Erschienen in: European Radiology 6/2014

01.06.2014 | Computed Tomography

Effective dose to patient measurements in flat-detector and multislice computed tomography: a comparison of applications in neuroradiology

verfasst von: Tobias Struffert, Michael Hauer, Rosemarie Banckwitz, Christoph Köhler, Kevin Royalty, Arnd Doerfler

Erschienen in: European Radiology | Ausgabe 6/2014

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Flat-detector CT (FD-CT) is used for a variety of applications. Additionally, 3D rotational angiography (3D DSA) is used to supplement digital subtraction angiography (DSA) studies. The aim was to measure and compare the dose of (1) standard DSA and 3D DSA and (2) analogous FD-CT and multislice CT (MSCT) protocols.

Methods

Using an anthropomorphic phantom, the effective dose to patients (according to ICRP 103) was measured on an MSCT and a flat-detector angiographic system using standard protocols as recommended by the manufacturer.

Results

(1) Evaluation of DSA and 3D DSA angiography protocols: ap.-lat. Standard/low-dose series 1/0.8 mSv, enlarged oblique projection 0.3 mSv, 3D DSA 0.9 mSv (limited coverage length 0.3 mSv). (2) Comparison of FD-CT and MSCT: brain parenchyma imaging 2.9 /1.4 mSv, perfusion imaging 2.3/4.2 mSv, temporal bone 0.2 /0.2 mSv, angiography 2.9/3.3 mSv, limited to the head using collimation 0.5/0.5 mSv.

Conclusion

The effective dose for an FD-CT application depends on the application used. Using collimation for FD-CT applications, the dose may be reduced considerably. Due to the low dose of 3D DSA, we recommend using this technique to reduce the number of DSA series needed to identify working projections.

