Open Access 09.10.2020 | Original Article

Effect of delayed entry on performance of the BACT/ALERT FAN PLUS bottles in the BACT/ALERT VIRTUO blood culture system

verfasst von: Mary Adamik, Anne Hutchins, Jasmin Mangilit, Betsy Katzin, Heather Totty, Parampal Deol

Erschienen in: European Journal of Clinical Microbiology & Infectious Diseases | Ausgabe 4/2021

Abstract

Delayed entry of patient blood culture samples into a microbial detection system is unavoidable at times, due to off-shift staffing or transporting samples to centralized laboratories. Pre-incubation time and temperature of blood culture bottles are the most critical factors impacting recovery and detection of microorganisms. A total of 1377 BACT/ALERT® (BTA) Fastidious Antimicrobial Neutralization (FAN® PLUS) bottles (FA PLUS, FN PLUS, and PF PLUS) were tested after delayed entry times of 24 and 36 h at 20–25 °C (room temperature, RT) prior to loading into the BACT/ALERT® VIRTUO® microbial detection system (VIRTUO). Clinically relevant organisms were inoculated into bottles with 5–84 colony forming units (CFU) per bottle, and human blood (0 to 10 mL), and then loaded into the VIRTUO. When bottles were loaded without delay, a mean time to detection (TTD) of 9.6 h was observed. For delayed bottles, the TTD reported by the VIRTUO was added to the 24-h and 36-h delay times and resulted in average time to results of 32.5 h and 42.5 h, respectively. The FAN PLUS bottles in conjunction with the VIRTUO produced acceptable results when delays up to 24 h at 20–25 °C occur in loading.
Hinweise

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

In 2017, the World Health Assembly (WHA) and World Health Organization (WHO) made sepsis a global health priority that causes approximately six million deaths worldwide each year [1]. The WHA resolved to improve the prevention, diagnosis, and management of sepsis. Increased mortality rates correlated with delays in diagnosis of blood stream infections (BSI) and subsequent delays in initiation of effective treatment [2]. Blood culture bottles continue to be the gold standard for isolating the infectious agent from a patient. Ideally, blood or sterile body fluids are inoculated directly into culture bottles at the patient’s bedside and then transported to a clinical microbiology laboratory without delay. However, some hospitals have microbiology labs that are closed or under-staffed during off-shift, weekends, and holidays, or, do not have the facilities or equipment to perform blood cultures in-house. Without an in-house laboratory, samples are collected and transported by courier at scheduled times to a central microbiology laboratory. During transport, delays can occur due to courier schedules and there may be temperature variations due to storage temperature or refrigeration of the samples.
Few studies have been conducted that investigate the effects of time between blood culture inoculation and loading of bottles onto a detection system on the recovery rates of microorganisms. Earlier delayed entry studies were conducted with smaller sample sizes, prior versions of the detection system BACT/ALERT® 3D (BTA 3D), and FAN blood culture bottles that contain charcoal [35]. The delayed entry performance of the recently launched BTA FA PLUS bottle, which contains adsorbent polymeric beads, has only been reported for a single non-fastidious organism, Escherichia coli, in bottles held at room temperature (RT) for 4 h, 8 h, or 12 h before loading [6]. The study described in this manuscript evaluated the performance of aerobic and anaerobic blood culture bottles seeded with a panel of microorganisms commonly found in BSIs, including fastidious bacteria, when held 24 h and 36 h at RT prior to loading into the VIRTUO. The VIRTUO is a next-generation BTA instrument that detects microbial growth and CO2 production by optically monitoring the reflectance of a colorimetric sensor in each bottle over time. The VIRTUO is equipped with new algorithms designed to enhance the sensitivity and specificity and deliver significantly faster TTDs for most microorganisms tested compared with the BTA 3D and BACTEC™ FX (BD Diagnostics, Franklin Lakes, NJ, USA) automated blood culture detection systems [79]. Two types of aerobic bottles, BTA FA PLUS and BTA PF PLUS, and anaerobic BTA FN PLUS bottles, were inoculated with a panel of clinically relevant microorganisms and loaded into the VIRTUO. Both aerobic and anaerobic bottles were tested as it is best practice to collect clinical specimens into both bottle types to increase the amount of blood drawn from a patient, which in turn leads to better recovery, as well as better detection of facultative anaerobes common in BSI [6]. This study demonstrates the performance of FAN PLUS bottles in conjunction with the VIRTUO when bottles encounter typical delay times in loading of 24 h and 36 h as reported in some clinical laboratories.

