Effect of metal artifact removal modes on the accuracy of linear measurement around titanium implants by applying different voltages: an original article

verfasst von: Daryoush Goodarzi Pour, Elham Emami Meybodi, Kosar Sodagar, Zahra Alsadat Emami Meybodi, Sanaz Safari

Erschienen in: BMC Medical Imaging | Ausgabe 1/2023

Abstract

Background

This study aims to evaluate the effects of the artifact removal algorithm on linear measurements of the buccal cortical plate by altering the voltage.

Methods

Ten titanium fixtures were inserted at the site of central, lateral, canine, premolars and molars of dry human mandibles. Vertical height of buccal plate was measured using a digital caliper as a gold standard. Mandibles were scanned with 54 and 58 kVp. Other parameters were constant. Images were reconstructed with none, low, medium and high artifact removal modes. Two Oromaxillofacial radiologists evaluated and measured the buccal plate height using Romexis software. Statistical package for the social sciences (SPSS) version 24 was used for data analysis.

Results

In medium and high modes, the difference between 54 and 58 kVp was significant (p < 0.001). No significance was noted by using low ARM (artifact removal mode) at the 54 kVp and 58 kVp.

Conclusion

Using artifact removal in low voltage decreases the accuracy of linear measurement and buccal crest visibility. By using high voltage, artifact removal would have no significant effect on accuracy of linear measurements.
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
KVP
Kilovoltage peak
CBCT
Cone-beam computed tomography
CT
Computed tomography
3D
Three dimension
ARM
Artifact removal modes
MAR
Metal artifact reduction

Background

Cone-beam computed tomography (CBCT) imaging is recommended for implant placement [13]. This modality enables three dimensional (3D) assessment of anatomical structures with an exposure dose less than that of other 3D imaging modalities such as computed tomography (CT) [2, 4, 5]. Nowadays, CBCT imaging modality is utilized more than before to assess peri-implant hard tissue, especially in the aesthetic zone, with the risk of resorption of a thin buccal plate [2, 6, 7]. However, quality reduction due to metal artifacts around high-density objects significantly decreases the accuracy of this technique [810]. This type of artifact is referred to as the beam hardening artifact and appears as dark bands and streaks and cupping artifact around metal objects [1012]. In fact, these objects absorb a high portion of X-ray photons due to high density and create artifacts [5, 10, 11]. On the other hand, the accuracy of linear measurements and visibility of anatomical structures such as the alveolar crest are highly important since they can significantly affect the treatment plan [1315].
There are different ways to minimize artifacts such as using anti-scatter guard, small field of view (FOV) and changing the exposure setting [5, 10, 1618]. Moreover, the manufacturers have introduced algorithms such as artifact removal modes (ARMs) to decrease the metal artifacts [2, 8, 10, 11, 19].
Although some studies have evaluated the efficacy of such algorithms, no previous study has assessed the effect of metal artifact removal (MAR) algorithms on the accuracy of linear measurements made on CBCT scans taken at different voltage values, which is the aim of this study [2, 8, 10, 16, 20, 21]. There is a possibility that these algorithms eliminate beneficial data [8].

