Open Access 01.03.2018 | Original Research

Enhanced Penetration for Axial Length Measurement of Eyes with Dense Cataracts Using Swept Source Optical Coherence Tomography: A Consecutive Observational Study

verfasst von: Nino Hirnschall, Ralph Varsits, Birgit Doeller, Oliver Findl

Erschienen in: Ophthalmology and Therapy | Ausgabe 1/2018

Abstract

Introduction

The aim of this study was to find cases in which the axial eye length could not be measured with partial coherence interferometry (PCI) technology and to assess if it could be measured using swept source optical coherence tomography (ss-OCT) technology.

Methods

All patients were measured at their pre-assessment visit 1 week prior to cataract surgery using conventional optical biometry (PCI technology, IOLMaster 500, Carl Zeiss Meditec AG, Jena, Germany). Patients in whom one or both eyes could not be measured using PCI technology were invited to participate in the study and to be measured with the ss-OCT (IOL Master 700, Carl Zeiss Meditec AG, Jena, Germany) device.

Results

Altogether, 1226 eyes of 613 patients were measured consecutively, and 78 eyes were not measured successfully with PCI technology. Among those with unsuccessfully measured eyes, 23 patients were willing to participate in the study, and two of those were also unsuccessfully measured with the ss-OCT device (8.7%, 2/23). However, 91.3% (21/23) of the eyes that were unsuccessfully scanned with PCI technology were measurable with the ss-OCT device. The estimated overall rate of unsuccessful scans with the ss-OCT device was 0.5% (6/1226) (\( p_{{\chi^{2} }} \) < 0.01).

Conclusion

ss-OCT technology significantly improves the rate of attainable axial eye length measurements, especially in eyes with posterior subcapsular cataracts, but also in eyes with dense nuclear cataracts, except for white cataracts.
Hinweise

Enhanced content

To view enhanced content for this article go to https://​doi.​org/​10.​6084/​m9.​figshare.​5890828.

Introduction

The introduction of partial coherence interferometry (PCI) has significantly improved axial eye length measurements in comparison to ultrasound measurements [1, 2]. This is because the accuracy of PCI measurements is significantly higher than the accuracy of applanation ultrasound measurements, and because the results of PCI are less dependent on the examiner [35]. However, conventional PCI technology has one important disadvantage: in contrast to ultrasound, dense nuclear and posterior subcapsular cataracts may reduce the signal-to-noise ratio of the measurement, resulting in an unsuccessful measurement [6, 7]. Although the introduction of a composite scan method has reduced this problem, 4.7% of eyes still cannot be measured with PCI technology [8].
Recently, swept source optical coherence tomography (ss-OCT) was introduced for optical biometry [9]. Among the significant differences between PCI and ss-OCT technology are the wavelength and the measurement setup employed in each.
The aim of this study was to find cases that could not be measured with PCI technology and to assess if they could be measured using ss-OCT technology instead.

Methods

This study included data from patients who underwent cataract surgery. The data were analyzed retrospectively. Patients with severe corneal scarring or head tremor were excluded from this study. All procedures involving human participants were performed in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards. Informed consent was obtained from all individual participants included in the study.
All patients were measured at their pre-assessment visit one week prior to cataract surgery using conventional optical biometry (PCI technology, IOLMaster 500, Carl Zeiss Meditec AG, Jena, Germany). At least five scans were performed with PCI technology in all cases. If the SNR of the composite scan was <2.0, the scan was classified as unsuccessful.
Patients in whom one or both eyes could not be measured were invited to participate in the study and to be measured with the ss-OCT (IOLMaster 700, Carl Zeiss Meditec AG) device. As this device was located at another location 40 min away by public transport, patients were asked to come for this additional appointment within the following three days. When successful, only one measurement (consisting of six scans) was performed with the ss-OCT (Fig. 1); otherwise, a maximum of five measurements (i.e., 30 scans in total) were performed to get one successful scan. In contrast to the PCI device, the ss-OCT device does not provide a composite scan or a signal-to-noise ratio. Therefore, each scan was analyzed separately and was defined as successful if the device was able to give a value for axial eye length. Regarding the protocol of the IOLMaster, a scan was only classified as successful if no warning was given by the device.

