Evaluation of a new suture material (Duramesh™) by measuring suture tension in small and large bites techniques for laparotomy closure in a porcine model

verfasst von: Y. Yurtkap, F. P. J. den Hartog, W. van Weteringen, J. Jeekel, G. J. Kleinrensink, J. F. Lange

Erschienen in: Hernia | Ausgabe 6/2020

Abstract

Purpose

After closure of laparotomies, sutures may pull through tissue due to too high intra-abdominal pressure or suture tension, resulting in burst abdomen and incisional hernia. The objective of this study was to measure the suture tension in small and large bites with a new suture material.

Methods

Closure of the linea alba was performed with small bites (i.e., 5 mm between two consecutive stitches and 5 mm distance from the incision) and large bites (i.e., 10 mm × 10 mm) with Duramesh™ size 0 (2 mm) and PDS II 2-0 in 24 experiments on six porcine abdominal walls. The abdominal wall was fixated on an artificial computer-controlled insufflatable abdomen, known as the ‘AbdoMan’. A custom-made suture tension sensor was placed in the middle of the incision.

Results

The suture tension was significantly lower with the small bites technique and Duramesh™ when compared with large bites (small bites 0.12 N (IQR 0.07–0.19) vs. large bites 0.57 N (IQR 0.23–0.92), p  < 0.025). This significant difference was also found in favour of the small bites with PDS II 2-0 (p  < 0.038). No macroscopic tissue failure was seen during or after the experiments.

Conclusion

Closure of the abdominal wall with the small bites technique and Duramesh was more efficient in dividing suture tension across the incision when compared to large bites. However, suture tension compared to a conventional suture material was not significantly different, contradicting an advantage of the new suture material in the prevention of burst abdomen and incisional hernia during the acute, postoperative phase.
Hinweise

Electronic supplementary material

The online version of this article (https://​doi.​org/​10.​1007/​s10029-020-02140-7) contains supplementary material, which is available to authorized users.
Y. Yurtkap and F. P. J. den Hartog have contributed equally and should be considered as first authors.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

Abdominal wound dehiscence (burst abdomen, ‘Platzbauch’) has an incidence of up to 4% and it is a feared early complication after abdominal surgery with sequelae like evisceration, prolonged hospitalization and high mortality rates [1]. In addition, incisional hernia is a common complication after midline incisions with a 5–30% incidence and may result in pain, reduced quality of life and high healthcare costs [24]. Several suture materials and techniques for the closure of the linea alba after midline incisions have been investigated, however, there is still a need for closure techniques that can prevent incisional hernia [5]. The current recommendation, also stemming from a recent randomized controlled trial, is to use the small bites technique (i.e., 5 millimetre (mm) tissue bites and 5 mm between two sutures) with slowly absorbable suture materials for the closure of the linea alba after midline laparotomy [6]. Nevertheless, this randomized controlled trial showed that the occurrence of an incisional hernia still persists in 13% after a 1-year follow-up [6]. This result confirms that the exact biomechanical basis underlying the superiority of the small bites technique remains unknown.
In a rodent model, the dynamic change of the surgical suture tension has been investigated with the use of a customised force sensor [7]. The development of a comparable suture force or tension sensor permitted researchers and surgeons to gather data on suture tension in various tissues, suture materials, suturing patterns and closure techniques. A suture may pull through tissue due to localized pressure or tension which may cut through tissue immediately resulting in a burst abdomen or an incisional hernia after a period of time from disturbed healing by infection and/or tissue necrosis [8, 9]. Dumanian and colleagues created a novel suture of uncoated mid-weight macroporous polypropylene mesh—named Duramesh™ suturable mesh suture—to reduce the occurrence of sutures pulling through tissue and to prevent incisional hernia formation [8].
The aim of this study was therefore to measure suture tension using the small bites technique and the newly developed Duramesh™ size 0 (2 mm) also in comparison with the large bites technique, by using an implantable suture tension sensor which was developed specifically for these experiments. Furthermore, the small and large bites techniques were compared with a conventional suture material, i.e., PDS II 2-0, as a control suture material. All experiments were performed in an ex vivo porcine abdominal wall using the artificial ‘AbdoMan’.

