Estimation of respiratory volume from thoracoabdominal breathing distances: comparison of two models of machine learning

verfasst von: Rémy Dumond, Steven Gastinger, Hala Abdul Rahman, Alexis Le Faucheur, Patrice Quinton, Haitao Kang, Jacques Prioux

Erschienen in: European Journal of Applied Physiology | Ausgabe 8/2017

Einloggen, um Zugang zu erhalten

Abstract

Purpose

The purposes of this study were to both improve the accuracy of respiratory volume (V) estimates using the respiratory magnetometer plethysmography (RMP) technique and facilitate the use of this technique.

Method

We compared two models of machine learning (ML) for estimating \(\widehat{V}_\mathrm{RMP}\): a linear model (multiple linear regression—MLR) and a nonlinear model (artificial neural network—ANN), and we used cross-validation to validate these models. Fourteen healthy adults, aged \(24.1 \pm 3.4\) years participated in the present study. The protocol was conducted in a laboratory test room. The anteroposterior displacements of the rib cage and abdomen, and the axial displacements of the chest wall and spine were measured using two pairs of magnetometers. \(\widehat{V}_\mathrm{RMP}\) was estimated from these four signals, and the respiratory volume was simultaneously measured using a spirometer (\(V_\mathrm{SP}\)) under lying, sitting and standing conditions as well as various exercise conditions (working on computer, treadmill walking at 4 and 6 km\(\mathrm{ \ h}^{-1}\), treadmill running at 9 and 12  km \(\mathrm{ \ h}^{-1}\) and ergometer cycling at 90 and 110 W).

Results

The results from the ANN model fitted the spirometer volume significantly better than those obtained through MLR. Considering all activities, the difference between \(\widehat{V}_\mathrm{RMP}\) and \(V_\mathrm{SP}\) (bias) was higher for the MLR model (\(0.00191 \pm 0.141\) L) than for the ANN model (\(0.00158 \pm 0.150\) L).

Conclusion

Our results demonstrate that this new processing approach for RMP seems to be a valid tool for estimating V with sufficient accuracy during lying, sitting and standing and under various exercise conditions.
Literatur
Zurück zum Zitat Augousti AT (1997) A theoretical study of the robustness of the isovolume calibration method for a two-compartment model of breathing, based on an analysis of the connected cylinders model. Phys Med Biol 42(2):283CrossRefPubMed Augousti AT (1997) A theoretical study of the robustness of the isovolume calibration method for a two-compartment model of breathing, based on an analysis of the connected cylinders model. Phys Med Biol 42(2):283CrossRefPubMed
Zurück zum Zitat Baba R, Mori E, Tauchi N, Nagashima M (2002) Simple exponential regression model to describe the relation between minute ventilation and oxygen uptake during incremental exercise. Nagoya J Med Sci 65(3–4):95–102PubMed Baba R, Mori E, Tauchi N, Nagashima M (2002) Simple exponential regression model to describe the relation between minute ventilation and oxygen uptake during incremental exercise. Nagoya J Med Sci 65(3–4):95–102PubMed
Zurück zum Zitat Baevsky RM, Baranov VM, Funtova II, Diedrich A, Pashenko AV, Chernikova AG, Drescher J, Jordan J, Tank J (2007) Autonomic cardiovascular and respiratory control during prolonged spaceflights aboard the international space station. J Appl Physiol 103(1):156–161CrossRefPubMed Baevsky RM, Baranov VM, Funtova II, Diedrich A, Pashenko AV, Chernikova AG, Drescher J, Jordan J, Tank J (2007) Autonomic cardiovascular and respiratory control during prolonged spaceflights aboard the international space station. J Appl Physiol 103(1):156–161CrossRefPubMed
