Erschienen in: European Radiology 8/2020

Open Access 27.03.2020 | Breast

Diagnostic performance of standard breast MRI compared to dedicated axillary MRI for assessment of node-negative and node-positive breast cancer

verfasst von: Sanaz Samiei, Marjolein L. Smidt, Sigrid Vanwetswinkel, Sanne M. E. Engelen, Robert-Jan Schipper, Marc B. I. Lobbes, Thiemo J. A. van Nijnatten

Erschienen in: European Radiology | Ausgabe 8/2020

Abstract

Objectives

To investigate whether breast MRI has comparable diagnostic performance as dedicated axillary MRI regarding assessment of node-negative and node-positive breast cancer.

Methods

Forty-seven patients were included. All had undergone both breast MRI and dedicated axillary MRI, followed by surgery. All included breast MRI exams had complete field of view (FOV) of the axillary region. First, unenhanced T2-weighted (T2W) and subsequent diffusion-weighted (DW) images of both MRI exams were independently analyzed by two breast radiologists using a confidence scale and compared to histopathology. ADC values were measured by two researchers independently. Diagnostic performance parameters were calculated on a patient-by-patient basis.

Results

T2W breast MRI had the following diagnostic performance: sensitivity of 50.0% and 62.5%, specificity of 92.3%, PPV of 57.1% and 62.5%, NPV of 90.0% and 92.3%, and AUC of 0.72 for reader 1 and 0.78 for reader 2. T2W dedicated axillary MRI had the following diagnostic performance: sensitivity of 37.5% and 62.5%, specificity of 82.1% and 92.3%, PPV of 44.6% and 50.0%, NPV of 87.8% and 91.4%, and AUC of 0.65 for reader 1 and 0.73 for reader 2. In both evaluations, addition of DW images resulted in comparable diagnostic performance. For both breast MRI and dedicated axillary MRI, there was no significant difference between mean ADC values of benign and malignant lymph nodes.

Conclusions

T2W breast MRI with complete FOV of the axillary region has comparable diagnostic performance as T2W dedicated axillary MRI regarding assessment of node-negative and node-positive breast cancer. Optimization of T2W breast MRI protocol by including a complete FOV of the axillary region can, therefore, be recommended in clinical practice.

Key Points

• Breast MRI with complete field of view of the axillary region has comparable diagnostic performance as dedicated axillary MRI regarding assessment of node-negative and node-positive breast cancer.
• Optimization of breast MRI protocol by including a complete field of view of the axillary region is recommended in clinical practice.
• For both breast MRI and dedicated axillary MRI, DW imaging (including ADC measurements) is of no added value.
Hinweise

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
2D
Two-dimensional
3D
Three-dimensional
ADC
Apparent diffusion coefficient
ALND
Axillary lymph node dissection
AUC
Area under the curve
CI
Confidence interval
DW
Diffusion-weighted
FOV
Field of view
IQR
Interquartile range
MRI
Magnetic resonance imaging
NPV
Negative predictive value
NST
Neoadjuvant systemic therapy
PPV
Positive predictive value
ROC
Receiver operating characteristic
ROI
Region of interest
SLNB
Sentinel lymph node biopsy
SNR
Signal-to-noise ratio
T2W
T2-weighted
TE
Echo time
TR
Repetition time

Introduction

In breast cancer, the presence and extent of axillary lymph node metastases is an important prognostic indicator and helps in determining the optimal treatment plan [14]. Accurate assessment of axillary lymph node involvement, therefore, plays a pivotal role in breast cancer treatment. Over the past years, the surgical staging procedures of the axilla have evolved from routine axillary lymph node dissection (ALND) toward less extensive procedures, such as the sentinel lymph node biopsy (SLNB).
Parallel to surgical advances, the imaging techniques for axillary lymph node staging have improved and become increasingly useful as a non-invasive diagnostic imaging modality to assess the axillary lymph node status. This is important since accurate preoperative axillary imaging can contribute to a more patient-tailored treatment strategy regarding axillary surgery. In clinical practice, breast magnetic resonance imaging (MRI) is predominantly used in the preoperative setting to evaluate (initial) tumor extent [57]. Previous studies suggested promising results of breast MRI for assessing the axillary lymph node status [8, 9]. Breast MRI enables radiologists to simultaneously assess the breast tumor and axillary lymph nodes in the same field of view (FOV) if the axillary region is completely visualized with an acceptable signal-to-noise ratio (SNR). However, in 40% of the standard breast MRI exams, no complete FOV of the axillary region is planned, limiting the assessment of extensive axillary lymph node involvement in the upper part of the axilla [10]. Consequently, dedicated axillary MRI has been investigated to improve axillary lymph node imaging by representing the complete axillary region. Several studies have suggested that using a dedicated axillary surface coil increases the diagnostic performance of MRI for assessing the axillary lymph node status [1115]. However, these results were based on studies, including patients that underwent only one of the two imaging protocols, not both.
To our knowledge, no previous study has compared standard breast MRI with dedicated axillary MRI within a single cohort of breast cancer patients who had undergone both MRI exams. Similar performance of axillary lymph node assessment on standard breast MRI, as opposed to dedicated axillary MRI, can make broad implementation possible given the existing breast MRI protocols. This study aimed to investigate whether standard breast MRI has a comparable diagnostic performance with that of dedicated axillary MRI regarding the assessment of node-negative and node-positive breast cancer using unenhanced T2-weighted (T2W) and diffusion-weighted (DW) images.

