Effect of augmentation techniques on the failure of pedicle screws under cranio-caudal cyclic loading

verfasst von: Richard Bostelmann, Alexander Keiler, Hans Jakob Steiger, Armin Scholz, Jan Frederick Cornelius, Werner Schmoelz

Erschienen in: European Spine Journal | Ausgabe 1/2017

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Augmentation of pedicle screws is recommended in selected indications (for instance: osteoporosis). Generally, there are two techniques for pedicle screw augmentation: inserting the screw in the non cured cement and in situ-augmentation with cannulated fenestrated screws, which can be applied percutaneously. Most of the published studies used an axial pull out test for evaluation of the pedicle screw anchorage. However, the loading and the failure mode of pullout tests do not simulate the cranio-caudal in vivo loading and failure mechanism of pedicle screws. The purpose of the present study was to assess the fixation effects of different augmentation techniques (including percutaneous cement application) and to investigate pedicle screw loosening under physiological cyclic cranio-caudal loading.

Methods

Each of the two test groups consisted of 15 vertebral bodies (L1–L5, three of each level per group). Mean age was 84.3 years (SD 7.8) for group 1 and 77.0 years (SD 7.00) for group 2. Mean bone mineral density was 53.3 mg/cm3 (SD 14.1) for group 1 and 53.2 mg/cm3 (SD 4.3) for group 2. 1.5 ml high viscosity PMMA bone cement was used for all augmentation techniques. For test group 1, pedicles on the right side of the vertebrae were instrumented with solid pedicle screws in standard fashion without augmentation and served as control group. Left pedicles were instrumented with cannulated screws (Viper cannulated, DePuy Spine) and augmented. For test group 2 pedicles on the left side of the vertebrae were instrumented with cannulated fenestrated screws and in situ augmented. On the right side solid pedicle screws were augmented with cement first technique. Each screw was subjected to a cranio-caudal cyclic load starting at 20–50 N with increasing upper load magnitude of 0.1 N per cycle (1 Hz) for a maximum of 5000 cycles or until total failure. Stress X-rays were taken after cyclic loading to evaluate screw loosening.

Results

Test group 1 showed a significant higher number of load cycles until failure for augmented screws compared to the control (4030 cycles, SD 827.8 vs. 1893.3 cycles, SD 1032.1; p < 0.001). Stress X-rays revealed significant less screw toggling for the augmented screws (5.2°, SD 5.4 vs. 16.1°, SD 5.9; p < 0.001). Test group 2 showed 3653.3 (SD 934) and 3723.3 (SD 560.6) load cycles until failure for in situ and cement first augmentation. Stress X-rays revealed a screw toggling of 5.1 (SD 1.9) and 6.6 (SD 4.6) degrees for in situ and cement first augmentation techniques (p > 0.05).