Materials and methods

Study design

Aerobic bottles, FA PLUS and PF PLUS (pediatric) (bioMérieux, Durham, NC), were seeded with aerobic and facultative anaerobic organisms (Table 1). Anaerobic bottles, FN PLUS (bioMérieux, Durham, NC), were seeded with anaerobic and facultative anaerobic organisms (Table 1). Testing was performed by the R&D Microbiology department at the bioMérieux, Durham, NC site.
Table 1
Delayed entry test panel
Bottle type
Microorganism
Blood volume (mL)
Strain(s)d
0
1
4
10
FA PLUS/PF PLUS
Acinetobacter baumanniia
x
 
x
x
19606, clinical isolate 102460, clinical isolate 105298
Candida albicans
x
 
x
x
14053
Enterococcus faecium
x
 
x
x
35667
Escherichia coli
x
 
x
x
25922
Haemophilus influenzaec
 
x
x
x
10211
Pseudomonas aeruginosaa
x
 
x
x
9027, 27853
Neisseria meningitidisb,c
  
x
x
13090
Staphylococcus aureus
x
 
x
x
25923
Streptococcus pneumoniaec
  
x
x
6305, 49619, clinical isolate 10076
FN PLUS
Bacteroides fragilis
 
x
 
x
25285
Clostridium perfringens
x
  
x
13124
Enterococcus faecium
x
  
x
35667
Escherichia coli
x
  
x
25922
Staphylococcus aureus
x
  
x
25923
Streptococcus pneumoniaec
   
x
6305, 49619, clinical isolate 10076
aNon-fermentative organisms
bTemperature-sensitive organisms
cOrganisms associated with pediatric samples [10]
dATCC strain, unless otherwise noted

Blood

Human blood from healthy willing donors was collected into blood collection tubes containing 0.35% sodium polyanetholesulfate (SPS) in 0.85% sodium chloride (Becton Dickenson, Franklin Lakes, NJ). Blood donors signed an informed consent form in accordance with company policies. The collected blood was pooled and then added aseptically to bottles on the day of collection as needed.
The FA PLUS and FN PLUS bottles were tested with blood (10 mL, recommended blood volume for adult samples), and without blood to represent sterile body fluids, which typically do not contain blood. The PF PLUS bottles were inoculated with 4 mL of blood, the recommended blood volume for the pediatric bottle. Fastidious organisms requiring blood, such as Bacteroides fragilis (B. fragilis) and Haemophilus influenzae (H. influenzae), were tested with a minimum blood volume of 1 mL, or in the case of Neisseria meningitidis (N. meningitidis), 4 mL of blood. Blood sterility was ensured by testing un-inoculated bottles containing only blood on each test date. These bottles also served as negative controls.

Organisms

Nine microorganisms were tested in FA PLUS and PF PLUS bottles, and six microorganisms in the FN PLUS bottles (Table 1). One isolate per microorganism was tested, except for the following: Acinetobacter baumannii (A. baumannii) and Streptococcus pneumoniae (S. pneumoniae) (two additional strains), and Pseudomonas aeruginosa (P. aeruginosa) (one additional strain). Each strain was tested in triplicate per bottle type, blood volume (depending on organism growth requirement), hold time, and instrument (three VIRTUOs) using a single inoculum suspension. Repeat testing was performed for a subset of bottles seeded with N. meningitidis due to an instrument loading error. Organisms were selected based on clinical significance, and as representative organisms that challenge the detection technology and algorithms of the VIRTUO, such as fastidious organisms, organisms sensitive to temperature fluctuations, or low CO2 producers [11, 12]. Multiple strains of A. baumannii, P. aeruginosa, and S. pneumoniae were tested due to observed recovery failures in previous delayed entry studies [35]. The organisms H. influenzae, N. meningitidis, and S. pneumoniae were selected due to their prevalence in pediatric bacteremic patients as well as their fastidious nature and poor recovery in CSF samples when subjected to long transit times and variation in temperatures [10, 13].
Each inoculum was prepared from fresh isolates on solid media. Aerobic microorganisms were suspended in tryptic soy broth (TSB) (bioMérieux, Durham, NC) and anaerobic microorganisms were suspended in Schaedler’s broth (Accumedia, Lansing, MI) to achieve an 85–90% transmittance at 660 nm, or an equivalent McFarland value (0.30 to 0.50) on a DensiCHEK Plus (bioMérieux, Durham, NC). Each concentrated suspension was then serially diluted to a target inoculum concentration of ≤ 200 CFU/mL such that upon inoculation with 0.5 mL, each bottle would receive ≤ 100 CFU/bottle. To determine the actual number of CFUs added to the bottles, a portion of each inoculum was plated on standard media. Enumeration demonstrated the actual range was 5–84 CFU/bottle and inocula were pure. Note that anaerobic organisms were prepared, inoculated into bottles, and plated for purity and enumeration in an anaerobic chamber.

