Open Access 01.12.2016 | Erratum

Erratum to: The accuracy and timeliness of a Point Of Care lactate measurement in patients with Sepsis

verfasst von: Fatene Ismail, William G. Mackay, Andrew Kerry, Harry Staines, Kevin D. Rooney

Erschienen in: Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine | Ausgabe 1/2016

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN
Hinweise
The online version of the original article can be found under doi:10.​1186/​s13049-015-0151-x.
The online version of the original article can be found at http://​dx.​doi.​org/​10.​1186/​s13049-015-0151-x.

Erratum

After publication of the original article [1], the authors noticed a mistake in Table 2. Patient numbers in lactate classification categories Low, Medium and High were incorrectly reported as forty one, seven and two respectively in the GEM 4000 column. The patient numbers that should have been reported in Table 2 are forty, three and seven. Table 2 appears in its correct form in this erratum.
Table 2
Lactate risk category classification determined by the i-STAT and the blood gas analysers
Risk category
i-STAT
GEM 4000
Risk category
i-STAT
OMNI S
Low <2.5 mmol/L (n = 40)
40
40
Low <2.5 mmol/L (n = 45)
45
45
Medium 2.5 – 3.99 mmol/L (n = 3)
3
3
Medium 2.5 – 3.99 mmol/L (n = 2)
2
2
High ≥4 mmol/L (n = 7)
7
7
High ≥4 mmol/L (n = 0)
0
0
The above table shows that all lactate samples (50) analysed on the i-STAT and the GEM 4000 analysers fell in the same lactate risk level categories
Similarly, all lactate samples (47) analysed on the i-STAT and the OMNI S fell in the same lactate risk level categories
Additionally, the incorrect patient numbers from Table 2 were reported in the ‘Results’ section. The ‘Comparison of i-STAT against GEM premier 4000’ sub-section should therefore have read as follows:
“Among the 50 samples (from 11 ICU patients) analysed on the GEM 4000 and the i-STAT; there were forty samples classified as low, three samples as medium and seven samples as high lactate risk level categories.”
This mistake does not alter the validity of the results and conclusions as all lactate samples analysed on the i-STAT and GEM 4000 analysers fell in the same lactate risk categories.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
download
DOWNLOAD
print
DRUCKEN
Metadaten
Titel
Erratum to: The accuracy and timeliness of a Point Of Care lactate measurement in patients with Sepsis
verfasst von
Fatene Ismail
William G. Mackay
Andrew Kerry
Harry Staines
Kevin D. Rooney
Publikationsdatum
01.12.2016