28.06.2016 | Otology

Electrophysiological detection of scalar changing perimodiolar cochlear electrode arrays: a long term follow-up study

verfasst von: Philipp Mittmann, I. Todt, A. Ernst, G. Rademacher, S. Mutze, S. Göricke, M. Schlamann, R. Ramalingam, S. Lang, F. Christov, D. Arweiler-Harbeck

Erschienen in: European Archives of Oto-Rhino-Laryngology | Ausgabe 12/2016

Einloggen, um Zugang zu erhalten

Abstract

The position of the cochlear electrode array within the scala tympani is essential for an optimal hearing benefit. An intraoperative NRT-ratio was established, which can provide information about the intraoperative intracochlear electrode array position for perimodiolar electrodes. The aims of this study were to verify the longterm reliability for the NRT-ratio in perimodiolar electrodes. In a retrospective controlled study in a Tertiary Referral Center the electrophysiological data sets of 123 patients with implanted Nucleus Contour Advance electrodes were enclosed. Intraoperative and up to 1 year follow-up Auto-NRTs were evaluated. A NRT-ratio was calculated by dividing the average Auto-NRT data from electrode 16 to 18 with the average from electrode 5 to 7. Using a flat panel tomography system, the position of the electrode array was certified radiological. 31 patients with perimodiolar electrodes with 1 year follow-up data were included in the study. Eleven patients showed regular follow-up NRT-ratio with a correlated and radiologically confirmed electrode position. 20 patients showed mismatches between the NRT-ratio and the radiological position. These patients were highly variable in terms of duration of deafness and neural spectrum disorders. The NRT-ratio can be used to determine the intracochlear position of the electrode array for perimodiolar electrodes. Intraoperatively the NRT-ratio predicts the array position within the cochlea highly reliable for perimodiolar electrodes. We showed that after 6 months and a year, the NRT-ratio remains unchanged in most of the cases and shows a good correlation to the radiological determined position of the array. Nevertheless, the condition of the neural structures is highly important for reproducible responses. Limited validity is given in patients with degenerative and structural neural disorders.
Literatur
2.
Zurück zum Zitat Kiefer J, Gstoettner W, Baumgartner W, Pok SM, Tillein J, Ye Q, von Ilberg C (2004) Conservation of low-frequency hearing in cochlear implantation. Acta Otolaryngol 124(3):272–280CrossRefPubMed Kiefer J, Gstoettner W, Baumgartner W, Pok SM, Tillein J, Ye Q, von Ilberg C (2004) Conservation of low-frequency hearing in cochlear implantation. Acta Otolaryngol 124(3):272–280CrossRefPubMed
5.
9.
Zurück zum Zitat Skinner MW, Holden TA, Whiting BR, Voie AH, Brunsden B, Neely JG, Saxon EA, Hullar TE, Finley CC (2007) In vivo estimates of the position of advanced bionics electrode arrays in the human cochlea. Ann Otol Rhinol Laryngol Suppl 197:2–24CrossRefPubMed Skinner MW, Holden TA, Whiting BR, Voie AH, Brunsden B, Neely JG, Saxon EA, Hullar TE, Finley CC (2007) In vivo estimates of the position of advanced bionics electrode arrays in the human cochlea. Ann Otol Rhinol Laryngol Suppl 197:2–24CrossRefPubMed
11.
Zurück zum Zitat Aschendorff A, Kubalek R, Turowski B, Zanella F, Hochmuth A, Schumacher M, Klenzner T, Laszig R (2005) Quality control after cochlear implant surgery by means of rotational tomography. Otol Neurotol 26(1):34–37CrossRefPubMed Aschendorff A, Kubalek R, Turowski B, Zanella F, Hochmuth A, Schumacher M, Klenzner T, Laszig R (2005) Quality control after cochlear implant surgery by means of rotational tomography. Otol Neurotol 26(1):34–37CrossRefPubMed
