Dual energy CT angiography: pros and cons of dual-energy metal artifact reduction algorithm in patients after endovascular aortic repair

verfasst von: Johannes Boos, Jieming Fang, Benedikt H. Heidinger, Vassilios Raptopoulos, Olga R. Brook

Erschienen in: Abdominal Radiology | Ausgabe 3/2017

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To evaluate the value of metal artifact reduction (MAR) post-processing and iodine MD images in fast kV-switching dual-energy computed tomography (DECT) in patients after endovascular aortic repair (EVAR).

Materials and methods

Twenty-four consecutive EVAR patients (age 76 ± 9 years, 7/24 (29%) with coils, 9/24 (37.5%) with 10 endoleaks) who underwent DECT angiography were included in this HIPAA-compliant, IRB-approved retrospective study. Monochromatic reconstructions included 55, 60, 65, 70, and 75 keV with and without MAR and iodine MD images. Near field, far field, and vessel artifacts were assessed subjectively (1 = none; 5 = severe) and objectively by measuring noise and contrast-to-noise ratio. Visibility of endoleak was evaluated (1 = optimal; 5 = not visible).

Results

MAR objectively decreased artifacts from EVAR stents in the near field (60.7 ± 25.4 HU vs. 70.1 ± 34.2; p = .002) and subjectively increased near field (3.2 ± 0.9 vs. 2.8 ± 0.6; p < .001), far field (2.2 ± 0.6 vs. 1.6 ± 0.6; p < .001), and vessel (3.1 ± 1.1 vs. 2.5 ± 0.9; p < .001) artifacts. Near-field artifacts from coils were reduced by the MAR objectively (72.4 ± 24.8 vs. 182.7 ± 57.3 HU; p < .001) and subjectively (4.5 ± 0.5 vs. 4.9 ± 0.4; p = .02). CNR of standard reconstructions was optimal at 60 keV (38.3 ± 16.8). Reconstructions without MAR and iodine MD images provided improved endoleak visualization in 6/10 (60%) of cases (median 1 for both) compared to MAR (median 3) (p < 0.001). However, MAR improved visualization in 1/10 (10%) cases due to endoleak location adjacent to a coil.

