Dose estimation of ultra-low-dose chest CT to different sized adult patients

verfasst von: Tony M. Svahn, Tommy Sjöberg, Jennifer C. Ast

Erschienen in: European Radiology | Ausgabe 8/2019

Einloggen, um Zugang zu erhalten

Abstract

Objectives

To evaluate the effect of patient size on radiation dose for standard CT (SD-CT), ultra-low-dose CT (ULD-CT) and two-view digital radiography (DR).

Methods

Dosimeters were distributed within the lungs of chest phantoms representing males of 65 kg and 82 kg (body mass indices 23 and 29). In contrast to SD-CT and DR which include automatic exposure control (AEC), the ULD scan employs a fixed mAs value. The phantoms were exposed to SD, ULD and DR while recording lung doses. Projected dose data were calculated from the phantoms. The resulting exposure settings were used in Monte Carlo programs to determine the effective dose for a standard-sized (BMI 24.2) adult male (170 cm/70 kg) and female (160 cm/59 kg). Patients previously examined by both ULD- and SD-CT were identified to determine post hoc size-specific dose estimates (SSDEs).

Results

ULD-CT dose was inversely related to patient size; average lung doses summarised in terms of patient size BMI23/29 are 5.2/8.1 (SD-CT), 0.56/0.35 (ULD-CT) and 0.05/0.13 mGy (DR), while the effective doses for these techniques on a standard-sized male were 2.9, 0.16 and 0.03 mSv and 2.3, 0.247 and 0.024 mSv for a standard-sized female respectively. SSDEs for 15 patients (averages: BMI 26, range 18–37) averaged 5.5 mGy (3.6–10) for SD-CT and 0.35 mGy (0.42–0.27) for ULD-CT.

Conclusions

The effective doses for a standard-sized male and female examined by ULD-CT are (respectively) ~ 6%/~ 11% of SD-CT and ~ 5/~ 10 times higher than DR. ULD-CT gave a lower radiation dosage to larger patients than DR. AEC is warranted in ULD-CT for improved dose consistency.