Method

Bottles inoculated with and without blood and/or organism were pre-incubated at specified times prior to loading in the VIRTUO. Control “No Delay” bottles were loaded immediately after inoculation. Test bottles were held 24 h and 36 h at 20–25 °C (RT), as it is not recommended to incubate at 37 °C or to refrigerate bottles when there is a delay in loading, per the IFU (instructions for use) of the BTA blood culture bottles. All bottles were subcultured to solid media upon removal from the VIRTUO. Due to the autolytic nature of S. pneumoniae, which can lead to no growth on solid media, a latex agglutination kit (Wellcogen, Kent, UK) was also used to confirm bottles were inoculated with S. pneumoniae. Colony morphologies (or agglutination test results) from positive bottles were observed to ensure purity of the inoculum and/or to confirm the intended microorganism was tested. Negative bottles remained in the instrument for 5 days until declared negative, and were then subcultured (or tested by agglutination) to verify the absence of an organism. Negative bottles with growth upon subculture or with a positive agglutination test would be deemed false negatives. The TTD for all bottles was recorded by the VIRTUO software.

Limitations

A limitation of this study is the use of seeded simulated blood cultures inoculated with a controlled amount of organism and supplemented with blood from willing healthy donors, as opposed to using clinical samples from known BSI patients.

Results

A total of 1377 bottles were tested in this delayed entry study. When aerobic and anaerobic bottles were loaded without delay, an overall detection rate of 100% (460/460) was observed. Similarly, when bottles were held for 24 h, 100% (458/458) of bottles were detected. Less than 100% was observed with bottles held for 36 h (Table 2). A decrease in the detection rate from 100.0 to 99.6% was observed as the hold time increased from no delay and 24 to 36 h. Using a chi-square equality of proportion test, no significant difference was detected between detection rates for the no delay, 24-h, and 36-h delayed hold times (p value = 0.2214). When bottles were loaded without delay, a mean TTD of 13.8 h was observed. The delay times of 24 h and 36 h resulted in average TTD of 8.5 h and 6.5 h, respectively (Table 2). The observed faster TTDs for the hold times of 24 h and 36 h demonstrate microbial growth occurred at RT prior to loading. The delay time needs to be added to the TTD to determine the time to patient results as any delay in loading may lead to a delay in patient care.
Table 2
Overall recovery at delay times and temperatures for FA PLUS, PF PLUS, and FN PLUS
Delayed entry time (h)
Delayed entry holding temperature (°C)
Avg. TTD (h)
Hold time + Avg. TTD (h)
TTD range + hold time (h)
Detection rate %
# (+)/n
No delay
No delay
13.8
13.8
8.5–40.5
100 (460/460)
24
20–25 °C
8.5
32.5
27.4–54.3
100 (458/458)
36
20–25 °C
6.5
42.5
37.5–65.1
99.6 (457/459)
Grand total
99.9 (1375/1377)a
aTwo bottles seeded with Acinetobacter baumannii were negative when held at RT for 36 h. One of these bottles was a false negative

