A 74-year-old man with paroxysmal atrial fibrillation, severe ischemic heart failure, after surgical coronary artery revascularization and implantation of cardiac resynchronization cardioverter-defibrillator was referred for left atrial appendage closure (LAAC). He was treated with dabigatran because of INR lability on warfarin. Nevertheless, the patient suffered acute life-threatening gastrointestinal bleeding a few months ago. The procedure of LAAC was performed under local anesthesia via the right femoral vein and with transoesophageal echocardiography (TOE) guidance. After transseptal puncture, a cardioangiography revealed a cone-shaped windsock left atrial appendage (LAA). The landing zone measured on fluoroscopy was 21 mm. A 25-mm Amplatzer Amulet device was deployed; however, it failed the stability tests. Therefore, remeasurements were performed, and finally, a 28-mm plug was chosen and successfully deployed without any leak. Despite a good outcome in the stability test, little compression of the lobe was observed (Fig. 1). The deployment of the occluder was suboptimal but was deemed as the best possible. Nonetheless, a routine transthoracic echocardiography the next day revealed embolization of the device into the apical area of the LV (Fig. 2), without any impairment of mitral valve function. The patient was asymptomatic and remained clinically stable. LAA occluder embolization is a rare but known complication and can be managed both surgically and percutaneously [1]. Under local anesthesia, common femoral artery access was obtained with an 80-cm 16F Check-Flo Performer Introducer (Cook Medical Inc.). A 7F JR 4.0 guiding catheter was placed through a sheath into the left ventricle, and a 30-mm Multisnare device (Braun Interventional Systems Inc.) was introduced to snare the occluder. Subsequently, the device was carefully moved into the left ventricular outflow tract and pulled across the aortic valve to the ascending aorta. The recapturing maneuver was performed in the abdominal aorta. TOE ruled out any mechanical damage to the aortic valve. No complication was observed in the 6 months follow-up. We obtained the following experiences from our procedure. There were three critical moments which allowed the procedure to be performed successfully and without complications. The first difficulty was to grab the device in the left ventricle. Secondly, the device was brought to the ascending aorta through the aortic valve. A firm and continuous pullback got the occluder stuck in the aortic valve and subsequently made it jump into the aorta (Movies 2 and 3). Finally, to prevent any cerebral events, the occluder was pulled back into the sheath in the descending aorta. A challenging anatomy of the LAA and a sinus rhythm are the principal risk factors of device embolization [2, 3]. However, some authors hypothesize that device embolization can occur without obvious explanation, and it can be associated with limitations of the method. In our case, prior to device release, four signs of proper device placement were observed. The distal part of the lobe was slightly over-compressed. Therefore, adhesion of the proximal part of the lobe to the LAA wall was incomplete, and this could be the cause of embolization The anatomy of the LAA might have been unfavorable for the Amplatzer Amulet device [3]. The patient presented in this manuscript is one of the few examples supporting the percutaneous approach for the retrieval of embolized LAA occluders as a safe and efficacious solution for such complication.
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
download
DOWNLOAD
print
DRUCKEN

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Innere Medizin

Kombi-Abonnement

Mit e.Med Innere Medizin erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Innere Medizin, den Premium-Inhalten der internistischen Fachzeitschriften, inklusive einer gedruckten internistischen Zeitschrift Ihrer Wahl.

Anhänge
Literatur
1.
Zurück zum Zitat Aminian A, Lalmand J, Tzikas A, Budts W, Benit E, Kefer J. Embolization of left atrial appendage closure devices: a systematic review of cases reported with the watchman device and the Amplatzer cardiac plug. Catheter Cardiovasc Interv. 2015;86(1):128–35.CrossRefPubMed Aminian A, Lalmand J, Tzikas A, Budts W, Benit E, Kefer J. Embolization of left atrial appendage closure devices: a systematic review of cases reported with the watchman device and the Amplatzer cardiac plug. Catheter Cardiovasc Interv. 2015;86(1):128–35.CrossRefPubMed
Metadaten
Titel
Device embolization into the LV following left atrial endage closure with an Amplatzer Amulet
verfasst von
Krzysztof Kaczmarek
Bartosz Czarniak
Piotr Jakubowski
Pawel Ptaszynski
Publikationsdatum
03.05.2018
Verlag
Springer US
Erschienen in
Journal of Interventional Cardiac Electrophysiology / Ausgabe 2/2018
Print ISSN: 1383-875X
Elektronische ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-018-0374-7

Weitere Artikel der Ausgabe 2/2018

Journal of Interventional Cardiac Electrophysiology 2/2018 Zur Ausgabe

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.