Methods

Suture tension sensor

An implantable suture tension sensor was developed using a Force Sensing Resistor (Interlink Electronics FSR 400, Interlink Electronics, Westlake Village, CA, USA) with an actuation force of approximately 0.2–20 N [10]. A three-dimensional (3D) model was developed for the enclosure of the suture tension sensor (Fig. 1a). The tension generated by the suture in the actuator notch is translated downward onto a circular surface, precisely and evenly pressing down on a force sensor within the suture tension sensor (Fig. 1b, c). An analog-to-digital converter (ADC) and an Arduino Uno controller (Arduino AG, Somerville, MA, USA) were used to read the raw output from the suture tension sensor. A custom-made program was written to create a live graph of the tension sensor data in Newtons (N).

Measuring model

The ‘AbdoMan’ was developed as an artificial simulation of the human abdominal wall by taking the muscle contractions and intra-abdominal pressure into account [11]. In these current experiments, only the intra-abdominal pressure was considered. A 3500 ml air-filled collecting bag was placed on a three-dimensional printed part in the shape of an abdominal wall. A laparoscopic insufflator (Olympus UHI-3 High Flow Insufflator, Olympus Corporation, Shinjuku, Japan) was used to apply insufflation pressures up to 20 millimetres of mercury (mmHg). After sensor placement and prior to insufflation, a baseline suture tension was measured over the course of one minute. Validation of the suture tension sensor was performed before these experiments, by varying the force applied to the suture tension sensor in a controlled manner, verifying whether the suture tension sensor would be able to correctly detect and measure these variations. The measured suture tension is relative to the baseline tension in the closed incision. The measurements are a derivative of the actual tension within the suture.

Duramesh™

Duramesh™ is a novel suturing concept, based on the principles of meshes, used in hernia repair, while providing the precision and flexibility of a suture [12]. It is a non-resorbable suture, made of polypropylene. The three-dimensional macroporous structure has a larger surface than standard sutures and it has been shown to stimulate better tissue integration in an in vivo porcine model [12].

Experimental set-up

Six porcine abdominal walls of female Yorkshire-Landrace pigs with comparable dimensions, ranging in weight from 30 to 40 kilograms (kg) were explanted directly after euthanasia and frozen at −20 Celcius (°C). Twenty-four hours prior to the experiments, the abdominal wall was thawed [13]. Two midline incisions of 5 cm each (i.e., cranial and caudal) were made through all layers of the abdominal wall. This number of midline incisions and their length was chosen, because this length would be the longest possible length compatible with all specimens. The abdominal wall was inversely placed with the peritoneum upwards and fixated onto the’AbdoMan’ (Fig. 2). The linea alba was closed with continuous sutures including all layers of the abdominal wall, including the peritoneum. Closure was performed using the small bites (i.e., 5 mm between two consecutive stitches and 5 mm distance from the incision) and the large bites (i.e., 10 mm between two consecutive stitches and 10 mm distance from the incision) techniques with Duramesh™ size 0, 2 mm (Duramesh™ Suturable Mesh, Inc., Dorado, Puerto Rico, USA) and PDS II 2-0 (Ethicon, Somerville, NJ, USA). Locations (i.e., cranial and caudal) were switched for small and large bites to randomize for location. Each experiment was repeated twice, using the existing incisions. The suture tension sensor was placed in the middle of the incision. All experiments were performed by a single researcher. Lastly, the collecting bag was insufflated to 20 mmHg for a duration of 30 minutes (min).
Finally, the suture tension was compared between:
1.
Small bites with Duramesh™ size 0 versus large bites with Duramesh™ size 0 (N = 12).
 
2.
Small bites with PDS II 2-0 versus large bites with PDS II 2-0 (N = 12).
 
3.
Small bites versus large bites with both materials.
 
4.
Small bites with PDS II 2-0 versus small bites with Duramesh™ size 0.
 
5.
Large bites with PDS II 2-0 versus large bites with Duramesh™ size 0.
 

Data analyses

Results are presented as median differences and interquartile ranges in suture tension. Statistical significance was assessed using a Wilcoxon rank-sum test for all samples comparing two different modalities (i.e., small bites, large bites, Duramesh™ size 0, and PDS II 2-0) after 25 min. p values lower than 0.05 were considered statistically significant. Python for Windows, version 3.5.1. (Python Software Foundation, Beaverton, USA) was used to perform all statistical analyses.

