Detection of intratracheal accumulation of thick secretions by using continuous monitoring of respiratory acoustic spectrum: a preliminary analysis

verfasst von: Young-Jin Moon, Allison J. Bechtel, Sung-Hoon Kim, Jae-Won Kim, Robert H. Thiele, Randal S. Blank

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 4/2020

Einloggen, um Zugang zu erhalten

Abstract

The accumulation of tracheobronchial secretions may contribute to a deterioration in pulmonary function and its early detection is important. In this study, we analyzed the respiratory sound spectrum in patients with intratracheal secretion, and compared acoustic characteristics before and after therapeutic endotracheal suctioning. After review of anesthetic records of liver transplant recipients, we included recipients with identified intratracheal secretion during surgery. Intraoperative breath sounds recorded through esophageal stethoscope were sampled in 20 s-period before and after suctioning of secretion and analyzed using fast Fourier transform. We also analyzed normal breath sounds from recipients without any respiratory problem as control group. The maximal power (dBmMax), total power from whole frequency range of 80–500 Hz (Pt), total power of each frequency range (80–200 Hz, P80–200; 200–300 Hz, P200–300; 300–400 Hz, P300–400; 400–500 Hz, P400–500), and their ratio (P80–200/Pt, P200–300/Pt, P300–400/Pt, P400–500/Pt) were compared. Breath sounds were obtained from 20 recipients; 9 pairs of breath sound before and after suctioning of secretion and 11 normal breath sounds. Patients with intratracheal secretion showed significantly higher P80–200, P200–300, P300–400, P400–500 when compared to the those of normal control patients (P = 0.003, P = 0.002, and P = 0.009, respectively), while dBmMax did not differ. Elimination of secretions attenuated P80–200, P200–300, P300–400, and P400–500 by 22.4%, 25.7%, 48.5%, and 15.3%, respectively (P = 0.002, 0.024, 0.009, and 0.016, respectively). Identifying the presence of intratracheal secretions with power ratio at 80–200 Hz and 300–400 Hz showed the highest area under the curve of 0.955 in receiver operating characteristic curve analysis. We suggest that spectral analysis of breath sounds obtained from the esophageal stethoscope might be a useful non-invasive respiratory monitor for accumulation of intratracheal secretion. Further prospective studies to evaluate the utility of acoustic analysis in surgical patients are warranted.
Literatur
2.
Zurück zum Zitat Haight C. Intratracheal suction in the management of postoperative pulmonary complications. Ann Surg. 1938;107:218–28.CrossRef Haight C. Intratracheal suction in the management of postoperative pulmonary complications. Ann Surg. 1938;107:218–28.CrossRef
3.
Zurück zum Zitat Checketts MR, Alladi R, Ferguson K, Gemmell L, Handy JM, Klein AA, Love NJ, Misra U, Morris C, Nathanson MH, Rodney GE, Verma R, Pandit JJ, Association of Anaesthetists of Great Britain and Ireland. Recommendations for standards of monitoring during anaesthesia and recovery 2015: Association of Anaesthetists of Great Britain and Ireland. Anaesthesia. 2016;71:85–93. https://doi.org/10.1111/anae.13316.CrossRefPubMed Checketts MR, Alladi R, Ferguson K, Gemmell L, Handy JM, Klein AA, Love NJ, Misra U, Morris C, Nathanson MH, Rodney GE, Verma R, Pandit JJ, Association of Anaesthetists of Great Britain and Ireland. Recommendations for standards of monitoring during anaesthesia and recovery 2015: Association of Anaesthetists of Great Britain and Ireland. Anaesthesia. 2016;71:85–93. https://​doi.​org/​10.​1111/​anae.​13316.CrossRefPubMed
