Erschienen in: Neuroradiology 8/2012

01.08.2012 | Interventional Neuroradiology

Effects of metal coverage rate of flow diversion device on neointimal growth at side branch ostium and stented artery: an animal experiment in rabbit abdominal aorta

verfasst von: Bo Hong, Kuizhong Wang, Qinghai Huang, Yi Xu, Xinggen Fang, Zhen Li, Jianmin Liu

Erschienen in: Neuroradiology | Ausgabe 8/2012

Einloggen, um Zugang zu erhalten

Abstract

Introduction

To access the effect of actual metal coverage rate (MCR) on neointimal growth at covered side branch ostium and stented artery after implantation of a flow diversion device.

Methods

Flow diverters (FDs) were implanted into abdominal aortas of 20 New Zealand rabbits. Four weeks and three months after FD implantation, the patency of side branches covered by the devices was assessed by angiography. The animals were sacrificed after angiography at 3 months postsurgery. The local actual MCR was measured under microscope and calculated. The extent of neointimal coverage at the ostia of branches and the neointima within the stent were examined by histology and scanning electron microscopy.

Results

No side branch occlusion was noted, either immediately after implantation or at follow-ups. At 3 months after implantation, the intimal coverage of branch ostia caused by a 30–40% MCR was not significantly different from that caused by an MCR ≤ 30% (p = 0.792), but it was significantly lower than that caused by an MCR ≥ 40% (p = 0.021). Neointimal thickness in the stented abdominal aorta was positively correlated to MCR (r = 0.523, p = 0.001). The neointima was composed predominantly of smooth muscle cells and collagen fibers.

