Erschienen in:
24.02.2018 | Original Article
Development of a novel shrouded impeller pediatric blood pump
verfasst von:
Talha Irfan Khan, Haris Sheh Zad, Ismail Lazoglu, Ozlem Yalcin
Erschienen in:
Journal of Artificial Organs
|
Ausgabe 2/2018
Einloggen, um Zugang zu erhalten
Abstract
The aim of this work was to analyze a shrouded impeller pediatric ventricular assist device (SIP-VAD). This device has distinctive design characteristics and parameter optimizations for minimization of recirculation flow and reduction in high-stress regions that cause blood damage. Computational Fluid Dynamics (CFD) simulations were performed to analyze the optimized design. The bench-top prototype of SIP-VAD was manufactured with biocompatible stainless steel. A study on the hydrodynamic and hemodynamic performance of the SIP-VAD was conducted with predictions from CFD and actual experimentation values, and these results were compared. The CFD analysis yielded a pressure range of 29–90 mmHg corresponding to flow rates of 0.5–3 L/min over 9000–11000 rpm. The predicted value of the normalized index of hemolysis (NIH) was 0.0048 g/100 L. The experimental results with the bench-top prototype showed a pressure rise of 30–105 mmHg for the flow speed of 8000–12000 rpm and flow rate of 0.5–3.5 L/min. The maximum difference between CFD and experimental results was 4 mmHg pressure. In addition, the blood test showed the average NIH level of 0.00674 g/100 L. The results show the feasibility of shrouded impeller design of axial-flow pump for manufacturing the prototype for further animal trials.