Key Points

Effective dose of FD-CT in comparison to MSCT is in comparable range.
Collimation decreases the dose of FD-CT effectively.
Effective dose of 3-D angiography is identical to 2-D DSA.
Different FD-CT programs have different dose.
Literatur
1.
Zurück zum Zitat Kalender WA, Kyriakou Y (2007) Flat-detector computed tomography (FD-CT). Eur Radiol 17:2767–2779CrossRefPubMed Kalender WA, Kyriakou Y (2007) Flat-detector computed tomography (FD-CT). Eur Radiol 17:2767–2779CrossRefPubMed
2.
Zurück zum Zitat Pfaff J, Struffert T, Gölitz P, Doerfler A (2013) Angiographic CT for intraprocedural monitoring of complex neuroendovascular procedures. AJNR Am J Neuroradiol 34:E77–E80CrossRefPubMed Pfaff J, Struffert T, Gölitz P, Doerfler A (2013) Angiographic CT for intraprocedural monitoring of complex neuroendovascular procedures. AJNR Am J Neuroradiol 34:E77–E80CrossRefPubMed
3.
Zurück zum Zitat Tanoue S, Kiyosue H, Kenai H, Nakamura T, Yamashita M, Mori H (2000) Three-dimensional reconstructed images after rotational angiography in the evaluation of intracranial aneurysms: surgical correlation. Neurosurgery 47:866–871CrossRefPubMed Tanoue S, Kiyosue H, Kenai H, Nakamura T, Yamashita M, Mori H (2000) Three-dimensional reconstructed images after rotational angiography in the evaluation of intracranial aneurysms: surgical correlation. Neurosurgery 47:866–871CrossRefPubMed
4.
Zurück zum Zitat von Gottberg P, Psychogios M, Schuetze G, Cohnen J, Zwaka P et al (2013) Feasibility of flat panel detector computed tomography for position assessment of external ventricular drainage. Neurol Neurochir Pol 47:32–42CrossRef von Gottberg P, Psychogios M, Schuetze G, Cohnen J, Zwaka P et al (2013) Feasibility of flat panel detector computed tomography for position assessment of external ventricular drainage. Neurol Neurochir Pol 47:32–42CrossRef
5.
Zurück zum Zitat Struffert T, Richter G, Engelhorn T, Doelken M, Goelitz P, Kalender WA et al (2009) Visualisation of intracerebral haemorrhage with flat-detector CT compared to multislice CT: results in 44 cases. Eur Radiol 19:619–625CrossRefPubMed Struffert T, Richter G, Engelhorn T, Doelken M, Goelitz P, Kalender WA et al (2009) Visualisation of intracerebral haemorrhage with flat-detector CT compared to multislice CT: results in 44 cases. Eur Radiol 19:619–625CrossRefPubMed
6.
Zurück zum Zitat Heran NS, Song JK, Namba K, Smith W, Niimi Y, Berenstein A (2006) The utility of DynaCT in neuroendovascular procedures. AJNR Am J Neuroradiol 27:330–332PubMed Heran NS, Song JK, Namba K, Smith W, Niimi Y, Berenstein A (2006) The utility of DynaCT in neuroendovascular procedures. AJNR Am J Neuroradiol 27:330–332PubMed
7.
Zurück zum Zitat Bartling SH, Gupta R, Torkos A, Dullin C, Eckhardt G, Lenarz T et al (2006) Flat-panel volume computed tomography for cochlear implant electrode array examination in isolated temporal bone specimens. Otol Neurotol 27:491–498CrossRefPubMed Bartling SH, Gupta R, Torkos A, Dullin C, Eckhardt G, Lenarz T et al (2006) Flat-panel volume computed tomography for cochlear implant electrode array examination in isolated temporal bone specimens. Otol Neurotol 27:491–498CrossRefPubMed
8.
Zurück zum Zitat Bozzato A, Struffert T, Hertel V, Iro H, Hornung J (2010) Analysis of the accuracy of high-resolution computed tomography techniques for the measurement of stapes prostheses. Eur Radiol 20:566–571CrossRefPubMed Bozzato A, Struffert T, Hertel V, Iro H, Hornung J (2010) Analysis of the accuracy of high-resolution computed tomography techniques for the measurement of stapes prostheses. Eur Radiol 20:566–571CrossRefPubMed
9.
Zurück zum Zitat Buhk JH, Lingor P, Knauth M (2008) Angiographic CT with intravenous administration of contrast medium is a noninvasive option for follow-up after intracranial stenting. Neuroradiology 50:349–354PubMedCentralCrossRefPubMed Buhk JH, Lingor P, Knauth M (2008) Angiographic CT with intravenous administration of contrast medium is a noninvasive option for follow-up after intracranial stenting. Neuroradiology 50:349–354PubMedCentralCrossRefPubMed
