Enhancing the precision of circular stapled colorectal anastomosis: could powered stapler technology provide the solution?

verfasst von: R. Mirnezami, A. Soares, M. Chand

Erschienen in: Techniques in Coloproctology | Ausgabe 7/2019

Begleitmaterial
Supplementary material 1 (MP4 110465 kb)
Hinweise

Electronic supplementary material

The online version of this article (https://​doi.​org/​10.​1007/​s10151-019-02031-9) contains supplementary material, which is available to authorized users.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

The use of circular stapling devices to facilitate low colorectal anastomosis (CRA) was first described in the 1970s. This approach is now standard practice and has consistently demonstrated equivalence in terms of safety and efficacy compared with hand-sewn anastomosis, with the added advantages of shorter anastomotic time and greater reproducibility. Briefly, the key elements of circular stapled CRA comprise: (1) selecting a stapler gun of appropriate size; (2) securing the anvil in the distal end of the mobilised colonic conduit; (3) performing a rendezvous manoeuvre to securely dock the anvil onto the end of the stapler gun, which has been advanced from below; and (4) firing the stapler to complete the anastomosis.
While there have been subtle modifications to staple design and configuration, there has been relatively little progress made in circular stapler functionality over the past few decades. Further advances in stapling technology could enhance the precision of CRA with the potential to positively impact on the incidence of anastomotic leakage and on the need for diverting stoma. However, refining the existing design requires the four key steps outlined above to be carefully considered to determine how and/or where the technology can be fine-tuned; sizing and anvil application are obligatory steps, while the rendezvous technique represents an elegant design that lends itself readily to both open and intra-corporeal CRA. The final and arguably most critical step in the technique, namely, discharging the stapler, represents a component in the pathway, where refinement may be both feasible and advantageous, and in this regard, the powered surgical stapler is emerging as a potential solution.
Powered surgical stapler devices were first introduced in 2011 and have been applied to a wide variety of clinical applications. More recently, powered circular stapling devices have been developed specifically intended for use in colorectal surgery. We provide a technical description of our initial experience with the ECHELON CIRCULAR™ Powered Stapler (Ethicon, Somerville, NJ, USA), which was employed during laparoscopic anterior resection carried out at a tertiary colorectal referral centre (University College London Hospital (UCLH), London, UK).

Technical description

Laparoscopic partial mesorectal excision (PME) was performed in a previously medically fit 68-year-old, male with a biopsy proven adenocarcinoma of the upper rectum, staged with magnetic resonance imaging (MRI) as T4a N2, with no evidence of distant metastases. Dissection was performed in the standard fashion with medial-to-lateral dissection, identification, and preservation of the left ureter and gonadal vessels, proximal ligation, and division of the inferior mesenteric vein and artery, followed by PME down to the intended level of rectal transection. An extended cuff of anterior abdominal wall parietal peritoneum was excised en bloc with the specimen, where the serosal surface was breached by tumour. The splenic flexure was mobilised fully before the rectum was divided intra-corporeally using an Echelon Flex 45 stapler. A 4 cm trans-umbilical incision was used to extract the specimen and apply the anvil to the distal end of the colonic conduit in the standard fashion. A CDH 31 ECHELON CIRCULAR™ Powered Stapler was used to construct a tension-free low CRA. Anastomotic donuts were checked and confirmed to be both intact and of adequate thickness. An air-leak test was performed on table which demonstrated no evidence of leak at the anastomotic site and the anastomosis was directly visualised by flexible sigmoidoscopy, confirming its integrity. In view of the favourable circumstances and adequacy of the anastomosis, the decision was taken not to construct a diverting ileostomy. The patient made an uneventful recovery in hospital and was discharged home with a final histological stage of pT3N1.

Discussion

A key challenge with the operation and handling of conventional, manual stapling devices is the force required to effect penetration of staples through tissue, which can be variable depending on the operative circumstances. In generating this force, the hands of the operating surgeon are prone to unwanted movements, which are further magnified along the length of the stapler shaft to the site of intended anastomosis when the stapler is being fired. These factors are believed to be, at least in part, responsible for technical errors that have been described during stapled CRA. For example, a study by Offodile et al. [1] provided data on 349 consecutive patients undergoing left-sided colon and rectal resections at a single centre and identified a circular stapled anastomotic malfunction rate of 9% (32/349 patients). Although data were comparable between the control group and patients who experienced anastomotic error, the latter had a significantly higher incidence of positive air-leak tests and proximal diversion (34% versus 16%; p = 0.0003).
It is logical from a tissue mechanics perspective that undue/erratic movement at the anastomotic site could result in microvascular trauma and fractures at the bowel interface, potentially compromising the anastomosis. Moreover, several previous studies have demonstrated that some surgeons are simply incapable of generating the grip-strength force required to fire circular staplers [2, 3]. Design improvements in circular stapling technology that enhance the device-to-user interface (i.e., improve handling, reduce force-to-fire requirements) whilst simultaneously addressing issues with device-to-tissue interaction (i.e., reducing unwanted movement at the anastomosis during construction) have the potential to decrease the rate of technical errors and lead to improvements in clinical outcome.
Over the past, few years powered surgical staplers have emerged on the market and these may offer a means of overcoming the force-to-fire barrier in stapled CRA. It is believed that reducing reliance on the surgeon to generate this force will help to improve the stability of the stapler during firing, potentially greatly reducing unwanted movements at the anastomotic site. It is also conceivable that the powered operation could result in more consistent compression of stapler onto tissue, potentially minimising the average compressive force experienced by target tissue compared with using a manual stapler.
Another new feature of this stapler is the staple configuration which is described as 3D. There is relatively sparse literature investigating the effect of staple configuration on anastomosis and those reports that are published are of linear stapled anastomoses. Foo et al. [4] have studied the outcomes of 340 patients and whether triple staple lines and enhanced configuration (3D) have an effect on clinical outcomes. They could not identify whether any benefit in patient morbidity was specifically associated with staple configuration, but there was an overall improvement when a stapler with triple staples and 3D configuration was used over conventional staplers. There was no significant change in the anastomotic leak rate. According to the manufacturer of the circular stapler in this report, the 3D stapling configuration and offsetting of the staple legs provide a more even compression of tissue and more optimal conditions for anastomotic healing. These benefits are in addition to the advantages of atraumatic Gripping Surface Technology that has separately been shown to provide gentler handling with a reduction in compressive forces on tissue [5].
The novel technology introduced in this stapler can confuse somewhat with automation being combined with claimed improved stapling technology. On one side, there are the obvious benefits which come with reduced movement at the time of firing. However, whether there is further optimisation of the anastomosis by the new stapling technology is more difficult to quantify or appreciate at first glance. Further clinical studies on this claim are welcomed.
Although the gains are likely to be incremental in the short term, we believe that in the longer term, we may see more significant clinical outcome improvement and hope that this will be reported in clinical studies. The powered device will compensate for differences in physical grip strength, unintentional movement, and level of experience, to create a more technically robust and reproducible anastomosis than is currently possible with manual alternatives.