Methods

In this experimental study, 10 titanium fixtures 4 × 8 mm (Super Line; Dentium, Implantium, Seoul, Korea) were inserted in two dry human mandibles obtained from the Anatomy Department of Tehran University of Medical Sciences which are used for scientific research and no patient have been involved in this study. The approval ID of Ethics committee is IR.TUMS.DENTISTRY.REC.1396.2970. An expert oral and maxillofacial surgeon placed dental implants at the alveolar crest level in the central and lateral incisors, canine, premolars, and molars (Fig. 1).
Both mandibles were fixed on an acrylic platform and then scanned by ProMax CBCT system (Planmeca, Helsinki, Finland) with a FOV of 8 × 8 cm. The exposure parameters were the same as 5 mA and 12.42 s at 54 and 58 kVp voltage values. Arc rotation was 270° and voxel size was 0.32 mm. There were four modes of artifact removal algorithms employed in the image reconstruction (Fig. 2): no artifact removal, medium, low and high artifact removal.
Digital caliper was used to measure the vertical height of buccal plate of each fixture as a gold standard. Cross sectional images that passes through the midline of each fixture were selected (Fig. 3). Two Oral and Maxillofacial radiologists measured the vertical buccal height blindly using Romexis 2.9.2 software under the standard lighting conditions on a 21 inches monitor (LG, Seoul, Korea). Observers could use any of the filters available in Romexis software.
Measurements were repeated at two-weeks intervals. They were requested to r SPSS version 24 (IBM, New York, USA) was used for data analysis. The difference between the measurements on software and the gold standard was calculated as the error rate. eport if the buccal crest couldn’t be observed.
Inter observer agreement was 0.83 and intra observer agreement was 0.86. Considering the high agreement, the mean error measured by two observers was considered as the overall error rate. The acceptable error rate was considered to be 0.5 mm. The generalized estimating equation model was used to analyze the accuracy of measurements. Sidak adjustment was applied for pairwise comparison of different modes. Also, the effect of voltage on each mode was evaluated. P < 0.05 was considered statistically significant.

Results

In both voltages, application of ARM decreased the measurement accuracy and also it could remove the buccal crest.
Table 1 presents measurement accuracy and buccal crest visibility. If the buccal crest becomes invisible for observers due to voxel elimination by using MAR, it is considered as “missing” and if the buccal crest remains visible, it is considered as valid. For accuracy, if the difference between measurement and gold standard is equal or less than 0.5 mm it is precise and if it is more than 0.5 mm it is imprecise.
Table 1
Visibility of buccal crest and buccal crest height measurement error compared to the gold standard
Imprecise,
 > 0.5 mm
Precise,
 ≤ 0.5 mm
Valid
Missing
Artifact Removal
Kvp
40%(4)
60%(6)
100%(10)
0%(0)
None
54
50%(5)
40%(4)
90%(9)
10%(1)
Low
60%(6)
10%(1)
70%(7)
30%(3)
Medium
50%(5)
20%(2)
70%(7)
30%(3)
High
50%(5)
50%(5)
100%(10)
0%(0)
None
58
60%(6)
40%(4)
100%(10)
0%(0)
Low
80%(8)
20%(2)
100%(10)
0%(0)
Medium
30%(3)
50%(5)
80%(8)
20%(2)
High
At 54 kVp, no significant difference was found without using the ARM and low ARM (P = 0.568) or between high ARM and medium ARM (P = 1.0). However, significant differences were noted between medium/high ARMs and no/low ARM (P < 0.05, Table 2).
Table 2
Pairwise comparison of accuracy of different artifact removal modes at 54 kVp
Artifact Removal
Mean difference
P-Value*
95% Confidence interval for difference
Lower
Upper
None
 Low
0.57
0.57
-0.42
1.57
 Medium
2.10
0.001
0.66
3.54
 High
2.10
<0.001
0.75
3.44
Low
 None
-0.57
0.57
-1.57
0.42
 Medium
1.53
0.01
0.20
2.86
 High
1.53
0.003
0.37
2.68
Medium
 None
-2.10
0.001
-3.54
-0.66
 Low
-1.53
0.01
-2.86
-0.20
 High
-0.01
1.00
-0.34
0.32
High
 None
-2.10
 < 0.001
-3.44
-0.75
 Low
-1.53
0.003
-2.68
-0.37
 Medium
0.01
1.00
-0.32
0.34
* Sidak adjustment
At 58 kVp, no significant difference was found in accuracy of linear measurements of buccal bone height between using and non-use of ARM (P > 0.05, Table 3).
Table 3
Pairwise comparison of accuracy of different artifact removal modes at 58 kVp
Artifact Removal
Mean difference
P-Value*
95% Confidence interval for difference
Lower
Upper
None
 Low
-0.05
1.00
-0.71
0.61
 Medium
0.19
0.95
-0.39
0.76
 High
-0.01
1.00
-0.57
0.56
Low
 None
0.05
1.00
-0.61
0.71
 Medium
0.24
0.76
-0.26
0.73
 High
0.04
1.00
-0.62
0.71
Medium
 None
-0.19
0.95
-0.76
0.39
 Low
-0.24
0.76
0.73
0.26
 High
-0.19
0.97
-0.83
0.45
High
 None
0.01
1.00
-0.56
0.57
 Low
-0.04
1.00
-0.71
0.62
 Medium
0.19
0.97
-0.45
0.83
* Sidak adjustment
By changing the voltage in low ARM, the accuracy of linear measurements did not change significantly (P = 0.351) but the difference was significant between two voltage values when medium and high ARMs were applied (P < 0.001), and the accuracy was higher at 58 kVp.