Zurück zum Zitat Bishop C (2006) Pattern recognition and machine learning. Springer, New York Bishop C (2006) Pattern recognition and machine learning. Springer, New York
Zurück zum Zitat Black AMS, Bambridge A, Kunst G, Millard RK (2001) Progress in non-invasive respiratory monitoring using uncalibrated breathing movement components. Physiol Meas 22(1):245CrossRefPubMed Black AMS, Bambridge A, Kunst G, Millard RK (2001) Progress in non-invasive respiratory monitoring using uncalibrated breathing movement components. Physiol Meas 22(1):245CrossRefPubMed
Zurück zum Zitat Bland JM, Altman D (1986) Statistical methods for assessing agreement between two methods of clinical measurement. The Lancet 327(8476):307–310CrossRef Bland JM, Altman D (1986) Statistical methods for assessing agreement between two methods of clinical measurement. The Lancet 327(8476):307–310CrossRef
Zurück zum Zitat Bokov P, Delclaux C (2016) Interprétation et utilisation des explorations fonctionnelles respiratoires de routine de ladulte : spirométrie, volumes non mobilisables, diffusion, hématose, test de provocation bronchique la métacholine et test de marche. La Revue de Médecine Interne 37(2):100–110CrossRefPubMed Bokov P, Delclaux C (2016) Interprétation et utilisation des explorations fonctionnelles respiratoires de routine de ladulte : spirométrie, volumes non mobilisables, diffusion, hématose, test de provocation bronchique la métacholine et test de marche. La Revue de Médecine Interne 37(2):100–110CrossRefPubMed
Zurück zum Zitat Chhabra SK (2015) Interpretation of spirometry: selection of predicted values and defining abnormality. Indian J Chest Dis Allied Sci 57(2):91–105PubMed Chhabra SK (2015) Interpretation of spirometry: selection of predicted values and defining abnormality. Indian J Chest Dis Allied Sci 57(2):91–105PubMed
Zurück zum Zitat Clarenbach CF, Senn O, Brack T, Kohler M, Bloch KE (2005) Monitoring of ventilation during exercise by a portable respiratory inductive plethysmograph. Chest 128(3):1282–1290CrossRefPubMed Clarenbach CF, Senn O, Brack T, Kohler M, Bloch KE (2005) Monitoring of ventilation during exercise by a portable respiratory inductive plethysmograph. Chest 128(3):1282–1290CrossRefPubMed
Zurück zum Zitat Consolazio CF, Nelson RA, Daws TA, Krzywicki HJ, Johnson HL, Barnhart RA (1971) Body weight, heart rate, and ventilatory volume relationships to oxygen uptakes. Am J Clin Nutr 24(9):1180–1185PubMed Consolazio CF, Nelson RA, Daws TA, Krzywicki HJ, Johnson HL, Barnhart RA (1971) Body weight, heart rate, and ventilatory volume relationships to oxygen uptakes. Am J Clin Nutr 24(9):1180–1185PubMed
Zurück zum Zitat Datta SR, Ramanathan NL (1969) Energy expenditure in work predicted from heart rate and pulmonary ventilation. J Appl Physiol 26(3):297–302PubMed Datta SR, Ramanathan NL (1969) Energy expenditure in work predicted from heart rate and pulmonary ventilation. J Appl Physiol 26(3):297–302PubMed
Zurück zum Zitat Dellaca RL, Aliverti A, Lo Mauro A, Lutchen KR, Pedotti A, Suki B (2015) Correlated variability in the breathing pattern and end-expiratory lung volumes in conscious humans. PLoS One 10(3):1–15CrossRef Dellaca RL, Aliverti A, Lo Mauro A, Lutchen KR, Pedotti A, Suki B (2015) Correlated variability in the breathing pattern and end-expiratory lung volumes in conscious humans. PLoS One 10(3):1–15CrossRef