Materials and methods

Patient population

All patients with histopathologically confirmed invasive breast cancer between August 2012 and December 2014, who had undergone both standard breast MRI and dedicated axillary MRI followed by either SLNB or ALND, were considered for inclusion. Exclusion criteria were neoadjuvant systemic therapy (NST) before axillary surgery and breast MRI exams with incomplete FOV or poor SNR of the axillary region. The local medical ethics committee waived the necessity to acquire informed consent due to the retrospective study design.

MRI acquisition

The breast MRI exams were performed with three different 1.5-T and 3.0-T scanners (Intera, Ingenia, and Achieva, Philips Healthcare), using a dedicated bilateral 16-channel breast coil with the patient in the prone position. The imaging protocol of 1.5-T breast MRI consisted of the following: (1) unenhanced three-dimensional (3D) T2W turbo spin-echo sequence without fat suppression (pixel size, 0.87 × 0.87 mm; repetition time (TR), 2000 ms; echo time (TE), 222 ms; echo train length, 92; flip angle, 90°; slice thickness, 2.0 mm); (2) contrast-enhanced T1W sequence; and (3) DW imaging sequence with fat suppression (b-values of 0, 150, and 800 s/mm2; pixel size, 1.28 × 1.28 mm; TR, 9670 ms; TE, 89 ms; echo train length, 68; flip angle, 90°; slice thickness, 3.0 mm). The imaging protocol of 3.0-T breast MRI consisted of the following: (1) unenhanced two-dimensional (2D) T2W turbo spin-echo sequence without fat suppression (pixel size, 0.59 × 0.59 mm; TR, 5294 ms; TE, 100 ms; echo train length, 27; flip angle, 90°; slice thickness, 2.0 mm); (2) contrast-enhanced T1W sequence; and (3) DW imaging sequence with fat suppression (b-values of 0, 150, and 800 s/mm2; pixel size, 1.25 × 1.25 mm; TR, 8683 ms; TE, 51 ms; echo train length, 43; flip angle, 90°; slice thickness, 3.0 mm). For standard breast MRI, the anatomic confines for an adequate FOV were between the humeral head and xiphoid process of the sternum. The dedicated axillary MRI exams were performed with a 3.0-T scanner (Achieva, Philips Healthcare), using a 32-channel cardiac coil with the patient in the supine position and ipsilateral arm elevated. The imaging protocol of dedicated axillary MRI consisted of the following: (1) unenhanced 3D T2W turbo spin-echo sequence without fat suppression (pixel size, 1.25 × 1.25 mm; TR, 2000 ms; TE, 155 ms; echo train length, 51; flip angle, 90°; slice thickness, 2.5 mm); (2) contrast-enhanced T1W; and (3) DW imaging sequence with fat suppression (b-values of 0, 500, and 800 s/mm2; pixel size, 1.38 × 1.38 mm; TR, 2110 ms; TE, 52 ms; echo train length, 71; flip angle, 90°; slice thickness, 3.0 mm). For dedicated axillary MRI, the anatomic confines for an adequate FOV were between the humeral head and inferior border of the scapula. The apparent diffusion coefficient (ADC) maps were automatically constructed for all DW images using the built-in MR software.

Image analysis

The breast and dedicated axillary MRI exams were analyzed by two dedicated breast radiologists independently with significant experience in breast imaging (M.B.I.L. [reader 1] and S.V. [reader 2] with 11 and 8 years of experience, respectively). Comparable with clinical practice, the radiologists were aware of the laterality of the breast tumor and the clinical tumor size assessed by MRI. However, the histopathological outcome was not provided. For qualitative assessment, first, the ipsilateral axillary lymph nodes were assessed on the unenhanced T2W images and subsequently on the DW images of the standard breast MRI and dedicated axillary MRI exams separately. Both readers scored the axillary lymph nodes on the MRI exams using a 5-point confidence scale, ranging from 0 (no lymph nodes) to 4 (definitely malignant) [16]. The additional information from the DW images was used to adjust the score of the axillary lymph nodes based on the T2W images. If the DW image was unavailable or of poor image quality, the score remained unchanged. Characteristics of a malignant lymph node were based on size and morphologic features including irregular margins, inhomogeneous cortex, perifocal edema, asymmetry, loss of fatty hilum, and/or the absence of chemical shift artifact [16, 17]. For quantitative assessment, the DW images were analyzed by two researchers (S.S. [reader 3] and T.J.A.v.N. [reader 4]) dedicated to axillary lymph node imaging. Both readers were blinded to the histopathological outcome. High signal intensity area in the ipsilateral axilla on the DW images was detected and compared with the T2W images to evaluate whether or not it was an axillary lymph node. If multiple lymph nodes were detected, the lymph node with the longest axis was identified [18, 19]. A region of interest (ROI) was manually drawn on the DW images at b = 800 s/mm2 on one representative slice and then copied to the corresponding ADC map [20, 21]. The whole lymph node region with evidently high signal intensity was delineated in the case of subcentimeter lymph nodes, and in larger lymph nodes only the cortex was delineated avoiding the (fatty) hilum and surrounding tissue [22]. After the delineations, a consensus meeting was held to confirm the lymph node’s delineation between the two readers. Subsequently, both readers independently measured the mean ADC of the largest lymph node. The quantitative assessment was performed in OsiriX (version 10.0, Pixmeo SARL).

Histopathological analysis

The lymph nodes obtained by axillary surgery were recorded as benign, isolated tumor cells (≤ 0.2 mm and/or < 200 cells in a single histological cross section), micrometastasis (0.2 ≤ 2.0 mm), or macrometastasis (> 2.0 mm) [23]. The isolated tumor cells and the micrometastases were considered negative, and macrometastases were considered positive axillary lymph nodes [24].