Discussion

This study revealed that using MAR algorithm reduced the accuracy of linear measurement. Low ARM caused lower accuracy compared to non-use of ARM. However, this difference was not statistically significant. Medium and high ARMs decreased the accuracy of linear measurements compared to low and no ARM and the difference was statistically significant.
The structure of this algorithm is that it eliminates the voxels with gray values higher than the threshold limit (which is 8000 for low, 4000 for medium and 3500 for high ARM) [11]. Therefore, some important details may be eliminated [11]. According to Bechara and Bezarra, evaluated the diagnostic accuracy of detection of root fracture, although MAR algorithm decreased the overall image artifact, it significantly caused lower diagnostic accuracy [20, 21]. As same as present study, Parsa et al. applied titanium implants which are more popular than ceramic or zirconium ones, reported that the algorithm could not significantly correct the voxel gray value of artifacts around fixtures [2]. No significant difference was found with (low, medium, high) and without MAR in two studies by Kamburoglu et al. for detection of buccal marginal defects, periodontal defects and furcation perforations [8, 11]. According to similar studies, MAR could not significantly improve the diagnostic accuracy for detection of dehiscence and fenestration around titanium implants [10, 16]. Application of MAR is probably not suitable for fine anatomical structures that require high spatial resolution. Another study found that MAR decreased beam hardening artifact and increased image quality [22]. But higher quality did not necessary mean higher accuracy. Queiroz et al. used cylindrical utility wax phantom, reported the positive effect when artifact generating object located at the center of FOV; However, at the periphery of FOV, MAR decreased the image quality [23]. Nikbin et al. reported a similar result [19].
This study also assessed the effect of voltage on MAR algorithm and showed that increasing the voltage significantly increased the accuracy of measurements in medium and high modes. But the difference was not significant in low mode. Similar studies showed that using higher voltage decreased metal artifact [2426]. Panjnoush et al. reported that increasing the voltage significantly decreased the metal artifact at the buccal surface of titanium rods [5]. Chindasombatjareon reported the same in both CT and CBCT [27]. Another studies found an inverse correlation between the severity of metal artifact and applied voltage [28, 29]. The majority of photons have medium level of energy and a small number of photons have maximum energy [5, 12, 30]. Thus, when hitting a high-density object such as metal, most of the photons are absorbed and a small number can pass through the object and create signals; this leads to beam hardening artifact [5, 12, 30]. By increasing the voltage, the mean energy of photons increases and consequently metal artifact would be minimized [5]. When the MAR algorithm is applied, depending on the selected mode and its threshold, voxels with higher gray value are eliminated [10, 16]. Moreover, in higher voltages, according to Bechara, the mean gray level and the gray level variation decrease while the contrast to noise ratio increases [22]. In the present in vitro study, using the voltage of 58 kVp yielded images with optimal quality and decreased the amount of artifact. With constant spatial resolution, FOV and arc rotation, increasing the voltage decreased the gray value variation and consequently by applying the high and medium ARMs, smaller details of image were lost. As a result buccal remained visible and the accuracy of linear measurement increased. However, since the low mode had a higher threshold, details was not significantly different in both voltages.
In the present study, we tried to simulate clinical condition. Hence the fixtures were located in places similar to those in the mouth. According to Misch, the canine and first molar sites are key locations for implant placement [31]. Depending on the quality of bone and location of mental foramen, first and second premolar areas may be suitable for placement of middle implants [31]; Moreover, anterior mandibular buccal crest is very thin and susceptible for resorption, which is a major complication of implant placement [31]. Thus, the central and lateral incisor, canine, premolar and first molar areas were chosen for placement of implants in our study. Moreover, other factors affecting beam hardening artifacts such as arc rotation, the reconstruction algorithm and X-ray configuration were constant in our study. Furthermore, we used dry mandible to better simulate the clinical condition comparing with homogenous phantom. Due to the fact that beam scattering may be variable when homogenous phantom was applied. However, the limitation of this study was that the phantom was positioned at the center of FOV but fixtures were located at the periphery. Another limitation was that there was no soft tissue simulation. Definitely, if soft tissue simulation was done, we could better generalize the results to the clinical situation, due to the fact that the artifacts in a cadaver or a patient may be different compared to dry mandible. But we decided to reduce the interfering factors to evaluate the pure effect of the artifact removal. The next step of the research is to evaluate the effect of other factors, including soft tissue.