Zurück zum Zitat Dua S (2014) Machine learning in healthcare informatics. Springer, BerlinCrossRef Dua S (2014) Machine learning in healthcare informatics. Springer, BerlinCrossRef
Zurück zum Zitat Emeriaud G, Eberhard A, Benchetrit G, Debillon T, Baconnier P (2008) Calibration of respiratory inductance plethysmograph in preterm infants with different respiratory conditions. Pediatr Pulmonol 43(11):1135–1141CrossRefPubMed Emeriaud G, Eberhard A, Benchetrit G, Debillon T, Baconnier P (2008) Calibration of respiratory inductance plethysmograph in preterm infants with different respiratory conditions. Pediatr Pulmonol 43(11):1135–1141CrossRefPubMed
Zurück zum Zitat Fekr AR, Radecka K, Zilic Z (2015) Design and evaluation of an intelligent remote tidal volume variability monitoring system in e-health applications. IEEE J Biomed Health Inform 19(5):1532–1548CrossRefPubMed Fekr AR, Radecka K, Zilic Z (2015) Design and evaluation of an intelligent remote tidal volume variability monitoring system in e-health applications. IEEE J Biomed Health Inform 19(5):1532–1548CrossRefPubMed
Zurück zum Zitat Fieselmann JF, Hendryx MS, Helms CM, Wakefield DS (1993) Respiratory rate predicts cardiopulmonary arrest for internal medicine in patients. J Gen Intern Med 8(7):354–360CrossRefPubMed Fieselmann JF, Hendryx MS, Helms CM, Wakefield DS (1993) Respiratory rate predicts cardiopulmonary arrest for internal medicine in patients. J Gen Intern Med 8(7):354–360CrossRefPubMed
Zurück zum Zitat Fisher DMMD (1995) When is the ex-premature infant no longer at risk for apnea? Anesthesiology 82(4):807–808CrossRefPubMed Fisher DMMD (1995) When is the ex-premature infant no longer at risk for apnea? Anesthesiology 82(4):807–808CrossRefPubMed
Zurück zum Zitat Folke M, Cernerud L, Ekstrm M, Hk B (2003) Critical review of non-invasive respiratory monitoring in medical care. Med Biol Eng Comput 41(4):377–383CrossRefPubMed Folke M, Cernerud L, Ekstrm M, Hk B (2003) Critical review of non-invasive respiratory monitoring in medical care. Med Biol Eng Comput 41(4):377–383CrossRefPubMed
Zurück zum Zitat Galvagno SM, Brayanov J, Corneille MG, Voscopoulos CJ, Sordo S, Ladd D, Freeman J (2015) Non-invasive respiratory volume monitoring in patients with traumatic thoracic injuries. Trauma 17(3):219–223CrossRef Galvagno SM, Brayanov J, Corneille MG, Voscopoulos CJ, Sordo S, Ladd D, Freeman J (2015) Non-invasive respiratory volume monitoring in patients with traumatic thoracic injuries. Trauma 17(3):219–223CrossRef
Zurück zum Zitat Gastinger S, Sefati H, Nicolas G, Sorel A, Gratas-Delamarche A, Prioux J (2010a) Estimates of ventilation from measurements of rib cage and abdominal distances: a portable device. Eur J Appl Physiol 109(6):1179–1189CrossRefPubMed Gastinger S, Sefati H, Nicolas G, Sorel A, Gratas-Delamarche A, Prioux J (2010a) Estimates of ventilation from measurements of rib cage and abdominal distances: a portable device. Eur J Appl Physiol 109(6):1179–1189CrossRefPubMed