Results

In total, 24 experiments were performed. No macroscopic tissue failure was visible during or after the experiments. Median suture tension was calculated for one point in time; i.e., at 25 min from the start of insufflation, when the suture tension had reached a plateau in all experiments. When considering the two suture materials individually, each showed a significant difference in suture tension between small and large bites, in favour of the small bites. Regarding Duramesh™ size 0: small bites 0.12 N (IQR 0.07–0.19) versus large bites 0.57 N (IQR 0.23–0.92), p  < 0.025 (Fig. 3). Regarding PDS II 2-0: small bites 0.15 N (IQR 0.05–0.31) versus large bites 0.56 N (IQR 0.37–0.98), p < 0.038 (Fig. 4).
Irrespective of the suture material used, the tension in the sutures was significantly lower when the linea alba had been closed with small bites when compared to the tension in the sutures when large bites had been applied (small bites 0.14 N (IQR 0.06–0.20) versus large bites 0.56 N (IQR 0.31–0.98), p < 0.0015, Fig. 5).
When only considering small bites, there was no significant difference in suture tension between Duramesh™ size 0 and PDS II 2-0 (Duramesh™ size 0 0.12 N (IQR 0.07–0.19) versus PDS II 2-0 0.15 N (IQR 0.05–0.31), p > 0.05, Supplemental Fig. 1). Similarly, when only considering large bites, there was no significant difference in suture tension between Duramesh™ size 0 and PDS II 2-0 (Duramesh™ size 0 0.57 N (IQR 0.23–0.92) versus PDS II 2-0 0.56 N (IQR 0.37–0.98), p > 0.05, Supplemental Fig. 2).

