Diaphragm echodensity in mechanically ventilated patients: a description of technique and outcomes

verfasst von: Benjamin Coiffard, Stephen Riegler, Michael C. Sklar, Martin Dres, Stefannie Vorona, W. Darlene Reid, Laurent J. Brochard, Niall D. Ferguson, Ewan C. Goligher

Erschienen in: Critical Care | Ausgabe 1/2021

Abstract

Background

Acute increases in muscle sonographic echodensity reflect muscle injury. Diaphragm echodensity has not been measured in mechanically ventilated patients. We undertook to develop a technique to characterize changes in diaphragm echodensity during mechanical ventilation and to assess whether these changes are correlated with prolonged mechanical ventilation.

Methods

Diaphragm ultrasound images were prospectively collected in mechanically ventilated patients and in 10 young healthy subjects. Echodensity was quantified based on the right-skewed distribution of grayscale values (50th percentile, ED50; 85th percentile, ED85). Intra- and inter-analyzer measurement reproducibility was determined. Outcomes recorded included duration of ventilation and ICU complications (including reintubation, tracheostomy, prolonged ventilation, or death).

Results

Echodensity measurements were obtained serially in 34 patients comprising a total of 104 images. Baseline (admission) diaphragm ED85 was increased in mechanically ventilated patients compared to younger healthy subjects (median 56, interquartile range (IQR) 42–84, vs. 39, IQR 36–52, p = 0.04). Patients with an initial increase in median echodensity over time (≥ + 10 in ED50 from baseline) had fewer ventilator-free days to day 60 (n = 13, median 46, IQR 0–52) compared to patients without this increase (n = 21, median 53 days, IQR 49–56, unadjusted p = 0.03). Both decreases and increases in diaphragm thickness during mechanical ventilation were associated with increases in ED50 over time (adjusted p = 0.03, conditional R2 = 0.80) and the association between increase in ED50 and outcomes persisted after adjusting for changes in diaphragm thickness.

Conclusions

Many patients exhibit increased diaphragm echodensity at the outset of mechanical ventilation. Increases in diaphragm echodensity during the early course of mechanical ventilation are associated with prolonged mechanical ventilation. Both decreases and increases in diaphragm thickness during mechanical ventilation are associated with increased echodensity.
Begleitmaterial
Hinweise
Benjamin Coiffard and Stephen Riegler are equally contributed to this work

Supplementary Information

The online version contains supplementary material available at https://​doi.​org/​10.​1186/​s13054-021-03494-9.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
ED50
echodensity at the 50th percentile
ED85
echodensity at the 85th percentile
HEA65
high echodensity area above grayscale value of 65
ICU
intensive care unit
IQR
interquartile ranges
SAPS II
simplified acute physiology score II
SAS
riker sedation-agitation scale
SOFA
sequential organ failure assessment

Background

Point-of-care ultrasound is used to evaluate skeletal muscle structure and function in critically ill patients [1, 2]. Sonography allows assessment of muscle quantity (thickness), muscle contractility (contractile thickening and shortening) and muscle quality (echodensity). Changes in muscle echodensity (also referred to as echogenicity) can be assessed using grayscale analysis to quantify changes in muscle echotexture [3, 4]. Muscle echodensity (the sonographic property of signal reflection) is normally low. Healthy muscle tissue usually appears dark, almost black, because it contains little fibrous tissue with minimal sound reflection. In disease, replacement of muscle with fat or fibrous tissue increases echodensity and muscle appears ‘brighter’ [57]. Moreover, increases in echodensity correlate with muscle fiber degeneration and necrosis in experimental animal models of muscle injury [8, 9]. Increases in echodensity develop with acute muscle injury in athletes, with progression of chronic muscular disease states such as muscular dystrophy, and with muscular inflammation, necrosis, and weakness in critically ill patients [8, 1012].
Diaphragm structure and function are known to deteriorate during mechanical ventilation. Both acute decreases and increases in diaphragm thickness during mechanical ventilation have been documented repeatedly using point-of-care ultrasound [1315]; these sonographic findings are associated with poor clinical outcomes [16]. It is unknown whether diaphragm echodensity changes during mechanical ventilation and whether this provides additional independent information about clinically important changes in muscle structure and function during mechanical ventilation and critical illness. Similar assessments have been performed for analyzing the quadriceps echodensity in intensive care unit (ICU) patients and we adapted these measurements to assess diaphragm echodensity [8].
In a previously published cohort study of diaphragm ultrasound [16], we undertook to develop a technique for assessing diaphragm echodensity in ventilated patients, to characterize the evolution of diaphragm echodensity over time during the early course of mechanical ventilation, and to assess its relation to clinical outcome.

Methods

Study population and setting

This study was a secondary analysis of a previously published cohort study in mechanically ventilated patients at Toronto General Hospital and St. Michael’s Hospital, located in Toronto, Canada. The study was designed to assess changes in diaphragm thickness over time; during the latter portion of the study we proposed to assess changes in diaphragm echodensity during mechanical ventilation and began collecting B mode images for echodensity analysis.
Patients were enrolled within 36 h of intubation but were ineligible if liberation from mechanical ventilation was expected within 24 h, or if they had received more than 48 h of mechanical ventilation in the past 6 months. The cohort is described in detail elsewhere [16].
Patients from this cohort were included in the present analysis if the following criteria were met: (a) stored B-mode images were available for analysis; (b) the entirety of the pleural and peritoneal membranes were clearly demarcated across the entire image, allowing the diaphragm to be distinguished from surrounding tissues with confidence; (c) images of acceptable quality (criteria a and b) were available for more than one study day; (d) use of consistent gain and frequency settings across the images obtained from the same patient over time. All diaphragm images up to day 5 were analysed.
In addition, 10 healthy subjects (non-smoking, no history of cardio-pulmonary disease) were enrolled to form a control group.

Measurement of diaphragm thickness and echodensity

The method to collect diaphragm ultrasound images, to measure diaphragm thickness, and to calculate diaphragm thickening fraction was described in the original paper [16, 17]. B-mode images were collected as close to end-expiration as possible. To standardize ultrasound gain and frequency for echodensity measurements, B-mode images were obtained after restoring the ultrasound device settings to start-up, pre-set default values. Over time, three different ultrasound machines were used (Phillips Sparq, Mindray, Fujifilm Sonosite), but the same machine was used for all images collected from each individual patient.
Diaphragm echodensity was quantified by performing a grayscale histogram analysis in ImageJ (National Institutes of Health, Bethesda, MD, USA) (Additional file 1: A). Grayscale histogram analysis to quantify echodensity has been described elsewhere [1820]; values range between 0 (black) and 255 (white). The analysis was performed using the trace method [20] by selecting the largest free-form area devoid of artifacts between (but excluding) the pleural and peritoneal membranes (Additional file 1: B). A grayscale frequency histogram was generated for the selected region. As depicted in Additional file 1: C, the distribution of echodensity for the selected region is right-skewed and was quantified using two different parameters: the 50th percentile (ED50), and the 85th percentile (ED85). The 50th and 85th percentile thresholds were chosen as they were deemed to represent the center and upper tail of the grayscale distribution, respectively, for each image. We also defined the percentage of pixels above a grayscale value of 65 (high echodensity area, HEA65). The value of 65 (upper limit of normal for echodensity) was determined based on the 95th percentile grayscale value of the average grayscale histogram obtained from the ten healthy subjects (see “Results”).

