Effects of extremity positioning on radiographic evaluation of femoral tunnel location with digitally reconstructed femoral lateral radiographs after anterior cruciate ligament reconstruction

Erschienen in: BMC Medical Imaging | Ausgabe 1/2015

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Abstract

Background

Radiographic imaging is a valuable tool in clinical practice for quick anatomical assessment. We aimed to radiographically assess (A) the anterior cruciate ligament (ACL) graft tunnel location after anatomic single-bundle (SB) reconstruction and (B) the effects of extremity positioning on the localization of the orifice of the tunnel in the distal femur in comparison with Blumensaat’s line (BL).

Methods

Three-dimensional computed tomography (3D CT) scan examinations of 22 knees of 22 subjects were evaluated. The 3D CT scan data was used to digitally reconstruct the true lateral radiographs. Graft tunnel location on the distal femoral shaft along the Blumensaat’s line and perpendicular to it were assessed on these radiographs. The femur was digitally rotated to simulate varus, valgus, internal rotation and external rotation in 5-degree increments from 0 to 20-degree. At each incremental rotated position of the femur, position of the ACL graft tunnel was calculated relative to BL and the difference from the true lateral x-ray was estimated.

Results

The position of the tunnel in the distal femur was 30.6 (±4.4) % along BL and 33.1 (±5.4) % perpendicular to BL. Ten and more degree of external, internal, valgus and varus rotations significantly affected the estimates of tunnel position (P < 0.05).

Conclusions

Femoral tunnel location can be reliably estimated from lateral radiographs after anatomic SB ACL reconstruction. Although, ten or more degree of rotations can introduce significant inaccuracies in tunnel location estimates, our study suggests that BL is overall reliable for assessing location of the distal femoral tunnel. Level of evidence: Level 2b (Retrospective Cohort Study).
Hinweise

Competing interests

There are no conflicts of interest. Financial support came from Medical Research Center, Hamad Medical Corporation, Doha, Qatar.

Authors’ contributions

PSM conceived of the study, and participated in its design and coordination and drafted the manuscript. PC participated in the design of the study and performed the statistical analysis. NA carried out the measurements on the images and helped in drafting the manuscript. SAH carried out the measurements on the images and helped in drafting the manuscript. All authors read and approved the final manuscript.
Abkürzungen
ACL
Anterior cruciate ligament
SB
Single-bundle
BL
Blumensaat’s line
3D CT
Three-dimensional computed tomography
3D
Three-dimensional
CT
Computerized tomography
MRI
Magnetic resonance imaging
2D
Two-dimensional
GE
General electric
RRs
Reconstructed radiographs
ICC
Intra-class correlation coefficient
%BL
The percentage of the length along BL
DP
Line perpendicular to Blumensaat’s line
%DP
The percentage of the length along the line 90-degree to BL
ANOVA
Analysis of variance
SEM
Standard error of measurement
CI
Confidence interval

Background

Although anterior cruciate ligament (ACL) reconstructions are routinely performed to treat ACL injuries, these injuries may have serious consequences [1]. Double-bundle, single-bundle (SB) augmentation and SB techniques are widely practiced to reconstruct the ACL [2]. The double-bundle and SB augmentation techniques are also referred to as anatomic double-bundle and anatomic SB techniques respectively. In anatomic SB ACL reconstruction the femoral tunnel is placed at the site of insertion of the native ACL [3]. Correct tunnel positioning is essential for an optimum clinical outcome in all these techniques [38]. Several landmarks and reference points are suggested to aid correct tunnel positioning [9, 10]. Blumensaat’s line (BL) and Bernard et al. technique [9] are commonly adopted to determine the location of the tunnels in the distal femoral shaft on plain x-rays. In a lateral plain x-ray of knee the BL is seen as a projection of the femoral intercondylar notch ceiling or dome [9]. The quadrant technique to estimate the position of femoral tunnel orifices on plain x-rays was first developed by Bernard et al. [9]. In Bernard et al. grid based technique, the ACL graft tunnel position in the distal femoral shaft is computed as a proportion or percentage of the length along the BL from the proximal or anterior femoral cortex to the distal or posterior, and the junction of BL and the distal or posterior cortex is considered as 0 % [9]. This technique is frequently used in both morphological research examinations and examinations reporting the result of ACL reconstruction [11]. For SB reconstruction, 25 % with a standard deviation of ±7 % was suggested as an ideal value [11, 12].
Three-dimensional (3D) imaging techniques, like computerized tomography (CT) and magnetic resonance imaging (MRI) are now commonly used to evaluate ACL injury and progress following ACL reconstruction [11]. Nevertheless two-dimensional (2D) imaging measurements, such as BL on plain x-rays, remain the most common technique for estimating the position of the distal femoral tunnel [10, 1318], due to its simplicity, easy access, availability, low radiation exposure and low costs. Spatial orientation of the knee joint in the X-ray setup may affect the comparative position of landmarks in a 2D image [11]. Hence, it is envisaged that 2D-radiography is less accurate than 3D-imaging techniques. However a few recent studies suggest similar levels of accuracy with 2D-radiography and 3D-imaging or intraoperative techniques in assessing tunnel/graft position [11, 1922]. There are very few studies evaluating tunnel placement by radiographs in anatomic single-bundle ACL reconstruction [19, 23]. Hence we (1) radiographically evaluated the influence of femoral rotation on the position of the tunnels in the distal femur relative to BL after anatomic SB ACL reconstruction; and (2) assessed clinical usefulness of radiographic evaluation of femoral tunnel and discuss its role as compared to that of the 3D-imaging techniques. We hypothesized that the radiographic projection of BL changes with femoral rotation, and that extremity malalignment (in comparison to a true lateral x-ray) will induce inaccuracies in estimating the graft tunnel position after SB ACL reconstruction.