Discussion

In this present study, small bites and the use of Duramesh™ resulted in a significantly lower suture tension when compared to the tension in the sutures in large bites in a porcine abdominal wall. The suture tension was measured with a custom-made suture tension sensor in an experimental set-up using the artificial ‘AbdoMan’. This significant difference was also found when the same comparison was made with the use of PDS II 2-0 as a control suture material. These findings were analogous to findings from a recent clinical study [6]. The superiority of the small bites technique is not limited to PDS II 2-0 sutures, but also holds for suture materials with an elaborate three-dimensional structure, such as the Duramesh™ size 0 and perhaps for other types of suture materials. However, in this present experiment Duramesh™ size 0 was neither superior, nor inferior, compared to PDS II 2-0. This finding makes Duramesh™ a viable option in choosing suture materials for abdominal wall closure. Nevertheless, most previous experiments with Duramesh™ have revolved around linearly pulling it until either tissue or suture failure [12, 14]. In this experimental set-up, Duramesh™ was tested to much weaker forces i.e., 20 mmHg. Simulating pulling to failure in this set-up would involve raising the intra-abdominal pressure to (much) higher levels than 20 mmHg, perhaps ranging in the hundreds of mmHg. Therefore, a tensile test would be more suitable for this kind of experiments to demonstrate a difference. When being pulled, the Duramesh™ size 0 stretched and flattened like a ribbon, which may be helpful in dividing the suture tension across the wound. When the Duramesh™ size 0 was pulled through the tissue, the structure of the suture was occasionally damaged. This damage may have compromised the integrity of its shape, thereby impairing its mechanism of action. That being said, the Duramesh™ size 0 or PDS II 2-0 never broke completely. It should be noted that only size 0 (2 mm) of the Duramesh™ was tested in these present experiments. The expectation is that the superiority of Duramesh™ will be clearer in an in vivo model, where tissue integration can be measured as well.
The relation between suture material and the development of burst abdomen has not been studied extensively. Van Ramshorst et al. found that failure of the knot was a significant cause in addition to other causes like ileus [1]. On the contrary, the effects of different suture materials, suture configurations and suture length to wound length ratio on the occurrence of incisional hernia have been extensively studied [5, 15]. Cooney et al. found the best performing bite separation and bite width to be 5 and 16 mm, respectively, in a biomechanical abdominal wall model [16]. As they also state, this is partly in agreement with the findings of the STITCH trial. However, they suggest that perhaps a small bite separation should be combined with a large bite depth rather than a small bite depth like the 5 mm used in the STITCH trial. Comparing these two modalities should be the next step in this present experimental setup. However, the optimal suture tension has not been studied largely [7]. The tension in a suture is the composite result of the type or resistance of tissue, the suture material used, the force applied by the surgeon during knotting, and the bite width and separation [7]. Proper closure of the abdominal wall involves the close approximation of tissue edges with sutures. If the sutures are too loose, however, the wound edges cannot be properly approximated and there will exist a risk for impaired healing and wound dehiscence. This theoretically would result in an increased risk for incisional hernia formation [1]. Another reason for insufficient suture tension could be the phenomenon of creep, in which the suture will be irreversibly elongated over time as a result of a continuous pulling force [10]. Inadequate abdominal fascial closure may also be seen in cases where the tension in the sutures is too high. The suture may cut through tissue and cause additional tissue damage, tissue necrosis or an incisional hernia [1]. This implies that the relation between suture tension and outcome is parabolic, allowing for the definition of a possibly optimal suture tension [17]. Nonetheless, it is not easy to obtain the ideal suture tension since this is subject to inter- and intra-surgeon variability [18]. It would be helpful to have a device attached to the suture needle or the suture material continuously measuring suture tension so that the surgeon would be able to apply the same suture tension with every knot or throw. While there is no currently available method to determine suture tension during suturing, the pore size of the Duramesh™ was macroscopically changed with higher tension, giving the surgeon feedback while suturing.
As almost in every in vitro study, this study also has limitations. One limitation in this study is the use of porcine abdominal walls instead of human abdominal walls, which were not available. However, in a recently published study, porcine tissue was demonstrated to be an appropriate surrogate for examining the human abdominal wall when it comes to the linea alba [19]. Another limitation was that the suture length to wound length ratio of at least 4:1 was established prior to the experiment. However, with the small bites technique, twice as many suture loops were placed than with the large bites technique. The present ex vivo experiments can be considered an acute postoperative model rather than a wound healing model. As a consequence it can be concluded that the new Duramesh™ size 0 suture seems to behave similar to a conventional suture like PDS II 2-0, that the advantages of small bite closure of the linea alba also apply to it, but that it cannot be expected to prevent the early development of burst abdomen and incisional hernia in a better way. One could propose that the three-dimensional, macroporous structure of the Duramesh™ would provide for a more profound tissue integration, allowing the tissue to grow through its individual threads and completely envelop the suture. This could hypothetically strengthen the wound healing and help prevent incisional hernia, something which has already been shown in in vivo experiments [8]. Finding a way to simulate wound healing, such as in animal models, would allow to focus on incisional hernia formation at a later point in time—after closure of the abdominal wall and during the healing process. In such an experimental setup the Duramesh™ would be expected to be more efficient when compared with conventional suture materials. In the future, this experimental setup and the suture tension sensor might be used for experiments with other suture materials and configurations.

Conclusion

The suture tension with the small bites technique and the use of Duramesh™ size 0 was significantly lower when compared with large bites in this model. Additionally, macroscopic tissue failure was not seen in either suture material during or after the experiment. Further research should be conducted to find out whether these findings are also valid in different stages of wound healing and in abdominal walls of different origins, shapes and sizes, as well as with the use of other types of sutures or suturing techniques.

Acknowledgements

We would like to thank Gregory Dumanian, M.D. for providing us Duramesh free of charge. Gregory Dumanian was not involved in study design, data analysis or construction of the manuscript. Additionally, we would like to thank J. Hagoort for proofreading the article.

Compliance with ethical standards

Conflict of interest

Y. Yurtkap, F.P.J. den Hartog, W. van Weteringen, J. Jeekel, G.J. Kleinrensink and J.F. Lange declare no conflict of interest related to the submitted work.

Ethical approval

For this type of study ethical approval is not required.

