28.11.2016 | Ausgabe 3/2017

Abdominal Radiology 3/2017

Dual energy CT angiography: pros and cons of dual-energy metal artifact reduction algorithm in patients after endovascular aortic repair

Zeitschrift:
Abdominal Radiology > Ausgabe 3/2017
Autoren:
Johannes Boos, Jieming Fang, Benedikt H. Heidinger, Vassilios Raptopoulos, Olga R. Brook

Abstract

Purpose

To evaluate the value of metal artifact reduction (MAR) post-processing and iodine MD images in fast kV-switching dual-energy computed tomography (DECT) in patients after endovascular aortic repair (EVAR).

Materials and methods

Twenty-four consecutive EVAR patients (age 76 ± 9 years, 7/24 (29%) with coils, 9/24 (37.5%) with 10 endoleaks) who underwent DECT angiography were included in this HIPAA-compliant, IRB-approved retrospective study. Monochromatic reconstructions included 55, 60, 65, 70, and 75 keV with and without MAR and iodine MD images. Near field, far field, and vessel artifacts were assessed subjectively (1 = none; 5 = severe) and objectively by measuring noise and contrast-to-noise ratio. Visibility of endoleak was evaluated (1 = optimal; 5 = not visible).

Results

MAR objectively decreased artifacts from EVAR stents in the near field (60.7 ± 25.4 HU vs. 70.1 ± 34.2; p = .002) and subjectively increased near field (3.2 ± 0.9 vs. 2.8 ± 0.6; p < .001), far field (2.2 ± 0.6 vs. 1.6 ± 0.6; p < .001), and vessel (3.1 ± 1.1 vs. 2.5 ± 0.9; p < .001) artifacts. Near-field artifacts from coils were reduced by the MAR objectively (72.4 ± 24.8 vs. 182.7 ± 57.3 HU; p < .001) and subjectively (4.5 ± 0.5 vs. 4.9 ± 0.4; p = .02). CNR of standard reconstructions was optimal at 60 keV (38.3 ± 16.8). Reconstructions without MAR and iodine MD images provided improved endoleak visualization in 6/10 (60%) of cases (median 1 for both) compared to MAR (median 3) (p < 0.001). However, MAR improved visualization in 1/10 (10%) cases due to endoleak location adjacent to a coil.

Conclusion

DECT with MAR reduced artifacts from coils and improved endoleak visualization in 1/10 (10%) cases due to location adjacent to a coil. However, MAR impaired endoleak visualization in 6/10 (60%) cases and should be reviewed combined with 60 keV standard reconstructions and iodine MD images.

Bitte loggen Sie sich ein, um Zugang zu diesem Inhalt zu erhalten

★ PREMIUM-INHALT
e.Med Interdisziplinär

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de. Zusätzlich können Sie eine Zeitschrift Ihrer Wahl in gedruckter Form beziehen – ohne Aufpreis.

Weitere Produktempfehlungen anzeigen
Literatur
Über diesen Artikel

Weitere Artikel der Ausgabe 3/2017

Abdominal Radiology 3/2017Zur Ausgabe

Classics in Abdominal Imaging

The stepladder sign

Classics in Abdominal Imaging

The “erased charcoal” sign

Classics in Abdominal Imaging

The northern exposure sign

  1. Das kostenlose Testabonnement läuft nach 14 Tagen automatisch und formlos aus. Dieses Abonnement kann nur einmal getestet werden.

Neu im Fachgebiet Radiologie

Meistgelesene Bücher aus der Radiologie

2016 | Buch

Medizinische Fremdkörper in der Bildgebung

Thorax, Abdomen, Gefäße und Kinder

Dieses einzigartige Buch enthält ca. 1.600 hochwertige radiologische Abbildungen und Fotos iatrogen eingebrachter Fremdmaterialien im Röntgenbild und CT.

Herausgeber:
Dr. med. Daniela Kildal

2011 | Buch

Atlas Klinische Neuroradiologie des Gehirns

Radiologie lebt von Bildern! Der vorliegende Atlas trägt dieser Tatsache Rechnung. Sie finden zu jedem Krankheitsbild des Gehirns Referenzbilder zum Abgleichen mit eigenen Befunden.

Autoren:
Priv.-Doz. Dr. med. Jennifer Linn, Prof. Dr. med. Martin Wiesmann, Prof. Dr. med. Hartmut Brückmann

Mail Icon II Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Radiologie und bleiben Sie gut informiert – ganz bequem per eMail.

Bildnachweise