Key Points

• For standard-sized patients, ULD-CT dose level is ~ 6%/~ 11% of SD-CT, and ~ 5/~ 10 times higher than DR. For larger patients, ULD-CT is currently being used clinically at lower dose levels than DR.
• Using ULD-CT should greatly reduce the risk of late effects from ionising radiation.
• AEC in ULD-CT is desirable for increased consistency in patient dose.
Literatur
1.
Zurück zum Zitat Båth M, Håkansson M, Börjesson S et al (2005) Nodule detection in digital chest radiography: effect of anatomical noise. Radiat Prot Dosimetry 114(1–3):109–113CrossRef Båth M, Håkansson M, Börjesson S et al (2005) Nodule detection in digital chest radiography: effect of anatomical noise. Radiat Prot Dosimetry 114(1–3):109–113CrossRef
2.
Zurück zum Zitat Blanchon T, Bréchot JM, Grenier PA et al (2007) Baseline results of the Depiscan study: a French randomized pilot trial of lung cancer screening comparing low dose CT scan (LDCT) and chest X-ray (CXR). Lung Cancer 58(1):50–58CrossRef Blanchon T, Bréchot JM, Grenier PA et al (2007) Baseline results of the Depiscan study: a French randomized pilot trial of lung cancer screening comparing low dose CT scan (LDCT) and chest X-ray (CXR). Lung Cancer 58(1):50–58CrossRef
3.
Zurück zum Zitat Terra-Filho M, Bagatin E, Nery LE et al (2015) Screening of miners and millers at decreasing levels of asbestos exposure: comparison of chest radiography and thin-section computed tomography. PLoS One 10(3):e0118585CrossRef Terra-Filho M, Bagatin E, Nery LE et al (2015) Screening of miners and millers at decreasing levels of asbestos exposure: comparison of chest radiography and thin-section computed tomography. PLoS One 10(3):e0118585CrossRef
4.
Zurück zum Zitat National Research Council (2006) Health risks from exposure to low levels of ionizing radiation: BEIR VII phase 2. The National Academies Press, Washington, DC, p 245 National Research Council (2006) Health risks from exposure to low levels of ionizing radiation: BEIR VII phase 2. The National Academies Press, Washington, DC, p 245
5.
Zurück zum Zitat International Commission on Radiological Protection (1975) Report of the Task Group on Reference Man. ICRP Publication 23. Pergamon Press, Oxford Adopted October 1974 International Commission on Radiological Protection (1975) Report of the Task Group on Reference Man. ICRP Publication 23. Pergamon Press, Oxford Adopted October 1974
6.
Zurück zum Zitat Zhu X, Yu J, Huang Z (2004) Low-dose chest CT: optimizing radiation protection for patients. AJR Am J Roentgenol 183(3):809–816CrossRef Zhu X, Yu J, Huang Z (2004) Low-dose chest CT: optimizing radiation protection for patients. AJR Am J Roentgenol 183(3):809–816CrossRef
7.
Zurück zum Zitat Ketelslegers E, Van Beers BE (2006) Urinary calculi: improved detection and characterization with thin-slice multidetector CT. Eur Radiol 16(1):161–165CrossRef Ketelslegers E, Van Beers BE (2006) Urinary calculi: improved detection and characterization with thin-slice multidetector CT. Eur Radiol 16(1):161–165CrossRef
8.
Zurück zum Zitat Kubo T, Ohno Y, Nishino M et al (2016) Low dose chest CT protocol (50 mAs) as a routine protocol for comprehensive assessment of intrathoracic abnormality. Eur J Radiol Open 3:86–94CrossRef Kubo T, Ohno Y, Nishino M et al (2016) Low dose chest CT protocol (50 mAs) as a routine protocol for comprehensive assessment of intrathoracic abnormality. Eur J Radiol Open 3:86–94CrossRef
9.
Zurück zum Zitat Rampinelli C, De Marco P, Origgi D et al (2017) Exposure to low dose computed tomography for lung cancer screening and risk of cancer: secondary analysis of trial data and risk-benefit analysis. BMJ 356:j347CrossRef Rampinelli C, De Marco P, Origgi D et al (2017) Exposure to low dose computed tomography for lung cancer screening and risk of cancer: secondary analysis of trial data and risk-benefit analysis. BMJ 356:j347CrossRef
10.