Conclusion

DECT with MAR reduced artifacts from coils and improved endoleak visualization in 1/10 (10%) cases due to location adjacent to a coil. However, MAR impaired endoleak visualization in 6/10 (60%) cases and should be reviewed combined with 60 keV standard reconstructions and iodine MD images.
Literatur
1.
Zurück zum Zitat Stavropoulos SW, Marin H, Fairman RM, et al. (2005) Recurrent endoleak detection and measurement of aneurysm size with CTA after coil embolization of endoleaks. J Vasc Interv Radiol JVIR 16:1313–1317CrossRefPubMed Stavropoulos SW, Marin H, Fairman RM, et al. (2005) Recurrent endoleak detection and measurement of aneurysm size with CTA after coil embolization of endoleaks. J Vasc Interv Radiol JVIR 16:1313–1317CrossRefPubMed
2.
Zurück zum Zitat Dong Y, Shi AJ, Wu JL, et al. (2015) Metal artifact reduction using virtual monochromatic images for patients with pedicle screws implants on CT. Eur Spine J 25:1754–1763CrossRefPubMed Dong Y, Shi AJ, Wu JL, et al. (2015) Metal artifact reduction using virtual monochromatic images for patients with pedicle screws implants on CT. Eur Spine J 25:1754–1763CrossRefPubMed
3.
Zurück zum Zitat Jia Y, Zhang J, Fan J, et al. (2015) Gemstone spectral imaging reduced artefacts from metal coils or clips after treatment of cerebral aneurysms: a retrospective study of 35 patients. Br J Radiol 88:20150222CrossRefPubMedPubMedCentral Jia Y, Zhang J, Fan J, et al. (2015) Gemstone spectral imaging reduced artefacts from metal coils or clips after treatment of cerebral aneurysms: a retrospective study of 35 patients. Br J Radiol 88:20150222CrossRefPubMedPubMedCentral
4.
Zurück zum Zitat Stefaniak K, Stanisic M, Gabriel M, Oszkinis G (2016) Diagnostic imaging methods applied in long-term surveillance after EVAR. Will computed tomography angiography be replaced by other methods? Postępy Kardiol Interwencyjnej 12:6–12PubMedPubMedCentral Stefaniak K, Stanisic M, Gabriel M, Oszkinis G (2016) Diagnostic imaging methods applied in long-term surveillance after EVAR. Will computed tomography angiography be replaced by other methods? Postępy Kardiol Interwencyjnej 12:6–12PubMedPubMedCentral
5.
Zurück zum Zitat Brook OR, Gourtsoyianni S, Brook A, et al. (2012) Spectral CT with metal artifacts reduction software for improvement of tumor visibility in the vicinity of gold fiducial markers. Radiology 263:696–705CrossRefPubMed Brook OR, Gourtsoyianni S, Brook A, et al. (2012) Spectral CT with metal artifacts reduction software for improvement of tumor visibility in the vicinity of gold fiducial markers. Radiology 263:696–705CrossRefPubMed
6.
Zurück zum Zitat Laks S, Macari M, Chandarana H (2010) Dual-energy computed tomography imaging of the aorta after endovascular repair of abdominal aortic aneurysm. Semin Ultrasound CT MR 31:292–300CrossRefPubMed Laks S, Macari M, Chandarana H (2010) Dual-energy computed tomography imaging of the aorta after endovascular repair of abdominal aortic aneurysm. Semin Ultrasound CT MR 31:292–300CrossRefPubMed
7.
Zurück zum Zitat Boos J, Sawicki LM, Lanzman RS, et al. (2016) Metal artifact reduction (MAR) based on two-compartment physical modeling: evaluation in patients with hip implants. Acta Radiol. doi:10.1177/0284185116633911 PubMed Boos J, Sawicki LM, Lanzman RS, et al. (2016) Metal artifact reduction (MAR) based on two-compartment physical modeling: evaluation in patients with hip implants. Acta Radiol. doi:10.​1177/​0284185116633911​ PubMed
8.
Zurück zum Zitat Sonoda A, Nitta N, Ushio N, et al. (2015) Evaluation of the quality of CT images acquired with the single energy metal artifact reduction (SEMAR) algorithm in patients with hip and dental prostheses and aneurysm embolization coils. Jpn J Radiol 33:710–716CrossRefPubMed Sonoda A, Nitta N, Ushio N, et al. (2015) Evaluation of the quality of CT images acquired with the single energy metal artifact reduction (SEMAR) algorithm in patients with hip and dental prostheses and aneurysm embolization coils. Jpn J Radiol 33:710–716CrossRefPubMed
9.