Conclusion

The actual MCR exhibited remarkable differences once FD was implanted in vivo. Significantly more intimal coverage at the side branch ostia could be induced when MCR was ≥40%. The neointimal thickness within the stent was positively correlated to device MCR.
Literatur
1.
Zurück zum Zitat Ahlhelm F, Roth C, Kaufmann R, Schulte-Altedorneburg G, Romeike BF, Reith W (2007) Treatment of wide-necked intracranial aneurysms with a novel self-expanding two-zonal endovascular stent device. Neuroradiology 49:1023–1028PubMedCrossRef Ahlhelm F, Roth C, Kaufmann R, Schulte-Altedorneburg G, Romeike BF, Reith W (2007) Treatment of wide-necked intracranial aneurysms with a novel self-expanding two-zonal endovascular stent device. Neuroradiology 49:1023–1028PubMedCrossRef
2.
Zurück zum Zitat Augsburger L, Farhat M, Reymond P, Fonck E, Kulcsar Z, Stergiopulos N, Rufenacht DA (2009) Effect of flow diverter porosity on intraaneurysmal blood flow. Klin Neuroradiol 19:204–214PubMedCrossRef Augsburger L, Farhat M, Reymond P, Fonck E, Kulcsar Z, Stergiopulos N, Rufenacht DA (2009) Effect of flow diverter porosity on intraaneurysmal blood flow. Klin Neuroradiol 19:204–214PubMedCrossRef
3.
Zurück zum Zitat Aurboonyawat T, Blanc R, Schmidt P, Piotin M, Spelle L, Nakib A, Moret J (2011) An in vitro study of silk stent morphology. Neuroradiology 53:659–667PubMedCrossRef Aurboonyawat T, Blanc R, Schmidt P, Piotin M, Spelle L, Nakib A, Moret J (2011) An in vitro study of silk stent morphology. Neuroradiology 53:659–667PubMedCrossRef
4.
Zurück zum Zitat Aurboonyawat T, Schmidt PJ, Piotin M, Blanc R, Spelle L, Moret J (2011) A study of the first-generation pipeline embolization device morphology using intraoperative angiographic computed tomography (ACT). Neuroradiology 53:23–30PubMedCrossRef Aurboonyawat T, Schmidt PJ, Piotin M, Blanc R, Spelle L, Moret J (2011) A study of the first-generation pipeline embolization device morphology using intraoperative angiographic computed tomography (ACT). Neuroradiology 53:23–30PubMedCrossRef
5.
Zurück zum Zitat Fiorella D, Kelly ME, Albuquerque FC, Nelson PK (2009) Curative reconstruction of a giant midbasilar trunk aneurysm with the pipeline embolization device. Neurosurgery 64:212–217PubMedCrossRef Fiorella D, Kelly ME, Albuquerque FC, Nelson PK (2009) Curative reconstruction of a giant midbasilar trunk aneurysm with the pipeline embolization device. Neurosurgery 64:212–217PubMedCrossRef
6.
Zurück zum Zitat Fiorella D, Woo HH, Albuquerque FC, Nelson PK (2008) Definitive reconstruction of circumferential, fusiform intracranial aneurysms with the pipeline embolization device. Neurosurgery 62:1115–1120PubMedCrossRef Fiorella D, Woo HH, Albuquerque FC, Nelson PK (2008) Definitive reconstruction of circumferential, fusiform intracranial aneurysms with the pipeline embolization device. Neurosurgery 62:1115–1120PubMedCrossRef
7.
Zurück zum Zitat Ionita CN, Dohatcu A, Sinelnikov A, Sherman J, Keleshis C, Paciorek AM, Hoffmann KR, Bednarek DR, Rudin S (2009) Angiographic analysis of animal model aneurysms treated with novel polyurethane asymmetric vascular stent (P-AVS): feasibility study. Proc Soc Photo Opt Instrum Eng 7262:72621H72621–72621H72610 Ionita CN, Dohatcu A, Sinelnikov A, Sherman J, Keleshis C, Paciorek AM, Hoffmann KR, Bednarek DR, Rudin S (2009) Angiographic analysis of animal model aneurysms treated with novel polyurethane asymmetric vascular stent (P-AVS): feasibility study. Proc Soc Photo Opt Instrum Eng 7262:72621H72621–72621H72610
8.
Zurück zum Zitat Ionita CN, Paciorek AM, Dohatcu A, Hoffmann KR, Bednarek DR, Kolega J, Levy EI, Hopkins LN, Rudin S, Mocco JD (2009) The asymmetric vascular stent: efficacy in a rabbit aneurysm model. Stroke 40:959–965PubMedCrossRef Ionita CN, Paciorek AM, Dohatcu A, Hoffmann KR, Bednarek DR, Kolega J, Levy EI, Hopkins LN, Rudin S, Mocco JD (2009) The asymmetric vascular stent: efficacy in a rabbit aneurysm model. Stroke 40:959–965PubMedCrossRef
9.