Specification of the ss-OCT Device

The ss-OCT (IOLMaster 700) device uses a wavelength of 1055 nm and has a scan depth of 44 mm and a scan width of 6 mm. Its resolution in tissue is 22 µm and its measurement speed is 2000 A-scans per second.

Statistical Analysis

Excel 12.2.0 was used for descriptive statistical analysis. Frequencies are given as percentages and significance testing of proportions was performed using Fisher’s exact test. p < 0.05 was considered to indicate statistical significance.

Results

Altogether, 1226 eyes of 613 patients were measured consecutively on different measurement days using PCI technology in this study. Among those patients, 65 [78 eyes = 6.4% (78/1226)] were not measured successfully (i.e., 52 patients in whom one eye could not be measured successfully and 13 patients in whom neither eye could be measured successfully). Twenty-three of those 65 patients were willing to participate in the study. The 23 patients were on average 68.4 years of age (SD 14.8; 32–88 years). The female:male distribution was 11:12. Eight patients had a best corrected visual acuity of finger counting or less, 9 patients had a best corrected visual acuity ranging between finger counting and 20/200 Snellen, and 5 patients had a best corrected visual acuity of better than 20/200 Snellen. Mean axial eye length (measured with ss-OCT) was 24.07 mm (SD 1.22 mm; 22.06–26.71).
Two of these 23 patients were measured unsuccessfully with ss-OCT technology (2/23; 8.7%). Table 1 presents the reasons for unsuccessful measurement with the PCI device and the ss-OCT device.
Table 1
Reasons for unsuccessful scans using the PCI device and the ss-OCT device
Reason
PCI
ss-OCT
Nuclear cataract
10
2 (20%)
PSC
9
0 (0%)
Nuclear cataract + PSC
4
0 (0%)
Total
23
2 (8.7%)
As 91.3% (21/23) of the unsuccessful scans using PCI technology were measurable with the ss-OCT device, the estimated rate of unsuccessful scans with the ss-OCT device was 0.5% (6/1226). This is a significant reduction (\( p_{{\chi^{2} }} \) < 0.01) of 92.2% (0.5/6.4) compared to PCI optical biometry.

Discussion

ss-OCT technology was shown to significantly increase the proportion of successful axial eye length measurements compared to PCI technology. In fact, all cases with posterior subcapsular cataract were measured successfully with ss-OCT technology.
Twenty percent of the eyes with dense nuclear or white cataracts that were not measurable with PCI technology were also not measurable with the ss-OCT device. In total, the number of unsuccessful scans was reduced from 6.4% to 0.6% when the ss-OCT device was used. Srivannaboon et al. [10] compared the same two devices in a different study setting. They used a much smaller sample size, but performed a direct comparison by measuring all patients with both devices. The main focus of their study was data comparability and the reproducibility of the ss-OCT device, which were found to be good. Similar results were obtained by Shammas et al. [11], Kunert et al. [12], and Kurian et al. [13].
The proportion of unsuccessful scans with PCI technology in Srivannaboon’s study was slightly lower than the proportion in our study (5% vs. 6.4%), possibly due to differences in the study populations or their smaller sample size. Similar to our findings, they observed a significantly higher proportion of successful scans in the ss-OCT group compared to the PCI group. Akman et al. [14] reported a higher proportion of unsuccessful scans (9.0%), but the proportion of eyes with nuclear cataracts and posterior subcapsular cataracts that were unsuccessfully measured by the PCI method but were successfully measured using the ss-OCT device was almost the same as in the present work.
The main reason for the higher success rate of ss-OCT technology is the higher wavelength used (1055 nm) compared to PCI technology (780 nm, IOLMaster), as a shorter wavelength results in a reduced penetration depth due to scattering [15]. This phenomenon is commonly known as Rayleigh scattering, which involves an inverse correlation between scattering and wavelength (to the fourth power).
The advantageous sensitivity of ss-OCT devices compared to time-domain OCT (basically the sum of multiple PCI scans) was recognized more than 10 years ago, but the wavelength used at that time was higher (1300 nm) [16]. Additionally, the feasibility of using ss-OCT technology for biometric measurements of the human eye was proven more than 10 years ago. However, those devices were prototypes and used a different wavelength (1310 nm) to that employed by ss-OCT devices today [17]. One of the first uses of ss-OCT technology for axial eye length measurements was published in 2009 by Chong et al. [18]. They used a similar wavelength to that of the investigated device (1060 nm) and a tunable filter, but they used a much shorter coherence length (28 mm to measure an axial eye length 20 mm in physical size). They were able to measure pigs’ eyes successfully in their study. There is a good correlation between cataract density and Rayleigh scattering, and it should be noted that because the amount of scattering is inversely proportional to the fourth power of the wavelength, a longer wavelength is significantly less prone to Rayleigh scattering [19].
One drawback of this study is that not all of the eyes that were unsuccessfully measured with PCI technology were measured with ss-OCT technology because the ss-OCT device was situated in another clinic some distance away. The geographical distance to that clinic was an obstacle for several patients—possibly even more so for patients with very dense cataracts, as they tended to be older and/or to have reduced mobility.