Human and animal rights

Animals had passed away prior to being included in this study (after ethical use in other experiments, not involved in this present study).
For this type of study formal consent is not required.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

Die Chirurgie

Print-Titel

Das Abo mit mehr Tiefe

Mit der Zeitschrift Die Chirurgie erhalten Sie zusätzlich Online-Zugriff auf weitere 43 chirurgische Fachzeitschriften, CME-Fortbildungen, Webinare, Vorbereitungskursen zur Facharztprüfung und die digitale Enzyklopädie e.Medpedia.

Bis 30. April 2024 bestellen und im ersten Jahr nur 199 € zahlen!

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Anhänge

Electronic supplementary material

Below is the link to the electronic supplementary material.
Literatur
1.
Zurück zum Zitat van Ramshorst GH et al (2010) Abdominal wound dehiscence in adults: development and validation of a risk model. World J Surg 34(1):20–27CrossRef van Ramshorst GH et al (2010) Abdominal wound dehiscence in adults: development and validation of a risk model. World J Surg 34(1):20–27CrossRef
2.
Zurück zum Zitat Jairam AP et al (2017) Prevention of incisional hernia with prophylactic onlay and sublay mesh reinforcement versus primary suture only in midline laparotomies (PRIMA): 2-year follow-up of a multicentre, double-blind, randomised controlled trial. Lancet 390(10094):567–576CrossRef Jairam AP et al (2017) Prevention of incisional hernia with prophylactic onlay and sublay mesh reinforcement versus primary suture only in midline laparotomies (PRIMA): 2-year follow-up of a multicentre, double-blind, randomised controlled trial. Lancet 390(10094):567–576CrossRef
3.
Zurück zum Zitat Poulose BK et al (2012) Epidemiology and cost of ventral hernia repair: making the case for hernia research. Hernia 16(2):179–183CrossRef Poulose BK et al (2012) Epidemiology and cost of ventral hernia repair: making the case for hernia research. Hernia 16(2):179–183CrossRef
4.
Zurück zum Zitat van Ramshorst GH et al (2012) Impact of incisional hernia on health-related quality of life and body image: a prospective cohort study. Am J Surg 204(2):144–150CrossRef van Ramshorst GH et al (2012) Impact of incisional hernia on health-related quality of life and body image: a prospective cohort study. Am J Surg 204(2):144–150CrossRef
5.
Zurück zum Zitat Henriksen NA et al (2018) Meta-analysis on materials and techniques for laparotomy closure: the match review. World J Surg 42(6):1666–1678CrossRef Henriksen NA et al (2018) Meta-analysis on materials and techniques for laparotomy closure: the match review. World J Surg 42(6):1666–1678CrossRef
6.
Zurück zum Zitat Deerenberg EB et al (2015) Small bites versus large bites for closure of abdominal midline incisions (STITCH): a double-blind, multicentre, randomised controlled trial. Lancet 386(10000):1254–1260CrossRef Deerenberg EB et al (2015) Small bites versus large bites for closure of abdominal midline incisions (STITCH): a double-blind, multicentre, randomised controlled trial. Lancet 386(10000):1254–1260CrossRef
7.
Zurück zum Zitat Klink CD et al (2011) Tension of knotted surgical sutures shows tissue specific rapid loss in a rodent model. BMC Surg 11:36CrossRef Klink CD et al (2011) Tension of knotted surgical sutures shows tissue specific rapid loss in a rodent model. BMC Surg 11:36CrossRef
8.
Zurück zum Zitat Souza JM et al (2015) In vivo evaluation of a novel mesh suture design for abdominal wall closure. Plast Reconstr Surg 135(2):322e–e330CrossRef Souza JM et al (2015) In vivo evaluation of a novel mesh suture design for abdominal wall closure. Plast Reconstr Surg 135(2):322e–e330CrossRef
9.