Zurück zum Zitat Larke FJ, Kruger RL, Cagnon CH et al (2011) Estimated radiation dose associated with low-dose chest CT of average-size participants in the National Lung Screening Trial. AJR Am J Roentgenol 197(5):1165–1169CrossRef Larke FJ, Kruger RL, Cagnon CH et al (2011) Estimated radiation dose associated with low-dose chest CT of average-size participants in the National Lung Screening Trial. AJR Am J Roentgenol 197(5):1165–1169CrossRef
11.
Zurück zum Zitat Mettler FA Jr, Huda W, Yoshizumi TT, Mahesh M (2008) Effective doses in radiology and diagnostic nuclear medicine: a catalog. Radiology 248(1):254–263CrossRef Mettler FA Jr, Huda W, Yoshizumi TT, Mahesh M (2008) Effective doses in radiology and diagnostic nuclear medicine: a catalog. Radiology 248(1):254–263CrossRef
12.
Zurück zum Zitat Fujita M, Higaki T, Awaya Y et al (2017) Lung cancer screening with ultra-low dose CT using full iterative reconstruction. Jpn J Radiol 35(4):179–189CrossRef Fujita M, Higaki T, Awaya Y et al (2017) Lung cancer screening with ultra-low dose CT using full iterative reconstruction. Jpn J Radiol 35(4):179–189CrossRef
13.
Zurück zum Zitat Schaal M, Severac F, Labani A, Jeung MY, Roy C, Ohana M (2016) Diagnostic performance of ultra-low-dose computed tomography for detecting asbestos-related pleuropulmonary diseases: prospective study in a screening setting. PLoS One 11(12):e0168979CrossRef Schaal M, Severac F, Labani A, Jeung MY, Roy C, Ohana M (2016) Diagnostic performance of ultra-low-dose computed tomography for detecting asbestos-related pleuropulmonary diseases: prospective study in a screening setting. PLoS One 11(12):e0168979CrossRef
14.
Zurück zum Zitat Brady SL, Kaufman RA (2012) Investigation of American Association of Physicists in Medicine report 204 size-specific dose estimates for pediatric CT implementation. Radiology 265(3):832–840CrossRef Brady SL, Kaufman RA (2012) Investigation of American Association of Physicists in Medicine report 204 size-specific dose estimates for pediatric CT implementation. Radiology 265(3):832–840CrossRef
16.
Zurück zum Zitat Shrimpton PC, Jones DG, Hillier MC, Wall BF, Le Heron JC, Faulkner K (1991) Survey of CT Practise in the UK. Part 2: dosimetric aspects (NRPB-R249). National Radiological Protection Borard, Chilton Shrimpton PC, Jones DG, Hillier MC, Wall BF, Le Heron JC, Faulkner K (1991) Survey of CT Practise in the UK. Part 2: dosimetric aspects (NRPB-R249). National Radiological Protection Borard, Chilton
17.
Zurück zum Zitat Stamm G, Nagel HD (2002) CT-expo--a novel program for dose evaluation in CT. Rofo 174(12):1570–1576CrossRef Stamm G, Nagel HD (2002) CT-expo--a novel program for dose evaluation in CT. Rofo 174(12):1570–1576CrossRef
18.
Zurück zum Zitat Tapiovaara M, Lakkisto M, Sermovaa A (1997) PCXMC: a PC-based Monte Carlo program for calculating patient doses in medical x-ray examinations. Finnish Centre for Radiation and Nuclear Safety, Helsinki Tapiovaara M, Lakkisto M, Sermovaa A (1997) PCXMC: a PC-based Monte Carlo program for calculating patient doses in medical x-ray examinations. Finnish Centre for Radiation and Nuclear Safety, Helsinki
19.
Zurück zum Zitat Cristy M, Eckerman K (1987) Specific absorbed fractions of energy at various ages from internal photon sources. Parts I-VII (ORNL/TM-8381). Oak Ridge National Laboratory, Oak Ridge, pp 1–74 Cristy M, Eckerman K (1987) Specific absorbed fractions of energy at various ages from internal photon sources. Parts I-VII (ORNL/TM-8381). Oak Ridge National Laboratory, Oak Ridge, pp 1–74
20.
Zurück zum Zitat Zankl M, Panzer W, Drexler G (1991) The calculation of dose from external photon exposures using reference human phantoms and Monte Carlo methods. VI. Organ doses from computed tomographic examinations. GSF-Forschungszentrum für Umwelt und Gesundtheit, Institut für Strahlenschutz, Neuherberg Zankl M, Panzer W, Drexler G (1991) The calculation of dose from external photon exposures using reference human phantoms and Monte Carlo methods. VI. Organ doses from computed tomographic examinations. GSF-Forschungszentrum für Umwelt und Gesundtheit, Institut für Strahlenschutz, Neuherberg