Methods

All 3D CT scan examinations of knee joints after anatomic single-bundle ACL reconstruction performed in our institute from May 2005 to November 2012 with available 3D CT scan data (22 knees of 22 different subjects) were evaluated after the study was approved by our Institutional Research Ethics Committee (Ethics Committee, Medical Research Center, Hamad Medical Corporation, Doha, Qatar). We received “Waiver of Informed Consent” from the Ethics Committee, as only anonymized images were studied retrospectively. According to van Eck et al. [24], anatomic ACL reconstruction is “the functional restoration of the ACL to its native dimensions, collagen orientation and insertion sites.” Out of the 22 subjects six were females. Subjects ranged in age from 19 to 46 years with average age of 26 years. Time between lesion occurrence and ACL reconstruction ranged from 6 to 13 months with an average time of 8 months. All 22 subjects underwent 3D-CT examination of the operated knee with a 64 slice helical scanner. A three-dimensional volumetric model of femur was created by first segmenting the femur from the adjoining soft tissues and then interposing its CT volume (described as the entire bone volume from external cortical surface to the entire internal tissue) with every single voxel illustrating radiographic density of the local bone. Radiographs were digitally reconstructed by using “Orthopaedics Review” application available under “3D viewer” tools for lower limbs on AW Server two – version aws 2.0–5.5 by GE (General Electric). These reconstructed radiographs (RRs) were generated by ray-tracing through the femoral CT volume model. Spatial orientation inside the virtual radiograph setup that this computer application used was same as that used while obtaining a regular knee x-ray, maintaining a distance of 1 meter between the source and the subject [25]. It rendered an image very similar to an actual x-ray, with prominent radiographic features (such as BL) clearly visible [Fig. 1a and b].
A “true lateral” x-ray was obtained by spatially manipulating the femoral 3D CT model inside the virtual imaging program. The Bernard et al. quadrant technique [9] was employed to estimate the ACL graft tunnel position in relation to the BL. A grid was marked onto the RR using a computer application called ImageJ (ImageJ version 1.47, National Institutes of Health, United States of America). The tunnel outline was observed and its midpoint was estimated. Four lines were drawn to form a grid and four distances (x, y, m and n) were measured along these lines. The “x” is the distance along BL from proximal or anterior aspect of femoral cortex to its distal or posterior aspect. The “y” is the distance along the line which is oriented 90-degree to the BL, from BL to the posterior aspect of the femoral cortex. The “m” is the distance from the proximal or anterior aspect of the femoral cortex to the midpoint of the tunnel along “x”, and the “n” is the distance from the BL to the midpoint of the tunnel along “y”. The midpoint of the ACL graft tunnel was illustrated as a percentage of two lengths along each axis. The percentage of the length along BL was called as “%BL” and computed as m/x. The %BL was measured from the distal or posterior to the proximal or anterior femoral condylar cortex, and 0 % denoted the junction of this line with the distal or posterior femoral condylar cortex. The percentage of the length along the line 90-degree to BL was called as %DP. The %DP was computed as n/y and measured from BL to the inferior cortex of the femoral condyle [Fig. 1a and b], with 0 % at the position on BL.
Subsequently, the femoral 3D CT reconstruction was moved spatially within the setup of virtual radiograph computer program to simulate varus, valgus, internal rotation and external rotation in 5-degree additions from the base level of 0 degree to 20-degree. An RR was produced at each 5-degree augmentation [Fig. 2]. The Bernard et al. technique as explained above was applied to each RR to estimate the ACL graft tunnel position in relation to the BL [11]. All parameters were measured by two radiologists to determine the inter-observer reliability. Final statistical analysis was performed after averaging the measurements by the two radiologists.

