Elevated serum levels of bromine do not always indicate pseudohyperchloremia
verfasst von:
Takayasu Ito, Eiji Ishikawa, Yoshikazu Matsuda, Akiko Tanoue, Mika Fujimoto, Hiroshi Matsuo, Tomohiro Murata, Takuya Hiramoto, Masaaki Ito, Shinsuke Nomura
Erschienen in:
Clinical and Experimental Nephrology
|
Ausgabe 5/2010
Einloggen, um Zugang zu erhalten
Abstract
Background
We encountered a case of bromism that was found to be due to pseudohyperchloremia. Hyperchloremia is known to be able to reveal existing bromism, but the fact that bromine (Br−) influences chloride (Cl−) in assays that use ion electrode machines is not widely known.
Methods
We assayed samples by an ion electrode method, using four types of machines. Different amounts of Cl− or Br− were added to each sample.
Results
With the addition of Cl− to the samples, the assayed Cl− concentrations were proportional to the amount of added Cl−. With the addition of Br− to the samples, the assayed Cl− concentrations, as measured by all machines, were increased, but the amounts of the increase differed significantly, and were not proportional to the amount of Br− added. In particular, in the machine most markedly influenced by additional Br−, the Cl− concentrations increased from 94.9 to 139.6 mEq/l with the addition of 10 mEq/l of Br−. Conversely, in the least influenced machine, Cl− values increased from 95.0 to 103.0 mEq/l with the addition of 10 mEq/l of Br−.
Conclusion
The influence on the Cl− assay of the addition of Br− varied significantly between different ion electrode machines. Clinical nephrologists therefore need to be able to recognize the characteristics of the specific machines used in their hospitals.