Delayed entry performance of aerobic bottles

Aerobic FA PLUS and PF PLUS bottles were evaluated using nine aerobic and facultative anaerobic microorganisms (Table 1). An overall detection rate of 99.7% (646/648) was observed for the FA PLUS bottles tested. All FA PLUS bottles were declared positive when loaded without delay, or with a 24-h delay. Two FA PLUS bottles, seeded with A. baumannii and held at RT for 36 h, were declared negative at 5 days. Of the two instrument-declared negative FA PLUS bottles, one was determined to be a false negative based on subculture. This was the only false negative observed out of 1377 bottles tested. For PF PLUS, a detection rate of 100% was observed for all bottles loaded (379/379) (Table 3).
Table 3
Result of delayed entry on organism TTD in FA PLUS, PF PLUS, and FN PLUS bottles on VIRTUO
Organism (strain #) actual CFU/bottle
Delayed entry time (h) and holding temperature (°C)
No delay
24 h (20–25 °C)
36 h (20–25 °C)
Avg. TTD (h)
TTD range (h)
# (+)/n
Detectionrate (%)
Avg. TTD (h)
TTD range (h)
# (+)/n
Detectionrate (%)
Avg.TTD (h)
TTDrange (h)
# (+)/n
Detectionrate (%)
Acinetobacter baumannii (19606) 11 CFU, (102460) 9 CFU, (105298) 5 CFU
12.0
10.1–14.7
81/81
100
6.0
3.8–10.0
81/81
100
3.8
2.7–7.7
79/81
97.5
Bacteroides fragilis (25285) 62 CFU
32.7
27.5–40.5
18/18
100
24.5
20.2–30.3
17/17
100
20.2
15.4–25.8
18/18
100
Candida albicans (14053) 26 CFU
26.9
23.0–30.5
26/26
100
12.4
10.0–14.0
27/27
100
5.8
2.8–9.8
27/27
100
Clostridium perfringens (13124) 47 CFU
10.3
9.1–13.0
18/18
100
6.8
5.5–9.6
18/18
100
3.1
2.8–3.7
18/18
100
Enterococcus faecium (35667) 12 CFU
10.6
9.2–15.2
45/45
100
4.2
3.4–5.5
45/45
100
2.4
1.5–3.2
45/45
100
Escherichia coli (25922) 16 CFU
9.7
9.0–10.9
45/45
100
4.1
3.5–5.5
45/45
100
2.3
1.7–2.9
45/45
100
Haemophilus influenzae (10211) 26 CFU
12.5
11.7–13.3
27/27
100
10.4
9.2–12.0
27/27
100
9.3
8.5–10.0
27/27
100
Neisseria meningitidis (13090) 40, 50 CFU
18.8
15.7–26.2
20/20
100
18.4
16.1–22.6
18/18
100
21.3
16.6–29.1
18/18
100
Pseudomonas aeruginosa (9027) 22 CFU, (27853) 30 CFU
14.3
11.2–16.3
54/54
100
7.2
6.1–9.7
54/54
100
4.3
3.2–6.0
54/54
100
Staphylococcus aureus (25923) 25 CFU
13.2
10.0–21.2
45/45
100
9.1
5.5–16.7
45/45
100
7.0
3.8–14.0
45/45
100
Streptococcus pneumoniae (6305) 11 CFU,(49619) 84 CFU, (10076) 9 CFU
11.3
8.5–14.7
81/81
100
9.2
7.7–14.1
81/81
100
8.8
6.4–13.0
81/81
100

Delayed entry performance of anaerobic bottles

Anaerobic FN PLUS bottles were tested with six microorganism species comprised of both strict and facultative anaerobes (Table 1). Seeded FN PLUS bottles demonstrated a 100% (350/350) detection rate at all delay conditions (Table 3).

Time to detection and % detection by organism

Ten of the 11 organisms tested were detected 100% of the time regardless of the delay time: B. fragilis, Candida albicans, Clostridium perfringens, Enterococcus faecium, E. coli, H. influenzae, N. meningitidis, P. aeruginosa, Staphylococcus aureus, and S. pneumoniae. The organism A. baumannii was negatively impacted by a delay in loading. Overall, the performance of FAN PLUS bottles, when held at room temperature for 24 h, demonstrated 100% detection (Table 3).