21.
Zurück zum Zitat International Committee on Radiological Protection (2007) The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103. Ann ICRP 37(2–4):1–332 International Committee on Radiological Protection (2007) The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103. Ann ICRP 37(2–4):1–332
22.
Zurück zum Zitat International Committee on Radiological Protection (1991) 1990 Recommendations of the International Commission on Radiological Protection. ICRP publication No. 60. Pergamom, Oxford International Committee on Radiological Protection (1991) 1990 Recommendations of the International Commission on Radiological Protection. ICRP publication No. 60. Pergamom, Oxford
23.
Zurück zum Zitat Rubin DL (2011) Informatics in radiology: measuring and improving quality in radiology: meeting the challenge with informatics. Radiographics 31(6):1511–1527CrossRef Rubin DL (2011) Informatics in radiology: measuring and improving quality in radiology: meeting the challenge with informatics. Radiographics 31(6):1511–1527CrossRef
24.
Zurück zum Zitat McCollough C, Bakalyar DM, Bostani M et al (2014) Use of water equivalent diameter for calculating patient size and size-specific dose estimates (SSDE) in CT: the report of AAPM task group 220. AAPM Rep 2014:6–23PubMedPubMedCentral McCollough C, Bakalyar DM, Bostani M et al (2014) Use of water equivalent diameter for calculating patient size and size-specific dose estimates (SSDE) in CT: the report of AAPM task group 220. AAPM Rep 2014:6–23PubMedPubMedCentral
25.
Zurück zum Zitat Weatherburn GC, Bryan S, Davies JG (2000) Comparison of doses for bedside examinations of the chest with conventional screen-film and computed radiography: results of a randomized controlled trial. Radiology 217(3):707–712CrossRef Weatherburn GC, Bryan S, Davies JG (2000) Comparison of doses for bedside examinations of the chest with conventional screen-film and computed radiography: results of a randomized controlled trial. Radiology 217(3):707–712CrossRef
26.
Zurück zum Zitat Fasola G, Belvedere O, Aita M et al (2007) Low-dose computed tomography screening for lung cancer and pleural mesothelioma in an asbestos-exposed population: baseline results of a prospective, nonrandomized feasibility trial—an Alpe-Adria thoracic oncology multidisciplinary group study (ATOM 002). Oncologist 12(10):1215–1224CrossRef Fasola G, Belvedere O, Aita M et al (2007) Low-dose computed tomography screening for lung cancer and pleural mesothelioma in an asbestos-exposed population: baseline results of a prospective, nonrandomized feasibility trial—an Alpe-Adria thoracic oncology multidisciplinary group study (ATOM 002). Oncologist 12(10):1215–1224CrossRef
27.
Zurück zum Zitat Carrillo MC, Alturkistany S, Roberts H et al (2013) Low-dose computed tomography (LDCT) in workers previously exposed to asbestos: detection of parenchymal lung disease. J Comput Assist Tomogr 37(4):626–630CrossRef Carrillo MC, Alturkistany S, Roberts H et al (2013) Low-dose computed tomography (LDCT) in workers previously exposed to asbestos: detection of parenchymal lung disease. J Comput Assist Tomogr 37(4):626–630CrossRef
28.
Zurück zum Zitat Båth M, Svalkvist A, von Wrangel A, Rismyhr-Olsson H, Cederblad A (2010) Effective dose to patients from chest examinations with tomosynthesis. Radiat Prot Dosimetry 139(1–3):153–158CrossRef Båth M, Svalkvist A, von Wrangel A, Rismyhr-Olsson H, Cederblad A (2010) Effective dose to patients from chest examinations with tomosynthesis. Radiat Prot Dosimetry 139(1–3):153–158CrossRef
29.
Zurück zum Zitat Dobbins JT 3rd, McAdams HP, Sabol JM et al (2017) Multi-institutional evaluation of digital tomosynthesis, dual-energy radiography, and conventional chest radiography for the detection and management of pulmonary nodules. Radiology 282(1):236–250CrossRef Dobbins JT 3rd, McAdams HP, Sabol JM et al (2017) Multi-institutional evaluation of digital tomosynthesis, dual-energy radiography, and conventional chest radiography for the detection and management of pulmonary nodules. Radiology 282(1):236–250CrossRef