Technical assessment of measurement reliability

To assess the repeatability and reproducibility of echodensity measurements, the principal analyzer trained a secondary analyzer using 30 randomly selected images. Both analyzers then analyzed an additional 30 randomly selected images. Intra-analyzer repeatability and inter-analyzer reproducibility of ED50 were quantified by the method of Bland and Altman [21].
To assess whether echodensity measurements were affected by the timing of the respiratory cycle (which may be difficult to standardize in B-mode), echodensity measurements were obtained on single frames representing end-expiration and end-inspiration in 15 randomly selected DICOM files capturing an entire respiratory cycle.

Patient characteristics and outcomes

Demographic data, comorbidities, admission diagnosis, baseline severity of illness (Simplified Acute Physiology Score [SAPS] II) [22], ventilator settings, arterial blood gas tensions, criteria for sepsis [23], Riker Sedation-Agitation Scale (SAS) [24], exposure to neuromuscular blockade, and Sequential Organ Failure Assessment (SOFA) [25] scores were extracted from the study database. The following outcomes were also extracted: extubation, reintubation, tracheostomy, ICU discharge, hospital discharge, and death. Liberation from ventilation was defined as separation from the ventilator (extubation or tracheostomy mask breathing for 24 h) without resumption of invasive ventilatory support during the ICU admission.
The duration of ventilation was the time from intubation until liberation from ventilation (or death). Ventilator-free days were computed to 60 days; patients who required more than 60 days of ventilatory support or who died on or before day 60 were assigned 0 days. Complications of acute respiratory failure were defined as the occurrence of any of the following events: reintubation, tracheostomy, prolonged ventilation (> 14 days), or death [16].
Investigators responsible for analysis of diaphragm ultrasound images were blinded to patient outcomes. Clinicians responsible for medical decisions including weaning were not aware of ultrasound measurement data. Routine weaning practices were similar across participating ICUs but were not uniformly standardized for the study [16].

Statistical analysis

Continuous variables were expressed as mean values (± standard deviation) or median values with interquartile ranges (IQR), according to the distribution (Shapiro–Wilk test). Discrete variables are expressed as percentage values.
Associations between clinical characteristics or outcomes and baseline median echodensity at baseline were assessed using linear or logistic regression models. Owing to the relatively lower number of patients in whom echodensity measurements were available, no multivariable adjustments were performed.
Patients in whom ED50 increased by at least 10 points in grayscale value from baseline at any time over the first 5 days of ventilation were classified as having an increase in echodensity. Patients who did not develop a 10-point increase at any time over the first five days were classified as unchanged. This threshold for categorization (10-point increase in ED50) was selected based on the inter-analyzer limit of agreement for ED50 measurement in this study (see “Results”). To mitigate against time-dependent confounding, patients were classified on the first day that the change in ED50 exceeded + 10 from baseline ED50. Comparisons between the two groups were performed using the Mann–Whitney test for continuous variables and using a chi-squared test for categorical variables. The relationship between the development of an increase in echodensity during mechanical ventilation and outcomes was assessed by linear and logistic regression. To address potential confounding, the association between echodensity and prolonged mechanical ventilation was adjusted for the initial change in diaphragm thickness from baseline. The relationship between decreases or increases in diaphragm thickness with ED50 change over time was fit using restricted cubic splines (3 knots) in linear regression, adjusted for daily fluid balance.
To address potential immortal time bias we performed a sensitivity analysis restricted to patients who remained ventilated at least 3 days (to only analyze diaphragm images while under mechanical ventilation which is the main suspected risk factor for diaphragm injury).
All statistical analyses were conducted using R version 3.5.1 (www.​R-project.​org).

Results

Population

Of 191 mechanically ventilated patients in the cohort, echodensity measurements were attempted for the last 41 patients. Of these, images were available for 34 patients comprising a total of 104 diaphragm images over the study period (4 patients had only one day of imaging, 1 patient was excluded due to inconsistent gain and frequency settings, 2 patients were excluded due to low-quality images). Patient characteristics are reported in Table 1. The healthy control group was comprised of 5 women and 5 men with a median age of 27 years, IQR 25–34 years.
Table 1
Baseline and change echodensity over time in relation to patient characteristics
Characteristics
Population
Relation with baseline ED50
Relation with change in ED50 from baseline
n = 34
Coefficient
CI 95%
p
 ≤ 10-point change in ED50, n = 21 (62%)
 > 10-point increase in ED50, n = 13 (38%)
p
Age, yr, mean (SD)
59 (16)
0.13
(− 0.46; 0.71)
0.68
59 (16)
59 (17)
0.94
Female sex, n (%)
10 (30)
9
(− 29; 11)
0.39
5 (24)
5 (38)
0.60
Co-morbidities, n (%)
 COPD
8 (24)
− 15
(− 37; 7)
0.20
4 (19)
4 (31)
0.62
 Asthma
3 (9)
35
(4; 67)
0.04
1 (5)
2 (15)
0.62
 ILD
10 (30)
− 20
(− 40; 0)
0.05
6 (29)
4 (31)
1.00
 OSA
3 (9)
− 8
(− 42; 26)
0.64
3 (14)
0 (0)
0.46
 CHF
2 (6)
− 15
(− 56; 21)
0.47
1 (5)
1 (8)
1.00
 Cirrhosis
5 (15)
34
(10; 58)
0.01
3 (14)
2 (15)
1.00
 CKD
6 (18)
5
(− 20; 29)
0.71
3 (14)
3 (23)
0.85
 Immunocompromise
10 (30)
16
(− 4; 37)
0.13
5 (24)
5 (38)
0.50
 Diabetes
10 (30)
4
(− 17; 25)
0.71
5 (24)
5 (38)
0.50
SOFA, mean over first 72 h, mean (SD)
11 (3)
0.49
(− 2.4; 3.4)
0.74
11 (3)
11 (3)
0.98
SAPS II, mean (SD)
52 (15)
− 0.09
(− 0.73; 0.55)
0.78
52 (16)
51 (13)
0.80
Primary cause of acute respiratory failure, n (%)
 Cardiovascular
2 (6)
Reference
1 (5)
1 (8)
0.52
  