Zurück zum Zitat Vogels RR et al (2017) Biocompatibility and biomechanical analysis of elastic TPU threads as new suture material. J Biomed Mater Res B Appl Biomater 105(1):99–106CrossRef Vogels RR et al (2017) Biocompatibility and biomechanical analysis of elastic TPU threads as new suture material. J Biomed Mater Res B Appl Biomater 105(1):99–106CrossRef
10.
Zurück zum Zitat Nout E et al (2007) Creep behavior of commonly used suture materials in abdominal wall surgery. J Surg Res 138(1):51–55CrossRef Nout E et al (2007) Creep behavior of commonly used suture materials in abdominal wall surgery. J Surg Res 138(1):51–55CrossRef
11.
Zurück zum Zitat Kroese LF et al (2017) The 'AbdoMAN': an artificial abdominal wall simulator for biomechanical studies on laparotomy closure techniques. Hernia 21(5):783–791CrossRef Kroese LF et al (2017) The 'AbdoMAN': an artificial abdominal wall simulator for biomechanical studies on laparotomy closure techniques. Hernia 21(5):783–791CrossRef
12.
Zurück zum Zitat Dumanian GA, Tulaimat A, Dumanian ZP (2015) Experimental study of the characteristics of a novel mesh suture. Br J Surg 102(10):1285–1292CrossRef Dumanian GA, Tulaimat A, Dumanian ZP (2015) Experimental study of the characteristics of a novel mesh suture. Br J Surg 102(10):1285–1292CrossRef
13.
Zurück zum Zitat Chan RW, Titze IR (2003) Effect of postmortem changes and freezing on the viscoelastic properties of vocal fold tissues. Ann Biomed Eng 31(4):482–491CrossRef Chan RW, Titze IR (2003) Effect of postmortem changes and freezing on the viscoelastic properties of vocal fold tissues. Ann Biomed Eng 31(4):482–491CrossRef
14.
Zurück zum Zitat Zhang T et al (2018) Rotator cuff repair with a novel mesh suture: an ex vivo assessment of mechanical properties. J Orthop Res 36(3):987–992PubMed Zhang T et al (2018) Rotator cuff repair with a novel mesh suture: an ex vivo assessment of mechanical properties. J Orthop Res 36(3):987–992PubMed
15.
Zurück zum Zitat Israelsson LA, Jonsson T, Knutsson A (1996) Suture technique and wound healing in midline laparotomy incisions. Eur J Surg 162(8):605–609PubMed Israelsson LA, Jonsson T, Knutsson A (1996) Suture technique and wound healing in midline laparotomy incisions. Eur J Surg 162(8):605–609PubMed
16.
Zurück zum Zitat Cooney GM et al (2018) Optimized wound closure using a biomechanical abdominal model. Br J Surg 105(4):395–400CrossRef Cooney GM et al (2018) Optimized wound closure using a biomechanical abdominal model. Br J Surg 105(4):395–400CrossRef
17.
Zurück zum Zitat Mioton LM, Dumanian GA (2018) Theoretic and evidence-based laparotomy closure with sutures and meshes. Plast Reconstr Surg 142(3 Suppl):117S–124SCrossRef Mioton LM, Dumanian GA (2018) Theoretic and evidence-based laparotomy closure with sutures and meshes. Plast Reconstr Surg 142(3 Suppl):117S–124SCrossRef
18.
Zurück zum Zitat Fischer L et al (2010) Variability of surgical knot tying techniques: do we need to standardize? Langenbecks Arch Surg 395(4):445–450CrossRef Fischer L et al (2010) Variability of surgical knot tying techniques: do we need to standardize? Langenbecks Arch Surg 395(4):445–450CrossRef
19.
Zurück zum Zitat Cooney GM et al (2017) The suture pullout characteristics of human and porcine linea alba. J Mech Behav Biomed Mater 68:103–114CrossRef Cooney GM et al (2017) The suture pullout characteristics of human and porcine linea alba. J Mech Behav Biomed Mater 68:103–114CrossRef
Metadaten
Titel
Evaluation of a new suture material (Duramesh™) by measuring suture tension in small and large bites techniques for laparotomy closure in a porcine model
verfasst von
Y. Yurtkap
F. P. J. den Hartog
W. van Weteringen
J. Jeekel
G. J. Kleinrensink
J. F. Lange
Publikationsdatum
21.02.2020
Verlag
Springer Paris
Erschienen in
Hernia / Ausgabe 6/2020
Print ISSN: 1265-4906
Elektronische ISSN: 1248-9204
DOI
https://doi.org/10.1007/s10029-020-02140-7

Weitere Artikel der Ausgabe 6/2020

Hernia 6/2020 Zur Ausgabe

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.