Statistical analysis

Median and interquartile range (IQR) were given if patient characteristics were not normally distributed. On the confidence scale, the lymph nodes with the scores 0–2 were categorized as benign and lymph nodes with the scores 3–4 were categorized as malignant. Histopathology of the axillary surgery, SLNB or ALND, served as the gold standard. Diagnostic performance parameters of T2W images and T2W with DW images were calculated for both standard breast MRI and dedicated axillary MRI (sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and area under the receiver operating characteristic (ROC) curve (AUC)). The diagnostic performance parameters were presented with 95% confidence intervals (CIs). The nonparametric DeLong test was used to calculate the comparison between two AUCs [25]. Two-sided p values of <  0.05 were considered statistically significant. The mean ADC was compared between benign and malignant lymph nodes using the Wilcoxon signed-rank test. For the quantitative analysis, the intraclass correlation coefficient was calculated between readers 3 and 4. The difference between readers 3 and 4 in the ADC measurement and 95% limits of agreement were computed for standard breast MRI and dedicated axillary MRI. Statistical analyses were performed by using R project software (version 3.5.1, R Foundation for Statistical Computing) and Statistical Package for the Social Sciences software (version 25, IBM).

Results

A total of 70 patients had undergone both standard breast MRI and dedicated axillary MRI. Twelve patients were excluded who had been treated with NST before axillary surgery. Eleven patients were excluded because of breast MRI exams with incomplete FOV of the axillary region. For final analyses, 47 patients (median age, 59 years; IQR, 51–66 years) were included. Thirty-nine (83.0%) patients had benign axillary lymph nodes at final pathology and 8 (17.0%) patients had malignant axillary lymph nodes. The median size of macrometastases was 11.5 mm (IQR, 8.0–25.5 mm). Patient characteristics are summarized in Table 1.
Table 1
Patient characteristics
Variable
Patients (n = 47)
Age (years) (median; IQR)
59 (51–66)
Clinical tumor size (mm) (median; IQR)
19 (13–28.5)
Clinical tumor stage (%)
  T1
25 (53.2)
  T2
21 (44.7)
  T3
1 (2.1)
Tumor type (%)
  Ductal
34 (72.3)
  Lobular
7 (14.9)
  Mixed ductal and lobular
3 (6.4)
  Other*
3 (6.4)
Tumor grade (%)
  1
14 (29.8)
  2
27 (48.9)
  3
10 (21.3)
Receptor status (%)
  ER+HER2-
38 (80.9)
  ER+HER2+
4 (8.5)
  ER-HER2+
1 (2.1)
  Triple negative
4 (8.5)
Breast surgery (%)
  Breast-conserving surgery
25 (53.2)
  Mastectomy
22 (46.8)
Axillary surgery (%)
  SLNB
43 (91.5)
  ALND
4 (8.5)
Abbreviations: IQR, interquartile range; ER, estrogen receptor; PR, progesterone receptor; HER2, human epidermal growth factor receptor 2; SLNB, sentinel lymph node biopsy; ALND, axillary lymph node dissection
*Other tumor types: adenoid cystic carcinoma, mucinous carcinoma, encapsulated papillary carcinoma
The T2W breast MRI had the following diagnostic performance: sensitivity of 50.0% and 62.5%, specificity of 92.3%, PPV of 57.1% and 62.5%, NPV of 90.0% and 92.3%, and AUC of 0.72 for reader 1 and 0.78 for reader 2. The addition of DW images resulted in comparable sensitivity (50.0%), specificity (92.3% and 94.9%), PPV (57.1% and 66.7%), NPV (90.0% and 90.2%), and AUC for reader 1 (0.73) and reader 2 (0.72). The comparison between the AUC values of T2W breast MRI and with the addition of DW images for reader 1 and reader 2 separately had a p value of 0.67 and 0.33, respectively. Two of the standard breast MRI exams had no DW images, and nine DW images had poor image quality assessed by the radiologists.
The T2W dedicated axillary MRI had the following diagnostic performance: sensitivity of 37.5% and 62.5%, specificity of 82.1% and 92.3%, PPV of 44.6% and 50.0%, NPV of 87.8% and 91.4%, and AUC of 0.65 for reader 1 and 0.73 for reader 2. The addition of DW images resulted in comparable sensitivity (25.0% and 62.5%), specificity (87.2% and 92.3%), PPV (28.6% and 62.5%), NPV (85.0% and 92.3%), and AUC for reader 1 (0.57) and reader 2 (0.78). The comparison between the AUC values of T2W dedicated axillary MRI and with the addition of DW images for reader 1 and reader 2 separately had a p value of 0.21 and 0.068, respectively. One of the dedicated axillary MRI exams had no DW image. The results of the diagnostic performance parameters of T2W and T2W with DW images can be found in Table 2 and the ROC curves are presented in Fig. 1.
Table 2
Diagnostic performance of standard breast MRI with complete FOV and dedicated axillary MRI
 