Zurück zum Zitat Gastinger S, Sorel A, Nicolas G, Gratas-Delamarche A, Prioux J (2010b) A comparison between ventilation and heart rate as indicator of oxygen uptake during different intensities of exercise. J Sports Sci Med 9(1):110–118PubMedCentralPubMed Gastinger S, Sorel A, Nicolas G, Gratas-Delamarche A, Prioux J (2010b) A comparison between ventilation and heart rate as indicator of oxygen uptake during different intensities of exercise. J Sports Sci Med 9(1):110–118PubMedCentralPubMed
Zurück zum Zitat Gastinger S, Sefati H, Nicolas G, Sorel A, Gratas-Delamarche A, Prioux J (2011) A new method to estimate energy expenditure from abdominal and rib cage distances. Eur J Appl Physiol 111(11):2823–2835CrossRefPubMed Gastinger S, Sefati H, Nicolas G, Sorel A, Gratas-Delamarche A, Prioux J (2011) A new method to estimate energy expenditure from abdominal and rib cage distances. Eur J Appl Physiol 111(11):2823–2835CrossRefPubMed
Zurück zum Zitat Gastinger S, Donnelly A, Dumond R, Prioux J (2014) A review of the evidence for the use of ventilation as a surrogate measure of energy expenditure. J Parenter Enter Nutr 38(8):926–938CrossRef Gastinger S, Donnelly A, Dumond R, Prioux J (2014) A review of the evidence for the use of ventilation as a surrogate measure of energy expenditure. J Parenter Enter Nutr 38(8):926–938CrossRef
Zurück zum Zitat Gilbert R, Ashutosh K Jr, JHA, Rana S, Peppi D (1977) Prospective study of controlled oxygen therapy: poor prognosis of patients with asynchronous breathing. Chest 1(4):456–462 Gilbert R, Ashutosh K Jr, JHA, Rana S, Peppi D (1977) Prospective study of controlled oxygen therapy: poor prognosis of patients with asynchronous breathing. Chest 1(4):456–462
Zurück zum Zitat Heyde C, Leutheuser H, Eskofier B, Roecker K, Gollhofer A (2014) Respiratory inductance plethysmography—a rationale for validity during exercise. Med Sci Sports Exerc 46(3):488–495CrossRefPubMed Heyde C, Leutheuser H, Eskofier B, Roecker K, Gollhofer A (2014) Respiratory inductance plethysmography—a rationale for validity during exercise. Med Sci Sports Exerc 46(3):488–495CrossRefPubMed
Zurück zum Zitat Heyde C, Mahler H, Roecker K, Gollhofer A (2015) A wearable respiratory monitoring device-the between-days variability of calibration. Int J Sports Med 36(1):29–34PubMed Heyde C, Mahler H, Roecker K, Gollhofer A (2015) A wearable respiratory monitoring device-the between-days variability of calibration. Int J Sports Med 36(1):29–34PubMed
Zurück zum Zitat John D, Staudenmayer J, Freedson P (2013) Simple to complex modeling of breathing volume using a motion sensor. Sci Total Environ 454455:184–188CrossRef John D, Staudenmayer J, Freedson P (2013) Simple to complex modeling of breathing volume using a motion sensor. Sci Total Environ 454455:184–188CrossRef
Zurück zum Zitat Khemani RG, Flink R, Hotz J, Ross PA, Ghuman A, Newth CJ (2015) Respiratory inductance plethysmography calibration for pediatric upper airway obstruction: an animal model. Pediatr Res 77(1–1):75–83CrossRefPubMed Khemani RG, Flink R, Hotz J, Ross PA, Ghuman A, Newth CJ (2015) Respiratory inductance plethysmography calibration for pediatric upper airway obstruction: an animal model. Pediatr Res 77(1–1):75–83CrossRefPubMed
Zurück zum Zitat Koegelenberg CF, Swart F, Irusen EM (2013) Guideline for office spirometry in adults, 2012. S Afr Med J 103(1):52–62CrossRef Koegelenberg CF, Swart F, Irusen EM (2013) Guideline for office spirometry in adults, 2012. S Afr Med J 103(1):52–62CrossRef