Discussion

Early detection of BSI is a critical factor in preventing septic shock mortality. A retrospective study showed that patient survival drops an average of 7.6% for every additional hour before administration of effective antimicrobial initiation within the first 6 h post-onset of septic shock hypotension [2]. Current standard of care for suspected sepsis patients includes collection of blood cultures, both aerobic and anaerobic, at the same time as administration of a broad spectrum antibiotics [14]. Early empirical therapy is critical in the treatment of patients with sepsis, but should be used for a limited time due to the potential of developing resistant bacterial strains [15, 16]. Organism identification and antimicrobial susceptibility test (AST) results are essential for antibiotic de-escalation, defined as a discontinuation or change in antimicrobial agents. De-escalation results in lower patient mortality and better antimicrobial stewardship [17, 18].
To ensure fast and accurate diagnostic results, blood culture bottles should be loaded into the detection system as soon as possible. Unfortunately, delays in loading of inoculated blood culture bottles are routinely encountered in clinical laboratories due to reduced off-shift staffing and transportation delays from satellite hospitals to a central laboratory. The majority of studies evaluating delayed entry were conducted prior to introduction of the BTA FAN PLUS bottles and the next-generation BTA detection system, the VIRTUO [4, 5]. Only one published study has evaluated the FA PLUS bottle in conjunction with the VIRTUO in a delayed entry study, but only a single organism and temperature was tested [6].
This study was designed to challenge the VIRTUO detection system using multiple aspects of the blood culture clinical workflow, such as different bottle types, blood volumes, inoculum concentrations of 5–84 CFU/bottle, and storage times defined in previous studies [1922]. Both pediatric and adult recommended blood volumes were tested. In addition, bottles were tested without blood to represent sterile body fluids. Compared with previous delayed entry studies, the FA PLUS, PF PLUS, and FN PLUS bottles had improved detection rates and faster TTD on the VIRTUO, when loaded without delay, and held 24 h at room temperature [4, 5]. These results support an improvement in the BTA detection system consisting of the optimized FAN PLUS bottle formulations and algorithm on VIRTUO [23, 24]. When bottles were delayed 36 h, the VIRTUO demonstrated a detection rate of less than 100% for A. baumannii, which highlights the importance of loading bottles into the instrument as soon as possible.
Since the detection technology is based on the production of CO2 by the organism, low CO2-producing organisms were used to test the system under delayed entry conditions. Additional strains of A. baumannii and P. aeruginosa were included due to previously reported false negatives during delayed entry testing when pre-incubated at 36 °C for greater than 8 h using the predicate FAN media, containing charcoal, on the BTA 3D [3, 5]. An improvement in detection for A. baumannii in aerobic bottles was observed with an overall detection rate of 99.2% on the VIRTUO, with one false negative observed and one true-negative bottle, both occurring when bottles were held at RT for 36 h. The detection rate for P. aeruginosa was 100% for all hold times and temperatures.
In addition, three strains of S. pneumoniae were included as false negatives have been observed, in previous studies, when FAN charcoal bottles were pre-incubated at 36 °C for longer than 24 h [4]. All three FAN PLUS bottle types inoculated with S. pneumoniae demonstrated 100% detection at all hold times and temperatures when tested on the VIRTUO.
The FN PLUS bottle demonstrated a 100% (350/350) detection rate when tested at every pre-incubation condition, organism, and blood volume. The FN PLUS bottle performance reinforces the best practice of collecting patient samples into both aerobic and anaerobic blood culture bottles to have the greatest chance of detecting a wide range of organisms, in particular, facultative anaerobes such as Staphylococcus spp., Streptococcus spp., and E. coli, which are common in BSIs.
This paper supports the conclusions from previous studies as well as the package insert recommendations to load bottles as soon as possible, but to store at room temperature if a delay is unavoidable [4, 12, 19, 25]. Additionally, guidelines from CLSI, ASM, and Cumitech recommend delivery to the lab within 2 h of collection and, if there is a delay in loading, storage at room temperature [2628]. When bottles are loaded immediately per the recommendations, a 100% detection rate was achieved for all FAN PLUS bottle types on the VIRTUO system.
Even though patients suspected of BSI are typically quite sick, the blood cultures collected from them are not necessarily treated with urgency [29]. Significant attention has been paid to certain aspects of blood culture such as media composition, sample volume, and detection systems methods; however, few studies have focused on the logistics of blood culture samples during transportation [30].
With the VIRTUO’s automated loading, it is easy for staff with minimal training to load bottles onto the instrument during off-shift staffing of the laboratory; bottles are placed on the conveyor and the instrument loads the bottles via a robotic arm. VIRTUO instruments could be placed at strategic locations throughout the hospital, or at point of care, where blood cultures could be loaded onto the instrument without delay. This workflow would reduce the time to patient results and subsequent identification of the microorganism, and result in a quicker time to de-escalation of empirical therapy [29]. This de-escalation in therapy not only benefits the patient with optimal care but also decreases the cost of unnecessary treatment, and reduces the risk of developing resistant organisms, which is becoming a threat worldwide [22].
Since the identity of the causative agent is not known when the patient sample is collected, it is imperative to load the bottle into the instrument as soon as possible after collection. However, when delays occur, the FAN PLUS bottles and the VIRTUO demonstrate acceptable performance when samples are held at room temperature for up to 24 h. Minimizing delays and the use of the FAN PLUS bottle paired with the VIRTUO will improve detection rates of organisms, such as A. baumannii, and allow faster reporting to the clinician. Additionally, FAN PLUS bottles have the capability to neutralize antimicrobials [31] allowing for the detection and recovery of microorganisms from patients already receiving antimicrobial therapy, which may account for the majority (82%) of blood draws [32]. Implementation of the BACT/ALERT detection system including FAN PLUS bottles and the VIRTUO supports the most robust and fastest time to diagnosis even when delays in loading occur, thus, improving the management of sepsis. Ultimately, the adoption of these recommended practices can improve patient care with effective, targeted therapy; decrease the overall cost of treatment; and reduce the incidence of antimicrobial resistance.

Code availability

Not applicable.

Compliance with ethical standards

Conflict of interest

All authors are employees of bioMérieux, Inc., manufacturer of the BACT/ALERT® VIRTUO® and BACT/ALERT® FAN PLUS blood culture bottles used in this manuscript.