10.
Zurück zum Zitat Struffert T, Kloska S, Engelhorn T, Deuerling-Zheng Y, Ott S, Doelken M et al (2011) Optimized intravenous Flat Detector CT for non-invasive visualization of intracranial stents: first results. Eur Radiol 21:411–418CrossRefPubMed Struffert T, Kloska S, Engelhorn T, Deuerling-Zheng Y, Ott S, Doelken M et al (2011) Optimized intravenous Flat Detector CT for non-invasive visualization of intracranial stents: first results. Eur Radiol 21:411–418CrossRefPubMed
11.
Zurück zum Zitat Struffert T, Deuerling-Zheng Y, Kloska S, Engelhorn T, Boese J, Zellerhoff M et al (2011) Cerebral blood volume imaging by flat detector computed tomography in comparison to conventional multislice perfusion CT. Eur Radiol 21:882–889CrossRefPubMed Struffert T, Deuerling-Zheng Y, Kloska S, Engelhorn T, Boese J, Zellerhoff M et al (2011) Cerebral blood volume imaging by flat detector computed tomography in comparison to conventional multislice perfusion CT. Eur Radiol 21:882–889CrossRefPubMed
12.
Zurück zum Zitat Kyriakou Y, Richter G, Dörfler A, Kalender WA (2008) Neuroradiologic applications with routine C-arm flat panel detector CT: evaluation of patient dose measurements. AJNR Am J Neuroradiol 29:1930–1936CrossRefPubMed Kyriakou Y, Richter G, Dörfler A, Kalender WA (2008) Neuroradiologic applications with routine C-arm flat panel detector CT: evaluation of patient dose measurements. AJNR Am J Neuroradiol 29:1930–1936CrossRefPubMed
13.
Zurück zum Zitat Geleijns J, Salvadó Artells M, de Bruin PW, Matter R, Muramatsu Y, McNitt-Gray MF (2009) Computed tomography dose assessment for a 160 mm wide, 320 detector row, cone beam CT scanner. Phys Med Biol 54:3141–3159PubMedCentralCrossRefPubMed Geleijns J, Salvadó Artells M, de Bruin PW, Matter R, Muramatsu Y, McNitt-Gray MF (2009) Computed tomography dose assessment for a 160 mm wide, 320 detector row, cone beam CT scanner. Phys Med Biol 54:3141–3159PubMedCentralCrossRefPubMed
14.
Zurück zum Zitat International Commission on Radiological Protection (2007) 2007 Recommendations of the ICRP. ICRP Publication 103. Ann. ICRP 37(2–4). Elsevier Science Ltd International Commission on Radiological Protection (2007) 2007 Recommendations of the ICRP. ICRP Publication 103. Ann. ICRP 37(2–4). Elsevier Science Ltd
18.
Zurück zum Zitat Paul J, Krauss B, Banckwitz R, Maentele W, Bauer RW, Vogl TJ (2012) Relationships of clinical protocols and reconstruction kernels with image quality and radiation dose in a 128-slice CT scanner: study with an anthropomorphic and water phantom. Eur J Radiol 81:e699–e703CrossRefPubMed Paul J, Krauss B, Banckwitz R, Maentele W, Bauer RW, Vogl TJ (2012) Relationships of clinical protocols and reconstruction kernels with image quality and radiation dose in a 128-slice CT scanner: study with an anthropomorphic and water phantom. Eur J Radiol 81:e699–e703CrossRefPubMed
19.
Zurück zum Zitat Paul J, Banckwitz R, Krauss B, Vogl TJ, Maentele W, Bauer RW (2012) Estimation and comparison of effective dose (E) in standard chest CT by organ dose measurements and dose-length-product methods and assessment of the influence of CT tube potential (energy dependency) on effective dose in a dual-source CT. Eur J Radiol 81:e507–e512CrossRefPubMed Paul J, Banckwitz R, Krauss B, Vogl TJ, Maentele W, Bauer RW (2012) Estimation and comparison of effective dose (E) in standard chest CT by organ dose measurements and dose-length-product methods and assessment of the influence of CT tube potential (energy dependency) on effective dose in a dual-source CT. Eur J Radiol 81:e507–e512CrossRefPubMed
20.
Zurück zum Zitat Mahnken AH, Sedlmair M, Ritter C, Banckwitz R, Flohr T (2012) Efficacy of lower-body shielding in computed tomography fluoroscopy-guided interventions. Cardiovasc Intervent Radiol 35:1475–1479CrossRefPubMed Mahnken AH, Sedlmair M, Ritter C, Banckwitz R, Flohr T (2012) Efficacy of lower-body shielding in computed tomography fluoroscopy-guided interventions. Cardiovasc Intervent Radiol 35:1475–1479CrossRefPubMed