Conclusions

In summary, ss-OCT technology significantly improves the rate of attainable axial eye length measurements, especially in eyes with posterior subcapsular cataracts, but also in eyes with dense nuclear cataracts, except for white cataracts. It is possible that the use of an even higher wavelength would increase the number of successful scans, but it may also reduce the scan resolution.

Acknowledgements

Funding

No funding or sponsorship was received for this study or the publication of this article.

Authorship

All named authors meet the International Committee of Medical Journal Editors (ICMJE) criteria for authorship for this article, take responsibility for the integrity of the work as a whole, and have given their approval for this version to be published. The authors have no proprietary or financial interest in any of the materials or equipment mentioned in this study.

Disclosures

Oliver Findl is a scientific advisor to Carl Zeiss Meditec AG. Nino Hirnschall gives presentations for Carl Zeiss Meditec AG. Birgit Doeller and Ralph Varsits have nothing to disclose.

Compliance with Ethics Guidelines

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards. Informed consent was obtained from all individual participants included in the study.

Data Availability

The datasets obtained and/or analyzed during the current study are available from the corresponding author on reasonable request.

Open Access

This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License (http://​creativecommons.​org/​licenses/​by-nc/​4.​0/​), which permits any noncommercial use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (https://​creativecommons.​org/​licenses/​by/​4.​0), which permits use, duplication, adaptation, distribution, and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
Literatur
1.
Zurück zum Zitat Drexler W, Findl O, Menapace R, et al. Partial coherence interferometry: a novel approach to biometry in cataract surgery. Am J Ophthalmol. 1998;126(4):524–34.CrossRefPubMed Drexler W, Findl O, Menapace R, et al. Partial coherence interferometry: a novel approach to biometry in cataract surgery. Am J Ophthalmol. 1998;126(4):524–34.CrossRefPubMed
2.
Zurück zum Zitat Goel S, Chua C, Butcher M, Jones CA, Bagga P, Kotta S. Laser vs ultrasound biometry—a study of intra- and interobserver variability. Eye. 2004;18(5):514–8.CrossRefPubMed Goel S, Chua C, Butcher M, Jones CA, Bagga P, Kotta S. Laser vs ultrasound biometry—a study of intra- and interobserver variability. Eye. 2004;18(5):514–8.CrossRefPubMed
3.
Zurück zum Zitat Findl O, Kriechbaum K, Sacu S, et al. Influence of operator experience on the performance of ultrasound biometry compared to optical biometry before cataract surgery. J Cataract Refract Surg. 2003;29(10):1950–5.CrossRefPubMed Findl O, Kriechbaum K, Sacu S, et al. Influence of operator experience on the performance of ultrasound biometry compared to optical biometry before cataract surgery. J Cataract Refract Surg. 2003;29(10):1950–5.CrossRefPubMed
4.
Zurück zum Zitat Findl O, Drexler W, Menapace R, Kiss B, Hitzenberger CK, Fercher AF. Optical biometry in cataract surgery. Dev Ophthalmol. 2002;34:131–40.CrossRefPubMed Findl O, Drexler W, Menapace R, Kiss B, Hitzenberger CK, Fercher AF. Optical biometry in cataract surgery. Dev Ophthalmol. 2002;34:131–40.CrossRefPubMed
5.
Zurück zum Zitat Rajan MS, Keilhorn I, Bell JA. Partial coherence laser interferometry vs conventional ultrasound biometry in intraocular lens power calculations. Eye. 2002;16(5):552–6.CrossRefPubMed Rajan MS, Keilhorn I, Bell JA. Partial coherence laser interferometry vs conventional ultrasound biometry in intraocular lens power calculations. Eye. 2002;16(5):552–6.CrossRefPubMed