Ethics approval

This study was conducted in accordance with policies and procedures of bioMérieux, Inc.
Informed consent was obtained from healthy blood donors in accordance with company policy. No patient data was obtained.
Not applicable.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Innere Medizin

Kombi-Abonnement

Mit e.Med Innere Medizin erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Innere Medizin, den Premium-Inhalten der internistischen Fachzeitschriften, inklusive einer gedruckten internistischen Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Organization, W.H. (2017) World Health Organization seventieth world health assembly Organization, W.H. (2017) World Health Organization seventieth world health assembly
2.
Zurück zum Zitat Kumar A et al (2006) Duration of hypotension before initiation of effective antimicrobial therapy is the critical determinant of survival in human septic shock. Crit Care Med 34(6):1589–1596CrossRef Kumar A et al (2006) Duration of hypotension before initiation of effective antimicrobial therapy is the critical determinant of survival in human septic shock. Crit Care Med 34(6):1589–1596CrossRef
3.
Zurück zum Zitat Klaerner HG et al (2000) Failure of an automated blood culture system to detect nonfermentative gram-negative bacteria. J Clin Microbiol 38(3):1036–1041CrossRef Klaerner HG et al (2000) Failure of an automated blood culture system to detect nonfermentative gram-negative bacteria. J Clin Microbiol 38(3):1036–1041CrossRef
4.
Zurück zum Zitat Seegmuller I et al (2004) Sensitivity of the BacT/ALERT FA-medium for detection of Pseudomonas aeruginosa in pre-incubated blood cultures and its temperature-dependence. J Med Microbiol 53(Pt 9):869–874CrossRef Seegmuller I et al (2004) Sensitivity of the BacT/ALERT FA-medium for detection of Pseudomonas aeruginosa in pre-incubated blood cultures and its temperature-dependence. J Med Microbiol 53(Pt 9):869–874CrossRef
5.
Zurück zum Zitat Sautter RL et al (2006) Effects of delayed-entry conditions on the recovery and detection of microorganisms from BacT/ALERT and BACTEC blood culture bottles. J Clin Microbiol 44(4):1245–1249CrossRef Sautter RL et al (2006) Effects of delayed-entry conditions on the recovery and detection of microorganisms from BacT/ALERT and BACTEC blood culture bottles. J Clin Microbiol 44(4):1245–1249CrossRef
6.
Zurück zum Zitat Miller N, Brassinne L, Allemeersch D (2018) Implementation of the new VIRTUO blood culture system: evaluation and comparison to the 3D system using simulated blood cultures. Acta Clin Belg 73(1):16–20CrossRef Miller N, Brassinne L, Allemeersch D (2018) Implementation of the new VIRTUO blood culture system: evaluation and comparison to the 3D system using simulated blood cultures. Acta Clin Belg 73(1):16–20CrossRef
7.
Zurück zum Zitat Somily AM et al (2018) Time-to-detection of bacteria and yeast with the BACTEC FX versus BacT/Alert Virtuo blood culture systems. Ann Saudi Med 38(3):194–199CrossRef Somily AM et al (2018) Time-to-detection of bacteria and yeast with the BACTEC FX versus BacT/Alert Virtuo blood culture systems. Ann Saudi Med 38(3):194–199CrossRef
8.
Zurück zum Zitat Kim SC et al (2019) Comparison of clinical performance between BacT/Alert Virtuo and BacT/Alert 3D blood culture systems. Ann Lab Med 39(3):278–283CrossRef Kim SC et al (2019) Comparison of clinical performance between BacT/Alert Virtuo and BacT/Alert 3D blood culture systems. Ann Lab Med 39(3):278–283CrossRef
9.
Zurück zum Zitat Menchinelli G et al (2019) In vitro evaluation of BACT/ALERT(R) VIRTUO(R), BACT/ALERT 3D(R), and BACTEC FX automated blood culture systems for detection of microbial pathogens using simulated human blood samples. Front Microbiol 10:221CrossRef Menchinelli G et al (2019) In vitro evaluation of BACT/ALERT(R) VIRTUO(R), BACT/ALERT 3D(R), and BACTEC FX automated blood culture systems for detection of microbial pathogens using simulated human blood samples. Front Microbiol 10:221CrossRef
10.
Zurück zum Zitat Pai S, Enoch DA, Aliyu SH (2015) Bacteremia in children: epidemiology, clinical diagnosis and antibiotic treatment. Expert Rev Anti-Infect Ther 13(9):1073–1088CrossRef Pai S, Enoch DA, Aliyu SH (2015) Bacteremia in children: epidemiology, clinical diagnosis and antibiotic treatment. Expert Rev Anti-Infect Ther 13(9):1073–1088CrossRef
11.