Statistical analysis

We computed ICC to assess inter-observer variability for %BL and %DP positions of ACL graft tunnel orifices. Mean, range and standard deviation of all ACL graft tunnel orifice positions of %BL and %DP were computed. We used repeated-measures ANOVA to determine if %BL and %DP locations of the ACL graft tunnel orifices changed with each 5-degree augmentation of varus, valgus, internal and external femoral rotation. When the repeated-measures ANOVA was significant (p < 0.05), we performed post hoc tests with Wilcoxon matched-pairs signed ranks test. In this test, rotation smaller than 10-degree was compared to rotation greater than 10-degree to determine if a 10-degree change in extremity alignment significantly altered ACL graft tunnel orifice positions of %BL and %DP. Pictorial presentations of the key results were made using appropriate statistical graphs. A two-sided P value <0.05 was considered to be statistically significant. All Statistical analyses were done using statistical packages SPSS 19.0 (SPSS Chicago, IL).

Results

The inter-observer variability and standard error of measurement (SEM) along with 95 % confidence interval (CI) of the measured ACL graft tunnel orifice positions in relation to the BL are presented in Table 1. Very good inter-observer agreement was observed under this study. Intra-class correlation coefficient for x (distance from posterior to anterior cortex along BL), y (distance from BL to femoral cortex), m (distance of tunnel from posterior femoral cortex to center of tunnel along BL), n (distance of tunnel from BL to center of tunnel perpendicular to BL), %BL (% BL of tunnel) and %DP (% DP of tunnel) values were 0.971 (95 % CI: 0.962–0.980), 0.930 (95 % CI: 0.906–0.950), 0.887 (95 % CI: 0.848–0.919), 0.758 (95 % CI: 0.675–0.827), 0.829 (95 % CI: 0.770–0.878) and 0.758 (95 % CI: 0.675–0.827) respectively.
Table 1
Inter-observer variability and standard error of measurement of quadrant methods for evaluation of tunnel aperture orifice after anatomic single-bundle anterior cruciate ligament reconstruction
 
ICCa
CIb
SEMc
x - Distance from posterior to anterior cortex along BL
0.971
0.962, 0.980
3.06
y - Distance from BL to femoral cortex
0.930
0.906, 0.950
2.46
m - Distance of tunnel from posterior femoral cortex to center of tunnel along BL
0.887
0.848, 0.919
3.73
n - Distance of tunnel from BL to center of tunnel perpendicular to BL
0.758
0.675, 0.827
2.80
%BL of tunnel
0.829
0.770, 0.878
1.69
%DP of tunnel
0.758
0.675, 0.827
2.36
BL Blumensaat’s line
aIntra-class correlation coefficient
b95 % confidence interval of the intra-class correlation coefficient
cStandard error of measurement
The mean ACL graft tunnel orifice positions for all 5-degree increases for all rotation types (external, internal, valgus and varus) are presented in Table 2. In our study the ACL graft tunnel position was 30.6 (±4.4) %BL and 33.1 (±5.4) %DP. The %BL measurements of tunnel position did not show any significant change for 0 to 20 degree of external, internal and valgus rotations (P > 0.05) [Fig. 3]. Fifteen and more degree of varus rotation resulted in a significant different position of the tunnel (only for %BL measurements) along Blumensaat’s line (P = <0.0001) [Fig. 3]. The %DP measurements of tunnel position did not show any significant change for 0 to 20 degree of varus rotation (P > 0.05) [Fig. 4]. Ten and more degree of external, internal and valgus rotations resulted in a significant different position of the tunnel (only for %DP measurements) along Blumensaat’s line (P < 0.0001) [Fig. 4]. We did not find any significant variability in the position of tunnel for each 5 degree increase of rotation amongst the 22 subjects in our study related to skeletal morphological differences.
Table 2
Tunnel positions relative to Blumensaat’s line
Rotation Type
Lateral
Valgus
Varus
Degree
0
5
10
15
20
5
10
15
20
 %BL
30.6 ± 4.3
30.8 ± 4.1
31.7 ± 4.5
32.1 ± 5.1
33.2 ± 5.5
31.7 ± 4.5
33.2 ± 4.6
34.8 ± 4.7
35.8 ± 5.1
 %DP
33.1 ± 5.4
31.4 ± 5.7
29.9 ± 5.1
28.6 ± 4.2
28.9 ± 5.3
33.1 ± 6.0
32.9 ± 5.8
31.8 ± 5.7
30.6 ± 7.5
Rotation Type
Lateral
External
Internal
Degree
0
5
10
15
20
5
10
15
20
 %BL
30.6 ± 4.3
29.7 ± 4.2
29.4 ± 4.3
29.9 ± 3.4
29.8 ± 4.1
31.4 ± 4.7
30.9 ± 5.2
30.1 ± 5.2
30.7 ± 5.3
 %DP
33.1 ± 5.4
30.8 ± 6.5
28.3 ± 5.0
26.4 ± 4.2
25.5 ± 5.7
31.4 ± 5.4
28.8 ± 5.8
27.1 ± 5.3
27.2 ± 7.9
BL Blumensaat’s line, DP perpendicular to Blumensaat’s line

