Determining the accuracy of zero-flux and ingestible thermometers in the peri-operative setting

verfasst von: James M. Jack, Helen Ellicott, Christopher I. Jones, Stephen A. Bremner, Ian Densham, C. Mark Harper

Erschienen in: Journal of Clinical Monitoring and Computing | Ausgabe 6/2019

Einloggen, um Zugang zu erhalten

Abstract

Accurately monitoring peri-operative core temperature is a cornerstone of good practice. Relatively invasive devices such as oesophageal temperature probes and pulmonary artery catheters facilitate this, but are inappropriate for many patients. There remains a need for accurate monitors of core temperature that can be used in awake patients. This study compared the accuracy of two core temperature thermometers that can be used for this purpose: the 3M Bair Hugger™ Temperature Monitoring System Zero Flux Thermometer and the CorTempR™ Wireless Ingestible Temperature Sensor. Readings were compared with the oesophageal probe, the current intraoperative standard. Thirty patients undergoing elective surgical procedures under general anaesthesia were recruited. The ingestible sensor was ingested prior to induction of anaethesia, and post induction, the zero-flux electrode attached above the right eyebrow and oesophageal probe inserted. During surgery, the temperature on each device was recorded every minute. Measurements were compared using Bland–Altman analysis. The ingestible sensor experienced interference from use of diathermy and fluoroscopy in the operating theatre, rendering 39% of its readings unusable. These were removed from analysis. With remaining readings the bias compared with oesophageal probe was + 0.42 °C, with 95% limits of agreement − 2.4 °C to 3.2 °C. 75.4% of readings were within ± 0.5 °C of the OTP reading. The bias for the zero flux electrode compared to oesophageal probe was + 0.02 °C with 95% limits of agreement − 0.5 °C to 0.5 °C. 97.7% of readings were within ± 0.5 °C of the oesophageal probe. The study findings suggest the zero-flux thermometer is sufficiently accurate for clinical use, whereas the ingestible sensor is not.
Trial registration The study was registered at http://​www.​clinicaltrials.​gov, NCT Number: NCT02121574.
Literatur
2.
Zurück zum Zitat Harper CM, McNicholas T, Gowrie-Mohan S. Maintaining perioperative normothermia. Brit Med J. 2003;326(7392):721–2.CrossRef Harper CM, McNicholas T, Gowrie-Mohan S. Maintaining perioperative normothermia. Brit Med J. 2003;326(7392):721–2.CrossRef
3.
Zurück zum Zitat Sessler DI. Perioperative heat balance. Anesthesiology. 2000;92(2):578–96.CrossRef Sessler DI. Perioperative heat balance. Anesthesiology. 2000;92(2):578–96.CrossRef
4.
Zurück zum Zitat NICE. Hypothermia: prevention and management in adults having surgery. NICE clinical guideline 65. London: National Institute for Health and Clinical Excellence; 2008 (updated 2016). NICE. Hypothermia: prevention and management in adults having surgery. NICE clinical guideline 65. London: National Institute for Health and Clinical Excellence; 2008 (updated 2016).
5.
Zurück zum Zitat Sessler D. Temperature monitoring and perioperative thermoregulation. Anesthesiology. 2008;109(2):318–38.CrossRef Sessler D. Temperature monitoring and perioperative thermoregulation. Anesthesiology. 2008;109(2):318–38.CrossRef
6.
Zurück zum Zitat Harper CM. The need for an accurate noninvasive thermometer. Anesth Analg. 2009;109(1):288.CrossRef Harper CM. The need for an accurate noninvasive thermometer. Anesth Analg. 2009;109(1):288.CrossRef
7.
Zurück zum Zitat Fox RH, Solman AJ, Isaacs R, et al. A new method for monitoring deep body temperature from the skin surface. Clin Sci. 1973;44:81–6.CrossRef Fox RH, Solman AJ, Isaacs R, et al. A new method for monitoring deep body temperature from the skin surface. Clin Sci. 1973;44:81–6.CrossRef
8.
Zurück zum Zitat Teunissen LP, Klewer J, de Haan A, et al. Noninvasive continuous core temperature measurement by zero heat flux. Physiol Meas. 2011;32:559–70.CrossRef Teunissen LP, Klewer J, de Haan A, et al. Noninvasive continuous core temperature measurement by zero heat flux. Physiol Meas. 2011;32:559–70.CrossRef
12.
Zurück zum Zitat Lancaster GA, Dodd S, Williamson PR. Design and analysis of pilot studies: recommendations for good practice. J Eval Clin Pract. 2004;10(2):307–12.CrossRef Lancaster GA, Dodd S, Williamson PR. Design and analysis of pilot studies: recommendations for good practice. J Eval Clin Pract. 2004;10(2):307–12.CrossRef
13.
Zurück zum Zitat Lees DE, Kim YD, Macnamara TE. Noninvasive determination of core temperature during anesthesia. South Med J. 1980;73(10):1322–24.CrossRef Lees DE, Kim YD, Macnamara TE. Noninvasive determination of core temperature during anesthesia. South Med J. 1980;73(10):1322–24.CrossRef
17.
Zurück zum Zitat Mäkinen M-T, Pesonen A, Jousela I, et al. Novel zero-heat-flux deep body temperature measurement in lower extremity vascular and cardiac surgery. J Cardiothorac Vasc Anesth. 2016;30(4):973–8.CrossRef Mäkinen M-T, Pesonen A, Jousela I, et al. Novel zero-heat-flux deep body temperature measurement in lower extremity vascular and cardiac surgery. J Cardiothorac Vasc Anesth. 2016;30(4):973–8.CrossRef
19.
Zurück zum Zitat Eshraghi Y, Nasr V, Parra-Sanchez I, et al. An evaluation of a zero-heat-flux cutaneous thermometer in cardiac surgical patients. Anesth Analg. 2014;119:543–9.CrossRef Eshraghi Y, Nasr V, Parra-Sanchez I, et al. An evaluation of a zero-heat-flux cutaneous thermometer in cardiac surgical patients. Anesth Analg. 2014;119:543–9.CrossRef
Metadaten
Titel
Determining the accuracy of zero-flux and ingestible thermometers in the peri-operative setting
verfasst von
James M. Jack
Helen Ellicott
Christopher I. Jones
Stephen A. Bremner
Ian Densham
C. Mark Harper
Publikationsdatum
08.01.2019
Verlag
Springer Netherlands
Erschienen in
Journal of Clinical Monitoring and Computing / Ausgabe 6/2019
Print ISSN: 1387-1307
Elektronische ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-019-00252-9

Weitere Artikel der Ausgabe 6/2019

Journal of Clinical Monitoring and Computing 6/2019 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.