Compliance with ethical standards

Conflict of interest

Manish Chand receives fees for teaching and consulting from Ethicon/Johnson and Johnson. The other authors declare that they have no conflict of interest.

Ethical approval

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.
The patient involved in this case was appropriately consented for this publication using the institution's policy for media consent.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

Die Chirurgie

Print-Titel

Das Abo mit mehr Tiefe

Mit der Zeitschrift Die Chirurgie erhalten Sie zusätzlich Online-Zugriff auf weitere 43 chirurgische Fachzeitschriften, CME-Fortbildungen, Webinare, Vorbereitungskursen zur Facharztprüfung und die digitale Enzyklopädie e.Medpedia.

Bis 30. April 2024 bestellen und im ersten Jahr nur 199 € zahlen!

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Anhänge

Electronic supplementary material

Below is the link to the electronic supplementary material.
Supplementary material 1 (MP4 110465 kb)
Literatur
1.
Zurück zum Zitat Offodile AC, Feingold DL, Nasar A, Whelan RL, Arnell TD (2010) High incidence of technical errors involving the EEA circular stapler: a single institution experience. J Am Coll Surg 210(3):331–335CrossRefPubMed Offodile AC, Feingold DL, Nasar A, Whelan RL, Arnell TD (2010) High incidence of technical errors involving the EEA circular stapler: a single institution experience. J Am Coll Surg 210(3):331–335CrossRefPubMed
2.
Zurück zum Zitat Kono E et al (2014) Ergonomic evaluation of a mechanical anastomotic stapler used by Japanese surgeons. Surg Today 44(6):1040–1047CrossRefPubMed Kono E et al (2014) Ergonomic evaluation of a mechanical anastomotic stapler used by Japanese surgeons. Surg Today 44(6):1040–1047CrossRefPubMed
3.
Zurück zum Zitat Kono E, Tomizawa Y, Matsuo T, Nomura S (2012) Rating and issues of mechanical anastomotic staplers in surgical practice: a survey of 241 Japanese gastroenterological surgeons. Surg Today 42(10):962–972CrossRefPubMed Kono E, Tomizawa Y, Matsuo T, Nomura S (2012) Rating and issues of mechanical anastomotic staplers in surgical practice: a survey of 241 Japanese gastroenterological surgeons. Surg Today 42(10):962–972CrossRefPubMed
4.
Zurück zum Zitat Foo CC, Chiu AHO, Yip J, Law WL (2018) Does advancement in stapling technology with triple-row and enhanced staple configurations confer additional safety? A matched comparison of 340 stapled ileocolic anastomoses. Surg Endosc 32(7):3122–3130CrossRefPubMed Foo CC, Chiu AHO, Yip J, Law WL (2018) Does advancement in stapling technology with triple-row and enhanced staple configurations confer additional safety? A matched comparison of 340 stapled ileocolic anastomoses. Surg Endosc 32(7):3122–3130CrossRefPubMed
5.
Zurück zum Zitat Fegelman E et al (2017) Evaluation of a powered stapler system with gripping surface technology on surgical interventions required during laparoscopic sleeve gastrectomy. J Laparoendosc Adv Surg Tech A 27(5):489–494CrossRefPubMedPubMedCentral Fegelman E et al (2017) Evaluation of a powered stapler system with gripping surface technology on surgical interventions required during laparoscopic sleeve gastrectomy. J Laparoendosc Adv Surg Tech A 27(5):489–494CrossRefPubMedPubMedCentral
Metadaten
Titel
Enhancing the precision of circular stapled colorectal anastomosis: could powered stapler technology provide the solution?
verfasst von
R. Mirnezami
A. Soares
M. Chand
Publikationsdatum
05.07.2019
Verlag
Springer International Publishing
Erschienen in
Techniques in Coloproctology / Ausgabe 7/2019
Print ISSN: 1123-6337
Elektronische ISSN: 1128-045X
DOI
https://doi.org/10.1007/s10151-019-02031-9

Weitere Artikel der Ausgabe 7/2019

Techniques in Coloproctology 7/2019 Zur Ausgabe

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

CME: 2 Punkte

Prof. Dr. med. Gregor Antoniadis Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

CME: 2 Punkte

Dr. med. Benjamin Meyknecht, PD Dr. med. Oliver Pieske Das Webinar S2e-Leitlinie „Distale Radiusfraktur“ beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

CME: 2 Punkte

Dr. med. Mihailo Andric
Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.