Discussion

The most important finding of our study is that the BL is quite reliable in evaluating graft tunnel position after SB ACL reconstruction. However ten and more degree of external, internal, valgus and varus rotations significantly affected estimates of radiographic tunnel position. Less than 10-degree of external, internal, valgus and varus rotations did not have any significant effect on estimates of tunnel position. Our study demonstrates that the locations of tunnels after SB ACL reconstructions can be reliably estimated on true lateral x-rays, thus affirming our primary objective. In case of SB reconstructions, ten and more degree of external, internal, valgus and varus rotations significantly affected estimates of tunnel position. Less than 10-degree of external, internal, valgus and varus rotations did not have any significant effect on estimates of tunnel position. This is contrary to our hypothesis. Similar impact of rotations on both the tunnel orifice position and the BL for less than 10-degree rotations is the likely explanation for it. Expressing the ACL graft tunnel position in percentage terms (%BL along the BL and %DP along the line 90-degree to the BL), may compensate these distortions.
Skeletal morphological differences amongst the subjects may also be a conducive factor [11, 26]. We however did not find any significant variability in the position of tunnel for each 5 degree increase of rotation amongst the 22 subjects in our study related to skeletal morphological differences.
The quadrant method by Bernard et al. is a 2D technique. Despite the invention of newer 3D imaging methods, this 2D technique for placement of ACL graft tunnel is extensively used as it is easily available, cost-effective and simple to perform. The BL is just a radiographic projection of an anatomical structure (intercondylar notch ceiling) and not a genuine anatomical structure [11]. This is the reason, why it could be influenced by extremity malalignment and by patient’s skeletal anatomy. Thus, when the femur is rotated in relation to the source of the X-ray, BL may not depict the cranial most aspect of the intercondylar notch ceiling.
From a clinical point of view, our study indicates that the BL is quite reliable in evaluating graft tunnel position after ACL reconstruction. However, if the lateral x-ray is taken with inappropriate extremity positioning, significant inaccuracies may occur. Therefore, we would advise implementation of methods that consider the individual subject’s anatomy and standardize extremity positioning. If a strict limb positioning algorithm is followed to restrict limb malpositioning within 5-degree of any rotation, the radiographic evaluation can be clinically as useful as 3D-imaging techniques. Intraoperative fluoroscopy may be used for misplacement prevention. 3D CT scan or MRI scanning may be utilized where available to yield higher accuracy [11, 27, 28].
van Eck et al. [11] were the first to study the effects of limb alignment on ACL graft tunnel positions estimated from plain radiographs. They studied these effects in double-bundle and nonanatomic single bundle reconstructions. They found that, after double-bundle reconstruction, valgus rotation greater than 10° significantly affected the assessment of tunnel position (P = 0.043) and after nonanatomic single-bundle reconstruction, internal rotation of more than 10° significantly affected the assessment of tunnel position (P = 0.043). We studied the effects of limb alignment on tunnel positions in anatomic single-bundle reconstructions, which to our knowledge, were not evaluated before. Lee et al. [19] evaluated femoral tunnel positioning using 3D CT and radiographs after single-bundle ACL reconstruction with modified transtibial technique and found these to be very useful. They concluded that, their modified transtibial technique is anticipated to provide more anatomical placement of the femoral tunnel than the former traditional transtibial techniques. Borbas et al. [20] used radiodense ligament markers for radiographic evaluation of ACL reconstruction. They found the application of radiodense ACL graft markers as straight-forward, non-expensive and potentially useful to identify the position of the graft and for diagnosis of graft failure on antero-posterior radiographs. On lateral radiographs, however, marker distances were highly variable and did not correlate with clinical ACL graft failure. Pietrini et al. [21] studied radiographic landmarks for tunnel positioning in double-bundle ACL reconstructions. This study defined the radiographic locations of the femoral and tibial bundle attachment sites of the native ACL and also defined a reliable and transferable protocol for identifying these sites on radiographs in relation to surrounding landmarks and digitally projected reference lines. Jenny et al. [22] compared the intraoperative navigated measurements of the location of the tibial and femoral tunnels during arthroscopic-assisted ACL reconstruction to the postoperative measurements performed on standard plain antero-posterior and lateral radiographs. They found significant correlation between intraoperative navigated and postoperative radiographic measurements only at the femur and good agreement between all navigated and radiographic measurements.
Performing the measurements on computer generated x-rays (RRs), rather than actual x-rays is one of the limitations of our study. Though the appearance and quality of the RRs is slightly different from regular x-rays, use of computer generated images provided precise definition of imitated rotation and true lateral position with a resolution of 1 degree accuracy. Moreover, our technique had very good inter-observer agreement. Comparatively smaller number of included subjects is another limitation of our study. More such studies with larger number of subjects may be needed to confirm the findings of our study.