Conclusion

Although the manufacturers introduced the MAR algorithm aiming to increase the diagnostic quality of images, applying the ARM would decrease the accuracy.
As clinical condition, if high voltage is applied, application of ARM would have no significant effect on the accuracy of linear measurements.

Acknowledgements

Not applicable.

Declarations

Ethics committee of Tehran University of Medical Sciences approved the study.
Not applicable.

Competing interests

The authors declare no competing interests.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Tyndall DA, Price JB, Tetradis S, Ganz SD, Hildebolt C, Scarfe WC. American Academy of Oral and Maxillofacial Radiology. Oral Surg Oral Med Oral Pathol Oral Radiol. 2012;113(6):817–26.CrossRefPubMed Tyndall DA, Price JB, Tetradis S, Ganz SD, Hildebolt C, Scarfe WC. American Academy of Oral and Maxillofacial Radiology. Oral Surg Oral Med Oral Pathol Oral Radiol. 2012;113(6):817–26.CrossRefPubMed
2.
Zurück zum Zitat Parsa A, Ibrahim N, Hassan B, Syriopoulos K, van der Stelt P. Assessment of metal artefact reduction around dental titanium implants in cone beam CT. Dentomaxillofac Radiol. 2014;43(7):20140019.CrossRefPubMedPubMedCentral Parsa A, Ibrahim N, Hassan B, Syriopoulos K, van der Stelt P. Assessment of metal artefact reduction around dental titanium implants in cone beam CT. Dentomaxillofac Radiol. 2014;43(7):20140019.CrossRefPubMedPubMedCentral
3.
Zurück zum Zitat Ludlow JB, Laster WS, See M, Bailey LTJ, Hershey HG. Accuracy of measurements of mandibular anatomy in cone beam computed tomography images. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2007;103(4):534–42.CrossRefPubMed Ludlow JB, Laster WS, See M, Bailey LTJ, Hershey HG. Accuracy of measurements of mandibular anatomy in cone beam computed tomography images. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2007;103(4):534–42.CrossRefPubMed
4.
Zurück zum Zitat Bornstein MM, Horner K, Jacobs R. Use of cone beam computed tomography in implant dentistry: current concepts, indications and limitations for clinical practice and research. Periodontol 2000. 2017;73(1):51–72.CrossRefPubMed Bornstein MM, Horner K, Jacobs R. Use of cone beam computed tomography in implant dentistry: current concepts, indications and limitations for clinical practice and research. Periodontol 2000. 2017;73(1):51–72.CrossRefPubMed
5.
Zurück zum Zitat Panjnoush M, Kheirandish Y, Kashani PM, Fakhar HB, Younesi F, Mallahi M. Effect of exposure parameters on metal artifacts in cone beam computed tomography. J Dent (Tehran). 2016;13(3):143.PubMed Panjnoush M, Kheirandish Y, Kashani PM, Fakhar HB, Younesi F, Mallahi M. Effect of exposure parameters on metal artifacts in cone beam computed tomography. J Dent (Tehran). 2016;13(3):143.PubMed
6.
Zurück zum Zitat Benic GI, Mokti M, Chen CJ, Weber HP, Hämmerle CH, Gallucci GO. Dimensions of buccal bone and mucosa at immediately placed implants after 7 years: a clinical and cone beam computed tomography study. Clin Oral Implants Res. 2012;23(5):560–6.CrossRefPubMed Benic GI, Mokti M, Chen CJ, Weber HP, Hämmerle CH, Gallucci GO. Dimensions of buccal bone and mucosa at immediately placed implants after 7 years: a clinical and cone beam computed tomography study. Clin Oral Implants Res. 2012;23(5):560–6.CrossRefPubMed
7.
Zurück zum Zitat Schropp L, Wenzel A, Spin-Neto R, Stavropoulos A. Fate of the buccal bone at implants placed early, delayed, or late after tooth extraction analyzed by cone beam CT: 10-year results from a randomized, controlled, clinical study. Clin Oral Implants Res. 2015;26(5):492–500.CrossRefPubMed Schropp L, Wenzel A, Spin-Neto R, Stavropoulos A. Fate of the buccal bone at implants placed early, delayed, or late after tooth extraction analyzed by cone beam CT: 10-year results from a randomized, controlled, clinical study. Clin Oral Implants Res. 2015;26(5):492–500.CrossRefPubMed
