Effect of optical fiber diameter and laser emission mode (cw vs pulse) on tissue damage profile using 1.94 µm Tm:fiber lasers in a porcine kidney model

Zeitschrift:
World Journal of Urology > Ausgabe 6/2020
Autoren:
Benedikt Becker, Dmitry Enikeev, Petr Glybochko, Leonid Rapoport, Mark Taratkin, Andreas J. Gross, Viktoriya Vinnichenko, Thomas R. W. Herrmann, Christopher Netsch
Wichtige Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Abstract

Purpose

To evaluate the ablation capacity using two Thulium fiber lasers (TFL) in a porcine kidney model.

Methods

All tissue samples were mounted on a motorized stage for a precise speed of cutting. A continuous wave (cw) TFL and a super pulsed (SP) TFL were used at power settings of 60 and 120 W with 200 and 600 µm laser fibers. After lactate dehydrogenase staining, histological evaluation was performed to measure the vaporization volume (VV), ablation depth (AD), thermo-mechanical damage zones (TMZ), coagulation zones (CZ) and the carbonization grade (CG).

Results

At 120 W, no significant differences were seen between 200 and 600 µm fibers utilizing the cw TFL regarding VV (24.6 vs. 28.2 mm3/s), AD (5.6 vs. 5.7 mm), TMZ (0 vs. 0 mm2) and CZ (18.1 vs. 12.3 mm2). Using the SP TFL, no significant differences between both fiber diameters with regard to VV (4 vs. 6.2 mm3/s), AD (2.7 vs. 3.4 mm), TMZ (1 vs. 2.6 mm2) and CZ (3.1 vs. 2.2 mm2) at 120 W were found, respectively. However, the VV of the cw TFL at 60 W was significantly less compared to 120 W using 200 and 600 µm fibers, respectively, whereas the SP TFL did not show significant differences between 60 and 120 W with regard to VV. SP TFL showed a consistently lower CG compared to cw TFL.

Conclusions

This experiment suggests that there is no significant difference using 200 or 600 µm laser fibers in cw or SP TFLs. However, the cw TFL produces a coagulation zone three to five times larger than the SP TFL regardless of the fiber diameter.

Bitte loggen Sie sich ein, um Zugang zu diesem Inhalt zu erhalten

★ PREMIUM-INHALT
e.Med Interdisziplinär

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag als Mediziner

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Jetzt e.Med zum Sonderpreis bestellen!

Weitere Produktempfehlungen anzeigen
Literatur
Über diesen Artikel

Weitere Artikel der Ausgabe 6/2020

World Journal of Urology 6/2020 Zur Ausgabe
  1. Sie können e.Med Gynäkologie & Urologie 14 Tage kostenlos testen (keine Print-Zeitschrift enthalten). Der Test läuft automatisch und formlos aus. Es kann nur einmal getestet werden.

Neu im Fachgebiet Urologie

Mail Icon II Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Urologie und bleiben Sie gut informiert – ganz bequem per eMail.

Bildnachweise