Conclusions

Femoral tunnel location can be estimated from lateral radiographs after anatomic SB ACL reconstruction. Although, ten and more degree of external, internal, valgus and varus rotations can introduce significant inaccuracies in tunnel location estimates, our study suggests that BL is overall reliable for assessing location of the distal femoral tunnel and more so if a strict limb positioning algorithm is followed to restrict limb malpositioning within 10-degree of any rotation.

Acknowledgements

The authors gratefully acknowledge Dr. Buddhavarapu Shankar Rao for providing us support and encouragement during the completion of the study. The authors also gratefully acknowledge Jayakumary Santhosh and Sini George, from Hamad Medical Corporation, Qatar for their technical and administrative assistance respectively. Financial support came from Medical Research Center, Hamad Medical Corporation, Doha, Qatar. The financial support was limited to technical and administrative manpower budget and publication charges.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.

Competing interests

There are no conflicts of interest. Financial support came from Medical Research Center, Hamad Medical Corporation, Doha, Qatar.

Authors’ contributions

PSM conceived of the study, and participated in its design and coordination and drafted the manuscript. PC participated in the design of the study and performed the statistical analysis. NA carried out the measurements on the images and helped in drafting the manuscript. SAH carried out the measurements on the images and helped in drafting the manuscript. All authors read and approved the final manuscript.
Literatur
1.
Zurück zum Zitat Mahajan PS, Chandra P, Negi VC, Jayaram AP, Hussein SA. Smaller anterior cruciate ligament diameter is a predictor of subjects prone to ligament injuries: an ultrasound study. Biomed Res Int. 2015;2015:845689.CrossRefPubMedPubMedCentral Mahajan PS, Chandra P, Negi VC, Jayaram AP, Hussein SA. Smaller anterior cruciate ligament diameter is a predictor of subjects prone to ligament injuries: an ultrasound study. Biomed Res Int. 2015;2015:845689.CrossRefPubMedPubMedCentral
2.
Zurück zum Zitat Ma Y, Deie M, Iwaki D, Asaeda M, Fujita N, Adachi N, et al. Balance ability and proprioception after single-bundle, single-bundle augmentation, and double-bundle ACL reconstruction. ScientificWorldJournal. 2014;2014:342012.PubMedPubMedCentral Ma Y, Deie M, Iwaki D, Asaeda M, Fujita N, Adachi N, et al. Balance ability and proprioception after single-bundle, single-bundle augmentation, and double-bundle ACL reconstruction. ScientificWorldJournal. 2014;2014:342012.PubMedPubMedCentral
3.
Zurück zum Zitat Bedi A, Maak T, Musahl V, O’Loughlin P, Choi D, Citak M, et al. Effect of tunnel position and graft size in single-bundle anterior cruciate ligament reconstruction: an evaluation of time-zero knee stability. Arthroscopy. 2011;27(11):1543–51.CrossRefPubMed Bedi A, Maak T, Musahl V, O’Loughlin P, Choi D, Citak M, et al. Effect of tunnel position and graft size in single-bundle anterior cruciate ligament reconstruction: an evaluation of time-zero knee stability. Arthroscopy. 2011;27(11):1543–51.CrossRefPubMed
4.
Zurück zum Zitat Rayan F, Nanjayan SK, Quah C, Ramoutar D, Konan S, Haddad FS. Review of evolution of tunnel position in anterior cruciate ligament reconstruction. World J Orthop. 2015;6(2):252–62.CrossRefPubMedPubMedCentral Rayan F, Nanjayan SK, Quah C, Ramoutar D, Konan S, Haddad FS. Review of evolution of tunnel position in anterior cruciate ligament reconstruction. World J Orthop. 2015;6(2):252–62.CrossRefPubMedPubMedCentral
5.