Zurück zum Zitat Ionita CN, Paciorek AM, Hoffmann KR, Bednarek DR, Yamamoto J, Kolega J, Levy EI, Hopkins LN, Rudin S, Mocco J (2008) Asymmetric vascular stent: feasibility study of a new low-porosity patch-containing stent. Stroke 39:2105–2113PubMedCrossRef Ionita CN, Paciorek AM, Hoffmann KR, Bednarek DR, Yamamoto J, Kolega J, Levy EI, Hopkins LN, Rudin S, Mocco J (2008) Asymmetric vascular stent: feasibility study of a new low-porosity patch-containing stent. Stroke 39:2105–2113PubMedCrossRef
10.
Zurück zum Zitat Kallmes DF, Ding YH, Dai D, Kadirvel R, Lewis DA, Cloft HJ (2007) A new endoluminal, flow-disrupting device for treatment of saccular aneurysms. Stroke 38:2346–2352PubMedCrossRef Kallmes DF, Ding YH, Dai D, Kadirvel R, Lewis DA, Cloft HJ (2007) A new endoluminal, flow-disrupting device for treatment of saccular aneurysms. Stroke 38:2346–2352PubMedCrossRef
11.
Zurück zum Zitat Kallmes DF, Ding YH, Dai D, Kadirvel R, Lewis DA, Cloft HJ (2009) A second-generation, endoluminal, flow-disrupting device for treatment of saccular aneurysms. AJNR Am J Neuroradiol 30:1153–1158PubMedCrossRef Kallmes DF, Ding YH, Dai D, Kadirvel R, Lewis DA, Cloft HJ (2009) A second-generation, endoluminal, flow-disrupting device for treatment of saccular aneurysms. AJNR Am J Neuroradiol 30:1153–1158PubMedCrossRef
12.
Zurück zum Zitat Kim M, Taulbee DB, Tremmel M, Meng H (2008) Comparison of two stents in modifying cerebral aneurysm hemodynamics. Ann Biomed Eng 36:726–741PubMedCrossRef Kim M, Taulbee DB, Tremmel M, Meng H (2008) Comparison of two stents in modifying cerebral aneurysm hemodynamics. Ann Biomed Eng 36:726–741PubMedCrossRef
13.
Zurück zum Zitat Kulcsar Z, Ernemann U, Wetzel SG, Bock A, Goericke S, Panagiotopoulos V, Forsting M, Ruefenacht DA, Wanke I (2010) High-profile flow diverter (silk) implantation in the basilar artery: efficacy in the treatment of aneurysms and the role of the perforators. Stroke 41:1690–1696PubMedCrossRef Kulcsar Z, Ernemann U, Wetzel SG, Bock A, Goericke S, Panagiotopoulos V, Forsting M, Ruefenacht DA, Wanke I (2010) High-profile flow diverter (silk) implantation in the basilar artery: efficacy in the treatment of aneurysms and the role of the perforators. Stroke 41:1690–1696PubMedCrossRef
14.
Zurück zum Zitat Kulcsar Z, Wetzel SG, Augsburger L, Gruber A, Wanke I, Andre Rufenacht D, Sadasivan C, Cesar L, Seong J, Rakian A, Hao Q, Tio FO, Wakhloo AK, Lieber BB, Fujimura N, Ohta M, Abdo G, Ylmaz H, Lovblad KO, Rufenacht DA (2009) Effect of flow diversion treatment on very small ruptured aneurysms. Neurosurgery 67:789–793CrossRef Kulcsar Z, Wetzel SG, Augsburger L, Gruber A, Wanke I, Andre Rufenacht D, Sadasivan C, Cesar L, Seong J, Rakian A, Hao Q, Tio FO, Wakhloo AK, Lieber BB, Fujimura N, Ohta M, Abdo G, Ylmaz H, Lovblad KO, Rufenacht DA (2009) Effect of flow diversion treatment on very small ruptured aneurysms. Neurosurgery 67:789–793CrossRef
15.
Zurück zum Zitat Lieber BB, Livescu V, Hopkins LN, Wakhloo AK (2002) Particle image velocimetry assessment of stent design influence on intra-aneurysmal flow. Ann Biomed Eng 30:768–777PubMedCrossRef Lieber BB, Livescu V, Hopkins LN, Wakhloo AK (2002) Particle image velocimetry assessment of stent design influence on intra-aneurysmal flow. Ann Biomed Eng 30:768–777PubMedCrossRef
16.
Zurück zum Zitat Lieber BB, Sadasivan C (2010) Endoluminal scaffolds for vascular reconstruction and exclusion of aneurysms from the cerebral circulation. Stroke 41(10 Suppl):S21–25PubMedCrossRef Lieber BB, Sadasivan C (2010) Endoluminal scaffolds for vascular reconstruction and exclusion of aneurysms from the cerebral circulation. Stroke 41(10 Suppl):S21–25PubMedCrossRef
17.
Zurück zum Zitat Lieber BB, Stancampiano AP, Wakhloo AK (1997) Alteration of hemodynamics in aneurysm models by stenting: influence of stent porosity. Ann Biomed Eng 25:460–469PubMedCrossRef Lieber BB, Stancampiano AP, Wakhloo AK (1997) Alteration of hemodynamics in aneurysm models by stenting: influence of stent porosity. Ann Biomed Eng 25:460–469PubMedCrossRef