Zurück zum Zitat Konno K, Mead J (1967) Measurement of the separate volume changes of rib cage and abdomen during breathing. J Appl Physiol 22(3):407–422PubMed Konno K, Mead J (1967) Measurement of the separate volume changes of rib cage and abdomen during breathing. J Appl Physiol 22(3):407–422PubMed
Zurück zum Zitat Krehel M, Schmid M, Rossi RM, Boesel LF, Bona GL, Scherer LJ (2014) An optical fibre-based sensor for respiratory monitoring. Sensors 14(7):13,088CrossRef Krehel M, Schmid M, Rossi RM, Boesel LF, Bona GL, Scherer LJ (2014) An optical fibre-based sensor for respiratory monitoring. Sensors 14(7):13,088CrossRef
Zurück zum Zitat Kurth CD, Cot CJ (2015) Postoperative apnea in former preterm infants general anesthesia or spinal anesthesiado we have an answer? Anesthesiology 123(1):15–17CrossRefPubMed Kurth CD, Cot CJ (2015) Postoperative apnea in former preterm infants general anesthesia or spinal anesthesiado we have an answer? Anesthesiology 123(1):15–17CrossRefPubMed
Zurück zum Zitat Leino K, Nunes S, Valta P, Takala J (2001) Validation of a new respiratory inductive plethysmograph. Acta Anaesthesiol Scand 45(1):104–111CrossRefPubMed Leino K, Nunes S, Valta P, Takala J (2001) Validation of a new respiratory inductive plethysmograph. Acta Anaesthesiol Scand 45(1):104–111CrossRefPubMed
Zurück zum Zitat Li G, Wei J, Huang H, Gaebler CP, Yuan A, Deasy JO (2015) Automatic assessment of average diaphragm motion trajectory from 4dct images through machine learning. Biomed Phys Eng Express 1(4):045,015CrossRef Li G, Wei J, Huang H, Gaebler CP, Yuan A, Deasy JO (2015) Automatic assessment of average diaphragm motion trajectory from 4dct images through machine learning. Biomed Phys Eng Express 1(4):045,015CrossRef
Zurück zum Zitat Liu GZ, Guo YW, Zhu QS, Huang BY, Wang L (2011) Estimation of respiration rate from three-dimensional acceleration data based on body sensor network. Telemed e-Health 17(9):705–711CrossRef Liu GZ, Guo YW, Zhu QS, Huang BY, Wang L (2011) Estimation of respiration rate from three-dimensional acceleration data based on body sensor network. Telemed e-Health 17(9):705–711CrossRef
Zurück zum Zitat Liu JJ, Huang MC, Xu W, Zhang X, Stevens L, Alshurafa N, Sarrafzadeh M (2015) Breathsens: a continuous on-bed respiratory monitoring system with torso localization using an unobtrusive pressure sensing array. IEEE J Biomed Health Inform 19(5):1682–1688CrossRefPubMed Liu JJ, Huang MC, Xu W, Zhang X, Stevens L, Alshurafa N, Sarrafzadeh M (2015) Breathsens: a continuous on-bed respiratory monitoring system with torso localization using an unobtrusive pressure sensing array. IEEE J Biomed Health Inform 19(5):1682–1688CrossRefPubMed
Zurück zum Zitat Liu S, Gao R, He Q, Staudenmayer J, Freedson P (2012) Improved regression models for ventilation estimation based on chest and abdomen movements. Physiol Meas 33(1):79CrossRefPubMedCentralPubMed Liu S, Gao R, He Q, Staudenmayer J, Freedson P (2012) Improved regression models for ventilation estimation based on chest and abdomen movements. Physiol Meas 33(1):79CrossRefPubMedCentralPubMed
Zurück zum Zitat van Loon K, van Zaane B, Bosch EJ, Kalkman CJ, Peelen LM (2015) Non-invasive continuous respiratory monitoring on general hospital wards: a systematic review. PLoS One 10(12):1–14 van Loon K, van Zaane B, Bosch EJ, Kalkman CJ, Peelen LM (2015) Non-invasive continuous respiratory monitoring on general hospital wards: a systematic review. PLoS One 10(12):1–14
