Skip to main content
Erschienen in: Archives of Orthopaedic and Trauma Surgery 10/2005

01.12.2005 | Original Article

Dual energy X-ray absorptiometry analysis of peri-prosthetic stress shielding in the Birmingham resurfacing hip replacement

verfasst von: J. A. Harty, B. Devitt, L. C. Harty, M. Molloy, A. McGuinness

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 10/2005

Einloggen, um Zugang zu erhalten

Abstract

Introduction: Numerous reports in the literature refer to the femoral neck fracture rate in hip resurfacing. The aim of this study was to determine the bone mineral density and evidence of stress shielding around the femoral component of the Birmingham resurfacing prosthesis. Material and methods: Twenty-eight patients with primary unilateral osteoarthritis had a Birmingham resurfacing prosthesis. DEXA analysis of the proximal femur and femoral neck was performed and compared with the opposite unaffected side. Results: Total periprosthetic bone mineral density was 0.49% greater than the control, but this did not achieve statistical significance. Although the BMD of the femoral neck was slightly increased on the prosthetic side (1.002 g/cm2) as opposed to the control side, this difference did not reach statistical significance. Conclusion: The Birmingham resurfacing prosthesis does not appear to reduce femoral neck bone mineral density in comparison to the normal femoral neck bone density. We conclude that femoral neck fractures are unlikely to be due to stress shielding related to the prosthesis.
Literatur
1.
Zurück zum Zitat Bobyn JD, Mortimer ES, Glassman AH et al (1992) Producing and avoiding stress shielding. Laboratory and clinical observations of noncemented total hip arthroplasty. Clin Orthop 274:79–96PubMed Bobyn JD, Mortimer ES, Glassman AH et al (1992) Producing and avoiding stress shielding. Laboratory and clinical observations of noncemented total hip arthroplasty. Clin Orthop 274:79–96PubMed
2.
Zurück zum Zitat Cohen B, Rushton N (1995) Accuracy of DEXA measurement of bone mineral density after total hip arthroplasty. J Bone Joint Surg [Br] 77B:479–483 Cohen B, Rushton N (1995) Accuracy of DEXA measurement of bone mineral density after total hip arthroplasty. J Bone Joint Surg [Br] 77B:479–483
3.
Zurück zum Zitat Jordan GR, Loveridge N, Bell KL et al (2003) Increased femoral neck cancellous bone and connectivity in coxarthrosis (hip osteoarthritis). Bone 32(1):86–95CrossRefPubMed Jordan GR, Loveridge N, Bell KL et al (2003) Increased femoral neck cancellous bone and connectivity in coxarthrosis (hip osteoarthritis). Bone 32(1):86–95CrossRefPubMed
4.
Zurück zum Zitat Kilgus DJ, Shimaoka EE, Tipton JS, Eberle RW (1993) Dual-energy X-ray absorptiometry measurement of bone mineral density around porous-coated cementless femoral implants. Methods and preliminary results. J Bone Joint Surg [Br] 75(2):279–287 Kilgus DJ, Shimaoka EE, Tipton JS, Eberle RW (1993) Dual-energy X-ray absorptiometry measurement of bone mineral density around porous-coated cementless femoral implants. Methods and preliminary results. J Bone Joint Surg [Br] 75(2):279–287
5.
Zurück zum Zitat Kiratli BJ, Heiner JP, McBeath AA, Wilson MA (1992) Determination of bone mineral density by dual X-ray absorptiometry in patients with uncemented total hip arthroplasty. J Orthop Res 10(6):836–844CrossRefPubMed Kiratli BJ, Heiner JP, McBeath AA, Wilson MA (1992) Determination of bone mineral density by dual X-ray absorptiometry in patients with uncemented total hip arthroplasty. J Orthop Res 10(6):836–844CrossRefPubMed
6.
Zurück zum Zitat Kroger H, Miettinen H, Arnala I et al (1996) Evaluation of periprosthetic bone using dual-energy X-ray absorptiometry: precision of the method and effect of operation on bone mineral density. J Bone Miner Res 11(10):1526–1530PubMedCrossRef Kroger H, Miettinen H, Arnala I et al (1996) Evaluation of periprosthetic bone using dual-energy X-ray absorptiometry: precision of the method and effect of operation on bone mineral density. J Bone Miner Res 11(10):1526–1530PubMedCrossRef
7.
Zurück zum Zitat McCarthy CK, Steinberg GG, Agren M et al (1991) Quantifying bone loss from the proximal femur after total hip arthroplasty. J Bone Joint Surg [Br] 73(5):774–778 McCarthy CK, Steinberg GG, Agren M et al (1991) Quantifying bone loss from the proximal femur after total hip arthroplasty. J Bone Joint Surg [Br] 73(5):774–778
8.
Zurück zum Zitat McMinn D, Treacy R, Lin K, Pynsent P (1996) Metal on metal surface replacement of the hip: experience of the McMinn prosthesis. Clin Orthop 329(Suppl):89–98CrossRef McMinn D, Treacy R, Lin K, Pynsent P (1996) Metal on metal surface replacement of the hip: experience of the McMinn prosthesis. Clin Orthop 329(Suppl):89–98CrossRef
9.
Zurück zum Zitat Nishii T, Sugano N, Masuhara K et al (1997) Longitudinal evaluation of time-related bone remodelling after cementless total hip arthroplasty. Clin Orthop 339:121–131PubMedCrossRef Nishii T, Sugano N, Masuhara K et al (1997) Longitudinal evaluation of time-related bone remodelling after cementless total hip arthroplasty. Clin Orthop 339:121–131PubMedCrossRef
10.
Zurück zum Zitat Venesmaa PK, Kroger HP, Miettinen HJ et al (2001) Monitoring of periprosthetic BMD after uncemented total hip arthroplasty with dual-energy X-ray absorptiometry—a 3-year follow-up study. J Bone Miner Res 16(6):1056–1061PubMedCrossRef Venesmaa PK, Kroger HP, Miettinen HJ et al (2001) Monitoring of periprosthetic BMD after uncemented total hip arthroplasty with dual-energy X-ray absorptiometry—a 3-year follow-up study. J Bone Miner Res 16(6):1056–1061PubMedCrossRef
11.
Zurück zum Zitat Watanabe Y, Shiba N, Matsuo S et al (2000) Biomechanical study of the resurfacing hip arthroplasty: finite element analysis of the femoral component. J Arthroplasty 15:505–511CrossRefPubMed Watanabe Y, Shiba N, Matsuo S et al (2000) Biomechanical study of the resurfacing hip arthroplasty: finite element analysis of the femoral component. J Arthroplasty 15:505–511CrossRefPubMed
Metadaten
Titel
Dual energy X-ray absorptiometry analysis of peri-prosthetic stress shielding in the Birmingham resurfacing hip replacement
verfasst von
J. A. Harty
B. Devitt
L. C. Harty
M. Molloy
A. McGuinness
Publikationsdatum
01.12.2005
Verlag
Springer-Verlag
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 10/2005
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-005-0059-4

Weitere Artikel der Ausgabe 10/2005

Archives of Orthopaedic and Trauma Surgery 10/2005 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.