8.
Zurück zum Zitat Kamburoğlu K, Yılmaz F, Yeta EN, Özen D. Assessment of furcal perforations in the vicinity of different root canal sealers using a cone beam computed tomography system with and without the application of artifact reduction mode: an ex vivo investigation on extracted human teeth. Oral Surg Oral Med Oral Pathol Oral Radiol. 2016;121(6):657–65.CrossRefPubMed Kamburoğlu K, Yılmaz F, Yeta EN, Özen D. Assessment of furcal perforations in the vicinity of different root canal sealers using a cone beam computed tomography system with and without the application of artifact reduction mode: an ex vivo investigation on extracted human teeth. Oral Surg Oral Med Oral Pathol Oral Radiol. 2016;121(6):657–65.CrossRefPubMed
9.
Zurück zum Zitat Nagarajappa AK, Dwivedi N, Tiwari R. Artifacts: the downturn of CBCT image. J Int Soc Prev Community Den. 2015;5(6):440.CrossRef Nagarajappa AK, Dwivedi N, Tiwari R. Artifacts: the downturn of CBCT image. J Int Soc Prev Community Den. 2015;5(6):440.CrossRef
10.
Zurück zum Zitat Sheikhi M, Behfarnia P, Mostajabi M, Nasri N. The efficacy of metal artifact reduction (MAR) algorithm in cone-beam computed tomography on the diagnostic accuracy of fenestration and dehiscence around dental implants. J Periodontol. 2020;91(2):209–14.CrossRefPubMed Sheikhi M, Behfarnia P, Mostajabi M, Nasri N. The efficacy of metal artifact reduction (MAR) algorithm in cone-beam computed tomography on the diagnostic accuracy of fenestration and dehiscence around dental implants. J Periodontol. 2020;91(2):209–14.CrossRefPubMed
11.
Zurück zum Zitat Kamburoğlu K, Kolsuz E, Murat S, Eren H, Yüksel S, Paksoy C. Assessment of buccal marginal alveolar peri-implant and periodontal defects using a cone beam CT system with and without the application of metal artefact reduction mode. Dentomaxillofac Radiol. 2013;42(8):20130176.CrossRefPubMedPubMedCentral Kamburoğlu K, Kolsuz E, Murat S, Eren H, Yüksel S, Paksoy C. Assessment of buccal marginal alveolar peri-implant and periodontal defects using a cone beam CT system with and without the application of metal artefact reduction mode. Dentomaxillofac Radiol. 2013;42(8):20130176.CrossRefPubMedPubMedCentral
12.
Zurück zum Zitat Schulze R, Heil U, Groβ D, Bruellmann D, Dranischnikow E, Schwanecke U, et al. Artefacts in CBCT: a review. Dentomaxillofac Radiol. 2011;40(5):265–73.CrossRefPubMedPubMedCentral Schulze R, Heil U, Groβ D, Bruellmann D, Dranischnikow E, Schwanecke U, et al. Artefacts in CBCT: a review. Dentomaxillofac Radiol. 2011;40(5):265–73.CrossRefPubMedPubMedCentral
13.
Zurück zum Zitat Cavalcanti M, Rocha S, Vannier M. Craniofacial measurements based on 3D-CT volume rendering: implications for clinical applications. Dentomaxillofac Radiol. 2004;33(3):170–6.CrossRefPubMed Cavalcanti M, Rocha S, Vannier M. Craniofacial measurements based on 3D-CT volume rendering: implications for clinical applications. Dentomaxillofac Radiol. 2004;33(3):170–6.CrossRefPubMed
14.
Zurück zum Zitat Jacobs R. Preoperative radiologic planning of implant surgery in compromised patients. Periodontol 2000. 2003;33(1):12–25.CrossRefPubMed Jacobs R. Preoperative radiologic planning of implant surgery in compromised patients. Periodontol 2000. 2003;33(1):12–25.CrossRefPubMed
15.