Reader 1
Reader 2
T2W
T2W with DW imaging*
T2W
T2W with DW imaging*
Standard breast MRI (n = 47)
  Sensitivity
50.0% (4/8)
50.0% (4/8)
62.5% (5/8)
50.0% (4/8)
[15.7–84.3]
[15.7–84.3]
[24.5–91.5]
[15.7–84.3]
  Specificity
92.3% (36/39)
94.9% (37/39)
92.3% (36/39)
92.3% (36/39)
[79.1–98.4]
[82.7–99.4]
[79.1–98.4]
[79.1–98.4]
  PPV
57.1% (4/7)
66.7% (4/6)
62.5% (5/8)
57.1% (4/7)
[18.4–90.1]
[22.3–95.7]
[24.5–91.5]
[18.4–90.1]
  NPV
90.0% (36/40)
90.2% (37/41)
92.3% (36/39)
90.0% (36/40)
[76.3–97.2]
[76.9–97.3]
[79.1–98.4]
[76.3–97.2]
  AUC
0.72
0.73
0.78
0.72
[0.53–0.92]
[0.54–0.92]
[0.60–0.97]
[0.53–0.91]
Dedicated axillary MRI (n = 47)
  Sensitivity
37.5% (3/8)
25.0% (2/8)
62.5% (5/8)
62.5% (5/8)
[8.5–75.5]
[3.2–65.1]
[24.5–91.5]
[24.5–91.5]
  Specificity
92.3% (36/39)
87.2% (34/39)
82.1% (32/39)
92.3% (36/39)
[79.1–98.4]
[72.6–95.7]
[66.5–92.5]
[79.1–98.4]
  PPV
50.0% (3/6)
28.6% (2/7)
44.6% (5/12)
62.5% (5/8)
[11.8–88.2]
[3.7–71.0]
[15.2–72.3]
[24.5–91.5]
  NPV
87.8% (36/41)
85.0% (34/40)
91.4% (32/35)
92.3% (36/39)
[73.8–95.9]
[70.2–94.3]
[76.9–98.2]
[79.1–98.4]
  AUC
0.65
0.57
0.73
0.78
[0.46–0.83]
[0.39–0.75]
[0.53–0.92]
[0.60–0.97]
Data in parentheses are absolute numbers. Data in brackets are 95% confidence intervals
*The additional information of the DW MR images was used to adjust the initial score based on the T2W MR images. If the DW image score was unavailable, the T2W image score remained unchanged
Abbreviations: FOV, field of view; T2W, T2-weighted; DW, diffusion-weighted; PPV, positive predictive value; NPV, negative predictive value; AUC, area under the curve
For standard breast MRI, the mean ADC values of benign lymph nodes were 0.608 × 10−3 mm2/s and 0.611 × 10−3 mm2/s for readers 3 and 4, respectively. The mean ADC values for malignant lymph nodes were 0.627 × 10−3 mm2/s and 0.556 × 10−3 mm2/s for readers 3 and 4, respectively. For both readers, there was no significant difference between the mean ADC value of benign and malignant lymph nodes (reader 3, p = 0.40; reader 4, p = 0.61) (Fig. 2a). The intraclass correlation coefficient was excellent (0.94). Since two of the standard breast MRI exams had no DW images, the corresponding ADC maps were also unavailable for those MRI exams. Three other ADC maps had poor image quality assessed by the readers. The mean difference in ADC measurement pairs was 0.0092 × 10−3 mm2/s with 95% limits of agreement of − 0.033 × 10−3 and 0.051 × 10−3 mm2/s.
For dedicated axillary MRI, the mean ADC values of benign lymph nodes were 0.879 × 10−3 mm2/s and 0.866 × 10−3 mm2/s for readers 3 and 4, respectively. The mean ADC values for malignant lymph nodes were 0.727 × 10−3 mm2/s and 0.838 × 10−3 mm2/s for readers 3 and 4, respectively. For both readers, there was no significant difference between the mean ADC values of benign and malignant lymph nodes (reader 3, p = 0.74; reader 4, p = 0.87) (Fig. 2b). The intraclass correlation coefficient was good (0.75). Since one of the dedicated axillary MRI exams had no DW image, the corresponding ADC map was also unavailable. The mean difference in ADC measurement pairs was − 0.0056 × 10−3 mm2/s with 95% limits of agreement of − 0.053 × 10−3 and 0.042 × 10−3 mm2/s. Examples of an axillary lymph node on T2W image, DW image, and ADC map for standard breast MRI and dedicated axillary MRI are shown in Figs. 3 and 4.

