To evaluate and compare the effect of decentration and tilt on the optical quality of monofocal and trifocal intraocular lenses (IOL).
Methods
Optical quality of a monofocal IOL (AcrySof IQ SN60WF; Alcon Laboratories, Inc., USA) and a trifocal IOL (AcrySof IQ PanOptix; Alcon Laboratories, Inc., USA) was assessed using an in vitro optical bench (OptiSpheric IOL R&D; Trioptics GmbH, Germany). At apertures of 3.0 mm and 4.5 mm, modulation transfer function (MTF) at spatial frequency of 50 lp/mm, MTF curve and the United States Air Force (USAF) resolution test chart of the two IOLs were measured and compared at their focus with different degrees of decentration and tilt. Optical quality at infinity, 60 cm and 40 cm and the through-focus MTF curves were compared when the two IOLs were centered at apertures of 3.0 mm and 4.5 mm. Spectral transmittance of the two IOLs was measured by the UV–visible spectrophotometer (UV 3300 PC; MAPADA, China).
Results
The SN60WF and the PanOptix filtered blue light from 400 to 500 nm. Both IOLs at the far focus and the PanOptix at the intermediate focus showed a decrease in optical quality with increasing decentration and tilt. The PanOptix demonstrated enhanced optical quality compared to the previous gradient at the near focus at a decentration range of 0.3–0.7 mm with a 3.0 mm aperture, and 0.5 mm with a 4.5 mm aperture, whereas other conditions exhibited diminished optical quality with increasing decentration and tilt at the focus of both IOLs. When the two IOLs were centered, the SN60WF had better optical quality at infinity, while the PanOptix had better optical quality at 60 cm and 40 cm defocus. The optical quality of the SN60WF exceeded that of the PanOptix at far focus, with a 3 mm aperture decentration up to 0.7 mm and a 4.5 mm aperture decentration up to 0.3 mm; this observation held true for all tilts, irrespective of aperture size. As both decentration and tilt increased, the optical quality of the SN60WF deteriorated more rapidly than that of the PanOptix at the far focal point.
Conclusions
The SN60WF showed a decrease in optical quality with increasing decentration and tilt. Optical quality of the PanOptix at the near focus increased in some decentration conditions and decreased in some conditions, while it showed a decrease at the other focuses with increasing decentration. While tilt only had a negative effect on optical quality. When both IOLs were centered, the PanOptix provided a wider range of vision, while the SN60WF provided better far distance vision. At the far focus, the SN60WF has better resistance to tilt than the PanOptix, but the optical quality degrades more quickly when decentered and tilted.
Die Auswertung von Gesichtsfeldbefunden ermöglicht in vielen Fällen eine genaue Topodiagnostik und kann auch häufig direkt zu einer konkreten Verdachtsdiagnose führen. Im klinischen Alltag ist dabei die 30° automatische Rasterperimetrie die …
Das Glaukom ist eine der Hauptursachen für Sehbehinderung und Blindheit in Deutschland und weltweit. Der mögliche Einfluss von Ernährung sowie Nahrungsergänzungsmitteln auf den Verlauf und die Therapie dieser Erkrankung werden im klinischen Alltag …
Das trockene Auge gilt bei Erwachsenen ab dem 50. Lebensjahr mittlerweile als „Volkskrankheit“. Deshalb ist das Interesse an Nahrungsmitteln und Nährstoffen mit antioxidativen oder antientzündlichen Eigenschaften ergänzend zu Augentropfen o.ä, in den letzten Jahren deutlich gestiegen.
Die mit dem humanen Leukozytenantigen(HLA)-B27 assoziierte, akute anteriore Uveitis (AAU) ist die häufigste Form nichtinfektiöser intraokulärer Entzündungen und wird als eigene klinische Entität gesehen. Es sind überwiegend junge Erwachsene im …