Conclusion

Augmentation of pedicle screws in general significantly increased the number of load cycles and failure load comparing to the nonaugmented control group. For the augmentation technique (cement first, in situ augmented, percutaneously application) no effect could be exhibited on the failure of the pedicle screws. By the cranio-caudal cyclic loading failure of the pedicle screws occurred by screw cut through the superior endplate and the characteristic “windshield-wiper effect”, typically observed in clinical practice, could be reproduced.
Literatur
1.
Zurück zum Zitat Office GFS (2002–2010) Total cost of illness in millions of Euro. Classification: years, gender, ICD10, provider (Primary source: Federal Statistical Office, Total Health Expenditure, Statistics). In., Berlin Office GFS (2002–2010) Total cost of illness in millions of Euro. Classification: years, gender, ICD10, provider (Primary source: Federal Statistical Office, Total Health Expenditure, Statistics). In., Berlin
2.
Zurück zum Zitat Masala S, Lunardi P, Fiori R, Liccardo G, Massari F, Ursone A, Simonetti G (2004) Vertebroplasty and kyphoplasty in the treatment of malignant vertebral fractures. J Chemother 16(Suppl 5):30–33CrossRef Masala S, Lunardi P, Fiori R, Liccardo G, Massari F, Ursone A, Simonetti G (2004) Vertebroplasty and kyphoplasty in the treatment of malignant vertebral fractures. J Chemother 16(Suppl 5):30–33CrossRef
3.
Zurück zum Zitat Melton LJ 3rd, Thamer M, Ray NF, Chan JK, Chesnut CH 3rd, Einhorn TA, Johnston CC, Raisz LG, Silverman SL, Siris ES (1997) Fractures attributable to osteoporosis: report from the National Osteoporosis Foundation. J Bone Mineral Res 12:16–23. doi:10.1359/jbmr.1997.12.1.16 CrossRef Melton LJ 3rd, Thamer M, Ray NF, Chan JK, Chesnut CH 3rd, Einhorn TA, Johnston CC, Raisz LG, Silverman SL, Siris ES (1997) Fractures attributable to osteoporosis: report from the National Osteoporosis Foundation. J Bone Mineral Res 12:16–23. doi:10.​1359/​jbmr.​1997.​12.​1.​16 CrossRef
4.
Zurück zum Zitat Papaioannou A, Watts NB, Kendler DL, Yuen CK, Adachi JD, Ferko N (2002) Diagnosis and management of vertebral fractures in elderly adults. Am J Med 113:220–228CrossRefPubMed Papaioannou A, Watts NB, Kendler DL, Yuen CK, Adachi JD, Ferko N (2002) Diagnosis and management of vertebral fractures in elderly adults. Am J Med 113:220–228CrossRefPubMed
5.
Zurück zum Zitat Becker S, Chavanne A, Spitaler R, Kropik K, Aigner N, Ogon M, Redl H (2008) Assessment of different screw augmentation techniques and screw designs in osteoporotic spines. Eur Spine Jy 17:1462–1469. doi:10.1007/s00586-008-0769-8 CrossRef Becker S, Chavanne A, Spitaler R, Kropik K, Aigner N, Ogon M, Redl H (2008) Assessment of different screw augmentation techniques and screw designs in osteoporotic spines. Eur Spine Jy 17:1462–1469. doi:10.​1007/​s00586-008-0769-8 CrossRef
7.
Zurück zum Zitat Chen LH, Tai CL, Lai PL, Lee DM, Tsai TT, Fu TS, Niu CC, Chen WJ (2009) Pullout strength for cannulated pedicle screws with bone cement augmentation in severely osteoporotic bone: influences of radial hole and pilot hole tapping. Clin Biomech (Bristol, Avon) 24:613–618. doi:10.1016/j.clinbiomech.2009.05.002 CrossRef Chen LH, Tai CL, Lai PL, Lee DM, Tsai TT, Fu TS, Niu CC, Chen WJ (2009) Pullout strength for cannulated pedicle screws with bone cement augmentation in severely osteoporotic bone: influences of radial hole and pilot hole tapping. Clin Biomech (Bristol, Avon) 24:613–618. doi:10.​1016/​j.​clinbiomech.​2009.​05.​002 CrossRef
9.
Zurück zum Zitat Zindrick MR, Wiltse LL, Widell EH, Thomas JC, Holland WR, Field BT, Spencer CW (1986) A biomechanical study of intrapeduncular screw fixation in the lumbosacral spine. Clin Orthop Relat Res (203):99–112 Zindrick MR, Wiltse LL, Widell EH, Thomas JC, Holland WR, Field BT, Spencer CW (1986) A biomechanical study of intrapeduncular screw fixation in the lumbosacral spine. Clin Orthop Relat Res (203):99–112
10.
Zurück zum Zitat Rohlmann A, Bergmann G, Graichen F (1997) Loads on an internal spinal fixation device during walking. J Biomech 30:41–47CrossRefPubMed Rohlmann A, Bergmann G, Graichen F (1997) Loads on an internal spinal fixation device during walking. J Biomech 30:41–47CrossRefPubMed
11.
Zurück zum Zitat Rohlmann A, Bergmann G, Graichen F, Weber U (1997) Comparison of loads on internal spinal fixation devices measured in vitro and in vivo. Med Eng Phys 19:539–546CrossRefPubMed Rohlmann A, Bergmann G, Graichen F, Weber U (1997) Comparison of loads on internal spinal fixation devices measured in vitro and in vivo. Med Eng Phys 19:539–546CrossRefPubMed