Zurück zum Zitat Orsini J et al (2012) Microbiological profile of organisms causing bloodstream infection in critically ill patients. J Clin Med Res 4(6):371–377PubMedPubMedCentral Orsini J et al (2012) Microbiological profile of organisms causing bloodstream infection in critically ill patients. J Clin Med Res 4(6):371–377PubMedPubMedCentral
12.
Zurück zum Zitat Wilms MC et al (2009) Effects of preincubation temperature on the detection of fastidious organisms in delayed-entry samples in the BacT/ALERT 3D blood culture system. J Microbiol Methods 79(2):194–198CrossRef Wilms MC et al (2009) Effects of preincubation temperature on the detection of fastidious organisms in delayed-entry samples in the BacT/ALERT 3D blood culture system. J Microbiol Methods 79(2):194–198CrossRef
13.
Zurück zum Zitat Gray LD, Fedorko DP (1992) Laboratory diagnosis of bacterial meningitis. Clin Microbiol Rev 5(2):130–145CrossRef Gray LD, Fedorko DP (1992) Laboratory diagnosis of bacterial meningitis. Clin Microbiol Rev 5(2):130–145CrossRef
14.
Zurück zum Zitat Levy MM, Evans LE, Rhodes A (2018) The surviving sepsis campaign bundle: 2018 update. Crit Care Med 46(6):997–1000CrossRef Levy MM, Evans LE, Rhodes A (2018) The surviving sepsis campaign bundle: 2018 update. Crit Care Med 46(6):997–1000CrossRef
15.
Zurück zum Zitat Buckman SA, Turnbull IR, Mazuski JE (2018) Empiric antibiotics for sepsis. Surg Infect 19(2):147–154CrossRef Buckman SA, Turnbull IR, Mazuski JE (2018) Empiric antibiotics for sepsis. Surg Infect 19(2):147–154CrossRef
16.
Zurück zum Zitat Ramsamy Y, Muckart DJ, Han KS (2013) Microbiological surveillance and antimicrobial stewardship minimise the need for ultrabroad-spectrum combination therapy for treatment of nosocomial infections in a trauma intensive care unit: an audit of an evidence-based empiric antimicrobial policy. S Afr Med J 103(6):371–376 Ramsamy Y, Muckart DJ, Han KS (2013) Microbiological surveillance and antimicrobial stewardship minimise the need for ultrabroad-spectrum combination therapy for treatment of nosocomial infections in a trauma intensive care unit: an audit of an evidence-based empiric antimicrobial policy. S Afr Med J 103(6):371–376
17.
Zurück zum Zitat Garnacho-Montero J et al (2014) De-escalation of empirical therapy is associated with lower mortality in patients with severe sepsis and septic shock. Intensive Care Med 40(32) Garnacho-Montero J et al (2014) De-escalation of empirical therapy is associated with lower mortality in patients with severe sepsis and septic shock. Intensive Care Med 40(32)
18.
Zurück zum Zitat Dellinger RP et al (2013) Surviving sepsis campaign: international guidelines for management of severe sepsis and septic shock, 2012. Intensive Care Med 39(2):165–228CrossRef Dellinger RP et al (2013) Surviving sepsis campaign: international guidelines for management of severe sepsis and septic shock, 2012. Intensive Care Med 39(2):165–228CrossRef
19.
Zurück zum Zitat Saito T et al (2009) Delayed insertion of blood culture bottles into automated continuously monitoring blood culture systems increases the time from blood sample collection to the detection of microorganisms in bacteremic patients. J Infect Chemother 15(1):49–53CrossRef Saito T et al (2009) Delayed insertion of blood culture bottles into automated continuously monitoring blood culture systems increases the time from blood sample collection to the detection of microorganisms in bacteremic patients. J Infect Chemother 15(1):49–53CrossRef
20.
Zurück zum Zitat Janapatla RP et al (2010) Effect of overnight storage of blood culture bottles on bacterial detection time in the BACTEC 9240 blood culture system. J Microbiol Immunol Infect 43(2):126–132CrossRef Janapatla RP et al (2010) Effect of overnight storage of blood culture bottles on bacterial detection time in the BACTEC 9240 blood culture system. J Microbiol Immunol Infect 43(2):126–132CrossRef
21.
Zurück zum Zitat Schmitz RPH et al (2013) Quality of blood culture testing - a survey in intensive care units and microbiological laboratories across four European countries. Crit Care 17(5):R248CrossRef Schmitz RPH et al (2013) Quality of blood culture testing - a survey in intensive care units and microbiological laboratories across four European countries. Crit Care 17(5):R248CrossRef
22.
