M, Nakashima K, Kubota S, Houël R, Hillion ML, Loisance D. The risk of reoperative heart valve procedures in octogenarian patients. J Heart Valve Dis. 2004;13(6):991–6.PubMed
8.
Zurück zum Zitat Dvir D, Webb JG, Bleiziffer S, Pasic M, Waksman R, Kodali S, Barbanti M, Latib A, Schaefer U, Rodés-Cabau J, Treede H, Piazza N, Hildick-Smith D, Himbert D, Walther T, Hengstenberg C, Nissen H, Bekeredjian R, Presbitero P, Ferrari E, Segev A, de Weger A, Windecker S, Moat NE, Napodano M, Wilbring M, Cerillo AG, Brecker S, Tchetche D, Lefèvre T, De Marco F, Fiorina C, Petronio AS, Teles RC, Testa L, Laborde JC, Leon MB, Kornowski R, Valve-in-Valve International Data Registry Investigators. Transcatheter aortic valve implantation in failed bioprosthetic surgical valves. JAMA. 2014;312:162–70.CrossRefPubMed Dvir D, Webb JG, Bleiziffer S, Pasic M, Waksman R, Kodali S, Barbanti M, Latib A, Schaefer U, Rodés-Cabau J, Treede H, Piazza N, Hildick-Smith D, Himbert D, Walther T, Hengstenberg C, Nissen H, Bekeredjian R, Presbitero P, Ferrari E, Segev A, de Weger A, Windecker S, Moat NE, Napodano M, Wilbring M, Cerillo AG, Brecker S, Tchetche D, Lefèvre T, De Marco F, Fiorina C, Petronio AS, Teles RC, Testa L, Laborde JC, Leon MB, Kornowski R, Valve-in-Valve International Data Registry Investigators. Transcatheter aortic valve implantation in failed bioprosthetic surgical valves. JAMA. 2014;312:162–70.CrossRefPubMed
9.
Zurück zum Zitat Jabbour RJ, Tanaka A, Finkelstein A, Mack M, Tamburino C, Van Mieghem N, de Backer O, Testa L, Gatto P, Purita P, Rahhab Z, Veulemans V, Stundl A, Barbanti M, Nerla R, Sinning JM, Dvir D, Tarantini G, Szerlip M, Scholtz W, Scholtz S, Tchetche D, Castriota F, Butter C, Søndergaard L, Abdel-Wahab M, Sievert H, Alfieri O, Webb J, Rodés-Cabau J, Colombo A, Latib A. Delayed coronary obstruction after transcatheter aortic valve replacement. J Am Coll Cardiol. 2018;71(14):1513–24.CrossRefPubMed Jabbour RJ, Tanaka A, Finkelstein A, Mack M, Tamburino C, Van Mieghem N, de Backer O, Testa L, Gatto P, Purita P, Rahhab Z, Veulemans V, Stundl A, Barbanti M, Nerla R, Sinning JM, Dvir D, Tarantini G, Szerlip M, Scholtz W, Scholtz S, Tchetche D, Castriota F, Butter C, Søndergaard L, Abdel-Wahab M, Sievert H, Alfieri O, Webb J, Rodés-Cabau J, Colombo A, Latib A. Delayed coronary obstruction after transcatheter aortic valve replacement. J Am Coll Cardiol. 2018;71(14):1513–24.CrossRefPubMed
10.
Zurück zum Zitat Ribeiro HB, Nombela-Franco L, Urena M, Mok M, Pasian S, Doyle D, DeLarochelliere R, Cote M, Laflamme L, DeLarochelliere H, Allende R, Dumont E, Rodes-Cabau J. Coronary obstruction following transcatheter aortic valve implantation. A systematic review. JACC Cardiovasc Interv. 2013;6:452–61.CrossRefPubMed Ribeiro HB, Nombela-Franco L, Urena M, Mok M, Pasian S, Doyle D, DeLarochelliere R, Cote M, Laflamme L, DeLarochelliere H, Allende R, Dumont E, Rodes-Cabau J. Coronary obstruction following transcatheter aortic valve implantation. A systematic review. JACC Cardiovasc Interv. 2013;6:452–61.CrossRefPubMed
11.