Discussion

This study showed that the diagnostic performance of unenhanced T2W standard breast MRI with complete FOV of the axillary region is comparable with unenhanced T2W dedicated axillary MRI regarding the assessment of node-negative and node-positive breast cancer. Concerning the diagnostic performance of both MRI exams, especially the relatively high NPV can be used for clinical decision making. For both standard breast MRI and dedicated axillary MRI, DW images and ADC measurements were of no added value. To our knowledge, this is the first study that has compared standard breast MRI and dedicated axillary MRI within a single cohort of breast cancer patients.
In this present study, all breast cancer patients underwent a standard breast MRI for the evaluation of disease extent, to identify multicentric or multifocal disease, or to identify the presence of additional breast lesions. The evaluation of all axillary lymph nodes can be limited if only breast coils are used especially the lymph nodes located in the upper part of the axilla. In this study, only breast MRI exams with complete FOV and sufficient SNR of the axillary region were included to make the comparison with dedicated axillary MRI exams as equal as possible. Dedicated axillary MRI exams may improve the visualization of lymph nodes in axillary levels ΙΙ and ΙΙΙ which may not be identified easily due to the location. However, previous research showed that up to 94% of the SLNs are located in axillary levels Ι and ΙΙ [2629]. This implies that visualizing axillary level ΙΙΙ, especially in the case of clinically node-negative findings, can be redundant in most cases.
In this study, the unenhanced MR sequence was used since it has been previously described that this sequence has the best anatomical presentation of the lymph nodes based on size and morphology [10, 12, 22]. In addition to the unenhanced T2W sequence, DW imaging has been investigated as an adjunct to help differentiate between benign and malignant lymph nodes. Previously published data about lymph node assessment on unenhanced breast MR imaging showed a sensitivity of 88%, a specificity of 82%, and an accuracy of 85%, and the addition of DW imaging resulted in a sensitivity of 84%, a specificity of 77%, and an accuracy of 80% [22]. The addition of DW imaging was insufficient to improve the diagnostic performance of axillary lymph node assessment, which is in line with our results for both standard breast and dedicated axillary MRI. Further studies evaluating DW imaging showed varying results of sensitivity and specificity of 51.3–94.7% and 90.0–91.8%, respectively [20, 30].
For the quantitative analysis, the ADC values were calculated from the DW imaging. Discordant results have been reported about the ADC values of lymph nodes. Previous research reported similar ADC values for benign and malignant lymph nodes [31], higher values in malignant nodes [32], and lower values in malignant nodes [30, 33, 34]. Our analyses for both standard breast and dedicated axillary MRI showed that the ADC values for benign and malignant lymph nodes are similar. These varying results can be possibly explained by the following: signal intensity on DW imaging can be influenced by lymph node changes like necrotic areas and inflammatory processes. These changes can cause artifacts on the ADC map and therefore not reflect the cellularity of the lymph node. Also, the variety of b-value combinations in different studies can significantly affect the ADC value of lymph nodes [35].
In addition to standard breast MRI, a dedicated axillary MRI was performed in a different session by using a surface coil on the patients’ axilla. With the surface coil, the complete axillary region is visualized in a coronal plane. An additional dedicated axillary MR image requires scanning time per axilla and bilateral evaluation to check for asymmetry is not possible. A few studies have evaluated the diagnostic performance of MR imaging with a dedicated axillary surface coil in the preoperative detection of axillary lymph node metastases [11, 12, 15, 36]. These studies reported sensitivity and specificity between 79.0 and 94.6% and 90.0 and 98.5%, respectively [11, 15, 36]. However, these results of dedicated axillary MR imaging are not comparable with those of the present study since contrast-enhanced T1W and T2*W sequence were used for analyses of axillary lymph node metastases.
Recent advances in the medical image analysis field have been made by introducing artificial intelligence for image recognition tasks [37]. Artificial intelligence with methods ranging from radiomics to convolutional neural networks can provide quantitative rather than qualitative imaging data in an automated fashion [37]. With artificial intelligence as a tool to assist radiology image workflow, the imaging assessment can be made more accurate and reproducible [37]. The assessment of axillary lymph nodes by combining the MR images and artificial intelligence can increase the diagnostic outcome. Future research should provide insights into this topic.
Our study has certain limitations. The number of included patients in this study was relatively small, so that potential minor differences between diagnostic performances could have gone undetected. Given the small sample size, further research on this topic is necessary. Also, the statistical analyses were not based on a node-by-node comparison between the visualized lymph nodes and their pathological findings. We assumed that the overall T2W and DWI score of the axilla (benign or malignant) was correlated with the pathology outcome. This was also the case for the identified lymph node with the longest axis on DWI and the corresponding ADC map. Further, two types of MRI systems (1.5 and 3.0 T) were used. However, the varying MR protocols could not be compared due to the small sample size. Finally, nine DW images of breast MRI were inevitably excluded because of movement or susceptibility artifacts, which could have influenced our DW imaging findings for standard breast MRI. A previous study has also suffered from the same issue [22].
In conclusion, the diagnostic performance of T2W standard breast MRI with complete FOV of the axillary region is comparable with that of the T2W dedicated axillary MRI regarding the assessment of node-negative and node-positive breast cancer. Optimization of T2W standard breast MRI protocol by including a complete FOV of the axillary region can, therefore, be recommended in clinical practice.

Compliance with ethical standards

Guarantor

The scientific guarantor of this publication is T.J.A. van Nijnatten.

Conflict of interest

Dr. M.L. Smidt received research funding from Servier for the microbiota study.
The other authors of this manuscript declare no relationships with any companies whose products or services may be related to the subject matter of the article.

Statistics and biometry

No complex statistical methods were necessary for this paper.
Written informed consent was waived by the Institutional Review Board.

Ethical approval

Institutional Review Board approval was obtained.