4.
Zurück zum Zitat Roguin A. Rene Theophile Hyacinthe Laennec (1781-1826): the man behind the stethoscope. Clin Med Res. 2006;4:230–5.CrossRef Roguin A. Rene Theophile Hyacinthe Laennec (1781-1826): the man behind the stethoscope. Clin Med Res. 2006;4:230–5.CrossRef
5.
Zurück zum Zitat Smith C. An endo-oesophageal stethoscope. Anesthesiology. 1954;15:566.CrossRef Smith C. An endo-oesophageal stethoscope. Anesthesiology. 1954;15:566.CrossRef
6.
Zurück zum Zitat Wipf JE, Lipsky BA, Hirschmann JV, Boyko EJ, Takasugi J, Peugeot RL, Davis CL. Diagnosing pneumonia by physical examination: relevant or relic? Arch Intern Med. 1999;159:1082–7.CrossRef Wipf JE, Lipsky BA, Hirschmann JV, Boyko EJ, Takasugi J, Peugeot RL, Davis CL. Diagnosing pneumonia by physical examination: relevant or relic? Arch Intern Med. 1999;159:1082–7.CrossRef
13.
Zurück zum Zitat Arnott PJ, Pfeiffer GW, Tavel ME. Spectral analysis of heart sounds: relationships between some physical characteristics and frequency spectra of first and second heart sounds in normals and hypertensives. J Biomed Eng. 1984;6:121–8.CrossRef Arnott PJ, Pfeiffer GW, Tavel ME. Spectral analysis of heart sounds: relationships between some physical characteristics and frequency spectra of first and second heart sounds in normals and hypertensives. J Biomed Eng. 1984;6:121–8.CrossRef
14.
Zurück zum Zitat Manecke GR Jr, Dilger JP, Kutner LJ, Poppers PJ. Auscultation revisited: the waveform and spectral characteristics of breath sounds during general anesthesia. Int J Clin Monit Comput. 1997;14:231–40.CrossRef Manecke GR Jr, Dilger JP, Kutner LJ, Poppers PJ. Auscultation revisited: the waveform and spectral characteristics of breath sounds during general anesthesia. Int J Clin Monit Comput. 1997;14:231–40.CrossRef
15.
Zurück zum Zitat Sandor L, Fenyes G. When is rigid internal fixation indicate din multiple injuries with head trauma? (author’s transl). Magy Traumatol Orthop Helyreallito Seb. 1981;24:163–6.PubMed Sandor L, Fenyes G. When is rigid internal fixation indicate din multiple injuries with head trauma? (author’s transl). Magy Traumatol Orthop Helyreallito Seb. 1981;24:163–6.PubMed
18.
Zurück zum Zitat Parsaei H, Vakily A, Shafiei AM. A wireless electronic esophageal stethoscope for continuous monitoring of cardiovascular and respiratory systems during anaesthesia. J Biomed Phys Eng. 2017;7:69–78.PubMedPubMedCentral Parsaei H, Vakily A, Shafiei AM. A wireless electronic esophageal stethoscope for continuous monitoring of cardiovascular and respiratory systems during anaesthesia. J Biomed Phys Eng. 2017;7:69–78.PubMedPubMedCentral
22.
Zurück zum Zitat Branson RD. Secretion management in the mechanically ventilated patient. Respir Care. 2007;52:1328–42 discussion 1342–1327.PubMed Branson RD. Secretion management in the mechanically ventilated patient. Respir Care. 2007;52:1328–42 discussion 1342–1327.PubMed
Metadaten
Titel
Detection of intratracheal accumulation of thick secretions by using continuous monitoring of respiratory acoustic spectrum: a preliminary analysis
verfasst von
Young-Jin Moon
Allison J. Bechtel
Sung-Hoon Kim
Jae-Won Kim
Robert H. Thiele
Randal S. Blank
Publikationsdatum
20.07.2019
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 4/2020
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-019-00359-z

Weitere Artikel der Ausgabe 4/2020

Journal of Clinical Monitoring and Computing 4/2020 Zur Ausgabe

Update AINS