Zurück zum Zitat McCaughey EJ, McLachlan AJ, Gollee H (2014) Non-intrusive real-time breathing pattern detection and classification for automatic abdominal functional electrical stimulation. Med Eng Phys 36(8):1057–1061CrossRefPubMed McCaughey EJ, McLachlan AJ, Gollee H (2014) Non-intrusive real-time breathing pattern detection and classification for automatic abdominal functional electrical stimulation. Med Eng Phys 36(8):1057–1061CrossRefPubMed
Zurück zum Zitat McCool FD, Paek D (1993) Measurements of ventilation in freely ranging subjects. Res Rep Health Eff Inst 1(59):1–17 McCool FD, Paek D (1993) Measurements of ventilation in freely ranging subjects. Res Rep Health Eff Inst 1(59):1–17
Zurück zum Zitat McCool FD, Kelly KB, Loring SH, Greaves IA, Mead J (1986) Estimates of ventilation from body surface measurements in unrestrained subjects. J Appl Physiol 61(3):1114–1119PubMed McCool FD, Kelly KB, Loring SH, Greaves IA, Mead J (1986) Estimates of ventilation from body surface measurements in unrestrained subjects. J Appl Physiol 61(3):1114–1119PubMed
Zurück zum Zitat McCool FD, Wang J, Ebi KL (2002) Tidal volume and respiratory timing derived from a portable ventilation monitor. Chest 122(2):684–691CrossRefPubMed McCool FD, Wang J, Ebi KL (2002) Tidal volume and respiratory timing derived from a portable ventilation monitor. Chest 122(2):684–691CrossRefPubMed
Zurück zum Zitat Millard RK (2002) Key to better qualitative diagnostic calibrations in respiratory inductive plethysmography. Physiol Meas 23(2):N1CrossRefPubMed Millard RK (2002) Key to better qualitative diagnostic calibrations in respiratory inductive plethysmography. Physiol Meas 23(2):N1CrossRefPubMed
Zurück zum Zitat Miller MR, Hankinson J, Brusasco V, Burgos F, Casaburi R, Coates A, Crapo R, Enright P, van der Grinten CPM, Gustafsson P, Jensen R, Johnson DC, MacIntyre N, McKay R, Navajas D, Pedersen OF, Pellegrino R, Viegi G, Wanger J (2005) Standardisation of spirometry. Eur Respir J 26(2):319–338CrossRefPubMed Miller MR, Hankinson J, Brusasco V, Burgos F, Casaburi R, Coates A, Crapo R, Enright P, van der Grinten CPM, Gustafsson P, Jensen R, Johnson DC, MacIntyre N, McKay R, Navajas D, Pedersen OF, Pellegrino R, Viegi G, Wanger J (2005) Standardisation of spirometry. Eur Respir J 26(2):319–338CrossRefPubMed
Zurück zum Zitat Miyata M, Burioka N, Sako T, Suyama H, Fukuoka Y, Tomita K, Higami S, Shimizu E (2004) A short daytime test using correlation dimension for respiratory movement in osahs. Eur Respir J 23(6):885–890CrossRefPubMed Miyata M, Burioka N, Sako T, Suyama H, Fukuoka Y, Tomita K, Higami S, Shimizu E (2004) A short daytime test using correlation dimension for respiratory movement in osahs. Eur Respir J 23(6):885–890CrossRefPubMed
Zurück zum Zitat Paek D, McCool FD (1992) Breathing patterns during varied activities. J Appl Physiol 73(3):887–893PubMed Paek D, McCool FD (1992) Breathing patterns during varied activities. J Appl Physiol 73(3):887–893PubMed
Zurück zum Zitat Petrović MD, Petrovic J, Daničić A, Vukčević M, Bojović B, Hadžievski L, Allsop T, Lloyd G, Webb DJ (2014) Non-invasive respiratory monitoring using long-period fiber grating sensors. Biomed Opt Express 5(4):1136–1144CrossRefPubMedCentralPubMed Petrović MD, Petrovic J, Daničić A, Vukčević M, Bojović B, Hadžievski L, Allsop T, Lloyd G, Webb DJ (2014) Non-invasive respiratory monitoring using long-period fiber grating sensors. Biomed Opt Express 5(4):1136–1144CrossRefPubMedCentralPubMed