Zurück zum Zitat Marmulla R, Wortche R, Muhling J, Hassfeld S. Geometric accuracy of the NewTom 9000 cone beam CT. Dentomaxillofac Radiol. 2005;34(1):28–31.CrossRefPubMed Marmulla R, Wortche R, Muhling J, Hassfeld S. Geometric accuracy of the NewTom 9000 cone beam CT. Dentomaxillofac Radiol. 2005;34(1):28–31.CrossRefPubMed
16.
Zurück zum Zitat De-Azevedo-Vaz SL, Peyneau PD, Ramirez-Sotelo LR, de Faria VK, Campos PSF, Haiter-Neto F. Efficacy of a cone beam computed tomography metal artifact reduction algorithm for the detection of peri-implant fenestrations and dehiscences. Oral Surg Oral Med Oral Pathol Oral Radiol. 2016;121(5):550–6.CrossRefPubMed De-Azevedo-Vaz SL, Peyneau PD, Ramirez-Sotelo LR, de Faria VK, Campos PSF, Haiter-Neto F. Efficacy of a cone beam computed tomography metal artifact reduction algorithm for the detection of peri-implant fenestrations and dehiscences. Oral Surg Oral Med Oral Pathol Oral Radiol. 2016;121(5):550–6.CrossRefPubMed
17.
Zurück zum Zitat Endo M, Tsunoo T, Nakamori N, Yoshida K. Effect of scattered radiation on image noise in cone beam CT. Med phys. 2001;28(4):469–74.CrossRefPubMed Endo M, Tsunoo T, Nakamori N, Yoshida K. Effect of scattered radiation on image noise in cone beam CT. Med phys. 2001;28(4):469–74.CrossRefPubMed
18.
Zurück zum Zitat Siewerdsen JH, Jaffray DA. Cone-beam computed tomography with a flat-panel imager: magnitude and effects of x-ray scatter. Med phys. 2001;28(2):220–31.CrossRefPubMed Siewerdsen JH, Jaffray DA. Cone-beam computed tomography with a flat-panel imager: magnitude and effects of x-ray scatter. Med phys. 2001;28(2):220–31.CrossRefPubMed
19.
Zurück zum Zitat Nikbin A, Dalili Kajan Z, Taramsari M, Khosravifard N. Effect of object position in the field of view and application of a metal artifact reduction algorithm on the detection of vertical root fractures on cone-beam computed tomography scans: an in vitro study. Imaging Sci Dent. 2018;48(4):245–54.CrossRefPubMedPubMedCentral Nikbin A, Dalili Kajan Z, Taramsari M, Khosravifard N. Effect of object position in the field of view and application of a metal artifact reduction algorithm on the detection of vertical root fractures on cone-beam computed tomography scans: an in vitro study. Imaging Sci Dent. 2018;48(4):245–54.CrossRefPubMedPubMedCentral
20.
Zurück zum Zitat Bechara B, Alex McMahan C, Moore W, Noujeim M, Teixeira F, Geha H. Cone beam CT scans with and without artefact reduction in root fracture detection of endodontically treated teeth. Dentomaxillofac Radiol. 2013;42(5):20120245.CrossRefPubMedPubMedCentral Bechara B, Alex McMahan C, Moore W, Noujeim M, Teixeira F, Geha H. Cone beam CT scans with and without artefact reduction in root fracture detection of endodontically treated teeth. Dentomaxillofac Radiol. 2013;42(5):20120245.CrossRefPubMedPubMedCentral
21.
Zurück zum Zitat Bezerra I, Neves F, Vasconcelos T, Ambrosano G, Freitas D. Influence of the artefact reduction algorithm of Picasso Trio CBCT system on the diagnosis of vertical root fractures in teeth with metal posts. Dentomaxillofac Radiol. 2015;44(6):20140428.CrossRefPubMedPubMedCentral Bezerra I, Neves F, Vasconcelos T, Ambrosano G, Freitas D. Influence of the artefact reduction algorithm of Picasso Trio CBCT system on the diagnosis of vertical root fractures in teeth with metal posts. Dentomaxillofac Radiol. 2015;44(6):20140428.CrossRefPubMedPubMedCentral
22.
23.
Zurück zum Zitat Queiroz PM, Santaella GM, Da Paz TD, Freitas DQ. Evaluation of a metal artefact reduction tool on different positions of a metal object in the FOV. Dentomaxillofac Radiol. 2017;46(3):20160366.CrossRefPubMedPubMedCentral Queiroz PM, Santaella GM, Da Paz TD, Freitas DQ. Evaluation of a metal artefact reduction tool on different positions of a metal object in the FOV. Dentomaxillofac Radiol. 2017;46(3):20160366.CrossRefPubMedPubMedCentral