13.
Zurück zum Zitat Seidman MD, Vivek P, Dickinson W (2005) Neural response telemetry results with the nucleus 24 contour in a perimodiolar position. Otol Neurotol 26(4):620–623CrossRefPubMed Seidman MD, Vivek P, Dickinson W (2005) Neural response telemetry results with the nucleus 24 contour in a perimodiolar position. Otol Neurotol 26(4):620–623CrossRefPubMed
15.
Zurück zum Zitat Shepherd RK, Hatsushika S, Clark GM (1993) Electrical stimulation of the auditory nerve: the effect of electrode position on neural excitation. Hear Res 66(1):108–120CrossRefPubMed Shepherd RK, Hatsushika S, Clark GM (1993) Electrical stimulation of the auditory nerve: the effect of electrode position on neural excitation. Hear Res 66(1):108–120CrossRefPubMed
16.
Zurück zum Zitat Wanna GB, Noble JH, Gifford RH, Dietrich MS, Sweeney AD, Zhang D, Dawant BM, Rivas A, Labadie RF (2015) Impact of intrascalar electrode location, electrode type, and angular insertion depth on residual hearing in cochlear implant patients: preliminary results. Otol Neurotol 36(8):1343–1348. doi:10.1097/MAO.0000000000000829 CrossRefPubMed Wanna GB, Noble JH, Gifford RH, Dietrich MS, Sweeney AD, Zhang D, Dawant BM, Rivas A, Labadie RF (2015) Impact of intrascalar electrode location, electrode type, and angular insertion depth on residual hearing in cochlear implant patients: preliminary results. Otol Neurotol 36(8):1343–1348. doi:10.​1097/​MAO.​0000000000000829​ CrossRefPubMed
17.
Zurück zum Zitat Boyer E, Karkas A, Attye A, Lefournier V, Escude B, Schmerber S (2015) Scalar localization by cone-beam computed tomography of cochlear implant carriers: a comparative study between straight and periomodiolar precurved electrode arrays. Otol Neurotol 36(3):422–429. doi:10.1097/MAO.0000000000000705 CrossRefPubMed Boyer E, Karkas A, Attye A, Lefournier V, Escude B, Schmerber S (2015) Scalar localization by cone-beam computed tomography of cochlear implant carriers: a comparative study between straight and periomodiolar precurved electrode arrays. Otol Neurotol 36(3):422–429. doi:10.​1097/​MAO.​0000000000000705​ CrossRefPubMed
18.
Zurück zum Zitat Mittmann P, Rademacher G, Mutze S, Hassepass F, Ernst A, Todt I (2015) Evaluation of the relationship between the NRT-ratio, cochlear anatomy, and insertions depth of perimodiolar cochlear implant electrodes. BioMed Res Int 2015:706253. doi:10.1155/2015/706253 CrossRefPubMedPubMedCentral Mittmann P, Rademacher G, Mutze S, Hassepass F, Ernst A, Todt I (2015) Evaluation of the relationship between the NRT-ratio, cochlear anatomy, and insertions depth of perimodiolar cochlear implant electrodes. BioMed Res Int 2015:706253. doi:10.​1155/​2015/​706253 CrossRefPubMedPubMedCentral
19.
20.
Zurück zum Zitat Mittmann P, Todt I, Wesarg T, Arndt S, Ernst A, Hassepass F (2015) Electrophysiological detection of intracochlear scalar changing perimodiolar cochlear implant electrodes: a blinded study. Otol Neurotol. doi:10.1097/MAO.0000000000000766 Mittmann P, Todt I, Wesarg T, Arndt S, Ernst A, Hassepass F (2015) Electrophysiological detection of intracochlear scalar changing perimodiolar cochlear implant electrodes: a blinded study. Otol Neurotol. doi:10.​1097/​MAO.​0000000000000766​
21.