Zurück zum Zitat Ahn JH, Jeong HJ, Ko CS, Ko TS, Kim JH. Three-dimensional reconstruction computed tomography evaluation of tunnel location during single-bundle anterior cruciate ligament reconstruction: a comparison of transtibial and 2-incision tibial tunnel-independent techniques. Clin Orthop Surg. 2013;5(1):26–35.CrossRefPubMedPubMedCentral Ahn JH, Jeong HJ, Ko CS, Ko TS, Kim JH. Three-dimensional reconstruction computed tomography evaluation of tunnel location during single-bundle anterior cruciate ligament reconstruction: a comparison of transtibial and 2-incision tibial tunnel-independent techniques. Clin Orthop Surg. 2013;5(1):26–35.CrossRefPubMedPubMedCentral
6.
Zurück zum Zitat Seo SS, Kim CW, Kim JG, Jin SY. Clinical results comparing transtibial technique and outside in technique in single bundle anterior cruciate ligament reconstruction. Knee Surg Relat Res. 2013;25(3):133–40.CrossRefPubMedPubMedCentral Seo SS, Kim CW, Kim JG, Jin SY. Clinical results comparing transtibial technique and outside in technique in single bundle anterior cruciate ligament reconstruction. Knee Surg Relat Res. 2013;25(3):133–40.CrossRefPubMedPubMedCentral
7.
Zurück zum Zitat Behrend H, Stutz G, Kessler MA, Rukavina A, Giesinger K, Kuster MS. Tunnel placement in anterior cruciate ligament (ACL) reconstruction: quality control in a teaching hospital. Knee Surg Sports Traumatol Arthrosc. 2006;14(11):1159–65.CrossRefPubMed Behrend H, Stutz G, Kessler MA, Rukavina A, Giesinger K, Kuster MS. Tunnel placement in anterior cruciate ligament (ACL) reconstruction: quality control in a teaching hospital. Knee Surg Sports Traumatol Arthrosc. 2006;14(11):1159–65.CrossRefPubMed
8.
Zurück zum Zitat Bird JH, Carmont MR, Dhillon M, Smith N, Brown C, Thompson P, et al. Validation of a new technique to determine midbundle femoral tunnel position in anterior cruciate ligament reconstruction using 3-dimensional computed tomography analysis. Arthroscopy. 2011;27(9):1259–67.CrossRefPubMed Bird JH, Carmont MR, Dhillon M, Smith N, Brown C, Thompson P, et al. Validation of a new technique to determine midbundle femoral tunnel position in anterior cruciate ligament reconstruction using 3-dimensional computed tomography analysis. Arthroscopy. 2011;27(9):1259–67.CrossRefPubMed
9.
Zurück zum Zitat Bernard M, Hertel P, Hornung H, Cierpinski T. Femoral insertion of the ACL. Radiographic quadrant method. Am J Knee Surg. 1997;10(1):14–21. discussion 21–2.PubMed Bernard M, Hertel P, Hornung H, Cierpinski T. Femoral insertion of the ACL. Radiographic quadrant method. Am J Knee Surg. 1997;10(1):14–21. discussion 21–2.PubMed
10.
Zurück zum Zitat Colombet P, Robinson J, Christel P, Franceschi JP, Djian P, Bellier G, et al. Morphology of anterior cruciate ligament attachments for anatomic reconstruction: a cadaveric dissection and radiographic study. Arthroscopy. 2006;22(9):984–92.CrossRefPubMed Colombet P, Robinson J, Christel P, Franceschi JP, Djian P, Bellier G, et al. Morphology of anterior cruciate ligament attachments for anatomic reconstruction: a cadaveric dissection and radiographic study. Arthroscopy. 2006;22(9):984–92.CrossRefPubMed
11.
Zurück zum Zitat van Eck CF, Wong AK, Irrgang JJ, Fu FH, Tashman S. The effects of limb alignment on anterior cruciate ligament graft tunnel positions estimated from plain radiographs. Knee Surg Sports Traumatol Arthrosc. 2012;20(5):979–85.CrossRefPubMed van Eck CF, Wong AK, Irrgang JJ, Fu FH, Tashman S. The effects of limb alignment on anterior cruciate ligament graft tunnel positions estimated from plain radiographs. Knee Surg Sports Traumatol Arthrosc. 2012;20(5):979–85.CrossRefPubMed
12.
Zurück zum Zitat Topliss C, Webb J. An audit of tunnel position in anterior cruciate ligament reconstruction. Knee. 2001;8:59–63.CrossRefPubMed Topliss C, Webb J. An audit of tunnel position in anterior cruciate ligament reconstruction. Knee. 2001;8:59–63.CrossRefPubMed
13.
Zurück zum Zitat Cole J, Brand Jr JC, Caborn DN, Johnson DL. Radiographic analysis of femoral tunnel position in anterior cruciate ligament reconstruction. Am J Knee Surg. 2000;13(4):218–22.PubMed Cole J, Brand Jr JC, Caborn DN, Johnson DL. Radiographic analysis of femoral tunnel position in anterior cruciate ligament reconstruction. Am J Knee Surg. 2000;13(4):218–22.PubMed
14.