Methodology

• Retrospective
• Diagnostic study
• Performed at one institution
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Oliveira M, Cortes J, Bellet M et al (2013) Management of the axilla in early breast cancer patients in the genomic era. Ann Oncol 24:1163–1170CrossRef Oliveira M, Cortes J, Bellet M et al (2013) Management of the axilla in early breast cancer patients in the genomic era. Ann Oncol 24:1163–1170CrossRef
2.
Zurück zum Zitat Carter CL, Allen C, Henson DE (1989) Relation of tumor size, lymph node status, and survival in 24,740 breast cancer cases. Cancer 63:181–187CrossRef Carter CL, Allen C, Henson DE (1989) Relation of tumor size, lymph node status, and survival in 24,740 breast cancer cases. Cancer 63:181–187CrossRef
3.
Zurück zum Zitat Fisher B, Bauer M, Wickerham DL et al (1983) Relation of number of positive axillary nodes to the prognosis of patients with primary breast cancer. An NSABP Update Cancer 52:1551–1557PubMed Fisher B, Bauer M, Wickerham DL et al (1983) Relation of number of positive axillary nodes to the prognosis of patients with primary breast cancer. An NSABP Update Cancer 52:1551–1557PubMed
4.
Zurück zum Zitat Banerjee M, George J, Song EY, Roy A, Hryniuk W (2004) Tree-based model for breast cancer prognostication. J Clin Oncol 22:2567–2575CrossRef Banerjee M, George J, Song EY, Roy A, Hryniuk W (2004) Tree-based model for breast cancer prognostication. J Clin Oncol 22:2567–2575CrossRef
5.
Zurück zum Zitat Plana MN, Carreira C, Muriel A et al (2012) Magnetic resonance imaging in the preoperative assessment of patients with primary breast cancer: systematic review of diagnostic accuracy and meta-analysis. Eur Radiol 22:26–38CrossRef Plana MN, Carreira C, Muriel A et al (2012) Magnetic resonance imaging in the preoperative assessment of patients with primary breast cancer: systematic review of diagnostic accuracy and meta-analysis. Eur Radiol 22:26–38CrossRef
6.
Zurück zum Zitat Berg WA, Gutierrez L, NessAiver MS et al (2004) Diagnostic accuracy of mammography, clinical examination, US, and MR imaging in preoperative assessment of breast cancer. Radiology 233:830–849CrossRef Berg WA, Gutierrez L, NessAiver MS et al (2004) Diagnostic accuracy of mammography, clinical examination, US, and MR imaging in preoperative assessment of breast cancer. Radiology 233:830–849CrossRef
7.
Zurück zum Zitat Sardanelli F, Boetes C, Borisch B et al (2010) Magnetic resonance imaging of the breast: recommendations from the EUSOMA working group. Eur J Cancer 46:1296–1316CrossRef Sardanelli F, Boetes C, Borisch B et al (2010) Magnetic resonance imaging of the breast: recommendations from the EUSOMA working group. Eur J Cancer 46:1296–1316CrossRef
8.
Zurück zum Zitat Kuijs VJ, Moossdorff M, Schipper RJ et al (2015) The role of MRI in axillary lymph node imaging in breast cancer patients: a systematic review. Insights Imaging 6:203–215CrossRef Kuijs VJ, Moossdorff M, Schipper RJ et al (2015) The role of MRI in axillary lymph node imaging in breast cancer patients: a systematic review. Insights Imaging 6:203–215CrossRef
9.
Zurück zum Zitat Mortellaro VE, Marshall J, Singer L et al (2009) Magnetic resonance imaging for axillary staging in patients with breast cancer. J Magn Reson Imaging 30:309–312CrossRef Mortellaro VE, Marshall J, Singer L et al (2009) Magnetic resonance imaging for axillary staging in patients with breast cancer. J Magn Reson Imaging 30:309–312CrossRef
10.
Zurück zum Zitat van Nijnatten TJA, Ploumen EH, Schipper RJ et al (2016) Routine use of standard breast MRI compared to axillary ultrasound for differentiating between no, limited and advanced axillary nodal disease in newly diagnosed breast cancer patients. Eur J Radiol 85:2288–2294CrossRef van Nijnatten TJA, Ploumen EH, Schipper RJ et al (2016) Routine use of standard breast MRI compared to axillary ultrasound for differentiating between no, limited and advanced axillary nodal disease in newly diagnosed breast cancer patients. Eur J Radiol 85:2288–2294CrossRef
11.
Zurück zum Zitat Kvistad KA, Rydland J, Smethurst HB, Lundgren S, Fjosne HE, Haraldseth O (2000) Axillary lymph node metastases in breast cancer: preoperative detection with dynamic contrast-enhanced MRI. Eur Radiol 10:1464–1471CrossRef Kvistad KA, Rydland J, Smethurst HB, Lundgren S, Fjosne HE, Haraldseth O (2000) Axillary lymph node metastases in breast cancer: preoperative detection with dynamic contrast-enhanced MRI. Eur Radiol 10:1464–1471CrossRef
12.
Zurück zum Zitat Schipper RJ, Paiman ML, Beets-Tan RG et al (2015) Diagnostic performance of dedicated axillary T2- and diffusion-weighted MR imaging for nodal staging in breast cancer. Radiology 275:345–355CrossRef Schipper RJ, Paiman ML, Beets-Tan RG et al (2015) Diagnostic performance of dedicated axillary T2- and diffusion-weighted MR imaging for nodal staging in breast cancer. Radiology 275:345–355CrossRef
13.
Zurück zum Zitat van Nijnatten TJA, Schipper RJ, Lobbes MBI et al (2018) Diagnostic performance of gadofosveset-enhanced axillary MRI for nodal (re)staging in breast cancer patients: results of a validation study. Clin Radiol 73:168–175CrossRef van Nijnatten TJA, Schipper RJ, Lobbes MBI et al (2018) Diagnostic performance of gadofosveset-enhanced axillary MRI for nodal (re)staging in breast cancer patients: results of a validation study. Clin Radiol 73:168–175CrossRef
14.