 Respiratory
7 (21)
15
(− 28; 58)
0.33
4 (19)
3 (23)
 
 Sepsis (nonpulmonary)
5 (15)
12
(− 33; 56)
 
3 (14)
2 (15)
 
 Neurologic
2 (6)
45
(− 8; 99)
 
1 (5)
1 (8)
 
 Postoperative
6 (18)
− 5
(− 48; 39)
 
6 (29)
0 (0)
 
 Post-transplantation
9 (26)
− 1
(− 42; 41)
 
5 (24)
4 (31)
 
 Other (hepatic, renal)
3 (9)
− 5
(− 54; 43)
 
1 (5)
2 (15)
 
 Use of NIV prior to intubation
4 (12)
3
(− 26; 32)
0.82
2 (10)
2 (15)
0.96
Primary cause of acute respiratory failure, n (%)
 Non-respiratory
15 (44)
Reference
8 (38)
7 (54)
0.59
  
 Respiratory
19 (56)
− 9
(− 28; 9)
0.34
13 (62)
6 (46)
 
Sepsis-3 criteria present in first 48 h, n (%)
31 (91)
16
(− 17; 49)
0.35
19 (90)
12 (92)
1.00
Comorbidity at baseline (≥ 1), n (%)
26 (76)
8
(− 14; 30)
0.47
14 (67)
12 (92)
0.19
Baseline PaO2/FiO2, mmHg
159 (116–241)
0.06
(− 0.02; 0.15)
0.17
138 (105–212)
168 (124–253)
0.29
Cumulative fluid balance on Day 1 of ventilation, L
0.8 (− 0.3–4.5)
− 0.2
(− 1.4; 1.0)
0.72
2.1 (0.1–5.1)
0.3 (− 0.7–0.9)
0.36
Cumulative fluid balance on Day 3 of ventilation, L, mean (SD)
4.1 (5.2)
1.2
(− 1; 3)
0.24
3.7 (6.2)
4.6 (3.5)
0.60
Initial diaphragm thickness, mm
2.3 (2.1–2.5)
16
(− 3; 35)
0.10
2.3 (2.1–2.6)
2.2 (1.9–2.4)
0.49
Diaphragm thickening fraction (%), mean over first 72 h
9 (6–15)
− 0.4
(− 1.5; 0.8)
0.56
9 (6–15)
10 (9–16)
0.31
Baseline ventilator settings
Mode of ventilaton, n (%)
 Controlled
32 (94)
Reference
20 (95)
12 (92)
1.0
  
 Partially assisted
2 (6)
− 27
(− 66; 12)
0.18
1 (5)
1 (8)
 
Ventilator settings
 Vt, ml/kg PBW
6.8 (6.2–7.2)
− 0.7
(− 1.6; 0.2)
0.15
7.0 (6.4–7.4)
6.7 (5.7–7.0)
0.18
 Peak airway pressure, cmH2O, mean (SD)
23 (5)
− 0.6
(− 2.0; 0.9)
0.45
23 (5)
24 (4)
0.52
 Set inspiratory pressure above PEEP, cmH2O*
14 (13–16)
1.2
(− 1.5; 3.9)
0.39
14 (13–16)
15 (14–18)
0.41
 PEEP, cmH2O
8 (6–10)
− 0.4
(− 3.2; 2.3)
0.75
8 (5–10)
8 (8–10)
0.84
 Frequency, min-1, mean (SD)
20 (5)
0.7
(− 1.0; 2.3)
0.43
20 (4)
20 (6)
0.80
 FiO2
0.5 (0.4–0.6)
− 0.2
(− 3.0; 2.6)
0.90
0.5 (0.4–0.6)
0.5 (0.4–0.8)
0.93
Arterial blood gas tensions
 pH
7.36 (7.31–7.39)
− 0.09
(− 1.4; 1.2)
0.90
7.36 (7.31–7.41)
7.37 (7.32–7.39)
0.84
 PaCO2, mmHg, mean (SD)
42 (7)
− 0.4
(− 1.7; 0.8)
0.50
42 (7)
41 (7)
0.54
 PaO2, mmHg, mean (SD)
104 (25)
− 0.6
(− 1.6; 0.3)
0.19
98 (25)
111 (24)
0.14
The respiratory primary cause of acute respiratory failure comprises lung transplantation, postoperative pulmonary endarterectomy, chronic lung rejection, exacerbation of interstitial lung disease, pneumonia, and aspiration; non-respiratory are the other diagnoses
CHF = chronic heart failure; CKD = chronic kidney disease; COPD = chronic obstructive pulmonary disease; ILD = interstitial lung disease; NIV = non-invasive ventilation; OSA = obstructive sleep apnea; PBW = predicted body weight; PEEP = positive end-expiratory pressure; SAPS = Simplified Acute Physiology Score; SAS = Riker Sedation-Agitation Scale; SOFA = Sequential Organ Failure Assessment
All numerical distributions are reported as median (interquartile range) unless otherwise noted
*Set inspiratory pressure above PEEP = peak airway pressure—PEEP

Intra- and inter-analyzer reproducibility of echodensity measurements

ED50 was mean 44 (standard deviation 27) across all mechanically ventilated patients at ICU admission. The average difference in ED50 between analyzers was -1.5 (limits of agreement -9 to 6) (Additional file 2: panel A). The average difference in echodensity between images within the same analyzer was -3 (limits of agreement -16–10) (Additional file 2: panel B). Within a single respiratory cycle, ED50 measured by the same analyzer differed between end-expiration and peak inspiration by mean -1 (standard deviation 4) (Additional file 2: panel C).
Reproducibility according to the ultrasound machines (Philips, Mindray, Sonosite) is described in Additional file 3.

