01.06.2015 | Ausgabe 6/2015

Surgical Endoscopy 6/2015

Evaluation of absorbable mesh fixation devices at various deployment angles

Zeitschrift:
Surgical Endoscopy > Ausgabe 6/2015
Autoren:
Ahmed M. Zihni, Jaime A. Cavallo, Dominic M. Thompson Jr., Nabeel H. Chowdhury, Margaret M. Frisella, Brent D. Matthews, Corey R. Deeken

Abstract

Background

Hernia repair failure may occur due to suboptimal mesh fixation by mechanical constructs before mesh integration. Construct design and acute penetration angle may alter mesh-tissue fixation strength. We compared acute fixation strengths of absorbable fixation devices at various deployment angles, directions of loading, and construct orientations.

Methods

Porcine abdominal walls were sectioned. Constructs were deployed at 30°, 45°, 60°, and 90° angles to fix mesh to the tissue specimens. Lap-shear testing was performed in upward, downward, and lateral directions in relation to the abdominal wall cranial–caudal axis to evaluate fixation. Absorbatack™ (AT), SorbaFix™ (SF), and SecureStrap™ in vertical (SSV) and horizontal (SSH) orientations in relation to the abdominal wall cranial–caudal axis were tested. Ten tests were performed for each combination of device, angle, and loading direction. Failure types and strength data were recorded. ANOVA with Tukey–Kramer adjustments for multiple comparisons and χ 2 tests were performed as appropriate (p < 0.05 considered significant).

Results

At 30°, SSH and SSV had greater fixation strengths (12.95, 12.98 N, respectively) than SF (5.70 N; p = 0.0057, p = 0.0053, respectively). At 45°, mean fixation strength of SSH was significantly greater than SF (18.14, 11.40 N; p = 0.0002). No differences in strength were identified at 60° or 90°. No differences in strength were noted between SSV and SSH with different directions of loading. No differences were noted between SS and AT at any angle. Immediate failure was associated with SF (p < 0.0001) and the 30° tacking angle (p < 0.01).

Conclusions

Mesh-tissue fixation was stronger at acute deployment angles with SS compared to SF constructs. The 30° angle and the SF device were associated with increased immediate failures. Varying construct and loading direction did not generate statistically significant differences in the fixation strength of absorbable fixation devices in this study.

Bitte loggen Sie sich ein, um Zugang zu diesem Inhalt zu erhalten

★ PREMIUM-INHALT
e.Med Interdisziplinär

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de. Zusätzlich können Sie eine Zeitschrift Ihrer Wahl in gedruckter Form beziehen – ohne Aufpreis.

Jetzt e.Med zum Sonderpreis bestellen!

Weitere Produktempfehlungen anzeigen
Literatur
Über diesen Artikel

Weitere Artikel der Ausgabe 6/2015

Surgical Endoscopy 6/2015 Zur Ausgabe
  1. Sie können e.Med Chirurgie 14 Tage kostenlos testen (keine Print-Zeitschrift enthalten). Der Test läuft automatisch und formlos aus. Es kann nur einmal getestet werden.


 

Neu im Fachgebiet Chirurgie

Mail Icon II Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Chirurgie und bleiben Sie gut informiert – ganz bequem per eMail.

Bildnachweise