Zurück zum Zitat Murray AD, Staff RT, Redpath TW et al (2002) Dynamic contrast enhanced MRI of the axilla in women with breast cancer: comparison with pathology of excised nodes. Br J Radiol 75:220–228CrossRef Murray AD, Staff RT, Redpath TW et al (2002) Dynamic contrast enhanced MRI of the axilla in women with breast cancer: comparison with pathology of excised nodes. Br J Radiol 75:220–228CrossRef
15.
Zurück zum Zitat Yoshimura G, Sakurai T, Oura S et al (1999) Evaluation of axillary lymph node status in breast cancer with MRI. Breast Cancer 6:249–258CrossRef Yoshimura G, Sakurai T, Oura S et al (1999) Evaluation of axillary lymph node status in breast cancer with MRI. Breast Cancer 6:249–258CrossRef
16.
Zurück zum Zitat Baltzer PA, Dietzel M, Burmeister HP et al (2011) Application of MR mammography beyond local staging: is there a potential to accurately assess axillary lymph nodes? Evaluation of an extended protocol in an initial prospective study. AJR Am J Roentgenol 196:W641–W647CrossRef Baltzer PA, Dietzel M, Burmeister HP et al (2011) Application of MR mammography beyond local staging: is there a potential to accurately assess axillary lymph nodes? Evaluation of an extended protocol in an initial prospective study. AJR Am J Roentgenol 196:W641–W647CrossRef
17.
Zurück zum Zitat Farshchian N, Tamari S, Farshchian N, Madani H, Rezaie M, Mohammadi-Motlagh HR (2011) Diagnostic value of chemical shift artifact in distinguishing benign lymphadenopathy. Eur J Radiol 80:594–597CrossRef Farshchian N, Tamari S, Farshchian N, Madani H, Rezaie M, Mohammadi-Motlagh HR (2011) Diagnostic value of chemical shift artifact in distinguishing benign lymphadenopathy. Eur J Radiol 80:594–597CrossRef
18.
Zurück zum Zitat Belli P, Costantini M, Bufi E et al (2015) Diffusion magnetic resonance imaging in breast cancer characterisation: correlations between the apparent diffusion coefficient and major prognostic factors. Radiol Med 120:268–276CrossRef Belli P, Costantini M, Bufi E et al (2015) Diffusion magnetic resonance imaging in breast cancer characterisation: correlations between the apparent diffusion coefficient and major prognostic factors. Radiol Med 120:268–276CrossRef
19.
Zurück zum Zitat Chung J, Youk JH, Kim JA et al (2014) Role of diffusion-weighted MRI: predicting axillary lymph node metastases in breast cancer. Acta Radiol 55:909–916CrossRef Chung J, Youk JH, Kim JA et al (2014) Role of diffusion-weighted MRI: predicting axillary lymph node metastases in breast cancer. Acta Radiol 55:909–916CrossRef
20.
Zurück zum Zitat Kim SH, Shin HJ, Shin KC et al (2017) Diagnostic performance of fused diffusion-weighted imaging using T1-weighted imaging for axillary nodal staging in patients with early breast cancer. Clin Breast Cancer 17:154–163CrossRef Kim SH, Shin HJ, Shin KC et al (2017) Diagnostic performance of fused diffusion-weighted imaging using T1-weighted imaging for axillary nodal staging in patients with early breast cancer. Clin Breast Cancer 17:154–163CrossRef
21.
Zurück zum Zitat Rahbar H, Conlin JL, Parsian S et al (2015) Suspicious axillary lymph nodes identified on clinical breast MRI in patients newly diagnosed with breast cancer: can quantitative features improve discrimination of malignant from benign? Acad Radiol 22:430–438CrossRef Rahbar H, Conlin JL, Parsian S et al (2015) Suspicious axillary lymph nodes identified on clinical breast MRI in patients newly diagnosed with breast cancer: can quantitative features improve discrimination of malignant from benign? Acad Radiol 22:430–438CrossRef
22.
Zurück zum Zitat Scaranelo AM, Eiada R, Jacks LM, Kulkarni SR, Crystal P (2012) Accuracy of unenhanced MR imaging in the detection of axillary lymph node metastasis: study of reproducibility and reliability. Radiology 262:425–434CrossRef Scaranelo AM, Eiada R, Jacks LM, Kulkarni SR, Crystal P (2012) Accuracy of unenhanced MR imaging in the detection of axillary lymph node metastasis: study of reproducibility and reliability. Radiology 262:425–434CrossRef
23.
Zurück zum Zitat Lester SC, Bose S, Chen YY et al (2009) Protocol for the examination of specimens from patients with invasive carcinoma of the breast. Arch Pathol Lab Med 133:1515–1538PubMed Lester SC, Bose S, Chen YY et al (2009) Protocol for the examination of specimens from patients with invasive carcinoma of the breast. Arch Pathol Lab Med 133:1515–1538PubMed
24.
Zurück zum Zitat Senkus E, Kyriakides S, Penault-Llorca F et al (2013) Primary breast cancer: ESMO clinical practice guidelines for diagnosis, treatment and follow-up. Ann Oncol 24(Suppl 6):vi7–v23CrossRef Senkus E, Kyriakides S, Penault-Llorca F et al (2013) Primary breast cancer: ESMO clinical practice guidelines for diagnosis, treatment and follow-up. Ann Oncol 24(Suppl 6):vi7–v23CrossRef
25.
Zurück zum Zitat DeLong ER, DeLong DM, Clarke-Pearson DL (1988) Comparing the areas under two or more correlated receiver operating characteristic curves: a nonparametric approach. Biometrics 44:837–845CrossRef DeLong ER, DeLong DM, Clarke-Pearson DL (1988) Comparing the areas under two or more correlated receiver operating characteristic curves: a nonparametric approach. Biometrics 44:837–845CrossRef
26.
Zurück zum Zitat Chung MA, DiPetrillo T, Hernandez S, Masko G, Wazer D, Cady B (2002) Treatment of the axilla by tangential breast radiotherapy in women with invasive breast cancer. Am J Surg 184:401–402CrossRef Chung MA, DiPetrillo T, Hernandez S, Masko G, Wazer D, Cady B (2002) Treatment of the axilla by tangential breast radiotherapy in women with invasive breast cancer. Am J Surg 184:401–402CrossRef