Zurück zum Zitat Ronnberg C et al (2013) Transport time for blood culture bottles: underlying factors and its consequences. Diagn Microbiol Infect Dis 76(3):286–290CrossRef Ronnberg C et al (2013) Transport time for blood culture bottles: underlying factors and its consequences. Diagn Microbiol Infect Dis 76(3):286–290CrossRef
23.
Zurück zum Zitat Totty H et al (2017) A controlled comparison of the BacT/ALERT(R) 3D and VIRTUO microbial detection systems. Eur J Clin Microbiol Infect Dis 36(10):1795–1800CrossRef Totty H et al (2017) A controlled comparison of the BacT/ALERT(R) 3D and VIRTUO microbial detection systems. Eur J Clin Microbiol Infect Dis 36(10):1795–1800CrossRef
24.
Zurück zum Zitat Jacobs MR et al (2017) Multicenter clinical evaluation of BacT/Alert Virtuo blood culture system. J Clin Microbiol 55(8):2413–2421CrossRef Jacobs MR et al (2017) Multicenter clinical evaluation of BacT/Alert Virtuo blood culture system. J Clin Microbiol 55(8):2413–2421CrossRef
25.
Zurück zum Zitat Akan OA, Yildiz E (2006) Comparison of the effect of delayed entry into 2 different blood culture systems (BACTEC 9240 and BacT/ALERT 3D) on culture positivity. Diagn Microbiol Infect Dis 54(3):193–196CrossRef Akan OA, Yildiz E (2006) Comparison of the effect of delayed entry into 2 different blood culture systems (BACTEC 9240 and BacT/ALERT 3D) on culture positivity. Diagn Microbiol Infect Dis 54(3):193–196CrossRef
26.
Zurück zum Zitat (2007) Principles and procedures for blood cultures; approved guideline. M47-A (2007) Principles and procedures for blood cultures; approved guideline. M47-A
27.
Zurück zum Zitat Miller JM et al (2018) A guide to utilization of the microbiology laboratory for diagnosis of infectious diseases: 2018 update by the Infectious Diseases Society of America and the American Society for Microbiology. Clin Infect Dis 67(6):e1–e94CrossRef Miller JM et al (2018) A guide to utilization of the microbiology laboratory for diagnosis of infectious diseases: 2018 update by the Infectious Diseases Society of America and the American Society for Microbiology. Clin Infect Dis 67(6):e1–e94CrossRef
28.
Zurück zum Zitat Baron EJ et al (2005) Cumitech 1C blood cultures IV Baron EJ et al (2005) Cumitech 1C blood cultures IV
29.
Zurück zum Zitat Weinbren MJ et al (2018) Optimization of the blood culture pathway: a template for improved sepsis management and diagnostic antimicrobial stewardship. J Hosp Infect 98(3):232–235CrossRef Weinbren MJ et al (2018) Optimization of the blood culture pathway: a template for improved sepsis management and diagnostic antimicrobial stewardship. J Hosp Infect 98(3):232–235CrossRef
30.
Zurück zum Zitat Kerremans JJ et al (2009) Needle-to-incubator transport time: logistic factors influencing transport time for blood culture specimens. J Clin Microbiol 47(3):819–822CrossRef Kerremans JJ et al (2009) Needle-to-incubator transport time: logistic factors influencing transport time for blood culture specimens. J Clin Microbiol 47(3):819–822CrossRef
31.
Zurück zum Zitat Lovern D et al (2016) Antimicrobial binding and growth kinetics in BacT/ALERT(R) FA Plus and BACTEC(R) aerobic/F Plus blood culture media. Eur J Clin Microbiol Infect Dis 35(12):2033–2036CrossRef Lovern D et al (2016) Antimicrobial binding and growth kinetics in BacT/ALERT(R) FA Plus and BACTEC(R) aerobic/F Plus blood culture media. Eur J Clin Microbiol Infect Dis 35(12):2033–2036CrossRef
32.
Zurück zum Zitat Zadroga R et al (2013) Comparison of 2 blood culture media shows significant differences in bacterial recovery for patients on antimicrobial therapy. Clin Infect Dis 56(6):790–797CrossRef Zadroga R et al (2013) Comparison of 2 blood culture media shows significant differences in bacterial recovery for patients on antimicrobial therapy. Clin Infect Dis 56(6):790–797CrossRef
Metadaten
Titel
Effect of delayed entry on performance of the BACT/ALERT FAN PLUS bottles in the BACT/ALERT VIRTUO blood culture system
verfasst von
Mary Adamik
Anne Hutchins
Jasmin Mangilit
Betsy Katzin
Heather Totty
Parampal Deol
Publikationsdatum
09.10.2020
Verlag
Springer Berlin Heidelberg
Erschienen in
European Journal of Clinical Microbiology & Infectious Diseases / Ausgabe 4/2021
Print ISSN: 0934-9723
Elektronische ISSN: 1435-4373
DOI
https://doi.org/10.1007/s10096-020-04042-z

Weitere Artikel der Ausgabe 4/2021

European Journal of Clinical Microbiology & Infectious Diseases 4/2021 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.