Zurück zum Zitat Henkin Y, Kaplan-Neeman R, Kronenberg J, Migirov L, Hildesheimer M, Muchnik C (2006) A longitudinal study of electrical stimulation levels and electrode impedance in children using the Clarion cochlear implant. Acta Otolaryngol 126(6):581–586. doi:10.1080/00016480500443391 CrossRefPubMed Henkin Y, Kaplan-Neeman R, Kronenberg J, Migirov L, Hildesheimer M, Muchnik C (2006) A longitudinal study of electrical stimulation levels and electrode impedance in children using the Clarion cochlear implant. Acta Otolaryngol 126(6):581–586. doi:10.​1080/​0001648050044339​1 CrossRefPubMed
23.
Zurück zum Zitat Mittmann P, Todt I, Wesarg T, Arndt S, Ernst A, Hassepass F (2015) Electrophysiological detection of scalar-changing perimodiolar cochlear electrode arrays: a six-month follow-up study. Audiol Neurootol 20(6):400–405. doi:10.1159/000441346 CrossRefPubMed Mittmann P, Todt I, Wesarg T, Arndt S, Ernst A, Hassepass F (2015) Electrophysiological detection of scalar-changing perimodiolar cochlear electrode arrays: a six-month follow-up study. Audiol Neurootol 20(6):400–405. doi:10.​1159/​000441346 CrossRefPubMed
24.
Zurück zum Zitat Kawano A, Seldon HL, Clark GM, Ramsden RT, Raine CH (1998) Intracochlear factors contributing to psychophysical percepts following cochlear implantation. Acta Otolaryngol 118(3):313–326CrossRefPubMed Kawano A, Seldon HL, Clark GM, Ramsden RT, Raine CH (1998) Intracochlear factors contributing to psychophysical percepts following cochlear implantation. Acta Otolaryngol 118(3):313–326CrossRefPubMed
26.
Zurück zum Zitat Henkin Y, Kaplan-Neeman R, Muchnik C, Kronenberg J, Hildesheimer M (2003) Changes over time in electrical stimulation levels and electrode impedance values in children using the Nucleus 24 M cochlear implant. Int J Pediatr Otorhinolaryngol 67(8):873–880CrossRefPubMed Henkin Y, Kaplan-Neeman R, Muchnik C, Kronenberg J, Hildesheimer M (2003) Changes over time in electrical stimulation levels and electrode impedance values in children using the Nucleus 24 M cochlear implant. Int J Pediatr Otorhinolaryngol 67(8):873–880CrossRefPubMed
28.
29.
Zurück zum Zitat Nadol JB Jr, Eddington DK (2004) Histologic evaluation of the tissue seal and biologic response around cochlear implant electrodes in the human. Otol Neurotol Off Publ Am Otol Soc Am Neurotol Soc Eur Acad Otol Neurotol 25(3):257–262CrossRef Nadol JB Jr, Eddington DK (2004) Histologic evaluation of the tissue seal and biologic response around cochlear implant electrodes in the human. Otol Neurotol Off Publ Am Otol Soc Am Neurotol Soc Eur Acad Otol Neurotol 25(3):257–262CrossRef
30.
Zurück zum Zitat Brown CJ, Abbas PJ, Etlert CP, O’Brient S, Oleson JJ (2010) Effects of long-term use of a cochlear implant on the electrically evoked compound action potential. J Am Acad Audiol 21(1):5–15CrossRefPubMedPubMedCentral Brown CJ, Abbas PJ, Etlert CP, O’Brient S, Oleson JJ (2010) Effects of long-term use of a cochlear implant on the electrically evoked compound action potential. J Am Acad Audiol 21(1):5–15CrossRefPubMedPubMedCentral
31.
Zurück zum Zitat Saunders E, Cohen L, Aschendorff A, Shapiro W, Knight M, Stecker M, Richter B, Waltzman S, Tykocinski M, Roland T, Laszig R, Cowan R (2002) Threshold, comfortable level and impedance changes as a function of electrode-modiolar distance. Ear Hear 23(1 Suppl):28S–40SCrossRefPubMed Saunders E, Cohen L, Aschendorff A, Shapiro W, Knight M, Stecker M, Richter B, Waltzman S, Tykocinski M, Roland T, Laszig R, Cowan R (2002) Threshold, comfortable level and impedance changes as a function of electrode-modiolar distance. Ear Hear 23(1 Suppl):28S–40SCrossRefPubMed
32.