Diaphragm echodensity in healthy subjects and ventilated patients at baseline

Compared to healthy subjects (Fig. 1), mechanically ventilated patients exhibited higher baseline ED50 (median 27, IQR 23–34, vs 40, IQR 23–55, p = 0.07 for comparison) and higher baseline ED85 (median 39, IQR 36–52, vs 56, IQR 42–84, p = 0.04 for comparison) at the time of ICU admission. The 95th percentile of the average grayscale distribution in the healthy subjects was 65. On this basis, a grayscale value of 65 was taken as the upper limit of normal for echodensity. Eighteen patients (53%) had abnormally increased echodensity at baseline (i.e. patients with more than 5% of pixels above a grayscale value of 65 and defined as HEA65 > 5%).
In mechanically ventilated patients, baseline ED50 was not associated with patient characteristics (Table 1). Baseline ED50 was not associated with clinical outcomes such as ventilator-free days to day 60, duration of ICU admission, or death (Table 2).
Table 2
Clinical Outcomes in relation to baseline and change echodensity over time
Outcome
Population
Relation with baseline ED50
Relation with change in ED50 from baseline
n = 34
Coefficient
CI 95%
p
 ≤ 10-point change in ED50, n = 21 (62%)
 > 10-point increase in ED50, n = 13 (38%)
p
Duration of ventilation (in ICU survivors), d
7 (4–11)
− 0.004
(− 0.02; 0.01)
0.55
5 (4–10)
10 (8–14)
0.10
Ventilator-free days to Day 60, d
51 (36–56)
− 0.001
(− 0.01; 0.01)
0.78
53 (49–56)
46 (0–52)
0.03
Duration of ICU admission (in ICU survivors), d
10 (5–15)
− 0.002
(− 0.02; 0.01)
0.72
10 (4–15)
11 (10–15)
0.18
Duration of hospitalization (in hospital survivors), d
26 (17–54)
0.003
(− 0.01; 0.01)
0.44
26 (16–53)
27.5 (22–72)
0.46
Maximal thickening fraction at first SBT (%), mean (SD)
29 (16)
0.001
(− 0.03; 0.03)
0.83
33 (18)
26 (13)
0.26
Complications of acute respiratory failure, n (%)
15 (44)
0.02
(− 0.01; 0.05)
0.27
7 (33)
8 (61)
0.21
Reintubation, n (%)
3 (9)
− 0.006
(− 0.07; 0.03)
0.81
1 (5)
2 (15)
0.66
Tracheostomy, n (%)
9 (26)
0.02
(− 0.01; 0.05)
0.28
5 (24)
4 (31)
0.96
Mechanical ventilation > 7 d, n (%)
18 (53)
0.02
(− 0.01; 0.05)
0.19
7 (33)
11 (85)
0.01
Readmission to ICU during same hospital admission, n (%)
1 (3)
− 0.04
(− 0.23; 0.04)
0.53
0 (0)
1 (8)
0.81
Death in ICU, n (%)
6 (18)
0.009
(− 0.02; 0.04)
0.55
2 (10)
4 (31)
0.26
Death in hospital, n (%)
7 (21)
0.01
(− 0.02; 0.04)
0.47
2 (10)
5 (38)
0.11
All numerical distributions are reported as median (interquartile range)

Evolution of diaphragm echodensity

A median of 3 images, IQR [24], were available per patient. The evolution of diaphragm echodensity varied widely among patients over time during mechanical ventilation (Additional files 4 and 5, and Fig. 2). Increased diaphragm echodensity (defined a priori as > 10-point increase in ED50 from baseline) developed in 13 patients (38%) at a median of day 3 after intubation, IQR [24]). In these patients, the median maximal increase in ED50 reached over the first 5 days of ventilation was + 18 points (IQR + 16 to + 26 points). There was no association between patient characteristics including severity of illness score, cumulative fluid balance, or ventilator settings and the development of increased echodensity (Table 1).
Both decreases and increases in diaphragm thickness from baseline were associated with increases in ED50 over time (p = 0.03, within-subjects R2 = 0.78), even after adjustment for cumulative fluid balance at day 3 (adjusted p = 0.03, within-subjects R2 = 0.80, Fig. 3).
In comparison to patients with unchanged echodensity, patients who developed increased echodensity had fewer ventilator-free days to day 60 (median 46, IQR 0–52, versus median 53, IQR 49–56, p = 0.03) and more frequently required ventilation for ≥ 7 days (85% vs. 33%, p = 0.01) (Table 2). There was a trend towards higher mortality in patients with increased ED50 (38% vs. 10%, p = 0.11). Patients who required mechanical ventilation for more than 7 days exhibited significant increases in ED50 on day 2 of mechanical ventilation (+ 40% increase from baseline, IQR + 4% to + 54%) in comparison to patients who required ventilation for less than 7 days (− 9% change in ED50 from baseline on day 2, IQR − 45% to + 4%, p = 0.007 for comparison) (Fig. 2). The association between increased echodensity and prolonged ventilation for ≥ 7 days persisted after adjusting for the change in diaphragm thickness from baseline (adjusted p = 0.01).
In a sensitivity analysis restricting to patients who remained ventilated for at least 3 days (n = 33) to address potential immortal time bias, increased ED50 developing within 3 days of ventilation (n = 8/33, 24%) was associated with fewer ventilator-free days in comparison to patients with unchanged echodensity (median 48 days, IQR [0–50] vs. median 52 days, IQR [44–56] respectively, p = 0.09). Among patients ventilated for at least 3 days, 88% of patients who developed increased ED50 within 3 days of ventilation (n = 8) required prolonged ventilation (≥ 7 days) whereas 44% of patients with no increase in ED50 (n = 25) required prolonged ventilation (p = 0.08).

