Distal pressure monitoring and titration with percent balloon volume: feasible management of partial resuscitative endovascular balloon occlusion of the aorta (P-REBOA)

verfasst von: Yosuke Matsumura, Akiko Higashi, Yoshimitsu Izawa, Shuji Hishikawa, Hiroshi Kondo, Viktor Reva, Shigeto Oda, Junichi Matsumoto

Erschienen in: European Journal of Trauma and Emergency Surgery | Ausgabe 4/2021

Einloggen, um Zugang zu erhalten

Abstract

Introduction

Resuscitative endovascular balloon occlusion of the aorta (REBOA) increases proximal arterial pressure, but may also induce life-threatening distal ischemia. Partial REBOA (P-REBOA) is thought to mitigate distal ischemia during aortic occlusion. However, feasible indicators of the degree of P-REBOA remain inconsistent. We hypothesised percent balloon volume could be a substitute for pressure measurements of gradients during P- REBOA. This study aimed to compare balloon volume and arterial pressure gradient, and analysed with intra-balloon pressure and balloon shape.

Methods

Proximal (carotid) and distal (femoral) arterial pressures were recorded and a 7-Fr REBOA catheter was placed in four swine. Total REBOA was defined as a cessation of distal pulse pressure and maximum balloon volume was documented. The balloon volume was titrated by 20% increments of maximum capacity to adjust the degree of P-REBOA. The distal/proximal arterial pressure gradient and the intra-balloon pressures were also recorded. The changes in shape and the cross-sectional area of the balloon were evaluated with computed tomography (CT) images.

Results

The proximal mean arterial pressure (MAP) plateaued after 60% balloon volume; meanwhile, distal pulse pressure was still left. The balloon pressure was traced with proximal MAP before contact with aortic wall. The balloon shape changed unevenly from “cone” to “spindle” shape, although the balloon cross-sectional area of the mid-segment linearly increased.

Conclusion

Monitoring distal pressure and titrating percent balloon volume is feasible to manage P-REBOA. In this experiment, 60% balloon volume was enough inflation to elevate central pressure allowing distal perfusion. The intra-balloon pressure was not reliable due to the strong influence of proximal MAP and uneven change of the balloon shape.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
3.
Zurück zum Zitat Brenner M, Inaba K, Aiolfi A, DuBose J, Fabian T, Bee T, et al. Resuscitative endovascular balloon occlusion of the aorta and resuscitative thoracotomy in select patients with hemorrhagic shock: early results from the american association for the surgery of Trauma's aortic occlusion in resuscitation for trauma and acute care surgery registry. J Am Coll Surg. 2018;226(5):730–40. https://doi.org/10.1016/j.jamcollsurg.2018.01.044.CrossRefPubMed Brenner M, Inaba K, Aiolfi A, DuBose J, Fabian T, Bee T, et al. Resuscitative endovascular balloon occlusion of the aorta and resuscitative thoracotomy in select patients with hemorrhagic shock: early results from the american association for the surgery of Trauma's aortic occlusion in resuscitation for trauma and acute care surgery registry. J Am Coll Surg. 2018;226(5):730–40. https://​doi.​org/​10.​1016/​j.​jamcollsurg.​2018.​01.​044.CrossRefPubMed
5.
Zurück zum Zitat Manzano-Nunez R, Orlas CP, Herrera-Escobar JP, Galvagno S, DuBose J, Melendez JJ, et al. A meta-analysis of the incidence of complications associated with groin access after the use of resuscitative endovascular balloon occlusion of the aorta in trauma patients. J Trauma Acute Care Surg. 2018;85(3):626–34. https://doi.org/10.1097/TA.0000000000001978.CrossRefPubMed Manzano-Nunez R, Orlas CP, Herrera-Escobar JP, Galvagno S, DuBose J, Melendez JJ, et al. A meta-analysis of the incidence of complications associated with groin access after the use of resuscitative endovascular balloon occlusion of the aorta in trauma patients. J Trauma Acute Care Surg. 2018;85(3):626–34. https://​doi.​org/​10.​1097/​TA.​0000000000001978​.CrossRefPubMed
Metadaten
Titel
Distal pressure monitoring and titration with percent balloon volume: feasible management of partial resuscitative endovascular balloon occlusion of the aorta (P-REBOA)
verfasst von
Yosuke Matsumura
Akiko Higashi
Yoshimitsu Izawa
Shuji Hishikawa
Hiroshi Kondo
Viktor Reva
Shigeto Oda
Junichi Matsumoto
Publikationsdatum
06.11.2019
Verlag
Springer Berlin Heidelberg
Erschienen in
European Journal of Trauma and Emergency Surgery / Ausgabe 4/2021
Print ISSN: 1863-9933
Elektronische ISSN: 1863-9941
DOI
https://doi.org/10.1007/s00068-019-01257-4

Weitere Artikel der Ausgabe 4/2021

European Journal of Trauma and Emergency Surgery 4/2021 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.