30.01.2018 | Musculoskeletal | Ausgabe 7/2018 Open Access

European Radiology 7/2018

Detection of early cartilage damage: feasibility and potential of gagCEST imaging at 7T

Zeitschrift:
European Radiology > Ausgabe 7/2018
Autoren:
Sander Brinkhof, Razmara Nizak, Vitaliy Khlebnikov, Jeanine J. Prompers, Dennis W.J. Klomp, Daniel B.F. Saris
Wichtige Hinweise

Electronic supplementary material

The online version of this article (https://​doi.​org/​10.​1007/​s00330-017-5277-y) contains supplementary material, which is available to authorized users.
A comment to this article is available online at https://​doi.​org/​10.​1007/​s00330-018-5457-4.

Abstract

Objectives

The purpose was to implement a fast 3D glycosaminoglycan Chemical Exchange Saturation Transfer (gagCEST) sequence at 7 T, test stability and reproducibility in cartilage in the knee in healthy volunteers, and evaluate clinical applicability in cartilage repair patients.

Methods

Experiments were carried out on a 7-T scanner using a volume transmit coil and a 32-channel receiver wrap-around knee coil. The 3D gagCEST measurement had an acquisition time of 7 min. Signal stability and reproducibility of the GAG effect were assessed in eight healthy volunteers. Clinical applicability of the method was demonstrated in five patients before cartilage repair surgery.

Results

Coefficient of variation of the gagCEST signal was 1.9%. The reproducibility of the GAG effect measurements was good in the medial condyle (ICC = 0.87) and excellent in the lateral condyle (ICC = 0.97). GAG effect measurements in healthy cartilage ranged from 2.6%-12.4% compared with 1.3%-5.1% in damaged cartilage. Difference in GAG measurement between healthy cartilage and damaged cartilage was significant (p < 0.05).

Conclusions

A fast 3D gagCEST sequence was applied at 7 T for use in cartilage in the knee, acquired within a clinically feasible scan time of 7 min. We demonstrated that the method has high stability, reproducibility and clinical applicability.

Key Points

• gagCEST measurements are stable and reproducible
• A non-invasive GAG measurement with gagCEST can be acquired in 7 min
• gagCEST is able to discriminate between healthy and damaged cartilage

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Jetzt e.Med zum Sonderpreis bestellen!

Sichern Sie sich jetzt Ihr e.Med-Abo und sparen Sie 50 %!

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Jetzt e.Med zum Sonderpreis bestellen!

Sichern Sie sich jetzt Ihr e.Med-Abo und sparen Sie 50 %!

Zusatzmaterial
Literatur
Über diesen Artikel

Weitere Artikel der Ausgabe 7/2018

European Radiology 7/2018 Zur Ausgabe