24.
Zurück zum Zitat Barrett JF, Keat N. Artifacts in CT: recognition and avoidance. Radiographics. 2004;24(6):1679–91.CrossRefPubMed Barrett JF, Keat N. Artifacts in CT: recognition and avoidance. Radiographics. 2004;24(6):1679–91.CrossRefPubMed
25.
Zurück zum Zitat Kataoka ML, Hochman MG, Rodriguez EK, Lin P-JP, Kubo S, Raptopolous VD. A review of factors that affect artifact from metallic hardware on multi-row detector computed tomography. Curr Probl Diagn Radiol. 2010;39(4):125–36.CrossRefPubMed Kataoka ML, Hochman MG, Rodriguez EK, Lin P-JP, Kubo S, Raptopolous VD. A review of factors that affect artifact from metallic hardware on multi-row detector computed tomography. Curr Probl Diagn Radiol. 2010;39(4):125–36.CrossRefPubMed
26.
Zurück zum Zitat Schulze RKW, Berndt D, d’Hoedt B. On cone-beam computed tomography artifacts induced by titanium implants. Clin Oral Implants Res. 2010;21(1):100–7.CrossRefPubMed Schulze RKW, Berndt D, d’Hoedt B. On cone-beam computed tomography artifacts induced by titanium implants. Clin Oral Implants Res. 2010;21(1):100–7.CrossRefPubMed
27.
Zurück zum Zitat Chindasombatjaroen J, Kakimoto N, Murakami S, Maeda Y, Furukawa S. Quantitative analysis of metallic artifacts caused by dental metals: comparison of cone-beam and multi-detector row CT scanners. Oral Radiol. 2011;27(2):114–20.CrossRef Chindasombatjaroen J, Kakimoto N, Murakami S, Maeda Y, Furukawa S. Quantitative analysis of metallic artifacts caused by dental metals: comparison of cone-beam and multi-detector row CT scanners. Oral Radiol. 2011;27(2):114–20.CrossRef
28.
Zurück zum Zitat Draenert F, Gebhart F, Berthold M, Gosau M, Wagner W. Evaluation of demineralized bone and bone transplants in vitro and in vivo with cone beam computed tomography imaging. Dentomaxillofac Radiol. 2010;39(5):264–9.CrossRefPubMedPubMedCentral Draenert F, Gebhart F, Berthold M, Gosau M, Wagner W. Evaluation of demineralized bone and bone transplants in vitro and in vivo with cone beam computed tomography imaging. Dentomaxillofac Radiol. 2010;39(5):264–9.CrossRefPubMedPubMedCentral
29.
Zurück zum Zitat Esmaeili F, Johari M, Haddadi P, Vatankhah M. Beam hardening artifacts: comparison between two cone beam computed tomography scanners. J Dent Res Dent clin Dent Prospects. 2012;6(2):49.PubMedPubMedCentral Esmaeili F, Johari M, Haddadi P, Vatankhah M. Beam hardening artifacts: comparison between two cone beam computed tomography scanners. J Dent Res Dent clin Dent Prospects. 2012;6(2):49.PubMedPubMedCentral
30.
Zurück zum Zitat Pauwels R, Araki K, Siewerdsen J, Thongvigitmanee SS. Technical aspects of dental CBCT: state of the art. Dentomaxillofac Radiol. 2015;44(1):20140224.CrossRefPubMed Pauwels R, Araki K, Siewerdsen J, Thongvigitmanee SS. Technical aspects of dental CBCT: state of the art. Dentomaxillofac Radiol. 2015;44(1):20140224.CrossRefPubMed
31.
Zurück zum Zitat Misch CE. Dental implant prosthetics-e-book. Elsevier Health Sciences; 2004. Misch CE. Dental implant prosthetics-e-book. Elsevier Health Sciences; 2004.
Metadaten
Titel
Effect of metal artifact removal modes on the accuracy of linear measurement around titanium implants by applying different voltages: an original article
verfasst von
Daryoush Goodarzi Pour
Elham Emami Meybodi
Kosar Sodagar
Zahra Alsadat Emami Meybodi
Sanaz Safari
Publikationsdatum
01.12.2023
Verlag
BioMed Central
Erschienen in
BMC Medical Imaging / Ausgabe 1/2023
Elektronische ISSN: 1471-2342
DOI
https://doi.org/10.1186/s12880-023-01025-2