18.
Zurück zum Zitat Liou TM, Li YC (2008) Effects of stent porosity on hemodynamics in a sidewall aneurysm model. J Biomech 41:1174–1183PubMedCrossRef Liou TM, Li YC (2008) Effects of stent porosity on hemodynamics in a sidewall aneurysm model. J Biomech 41:1174–1183PubMedCrossRef
19.
Zurück zum Zitat Liou TM, Li YC, Wang TC (2008) Hemodynamics altered by placing helix stents in an aneurysm at a 45 degrees angle to the curved vessel. Phys Med Biol 53:3763–3776PubMedCrossRef Liou TM, Li YC, Wang TC (2008) Hemodynamics altered by placing helix stents in an aneurysm at a 45 degrees angle to the curved vessel. Phys Med Biol 53:3763–3776PubMedCrossRef
20.
Zurück zum Zitat Lubicz B, Collignon L, Raphaeli G, Pruvo JP, Bruneau M, De Witte O, Leclerc X (2010) Flow-diverter stent for the endovascular treatment of intracranial aneurysms: a prospective study in 29 patients with 34 aneurysms. Stroke 41:2247–2253PubMedCrossRef Lubicz B, Collignon L, Raphaeli G, Pruvo JP, Bruneau M, De Witte O, Leclerc X (2010) Flow-diverter stent for the endovascular treatment of intracranial aneurysms: a prospective study in 29 patients with 34 aneurysms. Stroke 41:2247–2253PubMedCrossRef
21.
Zurück zum Zitat Lylyk P, Miranda C, Ceratto R, Ferrario A, Scrivano E, Luna HR, Berez AL, Tran Q, Nelson PK, Fiorella D (2009) Curative endovascular reconstruction of cerebral aneurysms with the pipeline embolization device: the Buenos Aires experience. Neurosurgery 64:632–642PubMedCrossRef Lylyk P, Miranda C, Ceratto R, Ferrario A, Scrivano E, Luna HR, Berez AL, Tran Q, Nelson PK, Fiorella D (2009) Curative endovascular reconstruction of cerebral aneurysms with the pipeline embolization device: the Buenos Aires experience. Neurosurgery 64:632–642PubMedCrossRef
22.
Zurück zum Zitat Sadasivan C, Cesar L, Seong J, Rakian A, Hao Q, Tio FO, Wakhloo AK, Lieber BB (2009) An original flow diversion device for the treatment of intracranial aneurysms: evaluation in the rabbit elastase-induced model. Stroke 40:952–958PubMedCrossRef Sadasivan C, Cesar L, Seong J, Rakian A, Hao Q, Tio FO, Wakhloo AK, Lieber BB (2009) An original flow diversion device for the treatment of intracranial aneurysms: evaluation in the rabbit elastase-induced model. Stroke 40:952–958PubMedCrossRef
23.
Zurück zum Zitat Sadasivan C, Cesar L, Seong J, Wakhloo AK, Lieber BB (2009) Treatment of rabbit elastase-induced aneurysm models by flow diverters: development of quantifiable indexes of device performance using digital subtraction angiography. IEEE Trans Med Imaging 28:1117–1125PubMedCrossRef Sadasivan C, Cesar L, Seong J, Wakhloo AK, Lieber BB (2009) Treatment of rabbit elastase-induced aneurysm models by flow diverters: development of quantifiable indexes of device performance using digital subtraction angiography. IEEE Trans Med Imaging 28:1117–1125PubMedCrossRef
24.
Zurück zum Zitat Sadasivan C, Lieber BB, Cesar L, Miskolczi L, Seong J, Wakhloo AK (2006) Angiographic assessment of the performance of flow divertors to treat cerebral aneurysms. Conf Proc IEEE Eng Med Biol Soc 1:3210–3213PubMed Sadasivan C, Lieber BB, Cesar L, Miskolczi L, Seong J, Wakhloo AK (2006) Angiographic assessment of the performance of flow divertors to treat cerebral aneurysms. Conf Proc IEEE Eng Med Biol Soc 1:3210–3213PubMed
25.
Zurück zum Zitat Seong J, Wakhloo AK, Lieber BB (2007) In vitro evaluation of flow divertors in an elastase-induced saccular aneurysm model in rabbit. J Biomech Eng 129:863–872PubMedCrossRef Seong J, Wakhloo AK, Lieber BB (2007) In vitro evaluation of flow divertors in an elastase-induced saccular aneurysm model in rabbit. J Biomech Eng 129:863–872PubMedCrossRef
26.
Zurück zum Zitat Szikora I, Berentei Z, Kulcsar Z, Marosfoi M, Vajda ZS, Lee W, Berez A, Nelson PK (2010) Treatment of intracranial aneurysms by functional reconstruction of the parent artery: the Budapest experience with the pipeline embolization device. AJNR Am J Neuroradiol 31:1139–1147PubMedCrossRef Szikora I, Berentei Z, Kulcsar Z, Marosfoi M, Vajda ZS, Lee W, Berez A, Nelson PK (2010) Treatment of intracranial aneurysms by functional reconstruction of the parent artery: the Budapest experience with the pipeline embolization device. AJNR Am J Neuroradiol 31:1139–1147PubMedCrossRef