27.
Zurück zum Zitat Rabinovitch R, Ballonoff A, Newman F, Finlayson C (2008) Evaluation of breast sentinel lymph node coverage by standard radiation therapy fields. Int J Radiat Oncol Biol Phys 70:1468–1471CrossRef Rabinovitch R, Ballonoff A, Newman F, Finlayson C (2008) Evaluation of breast sentinel lymph node coverage by standard radiation therapy fields. Int J Radiat Oncol Biol Phys 70:1468–1471CrossRef
28.
Zurück zum Zitat Schlembach PJ, Buchholz TA, Ross MI et al (2001) Relationship of sentinel and axillary level I-II lymph nodes to tangential fields used in breast irradiation. Int J Radiat Oncol Biol Phys 51:671–678CrossRef Schlembach PJ, Buchholz TA, Ross MI et al (2001) Relationship of sentinel and axillary level I-II lymph nodes to tangential fields used in breast irradiation. Int J Radiat Oncol Biol Phys 51:671–678CrossRef
29.
Zurück zum Zitat van Roozendaal LM, Schipper RJ, Smit LH et al (2015) Three-dimensional breast radiotherapy and the elective radiation dose at the sentinel lymph node site in breast cancer. Ann Surg Oncol 22:3824–3830CrossRef van Roozendaal LM, Schipper RJ, Smit LH et al (2015) Three-dimensional breast radiotherapy and the elective radiation dose at the sentinel lymph node site in breast cancer. Ann Surg Oncol 22:3824–3830CrossRef
30.
Zurück zum Zitat Fornasa F, Nesoti MV, Bovo C, Bonavina MG (2012) Diffusion-weighted magnetic resonance imaging in the characterization of axillary lymph nodes in patients with breast cancer. J Magn Reson Imaging 36:858–864CrossRef Fornasa F, Nesoti MV, Bovo C, Bonavina MG (2012) Diffusion-weighted magnetic resonance imaging in the characterization of axillary lymph nodes in patients with breast cancer. J Magn Reson Imaging 36:858–864CrossRef
31.
Zurück zum Zitat Nakai G, Matsuki M, Harada T et al (2011) Evaluation of axillary lymph nodes by diffusion-weighted MRI using ultrasmall superparamagnetic iron oxide in patients with breast cancer: initial clinical experience. J Magn Reson Imaging 34:557–562CrossRef Nakai G, Matsuki M, Harada T et al (2011) Evaluation of axillary lymph nodes by diffusion-weighted MRI using ultrasmall superparamagnetic iron oxide in patients with breast cancer: initial clinical experience. J Magn Reson Imaging 34:557–562CrossRef
32.
Zurück zum Zitat Kamitani T, Hatakenaka M, Yabuuchi H et al (2013) Detection of axillary node metastasis using diffusion-weighted MRI in breast cancer. Clin Imaging 37:56–61CrossRef Kamitani T, Hatakenaka M, Yabuuchi H et al (2013) Detection of axillary node metastasis using diffusion-weighted MRI in breast cancer. Clin Imaging 37:56–61CrossRef
33.
Zurück zum Zitat Luo NB, Su DK, Jin GQ et al (2013) Apparent diffusion coefficient ratio between axillary lymph node with primary tumor to detect nodal metastasis in breast cancer patients. J Magn Reson Imaging 38:824–828CrossRef Luo NB, Su DK, Jin GQ et al (2013) Apparent diffusion coefficient ratio between axillary lymph node with primary tumor to detect nodal metastasis in breast cancer patients. J Magn Reson Imaging 38:824–828CrossRef
34.
Zurück zum Zitat Kim EJ, Kim SH, Kang BJ, Choi BG, Song BJ, Choi JJ (2014) Diagnostic value of breast MRI for predicting metastatic axillary lymph nodes in breast cancer patients: diffusion-weighted MRI and conventional MRI. Magn Reson Imaging 32:1230–1236CrossRef Kim EJ, Kim SH, Kang BJ, Choi BG, Song BJ, Choi JJ (2014) Diagnostic value of breast MRI for predicting metastatic axillary lymph nodes in breast cancer patients: diffusion-weighted MRI and conventional MRI. Magn Reson Imaging 32:1230–1236CrossRef
35.
Zurück zum Zitat Dorrius MD, Dijkstra H, Oudkerk M, Sijens PE (2014) Effect of b value and pre-admission of contrast on diagnostic accuracy of 1.5-T breast DWI: a systematic review and meta-analysis. Eur Radiol 24:2835–2847CrossRef Dorrius MD, Dijkstra H, Oudkerk M, Sijens PE (2014) Effect of b value and pre-admission of contrast on diagnostic accuracy of 1.5-T breast DWI: a systematic review and meta-analysis. Eur Radiol 24:2835–2847CrossRef
36.
Zurück zum Zitat Li CM, Meng S, Yang XH, Wang J, Hu JN (2014) The value of T2*in differentiating metastatic from benign axillary lymph nodes in patients with breast cancer - a preliminary in vivo study. PLoS One 9 Li CM, Meng S, Yang XH, Wang J, Hu JN (2014) The value of T2*in differentiating metastatic from benign axillary lymph nodes in patients with breast cancer - a preliminary in vivo study. PLoS One 9
37.
Zurück zum Zitat LeCun Y, Bengio Y, Hinton G (2015) Deep learning. Nature 521:436–444CrossRef LeCun Y, Bengio Y, Hinton G (2015) Deep learning. Nature 521:436–444CrossRef
Metadaten
Titel
Diagnostic performance of standard breast MRI compared to dedicated axillary MRI for assessment of node-negative and node-positive breast cancer
verfasst von
Sanaz Samiei
Marjolein L. Smidt
Sigrid Vanwetswinkel
Sanne M. E. Engelen
Robert-Jan Schipper
Marc B. I. Lobbes
Thiemo J. A. van Nijnatten
Publikationsdatum
27.03.2020
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 8/2020
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-020-06760-6

Weitere Artikel der Ausgabe 8/2020

European Radiology 8/2020 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.