Zurück zum Zitat Linthicum FH Jr, Fayad JN (2009) Spiral ganglion cell loss is unrelated to segmental cochlear sensory system degeneration in humans. Otol Neurotol 30(3):418–422CrossRefPubMedPubMedCentral Linthicum FH Jr, Fayad JN (2009) Spiral ganglion cell loss is unrelated to segmental cochlear sensory system degeneration in humans. Otol Neurotol 30(3):418–422CrossRefPubMedPubMedCentral
33.
Zurück zum Zitat Shah P, Riphagen S, Beyene J, Perlman M (2004) Multiorgan dysfunction in infants with post-asphyxial hypoxic-ischaemic encephalopathy. Arch Dis Child Fetal Neonatal Ed 89(2):F152–F155CrossRefPubMedPubMedCentral Shah P, Riphagen S, Beyene J, Perlman M (2004) Multiorgan dysfunction in infants with post-asphyxial hypoxic-ischaemic encephalopathy. Arch Dis Child Fetal Neonatal Ed 89(2):F152–F155CrossRefPubMedPubMedCentral
35.
Zurück zum Zitat Kraaijenga VJ, Smit AL, Stegeman I, Smilde JJ, van Zanten GA, Grolman W (2015) Factors that influence outcomes in cochlear implantation in adults, based on patient related characteristics—a retrospective study. Clin Otolaryngol. doi:10.1111/coa.12571 Kraaijenga VJ, Smit AL, Stegeman I, Smilde JJ, van Zanten GA, Grolman W (2015) Factors that influence outcomes in cochlear implantation in adults, based on patient related characteristics—a retrospective study. Clin Otolaryngol. doi:10.​1111/​coa.​12571
36.
Zurück zum Zitat Kawashima H, Tsuji N (1987) Syndrome of microcephaly, deafness/malformed ears, mental retardation and peculiar facies in a mother and son. Clin Genet 31(5):303–307CrossRefPubMed Kawashima H, Tsuji N (1987) Syndrome of microcephaly, deafness/malformed ears, mental retardation and peculiar facies in a mother and son. Clin Genet 31(5):303–307CrossRefPubMed
37.
Zurück zum Zitat Brown CB, Ogg CS, Cameron JS, Bewick M (1974) High dose frusemide in acute reversible intrinsic renal failure. A preliminary communication. Scott Med J 19(Suppl 1):35–39PubMed Brown CB, Ogg CS, Cameron JS, Bewick M (1974) High dose frusemide in acute reversible intrinsic renal failure. A preliminary communication. Scott Med J 19(Suppl 1):35–39PubMed
40.
Zurück zum Zitat Nadol JB Jr, Thornton AR (1987) Ultrastructural findings in a case of Meniere’s disease. Ann Otol Rhinol Laryngol 96(4):449–454CrossRefPubMed Nadol JB Jr, Thornton AR (1987) Ultrastructural findings in a case of Meniere’s disease. Ann Otol Rhinol Laryngol 96(4):449–454CrossRefPubMed
41.
Zurück zum Zitat Momin SR, Melki SJ, Alagramam KN, Megerian CA (2010) Spiral ganglion loss outpaces inner hair cell loss in endolymphatic hydrops. Laryngoscope 120(1):159–165. doi:10.1002/lary.20673 PubMed Momin SR, Melki SJ, Alagramam KN, Megerian CA (2010) Spiral ganglion loss outpaces inner hair cell loss in endolymphatic hydrops. Laryngoscope 120(1):159–165. doi:10.​1002/​lary.​20673 PubMed
Metadaten
Titel
Electrophysiological detection of scalar changing perimodiolar cochlear electrode arrays: a long term follow-up study
verfasst von
Philipp Mittmann
I. Todt
A. Ernst
G. Rademacher
S. Mutze
S. Göricke
M. Schlamann
R. Ramalingam
S. Lang
F. Christov
D. Arweiler-Harbeck
Publikationsdatum
28.06.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
European Archives of Oto-Rhino-Laryngology / Ausgabe 12/2016
Print ISSN: 0937-4477
Elektronische ISSN: 1434-4726
DOI
https://doi.org/10.1007/s00405-016-4175-2

Weitere Artikel der Ausgabe 12/2016

European Archives of Oto-Rhino-Laryngology 12/2016 Zur Ausgabe

Update HNO

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert – ganz bequem per eMail.