Zurück zum Zitat Ribeiro HB, Webb JG, Makkar RR, Cohen MG, Kapadia SR, Kodali S, Tamburino C, Barbanti M, Chakravarty T, Jilaihawi H, Paradis JM, de Brito FS Jr, Canovas SJ, Cheema AN, de Jaegere PP, del Valle R, Chiam PT, Moreno R, Pradas G, Ruel M, Salgado-Fernandez J, Sarmento-Leite R, Toeg HD, Velianou JL, Zajarias A, Babaliaros V, Cura F, Dager AE, Manoharan G, Lerakis S, Pichard AD, Radhakrishnan S, Perin MA, Dumont E, Larose E, Pasian SG, Nombela-Franco L, Urena M, Tuzcu EM, Leon MB, Amat-Santos IJ, Leipsic J, Rodes-Cabau J. Predictive factors, management, and clinical outcomes of coronary obstruction following transcatheter aortic valve implantation: insights from a large multicenter registry. J Am Coll Cardiol. 2013;62:1552–62.CrossRefPubMed Ribeiro HB, Webb JG, Makkar RR, Cohen MG, Kapadia SR, Kodali S, Tamburino C, Barbanti M, Chakravarty T, Jilaihawi H, Paradis JM, de Brito FS Jr, Canovas SJ, Cheema AN, de Jaegere PP, del Valle R, Chiam PT, Moreno R, Pradas G, Ruel M, Salgado-Fernandez J, Sarmento-Leite R, Toeg HD, Velianou JL, Zajarias A, Babaliaros V, Cura F, Dager AE, Manoharan G, Lerakis S, Pichard AD, Radhakrishnan S, Perin MA, Dumont E, Larose E, Pasian SG, Nombela-Franco L, Urena M, Tuzcu EM, Leon MB, Amat-Santos IJ, Leipsic J, Rodes-Cabau J. Predictive factors, management, and clinical outcomes of coronary obstruction following transcatheter aortic valve implantation: insights from a large multicenter registry. J Am Coll Cardiol. 2013;62:1552–62.CrossRefPubMed
12.
Zurück zum Zitat Dvir D, Leipsic J, Blanke P, Ribeiro HB, Kornowski R, Pichard A, Rodés-Cabau J, Wood DA, Stub D, Ben-Dor, Maluenda G, Makkar RR, Webb JG. Coronary obstruction in transcatheter aortic valve-in-valve implantation: preprocedural evaluation, device selection, protection, and treatment. Circ Cardiovasc Interv. 2015;8(1):e002079. https://doi.org/10.1161/CIRCINTERVENTIONS.114.002079. Dvir D, Leipsic J, Blanke P, Ribeiro HB, Kornowski R, Pichard A, Rodés-Cabau J, Wood DA, Stub D, Ben-Dor, Maluenda G, Makkar RR, Webb JG. Coronary obstruction in transcatheter aortic valve-in-valve implantation: preprocedural evaluation, device selection, protection, and treatment. Circ Cardiovasc Interv. 2015;8(1):e002079. https://​doi.​org/​10.​1161/​CIRCINTERVENTION​S.​114.​002079.
13.
Zurück zum Zitat Ribeiro HB, Rodés-Cabau J, Blanke P, Leipsic J, Kwan Park J, Bapat V, Makkar R, Simonato M, Barbanti M, Schofer J, Bleiziffer S, Latib A, Hildick-Smith D, Presbitero P, Windecker S, Napodano M, Cerillo AG, Abdel-Wahab M, Tchetche D, Fiorina C, Sinning JM, Cohen MG, Guerrero ME, Whisenant B, Nietlispach F, Palma JH, Nombela-Franco L, de Weger A, Kass M, Sandoli de Brito F Jr, Lemos PA, Kornowski R, Webb J, Dvir D. Incidence, predictors, and clinical outcomes of coronary obstruction following transcatheter aortic valve replacement for degenerative bioprosthetic surgical valves: insights from the VIVID registry. Eur Heart J. 2018;39(8):687–95.CrossRefPubMed Ribeiro HB, Rodés-Cabau J, Blanke P, Leipsic J, Kwan Park J, Bapat V, Makkar R, Simonato M, Barbanti M, Schofer J, Bleiziffer S, Latib A, Hildick-Smith D, Presbitero P, Windecker S, Napodano M, Cerillo AG, Abdel-Wahab M, Tchetche D, Fiorina C, Sinning JM, Cohen MG, Guerrero ME, Whisenant B, Nietlispach F, Palma JH, Nombela-Franco L, de Weger A, Kass M, Sandoli de Brito F Jr, Lemos PA, Kornowski R, Webb J, Dvir D. Incidence, predictors, and clinical outcomes of coronary obstruction following transcatheter aortic valve replacement for degenerative bioprosthetic surgical valves: insights from the VIVID registry. Eur Heart J. 2018;39(8):687–95.CrossRefPubMed
14.
Zurück zum Zitat Pfeiffer S, Santarpino G, Fischlein T, Jessl J, Pauschinger M. Coronary obstruction following TAVI valve-in-valve: could we prevent it? Catheter Cardiovasc Interv. 2013;81(2):386.CrossRefPubMed Pfeiffer S, Santarpino G, Fischlein T, Jessl J, Pauschinger M. Coronary obstruction following TAVI valve-in-valve: could we prevent it? Catheter Cardiovasc Interv. 2013;81(2):386.CrossRefPubMed
Metadaten
Titel
Sudden death after valve-in-valve procedure due to delayed coronary obstruction: a case report
verfasst von
Angelo Buscaglia
Giacomo Tini
Gian Paolo Bezante
Claudio Brunelli
Manrico Balbi
Publikationsdatum
01.12.2018
Verlag
BioMed Central
Erschienen in
Journal of Medical Case Reports / Ausgabe 1/2018
Elektronische ISSN: 1752-1947
DOI
https://doi.org/10.1186/s13256-018-1785-x

Weitere Artikel der Ausgabe 1/2018

Journal of Medical Case Reports 1/2018 Zur Ausgabe