27.
Zurück zum Zitat Szikora I, Nelson PK, Berentei Z, Kulcsar Z, Marosfoi M, Berez A (2008) The potential of flow modification in the treatment of intracranial aneurysms. Interv Neuroradiol 14(Suppl 1):77–80PubMed Szikora I, Nelson PK, Berentei Z, Kulcsar Z, Marosfoi M, Berez A (2008) The potential of flow modification in the treatment of intracranial aneurysms. Interv Neuroradiol 14(Suppl 1):77–80PubMed
28.
Zurück zum Zitat van Rooij WJ, Sluzewski M (2010) Perforator infarction after placement of a pipeline flow-diverting stent for an unruptured A1 aneurysm. AJNR Am J Neuroradiol 31:E43–E44PubMedCrossRef van Rooij WJ, Sluzewski M (2010) Perforator infarction after placement of a pipeline flow-diverting stent for an unruptured A1 aneurysm. AJNR Am J Neuroradiol 31:E43–E44PubMedCrossRef
29.
Zurück zum Zitat Molyneux AJ, Kerr RS, Birks J, Ramzi N, Yarnold J, Sneade M, Rischmiller J (2009) Risk of recurrent subarachnoid haemorrhage, death, or dependence and standardised mortality ratios after clipping or coiling of an intracranial aneurysm in the international subarachnoid aneurysm trial (ISAT): long-term follow-up. Lancet Neurol 8:427–433PubMedCrossRef Molyneux AJ, Kerr RS, Birks J, Ramzi N, Yarnold J, Sneade M, Rischmiller J (2009) Risk of recurrent subarachnoid haemorrhage, death, or dependence and standardised mortality ratios after clipping or coiling of an intracranial aneurysm in the international subarachnoid aneurysm trial (ISAT): long-term follow-up. Lancet Neurol 8:427–433PubMedCrossRef
30.
Zurück zum Zitat Molyneux AJ, Kerr RS, Yu LM, Clarke M, Sneade M, Yarnold JA, Sandercock P (2005) International subarachnoid aneurysm trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised comparison of effects on survival, dependency, seizures, rebleeding, subgroups, and aneurysm occlusion. Lancet 366:809–817PubMedCrossRef Molyneux AJ, Kerr RS, Yu LM, Clarke M, Sneade M, Yarnold JA, Sandercock P (2005) International subarachnoid aneurysm trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised comparison of effects on survival, dependency, seizures, rebleeding, subgroups, and aneurysm occlusion. Lancet 366:809–817PubMedCrossRef
31.
Zurück zum Zitat Rhee K, Han MH, Cha SH (2002) Changes of flow characteristics by stenting in aneurysm models: influence of aneurysm geometry and stent porosity. Ann Biomed Eng 30:894–904PubMedCrossRef Rhee K, Han MH, Cha SH (2002) Changes of flow characteristics by stenting in aneurysm models: influence of aneurysm geometry and stent porosity. Ann Biomed Eng 30:894–904PubMedCrossRef
32.
Zurück zum Zitat Komatsu R, Ueda M, Naruko T, Kojima A, Becker AE (1998) Neointimal tissue response at sites of coronary stenting in humans: macroscopic, histological, and immunohistochemical analyses. Circulation 98:224–233PubMedCrossRef Komatsu R, Ueda M, Naruko T, Kojima A, Becker AE (1998) Neointimal tissue response at sites of coronary stenting in humans: macroscopic, histological, and immunohistochemical analyses. Circulation 98:224–233PubMedCrossRef
33.
Metadaten
Titel
Effects of metal coverage rate of flow diversion device on neointimal growth at side branch ostium and stented artery: an animal experiment in rabbit abdominal aorta
verfasst von
Bo Hong
Kuizhong Wang
Qinghai Huang
Yi Xu
Xinggen Fang
Zhen Li
Jianmin Liu
Publikationsdatum
01.08.2012
Verlag
Springer-Verlag
Erschienen in
Neuroradiology / Ausgabe 8/2012
Print ISSN: 0028-3940
Elektronische ISSN: 1432-1920
DOI
https://doi.org/10.1007/s00234-011-0984-6

Weitere Artikel der Ausgabe 8/2012

Neuroradiology 8/2012 Zur Ausgabe

Leitlinien kompakt für die Neurologie

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Update Neurologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.