6.
Zurück zum Zitat Freeman G, Pesudovs K. The impact of cataract severity on measurement acquisition with the IOLMaster. Acta Ophthalmol Scand. 2005;83(4):439–42.CrossRefPubMed Freeman G, Pesudovs K. The impact of cataract severity on measurement acquisition with the IOLMaster. Acta Ophthalmol Scand. 2005;83(4):439–42.CrossRefPubMed
7.
Zurück zum Zitat Kiss B, Findl O, Menapace R, et al. Biometry of cataractous eyes using partial coherence interferometry: clinical feasibility study of a commercial prototype I. J Cataract Refract Surg. 2002;28(2):224–9.CrossRefPubMed Kiss B, Findl O, Menapace R, et al. Biometry of cataractous eyes using partial coherence interferometry: clinical feasibility study of a commercial prototype I. J Cataract Refract Surg. 2002;28(2):224–9.CrossRefPubMed
8.
Zurück zum Zitat Hirnschall N, Murphy S, Pimenides D, Maurino V, Findl O. Assessment of a new averaging algorithm to increase the sensitivity of axial eye length measurement with optical biometry in eyes with dense cataract. J Cataract Refract Surg. 2011;37(1):45–9.CrossRefPubMed Hirnschall N, Murphy S, Pimenides D, Maurino V, Findl O. Assessment of a new averaging algorithm to increase the sensitivity of axial eye length measurement with optical biometry in eyes with dense cataract. J Cataract Refract Surg. 2011;37(1):45–9.CrossRefPubMed
9.
Zurück zum Zitat Fukuda S, Kawana K, Yasuno Y, Oshika T. Repeatability and reproducibility of anterior ocular biometric measurements with 2-dimensional and 3-dimensional optical coherence tomography. J Cataract Refract Surg. 2010;36(11):1867–73.CrossRefPubMed Fukuda S, Kawana K, Yasuno Y, Oshika T. Repeatability and reproducibility of anterior ocular biometric measurements with 2-dimensional and 3-dimensional optical coherence tomography. J Cataract Refract Surg. 2010;36(11):1867–73.CrossRefPubMed
10.
Zurück zum Zitat Srivannaboon S, Chirapapaisan C, Chonpimai P, Loket S. Clinical comparison of a new swept-source optical coherence tomography-based optical biometer and a time-domain optical coherence tomography-based optical biometer. J Cataract Refract Surg. 2015;41(10):2224–32.CrossRefPubMed Srivannaboon S, Chirapapaisan C, Chonpimai P, Loket S. Clinical comparison of a new swept-source optical coherence tomography-based optical biometer and a time-domain optical coherence tomography-based optical biometer. J Cataract Refract Surg. 2015;41(10):2224–32.CrossRefPubMed
11.
Zurück zum Zitat Shammas HJ, Ortiz S, Shammas MC, Kim SH, Chong C. Biometry measurements using a new large-coherence-length swept-source optical coherence tomographer. J Cataract Refract Surg. 2016;42(1):50–61.CrossRefPubMed Shammas HJ, Ortiz S, Shammas MC, Kim SH, Chong C. Biometry measurements using a new large-coherence-length swept-source optical coherence tomographer. J Cataract Refract Surg. 2016;42(1):50–61.CrossRefPubMed
12.
Zurück zum Zitat Kunert KS, Peter M, Blum M, et al. Repeatability and agreement in optical biometry of a new swept-source optical coherence tomography-based biometer versus partial coherence interferometry and optical low-coherence reflectometry. J Cataract Refract Surg. 2016;42(1):76–83.CrossRefPubMed Kunert KS, Peter M, Blum M, et al. Repeatability and agreement in optical biometry of a new swept-source optical coherence tomography-based biometer versus partial coherence interferometry and optical low-coherence reflectometry. J Cataract Refract Surg. 2016;42(1):76–83.CrossRefPubMed