Zurück zum Zitat Wuest W, May MS, Brand M, et al. (2015) Improved image quality in head and neck CT using a 3D iterative approach to reduce metal artifact. AJNR Am J Neuroradiol 36:1988–1993CrossRefPubMed Wuest W, May MS, Brand M, et al. (2015) Improved image quality in head and neck CT using a 3D iterative approach to reduce metal artifact. AJNR Am J Neuroradiol 36:1988–1993CrossRefPubMed
10.
Zurück zum Zitat Andersson KM, Nowik P, Persliden J, Thunberg P, Norrman E (2015) Metal artefact reduction in CT imaging of hip prostheses—an evaluation of commercial techniques provided by four vendors. Br J Radiol 88:20140473CrossRefPubMedPubMedCentral Andersson KM, Nowik P, Persliden J, Thunberg P, Norrman E (2015) Metal artefact reduction in CT imaging of hip prostheses—an evaluation of commercial techniques provided by four vendors. Br J Radiol 88:20140473CrossRefPubMedPubMedCentral
11.
Zurück zum Zitat Lee YH, Park KK, Song H-T, Kim S, Suh J-S (2012) Metal artefact reduction in gemstone spectral imaging dual-energy CT with and without metal artefact reduction software. Eur Radiol 22:1331–1340CrossRefPubMed Lee YH, Park KK, Song H-T, Kim S, Suh J-S (2012) Metal artefact reduction in gemstone spectral imaging dual-energy CT with and without metal artefact reduction software. Eur Radiol 22:1331–1340CrossRefPubMed
12.
Zurück zum Zitat Kuchenbecker S, Faby S, Sawall S, Lell M, Kachelrieß M (2015) Dual energy CT: how well can pseudo-monochromatic imaging reduce metal artifacts? Med Phys 42:1023–1036CrossRefPubMed Kuchenbecker S, Faby S, Sawall S, Lell M, Kachelrieß M (2015) Dual energy CT: how well can pseudo-monochromatic imaging reduce metal artifacts? Med Phys 42:1023–1036CrossRefPubMed
13.
Zurück zum Zitat Pessis E, Campagna R, Sverzut J-M, et al. (2013) Virtual monochromatic spectral imaging with fast kilovoltage switching: reduction of metal artifacts at CT. Radiographics 33:573–583CrossRefPubMed Pessis E, Campagna R, Sverzut J-M, et al. (2013) Virtual monochromatic spectral imaging with fast kilovoltage switching: reduction of metal artifacts at CT. Radiographics 33:573–583CrossRefPubMed
14.
Zurück zum Zitat Pessis E, Sverzut J-M, Campagna R, et al. (2015) Reduction of metal artifact with dual-energy CT: virtual monospectral imaging with fast kilovoltage switching and metal artifact reduction software. Semin Musculoskelet Radiol 19:446–455CrossRefPubMed Pessis E, Sverzut J-M, Campagna R, et al. (2015) Reduction of metal artifact with dual-energy CT: virtual monospectral imaging with fast kilovoltage switching and metal artifact reduction software. Semin Musculoskelet Radiol 19:446–455CrossRefPubMed
15.
Zurück zum Zitat Shinohara Y, Sakamoto M, Iwata N, et al. (1987) Usefulness of monochromatic imaging with metal artifact reduction software for computed tomography angiography after intracranial aneurysm coil embolization. Acta Radiol Stockh Swed 2014(55):1015–1023 Shinohara Y, Sakamoto M, Iwata N, et al. (1987) Usefulness of monochromatic imaging with metal artifact reduction software for computed tomography angiography after intracranial aneurysm coil embolization. Acta Radiol Stockh Swed 2014(55):1015–1023
16.
Zurück zum Zitat Ascenti G, Mazziotti S, Lamberto S, et al. (2011) Dual-energy CT for detection of endoleaks after endovascular abdominal aneurysm repair: usefulness of colored iodine overlay. AJR Am J Roentgenol 196:1408–1414CrossRefPubMed Ascenti G, Mazziotti S, Lamberto S, et al. (2011) Dual-energy CT for detection of endoleaks after endovascular abdominal aneurysm repair: usefulness of colored iodine overlay. AJR Am J Roentgenol 196:1408–1414CrossRefPubMed
17.
Zurück zum Zitat Buffa V, Solazzo A, D’Auria V, et al. (2014) Dual-source dual-energy CT: dose reduction after endovascular abdominal aortic aneurysm repair. Radiol Med (Torino) 119:934–941CrossRef Buffa V, Solazzo A, D’Auria V, et al. (2014) Dual-source dual-energy CT: dose reduction after endovascular abdominal aortic aneurysm repair. Radiol Med (Torino) 119:934–941CrossRef
18.