Zurück zum Zitat Shah AA, Brien A, Lowe WR. Radiographic results of femoral tunnel drilling through the anteromedial portal in anterior cruciate ligament reconstruction. Arthroscopy. 2010;26(12):1586–92.CrossRefPubMed Shah AA, Brien A, Lowe WR. Radiographic results of femoral tunnel drilling through the anteromedial portal in anterior cruciate ligament reconstruction. Arthroscopy. 2010;26(12):1586–92.CrossRefPubMed
15.
Zurück zum Zitat Illingworth DK, Hensler D, Working ZM, Macalena JA, Tashman S, Fu FH. A simple evaluation of anterior cruciate ligament femoral tunnel position: the inclination angle and femoral tunnel angle. Am J Sports Med. 2011;39(12):2611–8.CrossRefPubMed Illingworth DK, Hensler D, Working ZM, Macalena JA, Tashman S, Fu FH. A simple evaluation of anterior cruciate ligament femoral tunnel position: the inclination angle and femoral tunnel angle. Am J Sports Med. 2011;39(12):2611–8.CrossRefPubMed
16.
Zurück zum Zitat Pascual-Garrido C, Swanson BL, Swanson KE. Transtibial versus low anteromedial portal drilling for anterior cruciate ligament reconstruction: a radiographic study of femoral tunnel position. Knee Surg Sports Traumatol Arthrosc. 2013;21(4):846–50.CrossRefPubMed Pascual-Garrido C, Swanson BL, Swanson KE. Transtibial versus low anteromedial portal drilling for anterior cruciate ligament reconstruction: a radiographic study of femoral tunnel position. Knee Surg Sports Traumatol Arthrosc. 2013;21(4):846–50.CrossRefPubMed
17.
Zurück zum Zitat Zantop T, Wellmann M, Fu FH, Petersen W. Tunnel positioning of anteromedial and posterolateral bundles in anatomic anterior cruciate ligament reconstruction: anatomic and radiographic findings. Am J Sports Med. 2008;36:65–72.CrossRefPubMed Zantop T, Wellmann M, Fu FH, Petersen W. Tunnel positioning of anteromedial and posterolateral bundles in anatomic anterior cruciate ligament reconstruction: anatomic and radiographic findings. Am J Sports Med. 2008;36:65–72.CrossRefPubMed
18.
Zurück zum Zitat Vieira RB, de Pinho Tavares LA, Pace Lasmar RC, da Cunha FA, de Melo Lisboa LA. Radiological analysis on femoral tunnel positioning between isometric and anatomical reconstructions of the anterior cruciate ligament. Rev Bras Ortop. 2014;49(2):160–6.CrossRefPubMedPubMedCentral Vieira RB, de Pinho Tavares LA, Pace Lasmar RC, da Cunha FA, de Melo Lisboa LA. Radiological analysis on femoral tunnel positioning between isometric and anatomical reconstructions of the anterior cruciate ligament. Rev Bras Ortop. 2014;49(2):160–6.CrossRefPubMedPubMedCentral
19.
Zurück zum Zitat Lee SR, Jang HW, Lee DW, Nam SW, Ha JK, Kim JG. Evaluation of femoral tunnel positioning using 3-dimensional computed tomography and radiographs after single bundle anterior cruciate ligament reconstruction with modified transtibial technique. Clin Orthop Surg. 2013;5(3):188–94.CrossRefPubMedPubMedCentral Lee SR, Jang HW, Lee DW, Nam SW, Ha JK, Kim JG. Evaluation of femoral tunnel positioning using 3-dimensional computed tomography and radiographs after single bundle anterior cruciate ligament reconstruction with modified transtibial technique. Clin Orthop Surg. 2013;5(3):188–94.CrossRefPubMedPubMedCentral
20.
Zurück zum Zitat Borbas P, Wieser K, Rahm S, Fucentese SF, Koch PP, Meyer DC. Radiodense ligament markers for radiographic evaluation of anterior cruciate ligament reconstruction. Knee. 2014;21(6):1018–22.CrossRefPubMed Borbas P, Wieser K, Rahm S, Fucentese SF, Koch PP, Meyer DC. Radiodense ligament markers for radiographic evaluation of anterior cruciate ligament reconstruction. Knee. 2014;21(6):1018–22.CrossRefPubMed
21.
Zurück zum Zitat Pietrini SD, Ziegler CG, Anderson CJ, Wijdicks CA, Westerhaus BD, Johansen S, et al. Radiographic landmarks for tunnel positioning in double-bundle ACL reconstructions. Knee Surg Sports Traumatol Arthrosc. 2011;19(5):792–800.CrossRefPubMed Pietrini SD, Ziegler CG, Anderson CJ, Wijdicks CA, Westerhaus BD, Johansen S, et al. Radiographic landmarks for tunnel positioning in double-bundle ACL reconstructions. Knee Surg Sports Traumatol Arthrosc. 2011;19(5):792–800.CrossRefPubMed