Discussion

In this preliminary description of diaphragm echodensity in mechanically ventilated patients, we found that diaphragm echodensity measurements were feasible and highly reproducible in mechanically ventilated patients. Diaphragm echodensity appears to be increased in many mechanically ventilated patients (but not all) in comparison to younger healthy subjects. Increases in diaphragm echodensity developed during the early course of mechanical ventilation in a substantial proportion of patients and these increases were associated with prolonged mechanical ventilation. Both decreases and increases in diaphragm thickness were correlated with increases in echodensity over time. Taken together, these findings suggest that sonographic measurements of echodensity represent a novel early marker of potentially important structural changes in the diaphragm associated with critical illness.
The technique for measuring echodensity of the diaphragm evaluated in this study proved highly feasible and yields results with acceptable reproducibility within and between analyzers. Echodensity was not significantly influenced by the timing of image acquisition in the respiratory cycle. One of the challenges of conducting such image analysis is to identify a single parameter that best reflects the magnitude of echodensity. Most previous studies used the mean value of the grayscale intensity as the principal marker of echodensity [3, 8, 20, 2628]. However, the mean grayscale value may not be the most appropriate marker if the distribution of grayscale values is skewed. In our study, we employed a method similar to that used in CT-scan studies to analyze lung densitometry using the Hounsfield unit scale [29]. We assessed a number of different parameters including percentile grayscale values (50th and 85th percentiles, ED50 and ED85, respectively) and areas under the curve. Based on the 95th percentile of an average distribution curve produced by averaging the grayscale density distributions of healthy subjects, we identified a grayscale value of 65 as the upper limit of normal. On this basis, we reported HEA65 (‘high echodensity area’ above 65), as the proportion of grayscale values above 65. These various parameters correlated quite closely. As the distribution median is the simplest and most familiar measurement, we chose to use ED50 as the primary measure of echodensity in our analysis.
Multiple previous studies have demonstrated important structural changes in the diaphragm during the early course of mechanical ventilation: decreases in diaphragm thickness suggestive of disuse atrophy [1316] and increases in diaphragm thickness raising the hypothesis of diaphragm myotrauma from excess loading [16, 30]. These changes are associated with an impairment in diaphragm function and poor clinical outcome. To our knowledge, this is the first study to describe changes in echodensity of the diaphragm during mechanical ventilation. As we did not obtain diaphragm muscle biopsies or measure diaphragm function in these patients, the precise pathophysiological significance of these changes in echodensity—and the potential mechanistic basis for their association to prolonged ventilation—is unclear. However, radiological-pathological correlations obtained in studies of other muscles may provide insight.
Several considerations support the hypothesis that an increase in diaphragm echodensity may reflect injury to the diaphragm during critical illness. Studies in other skeletal muscles have shown a strong correlation between increased echodensity and inflammation or muscle injury confirmed with muscle biopsy [811]. In an experimental model of calf muscle injury in rats, the muscle degenerative phase was characterized by increased echogenicity in the injured area for up to 20 days. After the initial phase of muscle injury or myotrauma, muscle tissue may develop fibrotic changes affecting early or long-term function [9]. In neurodegenerative diseases, muscle fibrosis and dystrophy assessed by histology strongly correlated with increased echodensity [5, 26]. In mechanically ventilated patients, increases in quadriceps echodensity over time was correlated with inflammation and myonecrosis in quadriceps muscle biopsies [8]. In that study, half of the patients exhibited increased quadriceps echodensity as soon as day 1 or day 3 after ICU admission and this predicted the severity of myofiber necrosis by days 7 or 10.
Alternatively, it is possible that changes in echodensity may signify the accumulation of tissue edema [8] related to resuscitation. Changes in echodensity were unrelated to the fluid balance in this study, though edema resulting from capillary leak in the context of a systemic inflammatory response may not be perfectly captured by the fluid balance.
In the present study, an early increase of diaphragm echodensity over time during mechanical ventilation was associated with a longer duration of ventilation. It is interesting to note that changes in diaphragm echodensity appeared mostly by day 2 after ICU admission, highlighting a rapid process most likely related to the primary cause of the acute respiratory failure or possibly resulting from injurious (insufficient or excessive) respiratory effort under mechanical ventilation. Conversely, echodensity decreased from baseline in some patients over the first several days of mechanical ventilation, which could in theory suggest a rapid recovery of the initial early diaphragm injury. The possibility of injury to the lung from excess respiratory effort before intubation (referred to as patient self-inflicted lung injury [31]) has been the topic of considerable debate of late in the context of COVID-19 management. The potential for the patient to sustain an acute load-induced injury to the diaphragm before intubation is less widely appreciated. Previous studies have convincingly demonstrated that the diaphragm is vulnerable to acute load-induced injury (e.g., underassistance myotrauma from insufficient ventilator support) [32, 33] and Vassilakopoulos and colleagues have shown an acute increase in systemic inflammation derived from the diaphragm during resistive loading [34] that is relieved after initiating controlled ventilation [35]. We speculate that the increased echodensity observed at baseline in some patients in this study may potentially reflect load-induced injury sustained prior to intubation which may then progressively resolve with the institution of mechanical ventilatory support.
Increases in echodensity over time were associated with changes in diaphragm thickness, suggesting that both acute decreases and acute increases in diaphragm thickness may reflect underlying deleterious changes in muscle structure. We observed increased echodensity at admission (compared to healthy subjects) and early changes during the ICU course. Thus, early change in diaphragm echodensity could be an interesting early marker of diaphragm injury to be studied as part of future diaphragm-protective ventilation trials [36]. The relation between changes in echodensity and outcome was independent of the change in thickness (itself associated with prolonged ventilation [16]), suggesting that this measure may have added prognostic value compared to diaphragm thickness at a given time.
This study has several important limitations. Adjustable settings such as gain or depth can have an effect on the grayscale value of the image produced [27] and ideally should be standardized for all images taken in all patients. The angle at which the probe is held, the pressure with which the probe is held against the body, as well as the amount of subcutaneous adipose tissue or edema between the probe and the muscle are all recognized as potential factors that affect the image processing (and thus the grayscale value of the image produced) [3, 3739]. To minimize measurement noise within patients related to the ultrasound settings, we selected images with consistent gain and depth in each patient and analyzed echodensity changes over time expressed relative to the baseline value in each patient. However, it was impossible to obtain the exact same settings between patients, limiting the validity of comparison between patients.
Reproducibility was assessed within and between analyzers from a single diaphragm image. Inter-observer reproducibility for ED measurement from different ultrasound images remains to be determined in future studies. Because this was a secondary analysis of a convenience sample from a previous cohort study, a sample size calculation was not performed. We included all patients for whom we possessed stored images of high quality with consistent settings (i.e. unchanged gain or frequency settings between different study days). The results of this are hypothesis-generating and require confirmation in future prospective studies designed to further correlate changes in diaphragm echodensity to markers of systemic inflammation (cytokines, etc.), diaphragm histology, diaphragm muscle function, and clinical outcomes.
The systematic difference in echodensity between mechanically ventilated patients and healthy subjects may arise from differences in age or chronic comorbidity between groups, apart from any acute muscle injury in mechanically ventilated patients. Both comorbidity and age have been shown to affect muscle echodensity [8, 28, 40]. These factors might account for some of the observed differences between patients and healthy subjects in this study; to mitigate confounding in this comparison, we analyzed the change in echodensity within subjects over time and normalized patients to their baseline echodensity.

Conclusion

Diaphragm echodensity can be measured feasibly and reproducibly in mechanically ventilated patients. In this study, diaphragm echodensity was increased in mechanically ventilated patients compared to healthy subjects and increases in echodensity over time were associated with prolonged mechanical ventilation. Both decreases and increases in diaphragm thickness over time were associated with increases in echodensity over time. Diaphragm echodensity merits further investigation as a potential clinically relevant marker of muscle injury during critical illness.

Supplementary Information

The online version contains supplementary material available at https://​doi.​org/​10.​1186/​s13054-021-03494-9.

Acknowledgements

None.
The Research Ethics Boards at University Health Network (#12–5582, #13–5953, #15–8998) and St. Michael’s Hospital (#14–229) approved the study protocols. All patient enrolled in this study or their next of kin signed an informed consent.
Available.