12.
Zurück zum Zitat Rohlmann A, Graichen F, Bergmann G (2000) Influence of load carrying on loads in internal spinal fixators. J Biomech 33:1099–1104CrossRefPubMed Rohlmann A, Graichen F, Bergmann G (2000) Influence of load carrying on loads in internal spinal fixators. J Biomech 33:1099–1104CrossRefPubMed
13.
Zurück zum Zitat Wu JC, Chen CS, Yip SW, Hsu ML (2012) Stress distribution and micromotion analyses of immediately loaded implants of varying lengths in the mandible and fibular bone grafts: a three-dimensional finite element analysis. Int J Oral Maxillofac Implants 27:e77–e84PubMed Wu JC, Chen CS, Yip SW, Hsu ML (2012) Stress distribution and micromotion analyses of immediately loaded implants of varying lengths in the mandible and fibular bone grafts: a three-dimensional finite element analysis. Int J Oral Maxillofac Implants 27:e77–e84PubMed
14.
Zurück zum Zitat Felsenberg D, Gowin W (1999) Bone densitometry by dual energy methods. Der Radiologe 39:186–193CrossRefPubMed Felsenberg D, Gowin W (1999) Bone densitometry by dual energy methods. Der Radiologe 39:186–193CrossRefPubMed
17.
18.
Zurück zum Zitat Goel VK, Winterbottom JM, Weinstein JN (1994) A method for the fatigue testing of pedicle screw fixation devices. J Biomech 27:1383–1388CrossRefPubMed Goel VK, Winterbottom JM, Weinstein JN (1994) A method for the fatigue testing of pedicle screw fixation devices. J Biomech 27:1383–1388CrossRefPubMed
19.
Zurück zum Zitat Law M, Tencer AF, Anderson PA (1993) Caudo-cephalad loading of pedicle screws: mechanisms of loosening and methods of augmentation. Spine 18:2438–2443CrossRefPubMed Law M, Tencer AF, Anderson PA (1993) Caudo-cephalad loading of pedicle screws: mechanisms of loosening and methods of augmentation. Spine 18:2438–2443CrossRefPubMed
20.
Zurück zum Zitat Kueny RA, Kolb JP, Lehmann W, Puschel K, Morlock MM, Huber G (2014) Influence of the screw augmentation technique and a diameter increase on pedicle screw fixation in the osteoporotic spine: pullout versus fatigue testing. Eur Spine J. doi:10.1007/s00586-014-3476-7 Kueny RA, Kolb JP, Lehmann W, Puschel K, Morlock MM, Huber G (2014) Influence of the screw augmentation technique and a diameter increase on pedicle screw fixation in the osteoporotic spine: pullout versus fatigue testing. Eur Spine J. doi:10.​1007/​s00586-014-3476-7
21.
Zurück zum Zitat Fan H, Zhang R, Shen C, Dong F, Li Y, Song P, Gong C, Wang Y (2000) The biomechanical properties of pedicle screw fixation combined with trajectory bone cement augmentation in osteoporotic vertebrae. J Spin Disord Tech. doi:10.1097/BSD.1090b1013e3182a14870 Fan H, Zhang R, Shen C, Dong F, Li Y, Song P, Gong C, Wang Y (2000) The biomechanical properties of pedicle screw fixation combined with trajectory bone cement augmentation in osteoporotic vertebrae. J Spin Disord Tech. doi:10.​1097/​BSD.​1090b1013e3182a1​4870
24.
25.
Zurück zum Zitat Sarzier JS, Evans AJ, Cahill DW (2002) Increased pedicle screw pullout strength with vertebroplasty augmentation in osteoporotic spines. J Neurosurg 96:309–312PubMed Sarzier JS, Evans AJ, Cahill DW (2002) Increased pedicle screw pullout strength with vertebroplasty augmentation in osteoporotic spines. J Neurosurg 96:309–312PubMed
27.
Zurück zum Zitat Rohlmann A, Graichen F, Weber U, Bergmann G (2000) 2000 Volvo Award winner in biomechanical studies: monitoring in vivo implant loads with a telemeterized internal spinal fixation device. Spine 25:2981–2986CrossRefPubMed Rohlmann A, Graichen F, Weber U, Bergmann G (2000) 2000 Volvo Award winner in biomechanical studies: monitoring in vivo implant loads with a telemeterized internal spinal fixation device. Spine 25:2981–2986CrossRefPubMed
28.
Metadaten
Titel
Effect of augmentation techniques on the failure of pedicle screws under cranio-caudal cyclic loading
verfasst von
Richard Bostelmann
Alexander Keiler
Hans Jakob Steiger
Armin Scholz
Jan Frederick Cornelius
Werner Schmoelz
Publikationsdatum
27.03.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
European Spine Journal / Ausgabe 1/2017
Print ISSN: 0940-6719
Elektronische ISSN: 1432-0932
DOI
https://doi.org/10.1007/s00586-015-3904-3

Weitere Artikel der Ausgabe 1/2017

European Spine Journal 1/2017 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.