Zurück zum Zitat Maturen KE, Kaza RK, Liu PS, et al. (2012) “Sweet spot” for endoleak detection: optimizing contrast to noise using low keV reconstructions from fast-switch kVp dual-energy CT. J Comput Assist Tomogr 36:83–87CrossRefPubMedPubMedCentral Maturen KE, Kaza RK, Liu PS, et al. (2012) “Sweet spot” for endoleak detection: optimizing contrast to noise using low keV reconstructions from fast-switch kVp dual-energy CT. J Comput Assist Tomogr 36:83–87CrossRefPubMedPubMedCentral
19.
Zurück zum Zitat Maturen KE, Kleaveland PA, Kaza RK, et al. (2011) Aortic endograft surveillance: use of fast-switch kVp dual-energy computed tomography with virtual noncontrast imaging. J Comput Assist Tomogr 35:742–746CrossRefPubMed Maturen KE, Kleaveland PA, Kaza RK, et al. (2011) Aortic endograft surveillance: use of fast-switch kVp dual-energy computed tomography with virtual noncontrast imaging. J Comput Assist Tomogr 35:742–746CrossRefPubMed
20.
Zurück zum Zitat Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174CrossRefPubMed Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174CrossRefPubMed
21.
Zurück zum Zitat Huang JY, Kerns JR, Nute JL, et al. (2015) An evaluation of three commercially available metal artifact reduction methods for CT imaging. Phys Med Biol 60:1047–1067CrossRefPubMedPubMedCentral Huang JY, Kerns JR, Nute JL, et al. (2015) An evaluation of three commercially available metal artifact reduction methods for CT imaging. Phys Med Biol 60:1047–1067CrossRefPubMedPubMedCentral
22.
Zurück zum Zitat Han SC, Chung YE, Lee YH, et al. (2014) Metal artifact reduction software used with abdominopelvic dual-energy CT of patients with metal hip prostheses: assessment of image quality and clinical feasibility. AJR Am J Roentgenol 203:788–795CrossRefPubMed Han SC, Chung YE, Lee YH, et al. (2014) Metal artifact reduction software used with abdominopelvic dual-energy CT of patients with metal hip prostheses: assessment of image quality and clinical feasibility. AJR Am J Roentgenol 203:788–795CrossRefPubMed
23.
Zurück zum Zitat Ellmann S, Kammerer F, Brand M, et al. (2016) A novel pairwise comparison-based method to determine radiation dose reduction potentials of iterative reconstruction algorithms, exemplified through circle of willis computed tomography angiography. Invest Radiol 51:331–339PubMed Ellmann S, Kammerer F, Brand M, et al. (2016) A novel pairwise comparison-based method to determine radiation dose reduction potentials of iterative reconstruction algorithms, exemplified through circle of willis computed tomography angiography. Invest Radiol 51:331–339PubMed
24.
Zurück zum Zitat Goetti R, Winklehner A, Gordic S, et al. (2012) Automated attenuation-based kilovoltage selection: preliminary observations in patients after endovascular aneurysm repair of the abdominal aorta. AJR Am J Roentgenol 199:W380–385CrossRefPubMed Goetti R, Winklehner A, Gordic S, et al. (2012) Automated attenuation-based kilovoltage selection: preliminary observations in patients after endovascular aneurysm repair of the abdominal aorta. AJR Am J Roentgenol 199:W380–385CrossRefPubMed
25.
Zurück zum Zitat Sudarski S, Apfaltrer P, Nance JW, et al. (2013) Optimization of keV-settings in abdominal and lower extremity dual-source dual-energy CT angiography determined with virtual monoenergetic imaging. Eur J Radiol 82:e574–581CrossRefPubMed Sudarski S, Apfaltrer P, Nance JW, et al. (2013) Optimization of keV-settings in abdominal and lower extremity dual-source dual-energy CT angiography determined with virtual monoenergetic imaging. Eur J Radiol 82:e574–581CrossRefPubMed
Metadaten
Titel
Dual energy CT angiography: pros and cons of dual-energy metal artifact reduction algorithm in patients after endovascular aortic repair
verfasst von
Johannes Boos
Jieming Fang
Benedikt H. Heidinger
Vassilios Raptopoulos
Olga R. Brook
Publikationsdatum
28.11.2016
Verlag
Springer US
Erschienen in
Abdominal Radiology / Ausgabe 3/2017
Print ISSN: 2366-004X
Elektronische ISSN: 2366-0058
DOI
https://doi.org/10.1007/s00261-016-0973-7

Weitere Artikel der Ausgabe 3/2017

Abdominal Radiology 3/2017 Zur Ausgabe

Classics in Abdominal Imaging

A “doughnut” in the abdomen

Classics in Abdominal Imaging

The “hairy kidney” sign

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.