Competing interests

Dr. Goligher reports personal fees and non-financial support from Getinge, non-financial support from Timpel, outside the submitted work. The other authors declare that they have no competing interests.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Anhänge

Supplementary Information

Literatur
1.
Zurück zum Zitat Connolly B, MacBean V, Crowley C. Ultrasound for the assessment of peripheral skeletal muscle architecture in critical illness: a systematic review. Crit Care Med. 2015;43:897–905.CrossRef Connolly B, MacBean V, Crowley C. Ultrasound for the assessment of peripheral skeletal muscle architecture in critical illness: a systematic review. Crit Care Med. 2015;43:897–905.CrossRef
2.
Zurück zum Zitat Tuinman PR, Jonkman AH, Dres M, Shi ZH, Goligher EC, Goffi A, et al. Respiratory muscle ultrasonography: methodology, basic and advanced principles and clinical applications in ICU and ED patients-a narrative review. Intensive Care Med. 2020;46:594–605.CrossRef Tuinman PR, Jonkman AH, Dres M, Shi ZH, Goligher EC, Goffi A, et al. Respiratory muscle ultrasonography: methodology, basic and advanced principles and clinical applications in ICU and ED patients-a narrative review. Intensive Care Med. 2020;46:594–605.CrossRef
3.
Zurück zum Zitat Strasser EM, Draskovits T, Praschak M, Quittan M, Graf A. Association between ultrasound measurements of muscle thickness, pennation angle, echogenicity and skeletal muscle strength in the elderly. Age (Dordr). 2013;35:2377–88.CrossRef Strasser EM, Draskovits T, Praschak M, Quittan M, Graf A. Association between ultrasound measurements of muscle thickness, pennation angle, echogenicity and skeletal muscle strength in the elderly. Age (Dordr). 2013;35:2377–88.CrossRef
4.
Zurück zum Zitat Pillen S, Arts IM, Zwarts MJ. Muscle ultrasound in neuromuscular disorders. Muscle Nerve. 2008;37:679–93.CrossRef Pillen S, Arts IM, Zwarts MJ. Muscle ultrasound in neuromuscular disorders. Muscle Nerve. 2008;37:679–93.CrossRef
5.
Zurück zum Zitat Heckmatt JZ, Dubowitz V, Leeman S. Detection of pathological change in dystrophic muscle with B-scan ultrasound imaging. Lancet. 1980;1:1389–90.CrossRef Heckmatt JZ, Dubowitz V, Leeman S. Detection of pathological change in dystrophic muscle with B-scan ultrasound imaging. Lancet. 1980;1:1389–90.CrossRef
6.
Zurück zum Zitat Reimers CD, Schlotter B, Eicke BM, Witt TN. Calf enlargement in neuromuscular diseases: a quantitative ultrasound study in 350 patients and review of the literature. J Neurol Sci. 1996;143:46–56.CrossRef Reimers CD, Schlotter B, Eicke BM, Witt TN. Calf enlargement in neuromuscular diseases: a quantitative ultrasound study in 350 patients and review of the literature. J Neurol Sci. 1996;143:46–56.CrossRef
7.
Zurück zum Zitat Heckmatt JZ, Pier N, Dubowitz V. Real-time ultrasound imaging of muscles. Muscle Nerve. 1988;11:56–65.CrossRef Heckmatt JZ, Pier N, Dubowitz V. Real-time ultrasound imaging of muscles. Muscle Nerve. 1988;11:56–65.CrossRef
8.
Zurück zum Zitat Puthucheary ZA, Phadke R, Rawal J, McPhail MJ, Sidhu PS, Rowlerson A, et al. Qualitative ultrasound in acute critical illness muscle wasting. Crit Care Med. 2015;43:1603–11.CrossRef Puthucheary ZA, Phadke R, Rawal J, McPhail MJ, Sidhu PS, Rowlerson A, et al. Qualitative ultrasound in acute critical illness muscle wasting. Crit Care Med. 2015;43:1603–11.CrossRef
9.
Zurück zum Zitat Jiménez-Díaz F, Jimena I, Luque E, Mendizábal S, Bouffard A, Jiménez-Reina L, et al. Experimental muscle injury: correlation between ultrasound and histological findings. Muscle Nerve. 2012;45:705–12.CrossRef Jiménez-Díaz F, Jimena I, Luque E, Mendizábal S, Bouffard A, Jiménez-Reina L, et al. Experimental muscle injury: correlation between ultrasound and histological findings. Muscle Nerve. 2012;45:705–12.CrossRef
10.
Zurück zum Zitat Koppaka S, Shklyar I, Rutkove SB, Darras BT, Anthony BW, Zaidman CM, et al. Quantitative ultrasound assessment of duchenne muscular dystrophy using edge detection analysis. J Ultrasound Med. 2016;35:1889–97.CrossRef Koppaka S, Shklyar I, Rutkove SB, Darras BT, Anthony BW, Zaidman CM, et al. Quantitative ultrasound assessment of duchenne muscular dystrophy using edge detection analysis. J Ultrasound Med. 2016;35:1889–97.CrossRef
11.
Zurück zum Zitat Lee JC, Mitchell AWM, Healy JC. Imaging of muscle injury in the elite athlete. Br J Radiol. 2012;85:1173–85.CrossRef Lee JC, Mitchell AWM, Healy JC. Imaging of muscle injury in the elite athlete. Br J Radiol. 2012;85:1173–85.CrossRef
12.
Zurück zum Zitat Berry MJ, Files DC, Campos CL, Bakhru RN, Skaggs BM, Morris PE. Echogenicity is related to skeletal muscle strength in patients with acute respiratory failure. J Cardiopulm Rehabil Prev. 2019;39:E17–20.CrossRef Berry MJ, Files DC, Campos CL, Bakhru RN, Skaggs BM, Morris PE. Echogenicity is related to skeletal muscle strength in patients with acute respiratory failure. J Cardiopulm Rehabil Prev. 2019;39:E17–20.CrossRef
13.
Zurück zum Zitat Grosu HB, Lee YI, Lee J, Eden E, Eikermann M, Rose KM. Diaphragm muscle thinning in patients who are mechanically ventilated. Chest. 2012;142:1455–60.CrossRef Grosu HB, Lee YI, Lee J, Eden E, Eikermann M, Rose KM. Diaphragm muscle thinning in patients who are mechanically ventilated. Chest. 2012;142:1455–60.CrossRef
14.