30.
Zurück zum Zitat Lång K, Andersson I, Rosso A, Tingberg A, Timberg P, Zackrisson S (2016) Performance of one-view breast tomosynthesis as a stand-alone breast cancer screening modality: results from the Malmo Breast Tomosynthesis Screening Trial, a population-based study. Eur Radiol 26(1):184–190CrossRef Lång K, Andersson I, Rosso A, Tingberg A, Timberg P, Zackrisson S (2016) Performance of one-view breast tomosynthesis as a stand-alone breast cancer screening modality: results from the Malmo Breast Tomosynthesis Screening Trial, a population-based study. Eur Radiol 26(1):184–190CrossRef
31.
Zurück zum Zitat Han D, Heuvelmans MA, Oudkerk M (2017) Volume versus diameter assessment of small pulmonary nodules in CT lung cancer screening. Transl Lung Cancer Res 6(1):52–61CrossRef Han D, Heuvelmans MA, Oudkerk M (2017) Volume versus diameter assessment of small pulmonary nodules in CT lung cancer screening. Transl Lung Cancer Res 6(1):52–61CrossRef
32.
Zurück zum Zitat Alvare G, Gordon R (2017) Foxels for high flux, high resolution computed tomography (foxelct) using broad x-ray focal spots: Theory and two-dimensional fan beam examples. Radiol Diagn Imaging 1(1):1–40CrossRef Alvare G, Gordon R (2017) Foxels for high flux, high resolution computed tomography (foxelct) using broad x-ray focal spots: Theory and two-dimensional fan beam examples. Radiol Diagn Imaging 1(1):1–40CrossRef
33.
Zurück zum Zitat Oyama A, Kumagai S, Arai N et al (2018) Image quality improvement in cone-beam CT using the super-resolution technique. J Radiat Res 59(4):501–510CrossRef Oyama A, Kumagai S, Arai N et al (2018) Image quality improvement in cone-beam CT using the super-resolution technique. J Radiat Res 59(4):501–510CrossRef
34.
Zurück zum Zitat Naidu SG, Kriegshauser JS, Paden RG, He M, Wu Q, Hara AK (2014) Ultra-low-dose computed tomographic angiography with model-based iterative reconstruction compared with standard-dose imaging after endovascular aneurysm repair: a prospective pilot study. Abdom Imaging 39(6):1297–1303CrossRef Naidu SG, Kriegshauser JS, Paden RG, He M, Wu Q, Hara AK (2014) Ultra-low-dose computed tomographic angiography with model-based iterative reconstruction compared with standard-dose imaging after endovascular aneurysm repair: a prospective pilot study. Abdom Imaging 39(6):1297–1303CrossRef
35.
Zurück zum Zitat Rob S, Bryant T, Wilson I, Somani BK (2017) Ultra-low-dose, low-dose, and standard-dose CT of the kidney, ureters, and bladder: is there a difference? Results from a systematic review of the literature. Clin Radiol 72(1):11–15CrossRef Rob S, Bryant T, Wilson I, Somani BK (2017) Ultra-low-dose, low-dose, and standard-dose CT of the kidney, ureters, and bladder: is there a difference? Results from a systematic review of the literature. Clin Radiol 72(1):11–15CrossRef
36.
Zurück zum Zitat Thurley P, Crookdake J, Norwood M, Sturrock N, Fogarty AW (2018) Demand for CT scans increases during transition from paediatric to adult care: an observational study from 2009 to 2015. Br J Radiol 91(1083):20170467PubMed Thurley P, Crookdake J, Norwood M, Sturrock N, Fogarty AW (2018) Demand for CT scans increases during transition from paediatric to adult care: an observational study from 2009 to 2015. Br J Radiol 91(1083):20170467PubMed
Metadaten
Titel
Dose estimation of ultra-low-dose chest CT to different sized adult patients
verfasst von
Tony M. Svahn
Tommy Sjöberg
Jennifer C. Ast
Publikationsdatum
17.12.2018
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 8/2019
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-018-5849-5

Weitere Artikel der Ausgabe 8/2019

European Radiology 8/2019 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.