Neu im Fachgebiet Radiologie

Bald 5% der Krebserkrankungen durch CT verursacht

Die jährlich rund 93 Millionen CTs in den USA könnten künftig zu über 100.000 zusätzlichen Krebserkrankungen führen, geht aus einer Modellrechnung hervor. Damit würde eine von 20 Krebserkrankungen auf die ionisierende Strahlung bei CT-Untersuchungen zurückgehen.

Röntgen-Thorax oder LDCT fürs Lungenscreening nach HNSCC?

Personen, die an einem Plattenepithelkarzinom im Kopf-Hals-Bereich erkrankt sind, haben ein erhöhtes Risiko für Metastasen oder zweite Primärmalignome der Lunge. Eine Studie hat untersucht, wie die radiologische Überwachung aussehen sollte.

Statine: Was der G-BA-Beschluss für Praxen bedeutet

Nach dem G-BA-Beschluss zur erweiterten Verordnungsfähigkeit von Lipidsenkern rechnet die DEGAM mit 200 bis 300 neuen Dauerpatienten pro Praxis. Im Interview erläutert Präsidiumsmitglied Erika Baum, wie Hausärztinnen und Hausärzte am besten vorgehen.

Brustdichte nicht mit Multivitaminpräparat-Einnahme assoziiert

Der regelmäßige Gebrauch von Nahrungsergänzungsmitteln scheint nicht die mammografische Brustdichte zu erhöhen. In einer US-amerikanischen Studie jedenfalls ließ sich ein derartiger Zusammenhang nicht bestätigen.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.