22.
Zurück zum Zitat Jenny JY, Abane L. Navigated anterior cruciate ligament reconstruction: radiographic validation of a nonimage-based system. Orthopedics. 2012;35(10 Suppl):18–21.CrossRefPubMed Jenny JY, Abane L. Navigated anterior cruciate ligament reconstruction: radiographic validation of a nonimage-based system. Orthopedics. 2012;35(10 Suppl):18–21.CrossRefPubMed
23.
Zurück zum Zitat Farrow LD, Huston KL, Morris P, Hall ET, Kaar SG. A simple radiographic sign vertical femoral tunnel placement during anterior cruciate ligament reconstruction. San Francisco: American Academy of Orthopaedic Surgeons Annual Meeting; 2012. Farrow LD, Huston KL, Morris P, Hall ET, Kaar SG. A simple radiographic sign vertical femoral tunnel placement during anterior cruciate ligament reconstruction. San Francisco: American Academy of Orthopaedic Surgeons Annual Meeting; 2012.
24.
Zurück zum Zitat van Eck CF, Lesniak BP, Schreiber VM, Fu FH. Anatomic single- and double-bundle anterior cruciate ligament reconstruction flowchart. Arthroscopy. 2010;26(2):258–68.CrossRefPubMed van Eck CF, Lesniak BP, Schreiber VM, Fu FH. Anatomic single- and double-bundle anterior cruciate ligament reconstruction flowchart. Arthroscopy. 2010;26(2):258–68.CrossRefPubMed
25.
Zurück zum Zitat Anderst W, Zauel R, Bishop J, Demps E, Tashman S. Validation of three-dimensional model-based tibio-femoral tracking during running. Med Eng Phys. 2009;31(1):10–6.CrossRefPubMed Anderst W, Zauel R, Bishop J, Demps E, Tashman S. Validation of three-dimensional model-based tibio-femoral tracking during running. Med Eng Phys. 2009;31(1):10–6.CrossRefPubMed
26.
Zurück zum Zitat Amis AA, Beynnon B, Blankevoort L, Chambat P, Christel P, Durselen L, et al. Proceedings of the ESSKA scientific workshop on reconstruction of the anterior and posterior cruciate ligaments. Knee Surg Sports Traumatol Arthrosc. 1994;2(3):124–32.CrossRefPubMed Amis AA, Beynnon B, Blankevoort L, Chambat P, Christel P, Durselen L, et al. Proceedings of the ESSKA scientific workshop on reconstruction of the anterior and posterior cruciate ligaments. Knee Surg Sports Traumatol Arthrosc. 1994;2(3):124–32.CrossRefPubMed
27.
Zurück zum Zitat Forsythe B, Kopf S, Wong AK, Martins CA, Anderst W, Tashman S, et al. The location of femoral and tibial tunnels in anatomic double-bundle anterior cruciate ligament reconstruction analyzed by three-dimensional computed tomography models. J Bone Joint Surg Am. 2010;92(6):1418–26.CrossRefPubMed Forsythe B, Kopf S, Wong AK, Martins CA, Anderst W, Tashman S, et al. The location of femoral and tibial tunnels in anatomic double-bundle anterior cruciate ligament reconstruction analyzed by three-dimensional computed tomography models. J Bone Joint Surg Am. 2010;92(6):1418–26.CrossRefPubMed
28.
Zurück zum Zitat Kopf S, Forsythe B, Wong AK, Tashman S, Anderst W, Irrgang JJ, et al. Nonanatomic tunnel position in traditional transtibial single-bundle anterior cruciate ligament reconstruction evaluated by three-dimensional computed tomography. J Bone Joint Surg Am. 2010;92(6):1427–31.CrossRefPubMedPubMedCentral Kopf S, Forsythe B, Wong AK, Tashman S, Anderst W, Irrgang JJ, et al. Nonanatomic tunnel position in traditional transtibial single-bundle anterior cruciate ligament reconstruction evaluated by three-dimensional computed tomography. J Bone Joint Surg Am. 2010;92(6):1427–31.CrossRefPubMedPubMedCentral
Metadaten
Titel
Effects of extremity positioning on radiographic evaluation of femoral tunnel location with digitally reconstructed femoral lateral radiographs after anterior cruciate ligament reconstruction
Publikationsdatum
01.12.2015
Erschienen in
BMC Medical Imaging / Ausgabe 1/2015
Elektronische ISSN: 1471-2342
DOI
https://doi.org/10.1186/s12880-015-0093-9

Weitere Artikel der Ausgabe 1/2015

BMC Medical Imaging 1/2015 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.