Zurück zum Zitat Goligher EC, Fan E, Herridge MS, Murray A, Vorona S, Brace D, et al. Evolution of diaphragm thickness during mechanical ventilation. Impact of inspiratory effort. Am J Respir Crit Care Med. 2015;192:1080–8.CrossRef Goligher EC, Fan E, Herridge MS, Murray A, Vorona S, Brace D, et al. Evolution of diaphragm thickness during mechanical ventilation. Impact of inspiratory effort. Am J Respir Crit Care Med. 2015;192:1080–8.CrossRef
15.
Zurück zum Zitat Sklar MC, Dres M, Fan E, Rubenfeld GD, Scales DC, Herridge MS, et al. Association of low baseline diaphragm muscle mass with prolonged mechanical ventilation and mortality among critically Ill adults. JAMA Netw Open. 2020;3:e1921520.CrossRef Sklar MC, Dres M, Fan E, Rubenfeld GD, Scales DC, Herridge MS, et al. Association of low baseline diaphragm muscle mass with prolonged mechanical ventilation and mortality among critically Ill adults. JAMA Netw Open. 2020;3:e1921520.CrossRef
16.
Zurück zum Zitat Goligher EC, Dres M, Fan E, Rubenfeld GD, Scales DC, Herridge MS, et al. Mechanical ventilation-induced diaphragm atrophy strongly impacts clinical outcomes. Am J Respir Crit Care Med. 2018;197:204–13.CrossRef Goligher EC, Dres M, Fan E, Rubenfeld GD, Scales DC, Herridge MS, et al. Mechanical ventilation-induced diaphragm atrophy strongly impacts clinical outcomes. Am J Respir Crit Care Med. 2018;197:204–13.CrossRef
17.
Zurück zum Zitat Goligher EC, Laghi F, Detsky ME, Farias P, Murray A, Brace D, et al. Measuring diaphragm thickness with ultrasound in mechanically ventilated patients: feasibility, reproducibility and validity. Intensive Care Med. 2015;41:642–9.CrossRef Goligher EC, Laghi F, Detsky ME, Farias P, Murray A, Brace D, et al. Measuring diaphragm thickness with ultrasound in mechanically ventilated patients: feasibility, reproducibility and validity. Intensive Care Med. 2015;41:642–9.CrossRef
18.
Zurück zum Zitat Harris-Love MO, Seamon BA, Teixeira C, Ismail C. Ultrasound estimates of muscle quality in older adults: reliability and comparison of Photoshop and ImageJ for the grayscale analysis of muscle echogenicity. PeerJ. 2016;4:e1721.CrossRef Harris-Love MO, Seamon BA, Teixeira C, Ismail C. Ultrasound estimates of muscle quality in older adults: reliability and comparison of Photoshop and ImageJ for the grayscale analysis of muscle echogenicity. PeerJ. 2016;4:e1721.CrossRef
19.
Zurück zum Zitat Parry SM, El-Ansary D, Cartwright MS, Sarwal A, Berney S, Koopman R, et al. Ultrasonography in the intensive care unit setting can be used to detect changes in the quality of muscle and is related to muscle strength and function. J Crit Care. 2015;30(1151):e9-14. Parry SM, El-Ansary D, Cartwright MS, Sarwal A, Berney S, Koopman R, et al. Ultrasonography in the intensive care unit setting can be used to detect changes in the quality of muscle and is related to muscle strength and function. J Crit Care. 2015;30(1151):e9-14.
20.
Zurück zum Zitat Sarwal A, Parry SM, Berry MJ, Hsu FC, Lewis MT, Justus NW, et al. Interobserver reliability of quantitative muscle sonographic analysis in the critically Ill population. J Ultrasound Med. 2015;34:1191–200.CrossRef Sarwal A, Parry SM, Berry MJ, Hsu FC, Lewis MT, Justus NW, et al. Interobserver reliability of quantitative muscle sonographic analysis in the critically Ill population. J Ultrasound Med. 2015;34:1191–200.CrossRef
21.
Zurück zum Zitat Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;327:307–10.CrossRef Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;327:307–10.CrossRef
22.
Zurück zum Zitat Le Gall JR, Lemeshow S, Saulnier F. A new Simplified Acute Physiology Score (SAPS II) based on a European/ North American multicenter study. JAMA. 1993;270:2957–63.CrossRef Le Gall JR, Lemeshow S, Saulnier F. A new Simplified Acute Physiology Score (SAPS II) based on a European/ North American multicenter study. JAMA. 1993;270:2957–63.CrossRef
23.
Zurück zum Zitat Singer M, Deutschman CS, Seymour CW, Shankar-Hari M, Annane D, Bauer M, et al. The third international consensus definitions for sepsis and septic shock (Sepsis-3). JAMA. 2016;315:801–10.CrossRef Singer M, Deutschman CS, Seymour CW, Shankar-Hari M, Annane D, Bauer M, et al. The third international consensus definitions for sepsis and septic shock (Sepsis-3). JAMA. 2016;315:801–10.CrossRef
24.
Zurück zum Zitat Riker RR, Picard JT, Fraser GL. Prospective evaluation of the Sedation-Agitation Scale for adult critically ill patients. Crit Care Med. 1999;27:1325–9.CrossRef Riker RR, Picard JT, Fraser GL. Prospective evaluation of the Sedation-Agitation Scale for adult critically ill patients. Crit Care Med. 1999;27:1325–9.CrossRef
25.
Zurück zum Zitat Vincent JL, Moreno R, Takala J, Willatts S, De Mendonça A, Bruining H, et al. The SOFA (Sepsis-related Organ Failure Assessment) score to describe organ dysfunction/failure. On behalf of the Working Group on Sepsis-Related Problems of the European Society of Intensive Care Medicine. Intensive Care Med. 1996;22:707–710. Vincent JL, Moreno R, Takala J, Willatts S, De Mendonça A, Bruining H, et al. The SOFA (Sepsis-related Organ Failure Assessment) score to describe organ dysfunction/failure. On behalf of the Working Group on Sepsis-Related Problems of the European Society of Intensive Care Medicine. Intensive Care Med. 1996;22:707–710.
26.
Zurück zum Zitat Reimers K, Reimers CD, Wagner S, Paetzke I, Pongratz DE. Skeletal muscle sonography: a correlative study of echogenicity and morphology. J Ultrasound Med. 1993;12:73–7.CrossRef Reimers K, Reimers CD, Wagner S, Paetzke I, Pongratz DE. Skeletal muscle sonography: a correlative study of echogenicity and morphology. J Ultrasound Med. 1993;12:73–7.CrossRef
27.