Zurück zum Zitat Poole KA, Thompson JR, Hallinan HM, Beardsmore CS (2000) Respiratory inductance plethysmography in healthy infants: a comparison of three calibration methods. Eur Respir J 16(6):1084–1090CrossRefPubMed Poole KA, Thompson JR, Hallinan HM, Beardsmore CS (2000) Respiratory inductance plethysmography in healthy infants: a comparison of three calibration methods. Eur Respir J 16(6):1084–1090CrossRefPubMed
Zurück zum Zitat Raoufy MR, Hajizadeh S, Gharibzadeh S, Mani AR, Parivash E, Masjedi MR (2013) Nonlinear model for estimating respiratory volume based on thoracoabdominal breathing movements. Respirology 18(1):108–116CrossRefPubMed Raoufy MR, Hajizadeh S, Gharibzadeh S, Mani AR, Parivash E, Masjedi MR (2013) Nonlinear model for estimating respiratory volume based on thoracoabdominal breathing movements. Respirology 18(1):108–116CrossRefPubMed
Zurück zum Zitat Raoufy MR, Ghafari T, Darooei R, Nazari M, Mahdaviani SA, Eslaminejad AR, Almasnia M, Gharibzadeh S, Mani AR, Hajizadeh S (2016) Classification of asthma based on nonlinear analysis of breathing pattern. PLoS One 11(1):1–17CrossRef Raoufy MR, Ghafari T, Darooei R, Nazari M, Mahdaviani SA, Eslaminejad AR, Almasnia M, Gharibzadeh S, Mani AR, Hajizadeh S (2016) Classification of asthma based on nonlinear analysis of breathing pattern. PLoS One 11(1):1–17CrossRef
Zurück zum Zitat Retory Y, Niedzialkowski P, de Picciotto C, Bonay M, Petitjean M (2016) New respiratory inductive plethysmography (rip) method for evaluating ventilatory adaptation during mild physical activities. PLoS One 11(3):1–12CrossRef Retory Y, Niedzialkowski P, de Picciotto C, Bonay M, Petitjean M (2016) New respiratory inductive plethysmography (rip) method for evaluating ventilatory adaptation during mild physical activities. PLoS One 11(3):1–12CrossRef
Zurück zum Zitat Robertson HT, Altemeier WA, Glenny RW (2000) Physiological implications of the fractal distribution of ventilation and perfusion in the lung. Ann Biomed Eng 28(8):1028–1031CrossRefPubMed Robertson HT, Altemeier WA, Glenny RW (2000) Physiological implications of the fractal distribution of ventilation and perfusion in the lung. Ann Biomed Eng 28(8):1028–1031CrossRefPubMed
Zurück zum Zitat Sabil A (2005) Respiratory inductance plethysmography without calibration. Developments in respiratory exploration, surveillance and assistance. PhD thesis, Université Joseph-Fourier - Grenoble I Sabil A (2005) Respiratory inductance plethysmography without calibration. Developments in respiratory exploration, surveillance and assistance. PhD thesis, Université Joseph-Fourier - Grenoble I
Zurück zum Zitat Sartene R, Dartus C, Bernard JL, Mathieu M, Goldman MD (1993) Comparison of thoracoabdominal calibration methods in normal human subjects. J Appl Physiol 75(5):2142–2150PubMed Sartene R, Dartus C, Bernard JL, Mathieu M, Goldman MD (1993) Comparison of thoracoabdominal calibration methods in normal human subjects. J Appl Physiol 75(5):2142–2150PubMed
Zurück zum Zitat Saunders SW, Rath D, Hodges PW (2004) Postural and respiratory activation of the trunk muscles changes with mode and speed of locomotion. Gait Posture 20(3):280–290CrossRefPubMed Saunders SW, Rath D, Hodges PW (2004) Postural and respiratory activation of the trunk muscles changes with mode and speed of locomotion. Gait Posture 20(3):280–290CrossRefPubMed