13.
Zurück zum Zitat Kurian M, Negalur N, Das S, et al. Biometry with a new swept-source optical coherence tomography biometer: repeatability and agreement with an optical low-coherence reflectometry device. J Cataract Refract Surg. 2016;42(4):577–81.CrossRefPubMed Kurian M, Negalur N, Das S, et al. Biometry with a new swept-source optical coherence tomography biometer: repeatability and agreement with an optical low-coherence reflectometry device. J Cataract Refract Surg. 2016;42(4):577–81.CrossRefPubMed
14.
Zurück zum Zitat Akman A, Asena L, Gungor SG. Evaluation and comparison of the new swept source OCT-based IOLMaster 700 with the IOLMaster 500. Br J Ophthalmol. 2016;100(9):1201–5.CrossRefPubMed Akman A, Asena L, Gungor SG. Evaluation and comparison of the new swept source OCT-based IOLMaster 700 with the IOLMaster 500. Br J Ophthalmol. 2016;100(9):1201–5.CrossRefPubMed
15.
Zurück zum Zitat Povazay B, Hermann B, Unterhuber A, et al. Three-dimensional optical coherence tomography at 1050 nm versus 800 nm in retinal pathologies: enhanced performance and choroidal penetration in cataract patients. J Biomed Opt. 2007;12(4):041211.CrossRefPubMed Povazay B, Hermann B, Unterhuber A, et al. Three-dimensional optical coherence tomography at 1050 nm versus 800 nm in retinal pathologies: enhanced performance and choroidal penetration in cataract patients. J Biomed Opt. 2007;12(4):041211.CrossRefPubMed
16.
Zurück zum Zitat Choma M, Sarunic M, Yang C, Izatt J. Sensitivity advantage of swept source and Fourier domain optical coherence tomography. Opt Express. 2003;11(18):2183–9.CrossRefPubMed Choma M, Sarunic M, Yang C, Izatt J. Sensitivity advantage of swept source and Fourier domain optical coherence tomography. Opt Express. 2003;11(18):2183–9.CrossRefPubMed
17.
Zurück zum Zitat Yasuno Y, Madjarova VD, Makita S, et al. Three-dimensional and high-speed swept-source optical coherence tomography for in vivo investigation of human anterior eye segments. Opt Express. 2005;13(26):10652–64.CrossRefPubMed Yasuno Y, Madjarova VD, Makita S, et al. Three-dimensional and high-speed swept-source optical coherence tomography for in vivo investigation of human anterior eye segments. Opt Express. 2005;13(26):10652–64.CrossRefPubMed
18.
Zurück zum Zitat Chong C, Suzuki T, Totsuka K, Morosawa A, Sakai T. Large coherence length swept source for axial length measurement of the eye. Appl Opt. 2009;48(10):D144–50.CrossRefPubMed Chong C, Suzuki T, Totsuka K, Morosawa A, Sakai T. Large coherence length swept source for axial length measurement of the eye. Appl Opt. 2009;48(10):D144–50.CrossRefPubMed
19.
Zurück zum Zitat van den Berg TJ, Coppens JC. Conversion of lens slit lamp photographs into physical light-scattering units. Invest Ophthalmol Vis Sci. 1999;40(9):2151–7.PubMed van den Berg TJ, Coppens JC. Conversion of lens slit lamp photographs into physical light-scattering units. Invest Ophthalmol Vis Sci. 1999;40(9):2151–7.PubMed
Metadaten
Titel
Enhanced Penetration for Axial Length Measurement of Eyes with Dense Cataracts Using Swept Source Optical Coherence Tomography: A Consecutive Observational Study
verfasst von
Nino Hirnschall
Ralph Varsits
Birgit Doeller
Oliver Findl
Publikationsdatum
01.03.2018
Verlag
Springer Healthcare
Erschienen in
Ophthalmology and Therapy / Ausgabe 1/2018
Print ISSN: 2193-8245
Elektronische ISSN: 2193-6528
DOI
https://doi.org/10.1007/s40123-018-0122-1

Weitere Artikel der Ausgabe 1/2018

Ophthalmology and Therapy 1/2018 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.