Zurück zum Zitat Zaidman CM, Holland MR, Anderson CC, Pestronk A. Calibrated quantitative ultrasound imaging of skeletal muscle using backscatter analysis. Muscle Nerve. 2008;38:893–8.CrossRef Zaidman CM, Holland MR, Anderson CC, Pestronk A. Calibrated quantitative ultrasound imaging of skeletal muscle using backscatter analysis. Muscle Nerve. 2008;38:893–8.CrossRef
28.
Zurück zum Zitat Maurits NM, Bollen AE, Windhausen A, De Jager AE, Van Der Hoeven JH. Muscle ultrasound analysis: normal values and differentiation between myopathies and neuropathies. Ultrasound Med Biol. 2003;29:215–25.CrossRef Maurits NM, Bollen AE, Windhausen A, De Jager AE, Van Der Hoeven JH. Muscle ultrasound analysis: normal values and differentiation between myopathies and neuropathies. Ultrasound Med Biol. 2003;29:215–25.CrossRef
29.
Zurück zum Zitat Mascalchi M, Camiciottoli G, Diciotti S. Lung densitometry: why, how and when. J Thorac Dis. 2017;9:3319–45.CrossRef Mascalchi M, Camiciottoli G, Diciotti S. Lung densitometry: why, how and when. J Thorac Dis. 2017;9:3319–45.CrossRef
30.
Zurück zum Zitat Goligher EC, Brochard LJ, Reid DW, Fan E, Saarela O, Slutsky AS, et al. Diaphragmatic myotrauma: a mediator of prolonged ventilation and poor patient outcomes in acute respiratory failure. Lancet Respir Med. 2019;7:90–8.CrossRef Goligher EC, Brochard LJ, Reid DW, Fan E, Saarela O, Slutsky AS, et al. Diaphragmatic myotrauma: a mediator of prolonged ventilation and poor patient outcomes in acute respiratory failure. Lancet Respir Med. 2019;7:90–8.CrossRef
31.
Zurück zum Zitat Brochard L, Slutsky A, Pesenti A. Mechanical ventilation to minimize progression of lung injury in acute respiratory failure. Am J Respir Crit Care Med. 2017;195:438–42.CrossRef Brochard L, Slutsky A, Pesenti A. Mechanical ventilation to minimize progression of lung injury in acute respiratory failure. Am J Respir Crit Care Med. 2017;195:438–42.CrossRef
32.
Zurück zum Zitat Jiang TX, Reid WD, Road JD. Delayed diaphragm injury and diaphragm force production. Am J Respir Crit Care Med. 1998;157:736–42.CrossRef Jiang TX, Reid WD, Road JD. Delayed diaphragm injury and diaphragm force production. Am J Respir Crit Care Med. 1998;157:736–42.CrossRef
33.
Zurück zum Zitat Orozco-Levi M, Lloreta J, Minguella J, Serrano S, Broquetas JM, Gea J. Injury of the human diaphragm associated with exertion and chronic obstructive pulmonary disease. Am J Respir Crit Care Med. 2001;164:1734–9.CrossRef Orozco-Levi M, Lloreta J, Minguella J, Serrano S, Broquetas JM, Gea J. Injury of the human diaphragm associated with exertion and chronic obstructive pulmonary disease. Am J Respir Crit Care Med. 2001;164:1734–9.CrossRef
34.
Zurück zum Zitat Vassilakopoulos T, Hussain SN. Ventilatory muscle activation and inflammation: cytokines, reactive oxygen species, and nitric oxide. J Appl Physiol. 1985;2007(102):1687–95. Vassilakopoulos T, Hussain SN. Ventilatory muscle activation and inflammation: cytokines, reactive oxygen species, and nitric oxide. J Appl Physiol. 1985;2007(102):1687–95.
35.
Zurück zum Zitat Hillas G, Perlikos F, Toumpanakis D, Litsiou E, Nikolakopoulou S, Sagris K, et al. Controlled mechanical ventilation attenuates the systemic inflammation of severe chronic obstructive pulmonary disease exacerbations. Am J Respir Crit Care Med. 2016;193:696–8.CrossRef Hillas G, Perlikos F, Toumpanakis D, Litsiou E, Nikolakopoulou S, Sagris K, et al. Controlled mechanical ventilation attenuates the systemic inflammation of severe chronic obstructive pulmonary disease exacerbations. Am J Respir Crit Care Med. 2016;193:696–8.CrossRef
37.
Zurück zum Zitat Mourtzakis M, Parry S, Connolly B, Puthucheary Z. Skeletal muscle ultrasound in critical care: a tool in need of translation. Ann Am Thorac Soc. 2017;14:1495–503.CrossRef Mourtzakis M, Parry S, Connolly B, Puthucheary Z. Skeletal muscle ultrasound in critical care: a tool in need of translation. Ann Am Thorac Soc. 2017;14:1495–503.CrossRef
38.
Zurück zum Zitat Ihnatsenka B, Boezaart AP. Ultrasound: Basic understanding and learning the language. Int J Shoulder Surg. 2010;4:55–62.CrossRef Ihnatsenka B, Boezaart AP. Ultrasound: Basic understanding and learning the language. Int J Shoulder Surg. 2010;4:55–62.CrossRef
39.
Zurück zum Zitat Young HJ, Jenkins NT, Zhao Q, Mccully KK. Measurement of intramuscular fat by muscle echo intensity. Muscle Nerve. 2015;52:963–71.CrossRef Young HJ, Jenkins NT, Zhao Q, Mccully KK. Measurement of intramuscular fat by muscle echo intensity. Muscle Nerve. 2015;52:963–71.CrossRef
40.
Zurück zum Zitat Lori S, Lolli F, Molesti E, Bastianelli M, Gabbanini S, Saia V, et al. Muscle-ultrasound evaluation in healthy pediatric subjects: Age-related normative data. Muscle Nerve. 2018;58:245–50.CrossRef Lori S, Lolli F, Molesti E, Bastianelli M, Gabbanini S, Saia V, et al. Muscle-ultrasound evaluation in healthy pediatric subjects: Age-related normative data. Muscle Nerve. 2018;58:245–50.CrossRef
Metadaten
Titel
Diaphragm echodensity in mechanically ventilated patients: a description of technique and outcomes
verfasst von
Benjamin Coiffard
Stephen Riegler
Michael C. Sklar
Martin Dres
Stefannie Vorona
W. Darlene Reid
Laurent J. Brochard
Niall D. Ferguson
Ewan C. Goligher
Publikationsdatum
01.12.2021
Verlag
BioMed Central
Erschienen in
Critical Care / Ausgabe 1/2021
Elektronische ISSN: 1364-8535
DOI
https://doi.org/10.1186/s13054-021-03494-9

Weitere Artikel der Ausgabe 1/2021

Critical Care 1/2021 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.