Zurück zum Zitat Smith JC, Mead J (1986) Three degree of freedom description of movement of the human chest wall. J Appl Physiol 60(3):928–934PubMed Smith JC, Mead J (1986) Three degree of freedom description of movement of the human chest wall. J Appl Physiol 60(3):928–934PubMed
Zurück zum Zitat Stagg D, Goldman M, Davis JN (1978) Computer-aided measurement of breath volume and time components using magnetometers. J Appl Physiol 44(4):623–633PubMed Stagg D, Goldman M, Davis JN (1978) Computer-aided measurement of breath volume and time components using magnetometers. J Appl Physiol 44(4):623–633PubMed
Zurück zum Zitat Stick SM, Ellis E, Lesouf PN, Sly PD (1992) Validation of respiratory inductance plethysmography (respitrace) for the measurement of tidal breathing parameters in newborns. Pediatr Pulmonol 14(3):187–191CrossRefPubMed Stick SM, Ellis E, Lesouf PN, Sly PD (1992) Validation of respiratory inductance plethysmography (respitrace) for the measurement of tidal breathing parameters in newborns. Pediatr Pulmonol 14(3):187–191CrossRefPubMed
Zurück zum Zitat Suaste-Gmez E, Hernndez-Rivera D, Snchez-Snchez AS, Villarreal-Calva E (2014) Electrically insulated sensing of respiratory rate and heartbeat using optical fibers. Sensors 14(11):21,523CrossRef Suaste-Gmez E, Hernndez-Rivera D, Snchez-Snchez AS, Villarreal-Calva E (2014) Electrically insulated sensing of respiratory rate and heartbeat using optical fibers. Sensors 14(11):21,523CrossRef
Zurück zum Zitat Suki B, Alencar AM, Frey U, Ivanov PC, Buldyrev SV, Majumdar A, Eugene Stanley H, Dawson CA, Krenz GS, Mishima M (2003) Fluctuations, noise and scaling in the cardio-pulmonary system. Fluct Noise Lett 03(01):R1–R25CrossRef Suki B, Alencar AM, Frey U, Ivanov PC, Buldyrev SV, Majumdar A, Eugene Stanley H, Dawson CA, Krenz GS, Mishima M (2003) Fluctuations, noise and scaling in the cardio-pulmonary system. Fluct Noise Lett 03(01):R1–R25CrossRef
Zurück zum Zitat Vannucci A (2017) Can spirometry be a new tool to predict the difficult airway? J Clin Monit Comput :1–3 Vannucci A (2017) Can spirometry be a new tool to predict the difficult airway? J Clin Monit Comput :1–3
Zurück zum Zitat Wilhelm FH, Roth WT, Sackner MA (2003) The lifeshirt: an advanced system for ambulatory measurement of respiratory and cardiac function. Behav Modif 27(5):671–691CrossRefPubMed Wilhelm FH, Roth WT, Sackner MA (2003) The lifeshirt: an advanced system for ambulatory measurement of respiratory and cardiac function. Behav Modif 27(5):671–691CrossRefPubMed
Zurück zum Zitat Zhang X, Bruce EN (2000) Fractal characteristics of end-expiratory lung volume in anesthetized rats. Ann Biomed Eng 28(1):94–101CrossRefPubMed Zhang X, Bruce EN (2000) Fractal characteristics of end-expiratory lung volume in anesthetized rats. Ann Biomed Eng 28(1):94–101CrossRefPubMed
Metadaten
Titel
Estimation of respiratory volume from thoracoabdominal breathing distances: comparison of two models of machine learning
verfasst von
Rémy Dumond
Steven Gastinger
Hala Abdul Rahman
Alexis Le Faucheur
Patrice Quinton
Haitao Kang
Jacques Prioux
Publikationsdatum
13.06.2017
Verlag
Springer Berlin Heidelberg
Erschienen in
European Journal of Applied Physiology / Ausgabe 8/2017
Print ISSN: 1439-6319
Elektronische ISSN: 1439-6327
DOI
https://doi.org/10.1007/s00421-017-3630-0

Weitere Artikel der Ausgabe 8/2017

European Journal of Applied Physiology 8/2017 Zur Ausgabe