21.
Zurück zum Zitat Kim S, Sopko D, Toncheva G, Enterline D, Keijzers B, Yoshizumi TT (2012) Radiation dose from 3D rotational X-ray imaging: organ and effective dose with conversion factors. Radiat Prot Dosim 150:50–54CrossRef Kim S, Sopko D, Toncheva G, Enterline D, Keijzers B, Yoshizumi TT (2012) Radiation dose from 3D rotational X-ray imaging: organ and effective dose with conversion factors. Radiat Prot Dosim 150:50–54CrossRef
22.
Zurück zum Zitat Bai M, Liu X, Liu B (2013) Effective patient dose during neuroradiological C-arm CT procedures. Diagn Interv Radiol 19:29–32PubMed Bai M, Liu X, Liu B (2013) Effective patient dose during neuroradiological C-arm CT procedures. Diagn Interv Radiol 19:29–32PubMed
23.
Zurück zum Zitat Tsapaki V, Vano E, Muavrikou I, Nueofotistou V, Gallego JJ, Fernandez JM et al (2008) Comparison of patient dose in two-dimensional carotid arteriography and three-dimensional rotational angiography. Cardiovasc Intervent Radiol 31:477–482CrossRefPubMed Tsapaki V, Vano E, Muavrikou I, Nueofotistou V, Gallego JJ, Fernandez JM et al (2008) Comparison of patient dose in two-dimensional carotid arteriography and three-dimensional rotational angiography. Cardiovasc Intervent Radiol 31:477–482CrossRefPubMed
24.
Zurück zum Zitat Schueler BA, Kallmes DF, Cloft HJ (2005) 3D cerebral angiography: radiation dose comparison with digital subtraction angiography. AJNR Am J Neuroradiol 26:1898–1901PubMed Schueler BA, Kallmes DF, Cloft HJ (2005) 3D cerebral angiography: radiation dose comparison with digital subtraction angiography. AJNR Am J Neuroradiol 26:1898–1901PubMed
25.
Zurück zum Zitat Bai M, Liu B, Mu H, Liu X, Jiang Y (2012) The comparison of radiation dose between C-arm flat-detector CT (DynaCT) and multi-slice CT (MSCT): a phantom study. Eur J Radiol 81:3577–3580CrossRefPubMed Bai M, Liu B, Mu H, Liu X, Jiang Y (2012) The comparison of radiation dose between C-arm flat-detector CT (DynaCT) and multi-slice CT (MSCT): a phantom study. Eur J Radiol 81:3577–3580CrossRefPubMed
26.
Zurück zum Zitat Struffert T, Deuerling-Zheng Y, Kloska S, Engelhorn T, Strother CM, Kalender WA et al (2010) Flat detector CT in the evaluation of brain parenchyma, intracranial vasculature, and cerebral blood volume: a pilot study in patients with acute symptoms of cerebral ischemia. AJNR Am J Neuroradiol 31:1462–1469CrossRefPubMed Struffert T, Deuerling-Zheng Y, Kloska S, Engelhorn T, Strother CM, Kalender WA et al (2010) Flat detector CT in the evaluation of brain parenchyma, intracranial vasculature, and cerebral blood volume: a pilot study in patients with acute symptoms of cerebral ischemia. AJNR Am J Neuroradiol 31:1462–1469CrossRefPubMed
27.
Zurück zum Zitat Diekmann S, Siebert E, Juran R, Roll M, Deeg W, Bauknecht HC et al (2010) Dose exposure of patients undergoing comprehensive stroke imaging by multidetector-row CT: comparison of 320-detector row and 64-detector row CT scanners. AJNR Am J Neuroradiol 31:1003–1009CrossRefPubMed Diekmann S, Siebert E, Juran R, Roll M, Deeg W, Bauknecht HC et al (2010) Dose exposure of patients undergoing comprehensive stroke imaging by multidetector-row CT: comparison of 320-detector row and 64-detector row CT scanners. AJNR Am J Neuroradiol 31:1003–1009CrossRefPubMed
Metadaten
Titel
Effective dose to patient measurements in flat-detector and multislice computed tomography: a comparison of applications in neuroradiology
verfasst von
Tobias Struffert
Michael Hauer
Rosemarie Banckwitz
Christoph Köhler
Kevin Royalty
Arnd Doerfler
Publikationsdatum
01.06.2014
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 6/2014
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-014-3136-7

Weitere Artikel der Ausgabe 6/2014

European Radiology 6/2014 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.